Добавил:
Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Wiley - Encyclopedia of Medical Devices and Instrumentation - Vol. 5.pdf
Скачиваний:
278
Добавлен:
10.08.2013
Размер:
15 Мб
Скачать
☆

418 PROSTATE SEED IMPLANTS

PET/CT scanner for clinical oncology. J Nucl Med 2000; 41(8):1369–1379.

29.Reba RC. PET and SPECT: Opportunities and challenges for psychiatry. J Clin Psychiatry 1993;54:26–32.

30.Fahey FH. Data acquisition in PET imaging. J Nucl Med 2002; 30(2):39–49.

31.Castiglioni I, Cremonesi O, Gilardi MC, Bettinardi V, Rizzo G, Savi A, Bellotti E, Fazio F. Scatter correction techniques in 3D PET: A Monte Carlo evaluation. IEEE Trans Nucl Sci 1999; 46(6):2053–2058.

32.Schmand M, Eriksson L, Casey ME, Andreaco MS, Melcher C, Wienhard K, Flugge G, Nutt R. Performance results of a new DOI detector block for a high resolution PET-LSO research tomograph HRRT. IEEE Trans Nucl Sci 1998;45(6):3000– 3006.

33.Shao Y, Silverman RW, Farrell R, Cirignano L, Grazioso R, Shah KS, Vissel G, Clajus M, Tumer TO, Cherry SR. Design studies of a high resolution PET detector using APD arrays. IEEE Trans Nucl Sci 2000;47(3):1051–1057.

34.Meng LJ, Ramsden D. Performance results of a prototype depth-encoding PET detector. IEEE Trans Nucl Sci 2000; 47(3):1011–1017.

35.Kontaxakis G, Strauss LG, Thireou T, Ledesma-Carbayo MJ, Santos A, Pavlopoulos S, Dimitrakopoulou-Strauss A. Iterative image reconstruction for clinical PET using ordered subsets, median root prior and a We-based interface. Mol Imag Biol 2002;4(3):219–231.

36.Townsend DW. From 3-D positron emission tomography to 3-D positron emission tomography/computed tomography: What did we learn? Mol Imaging Biol 2004;6(5):275–290.

37.Phelps ME. PET: The merging of biology and imaging into molecular imaging. J Nucl Med 2000;41:661–681.

See also COMPUTED TOMOGRAPHY; IMAGING DEVICES; RADIOPHARMACEUTICAL DOSIMETRY.

PROSTATE SEED IMPLANTS

MARCO ZAIDER

Department of Medical Physics,

Memorial Sloan Kettering

Cancer Center

New York

DAVID A. SILVERN

Medical Physics Unit, Rabin

Medical Center

Petah Tikva, Israel

INTRODUCTION

Prostate seed implantation is a radiation therapy procedure by means of which small radioactive sources (colloquially referred to as ‘‘seeds’’) are permanently implanted in the tumor-bearing tissue. In the absence of more specific information concerning the location within the prostate of tumor deposits, the goal of the implant is to deliver a minimum dose to the entire prostate gland while minimizing the dose to any adjacent healthy tissues, in particular, the urethra and rectum. As cause-specific death in prostate cancer is predominantly the result of distant metastasis (and not local failure), the raison d’eˆtre of prostate implantation must be the notion of some causal link, as opposed to

mere association, between local control and distant disease (1–3). Whether this is indeed the case, remains at this time controversial (4).

The absorbed dose in the target as well as its spatial and temporal configuration is the only treatment tool available to the radiation oncologist. Consequently, patient eligibility for permanent prostate implants is based on the physician’s ability to physically deliver the dose to the target, in other words, placing the seeds at relevant locations within or near the gland. Guidelines for patient selection consist of stage (T1–T2 disease) and prostate volume (less than about 50 cm3). (Staging refers to the size and location of the tumor, whether tumor cells have spread to the lymph nodes, whether cancer cells have metastized to other parts of the body and to the abnormality of the tumor cells – this latter is referred to as grade and quantified by the Gleason score. Thus, T1 refers to a clinically inapparent tumor (not visible or palpable), and T2 refers to a low-grade tumor confined within the prostate. Pretreatment with androgenablation therapy is sometimes used to reduce the prostate volume.) Counter-indications for brachytherapy are short life expectancy (<5 years), the presence of metastatic disease, prior transurethral resection of the prostate (TURP), prostatitis, acute voiding symptoms, and inflammatory bowel disease (5).

The treatment may be delivered as monotherapy (implant alone) or in combination with external beam radiation therapy (EBRT) depending on whether the disease is confined (in which case monotherapy is administered) or extends beyond the prostate. Indications for combined treatment are extracapsular extension (ECE) and/or seminal vesicle invasion (SVI). It has been suggested that the patient’s prognosis category as defined by pretreatment stage, Gleason score, and prostate specific antigen (PSA) may be taken as an indication of the likelihood that the disease did not spread outside the prostate; essentially, the lower the risk, the larger the probability of a confined tumor. Two classification schemes are currently in use. According to one proposal, low-risk patients are those with T1–T2b stage, Gleason 2–6, and PSA of less than 10 ng/mL. Intermediate-risk patients have one unfavorable factor: PSA larger than 10 ng/mL, Gleason score 7 or larger, or T2c or greater. High-risk patients have at least two unfavorable risk factors. The other sorting idea considers the risk low when PSA 10 ng/mL, Gleason <7, and T1a or T2a; intermediate when 10.1 < PSA 20, Gleason ¼ 7 or T2b, and high otherwise (6,7).

The implant is a three-step process: First, using an imaging study of the prostate, the treatment planner calculates the number and location of seeds within the prostate volume that will result in a dose distribution in agreement with the prescribed constraints. The implantation is performed as a one-off outpatient surgical procedure. A postimplant evaluation, based on which the dosimetric quality of the implant is assessed, follows the treatment.

As with the other treatment choices (radical prostatectomy, EBRT), the survival benefits of transperineal permanent interstitial implantation among men with early, localized prostate cancer are uncertain (8,9). This state of affairs is compounded by the fact that the treatment of prostate cancer by either modality is accompanied by the

risk of (occasionally permanent) rectal and urinary toxicity, as well as sexual disfunction (10–12). (Brachytherapy may be less likely to result in impotence in urinary incontinence than other forms of treatment.) Thus, for many patients with early, localized prostate cancer (the typical brachytherapy candidate), the decision to undergo a treatment of questionable benefit yet tangibly impacting on their quality of life (QOL) is understandably difficult; as a result, diminishing the risk of complications, and at the same time maintaining good dosimetric coverage of the tumor, remains the overriding concern in prostate brachytherapy.

