Добавил:
Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Gary S. Coyne - A Practical Guide to Materials, Equipment, and Technique.pdf
Скачиваний:
142
Добавлен:
15.08.2013
Размер:
8.47 Mб
Скачать

Volume 2.3

107

wall about a centimeter above the calibration line, being careful not to splash. The tip may touch the wall.

3.When the level is just below the calibration line, stop for a few minutes to let the liquid drain from the walls of the container.

4.Finish the fill using Steps 2 and 3 again. If you overshoot the calibration line, a pasture pipette can be used to draw out excess fluid. The surface tension in the tip of the pipette should be sufficient to draw out the fluid.

For emptying a to deliver cylinder (Style 1 only), the following procedure should be followed:

1.Slowly incline the cylinder to provide a steady stream of liquid from the spout. Be careful not to splash.

2.Continue inclining the cylinder until it is vertical and hold for about half a minute.

3.Touch the drop at the tip of the spout to the wall of the receiving container.

4.Remove the cylinder horizontally from the wall of the receiving container with no vertical motion.

2.3.12Pipettes

There are three types of volumetric pipettes: volumetric transfer, measuring, and serological. The differences are based on whether the volume within the pipette is subdivided and if the volume in the tip is included in the calibration (see Fig. 2.19).

The serological pipette is used to dispense varying volumes of liquid similar to the burette. Like a burette, the serological pipette is calibrated volume does not include the tip region. Volumetric pipettes are designed to dispense one volume of liquid, whereas the measuring pipette is calibrated to dispense varying volumes. Both of these pipettes are designed to include the tip region in their entire dispensed volumes.

There are two different methods for completely draining a volumetric transfer or measuring pipette, and each method requires a different pipette design. One method is to leave the tip against the side of the receiving vessel and let the material drain into the vessel. This method leaves a small amount of solution remaining at the tip. The difference between this remaining portion and what has been transferred has been accounted for in the pipette's calibration. The other method of pipette draining requires the user to blow out the remaining liquid in the pipette; after all the liquid has drained out naturally, an extra burst of air is applied to the end of the pipette to expel any remaining fluid out of the tip. All blow-out

108

Measurement

Serological

pipette

Measuring

pipette

Volumetric transfer pipette

Fig. 2.19 The three types of pipettes.

pipettes are identified by either (a) an opaque band one-quarter of an inch wide or

(b) two bands one-quarter inch apart at the end of the pipette.

There are pros and cons to both types of pipettes. The decision to use one or the other depends on the reliability, and repeatability, of your laboratory technique as well as the nature of the liquid you are pipetting. All non-blow-out pipettes are calibrated with water. Thus if your liquid has different surface tension or viscosity characteristics than water, your measurements will not be accurate. On the other hand, not everyone will exert an equal amount of blow-out force. Thus different people may deliver different volumes using the same equipment.

Pipettes often have color-coding bands on their ends to help identify them. Although the colors are designated by the ASTM, they do not necessarily specify a specific volume. Their volumes do not necessarily correspond to any other pipette design. In fact, a given color may identify two volumes for any given type of pipette: however, pipettes with the same color band are always distinctively different in volume.

Traditionally, pipettes were filled by sucking fluid into the tube with the mouth. This procedure, however, carries the risk of getting chemicals in the mouth. To protect the user, some pipette ends are designed to receive a cotton plug (see Fig. 2.20).

The cotton plug allows gases past, but inhibits the flow of liquids. Thus, if a user sucked too hard, the liquid would clog the cotton plug and prevent the liquid from

Jl

Note

constriction

Regular

Cotton Plug

end

end

Fig. 2.20 Regular and cotton plug ends for pipettes.

Volume 2.3

109

reaching the user's mouth. Although people still suck liquids into pipettes, the use of pipette fillers, or even rubber bulbs, are strongly recommended.

Volumetric Transfer. These pipettes are used to deliver a (single) specified volume. They are characterized by a bulb mid-span on the tube (see Fig. 2.19). The bulb is used to achieve greater capacity in the pipette and maintain the general length throughout the different volumes.

Volumetric pipettes are used solely to deliver in both drain and blow-out models. They are calibrated in both Class A and Class B tolerances based on guidelines established in the ASTM Designation E 969 - 83. Tolerances are provided in Table 2.16.

The color-coding band(s) on the end of the volumetric pipette are used for quick identification purposes only. The repetition of the colors is sufficiently separated by the size of the pipette to not confuse (for instance) the 10-mL and 50-mL sizes.

