Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5196_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
62 Мб
Скачать
CHAPTER 11 Equipment Design for Radiation Protection
207
• Digital radiography (DR)
• Digital fluoroscopy (DF)
• Digital mammography (DM)

Digital Imaging Overview

All non-analog or digital images have many similar prop­erties regardless of the engineering technology with which they are acquired, simply because they are all nu­merical approximations of non-digital (analog) signals. Even though these signals may be produced by vast num­bers of x-ray photons, the overall strength of the signal is recorded as one number at each location on the image receptor. All electronic radiography devices have inherent limitations with respect to both spatial and contrast reso­lution due to the finite dimensions of their detector ele­ments or picture elements (pixels). All digital imaging devices are also subject to certain artifacts (i.e., effects seen in an image system that were introduced by the tech­nology used in acquiring the image or images) such as aliasing, moiré patterns, and contouring (see Box 11.3).
These unwanted effects occur because digital images are produced collectively by an array or matrix of ele­ments that have finite dimensions and are always subject to random quantum noise effects (mottle), which grow in importance as the matrix elements become finer.
With any of these modes of computer-processed radi­ography, the latent (unprocessed) image formed by x-ray photons on a radiation detector is an electronic image.6 Because this anatomic information is subsequently col­lected by a computer and shown on its display, it is called a digital image.7 The familiar radiographic densities then appear as levels of brightness associated with shades of gray. The shades of gray that are displayed on a computer screen constitute the contrast in the image. The number of different shades of gray that can be stored in memory and subsequently displayed is known as the grayscale. All digital images are composed of numerical data that can be easily manipulated by a computer.
The numeric values of the digital image are aligned in a fixed number of rows and columns (an array) that form many individual miniature square boxes, each of which corresponds to a particular place in the image. These in­dividual boxes collectively constitute the image matrix. Each miniature square box in this matrix is a picture ele- ment, or pixel. The pixels collectively produce a two- dimensional representation of the information contained in a volume of tissue.8 The size of the pixels determines
7
BOX 11.3 Image Aliasing, Morié
Patterns, and Contouring Defects
In computer graphics, aliasing is an image distortion that shows up most simply as the jagged, or saw­toothed, appearance of curved or diagonal lines. Moiré patterns occur when an object that is being im­aged contains many fine, repetitive details. As a result, strange-looking wavy patterns are overlayed on the im­age. This is shown in the figure below.
(From Carter CE, Vealé BL: Digital radiography and PACS, ed 3, Philadelphia, 2019, Elsevier.)
Contouring artifacts usually show up as patterns of small blocks in an otherwise smooth image. An exam­ple of this are adjacent groups of rectangular or block­like images of different intensity or brightness superim­posed on what is really a clear smooth sky. These blocks become very much more noticeable as the im­age is viewed on larger devices or monitors.
the sharpness of the image. The resolution is finer when pixels are smaller. One example of a standard matrix size is 512 pixels high by 512 pixels wide, or simply 512 3 512. A range of other matrix sizes are also used, such as 1024 3 1024. For the same field of view (FOV), the latter cor­responds to four times as many elements distributed over the same area. The pixels are therefore smaller, which
208
CHAPTER 11 Equipment Design for Radiation Protection
Indirect Conversion
Scintillator
Photodiode
Thin-film
transistor array
Electrical signals
Fig. 11.13 Some large area detectors provide indirect conversion of x-ray energy to electrical charge through
intermediate steps involving photodiodes or charge-coupled devices. Other area detectors provide direct conversion of x-ray energy to electrical charge through the use of a photoconductor. (From Hendee WR, Rit­enour ER: Medical imaging physics, ed 4, Chicago, 2002, John Wiley & Sons.)
Scintillator
Charge-coupled-
device array
Electrical signals
leads to improved patient image detail. It should be noted that such resolution increases, however, are associated with very substantial increases in the quantity of data to be computer processed.
Digital image receptors used in DR convert the en­ergy of x-rays into electrical signals. The image receptor is divided into small detector elements that make up the pixels of the digital image. There are various types of these image receptors. Some use a scintillator, such as amorphous silicon,* to convert the x-ray energy into visible light. The visible light is then transformed into electrical signals by an assortment of transistors or an array of charge-couple devices (CCDs), such as those found in video cameras. Other systems use a photocon- ductor, such as amorphous selenium,* to convert the x-ray energy directly into electrical signals that are then read by an ordered grouping (matrix) of transistors. In these systems, the number and size of small transistors or CCDs determine the number and size of pixels in the digital image. Advances in materials technology have resulted in pixel sizes as small as 50 micrometers, which approaches the resolution of screen-film imaging sys­tems (Fig. 11.13).
*Amorphous refers to a non-crystalline grouping of silicon atoms ( in a regular geometric pattern, the silicone atoms or the sele­nium atoms are distributed in a continuous random fashion.
Si28) or selenium atoms (34Se79) in which, rather than
14
Direct Conversion
Photoconductor
Thin-film
transistor array
Electrical signals
Digital images can quickly be accessed via a PACS* network at multiple workstations at the same time, thus allowing image viewing to be very convenient for physicians providing patient care. Patient infor­mation and reports can be included in the patient’s imaging file, along with records from other imaging modalities.
9