In this article, we shall provide a step-by-step tutorial on permanent prostate implants, (by necessity) as practiced at the Memorial Sloan Kettering Cancer Center (MSKCC).

PREPLANNING OR INTRAOPERATIVE PLANNING?

Two modalities are currently in use for planning prostate implants. Preplanning refers to the situation where the treatment plan is completed several weeks before the actual implantation. The American Brachytherapy Society (ABS) discourages this approach, essentially because of well-recognized problems that may develop at the time the plan is implemented (5). For instance:

1.It is difficult to duplicate the patient position in the operating room (OR) to match the patient position during the preplanning simulation.

2.Patient geometry can change over time; for instance, urine in the bladder or feces in the rectum may swell the prostate.

3.A pretreatment plan may prove impossible to implement due to the needle site being blocked by the pubic symphysis.

The alternative to preplanning, which we and others strongly advocate, is to perform the plan in the OR using ultrasound (US) images of the treatment volumes acquired with the patient on the operating table in the implantation position (13–19). The ability to obtain a dose-optimized plan within a reasonable time (say, 10 minutes or less) is the key to intraoperative approach. Clearly, a manual (i.e., trial and error) approach to finding optimal seed positions in the OR will not do. The sine qua non condition of OR-based planning is then the availability of a compu- ter-optimized planning technique, as described below.

ISOTOPE SELECTION AND DOSE PRESCRIPTION

The two isotopes commonly used for permanent prostate

implants are 125I (mean photon energy Eg ¼ 27 keV, half life T1/2 ¼ 60 days) and 103Pd (Eg ¼ 21 keV, T1/2 ¼ 17 days). An important property these isotopes share is their

low effective energies. At these energy levels, practically all decay radiation emitted by the implanted sources is absorbed in the patient’s body. This enables the patient to be discharged shortly after the implant procedure without fear of posing a radiation hazard to the general public or to the patient’s family.

PROSTATE SEED IMPLANTS

419

Dose prescription in prostate brachytherapy makes use of the concept of minimum peripheral dose (mPD), which is defined as the largest isodose surface that completely surrounds the clinical target (20). The total dose prescrip-

tion for patients treated at MSKCC is 144 Gy (mPD) for 125I and 140 Gy for 103Pd (21). The initial dose rate D˙ð0Þ

corresponding to these values can be calculated from

T1=2 ˙

 

Dtotal ¼ lnð2Þ Dð0Þ

(1)

Thus, for 125I, one has D_ ð0Þ ¼ 7 cGy=h, whereas for 103Pd, the corresponding value is 24 cGy/h. Based on radiobiological considerations it was hypothesized that the higher initial dose rate of 103Pd may be more appropriate for rapidly proliferating tumor cells. Gleason score is taken as marker for such cells, hence, the notion that 103Pd should preferentially be used for high-grade tumors. Retrospective studies have failed to demonstrate any clinical (22,23) or dosimetric difference between the two isotopes. (One may also note that in the United States, the price of a typical 125I seed is about half the price of a 103Pd seed; as well, a typical implant would use, say, 75 125I seeds as against some 100–110 103Pd seeds.)

A second difference between the two isotopes is the potential for somewhat larger relative biological effectiveness (RBE) of the 103Pd compared with 125I (24,25).

THE PHYSICAL CHARACTERISTICS OF 125I AND 103Pd

125I is produced in a reactor by irradiating 124Xe with neutrons to form 125Xe. 125Xe has a 16.9 h half-life and

decays to 125I via electron capture. 125I in turn decays to an excited state of 125Te via electron capture. The excited 125Te nucleus immediately releases its excess energy in the form of a 35.5 keV gamma photon in 6.66% of the transformations, with the energy of the remaining 93.34% of the transformations being released as internal conversion electrons (26). The rearrangement of the orbital electrons results in the emission of characteristic X rays and Auger electrons. The Auger electrons are blocked by the encapsulation of the source and do not directly contribute to patient dose. As a result of the lower energy

characteristic X-ray emissions, the average photon energy of 125I is 28 keV.

103Pd is produced in a reactor by irradiating stable 102Pd with neutrons. 103Pd decays with a 17 day half-life to excited states of 103Rh via electron capture. The excited 103Rh nuclei lose nearly all of their excess energy via internal conversion (26). The rearrangement of the orbital electrons results in the generation of characteristic X rays and Auger electrons. As is the case with 125I sources, the Auger electrons are blocked by the encapsulation and do not directly contribute to patient dose.

The radioactive sources must be fabricated to highquality control standards. The sources must remain biocompatible in vivo for decades. To prevent the internal contents of the sources from diffusing into the body tissues, the integrity of the source encapsulation must also remain sound for a period of decades. The physical size of the sources must be sufficiently small to allow their interstitial

420 PROSTATE SEED IMPLANTS

implantation without causing undue tissue trauma or interfering with physiologic function. In addition to biocompatibility issues, the sources must be physically strong enough to maintain their shape and integrity during sterilization, routine handling, and implantation. As well as to the need for physical ruggedness, the encapsulation must be made of a material that will not absorb an undue fraction of the emitted photons. For the purpose of source localization on radiographs and computed tomography (CT) images, it is desirable for the sources to be radioopaque while causing minimal imaging artifacts.

For meeting the aforementioned requirements, titanium is the material of choice for source encapsulation. Titanium is as strong as steel but 45% lighter. It is only 60% heavier than aluminum but twice as strong. This metal is also highly resistant to sea water. As human tissues have a high degree of salinity, titanium can maintain its integrity in vivo for the remainder of the patient’s life. Titanium has an atomic number of 22, low enough not to cause serious artifacts on CT images. Due to the high strength of titanium, the encapsulation can be made thin, minimizing absorption of the emitted photons. The main drawback of titanium is that it is expensive, significantly adding to the cost of the radioactive source.

The actual internal structure of the radioactive sources is vendor specific. The thickness of the titanium encapsulation ranges from 0.04 to 0.06 mm (26). In the classic Amersham 6711 source, the radioactive material is adsorbed to the surface of a silver rod. A schematic representation of this source is shown in Fig. 1. The silver serves as a radio-opaque marker. In other source models, the radioactive material is adsorbed onto resin beads, impregnated in ceramic material, or coated by other means onto various substrates. Most 125I and 103Pd radioactive sources include radio-opaque marker material. Gold, silver, tung-

sten, and lead are the commonly employed marker materials used in 125I and 103Pd sources. The differences in the

internal structures of the radioactive sources result in vendor-specific dose distributions. These differences result in differing photon energy spectra, source anisotropy, and self-absorption properties. The dosimetric properties of 125I and 103Pd sources are discussed in the next section.

