Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5196_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
62 Мб
Скачать
CHAPTER 11 Equipment Design for Radiation Protection
217
Fluoroscopic Exposure Control Switch. According to
state and federal regulations, the fluoroscopic exposure control switch (e.g., the foot pedal) must be of the dead- man type, meaning that only continuous pressure ap­plied by the operator can keep the switch activated and the fluoroscopic tube emitting x-radiation. This clearly implies that the exposure automatically terminates if the person operating the switch becomes incapacitated (e.g., has a medical emergency) or, for any reason, re­moves his or her foot from the pedal.

Mobile Fluoroscopic Systems

Radiation Safety Features of Mobile C-arm Fluoroscopy.
portable x-ray unit that has a C-shaped multiorient­able gantry arm. An x-ray tube is attached to one end of its curved arm and an II attached to the other end. C-arm fluoroscopes (Fig. 11.20) are frequently used in the operating room for orthopedic procedures (e.g., pinning of a fractured hip). They are also employed quite commonly for:
• Cardiac imaging
• Interventional procedures
A mobile C-arm fluoroscopic unit is a
The use of C-arm fluoroscopy in interventional pro­cedures carries the potential for a relatively large patient radiation dose. C-arm fluoroscope operators, if stand­ing close to the patient, could also receive a significant increase in occupational exposure from patient scatter radiation during such cases. For these reasons, equip­ment operators, including attending physicians, must have appropriate education and training to ensure that they will be able to follow guidelines for safe C-arm operation and also meet radiation safety protocols es­sential to patient and personnel safety. A necessary component of this is the wearing of wrap-around lead aprons and thyroid shields by all involved personnel throughout each procedure.
Mobile fluoroscopic units are required to have a minimal source-to-end of collimator assembly distance of 30 cm (12 inches). Some type of spacer or collimator extension is customarily installed to prevent any part of the patient from coming closer than 30 cm (12 inches) to the tube target. During C-arm fluoroscopy procedures, the patient-II distance should be as short as possible (Fig. 11.21). This reduces patient entrance dose. For fur- ther dose-reduction purposes it is preferable to position the C-arm so that the x-ray tube is under the patient. With the x-ray tube in this position, scatter radiation is attenuated (Fig. 11.22) toward personnel above the knee level. However, when the x-ray tube is positioned over the patient, scatter radiation is not attenuated by the table, and therefore it is more intense, and radiation exposure of personnel increases correspondingly.
Fig. 11.20 C-arm fluoroscope and monitor.

RADIATION SAFETY FEATURES OF DIGITAL FLUOROSCOPIC EQUIPMENT

Digital Fluoroscopy (DF)

Pulsed Progressive Systems. Various methods are
used to obtain digital images in fluoroscopic equipment. The electrical signal from the video camera attached to the output phosphor may be digitized. In newer equip­ment, the analog TV camera may be replaced by a digital device such as a CCD camera or other digital detector. In whatever manner the digital image is acquired, the use of digital technology offers the possibility of several meth­ods of dose reduction. One such method utilizes a brief high-intensity pulse of radiation to create an entire im­age on the output phosphor. The lines composing the image are progressively scanned (i.e., from left to right
218
CHAPTER 11 Equipment Design for Radiation Protection
TELEVISION CAMERA
IMAGE INTENSIFIER
PATIENT
TABLE
SPACER CONE
X-RAY TUBE
Fig. 11.21 To reduce the patient’s entrance dose during C-arm fluoroscopy, the patient–image intensifier
distance should be as short as possible. (Courtesy Mark Rzeszotarski.)
WEAK
SCATTER
REGION
C-ARM ASSEMBLY
CONTROL
UNIT
IMAGE INTENSIFIER
STRONG
BACKSCATTER
REGION
INTENSE
PRIMARY
BEAM
Fig. 11.22 To reduce scatter radiation during C-arm fluoroscopy, position the C-arm so that the x-ray tube is
under the patient whenever possible. (Courtesy Mark Rzeszotarski.)
followed by right to left and so on, usually referred to as raster scanning, from top to bottom) to provide the im­age that appears on a monitor during a brief period (one-sixtieth of a second). The x-ray beam is deactivated while the image is being scanned, thereby decreasing patient dose, and then pulsed back on for the next image. These systems are known as pulsed progressive systems and are commonly used to lower patient dose.
SPACER
X-RAY
TUBE
CONE
Last Image Hold. Another dose-reduction technique
that is particularly effective in DF systems is the last im­age hold feature. For this feature, the image formed
when the x-ray tube was last energized remains on the monitor so that no further radiation exposure is needed to regenerate it. In a digital system, using embedded processing software, this retained image could also be composed of several frames of information that have
CHAPTER 11 Equipment Design for Radiation Protection
219
been added together to reduce the effect of quantum or random noise that would be particularly apparent in a single frame.