Computed Radiography (CR)

Computed radiography CR is the descriptive term for
those x-ray systems that generate images using the process of photostimulable luminescence (PSL). In this technique, the energy of the conventional diagnostic x-ray beams passing through a patient is not projected onto x-ray film within a cassette but rather is deposited onto a modified crystalline material known as a photo­stimulable phosphor (PSP).** The absorbed energy
*PACS refers to a picture archiving and communication system imaging technology which provides throughout an authorized and secure computer network both economical storage and con­venient remote access to images from multiple modalities. The universal format for PACS image storage and transfer is called dicom (Digital Imaging and Communications in Medicine.) **Europium-activated barium fluorohalide (BaFX:Eu) is the most commonly employed phosphor in which “Xrepresents a
combination of a bromide (e.g., potassium bromide KBr) and an iodide (e.g., sodium iodide NaI), typically 85% and 15%, respec-
1
tively.
CHAPTER 11 Equipment Design for Radiation Protection
209
produces a latent (undeveloped) image composed of PSP molecules with electrons that have been excited (energy raised) into fixed states, which are also known as local potential energy traps.
The number and locations of these trapped excited electrons is a molecular photo of the originally pro­jected x-ray energy. The electrons are released from their traps by gaining added energy from absorbing external light of a specific wavelength (photostimula­tion). These electrons subsequently lose their excess energy by the release of the photostimulable light (PSL) they absorbed. An image-reading unit employing a helium-neon laser is used to scan the light, which then can be measured and enumerated to create a digital im­age. A computer stores the digital image for visual dis­play on a monitor. If desired, the image can be printed on a laser film when a hard copy is needed. While the digital image is displayed on a monitor, the radiogra­pher, by manipulating the computer mouse (Fig. 11.14) or scrolling a touch screen cursor, can adjust the image to the correct:
• Size
• Brightness
• Contrast
After adjustments have been completed, the image or images are sent through a PACS system for review by a
radiologist. In CR, the receptor can be erased with white light and reused to acquire another projection.
CR was useful as an interim step in converting older imaging equipment to digital techniques, since the CR plate or cassette could be substituted for the film-screen cassette in older equipment with no further modifica­tions. However, the speed of image acquisition and dose efficiency considerations indicate that there will be an eventual conversion of all equipment to DR in the future.
Kilovoltage. The kVp controls the penetrating ability of
the x-ray beam as it passes through human anatomy and also affects radiographic contrast in the image. The size of the part or area of the body to be imaged and the type of subject contrast desired in the completed image determine kVp selection.2 In CR imaging, unlike film­screen imaging, it has been found that with respect to kilovoltage, there is a substantial amount of flexibility available for selection of the degree of desired subject contrast. Regulatory standards require that technique charts indicating optimal kVp values for all CR projec­tions must be available in the x-ray room near the oper­ating console for radiographers.
X-Ray Beam Collimation. For the computer to form a
CR image correctly, the body area or part being irradiated must be accurately positioned in or near the center of the CR image receptor. In practical application, only one projection per image is obtained on a CR imaging plate.
Fig. 11.14 The radiographer at the monitor uses the mouse to
adjust the computed radiography image of the body part to the proper size, density, and contrast before electronically sending the image for reading.
Use of Radiographic Grids. The CR imaging plate can
absorb low-energy scattered photons; therefore, it is sen­sitive to scatter radiation both before and after it is sensi­tized by exposure to a radiographic beam.3 Because of this increased sensitivity, a radiographic grid should be used more frequently. For chest radiography, Carlton and Adler advocated the use of a grid for optimum images when chest measurements exceed 24 to 26 cm.3 Some CR imaging manufacturers recommended the use of a grid for particular radiographic projections that require rela­tively high-kVp settings. Grid selection depends on sev­eral factors: the size of the anatomic features to be imaged, kVp selected, amount of scatter removal pre­ferred, and grid frequency (lines per centimeter or inch).
It is customary and, for the best image quality, neces­sary to use a grid for anatomy sections more than 10 cm thick or for techniques that exceed 70 kVp. This need remains true with both CR and DR. In general, the
3
210
CHAPTER 11 Equipment Design for Radiation Protection
problem a radiographer faces with CR is that the mAs required and, consequently, the patient dose received usually are higher than for non-CR. The addition of a grid will only further increase that dose. Many quality assurance teams, however, realize that CR, because of its wider exposure latitude, reduces the need for grid use on the pediatric population. As a result, satisfactory non-grid pediatric protocols have been developed and implemented.

