Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5196_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
62 Мб
Скачать
CHAPTER 14 X-Ray Breast Imaging: Methods and Radiation Safety Aspects
Fig. 14.10 Operational views of a DBT system. (From Siemens.)
TABLE 14.1 Physical Characteristics of Five Manufacturer’s Digital Breast Tomosynthesis
10
Units
Manufacturers: Fuji GE Hologic IMS Siemens
Anode Material: W Mo or Rh W W W Filter Material: Al or Rh Mo or Rh Ag Ag Rh Detector: a-Se FPD* CsI FPD** a-Se FPD a-Se FPD a-Se FPD Pixel size (mm): 150/100 100/50 100/140 85 85 Pixel shape: Hexagonal Square Square Square Square Tube motion: Continuous Step and shoot Continuous Step and shoot Continuous Sweep angle (°): 15/ 40 25 15 40 50 No. of projections: 15 9 15 13 25 Dose/projection: Uniform Uniform Uniform Variable Uniform Antiscatter grid: No Yes No No No
*amorphous selenium flat panel detector **Cesium iodide flat panel detector
287
BOX 14.1 Siemens Mammomat DBT
Twenty-five views are acquired in an angular range from
-25° to 125°. The exposure release button on the control box or on the foot or hand switch must be pressed and held during all exposures. Following proper positioning and compression of the patient’s breast, the steps of the image acquisition process are:
1. Initially, the swivel arm is in the 0° position.
2. The first view is acquired for automatic exposure control
settings.
3. Swivel arm then moves to -25° and a second exposure
is taken.
4. The swivel arm subsequently covers the entire angular
range from -25° to 125° while an exposure is taken at every 2° for a total of 50 projections.
5. Swivel arm returns to the start position.
288
Face shield
Compression plate
Fig. 14.11 Closeup of breast positioning and detector assemblies.
(From Siemens.)
CHAPTER 14 X-Ray Breast Imaging: Methods and Radiation Safety Aspects
To ensure radiation safety, a restricting face shield must
be used during tomographic examinations (Fig. 14.11).
The face shield is present because there is a risk that the patient’s head can drift into the x-ray beam path if not prevented from doing so. Note that the face shield also moves when the swivel arm moves. The patient’s head must therefore not lean against the face shield during the arc rotation.