Digital Subtraction Angiography (DSA) and Interventional Systems

Interventional Procedures. Digital techniques have
had a major impact on interventional fluoroscopy. The goal in interventional procedures is to locate high­contrast, small objects such as catheters, stents,* or electrical leads for electronic implants. Not only are these devices in the millimeter-size range, but they are also moved within the patient as part of the procedure to implant the devices or to inject liquids such as glues,** steroids, or anesthetics. Contrast material is sometimes injected to improve the visibility of the de­vices or the leading edge of catheters. The visualization of high-contrast, small objects with a digital imaging system requires that the pixel elements reveal very small regions of the patient. In addition, images must be up­dated at a reasonable rate to reveal the location of ob­jects in the patient as the objects are moved. In the latest DF systems, the II has been replaced by a “flat-panel” detector that is similar to those used in DR. A flat-panel digital detector for fluoroscopy meets the same techni­cal requirements as previously described (direct or indi­rect conversion of x-ray energy to stored electronic val­ues) but is also capable of recording images rapidly, as many images or “frames” as are required per second to be perceived as motion and presented in a live digital video format. The advent of these techniques has al­lowed the performance of more complicated proce­dures that are sometimes associated with much longer exposure times for patients. Some of the data related to these lengthy irradiations are provided in Chapter 2. It is still the responsibility of the imaging team to make sure that acceptable risk versus benefit principles are applied. While the radiation risk may be increased com­pared to other imaging techniques, the benefit of repair of life-threatening conditions must also be considered.
*A stent is a tubular support that can be placed inside a blood vessel, canal, or duct to create a pathway or relieve an ob­struction.
**Injection of glues or glue embolization is an interventional technique that uses a chemical called n-butyl cyanoacrylate, which is usually shortened to NBCA, to block problematic arteries.
Digital Subtraction Angiography. One of the most sig-
nificant advances in digital imaging in vascular and inter­ventional fluoroscopy over the past 50 years has been
Digital Subtraction Angiography (DSA). In vascular flu-
oroscopy, the objective is the visualization of blood vessels. This may be necessary to complete an interventional pro­cedure, or it may be used to reveal vascular problems such as occlusion (blockage), stenosis (narrowing), or aneu­rysms (abnormal bulging). It usually includes the injec­tion of contrast material to enhance vascular appearance. DSA implies the performance of arithmetic operations on the pixel values of digital images. Since the individual im­ages or frames of the digital video are available as a matrix of signal (voltage) values that are stored in a computer, pairs of images may be subtracted. In image subtraction, the digital values recorded for each corresponding pixel of an image pair are subtracted from each other. This process produces a resultant image that “only” consists of features
that have changed between the initial two images.
In DSA, a “mask” image (simply a reference image of the same area before the contrast is administered) is obtained prior to the injection of contrast material. Subsequent images may be subtracted from the initial mask so that the main difference between pairs of im­ages (the movement of contrast material) is empha­sized. Therefore, the contrast material that lines the vessels of interest enhances the appearance of the vessels revealing features such as narrowing, or stenosis. Inter­ventional procedures may use this information to guide the insertion of stents to open occluded blood vessel. For complex interventional procedures, this increased vascular contrast allows visualization of small vessels surrounding arteriovenous malformations.* Additional clarity may be obtained by remasking (redefining the image from which subsequent images are subtracted) when catheters approach the malformation.
Roadmapping. In situations where the vasculature is
complex or overlying structures provide a confusing background, the location of the catheter may be difficult to determine visually. More contrast material could be injected to improve visibility, but the image would still be cluttered, and there are limits to the amount of con­trast material that can safely be injected into the patient
*An abnormal tangle of blood vessels connecting arteries and veins, which disrupts normal blood flow and oxygen circulation.
220
CHAPTER 11 Equipment Design for Radiation Protection
Fig. 11.23 Roadmapping: First, the digital subtraction angiography (DSA) of the vascular structure is per-
formed. The postcontrast frame associated with maximum vessel opacification becomes the roadmap mask; subsequent digital fluoroscopic (Fluoro) images are subtracted from the roadmap mask. The result in live fluoroscopic images of the inserted catheter or wire overlaid on a static image of the vasculature (with dis­tracting underlying tissue removed). (From Pooley RA, McKinney JM, Miller DA, The AAPM/RSNA physics tutorial for residents: digital fluoroscopy, Radiographics 21(2): 521–534, 2001).
so as to avoid potential kidney damage. A technique known as roadmapping is useful in this situation. If the patient is not moving and the FOV is not altered, a static image of the vasculature may be obtained through sub­traction, pre- and postcontrast injection, as in standard DSA. This static image may be preserved as a lower­contrast “ghost” image or as a reversed-contrast lighter image. In either case, it serves the function of a station­ary map over which the outline of the catheter can be shown as it advances to the position where it is needed.
Fig. 11.23 illustrates a roadmap procedure used for a
catheter approaching an aneurism.