DOSIMETRY

Before any radioactive source can be employed in a clinical implant procedure, the dose distribution around the source

I-125 adsorbed

0.05 mm

on silver rod

titanium

0.8 mm

0.5 mm

3.0 mm

4.5 mm

Figure 1. Schematic representation of an Amersham 6711 125I source.

in question must be known. One obvious requirement is that the accuracy of the dose calculations be as high as possible. Another useful requirement is wide acceptance of the dosimetric formalism applied. A universal dose formalism simplifies comparison of treatment outcomes of implants performed by different institutions and helps establish universal treatment protocols.

To assure accurate dosimetric calculations, the following physical properties must be ascertained:

1.Photon interactions with structures inside the source and encapsulation.

2.Geometric distribution of radioactive material within the source.

3.Photon interactions with the tissues encompassing the source.

4.Reduction of radiation intensity as a function of distance from the source.

Photon interactions within the source have a noticeable effect on the shape, intensity, and energy spectrum of the dose distribution surrounding the source. These internal photon interactions affect the anisotropy of the emitted radiation. (Source anisotropy is a measure of angular dependence of the dose distributions surrounding the source.) For a point source with no asymmetric selfabsorption, dose would only be a function of distance from the source without any angular dependence. Photon interactions inside the source also influence the energy spectrum of the emitted photons. Depending on the physical construction and materials used to fabricate the source, the emission spectra will vary among different source models. Lower energy photons will be preferentially attenuated compared with higher energy photons. The degree of filtration is dependent on the construction of the source. Another alteration to the intrinsic spectrum of the radioactive material is the generation of characteristic X rays by photoelectric interaction with the materials inside the source. One noteworthy example of spectral variations among different 125I source models is the use of silver as the radio-opaque marker used by certain manufacturers. Silver has a K-edge that occurs at 25 keV (27,28). As the intrinsic photon emissions of 125I are in this energy range, the photoelectric cross section for interaction with the silver marker is high. As a result, I125 sources using silver as the radio-opaque marker will have a local peak around 25 keV in their emission spectra. I125 sources using another element for radio-opacity will not exhibit a 25 keV peak in their emission spectra. Differences in source construction also influence the degree of Compton scattering, further altering the emission spectra. The geometric distribution of the radioactive material inside the source will affect the angular dependence of the dose distributions around the source as well as the dose reduction as a function of distance from the source. This effect will be investigated when source geometry factors are discussed later in this section.

Another important factor that must be calculated by a dose formalism is tissue attenuation as a function of distance from the source. Tissue attenuation is a function of how the emission photons interact with the tissues. When using I125 or 103Pd sources, the predominant interactions

are Compton scattering and photoelectric effect. In soft tissue, the probabilities of photoelectric and Compton interactions are equal at a photon energy of about 25 keV (28). The dose decreases with distance from the source due to these photon interactions.

The simplest dosimetric formalism is to assume pointsource geometry, neglecting angular dependencies on the dose distributions. In this formalism, the dose is strictly a function distance from the source. Two components are assumed to contribute to the resulting dose at a given point. One component is the dose falloff due the geometric shape of the radioactive source. For a point source, this falloff component is the inverse square law, namely 1/r2. This dose falloff occurs irrespective of any photon interactions in the medium surrounding the source. It is solely dependent on the photon fluence intensity being geometrically reduced by the inverse square law. The second component of dose falloff results from photon interactions in the surrounding medium. One analytical approach to modeling this phenomenon is to assume that the dose reduction follows a simple exponential function, namely F(r) ¼ e mr where m is an average linear attenuation coefficient for the energy spectrum of the emitted photons. It is also assumed that dose is directly proportional to the source activity A. The equation for calculating the dose using this formalism is

DðrÞ ¼ GA fmedTav

e mr

(2)

r2

where D represents the dose at distance r and A is the source activity. The factors G and fmed are the exposure rate constant and the tissue f-factor, respectively. The f-factor converts the exposure in air to absorbed dose in a small piece of tissue just large enough to assure electronic equilibrium. Tav is the average life an isotope atom exists before undergoing a nuclear transformation and m is the effective linear attenuation coefficient. This analytical equation suffers from two drawbacks, the first being that this exponential equation is only rigorously correct for narrow-beam geometries. In deriving this equation, it is implicitly assumed that all photons that interact with the medium are removed from the beam and that no further interactions of scattered photons occur in the path of the beam. The geometry of a radioactive source in a medium is clearly not narrow-beam. Compton photons do in fact interact with the medium in the beam and hence contribute to dose. A second drawback of using an exponential is the implicit assumption that the photons are mono-energetic, clearly not the case for either 125I or 103Pd emissions. Although this formalism is not rigorously correct for the reasons stated, it has been used for many years in brachytherapy treatment planning. By empirically determining values for G and m, the discrepancies of calculated and measured doses could be made acceptably small. In the years that this equation was used, modern instrumentation was not available for precise dose measurement, computers were slow, and the standards of conformance were not as stringent as today.

The accuracy of the formalism can be improved by replacing the exponential equation with a data table. The values in the table consist of experimentally measured doses in water at known distances, multiplied by the

PROSTATE SEED IMPLANTS

421

square of the distance. Photon interactions in water are similar to those in soft tissue. The dose at any arbitrary distance from the source is calculated via linear interpolation of the tabulated data and by dividing by the square of the distance. Using tabulated data solves the two problems associated with the exponential function. As tabulated data were derived from measurements in the true broadbeam geometry of the source, the formalism inherently accounts for the dose occuring from Compton-scattered photons as well as the primary photons. As a table could be created for any source model, the data will inherently account for the energy spectrum of the emitted photons. An added benefit provided by the table is that it accounts for the selective tissue filtration of lower energy photons with increasing depths. As the emission spectra vary among different source models, the tissue filtration will also likewise vary.

A more sophisticated formalism can be developed by accounting for the geometric distribution of the radioactive material inside the source. The most natural extension of the formalism is the assumption that the radioactive material is distributed as a line source. Although the distribution of the radioactive material in many commercially available sources is not a true uniform line, the line source model is still a more accurate and realistic representation than the point source model. Applying a line source model significantly complicates the dosimetric computations. When applying the point source model, only the distance from the source to the calculation point need be known to uniquely determine the dose. By contrast, using a line source model requires that the angular orientation in addition to the distance of each dose calculation point in relation to the source be known. The angular orientation of the calculation point with respect to the source needs to be calculated using analytic geometry.

The angular dependence of the dose distribution around a source is the result of two separate processes. The first results from the distribution of the radioactive material inside the source, and the second results from the angular dependence of attenuation of the photons within the source. To develop a formalism based on a line source, it is easiest to start by calculating the dose to a differential piece of tissue in empty space. It is assumed that there is no attenuation in the tissue and that electronic equilibrium exists. As tissue effects will be introduced at a later stage, this assumption does not compromise the rigor of the development of the formalism. At this stage, the selfabsorption and tissue attenuation will not be considered.