Digital Radiography (DR)

DR systems, unlike CR, use as image receptors active­matrix flat panels consisting of a detection layer deposited over an array of thin-film transistors and photodiodes. Current state-of-the-art digital sensors are more sensitive or responsive to diagnostic energy x-rays than radio­graphic film, and thus much less radiation (up to 70% less) is often only required to produce a digital image.
Digital Radiography Systems Advantages and Disadvantages.
over CR. Some of these include:
• Lower doses—the PSP plates used in CR have a lower efficiency of detection compared to DR detectors. Thus, a higher radiation dose with CR is needed to obtain adequate image quality.
• Greater ease of use and faster patient throughput
Immediate imaging results: CR requires the cassette be removed from the x-ray machine and then placed into a reader. This is a labor-intensive step that re­quires the technologist to leave the patient and workstation with each imaging procedure, even if for a short time.
• Additional image manipulation
• Much less overall maintenance (e.g., avoidance of CR cassette testing, cleaning, and storage space requirements) There are, however, several disadvantages of DR rela-
tive to CR, namely10:
• DR is much more costly
• CR is compatible with a wide range of preinstalled traditional systems
• Multiple cassette sizes with CR allow for greater flex­ibility than the single detector size of DR
• The CR PSP imaging plates are subject to mechanical damage and also to chemical oxidation but can be replaced without great expense. DR receptors, while well protected from mechanical damage, do not last
DR systems offer several advantages
forever; they experience gradual aging processes, and their replacement cost is high.
An additional prospective disadvantage of DR relative to CR is that most DR systems either do not allow the user to change the grid to accommodate the imaging task or have a preinstalled grid that is not easily accessi­ble. These conditions can result in the use of grids for pediatric imaging, thereby unnecessarily increasing a child’s radiation dose. Because of this, DR facilities need to collaborate with their radiation safety officer (RSO) and physics group more than ever before to ensure the highest-quality low-radiation dose imaging for the smallest patients. During acceptance testing of a new DR system, a technologist should inquire whether or not grids are removable from the digital imaging equipment.