RADIATION SAFETY FOR HIGH-LEVEL CONTROL INTERVENTIONAL PROCEDURES

Justification for Use of High-Level Control (HLC) Interventional Procedures
As mentioned previously, interventional procedures are invasive actions performed by a physician with the aid of
fluoroscopic imaging. The interventional physician, usu­ally a radiologist or a cardiologist, inserts catheters into vessels or directly into patient tissues for the purpose of:
• Drainage
• Biopsy
• Alteration of vascular occlusions (obstructions) or malformations For these procedures, high-level control fluoros-
copy (HLCF) is often employed. HLCF is an operating
mode for state-of-the-art fluoroscopic equipment in which entrance radiation levels are substantially higher than those normally employed in routine procedures. The higher entrance exposure rate or entrance dose rates allows the examination of smaller and lower­contrast objects that are not ordinarily discerned during standard fluoroscopy. HLCF therefore is used for those interventional procedures in which visualization of fine catheters or other not easily seen structures is crucial. An audible signal constantly reminds personnel that the HLC is engaged. It is strongly recommended that pulsed
mode and frequent last image hold be the standard meth­ods of operation when HLCF is being used.
CHAPTER 11 Equipment Design for Radiation Protection
221

Public Health Advisory About the Dangers of Overexposure of Patients and Exposure Rate Limits