The angular dependence of dose resulting from the geometric distribution of the radioactive material can be modeled by a line source. A schematic representation of a line source is shown in Fig. 2.

For any point P one can define a two-dimensional coordinate system as shown in Fig. 2. With such a coordinate system defined, the dose to any point P from the line source can be calculated. It is seen from Fig. 2 that the following formulation holds:

X ¼ r cos u; Y ¼ r sin u;

r ¼ Y=sin u; tan u ¼ Y=X;

X ¼ Y=tan u

ð3Þ

422 PROSTATE SEED IMPLANTS

Y

P

r

X

Active Source Length L

 

Figure 2. Dose to a point P from a line source. Physical length includes encapsulation. Active length only includes length of radioactive material.

The differential length dX is calculated to be

dX ¼ Yðtan uÞ 2ðsec uÞ 2du ¼ Yðsin uÞ 2du

(4)

The total activity A of the line source is assumed to be uniformly deposited along the active length L of the source. Under ideal conditions, uthe dose to a differential piece of tissue at point P in free space is given by

Tav

 

DðrÞ ¼ AG fmed r2

(5)

To calculate the dose from the line source shown in Fig. 2, the source can be considered a continuum of differential point sources resulting in

 

 

 

A

 

dX

 

 

A

 

d0

 

 

dD ¼

 

G

fmedTav

 

 

¼

 

 

G

fmedTav

 

 

(6)

 

L

r2

L

Y

The total dose to point P is thus

 

 

 

 

 

 

 

A

 

 

 

02

d0

 

 

A

 

 

 

 

02 01

 

D

 

G fmedTavZ

 

 

G

fmedTav

(7)

 

 

 

r sin 0

 

¼ L

01

Y ¼ L

 

 

By using equation 7, it is now possible to account for the linear distribution of radioactive material inside the source. In clinical brachytherapy, however, one is generally interested in knowing the dose when the source is in the patient and not in free space. This means that the photon interactions with the surrounding tissues must be factored into the analysis. Additionally, the self-absorp- tion of the source must be taken into account. Unlike the linear distribution of the radioactive material, the selfabsorption and tissue interactions are not amenable to a rigorous treatment. These data must be derived either from direct measurement or from Monte Carlo simulation, or a combination of the two. The data tables would need to be two-dimensional to account for the distance from the calculation point to the source as well as the angular orientation of the calculation point to the center of the source. As the active length of the source is known, the values of u1 and u2 are uniquely determined. Once these data tables are available, the component of the dose due to geometric distribution of the radioactive material can be removed from the data by dividing the tabulated doses by the ideal line source dependency. Reviewing equation 7, one can separate the terms involving the linear distribu-

Physical Source Length

tion of the radioactive material. Doing this yields the following:

G

02 01

(8)

¼ Lr sin 0

 

 

where G is referred to as the geometry factor. Dividing the tabulated doses by G removes the line source contribution. Removing the geometry factor reduces the dependency of the data on distance from the source, enabling the use of smaller data tables for attaining the same degree of accuracy. The data tables still account for the tissue interactions and self-absorption. For permanent implant brachytherapy, the data tables can be in the form of total dose per unit source activity (or air kerma strength) because the implant time is infinite and the isotope half-life is known. The new formalism can be summarized as follows:

1.For each point P in space, define a two-dimensional coordinate system and compute the distances to the centers of each implanted source and compute the respective angles u, u1, and u2.

2.Use equation 8 to calculate the geometry factors at point P for each implanted source.

3.Interpolate the data tables based on the distances and angles for each implanted source calculated in step 1.

4.For each implanted source, multiply the respective geometry factor, source activity (or air kerma strength), and interpolated table value together. Each product represents the dose to point P for each implanted source.

5.Sum the individual doses to obtain total dose at point P.

These calculations are tedious and time consuming if performed by hand. The only feasible way forward is to code the formalism in software and have the computer perform the computations. In this way, a finely spaced three-dimensional data grid of calculation points can be rapidly calculated. The dose at any arbitrary point in space can be calculated by interpolating the three-dimensional grid of calculated doses. This will be discussed in a later section devoted to treatment planning.

The dose formalisms discussed thus far are but a small sample of the calculation methods used in brachytherapy

treatment planning. They were discussed to serve as illustrative examples and to provide a brief introduction to the physics involved in the development of dosimetry formalisms. Over the years, many different dosimetry formalisms have been proposed and implemented. By the early 1990s, treatment planning computers have become ubiquitous in radiotherapy departments. With many such systems in use, questions began to arise regarding the accuracy, consistency, and general agreement of the calculated doses generated by these systems. It was obvious that there were differences in the values calculated by these treatment planning systems as each system implemented its own dosimetric algorithm. During this same period, researchers also proposed new formalisms based on physical quantities not widely used in brachytherapy treatment planning. As a result of these issues, the American Association of Physicists in Medicine (AAPM) decided that the time had come to develop a standardized brachytherapy formalism. The adoption of such a formalism would standardize the dosimetric calculations, reducing the differences in values calculated by different treatment planning systems. The doses calculated among different institutions will be in closer agreement, enabling more realistic comparisons of different brachytherapy protocols.

The AAPM Radiation Therapy Committee Task Group 43 established a new recommended formalism. This formalism was published in 1995(26). The Task Group 43 (TG43) formalism is a radical departure from most established methodologies used up to that time. In following the new formalism, changes in dosimetric values in some instances were of sufficient magnitude to mandate changes in prescription doses for maintaining clinical consistency with older formalisms.

The most fundamental change recommended by TG43 is to use air kerma strength Sk in lieu of activity. The activity of a radioactive source is the number of disintegrations per unit time. As a fundamental unit, activity in and of itself does not provide any information regarding the nature of the energy deposition of the decay emissions. Traditional dosimetry formalisms need to use exposure rate constants and f-factors to convert from activity to exposure in air to dose deposited in tissue. The exposure rate constant is also isotope-specific.

Air kerma strength is defined as follows. A mass of air, dm, is placed a distance rref from the source along the perpendicular bisector. The source and air mass are in a vacuum. Air kerma strength is the kerma rate in mass dm multiplied by the square of the distance rref. Unlike activity, which is only applicable to radioactive isotopes, air kerma strength can be applied to any source of uncharged particles. For example, the radiation output from a linear accelerator or X-ray tube can also be quantified in terms of air kerma strength.

Another new quantity used by TG43 is the dose rate constant L defined as the dose rate to water at a distance of 1 cm from the perpendicular bisector of the source. Using the dose rate constant represents another departure from basing absorbed dose on exposure to air. It is a trend in the medical physics community to move toward basing dosimetric calculations and measurements on dose to water.