RADIATION DOSAGE

The x-ray spectrum used for tomosynthesis is generally similar to that employed in standard digital mammog­raphy, and the selected peak kilovoltage (kVp) depends on the thickness and density of the compressed breast. If the x-ray energy is increased slightly, thereby sacrific­ing some degree of image quality, then it is possible to reduce the absorbed dose just by using less mA. An­other option, especially for thicker breasts, for reducing dose while maintaining image quality, is to have addi­tional filtering of the x-ray beam. This inherently in­creases the mean beam energy and consequent net ra- diation penetration. It can be achieved by selecting in place of the standard molybdenum/molybdenum (Mo/ Mo) or molybdenum/rhodium (Mo/Rh) x-ray target/ filter combinations, other pairings such as tungsten/ rhodium (W/Rh), tungsten/aluminum (W/Al), or tungsten/silver (W/Ag).
For mammography, a useful and practical implementa­tion of tomography was not possible until the development of digital flat panel wide field detectors which initiated full
field digital mammography (FFDM). Subsequently, FFDM
was coupled with tomographic motion. With this, any de­sired image slice and slice thickness could be computer recon­structed from the stored multiple angle projection images,
and all of this was obtained with just one angular sweep of the x-ray tube.
12,19
The breast is composed of three types of tissue: glan­dular, adipose (fatty tissue), and skin. Because statistics stemming from a large database assembled from breast cancer screenings by many institutions has repeatedly shown that the development of breast cancer in adipose tissue is rare, mammographic radiation dosimetry is pre­dominantly concerned with the dose deposited in the glandular tissue of the breast. Therefore, the parameter that has been chosen for both standard and 3D digital mammography to estimate and represent effective dose in x-ray breast imaging is the mean glandular dose (MGD).
MGD is used to characterize the absorbed dosage to the radiosensitive fibroglandular breast tissue. Quanti­tively, it is determined from measurements which em­ploy a special design (thin window circular parallel plate) ionization chamber to measure the air kerma or exposure incident on the breast or on a breast-equivalent phantom. To obtain the MGD from such measure­ments, the ion chamber readings must be multiplied by special conversion factors derived from data generated by complex randomized computer modeling of radia­tion interaction processes in the breast. These factors are referred to as the normalized glandular dose coeffi- cients. For each ion chamber measurement, there is a particular coefficient value that is directly associated with the x-ray beam quality (HVL) employed as deter­mined by the x-ray tube target/filter combination and the selected kVp. Tables for the latest determined coef­ficients are listed in the current American College of Radiology Mammography Quality Assurance Manual.
The angular range and number of exposures taken over the x-ray tube arc during the DBT scan are addi­tional variables that need to be optimized with respect to balancing patient radiation dose and acceptable im­age quality. In general, it would seem that taking more exposures during a DBT procedure will generate recon­structions with fewer artifacts. However, doing this must be weighed against the consideration that for a
CHAPTER 14 X-Ray Breast Imaging: Methods and Radiation Safety Aspects
289
patient receiving a full mammographic examination*, it is still desired to limit the total MGD to 3 mGy (0.3 cGy). More DBT angles will then mandate lesser irradiation per exposure and consequently smaller de­tector signals for each of the individual tomographic projections. For low enough exposures, image receptor inherent noise will start to become a non-negligible component of the raw image and thereby may notice­ably degrade reconstructed image quality. A greater number of exposures also increases raw data size which produces longer reconstruction times.
During a DBT acquisition process, the total delivered radiation dose is apportioned among the multiple single x-ray views with every projection contributing about only 5% to 10% of a normal single-view mammogram absorbed dose.
12,19
Because each image voxel is ulti- mately overall probed or irradiated by essentially the same number of X-ray quanta as in a standard 2D mam­mography acquisition, a quality tomosynthesis scan, in total, should deliver approximately the same MGD as a conventional mammogram. A major prerequisite for this to be achieved, however, is that the image receptor has a high detective quantum efficiency** (DQE). Rapid imaging processing is also another requirement for the DBT detector.
Usually it is assumed that within a normal breast there is a homogeneous mixture of adipose and glandu­lar tissue surrounded by a layer of skin. In reality, it is
*A total or full mammographic examination in a facility that has a DBT unit usually includes the standard stationary 2D x-ray projections as well as the DBT series of projections. **Detective quantum efficiency, or DQE, refers to how effi­ciently a detection system translates incident x-ray photons into a useful signal relative to random noise within that image. In medical radiography, the listed DQE describes percentage­wise the degree to which an x-ray imaging system can produce an image with a high signal-to-noise ratio (SNR) relative to that delivered by an ideal detector whose efficiency is by defi­nition taken to be 100%. This specification can also be consid­ered an alternate measure of the radiation dose efficiency of a detector, since the needed amount of radiation exposure to a patient decreases as the system DQE is increased for the same image SNR and physical exposure conditions. Compared with film/screen imaging, a digital detector with high DQE has the potential to deliver significant object-detectability improve­ments at the same equivalent dose or an identical degree of detectability at a lower patient dose.
BOX 14.2 Typical Mean Glandular
Dose Values From Breast Phantom Measurements for Individual and Combined DBT and Standard 2D Exposures
Radiation dose for a single standard 2D mammogram is
about 1.2 to 1.4 mGy or 1.3 mGy (130 mrads) on average.
Radiation dose for a 3D or DBT sequence of projec-
tions is about 1.3 to 1.5 mGy or 1.4 mGy (140 mrads) on average (no grid)
Total MGD for the combination is thus about 2.7 mGy
or 270 mrads. (ACR recommends that total MGD not exceed 3 mGy (300 mrads) per examination)
found that the glandular dose deposited in various re­gions can differ considerably and that, especially, is why the mean glandular dose was adopted as a useful and practical parameter of mammographic absorbed dose. Some typical numerical values for mammographic pro­cedures are shown in Box 14.2.