Many of the fluoroscopically guided therapeutic inter­ventional procedures have the potential for substantial patient exposure. On September 30, 1974, the US Food and Drug Administration (FDA) issued a public health advisory to alert health care workers to the probability and dangers of overexposure of patients through the use of high-level fluoroscopy. The FDA x-ray equipment standards, issued in 1994, limited the tabletop exposure rate of fluoroscopic equipment for routine procedures to a maximum of 10 R/min unless an HLC was present, in which case routine fluoroscopy maximum tabletop exposure rates were set at 5 R/min when the system was not in HLC mode and unlimited when it was in HCL
mode.11 The authors of these 1994 standards believed
that the unrestricted high-level capability was necessary for certain vital situations involving therapeutic inter­ventional procedures in which the potential risks to the patient of much increased radiation exposure would be subordinate to a successful medical outcome of an intervention. As a safeguard, however, HCL was re­quired to have manual continuous positive pressure on a special high-level foot pedal, accompanied by a steady audible signal to remind personnel that the high-level fluoroscopic mode was in use. In this operational mode, patient entrance exposure rates have been estimated to range from 20 to 120 R/min (176 to 1056 mGya/min entrance or skin dose rates). When the rule was issued, total patient irradiation was limited by the heat-loading capabilities of the x-ray tube. Amazingly, the thinking or belief at that time was that the tube would reach its heat limit before any detectable patient early tissue reactions could occur. By the early 1990s, however, advances in x-ray tube heat dissipation technology and the in­creased frequency of vascular interventional procedures that require long fluoroscopic times (Box 11.5) had created a situation in which very serious skin reactions had been reported in some patients. Radiogenic skin injuries such as erythema (diffuse reddening) and des- quamation (sloughing off of skin cells) are early tissue reactions in which the severity of the disorder increases with radiation dose. As the data in Table 11.2 demon­strate, a half-hour of total high-level beam-on time at one location on a patient’s skin is sufficient to produce
BOX 11.5 Procedures Involving
Extended Fluoroscopic Time
Percutaneous transluminal angioplasty
Radiofrequency cardiac catheter ablation
Vascular embolization
Stent and filter placement
Thrombolytic and fibrinolytic procedures
Percutaneous transhepatic cholangiography
Endoscopic retrograde cholangiopancreatography
Transjugular intrahepatic portosystemic shunt
Percutaneous nephrostomy
Biliary drainage
Urinary or biliary stone removal
From the US Food and Drug Administration (FDA): Public
health advisory: avoidance of serious x-ray-induced skin injuries to patients during fluoroscopically guided procedures,
Rockville, MD, September 30, 1994, FDA.
erythema. In this case, the skin injury does not usually appear for approximately 10 days. Because manifesta­tions of skin injury are delayed, a radiologist would not typically be the first person to observe the onset of the symptoms. Therefore, patient monitoring, radiation dosimetry, and accurate record-keeping are essential for the future medical management of adverse reactions. The FDA has recommended that a notation be placed in the patient’s record if a skin dose in the range of 1 to 2 Gy
is received. The location of the area of the patient’s
t
skin that received the absorbed dose should also be noted using:
• A diagram
• Annotated photograph
• Narrative description Since 2000, however, alarmed state regulatory agencies
have imposed a restriction on high-level radiation entrance rates. With the II at a distance of 30 cm (12 inches) above the tabletop, the maximum continuous fluoroscopic entrance exposure rate permitted is 20 R/min (176 mGya/min entrance dose rate). This is not true, though, for pulsed mode for which the instantaneous exposure rates values can be very much higher.
Use of Fluoroscopic Equipment by Non-radiologist Physicians
Fluoroscopic devices are capable of subjecting the patient, the physician, and other personnel close to the fluoroscopic equipment to substantial amounts of
222
TABLE 11.2 Radiation-Induced Skin Injuries
Effect
Early transient erythema 2 1.7 0.17 Hours Temporary epilation 3 2.5 0.25 3 weeks Main erythema 6 5.0 0.50 10 days Permanent epilation 7 5.8 0.58 3 weeks Dry desquamation 10 8.3 0.83 4 weeks Dermal atrophy 11 9.2 0.92 0.14 weeks Telangiectasias 12 10.0 1.00 0.52 weeks Moist desquamation 15 12.5 1.25 4 weeks Late erythema 15 12.5 1.25 6–10 weeks Dermal necrosis 18 15.0 1.50 0.10 weeks Secondary ulceration 20 16.7 1.67 0.6 weeks
*Time required to deliver the typical threshold dose at the specified dose rate.
The unit for absorbed dose is the gray (Gyt) in the International System of Units. ‡Time after single irradiation to observation of effect. Modified from Wagner LK, Eifel PJ, Geise RA: Potential biological effects following high x-ray dose interventional procedures, J Vasc Interv Radiol 5:71, 1994.