PROSTATE SEED IMPLANTS

423

The dose rate constant replaces the exposure rate constant used in older brachytherapy formalisms.

For modeling the contribution of the geometric distribution of the radioactive material inside the source, TG43 endorses the use of either a point source representation (inverse square law) or a line source representation (equation 8). These geometric factors have been in use before the advent of the TG43 report.

The interactions of the emitted photons in tissues are modeled by a new function defined by TG43, namely the radial dose function g(r). The function g(r) is given by the following equation:

g r

dDðr; 00Þ=dt Gðr0; 00

Þ

(9)

ð Þ ¼ dDðr0; 00Þ=dt Gðr; 00Þ

 

where D is the dose at a point along the perpendicular bisector at a distance r from the source and G is the geometry factor that is equal to either the inverse square law or equation 8. In equation 9, u0 is equal to p/2. The radial distance r0 is the reference distance to which all TG43 parameters are normalized. In practice, r0 is taken to be 1 cm. It must be stressed that g(r) is only defined along the perpendicular bisector of the source. It can be seen that when r is equal to r0, g(r) equals 1. The factors in equation 9 involving G remove the dependency of the geometric distribution of radioactive material inside the source on g(r). The radial dose function models the interactions of the photons with tissues along the transverse axis of the source. In practical implementations of TG43, g(r) is based on tabulated values, which in turn are based on measured data or Monte Carlo simulations. In some implementations, analytic functions are fit to the data, whereas in others, the value of the radial dose function is calculated via interpolation.

Another phenomenon that needs to be modeled is the anisotropic nature of the radiation resulting from photon interactions inside the source. If there were no interactions of the radiation within the source, all of the angular dependence of the radiation pattern would result from the geometry factor, assuming the source is in a homo-

geneous medium. In reality, however, self-absorption is significant, especially at the low energies of 125I and 103Pd.

In the original TG43 report, three methods of modeling source anisotropy were proposed. The most general of these models is the source anisotropy function F(r, u). This function is defined as follows:

F

r;

dDðr; 0Þ=dt Gðr; 00Þ

(10)

0Þ ¼ dDðr; 00Þ=dt Gðr; 0Þ

ð

 

 

This function in effect is the ratio of the dose rate at an arbitrary distance from the source r and angle u multiplied by the ratio of the geometry factor at a distance r but on the transverse axis and the geometry factor at the location r and u. This function quantifies the angular variation of the dose distribution removing the contribution of the geometry function. In most practical implementations, the anisotropy function is calculated via bilinear interpolation of a two-dimensional data table.

A second method for modeling source anisotropy is to represent source anisotropy as a function of only the

424 PROSTATE SEED IMPLANTS

distance from the source. In this case, the dose rate is averaged over all values of solid angle from 0 through 4p steradians. The one-dimensional function is referred to as the anisotropy factor fan(r). By taking advantage of the cylindrical symmetry of radioactive sources, the solid angle integral for defining this factor reduces to the following:

 

 

1

 

 

p

dDðr; 0Þ

 

 

 

 

 

f r

Þ ¼

 

 

Z

sin

0

d

0

(11)

2dD r;

 

=dt

 

anð

0Þ

dt

 

 

 

 

ð

 

0

 

 

 

 

 

 

One fundamental difference between defining fan(r) and F(r,u) should be pointed out. The geometry factor is not removed from fan(r) as is the case for F(r, u). From a numerical standpoint, the average value of the geometry factor taken over 4p steradians is nearly equal to the nominal value of G(r,u0). This is especially true for distances greater than the active length of the source. Moreover, fan(r) is an average value that will result in an inevitable error in the actual dose calculation. Factoring out the geometry factor would not significantly improve the accuracy of the dosimetry. If the geometry factor contribution was in fact removed, two anisotropy factors would need to be calculated, one for point source geometry and the other for line source. The main motivating factor for defining fan(r) is to accommodate existing treatment planning systems that do not support two-dimensional dose calculations. Given a choice, using F(r,u) is preferred as it is rigorous.

A third anisotropy correction method prescribed by the original TG43 report is the use of an anisotropy constant fan that is an average value independent of distance. The original TG43 report was updated in 2004. In the updated report, use of anisotropy constants was made obsolescent and is no longer considered consistent with current standards of practice. A more detailed discussion of the TG43 update is discussed below.

Application of the TG43 formalism can be summarized by the following equations where all of the terms have been discussed:

dD

r;

0Þ ¼

S

L

Gðr; 00Þ

g r

F

r;

0Þ

(12)

ð

 

K

 

Gðr0; 00Þ

ð Þ

ð

 

 

Equation 12 is the rigorous implementation of the original TG43 formalism. A simpler implementation of equation 12 for use in treatment planning systems not supporting twodimensional dose calculations is the following:

dD

r;

0Þ ¼

S

L

Gðr; 00Þ

g r

f

r

(13)

Gðr0; 00Þ

ð

 

K

 

ð Þ

 

anð Þ

 

The third equation using the anisotropy constant is as follows:

dD

r;

0Þ ¼

S

L

Gðr; 00Þ

g r f

 

(14)

ð

 

K

 

Gðr0; 00Þ

ð Þ

an

 

Use of equation 18 is no longer recommended in the updated TG43 protocol. All treatment planning systems need to be capable of performing one-dimensional calculations. It is thus always possible to use the anisotropy factor fan(r). In some treatment planning systems, the product of

the radial dose function and the anisotropy factor may need to be used if only one distance-dependent term is used for calculating the dosimetry.

To end this section, a brief discussion of the updated TG43 formalism is presented. This update was published in 2004, 9 years after the original TG43 report (29,30). In the original formalism, the same radial dose functions were used irrespective of geometry factor. In the new formalism, two sets of radial dose functions are recommended. One radial dose function is to be used for point source geometry, whereas the other is to be used for line source geometry. Although the updated and original reports define the radial dose function using the same equation, the published values could be used for both geometry factors. To improve the consistency of the dose calculations, two sets of radial dose functions are now recommended. The choice of geometry factor dictates which radial dose function is to be used. Further refinements were also made in the tabulated values presented. In the intervening 9 years between the original report and the update, several new radioactive sources were introduced. The updated report includes published data for use with the newer source models. In the new report, the use of anisotropy constants is no longer recommended.