To fill a volumetric measuring pipette, draw the solution to just above the volumetric level, then let the solution fall to the calibration mark. Remove the pipette

Table 2.16 Requirements for Volumetric Transfer Pipettes*

 

ClassA

 

Class B

 

 

 

Nominal Capacity (mL)

Capacity Tolerance (mL)

Minimum Outflow Time

(sec/

Capacity Tolerance (mL)

Minimum

Outflow Time

(sec)*

Color-Coded Band

0.5

+0.006

 

5

+0.012

 

3

 

Black (2)

1

±0.006

 

10

±0.012

 

3

 

Blue

2

±0.006

 

10

±0.012

 

3

 

Orange

3

+0.01

 

10

+0.02

 

5

 

Black

4

±0.01

 

10

±0.02

 

5

 

Red (2)

5

+0.01

 

10

±0.02

 

8

 

White

10

±0.02

 

15

±0.04

 

8

 

Red

15

±0.03

 

25

+0.06

 

10

 

Green

20

±0.03

 

25

+0.06

 

10

 

Yellow

25

±0.03

 

25

+0.06

 

15

 

Blue

50

±0.05

 

25

+0.10

 

15

 

Red

100

±0.08

 

30

+0.16

 

20

 

Yellow

" From the ASTM document E 969, Table 1. Omitted were columns titled "Length of Delivery Tube, mm (Min and Max)," "Inside Diameter at Capacity Mark, mm (Min and Max)," and "Maximum Distance Between Bulb and Graduation Mark, mm." Reprinted with permission

* Maximum outflow time for A and B shall be 60 seconds.

110

 

 

 

 

 

 

 

 

 

 

Measurement

 

 

Table 2.17 Requirements for Measuring Pipettes"

 

 

Capacity

 

 

 

 

 

Outflow time (sec)

 

 

 

Accuracy

Graduations mL

Class A

Class B

Color of codingbands

Capacity

(±tolerance, mL)

 

 

 

 

%

 

§

«

s

J5

 

 

 

i

 

<

 

111

.5

s

S

a

 

B

 

 

 

U

 

S*S

|

 

S

§

 

 

 

2

 

 

 

 

 

 

 

 

s

 

 

 

 

 

 

5 l

 

-

 

 

0.1

-

0.005

0.01

 

-

0.5

10

White

 

0.01

0.1B

-

0.005

0.001

 

0.1

 

-

-

0.5

10

2-Green

0.2

-

0.008

0.01

 

0.02

-

-

0.5

8

Black

0.2*

-

0.008

0.001

 

0.02

-

-

0.5

8

2-Blue

0.5*

-

0.01

0.01

 

0.10

-

-

2

6

2-Yellow

1.0

0.01

0.02

0.10

 

0.10

20

60

3

22

Red

1.0

0.01

0.02

0.01

 

0.10

20

60

3

22

Yellow

2.0

0.01

0.02

0.10

 

0.20

20

60

3

22

Green

2.0*

-

0.02

0.01

 

0.20

-

-

3

22

2-White

5.0

0.02

0.04

0.10

 

1.0

 

30

60

8

20

Blue

10.0

0.03

0.06

0.10

 

1.0

 

30

60

8

30

Orange

25.0c

0.05

0.10

0.10

 

1.0

 

40

70

15

35

White

50.0c

-

0.16

0.20

 

2.0

 

-

-

20

40

Black

" From the ASTM document E 1293, Table 1. Omitted was the column titled "Outside Diameter of Grad. Portion min., mm." Reprinted with permission.

* Style 2 only.

c Style 1 only.

from where you drew the fluid, and wipe the tip of the pipette with a laboratory tissue to remove any solution on the outside of the tip so it is not included with the dispensed liquid from the inside of the pipette. When draining a volumetric pipette, let the tip touch the side of the receiving container and let the fluid flow. After emptying the pipette, count to two (to allow for any remaining fluid to flow to the bottom), and remove the tip sideways away from the receiving wall. Do not remove the tip with an upward or downward motion.

If the end of the pipette indicates the pipette is of blow-out design, gently (but firmly) provide (by lips or by the pipette filler) extra air pressure to "blow out" the last drop of liquid. Do not maintain a long continuous blow, especially if you are using the mouth to blow out the excess because you may contaminate your solution.

Measuring and Serological Pipettes. These pipettes have graduations along their sides. The difference is that the graduations on measuring pipettes stop before reaching the taper of the tip while the stated volumes of serological

Volume 2.3

111

pipettes include the contents of the tip (see Fig. 2.19). Measuring pipettes should never be drained or blown out when delivering solution because the extra volume in the tip is not part of the pipette's calculated volume. Some serological pipettes are blow-outs, and others are not. You need to examine the end for the one-quar- ter-inch opaque mark

Measuring pipettes come in both Class A and Class B and are available in two styles: Style 1 is a standard taper tip, and Style 2 is a long taper tip (Class B only). Serological pipettes are only made to Class B tolerances and have no special styles. Volumes, tolerances, and other data of measuring pipettes are provided in Table 2.17. Information for serological pipettes is provided in Table 2.18. All calibrations were made at 20°C.

