Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5196_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
62 Мб
Скачать
CHAPTER 15 Management of Imaging Personnel Radiation Dose
297
Fig. 15.2 (A) A lead apron protects occupationally exposed personnel from scattered radiation. (B) A lead
mobile x-ray barrier of 0.5 or 1.0 mm lead equivalent provides protection from scattered radiation. It may be used during special procedures, in the operating room, and in cardiac units.
Depending on the energy range of the radiation for a specific procedure, an apron containing the lead equiva­lent of 0.5 mm or an apron containing the minimum re­quired lead equivalent of 0.25 mm may be sufficient for use. The standard 0.5 mm lead equivalent apron, which has traditionally been worn during routine fluoroscopic procedures, weighs 3 to 7 kg. In contrast, the 0.25 mm minimum lead equivalent apron can weigh 1 to 5 kg.2 Some physical attributes of protective lead aprons, includ­ing the percentage of x-ray attenuation at selective peak kilovoltages (kVps), are listed in Table 15.1.
If any personnel could have the posterior surface of their body turned toward the x-ray source during a ra­diologic procedure, a wraparound style apron would be desirable to afford the best protection. When taking into consideration both the amount of protection provided
BA
TABLE 15.1 Physical Attributes of
Protective Lead Aprons
PERCENTAGE X-RAY ATTENUATION
KILOVOLTS AT
Lead Equivalent Thickness (mm)
0.25 1–5 97 66 51
0.50 3–7 99.9 88 75
1.00 5–12 99.9 99 94 At 100 kVp, x-ray attenuation for a 0.50-mm lead equiv-
alent apron and a 1-mm lead equivalent apron is 75% and 94%, respectively.
Modified from Bushong SC: Radiologic science for technolo- gists: physics, biology and protection, ed 11, St. Louis, 2017, Elsevier.
Weight
(kg)
PEAK
50 75 100
298
CHAPTER 15 Management of Imaging Personnel Radiation Dose
by an apron and its weight, the 0.5 mm lead equivalent apron provides a good compromise for general use.
All protective apparel must be stored correctly when not in use to preserve the integrity of the garment. Lead aprons should be hung on racks or draped over a bar designed for storage to prevent unnecessary damage. They are never to be folded or crunched up in any fashion because this will lead to cracks or breaks in the lead­impregnated material, thereby compromising the device’s effectiveness for protection from radiation. Regulatory requirements state that all aprons be inspected annually for cracks or other defects either by fluoroscopy or by imaging the apparel with a high kVp technique.

Technical Exposure Factors

Technical exposure factors can influence the quantity of scattered radiation produced and thereby reaching im­aging personnel. For lower kVps, more mA is needed to secure a high-quality image, and therefore more signifi­cant amounts of low-energy photons are present. These characteristics of the x-ray beam lend themselves to the production of increased large-angle scatter radiation. Conversely, higher kVp techniques:
• Increase the mean energy of the photons comprising
the radiographic beam, leading to a decrease in large
angle scatter
• Require lower incident photon beam intensity (i.e.,
lower milliampere-seconds [mAs])
Therefore, with higher selected kVp values, less side­scattered radiation is available to strike imaging personnel, and the potential EqD is reduced.

Patient Restraint

Radiographers must never stand in the primary (useful) beam to restrain a patient during a radiographic exposure (Fig. 15.3A). When patient restraint is necessary, me- chanical restraining devices should be used to immobilize the patient whenever possible. If mechanical means of restraint are not feasible, nonoccupationally exposed per­sons, wearing appropriate protective apparel, are to per­form this function. These individuals should be posi­tioned so that their lead-protected torsos are not struck by the primary, or direct, beam (Fig. 15.3B). Holding patients may be necessary when they are unable to support them­selves. For example, a weak elderly male or female patient may be unable to stand without assistance and raise both arms above the head for a lateral chest x-ray examination. In this situation, a nonoccupationally exposed person
(relative or friend) equipped with a lead apron can hold the patient in position during the exposure. A mechanical restraining device is often used to hold an infant in the upright position for chest images to be obtained. If such a device is not available, the child has to be physically held (usually by a parent) during the exposure. Pregnant women, however, are never to be permitted to assist in holding a patient at that time.