ULTRASOUND-GUIDED IMPLANTATION TECHNIQUE

The implantation proceeds as follows. The patient is placed in the extended lithotomy position. An ultrasound probe is positioned in the rectum, and needles are inserted along the periphery of the prostate using a perineal template as a guide. Trans-axial images of the prostate are acquired at 0.5 cm spacing (from base to apex), transferred to the treatment planning system using a PC-based video capture system, and calibrated. For each US image, prostate and urethra contours as well as the anterior position of the rectal wall are entered. Needle positions are identified on the ultrasound images and correlated with the US template locations. The contours, dose reference points, needle coordinates, and data describing the isotope/activity available along with predetermined dose constraints and their respective weights serve as input for the dose optimization algorithm.

TREATMENT PLANNING

A commissioned treatment planning system is the minimum equipment required for performing brachytherapy treatment planning. State-of-the-art treatment planning systems are based on modern computer hardware and software. Practically all modern brachytherapy treatment systems implement the TG43 dosimetry formalism for lowenergy radioactive sources. These systems provide interfaces to imaging scanners such as CT, US, and magnetic resonance imaging. Having a scanner interface presents the opportunity of superimposing the dosimetric information on the anatomical images. It is a simple matter to superimpose dosimetric and anatomic information on a series of images. In most situations, the anatomical images are parallel to one another. The treatment planning computer calculates the doses to a three-dimensional grid of

PROSTATE SEED IMPLANTS

425

points spatially registered with the anatomic images. After the dosimetric calculations are completed, the dose distributions are represented as a series of colored contour lines commonly known as isodoses. Isodoses are analogous to the contour lines of a mountain on a topographical map. In the case of a mountain, the contour lines represent locations of equal elevation. In a similar manner, isodoses represent the locus of points of equal dose. An example of isodose contour lines superimposed on an axial ultrasound of the prostate is shown in Fig. 3.

In Fig. 3, the outermost yellow contour line represents 50% of the prescribed dose. This is the lowest dose level shown in the figure. Similarly, the innermost lavender contour represents 150% of the prescribed dose. Usually, but not always, the inner isodoses represent higher doses than the outer contours. On a topographical map of a mountain, the inner contour lines represent higher elevations than the outer contours.

Viewing isodose contours superimposed on anatomical images provides a means for visually evaluating the conformity of the dose distributions with the target anatomy. Ideally, the prescription isodose line should exactly conform to the shape of the target volume. This goal is not generally achievable although it can be approached. In Fig. 3, the green contour is the prescription isodose line. There is close conformity of the green isodose line with the prostate. In addition to being able to visually evaluate the conformity of the prescription isodose with the anatomy, it is also possible to ascertain doses received to critical structures. In the case of the prostate, the critical structures where excessive doses should be avoided are the rectum and urethra. Referring to Fig. 3, it is observed that the urethra receives less than 120% of the prescription dose. To properly evaluate a treatment plan, the isodoses on all images must be reviewed. Some slices may manifest excel-

Figure 3. Isodoses superimposed on an ultrasound image of the prostate. The thick white contour line delineates the prostate. The inner isodose lines depict higher dose levels than outer contours. Note the conformity of the green prescription isodose with the shape of the prostate. The inner white contour is the urethra. Note avoidance of the 120% isodosecontour with the urethra, a radiosensitive critical structure.

lent conformity and satisfactory critical structure doses, whereas reviewing other images may reveal unsatisfactory dose distributions.

In traditional prostate brachytherapy treatment planning, the physicist manually selects source and needle locations in an attempt to optimize the treatment plan. This entails maximizing the conformity of the prescription isodose with the prostate and minimizing the doses received by the critical structures. This is a tedious, time-consuming process because the conformity and critical structure conditions must be simultaneously met on all slices. Often, improvement on one slice degrades the dosimetry on another. There is no universal standard regarding what constitutes an acceptable treatment plan. There are differing opinions both among individual physicians and treatment centers. These differences generally pertain to acceptable dose values covering the prostate and critical structures. Despite these differences, however, the aim of treatment planning is to maximize the coverage of the prescription dose to the prostate and minimize the doses to the critical structures.

A popular graphical method for evaluating treatment plans is known as a cumulative dose volume histogram, commonly referred to as a DVH. A cumulative DVH is a graph of the volume of a target or critical structure as a function of minimum dose received by the volume. A typical cumulative DVH plot is shown in Fig. 4. Referring to this figure, it is observed that at low doses, the volume covered is essentially 100%. This shows that the target as well as the adjacent critical structures receive a significant dose. In Fig. 4, the blue graph represents the DVH for the urethra. It is observed in this graph that around 95% of the urethra receives doses exceeding 50% of the prescription. As the urethra traverses the prostate, it is physically impossible for the urethra not to receive a significant dose

426 PROSTATE SEED IMPLANTS

Figure 4. Typical cumulative DVH for a prostate implant.

without sacrificing the dose to the prostate. In welldesigned treatment plans, the urethral doses are minimized. The specific criteria depend on the preferences of the physician or institution.

As was briefly indicated, manually optimizing prostate treatment plans is a time-consuming process. To address this issue, the authors developed an automated treatment planning and optimization system. The optimizer used in this system is based on a genetic algorithm (GA) (13,19). GAs are a genre of optimization algorithms that attempt to find the optimal solution of a problem based on evolutionary natural selection (31–34). In the case of the intraoperative optimizer, a population of bit streams is used to represent a population of treatment plans. Each bit stream is a one-dimensional array of numbers whose values are either one or zero. The length of the arrays is equal to the total number of potential source locations in the implant. The three-dimensional coordinates of each potential source location are known and fixed. If a source is present at a potential site, the bit for the corresponding site location is one. Conversely, if a source is not present at a potential site, the corresponding bit in the stream is set to zero. In this way, the treatment planning problem is encoded in a form amenable to genetic optimization.

To determine ‘‘how good’’ is the plan represented by a bit stream, an objective function is defined. The job of the objective function is to calculate a score, representing a figure of merit. The larger the score, the better the treatment plan, as determined by the criteria defined in the objective function. For instance (19), let PD be the prescription dose; the prostate score is the number of (uniformity) points that satisfy PD D 1.6 PD, the urethral score is the number of points in the urethra for which D 1.2 PD, and the rectal score sums up all points in the rectum for which D 0.78 PD. From this one obtains

a raw score:

Raw score ¼ 5 ðprostate scoreÞ þ 35 ðrectal scoreÞ

þ 50 ðurethral scoreÞ

and a final score (used in the optimization algorithm), which is a linear function of the raw score [¼A (raw score) þ B].

The objective function is defined to increase the score for greater coverage of the prescription dose to the prostate. On the other hand, as the doses to the rectum and urethra increase, the score is decreased. As can be observed, there are conflicting requirements and the score reflects all of these conditions.