CHAPTER 11 Equipment Design for Radiation Protection
HOURS OF FLUOROSCOPIC “ON TIME” TO REACH THRESHOLD*
Typical Threshold
Absorbed Dose
(Gy
)
t
Usual Fluoroscopic
Dose Rate of
0.02 Gy
/min
a
High-Level
Dose Rate of
0.2 Gya/min
Time to Onset of
Effect
direct and/or indirect ionizing radiation. These devices include:
• C-arm fluoroscopes
• Fluoroscopes on stationary equipment with HCL
mode used for interventional procedures
• Biplane (dual x-ray tubes) interventional fluoro-
scopic systems*
Ongoing education and training in the safe use of fluoroscopic equipment should be mandatory for non­radiologist physicians and equipment operators. Many institutions have therefore established through their RSOs radiation safety accreditation programs that must be taken and passed before non-radiologist physicians are permitted to use such devices.
Some of the causes of elevated radiation exposures to personnel during interventional procedures include:
• Operation of the fluoroscopic tube for long periods
in continuous mode in place of pulsed mode
• Failure to use the protective curtain or floating
shields on the stationary fluoroscopic equipment’s II
as a means of protection
*Biplane x-ray fluoroscopy systems consist of a double c-arm independent configuration with each c-arm having its own x-ray source and corresponding detector panel. The two c-arms are in the mechanical arrangement that allows them to generate multiple oblique dual-angle projections.
Monitoring and documenting procedural fluoro­scopic time are essential. The responsibility for this documentation has generally belonged to the radiographer assisting with the procedure. With newer systems, the computer software automatically produces a record of both accumulated continuous and pulsed radiation time, as well as estimates of de­livered dose and dose area product. In addition, if a physician loses track of how long a procedure is tak­ing and how much radiation is being delivered to a localized area of the patient’s body, it becomes the radiographer’s ethical responsibility to call this to the physician’s attention in the interest of the safety of all concerned. In the event of a critical situation in which there is markedly excessive fluoroscopic irra­diation time, the radiographer is responsible for no­tifying an appropriate supervisor, who should then implement the imaging facility’s relevant established protocol.
The National Cancer Institute and the Society of Interventional Radiology have conjointly designed guidelines to assist physicians in developing strategies that will enable them to fulfill their interventional clinical objectives while controlling patient radiation dose and minimizing exposure to occupationally exposed personnel. These strategies are listed in
Box 11.6.
CHAPTER 11 Equipment Design for Radiation Protection
223
BOX 11.6 Strategies to Manage Radiation Dose to Patients, Operators, and Staff During
Interventional Fluoroscopy
Immediate Long-Term
Optimize Dose to Patient
Use proper radiologic technique:
Maximize distance between x-ray tube and patient
Minimize distance between patient and image receptor
Limit use of electronic magnification
Control fluoroscopic time:
Limit use to necessary evaluation of moving structures
Employ last image hold function to review findings
Control images:
Limit acquisition to essential diagnostic and documenta-
tion purposes
Reduce dose:
Reduce field size (collimate) and minimize field overlap
Use pulsed fluoroscopy and low frame rate
Minimize Dose to Operators and Staff
Keep hands out of the beam Use movable shields Maintain awareness of body position relative to the x-ray
beam:
Horizontal x-ray beam: operator and staff should stand on
the side of the image receptor
Vertical x-ray beam: the image receptor should be above
the table
Wear adequate protection:
Protective, well-fitted lead apron
Leaded glasses
From the National Cancer Institute, Division of Cancer Epidemiology and Genetics, Radiation Epidemiology Branch: Interventional fluoroscopy: reducing radiation risks for patients and staff, NIH Publication No. 05-5286, Rockville, MD, 2005, National Institutes of Health.
Include medical physicist in decisions:
Machine selection and maintenance
Incorporate dose-reduction technologies and dose-
measurement devices in equipment
Establish a facility quality improvement program that
includes an appropriate x-ray equipment quality assurance program, overseen by a medical physicist, which includes equipment evaluation/inspection at appropriate intervals
Improve ergonomics of operations and staff:
Train operators and staff in ergonomically good position-
ing for use of fluoroscopy equipment; periodically assess their practice
Identify and provide the ergonomically best personal
protective gear for operators and staff
Urge manufacturers to develop ergonomically improved
personal protective gear
Recommend research to improve ergonomics for per-
sonal protective gear