PROTECTION FOR PREGNANT PERSONNEL

Imaging Department Protocol

Pregnant staff members should be able to continue per­forming their duties without interruption of employ­ment if they follow established radiation safety practices. Most health care facilities have policies for protecting pregnant personnel from radiation. Under these policies, an imaging professional who becomes pregnant first in­forms her supervisor. After this voluntary declaration has been made, the health care facility officially recog­nizes the pregnancy. The facility, through its radiation safety officer:
• Provides essential counseling
• Furnishes an appropriate additional radiation dosim­eter for monitoring of any possible radiation exposure to the embryo-fetus This additional dosimeter is to be worn at the waist
level during all radiation procedures. When a protective lead apron is used, the dosimeter should be worn at waist level beneath the apron. The purpose of this additional monitor is to ensure that the monthly EqD to the embryo-fetus does not exceed 0.5 mSv. This EqD limit excludes:
• Medical radiation
• Natural background radiation This practice is designed to significantly lower the
total lifetime risk of leukemia and other malignancies in persons exposed in utero.

Acknowledgment of Counseling and Understanding of Radiation Safety Measures

After receiving radiation safety counseling, the preg­nant radiologic technologist must read and sign a form acknowledging that she has received counseling and understands the practices to be followed to ensure the safety of the embryo-fetus. For monitoring of pregnant personnel, a separate monthly report is provided to
CHAPTER 15 Management of Imaging Personnel Radiation Dose
299
Fig. 15.3 (A) The radiographer should never stand in the primary (useful) beam to restrain the patient. (B) A
nonoccupationally exposed person restraining a patient during a radiographic exposure should wear a lead apron, gloves, and thyroid shield and stand outside the primary beam.
specifically document the exposure of the worker and the embryo-fetus. A copy of this report is sent to the facility’s radiation safety officer.

Protective Maternity Apparel

Protective maternity apparel, when needed, should be available for pregnant radiologists and radiographers. Specially designed maternity protective aprons consist of 0.5 mm lead equivalent over their entire length and width. They have an extra 1 mm lead equivalent protec­tive panel that runs transversely across the width of the apron to provide added safety for the embryo-fetus.
Wraparound protective aprons of 0.5 mm lead equivalent can also be used during pregnancy. The over­all physical size of the apron must be appropriate for the pregnant worker to ensure safety and provide reason­able comfort.
BA

Work Schedule Alteration

In accordance with ALARA guidelines, work schedules are designed to distribute radiation exposure risk evenly to all employees. If a declared pregnant radiog­rapher is reassigned to a lower radiation exposure risk area (e.g., removed from interventional fluoroscopy and assigned to general radiography), then the re­maining radiographers in the higher-risk area who must fill in can be subject to increased risk. Therefore, the declared pregnant radiographer does not necessar­ily need to be reassigned to a lower radiation exposure position as a direct consequence of a declared preg­nancy. However, it is imperative that, while remaining in her current position, the EqD to the embryo-fetus does not exceed the NCRP recommended monthly EqD limit of 0.5 mSv or a total of 5.0 mSv during the entire pregnancy.
300
CHAPTER 15 Management of Imaging Personnel Radiation Dose

BASIC PRINCIPLES OF RADIATION PROTECTION FOR PERSONNEL EXPOSURE REDUCTION

As previously stated, the three basic principles of radia­tion protection are:
• Time
• Distance
• Shielding Occupational radiation exposure of imaging personnel
can be minimized by the use of these cardinal principles.

Time

The amount of radiation a worker receives at a particular location is directly proportional to the length of time the individual is in the path of ionizing radiation. During fluoroscopy, the reduced exposure time will decrease both:
• Patient exposure
• Personnel exposure For this reason, fluoroscopic x-ray units are equipped
with 5 minute timers to alert the radiologist or other authorized equipment operator that a specific amount of time has elapsed. To minimize radiation exposure, a radiographer should be present in a fluoroscopy room only when needed to perform appropriate patient care and to fulfill the duties associated with the procedure. Otherwise, the radiographer should remain behind a protective barrier.