At the beginning of execution, the bit streams are randomly assigned arbitrary bit patterns. During execution of the GA, these bit streams are gradually ‘‘evolved’’ into higher scoring streams, indicating improvements in the treatment plans they represent. In an attempt to improve the scores of a population of treatment plans, the GA mimics biological evolutionary processes. Crossover, mutation, and survival-of-the-fittest are simulated by the GA in an attempt to maximize the scores received by the population. Two bit streams in the population are selected to act as ‘‘parents.’’ Crossover is implemented by interchanging corresponding randomly selected bits between the parents. Mutation is affected by inverting a very small ( 1%) number of bits in the two bit streams. The modified bit streams are evaluated and replace the two lowest-scoring chromosomes in the population.

In biological evolution, the fitter species has a higher likelihood of survival compared with the lesser fit species. This means that although the lesser fit species has a nonzero probability of survival, species that are better suited to their environments have higher survival probabilities.

PROSTATE SEED IMPLANTS

427

Natural selection is simulated in the GA by favoring higher scoring bit streams to act as parents. In 50% of the parent selections, the two highest scoring bit streams are selected to act as parents. In the other 50% of the selections, two bit streams are randomly selected, irrespective of their score. The parents undergo crossover and mutation and replace the two lowest-scoring bit streams in the population. After this process is repeated several thousands of times, a population of high scoring bit streams is generated. This in turn translates into a population of optimized treatment plans. The highest scoring treatment plan is selected to be used for the implant.

The GA can generally generate a treatment plan in under 5 min. Actual execution times depend on the size of the prostate and the number of needles used for implanting the radioactive sources.

Other optimization algorithms used in prostate brachytherapy are simulated annealing (35,36) and branch- and-bound (13–15).

SECONDARY DOSE VERIFICATION

Once the plan is approved by the clinician a second physicist must independently verify and formally approve the plan. An estimate of the total number of seeds can be obtained with the following equations, which give—for a given average dimension of the volume to be implanted, davg—the total source strength SK required (37):

For a 125I permanent implant and a prescription dose of 144 Gy:

is elevated may contain clinically significant cancer (38– 42). In general, ‘‘clinically significant’’ refers to cancer cells that are fast proliferating and/or radioresistant (features associated with local failure) or of high grade and thus with a potentially larger probability of distant dissemination. There is limited evidence of a correlation between choline levels and histological grade (Gleason score). As well, biochemical arguments have been invoked to support the expectation that a larger value of this ratio is expected to reflect an increased rate of cell proliferation although no direct proof exists yet.

As applied at MSKCC, the implementation of an MRSbased dose escalation amounts to increasing the dose to the MRS-positive voxels to 200% of the prescription dose (with no upper limit) while keeping the urethral and rectal dose within the usual range of constraints (14).

POST-IMPLANT ANALYSIS

At MSKCC, post-implant evaluation is performed the day of the implant. The number of seeds implanted is confirmed with a pair of planar radiographic films. A CT study (3 mm slices) is then obtained, and anatomical structures are marked out on each slice. The coordinates of the center of each seed are determined by using appropriate computer software (Interplant Post-implant Analysis System, Version 1.0: Burdette Medical Systems, Inc., Champaign, IL). With the seeds thus identified, (DVHs) are calculated for each structure of interest and compared with the original plan.

 

 

 

 

 

8

5:709

 

davg

 

davg

 

3 cm

CLOSING WORDS: KNOWN PROBLEMS AND POSSIBLE

 

 

Sk

 

 

 

 

>

 

 

 

cm

 

 

 

 

 

 

 

 

 

>

 

 

 

 

 

 

2:2

 

 

(15)

FIXES

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

U

>

 

 

 

 

 

 

 

 

 

 

 

 

 

>

1:524

davg

davg > 3 cm

 

 

 

 

¼ <

 

 

 

 

 

 

>

 

 

 

 

 

 

 

 

 

 

 

For a

 

 

 

 

>

 

 

 

cm

 

 

 

 

Permanent brachytherapy prostate implants are now a

Pd permanent implant and a prescription dose of

well-accepted treatment modality for early stage prostate

 

103

 

 

 

:

 

 

 

 

 

 

 

 

 

 

 

140 Gy:

 

 

 

 

 

 

 

 

 

 

 

 

cancer. The two major limitations of this procedure are

 

 

 

 

 

8

 

 

davg

 

 

 

 

higher incidence of urethral complications (when com-

 

 

 

 

 

29:41

 

davg

 

3 cm

pared with external-beam radiotherapy) and — for some

 

 

Sk

>

 

 

 

 

cm

 

 

 

 

patients — lower than prescribed delivered doses. In this

 

 

 

 

 

>

 

 

 

 

 

 

 

2:56

 

 

(16)

section, we list several unresolved issues that may be

 

 

U

>

 

 

 

 

cm

 

 

 

 

 

>

 

 

 

 

 

 

 

 

 

 

 

>

 

 

 

 

 

 

 

 

 

 

 

 

 

 

:

5:395

davg

 

davg > 3 cm

responsible for this state of affairs and suggest possible

 

 

¼ <

 

 

 

 

 

 

>

 

 

 

 

 

 

 

 

 

 

 

solutions.

One evaluates the total number of radioactive sources

Post-implant evaluations of permanent prostate

needed by dividing the total required source strength SK

implants often indicate significant differences between

by the single-seed strength used in that implant. In gen-

the intended plan and its actual implementation. Although

eral, one seeks agreement of 10% or better between the

an experienced physician can minimize the magnitude of

planned and the expected number of seeds.

these differences, many factors controlling execution of the

The plan verification must also determine that the

plan (e.g., bleeding, tissue swelling) are subject to random

correct prescription dose was used and that input data

fluctuations. This often leads to a higher than intended

to the planning software was correctly entered.

dose to urethra and rectum and/or lower or higher doses to

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the prostate, especially at the periphery of the gland. In our

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

view, this discrepancy represents the most important

DOSE ESCALATION TO PROSTATE SUBVOLUMES

obstacle and challenge that currently needs to be overcome

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

to achieve consistent application of a low urethral and

If information is available on tumor burden in specific

rectal dose range and thereby reduce morbidity after pros-

subvolumes of the prostate, dose escalation to these voxels

tate brachytherapy. In a series of recent articles the con-

is recommended. For instance, it has been hypothesized

cept of intraoperative dynamic dosimetric optimization has

that regions of the prostate where the choline/citrate ratio,

been proposed (43–46). The idea is to re-optimize the plan

as determined by magnetic resonance spectroscopy (MRS),

several times during the implantation based on the actual

428 PROSTATE SEED IMPLANTS

positions of the seeds already implanted. The key problem is obtaining in real time (and within a reasonable time interval — 5 min or less) the coordinates of the implanted seeds in the system of reference used for planning. Visualization of the actual seed positions on the intraoperative ultrasound image is difficult, if not impossible, to achieve because of significant artifacts noted on the ultrasound image from needles and/or hemorrhage within the gland.