DBT SUMMARY

Breast tomosynthesis is a three-dimensional imaging technology that involves acquiring a dozen or more im­ages of a stationary compressed breast at multiple angles of incidence during a partial rotational scan of the x-ray tube about an axis located within the breast. With these multiple diverse x-ray projections or views, objects at varying depths in the breast will be projected onto differ­ent locations on the detector. At any desired depth in the compressed breast, a reconstructed plane can be obtained by using the Shift and Add method or other techniques to properly combine all registered x-ray projection views for that location. Thus, in a particular reconstructed or syn­thesized tomosynthesis plane, a structure that is actually located at the corresponding depth will be in-focus, whereas structures lying above or below that plane are blurred.
Because there is a practical and technical limit to the angular extent of the projection acquisitions, DBT does not have isotropic spatial resolution, i.e., there will be a very high spatial resolution within the planes parallel to the detector (x-y), and a considerably less fine plane by plane discrimination in the perpendicular direction (z-axis). The depth resolution, however, that has been achieved in present 3D systems is deemed to be good
290
CHAPTER 14 X-Ray Breast Imaging: Methods and Radiation Safety Aspects
enough to substantially reduce the 2D mammography is­sue of tissue superposition, thereby noticeably lowering its negative impact on sensitivity and specificity.
23
To keep the radiation dose to the breast from a single DBT procedure at levels comparable to a standard 2D mammography breast examination (e.g., a mediolateral view), the total x-ray tube radiation output is spread out over multiple projections, so that each individual pro­jection delivers only a small fraction of the radiation dose of a standard 2D mammogram. This necessitates the usage of detectors with a high DQE at low air kerma or exposure levels and has led to the development of detectors specifically designed for DBT imaging.*
*Besides meeting the ordinary requirements for 2D digital mam­mography, detectors for DBT need to have additional capabilities. These include: (i) faster reading time, to keep the total acquisition time of all projections to a minimum; (ii) minimal ghosting (the reduction of sensitivity caused by previous exposure history of the detector) and minimal lag (the carryover of signal from a previous image); and (iii) minimal reduction in detective quan­tum efficiency at much lower exposures (a consequence of the need to divide the total exposure over multiple projections).
23,24
Following irradiation, the individual images are reconstructed using sophisticated algorithms into a large number of thin (,1 mm) or much fewer slab-like (,1 cm) high-resolution slices that can be displayed individually or in a ciné mode. Note that additional acquisitions at a different focal distance are not re­quired to enhance the visibility of objects at any desired depth—one set of acquired data can be reprocessed to generate the entire 3D volume set. It should always be kept in mind that “there is no free meal in tomosynthesis -
imaging parameters are a trade-off between z-axis accuracy, spatial resolution and importantly, radiation exposure to the breast.”16 In conclusion, digital breast
tomosynthesis provides the following benefits relative to other modalities:
• Enhances conspicuity of abnormalities by minimiz-
ing the obscuring effects of overlying structures
• Permits cross-sectional imaging (i.e., planes of view)
with high resolution
• A lower radiation dose as compared to CT mam-
mography
• Lower cost as compared with CT and magnetic reso-
nance imaging