Distance

Distance is the most effective means of protection from
ionizing radiation. Because there is a significant de­crease in the radiation level as a consequence of the dispersion or spread of the radiation beam with dis­tance, imaging personnel will receive significantly less radiation exposure by standing farther away from a source of radiation.
More
distance
2 d
3 d
4 d
Fig. 15.4 As the distance between the source of radiation and
any given measurement point increases, radiation intensity (quantity) measured at that point decreases by the square of the relative change in distance between the new location and the old.
Less
intensity
(quantity of
radiation)
1
4 intensity
1
9 intensity
1
16 intensity
To be more explicit, as the distance between the ra­diation source and a measurement point increases, the intensity, or quantity of radiation measured at the more distant position decreases by the square of the ratio of the original distance from the source to the new distance from the source (Fig. 15.4). This lowering of radiation intensity physically occurs because the total x-rays emit­ted are spread over a new area that has increased by the square of the relative distance change. For example, when the distance from the x-ray target, a point source* of radiation, is doubled, the radiation at the new location spans an area four times larger than the original area. However, because the same amount of radiation exists to cover this larger area, the intensity at the new distance consequently decreases by a factor of four (Fig. 15.5).
The formula for ISL is shown in the equation in
Box 15.2. A mathematical example is also provided. The
inverse square law should be used whenever possible to reduce the radiographer’s exposure from sources of
Application of the Inverse Square Law. When light is
emitted from a source, such as a flashlight, the intensity decreases quickly with the distance from the source. X-rays act in the same manner, as the distance from the source increases, the intensity decreases. The inverse
square law (ISL) expresses the relationship between dis-
tance and intensity (quantity) of radiation and is an im­portant tool for limiting the dose received by personnel. The law is simply stated as: “The intensity of radiation is inversely proportional to the square of the distance from the source.”
*Point source: To correctly treat finite-sized radioactive sources as “point sources” so as to facilitate the calculation of exposure rates at points of interest, it is necessary that the distance of such points from the radioactive source be at least equal to 10 times the largest dimension of the source. As an example of this concept, consider a spherical radioac­tive source whose diameter is 2.5 cm. Then the smallest distance away from this source at which it may be accurate enough to regard it mathematically as a “point source” will be 25 cm.
CHAPTER 15 Management of Imaging Personnel Radiation Dose
301
Point
source of
x-rays
d
At d (1 m),
area x
Fig. 15.5 When the distance from a point source of radiation
is doubled, the radiation at the new location spans an area four times larger than the original area. However, the intensity at the new distance is only one-fourth of the original intensity.
x
d
x
2
At 2d (2 m),
area 4x
2x
2x
2
BOX 15.2 Inverse Square Law Formula
and Example
2
II(d )
122
5
2
(d )
1
where I1 expresses the exposure (intensity) at the origi­nal distance, I new distance, d the source of radiation, and d tance from the source of radiation.
Example: If a radiographer stands 1 m away from an
x-ray tube and is subject to an exposure rate dose* of 2 mGy
a
moves to a position located 2 m from the x-ray tube?
Answer:
*Exposure rate dose, given in units of mGya per hour, is the same quantity as air kerma rate.
expresses the exposure (intensity) at the
2
expresses the original distance from
1
per hour, what will it be if the same radiographer
2I4
(Cros s-multiply )
1
2
I 0.5 mGy /hr
2 a
expresses the new dis-
2
2
II(d )
122
2
(d )
1
2
2I(2)
2
(1)
2
4I 2
2
x-radiation. (This law also may be applied to sources of gamma and neutron radiation.)
Conversely, the ISL also implies that if a radiogra­pher moves closer to a source of radiation, this indi­vidual’s radiation exposure can dramatically increase.
For example, if the radiographer stands 2 m away from an x-ray source instead of 6 m away, the radiog­rapher’s radiation exposure increases by a factor of (6/2)2 5 9.