One can reconstruct seed coordinates from fluoroscopic images taken at three different angles (43,44) or, equivalently, from a CT study obtained in the OR for instance, using a C-arm with CT capabilities. CT-based systems that perform seed segmentation do exist (e.g., Variseed from Varian Medical Systems, Inc, Charlottesville, VA; Interplant Post-implant Analysis System, Burdette Medical Systems, Inc., Champaign, IL), but at this time they do not seem to have the capability of performing this task on the fly and at the same time maintain the required seeddetection reliability.

A second problem concerns the effect of changes in prostate volume (edema shrinkage) as well as seed migration after implantation and the effect this has on a treatment plan that is based on the geometry of the target at the time of implantation. A method of planning that incorporates temporal changes in the target-seed configuration during dose delivery and makes use of the concept of effective volume has been developed by Lee and Zaider (47).

The preceding enumeration of problems has been brief and (admittedly) selective, but we hope to motivate the reader to take a careful look at these important issues. The desideratum of dosimetric conformality in permanent prostate implants remains a topic of active interest in the brachytherapy community, and no doubt the last word on this subject has not yet been spoken.

BIBLIOGRAPHY

1.Coen JJ, Zietman AL, Thakral H, Shipley WU. Radical radiation for localized prostate cancer: Local persistence of disease results in a late wave of metastases. J Clin Oncol 2002;20:3199–3205.

2.Zagars GK, vonEschenbach AC, Ayala AG, Schultheiss TE, Sherman NE. The influence of local-control on metastatic dissemination of prostate-cancer treated by external beam megavoltage radiation-therapy. Cancer 1991;68:2370–2377.

3.Valicenti R, Lu JD, Pilepich M, Asbell S, Grignon D. Survival advantage from higher-dose radiation therapy for clinically localized prostate cancer treated on the radiation therapy oncology group trials. J Clin Oncol 2000;18:2740–2746.

4.Logothetis C. Challenge of locally persistent prostate cancer: An unresolved clinical dilemma. J Clin Oncol 2000;20:3187.

5.Nag S, Shasha D, Janjan N, Petersen I, Zaider M. The American Brachytherapy Society recommendations for brachytherapy of soft tissue sarcomas. Int Radiat Oncol Biol Phys 2000;49:1033–1043.

6.D’Amico AV, Whittington R, Malkowicz SB, Fondurulia J, Chen MH, Kaplan I, Beard CJ, Tomaszewski JE, Renshaw AA, Wein A, Coleman CN. Pretreatment nomogram for pros- tate-specific antigen recurrence after radical prostatectomy or external-beam radiation therapy for clinically localized prostate cancer. J Clin Oncol 1999;17:168–172.

7.Moul JW, Connelly RR, Lubeck DP, Bauer JJ, Sun L, Flanders SC, Grossfeld GD, Carroll PR. Predicting risk of prostate

specific antigen recurrence after radical prostatectomy with the center for prostate disease research and cancer of the prostate strategic urologic research endeavor databases. J Urol 2001;166:1322–1327.

8.Albertsen PC, Hanley JA, Gleason DF, Barry MJ. Competing risk analysis of men aged 55 to 74 years at diagnosis managed conservatively for clinically localized prostate cancer. JAMA 1998;280:975–980.

9.Chodak GW. Comparing treatments for localized prostate cancer-persisting uncertainty. JAMA 1998;280:1008–1010.

10.Jani AB, Hellman S. Early prostate cancer: Clinical decisionmaking. Lancet 2003;361:1045–1053.

11.Sandhu AS, Zelefsky MJ, Lee HJ, Lombardi D, Fuks Z, Leibel SA. Long-term urinary toxicity after 3-dimensional conformal radiotherapy for prostate cancer in patients with prior history of transurethral resection. Int J Radiat Oncol Biol Phys 2000;48:643–647.

12.Zelefsky MJ, Hollister T, Raben A, Matthews S, Wallner KE. Five-year biochemical outcome and toxicity with transperineal CT-planned permanent I-125 prostate implantation for patients with localized prostate cancer. Int J Radiat Oncol Biol Phys 2000;47:1261–1266.

13.Lee EK, Gallagher RJ, Silvern D, Wuu CS, Zaider M. Treatment planning for brachytherapy: An integer programming model, two computational approaches and experiments with permanent prostate implant planning. Phys Med Biol 1999;44:145–165.

14.Zaider M, Zelefsky MJ, Lee EK, Zakian KL, Amols HI, Dyke J, Cohen G, Hu Y, Endi AK, Chui C, Koutcher JA. Treatment planning for prostate implants using magnetic-resonance spectroscopy imaging. Int J Radiat Oncol Biol Phys 2000;47:1085–1096.

15.Gallagher RJ, Lee EK. Mixed integer programming optimization models for brachytherapy treatment planning. Proc/ AMIA Annu Fall Symp 1997; 278–282.

16.Lee EK, Gallagher RJ, Silvern D, Wuu CS, Zaider M. Treatment planning for brachytherapy: An integer programming model, two computational approaches and experiments with permanent prostate implant planning. Phys Med Biol 1999;44:145–165.

17.Lee EK, Zaider M. Mixed integer programming approaches to treatment planning for brachytherapy. Ann Operat Res Optimizat Med Ann Operat Res 2002;119:147–163.

18.Lee EK, Zaider M. Intraoperative dynamic dose optimization in permanent prostate implants. Int J Radiat Oncol Biol Phys 2003;56:854–861.

19.Silvern DA. Automated OR prostate brachytherapy treatment planning using genetic optimization. 1998.

20.Nag S, Shasha D, Janjan N, Petersen I, Zaider M. The American Brachytherapy Society recommendations for brachytherapy of soft tissue sarcomas. Int J Radiat Oncol Biol Phys 2001;49:1033–1043.

21.Zelefsky MJ, Cohen G, Zakian KL, Dyke J, Koutcher JA, Hricak H, Schwartz L, Zaider M. Intraoperative conformal optimization for transperineal prostate implantation using magnetic resonance spectroscopic imaging. Cancer J 2000;6:249–255.

22.Potters L. Permanent prostate brachytherapy: Lessons learned, lessons to learn. Oncol-New York 2000;14:981– 991.

23.Cha CM, Potters L, Ashley R, Freeman K, Wang XH, Waldbaum R, Leibel S. Isotope selection for patients undergoing prostate brachytherapy. Int J Radiat Oncol Biol Phys 1999;45:391–395.

24.Ling CC, Li WX, Anderson LL. The relative biological effectiveness of I-125 and Pd-103. Int J Radiat Oncol Biol Phys 1995;32:373–378.