Repeat Rates in Digital Imaging

Because image contrast and overall light intensity may be manipulated after image acquisition, digital imag­ing eliminates the need for almost all repeat images required as a result of improper technique selection (Fig. 11.15). However, repeat rates for reasons of poor positioning are not lowered. Since, for DR, the image receptor is part of the imaging equipment and does not need to be removed for processing, the technolo­gist can simply view the image on a monitor at the control panel. This raises a concern regarding the number of repeats required due to mispositioning. There is no direct “penalty” to a technologist for repeating images because of poor technique, so the examination and resultant radiation exposure could potentially be delivered multiple times, instead of just once, without the knowledge of supervisors.
Early reports have indicated an increase in repeat rates in DR imaging. The increase is fundamentally due to the ease of repeating an image. It appears that technologists are in many cases seeking to obtain the perfect image, without the consideration of increased exposure to the patient.
Digital imaging systems, both DR and CR, separate the process of acquisition from the display and thereby permit a relatively large degree of viewing manipulation of the raw images by the radiographer and/or the radi­ologist upon examination. Since digital devices also typically have wide exposure latitude,* this can result in
*Exposure latitude refers to how much an image receptor can be overexposed or underexposed and still be able to yield a useful image.
CHAPTER 11 Equipment Design for Radiation Protection
1.6 mAs/70 kVp 3.2 mAs/70 kVp 6.4 mAs/70 kVp 12.5 mAs/70 kVp 25 mAs/70 kVp
A
211
2.5 mAs/70 kVp 5 mAs/70 kVp 10 mAs/70 kVp 40 mAs/70 kVp 80 mAs/70 kVp
B
Fig. 11.15 (A) The images obtained with a screen-film image receptor system illustrate how changing techni-
cal exposure factors greatly affects film image quality. (B) Computed radiography (CR) images obtained through the same technique ranges as those used for (A) have much less effect on image quality because “CR image contrast is constant, regardless of radiation exposure.” (From Betsy Shields, Presbyterian Hospi­tal, Charlotte, North Carolina. In Bushong SC: Radiologic science for technologists: physics, biology and protection, ed 11, St. Louis, 2017, Elsevier.)
a wide range of received patient doses, from very low to extremely high.
An “appropriate” patient exposure is the lowest value that is needed to provide a resultant image of sufficient quality for a radiologist to confidently make an accurate differential diagnosis. However, except for extreme overexposure or underexposure, both DR and CR images can be manipulated, because of the separation
between acquisition and display, to exhibit satisfactory contrast resolution sensitivity. This is predominately due to the ability of digital detector systems to adjust or rescale the received projection data to a useful grayscale viewing scope or range. Thus, it may be so, in many situations, that the patient has needlessly received a larger radiation exposure often without the knowledge of anyone involved in the diagnostic reading of the case.
212
In some exceptional situations, an overexposure factor of three or more might happen. In general, it has quite often been discovered that a phenomenon known as “dose creep” occurs due to the lack of negative impact observed when the patient’s anatomy is overexposed but the display still exhibits quality images.
To remedy potential overexposure and thereby adhere to as low as reasonably achievable (ALARA) practices, either each image taken by a digital system could be monitored by an independent quality control technolo­gist at a separate monitor, which might be quite tedious, or a quality control exposure counting system could be devised whereby for each technologist the number of images per examination is compared with the quantity ordered.
Alternatively, there is, however, a safety feature in­stalled with most digital detector systems: an exposure index (EI) indicator, derived from data collected with anthropomorphic phantoms, that supplies the relative speed and sensitivity of the digital receptor to incident x-rays. The EI value ideally provides a guide to the tech­nologist regarding the proper radiographic techniques to select that will yield an optimal image for a specific examination in terms of both acceptable image quality and patient radiation dose. Usually, during commis­sioning of a new digital system, a table of appropriate EIs will be determined for that system.
CHAPTER 11 Equipment Design for Radiation Protection

RADIATION SAFETY FEATURES OF FLUOROSCOPIC EQUIPMENT, DEVICES, AND ACCESSORIES

Fluoroscopic Procedures and Patient Irradiation Rates

Fluoroscopy is a continuous irradiation process that demonstrates dynamic motion of or through selected anatomic structures (e.g., a stomach filled with barium sulfate and air during an upper gastrointestinal series) by generating and displaying real-time imaging of those structures on a monitor that works in conjunction with an image signal amplification system. Fluoroscopic pro­cedures (Fig. 11.16) produce the largest patient radia­tion exposure rates in diagnostic radiology. Therefore, the referring physician should carefully evaluate whether the potential benefit to the patient, in terms of informa­tion gained, outweighs any adverse somatic or genetic
Fig. 11.16 Fluoroscopic procedures produce the largest pa-
tient radiation exposure rate in diagnostic radiology.
effects of the examination. If the fluoroscopic procedure is necessary, every reasonable effort must be taken to minimize patient exposure time.