Shielding

When it is not possible to use the principles of time and/ or distance to minimize occupational radiation exposure,
shielding of appropriate thickness may be used to provide
adequate protection from radiation. The most common materials used for structural protective barriers are:
• Lead
• Concrete Accessory protective devices are made of lead-
impregnated vinyl. These accessory devices include:
• Aprons
• Gloves
• Thyroid shields
• Protective eyeglasses This apparel is to be used when it is not possible to
remain wholly behind either a stationary or movable protective barrier. The ability of materials to attenuate radiation depends on their atomic number, density, and thickness.
Protective Structural Shielding. Structural barriers
such as walls and doors in an x-ray room have been designed to provide radiation shielding for both:
• Imaging department personnel
• The general public These barriers are necessary to ensure that occupa-
tional and nonoccupational annual EfD limits are not exceeded. Lead impregnated drywalls of appropriate thickness are used in the walls of the radiography or fluoroscopy room to provide adequate shielding. A qualified medical physicist determines the exact lead requirements for a particular imaging facility. Although radiographers should understand the concept of shield­ing, they are not responsible for determining barrier thickness.
Primary protective barrier. The purpose of a pri-
mary protective barrier is to prevent direct, or unscat-
tered, radiation from reaching personnel or members of the general public on the other side of the barrier. The primary beam consists of the x-ray photons that follow straight-line paths through all sets of collimator shut­ters. Primary protective barriers are located perpendic­ular to the undeflected line of travel of the x-ray beam (Fig. 15.6).
302
CHAPTER 15 Management of Imaging Personnel Radiation Dose
Secondary
protective
barrier
Fig. 15.6 Protective barriers are lined with lead to protect per-
sonnel and the general public from radiation. The primary pro­tective barrier is located perpendicular to the undeflected line of travel of the x-ray beam. The walls that are not in the direct line of travel of the primary beam are called secondary protec- tive barriers because they are designed to shield only against secondary (leakage and scattered) radiation.
0.8 mm (
Primary
protective
barrier
1
32-inch) rolled
lead sheet
Leakage radiation
Collimator
Primary (useful)
beam
Scattered
radiation
Exit or image
formation
radiation
X-ray tube
(peak energy
130 kVp)
1.5 to 2.1 meters (5 to 7 feet)
1.6 mm (1⁄16-inch) rolled lead sheet
If the peak energy of the beam is 120 kVp, the pri-
mary protective barrier in a typical installation:
• Contains of 1.6 mm (1/16 inch) lead
• Extends 2.1 m upward from the floor of the x-ray room, when the x-ray tube is 1.5 to 2.1 m from the wall in question
Secondary protective barrier. Secondary radiation
consists of radiation that has been deflected from the primary beam. Leakage from the tube housing (photons that pass through the housing) and scatter (primarily from the patient) make up the secondary radiation. A
secondary protective barrier protects against leakage
and scatter radiation. Any wall or barrier that is never struck by the primary x-ray beam is classified as a secondary barrier (see Fig. 15.6). This does not mean that secondary radiation cannot strike primary barriers as well. A secondary barrier should overlap the primary protective barrier by approximately 1.3 cm (1/2 inch).
In a typical installation, the secondary barrier consists of 0.8 mm (1/32 inch) of lead.
Radiographic and fluoroscopic exposures should be made only when the doors to x-ray rooms are closed. This practice affords a substantial degree of protection for persons in areas adjacent to the room door because in most facilities room doors have attenuation for di­agnostic energy x-rays equivalent to that provided by
0.8 mm (1/32 inch) of lead.
Control-booth barrier. X-ray rooms housing per-
manent radiographic equipment contain a control-
booth barrier for the protection of the radiographer.
This barrier must:
• Extend at least 2.1 m upward from the floor
• Be permanently secured to the floor
Diagnostic x-rays should scatter a minimum of two times before reaching any area behind this barrier. Be­cause this booth is situated so that it intercepts leakage and scattered radiation only, it may be regarded as a secondary protective barrier. To ensure maximum pro­tection during radiographic exposures, personnel must remain entirely behind the barrier. The radiographer may observe the patient through the lead glass window* in the booth (Fig. 15.7). This window typically consists of 1.5 mm (1/16-inch) lead equivalent. With the appro­priate degree of shielding in the barrier, the radiogra­pher’s exposure will not exceed a maximum allowance of 1 mSv (100 mrem) per week; in actual practice in a well­designed facility, exposure should not exceed 0.02 mSv (2 mrem) per week. For further protection, the exposure cord, if one is present, must be short enough that the exposure switch can be operated only when the radiog­rapher is completely behind the control-booth barrier.