S U M M A R Y

• Nonpalpable breast cancer may be detected through mammography.
• Flattening and reducing breast thickness prior to a standard mammographic radiation examination is necessary for producing a sharper image and reduc­ing patient dose.
• Protocols founded upon compression pressure which takes into account breast surface area will be a more realistic guide to better and less uncomfortable mammograms.
• Two x-ray views of each breast (craniocaudal direction [CC] and mediolateral oblique direction [MLO]) are taken in a standard screening mammogram evaluation.
• The American College of Radiology, the American Cancer Society, and the American Medical Associa­tion advocate annual mammography screening or mammography screening at least every other year for women age 40 to 49.
• Federal regulations state that the mean dose to the glandular tissue (MGD) of a 4.5-cm compressed breast using a digital mammography system should not exceed 3 mGyt per view.
• Studies have shown that well-calibrated mammo­graphic systems are capable of providing excellent imaging performance with an average glandular dose of not more than 2 mGyt.
22
• Dose reduction in mammography can be achieved by limiting the number of anatomical projections taken.
• Axillary projections in mammography should only be done on request of the radiologist.
• Metallic elements such as molybdenum (Z 5 42) and rhodium (Z 5 45) are commonly employed as filters in mammography enabling lower incident x-ray energy ranges that are more effective for breast imaging.
• Beryllium (Z 5 4) takes the place of glass in the win­dow of the low-kVp mammographic x-ray tube.
• Digital mammography units with image gray-level manipulation offer visualization improvement for patients with dense breasts.
• The method used for generating an in-focus two­dimensional image of a slice or cross-section through a three-dimensional object is called Tomography.
CHAPTER 14 X-Ray Breast Imaging: Methods and Radiation Safety Aspects
291
• Standard digital mammography can lead to both false positives and false negatives because the mul­tiple tissues and structures of the compressed breast appear overlapped in the acquired projection image.
• Digital Breast Tomography (DBT) is an imaging technique, which generates multiple planar images acquired while the x-ray tube rotates within a limited arc above the patient’s breast.
• The introduction of full field digital detector mam­mography (FFDM) was the primary technical break­through that enabled the wide-spread usage of DBT.
• Because in Digital Breast Tomography a third axis (z-direction) is stereoscopically extracted from the overall two-dimensional data, DBT is often referred to as “3D’ mammography when mammographic im­ages are reconstructed at arbitrary angles from the many different x-ray projections.
• Each DBT arc acquisition sequence may typically consist of from 15 to 25 separate x-ray projections. From the data generated by these, many nonblurred planar views can be obtained either individually or in a continuous sequence thereby generating a “movie”.
• The ability of DBT to decrease both false positives and false negatives as well as better image evalua­tion of dense breasts is due mainly to the enhanced z-axis resolution achievable by the tomographic method.
• The conventional method of image reconstruction used in digital tomosynthesis is the Shift and Add Algorithm (SAA).
• Artifacts present in DBT are primarily due to patient motion, the method of image acquisition, and im­perfect z-axis or depth resolution.
• With DBT used alone, a compression force only great enough to securely retain the breast in a stable posi­tion during the procedure may suffice.
• For thicker breasts, a method of reducing patient dose while maintaining image quality is to employ addi­tional filtering of the x-ray beam. This can be achieved by selecting in place of the standard molybdenum/ molybdenum (Mo/Mo) or molybdenum/rhodium (MoRh) x-ray target/filter combinations, other pair­ings such as tungsten/rhodium (W/Rh), tungsten/ aluminum (W/Al), or tungsten/silver (W/Ag).

D I S C U S S I O N Q U E S T I O N S

1. What filters are recommended for use with a molyb-
denum anode when a mammographic examination is performed on a patient with larger or dense breasts? Why are they recommended?
2. Why is the age recommendation for screening mam-
mography so controversial?
3. What tissues in the human body compose breast tissue?
4. What is mean glandular dose (MGD) used for?
5. What does detective quantum efficiency (DQE) refer
to?

R E V I E W Q U E S T I O N S

1. Studies have shown that well-calibrated mammo-
graphic systems are capable of providing excellent imaging performance with an average glandular dose of not more than:
A. 10 mGy B. 7 mGy C. 5 mGy D. 2 mGy
t
t
t
t
6. How can contrast in the radiographic image be
enhanced during digital mammography?
7. What is digital tomosynthesis?
8. What are the advantages of digital breast tomosyn-
thesis over conventional digital radiography of the breast?
9. How do artifacts result from the motion of the
x-ray tube during DBT?
10. How can radiation dose for the patient be reduced
during DBT?
2. Digital mammography units, which can enhance
contrast with image gray-level manipulation, offer substantial improvement for patients with:
A. Thin breast tissue B. Sparse breast tissue C. Porous breast tissue D. Dense breast tissue
292
CHAPTER 14 X-Ray Breast Imaging: Methods and Radiation Safety Aspects
3. Axillary projections of the breast should be done
only at the request of the:
A. Administrator of the imaging facility B. Patient C. Radiologist D. Radiographer
4. What material takes the place of glass in the window
of the low-kVp-producing mammographic x-ray tube?
A. Aluminum B. Beryllium C. Copper D. Lead
5. For DBT imaging the following is not true:
A. A lesser number of x-ray tube angular projections
leads to a thinner in-focus plane or improved z-axis resolution
B. Increasing the number of x-ray projections for
image reconstruction increases the blurring of out of plane structures for a particular plane
C. Using a smaller range of x-ray tube angles yields
an overall sharper breast image
D. All of the above
6. For DBT imaging the following is true:
A. The x-ray portion of the examination procedure
will last at least an hour
B. Because of the movement of the gantry it is nec-
essary that there be greater breast compression than for 2D mammography
C. The mean glandular radiation dose will exceed
3 mGy because of the many x-ray exposures taken during the examination
D. None of the above
7. With regard to spatial resolution of breast imaging,
which of the following is correct:
A. DBT is far inferior to CT in x-y plane image
resolution
B. The smaller the DQE value of the unit’s detector,
the better will be the quality of the mammo­graphic image
C. Increasing the number of x-ray tube angular
projections can lead to a more accurate posi­tional display of microcalcifications
D. All the above
8. A discrete image area element is termed a:
A. gantry B. kerma C. pixel D. voxel
9. A 3D image technology that involves acquiring a
dozen or more images of a stationary compressed breast at multiple angles of incidence during a par­tial rotational scan of the x-ray tube about an axis located within the breast is called:
A. Breast tomosynthesis B. 2D mammography C. Halo tomography D. Standard digital mammography
10. Which of the following can reduce dose in standard
mammography?
A. Increasing kVp to a minimum of 70 B. Not applying compression to the breast during
imaging
C. Limiting the number of projections taken D. Removing filtration from the x-ray tube in a
dedicated mammographic unit
Management of Imaging Personnel Radiation
Dose During Diagnostic X-Ray Procedures