Fluoroscopic Imaging Systems: Non-digital

Brightness of the Fluoroscopic Image and Patient Absorbed Dose.
Non-digital fluoroscopy (Fig. 11.17) involves the use of a signal amplification device called an image intensi- fier (II) tube (depicted in Fig. 11.18 and discussed in great detail in the next section). This apparatus mark­edly increases the brightness or intensity of the real­time image produced on a screen during fluoroscopy. II fluoroscopy has three significant benefits, which are listed in Box 11.4
The x-ray image intensifier (II) converts the pattern of x-rays transmitted through the patient into a corre­sponding and amplified visible light pattern. With an II the overall illumination of the raw fluoroscopic image is increased to roughly 10,000 times that of the image ob­tained by non-II fluoroscopic systems* while operating
*A pre-image intensification fluoroscopic system consisted of the fluoroscopic tube, mounted beneath the radiographic ta­ble, producing x-rays that passed through the tabletop and the patient before striking a zinc-cadmium sulfide (ZnCdS) screen. The latter then phosphoresced, producing an image of the anatomy of interest. Unfortunately, this image was very dim, and to be adequately discerned by the radiologist, a pho­tographic dark-room situation always had to be created.
CHAPTER 11 Equipment Design for Radiation Protection
Ceiling-mounted radiographic X-ray tube
213
Flat panel monitors
Image intensifier
Spot-film cassette
Bucky slot cover
Fig. 11.17 Image intensification fluoroscopy unit. The x-ray tube used in this unit is mounted beneath the
unit’s radiographic table, which supports the patient. The image intensifier and other image detection devices are then drawn forward and placed over the patient on the table to perform the examination. Other fluoro­scopic equipment arrangements are possible. (From Bushong SC: Radiologic science for technologists: phys- ics, biology and protection, ed 10, St. Louis, 2013, Elsevier.)
tube
Variable aperture collimator
Protective curtain
Fluoroscopic x-ray tube under table
Technologist control
Cassette tray for overhead radiography
’s
under the same conditions. This very large increase in brightness has dramatically improved the radiologist’s perception of the fluoroscopic image and at the same time enabled a decrease in patient absorbed dose. Image intensification fluoroscopy requires much smaller milli­amperage than does pre-II fluoroscopy (approximately 1 to 1.5 mA and even less can now be used for many pro­cedures whereas 3 to 5 mA and more was usually re­quired before image intensification fluoroscopy). The resultant decrease in exposure rate, for similar examina- tion durations, yields a sizable dose reduction for the patient.
Image Intensifier Tubes and Magnification.
An II tube is an “electronic device that receives the image-forming x-ray beam and converts it into a visible­light picture of high intensity.”1 A simple diagram of this tube with components labeled is shown in Fig. 11.18. Magnification, or multifield, features are present in the vast majority of II systems. They are also found in digital
fluoroscopy (DF) units (see the discussion on DF pre-
sented later in this chapter). Multifield image intensifica­tion tubes vary in size, but the 30/25/20 cm (12/10/8 inch) diameter tri-field model is typical for general­purpose fluoroscopic units. However, other sizes and
214
CHAPTER 11 Equipment Design for Radiation Protection
Output phosphor
Anode
Focal point
Electrostatic
lenses
Electrons
Glass
envelope
Fig. 11.18 Basic components of an image intensifier tube.
(From Bushong SC: Radiologic science for technologists: phys­ics, biology and protection, ed 10, St. Louis, 2013, Elsevier.)
Photocathode
Input
phosphor
BOX 11.4 Benefits of Image
Intensification Fluoroscopy
Increased image brightness
Saving of time for the radiologist
Patient dose reduction
If magnification in the fluoroscopic image is desired, the viewing mode can be changed to the smaller 25 cm mode (10 inches) or even less (e.g., 20 cm or 8 inches in many new systems). With this selection, the voltage on the electrostatic focusing lenses is increased, thereby causing the focal point of the electrons to move closer to the input phosphor surface or a greater distance away from the output phosphor.1 As a result, only photoelec­trons from the central 25 cm, or 20 cm, diameter por­tion of the input phosphor, instead of its entire surface area, actually reach the output phosphor of the image intensifier. This added distance from the focal point lo­calization of the photoelectrons to the output phosphor surface, however, creates a larger image but with a de- creased FOV (Fig. 11.19).
In magnification mode, the quality of the enlarged image, if there are no other changes, as viewed on a monitor, is somewhat degraded. This reduction in image clarity occurs because of the decrease in minifica- tion gain (i.e., an increase in brightness) when fewer photoelectrons are available to strike the output phos­phor on the image intensifier. Therefore, the resultant image is dimmer. Because it is necessary and desirable to maintain a constant level of light intensity on the
25-cm
focal point
30-cm
focal point