Clear lead–acrylic secondary protective barrier.
Clear lead–acrylic material impregnated with approxi­mately 30% lead by weight may be fashioned into an effective secondary protective barrier, such as for the control booth (Fig. 15.8). This creates a modern appear­ance for the facility and permits a panoramic view, allow­ing diagnostic imaging personnel to observe the patient more completely. Modular or movable x-ray barriers:
• Are shatter-resistant
• Can extend 2.1 m upward from the floor
• Are available in lead equivalency from 0.3 to 2 mm
* The wall barrier of the control booth should be at least 46 cm (18 inches) beyond the edge of the view window.
CHAPTER 15 Management of Imaging Personnel Radiation Dose
303
Fig. 15.7 While making a radiographic exposure with a station-
ary radiographic unit, the radiographer must remain completely within the control-booth barrier (behind the fixed protective bar­rier) for safety. The radiographer may observe the patient through the lead glass observation window in the control booth.
Clear lead–acrylic overhead protective barrier.
Clear lead–acrylic protective barriers also can be used as overhead x-ray barriers providing an open view during special procedures and cardiac catheterization (Fig. 15.9). This shielding typically offers 0.5 mm lead equivalency protection.
Accessory Protective Devices. Accessory protective
shielding includes aprons, gloves, and thyroid shields made of lead-impregnated vinyl. These protective garments are available in a variety of:
• Shapes
• Sizes
• Thicknesses As lead equivalent thickness increases, attenuation
of the x-ray beam also increases when kVp remains the same.
Fig. 15.8 A clear lead acrylic secondary protective barrier
impregnated with approximately 30% lead lends a modern appearance to the facility.
Fig. 15.9 A clear lead acrylic overhead protective barrier used
during special procedures and cardiac catheterization. (From Fluke Biomedical.)
304
Fig. 15.10 A lead apron, gloves, and thyroid shield protect the
radiographer from scattered radiation.
CHAPTER 15 Management of Imaging Personnel Radiation Dose
Requirements for lead aprons and gloves. If the ra-
diographer’s hands will be near the x-ray beam, leaded gloves should be used. A suitable lead apron is to be worn whenever the radiographer cannot remain behind a pro­tective barrier during an exposure (Fig. 15.10). Histori- cally, from regulatory doctrine, if the peak energy of the x-ray beam was 100 kVp, then a protective apron’s attenu­ation must be equivalent to at least a 0.25 mm thickness of lead. An apron of 0.5 mm lead equivalent, however, af­fords much higher security and is the most widely used and recommended thickness in diagnostic imaging and is the minimum lead equivalent required for a protective garment worn by occupationally exposed individuals dur­ing fluoroscopic or interventional procedures. Regardless of the regulatory mention of 0.25 mm thicknesses of lead for some purposes, the need for 0.5 mm lead equivalent for fluoroscopy and interventional cases and the recom­mendations that 0.5 mm lead aprons are desirable for all purposes have prompted most facilities to stock 0.5 mm lead aprons only. This eliminates the possibility of person­nel inadvertently selecting the wrong apron. Therefore
0.5 mm has become the all-purpose apron of choice and in many cases, should also be in a wraparound style. A lead apron with 0.25 mm of lead is, however, very appro­priate for use in mammography.
Neck and thyroid shield. A neck and thyroid shield
(Fig. 15.11) are employed to guard the thyroid area of occupationally exposed personnel during:
• General fluoroscopy
• X-ray special procedures The neck and thyroid shield should be a minimum
of 0.5 mm lead equivalent.
Fig. 15.11 The neck and thyroid gland can be protected from
radiation exposure through the use of a 0.5-mm lead equivalent protective shield.
Fig. 15.12 Eyeglasses protect the lens of the eyes during gen-
eral fluoroscopy and special procedures. (Shown are glasses with wraparound frames; other styles are also available.)
Protective eyeglasses. Scatter radiation to the lens
of the eyes of diagnostic imaging personnel can be sub­stantially reduced by the use of protective eyeglasses (Fig. 15.12), fitted with optically clear lenses that con­tain a minimal lead equivalent protection level of 0.35 mm. Side shields on the glasses are also available and useful for procedures that require turning of the head. A wraparound frame containing optically clear lenses with 0.5 mm lead equivalent may also be acquired.