O B J E C T I V E S

After completing this chapter, the reader will be able to perform the following:
• Define all key terms.
• State the annual occupational effective dose limit for whole-body exposure of diagnostic imaging personnel during routine operations.
• Explain why occupational exposure of diagnostic imaging personnel is limited and state the reason for allowing a much larger equivalent dose for radiation workers than for the population as a whole.
• Identify the type of x-radiation that poses the most significant occupational hazard in diagnostic radiology and explain the various ways this hazard can be reduced.
• Explain how the various methods and techniques that reduce patient exposure during a diagnostic examination can also minimize exposure for the radiographer and any other personnel.
• Discuss the responsibilities of the employer for protecting declared pregnant diagnostic imaging personnel from radiation exposure.
• Describe the three underlying principles of radiation protection that can be used for personnel exposure reduction.
• State and explain the inverse square law, and solve mathematical problems applying this concept.
15
• Differentiate between a primary and a secondary protective barrier, and list examples of such barriers.
• Describe the construction of protective structural shielding.
• Discuss the protective garments that may be worn to reduce whole-body or partial-body exposure.
• List and explain the methods and devices that may be used to reduce exposure for personnel during routine fluoroscopic and interventional examinations.
• Specify the techniques that are useful for reducing the radiographer’s exposure when performing a mobile radiographic examination.
• Explain the variation in dose rate caused by scatter radiation near the entrance and exit surfaces of the patient during C-arm fluoroscopy and discuss dose reduction methods for C-arm operators.
• List the three categories of radiation sources that may be generated in an x-ray room, list the considerations on which the design of radiation-absorbent barriers should be based, and explain the importance of each.
• Differentiate between a controlled area and an uncontrolled area.
• Discuss current approaches to shielding design.
• Describe radiation caution signage.
C H A P T E R O U T L I N E
Annual Limits for Occupationally Exposed Personnel
Effective Dose Limits Annual Occupational and Nonoccupational Dose Limits Allowance for a Larger Equivalent Dose for Radiation
Workers
ALARA Concept Dose-Reduction Methods and Techniques
Repeats in Digital Imaging
The Patient as a Source of Scattered Radiation Scattered Radiation—Occupational Hazard Filtration of the Diagnostic X-Ray Beam Protective Apparel Technical Exposure Factors Patient Restraint
Protection for Pregnant Personnel
Imaging Department Protocol
293
294
CHAPTER 15 Management of Imaging Personnel Radiation Dose
Acknowledgement of Counseling and Understanding
of Radiation Safety Measures Protective Maternity Apparel Work Schedule Alteration
Basic Principles of Radiation Protection for Personnel
Exposure Reduction
Time Distance Shielding
X-Ray Tube Housing Cables Protection During Fluoroscopic Procedures
Personnel Protection Dose-Reduction Techniques Remote Control Fluoroscopic Systems Protective Curtain Bucky Slot Shielding Device Rotational Scheduling of Personnel
Protection During Mobile X-Ray Examinations
Use of Protective Garments Distance as a Means of Protection
Protection During C-Arm Fluoroscopy
Personnel Exposure Resulting From Scattered
Radiation