magnification modes are also available.* When the normal viewing mode of 30 cm (12 inches) is used, photoelectrons emitted from the entire surface of a con­vex shaped cesium iodide (CsI) phosphor (i.e., the input surface when the x-ray photons passing through the patient first strike the II assembly) are accelerated and converge on a focal point from which they subse­quently spread out again and advance to the surface of a zinc-cadmium sulfide output phosphor as shown in
Fig. 11.18.
*Many current II systems are available with four selectable viewing modes or magnifications: e.g., 20 cm, 17 cm, 15 cm, 12 cm: 12”, 9”, 6”, 4” are other available selections.
25 cm
17 cm
Fig. 11.19 A 30/25/20 image intensifier tube produces a mag-
nified image in 25 cm mode, whereas the 20 cm mode pro­duces an image that is even more highly magnified. (From Bushong SC: Radiologic science for technologists: physics, bi- ology and protection, ed 10, St. Louis, 2013, Elsevier.)
CHAPTER 11 Equipment Design for Radiation Protection
215
view monitor, fluoroscopic mA increases automatically to counter this illuminance loss. The overall quality of the image will now be enhanced relative to the larger­diameter modes because a greater number of x-ray photons are being used to form the magnified image. This image will have a more even appearance (less noise), and it will be possible to distinguish among similar tissues more readily because of improved con­trast. However, the increase in tube mA necessarily raises the dose to the patient. Thus, magnification modes are only used when diagnostically needed.
An image intensifier can also be designed to inter- actively change the input FOV from a large to a smaller area. If the input FOV is halved, then the region of the patient being observed is also halved, which results in two-fold magnification of the image.
Pulsed Fluoroscopy. Pulsed, or intermittent, fluoros-
copy involves manual or automatic periodic activation of the fluoroscope x-ray tube by the fluoroscopist, rather than continuous activation. This practice:
• Significantly decreases patient dose, especially in
long procedures
• Helps extend the life of the tube
In pulsed mode the system software automatically turns the radiation beam on and off at an operator­selected repetition rate. The most common rates are 30, 15, and 7.5 pulses per second, with each radiation pulse lasting no more than 10 milliseconds. Shorter pulse durations lessen the effects of any patient motion thereby improving the sharpness of the image but de­crease the signal-to-noise ratio because of the smaller number of x-rays involved. A 30 p/s rate would typi­cally be used for imaging studies involving very rapid anatomic or process motion (e.g., interventional cath­eter procedures) to achieve acceptable temporal resolu­tion. Barium swallow studies, on the other hand, could make use of 7.5 p/s. Many systems include a last image hold feature that allows the fluoroscopist to momen­tarily halt the radiation and review the most recent image before giving the patient another pulse of radia­tion. Frequently utilizing the last image hold feature during long procedures will noticeably reduce patient absorbed dose.
Limiting Fluoroscopic Field Size. The radiologist must
limit the size of the fluoroscopic field to include only the
area of clinical interest. This involves visually moving the shutters placed between the x-ray tube and the pa­tient to define the desired field of view. When fluoro­scopic field size is limited, patient integral dose (i.e., volumetric dose) decreases substantially. For lengthy examinations, however, it is possible to spread out the patient entrance irradiation area and thereby minimize the possibility for skin effects while maintaining the same anatomic field of view. This spreading out is ac­complished by moving or rotating the patient so that the radiation enters at multiple portions of the patient surface instead of just the same limited region during the procedure.
The selected primary beam length and width must always be confined within the image receptor bound­ary. Regardless of the distance from the x-ray source to the image receptor, the useful beam should not extend outside the image receptor. Ideally, visible bor­ders ought to appear on the image monitor. If this is not so, a patient could receive a substantial radiation dose in certain procedures to sensitive areas adjacent to the study area. Thus, this conformity (alignment and congruence) between the x-ray field and the input phosphor of the II is an essential item of regulatory concern.
Radiation Delivery Factors
Selection of technique exposure factors for adult patients. During manual fluoroscopic procedures, the
fluoroscopist must select technical exposure factors that will minimize patient dose. Increasing the kVp and filtration reduces the patient exposure and dose rate. Most fluoroscopic examinations employ a range of 75 to 110 kVp for adult patients, depending on the body area being examined. This kVp range produces the cor­rect level of fluoroscopic image brightness. Lower kVp increases patient dose because a less penetrating x-ray beam necessitates the use of a higher milliamperage to obtain adequate image intensity. The operator can fur­ther limit excessive entrance irradiation of the patient by ensuring that the x-ray SSD is not less than 38 cm (15 inches) for stationary (fixed) fluoroscopes and not less than 30 cm (12 inches) for mobile fluoroscopes (C­arms). A 30-cm (12-inch) minimal distance is required, but an increased distance to 38 cm (15 inches) is pre­ferred for all current systems. In addition, positioning the II input phosphor surface as close as is practical to