X-RAY TUBE HOUSING CABLES

While the x-ray tube housing is massive enough to pro­vide a significant degree of shielding from secondary
CHAPTER 15 Management of Imaging Personnel Radiation Dose
305
radiation, it is also designed to protect the operator from the hazard of electric shock. Because of this, the radiographer must be both observant* and careful when handling this piece of equipment and its adjoining part, the collimator. While manipulating the tube housing assembly for a radiographic examination, the radiogra­pher should avoid rough handling or severely bending the high-tension cables that connect to the positive and negative terminals of the x-ray tube. No one should ever
touch the tube housing or high-tension cables while a radiographic exposure is in progress.

PROTECTION DURING FLUOROSCOPIC PROCEDURES

Personnel Protection

To ensure protection from scattered radiation emanat­ing from the patient during a fluoroscopic examination, the radiographer should:
• Stand as far away from the patient as is practical
• Move closer to the patient only when assistance is required A protective apron of at least 0.5 mm lead equivalent
must be worn during all fluoroscopic procedures. Pro­tective lead gloves of at least 0.25 mm lead equivalent should be worn whenever the hands must be placed near the fluoroscopic field (Fig. 15.13). Imaging personnel
*Observant means looking at the physical condition of the cables and cable coverings to discern undue wear or stress effects that could lead to a serious hazard.
assisting during a fluoroscopic examination also should wear thyroid shields of 0.5 mm lead equivalent, espe­cially if they are standing close to the patient being ex­amined (Fig. 15.14). If immediate assistance during a fluoroscopic examination is not required, the radiogra­pher should either stand behind the radiologist, who is also wearing protective apparel, or stand behind the control-booth barrier until services are required. A wrap-around protective apron is recommended to pro­tect personnel who must move around the x-ray room during a fluoroscopic examination.

Dose-Reduction Techniques

Many of the methods and devices that reduce the ra­diographer’s exposure when operating stationary (fixed) radiographic equipment also reduce the dose received during a fluoroscopic procedure. These methods and devices include:
• Adequate beam collimation
• Adequate filtration
• Control of technical exposure factors
• Appropriate source-to-skin distance
• Diagnostic-type protective x-ray tube housing To ensure adequate protection for both the radiogra-
pher and the radiologist, some additional requirements are included in the federal government specifications for the use of fluoroscopic equipment. An example is use of a cumulative timing device that produces an audible signal after 5 minutes of total beam-on time has been exceeded.
Thyroid shield
Protective
Protective curtain,
or sliding panel
Fig. 15.14 Scattered radiation produced during a fluoroscopic
examination can be absorbed by a protective curtain or sliding panel, with a minimum of 0.25 mm lead equivalent placed be­tween the fluoroscopist and the patient.Fig. 15.13 Lead gloves.
lead
apron
306
CHAPTER 15 Management of Imaging Personnel Radiation Dose

Remote-Control Fluoroscopic Systems

The remote-control unit provides imaging personnel with the best radiation protection opportunity. Remote­control systems permit the radiologist, and assisting radiographer, to remain at a control console located behind a protective barrier until their presence within the room is needed. This system also further improves imaging personnel safety because the added distance from the x-ray tube makes use of the inverse square law.

Protective Curtain

A protective curtain, or sliding panel, with a minimum of
0.25 mm lead equivalent should typically be positioned between the fluoroscopist and the patient to intercept scattered radiation above the tabletop (see Fig. 15.14).

Bucky Slot Shielding Device

A Bucky slot shielding device of at least 0.25 mm lead equivalent must automatically cover the Bucky slot open­ing in the side of the x-ray table during a standard fluo­roscopic examination when the Bucky tray is positioned at the foot end of the table (Fig. 15.15). This shielding device protects the radiologist and radiographer at the gonadal level. Without this device and the protective cur­tain in place, the exposure dose rate to the fluoroscopist could markedly exceed 1 mGya/hr at a standard distance of 0.6 m from the side of the x-ray table.

Rotational Scheduling of Personnel

Diagnostic imaging personnel are potentially subjected to the highest occupational exposure during:
• Fluoroscopy: fixed and mobile
• Mobile radiography
• Special procedures
• Interventional surgery Scheduling radiographers to spend less time in these
higher radiation tasks by arranging assignments to clinical areas in a rotational pattern can decrease this exposure. This practice, therefore, uses the cardinal safety principle of time as a means of additional radia­tion protection.

PROTECTION DURING MOBILE X-RAY EXAMINATIONS

Use of Protective Garments

Mobile radiographic systems create special radiation pro­tection considerations for the radiographer. Some states require radiographers to wear lead aprons whenever they are performing mobile radiographic or fluoroscopic ex­aminations. A protective apron should be assigned to each mobile unit so that it is immediately available for the radiographer.

Distance as a Means of Protection

The most recent mobile units are equipped with a re­mote-control exposure device. This permits the radiog­rapher to leave the immediate vicinity and uses the cardinal principle of distance as an effective means of protection from radiation. Most mobile units are not remotely controlled. For those units the cord leading to the exposure switch must be long enough to permit the radiographer to stand at least 2 m from the:
• Patient
• X-ray tube
• Useful beam
Head end of radiographic
table during
fluoroscopy
Bucky slot
shielding
device
Fig. 15.15 To provide protection at the gonadal level for the fluoroscopist, the Bucky slot shielding device
should be at least 0.25 mm lead equivalent.
Foot end of radiographic table during fluoroscopy
Bucky tray