K E Y T E R M S

Broad-beam x-ray transmission
factor (B)
Bucky slot shielding device control-booth barrier controlled area cumulative effective dose
(CumEfD) limit
distance inverse square law (ISL) leakage radiation occupancy factor (T) primary protective barrier primary radiation scatter radiation
Need for Protective Apparel for all Personnel and
Monitoring of Imaging Personnel Positioning of the C-Arm Fluoroscope Exposure Reduction for Personnel
Protection During High-Level Control Interventional
Procedures
Increased Importance of Radiation Safety Technique Knowledge of Dose Reduction Techniques Required
by the Radiographer How the Radiologist or Other Interventional Physician
Can Reduce Radiation Exposure Extremity Monitoring
Diagnostic X-Ray Suite Protection Design
Requirement for Radiation-Absorbent Barriers Reason for Overshielding Radiation Shielding Categories Calculation Considerations Calculating Barrier Shielding Requirements Current Approaches to Shielding
Radiation Caution Signs
Beam-On Indicator Sign General Posting
Summary
secondary protective barrier shielding time uncontrolled area use factor (U) workload (W)
Some x-ray procedures increase the radiographer’s risk of exposure (Box 15.1) due to scatter radiation. This chapter presents an overview of methods that can be used to reduce exposure for imaging profes­sionals during diagnostic x-ray procedures. Also, a
BOX 15.1 Imaging Procedures That
Increase the Radiographer’s Risk of Exposure
General fluoroscopy
Interventional procedures that employ high-level control
fluoroscopy (HLCF)
Mobile examinations
C-arm fluoroscopy
brief explanation of the design of a diagnostic x-ray suite is presented.

ANNUAL LIMIT FOR OCCUPATIONALLY EXPOSED PERSONNEL

Effective Dose Limits

Federal government standards, following a recommenda­tion of the National Council on Radiation Protection and Measurements (NCRP), permit diagnostic imaging per­sonnel to receive an “annual occupational effective dose (EfD) of 50 millisievert (mSv)”1 for whole-body exposure during routine operations. However, in keeping with the ALARA (as low as reasonably achievable) philosophy and careful supervision of personnel cumulative radiation
CHAPTER 15 Management of Imaging Personnel Radiation Dose
295
exposure records, no radiographer should approach this EfD level. The dose level referred to here, includes only occupational dose and does not include personal medical exposure that an employee may receive or the back­ground exposure that all people receive.
To ensure that the lifetime risk of occupationally exposed personnel is not exceeded, an additional rec­ommendation is that the lifetime EfD in mSvs should not exceed 10 times the person’s age in years. Hence a
cumulative effective dose (CumEfD) limit has been
established for the whole body.

Annual Occupational and Nonoccupational Effective Dose Limits

The annual occupational EfD limit of 50 mSv (5 rem) is an upper boundary limit. It is much higher than the an­nual EfD limit allowed for individual members of the general population not occupationally exposed. That limit is:
• 1 mSv (100 mrem) for continuous or frequent exposures
from artificial sources other than medical irradiation
and natural background radiation
• 5 mSv (500 mrem) for infrequent annual exposure
The 1 mSv annual EfD limit set for members of the general public is intended to limit that exposure to rea­sonable levels of risk that are comparable with risks from other familiar sources—i.e., about 1024 to 1026 annually1 (1024 to 1026 means an excess cancer risk of 1 chance in 10,000 to 1 chance in 1 million per year). The 5 mSv maximum annual EfD limit recommendation is made because annual exposures above the 1 mSv rec­ommendation, need not be regarded as especially haz­ardous, provided the average exposure to individuals in these groups does not exceed an average annual EfD of about 1 mSv.1 Both these limits will maintain the annual equivalent dose to organs and tissues below levels of concern for tissue reactions.
1
1
1

Allowance for a Larger Equivalent Dose for Radiation Workers

Valid reasons exist for permitting radiation workers to accumulate a larger equivalent dose (EqD). Among the most important of these reasons is that the workforce in radiation-related jobs is small when compared with the population as a whole. Therefore, the expectation of any measurable increase in disease in the popula­tion, in individuals, or impact upon the gene pool is negligible. Thus, the amount of radiation received by this workforce can be substantially greater than the
amount received by the general public without altera­tion in the genetically significant dose, the average annual gonadal EqD to members of the population who are of childbearing age. Although the radiogra­pher and other diagnostic imaging personnel are allowed to absorb more radiation, the EqD received must be minimized whenever possible, reducing the potential for:
• Somatic damage
• Genetic damage