Class A pipettes have slower outflow times than Class B pipettes. This slowness provides the user with more reaction time to control liquid flow and thereby

Table 2.18 Requirements for Serological Pipettes"

 

 

 

Graduations (mL)

Capacity (mL)

Capacity Tolerance (imL)

Least Value

Main Numbered

Interval Graduated 0.0 to at least

Style ][&II Outl low Tirne

_a

Style III Nominal Tip Opening (mm)

ColorCoding Band

0.1

0.005

0.01

0.01

0.09

0.5

3

 

White

0.1*

0.005

0.01

0.01

0.09

0.5

3

 

2 Green

0.2

0.008

0.001

0.02

0.19

0.5

3

 

Black

0.2*

0.008

0.001

0.01

0.19

0.5

3

 

2 Blue

0.25c

0.008

0.0125

0.05

0.25

0.5

3

 

 

0.5

0.01

0.01

0.05

0.45

0.5

3

 

2 Black

0.5

0.01

0.05

0.05

0.45

0.5

3

 

2 Yellow

0.60*

0.01

0.15

0.15

0.45

0.5

3

 

 

1.0

0.02

0.01

0.1

0.95

1

5

 

Yellow

1.0

0.02

0.1

0.1

0.9

1

5

2.0

Red

2.0

0.02

0.01

0.02

1.9

1

5

2.5

2 White

2.0

0.02

0.1

0.2

1.9

1

5

2.5

Green

5.0

0.04

0.1

1.0

4.5

3

10

3.0

Blue

10.0

0.06

0.1

1.0

9.5

5

15

3.0

Orange

25.0

0.10

0.1

1.0

23.0

5

15

3.0

White

a From the ASTM document E 1044, Table 1. Omitted was the column titled "Outside Diameter of Graduated Portion, min., mm." Reprinted with permission.

* Kahn serological

pipettes, calibrated to tip.

c Kahn serological

pipe, calibrated to base.

112

 

Measurement

Table 2.19 Tolerances of Disposable (Serological) Pipettes* b

Capacity (cm3)

Glass

Plastic

0.1

+7%

0.2

+6%

0.5

±3%

+3%

1.0

±3%

+3%

2.0

±3%

+3%

5.0

+ 3%

+3%

10.0

+3%

+3%

25.0

+3%

aFrom ASTM Tables E 714 (Standard Specification for Disposable Glass Serological Pipettes), Table 2 (omitted was the column titled "Coefficient of Variation"), and E 934 (Standard Specification for Serological Pipette, Disposable Plastic), Table 2 [omitted was the column titled "Coefficient of Variation (Equal to or less than 1.5%)"]. Reprinted with permission.

bApplies to the stated capacity, not individual calibrations.

achieve better accuracy. It also allows the fluid draining from the walls to keep up with the fluid being dispensed from the tip. Otherwise you may dispense what you thought was 5 mL of liquid from a pipette, but after the liquid in the pipette settled, you discover only 4.9 mL of solution had been dispensed.

To fill a measuring or serological pipette, draw the solution to just above the volumetric level, then let the solution fall to the calibration mark. Remove the pipette from where you drew the fluid, and wipe the tip with a laboratory tissue to remove any excess solution from the outside of the pipette so that it is not included with the calibrated liquid.

When dispensing fluid from a measuring pipette, let the tip touch the side of the receiving container and let the fluid flow. If you are dispensing the liquid by hand, you need to control the flow rate by placing your thumb on the end of the pipette. However, never let your thumb wander away from the end of the pipette because you will need it to stop the fluid flow. If you completely drain the pipette, count to two (to allow for any remaining fluid to flow to the bottom), and remove the tip sideways away from the receiving wall. Do not remove the tip with an upward or downward motion.

There are three different styles of serological pipettes. Style I has a standard end piece. Style II has an end that can receive a cotton plug (see Fig. 2.20). Style III has the same type of end as Style II, but also has a larger tip opening to speed the emptying process.

Pipettes are also made as disposable serological pipettes. They are just as easy to use as regular pipettes and are typically made out of plastics or soda-lime glass. They have very low-quality standards, and their calibrations are not required to have the same permanency as regular pipettes. Thus, if you try to wash them, the

Соседние файлы в предмете Химия