S U M M A R Y

• A diagnostic-type tube housing protects the patient and imaging personnel from off-focus, or leakage, radiation by restricting the emission of the x-rays to the area of the useful, or primary, beam.
• Leakage radiation from the tube housing measured at 1 m from the x-ray source must not exceed 0.88 mGya/h (100 mR/h) when the tube is operated at its highest kilovoltage at the highest current (mA) that allows continuous operation.
• The control panel, or console, must be located be­hind a suitable protective barrier that has a radia­tion-absorbent window that permits satisfactory observation of the patient during any procedure.
This panel must indicate the conditions of exposure and provide indication when the x-ray tube is ener-
1
gized.
• The radiographic examination tabletop must be of uniform thickness, and for under-table tubes as used in fluoroscopy, the patient support surface also should be as radiolucent as possible, which will lead to reducing the patient’s radiation dose.
• Radiographic equipment must have an source­to-image receptor distance (SID) indicator.
• X-ray beam limitation devices must be used to con­fine the useful beam before it enters the anatomic area of clinical interest.
224
CHAPTER 11 Equipment Design for Radiation Protection
• The patient’s skin surface should always be at least 15 cm below the collimator to minimize exposure to the epidermis.
• Good coincidence between the x-ray beam and the light-localizing beam of the collimator is necessary; both alignment and length and width dimensions of the two beams must correspond to within 2% of the SID.
• Exposure to the patient’s skin may be reduced through proper filtration of the radiographic beam.
• Compensating filters are used in radiography to provide uniform imaging of body parts when con­siderable variation in thickness or tissue composi­tion exists.
• Diagnostic x-ray units must have consistent expo­sure reproducibility, that is, the ability to duplicate specific radiographic exposures for any given combi­nation of kVp, mA, and time.
• Exposure linearity is essential. When a change is made from one mA station to a neighboring mA sta­tion, the most that linearity can vary is 10%.
• Radiographic grids increase patient dose in radiog­raphy. Their use for the examination of thicker body parts is a fair compromise because they remove scat­tered radiation emanating from the patient that would otherwise degrade the completed image.
• Because of increased sensitivity of photostimulable phosphor to scatter radiation before and after exposure to a radiographic beam, a grid may be used more fre­quently during computed radiography (CR) imaging. The use of a grid does increase patient dose but sig­nificantly improves radiographic contrast and visibility of detail.
• To limit the effects of inverse fall-off of radiation intensity with distance during a mobile radiographic examination, an x-ray source-skin distance (SSD) of at least 30 cm (12 inches) must be used.
• With digital radiography, the image receptor is divided into small detector elements that make up the two­dimensional picture elements, or pixels, of the image.
• Radiographers must always strive to select correct technical exposure factors from the beginning to avoid overexposing patients when digital images are obtained.
• The digital image ultimately acquired in both CR and digital radiography (DR) can be displayed on a monitor for viewing and evaluation and can also be printed if a hard copy is desired.
• Fluoroscopic procedures produce the largest patient radiation rates levels in diagnostic radiology. There­fore, multiple safety guidelines should always be ad­hered to.
• During C-arm fluoroscopic procedures, the pa­tient-image intensifier distance should be as short as possible.
• During digital fluoroscopy, the use of pulsed pro­gressive systems with last image hold and roadmap­ping will significantly lower patient dose.
• High-level control fluoroscopy (HLCF) is often en­gaged in interventional procedures and produces entrance radiation rate levels that are substantially higher than those allowed for routine fluoroscopic procedures.
• If skin dose is received in the range of 1 to 2 Gyt, the Food and Drug Administration (FDA) requires that a notation be placed in the patient’s record.

G E N E R A L D I S C U S S I O N Q U E S T I O N S

1. What are the x-ray tube housing construction require-
ments when a tube is operated at its highest voltage at the highest current that allows continuous operation?
2. What must the control panel, or console, indicate?
3. How do light-localizing variable-aperture rectan-
gular collimators reduce the amount of scattered radiation being produced during a radiographic examination?
4. How does filtration of the radiographic beam re-
duce exposure to the patient’s skin and superficial tissues?
5. When should a radiological physicist measure the
6. What are acceptable HVL values for radiographic
7. What is exposure linearity?
8. What is the effective metric equivalent unit for the
9. Why is the use of a radiographic grid acceptable if
10. What can a radiographer do to avoid overexposing
half-life of a diagnostic x-ray beam?
and fluoroscopic systems?
English unit 40 inches?
its use increases patient dose?
the patient when a CR system is used?
CHAPTER 11 Equipment Design for Radiation Protection
225
11. What effect does the use of pulsed, or intermittent,
fluoroscopy have on patient dose?
12. Why is there concern over the use of mobile C-arm
fluoroscopes during surgical, vascular, interven­tional, and other potentially lengthy procedures?
13. To what does the scientific term luminance refer?
14. What strategies can physicians employ during inter-
ventional fluoroscopic procedures to control patient radiation dose and minimize exposure of occupa­tionally exposed personnel and any other assisting personnel?
15. During digital fluoroscopy, how does the use of a
pulsed progressive system lower patient dose?
16. How is the appearance of vascular structures im-
proved by digital subtraction angiography?