ALARA CONCEPT

The best manner for radiographers and radiologists to conscientiously employ the ALARA principle is to use all appropriate radiation-control procedures to mini­mize their exposure levels. Personnel should faithfully employ procedures such as:
• The principles of time, distance, and shielding
• Adequately collimating the radiographic beam
Fig. 15.1)

DOSE-REDUCTION METHODS AND TECHNIQUES

Methods and techniques that reduce patient exposure can also reduce exposure for the radiographer, thereby limiting occupational exposure. Whenever a repeat im­age is performed because of human or mechanical error, the patient receives a double dose of primary radiation, while also increasing the radiographer’s potential for exposure to scattered radiation.

Repeats in Digital Imaging

Because image contrast and overall brightness in digital imaging can be manipulated after image acquisition, the need for all repeats as a result of improper technical exposure factors has been eliminated. However, repeats necessitated by mispositioning can still occur, causing additional radiation exposure to both the patient and possibly the radiographer. The radiographer must take the needed time to accurately position the patient prior to the exposure.

The Patient as a Source of Scattered Radiation

During any diagnostic x-ray examination, the patient becomes a source of scattered radiation as a consequence of the Compton interaction process. At a 90-degree angle
296
CHAPTER 15 Management of Imaging Personnel Radiation Dose
decreases the number of x-ray photons available to undergo Compton scatter. Because scatter is reduced, the radiographer’s potential for occupational exposure is decreased.

Filtration of the Diagnostic X-Ray Beam

When a radiographic beam is adequately filtered, non­useful low-energy photons are removed from the pri-
Collimator
Adjustable lead
shutters
Area of clinical
interest
mary beam. Without proper filtration, a relatively high percentage of the customarily excluded low-energy photons will interact with the tissues of the patient’s body. Some of these photons undergo Compton scatter. The radiographer’s EqD could, therefore, increase as a result of exposure to this excess scattered radiation. Most of these low-energy photons, however, are ab­sorbed in the patient, thereby increasing the patient’s absorbed dose and contributing no useful diagnostic information to the image. Thus, filtration primarily benefits the patient.
Image receptor
Fig. 15.1 Radiographic beam collimation (restricting the x-ray
beam to the area of clinical interest) limits the production of scattered radiation. This radiation-control procedure helps keep the radiographer’s occupational exposure as low as reasonably achievable (ALARA).
to the primary x-ray beam, at a distance of 1 m, the scat­tered x-ray intensity is generally approximately 1/1000th of the intensity of the primary x-ray beam. This charac­teristic should always be kept in mind as an additional method of radiation protection.

Scattered Radiation—Occupational Hazard

Because scattered radiation poses the most significant occupational hazard in diagnostic radiology, the use of any device or imaging technique that lessens the amount of scattered radiation will significantly reduce occupa­tional exposure of diagnostic imaging personnel. Beam constraint devices, such as automatic collimation, or positive beam limitation, restrict the dimensions of the radiographic beam so that its margins do not extend beyond the image receptor. This reduction in beam size

Protective Apparel

Protective lead aprons (Fig. 15.2A) and, in their absence, shielded barriers (Fig. 15.2B) function as gonadal shields for diagnostic imaging personnel. These devices protect personnel from scatter and leakage radiation, which are types of secondary radiation.
Similar to protective gloves that are used to cover the hands of radiologists or radiographers when they must be in or near the primary x-ray beam, lead aprons are avail­able in various thicknesses such as 0.25, 0.5, and 1 mm of lead equivalent.2 Higher lead equivalents in protective ap­parel provide greater protection from radiation exposure. However, for practical use in the clinical setting, the weight of the garment and the approximate length of time that it will be worn must also be considered. An apron containing 1 mm lead equivalent may weigh as much as 12 kg.2 Wearing this protective device for a lengthy proce­dure can, therefore, result in considerable back strain.*
*To reduce the possibility of back or neck problems, other materials may be used in the protective apron to lessen its weight, Some garments, for example, are impregnated with tin2 or similar metals because the electron shell structures of these substances offer advantages in terms of a more probable photoelectric interaction attenuation than does lead in the lower diagnostic x-ray energy range.