216
CHAPTER 11 Equipment Design for Radiation Protection
the patient will further reduce the patient’s entrance exposure and dose rate.
Selection of technique factors for children. Quality
radiation safety practices with fluoroscopic procedures for children necessitates a decrease in kVp by as much as 25%. The kVp chosen should depend on anatomic part thickness, just as it does in radiography. In addition to decreasing kVp, maintaining SSD and minimizing the height of the II entrance surface above the patient will further limit excessive entrance irradiation of the pedi­atric patient.
Filtration. The function of a filter in fluoroscopy, as
in radiographic procedures, is to reduce the patient’s skin absorbed dose from soft x-rays. Adequate layers of aluminum equivalent material placed within the collimator assembly in the path of the useful beam remove the more harmful lower-energy photons from the beam by absorbing them. A minimum of 2.5 mm total aluminum equivalent filtration must be perma­nently installed in the path of the useful beam of the fluoroscopic unit. With current systems, a total alu­minum equivalent filtration of at least 3.0 mm or markedly higher is typical. Although increases in fil­tration cause a loss of fluoroscopic image brightness, using higher kVps can somewhat compensate. As in radiography, the minimum permanent filtration of the x-ray beam is federally mandated, and therefore the HVL of the beam must be measured to confirm agreement with regulatory standards. In routine fluo­roscopy, an x-ray beam HVL of 3 to 4.5 mm alumi­num is considered acceptable when kVp ranges from 80 to 100.
Cumulative Timing Device. A cumulative timer must
be provided and used with each fluoroscopic unit. This resettable device measures the collective x-ray beam-on time and sounds an audible alarm or, in some cases, temporarily interrupts the radiation until it is reset after the fluoroscope has been activated for 5 minutes. It also serves to alert the fluoroscopist to the amount of time the patient has been receiving x-ray exposure. Activat­ing the fluoroscope for shorter periods will not only cause the patient to receive less radiation exposure but also the fluoroscopist and the radiographer. Total fluo­roscopic beam-on time should be documented for every fluoroscopic procedure.
Entrance Irradiation Rate Limitations. Current federal
standards limit entrance skin irradiation rates* of general-purpose intensified fluoroscopic units with maximum technique factors engaged to a maximum of 88 mGya per minute entrance absorbed dose rate (or 10 R/min entrance exposure rate), as measured at the tabletop using an average phantom, with the II entrance surface at a prescribed 30 cm (12 inches) above the table­top. This limit for conventional fluoroscopic systems has been imposed to provide protection against patients accidentally receiving skin-damaging entrance dose levels in short periods. Fluoroscopic units equipped with high-level control (HLC), however, may produce a skin entrance dose rate as great as 176 mGya per minute (20 R/min entrance exposure rate). Because certain lengthy fluoroscopic procedures can result in the largest patient doses in diagnostic x-ray imaging, sometimes reaching the level of therapeutic doses with potential physical damage to the patient, a concerted effort must be made to keep fluoroscopic exposure rates and cumu­lative exposure times within established limits.
Primary Protective Barrier. A primary protective bar-
rier (i.e., shielding in the direct line of the transmitted
patient irradiation) of 2 mm lead equivalent is required by regulatory guidelines for a fluoroscopic unit. The II assembly or a digital detector is designed to provide this barrier to direct radiation. The assembly must be physi­cally joined with the x-ray tube and interlocked so that the fluoroscopic x-ray tube cannot be activated when the II or other detection system is in a parked, un­aligned, and unconnected position.
*One must be precise in the terminology used for entrance or tabletop fluoroscopy radiation levels, e.g., there is the term, entrance or tabletop exposure rate, which has been and still is specified by regulatory agencies and medical physicists in R/min, and on the other hand there is the usage of mGya/min which is correctly described as entrance or tabletop absorbed dose rate and sometimes even as exposure dose rate. Fluoro­scopic dose rate and exposure rate are not the same. There is a numerical conversion factor existing between them, called the rads (cGy) per roentgen factor “f,” which is about 0.88. The f factor reflects the fact that not all incident or entrance x-ray photons interact with and deliver energy to surface tissue at­oms. Thus, an entrance exposure rate of 10 R/min in reality corresponds to an entrance absorbed dose rate of about 8.8 cGya/min or 88 mGya/min.