R E V I E W Q U E S T I O N S

1. The radiographic beam should be collimated so that
it is which of the following?
A. Slightly larger than the image receptor B. No larger than the image receptor C. Twice as large as the image receptor D. Four times as large as the image receptor
2. Both alignment and length and width dimensions of
the radiographic and light beams must correspond to within:
A. 1% of the SID B. 2% of the SID C. 5% of the SID D. 10% of the SID
3. What is the function of a filter in diagnostic radiology?
A. To permit only alpha rays to reach the patient’s skin B. To permit only beta particles to interact with the
atoms of the patient’s body
C. To decrease the x-radiation dose to the patient’s
skin and superficial tissue
D. To remove gamma radiation from the useful beam
4. HVL may be defined as the thickness of a designated
absorber required to do which of the following?
A. Increase the intensity of the primary beam by
50% of its initial value
B. Increase the intensity of the primary beam by
25% of its initial value
C. Decrease the intensity of the primary beam by
50% of its initial value
D. Decrease the intensity of the primary beam by
25% of its initial value
17. How is the progression of a catheter through com-
plicated vasculature improved by roadmapping?
18. If a radiographer is operating a C-arm fluoroscope
during an interventional procedure, what is the consequence if the operator stands too close to the patient while the x-ray beam is energized?
19. If during a lengthy fluoroscopic procedure, a non-
radiologist physician loses track of how long the procedure is taking and how much radiation is be­ing delivered to a localized area of the patient’s body, what is the radiographer’s ethical responsi­bility?
20. What is an arteriovenous malformation and what
imaging procedure would be used to visualize the anatomical structures of interest?
5. The photostimulable phosphor in the computed ra-
diography imaging plate is more sensitive to scatter radiation before and after it is sensitized through exposure to a radiographic beam. Because of this increased sensitivity, which of the following is true?
1. Five millimeters of added aluminum equiva­lent filtration must always be used during routine CR imaging.
2. A radiographic grid may be used more fre­quently during CR imaging.
3. Any source-to-image receptor distance can be used during CR imaging without adjustment in technical exposure factors.
A. 1 only B. 2 only C. 3 only D. 1, 2, and 3
6. To minimize skin exposure to electrons produced by
photon interaction with the collimator, how far below the collimator should the patient’s skin surface be?
A. At least 1 cm below B. At least 5 cm below C. At least 10 cm below D. At least 15 cm below
7. Which of the following aluminum equivalents for
total permanent filtration meets the minimum re­quirement for mobile diagnostic and fluoroscopic equipment?
A. 0.5 mm aluminum equivalent B. 1.0 mm aluminum equivalent
226
CHAPTER 11 Equipment Design for Radiation Protection
C. 2.0 mm aluminum equivalent D. 2.5 mm aluminum equivalent
8. The trough, or bilateral wedge, filter, which is used in
some dedicated chest radiographic units, is an ex­ample of which of the following?
A. Compensating filter B. Filter used in all digital imaging systems C. Filter used in all dedicated mammographic units D. Filter used in all computed tomography systems
9. To decrease patient exposure during fluoroscopic
procedures, the fluoroscopist can:
1. Limit the size of the fluoroscopic field to include only the area of anatomy that is of clinical interest
2. Employ the practice of pulsed fluoroscopy to reduce the overall length of exposure
3. Choose to use a conventional fluoroscope in­stead of an image intensification fluoroscope
A. 1 and 2 only B. 1 and 3 only C. 2 and 3 only D. 1, 2, and 3
10. A diagnostic-type protective tube housing must be
constructed so that leakage radiation measured at a distance of 1 m from the x-ray source does not ex­ceed ______ when the tube is operated at its highest voltage at the highest current that allows continu­ous operation.
A. 5 Gya/hr B. 3 Gya/hr C. 1 Gya/hr D. 0.1 Gya/hr
11. Which of the following changes might improve the
visibility of vessels in an interventional angiography study but may cause harm to the patient?
A. Digital subtraction angiography B. Increase the amount of contrast material
C. Use roadmapping D. Use pulsed fluoroscopy
12. Pulsed, or intermittent, fluoroscopy involves man-
ual or automatic periodic activation of the fluoro­scopic x-ray tube by the fluoroscopist, rather than continuous activation. Which of the following are benefits of this practice?
1. Automatically magnifies the fluoroscopic image
2. Helps to extend the life of the x-ray tube
3. Significantly decreases patient dose, especially
in long procedures
A. 1 and 2 only B. 1 and 3 only C. 2 and 3 only D. 1, 2, and 3
13. A resettable device on a fluoroscopic unit that mea-
sures the collective x-ray beam-on time and sounds an audible alarm or in some cases interrupts the exposure after the fluoroscope has been activated for 5 minutes is termed a:
A. Cumulative timer B. Interruption timer C. Interventional timer D. Subtraction timer
14. Non-digital fluoroscopy involves the use of a signal
amplification device called an:
A. Flat panel detector B. Image intensifier tube C. Photostimulable phosphor D. Progressive subtraction device
15. When changing one mA station to a neighboring
mA station, the most linearity can vary is:
A. 5% B. 10% C. 20% D. 50%