Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3585_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
102 Мб
Скачать
160
ab
2 Stenting ofBifurcation Lesions
P
SB
Fig. 2.6 Positioning a dedicated bifurcation stent. A dedicated stent
positioned optimally for access to the SB (a) and a stent with the carina positioned on the opposite side (b)
Fig. 2.7 Failure to optimally position a dedicated stent. It may be
impossible to place the center of the petal of a dedicated stent exactly at the center of the SB lumen, as shown on the left
MB
Fig. 2.9 Boston Scientic Petal stent. The stent is mounted on the
dedicated stent delivery system shown on the left. Ination of the spher­ical balloon at the SB ostium extends the petal obliquely into the SB.Subsequently, KBI is done to optimally reshape and align the petal
a
b1 b2 b3 b4
carina
Fig. 2.8 Incorrect deployment of a dedicated stent proximal to the
carina. It is necessary to reinsert a guidewire/balloon into the SB through a cell closer to the carina and perform KBI again
Fig. 2.10 Abbott Xience SBA stent (a). This dedicated stent is suitable
for provisional stenting, which is simply accomplished by simultaneous stent implantation and KBI.The stent delivery system mounts a single stent on two balloons and accommodates the SB guidewire near the tip (b1). The stenting procedure includes the following steps: advancing the stent until the SB balloon is just beyond the carina, releasing the SB guidewire tip and withdrawing the wire, pulling the entire system back slightly, advancing the guidewire with a reshaped tip into the SB (b2), advancing the stent until its carina meets the vascular carina (b3), and performing local KBI (b4)
2.1 Dedicated Bifurcation Stents
161
Fig. 2.11 Advanced Bifurcation Systems stent
Fig. 2.12 Medtronic Y stent
162
2 Stenting ofBifurcation Lesions
a b
Fig. 2.13 Tryton stent (a, b)
Fig. 2.14 STENTYS stent
2.2 Optimizing General-Purpose Stents forBifurcation Lesions
Fig. 2.15 AXXESS stent
2.2 Optimizing General-Purpose Stents forBifurcation Lesions
As shown in Fig. 2.16, the ideal bifurcation stent should achieve excellent apposition of a single layer of struts over the entire vessel wall at the bifurcation without causing SB or MB jailing. A general-purpose stent that can provide the ideal bifurcation scaffold needs to meet the following requirements:
1. High exibility (conformability): Bifurcation lesions are
often tortuous (Fig.2.17), and many side branches have a steep take-off angle (Fig.2.18). Hence, a stent for bifur­cation lesions requires high conformability. Highly con­formable stents usually have little edge effect. In general, stents with fewer links are more conformable.
2. No risk of SB jailing after optimal KBI: The optimal stent
for a bifurcation lesion can dilate the lesion correctly without causing SB jailing if optimal KBI is performed, which is dened as ination of appropriately sized kiss­ing balloons at the optimal pressure that does not injure the proximal MB after SB (re)wiring through the most distal cell near the carina.
3. Minimum risk of malapposition: The optimal stent for
bifurcation lesions must have high overall conformability as well as being easily deployed by balloon ination. Ease of stent expansion is inuenced by the design as well as the material used to make it.
To achieve optimal apposition in a bifurcation lesion, a stent with fewer links and with more and longer crowns is preferred. In addition, a stent made of a material with higher plasticity (e.g., stainless steel or platinum­chromium alloy) is better able to undergo plastic defor­mation when implanted at a complicated bifurcation
163
lesion and thus achieve optimal apposition. A stent made of cobalt-chromium alloy, which has lower plasticity (i.e., higher elasticity), may be slightly inferior with respect to ne plastic deformation at bifurcation lesions.
However, a stent with higher elasticity can undergo greater distortion before reaching the limit of elasticity, which makes the occurrence of fracture (especially acute fracture) or dynamic deformation less likely. Thus, the balance between elasticity and plasticity of the stent mate­rial leads to signicant trade-offs for stent performance.
4. Little or no risk of fracture: Stents with no risk of fracture have long been awaited, but no such stent has been devel­oped so far. I consider that stent fractures can be classied as “good” or “bad.” It would be ideal to completely pre­vent stent fractures, but this is impossible. Therefore, stents should be designed so that any fracture is likely to be a good fracture.
5. Drug-eluting: Unlike metal stents, BVS cannot be used extensively for bifurcation lesions at present because per­forming KBI leads to collapse of the scaffold. DES will remain the standard type of stent for bifurcation lesions, at least for a while.
6. Polymer-free: A polymer coating may peel off after KBI or culotte stenting. The peeling polymer can lead to thrombosis, even if it is bioresorbable, and there is less drug release from the part of the stent where the coating has been lost. Therefore, a polymer-free DES is prefera­ble for bifurcation stenting.
7. Proximal edge structure compatible with the aorto-ostial anatomy: The aortic ostium, whether normal or occluded/ stenosed, can usually be dilated to its original diameter, although its anatomy may prevent this. When treating an LMT bifurcation lesion, it is often necessary to dilate the bifurcation through the LMT ostium. Since the ostium can be as large as 6mm in diameter, a stent for an LMT bifurca­tion lesion has to be expandable to a large diameter. On the other hand, the part of the stent implanted outside the ostium may be deformed by a guiding catheter or other device. To prevent such stent deformation that could impair the system integrity, the stent must have two links within 2–3mm of the proximal edge. Two links are both necessary and sufcient for preserving stent conformability and preventing defor­mation, while three or more links are not required.
A stent that meets all of requirements (1) through (4) above should have no links, but it is virtually impossible to implant a no-link stent without marked deformation. Hence, the ideal stent for bifurcation lesions should have links that can prevent excessive deformation during implantation and lose these links as soon as possible after implantation.
If all of the stent links facing the SB ostium can be fractured during KBI, there is no risk of SB jailing. Thus, I would like to discuss the concept of a stent with tempo­rary (easily fractured) links later.
164
ab
Fig. 2.16 Optimal bifurcation stents. Whether
bifurcation stenting is performed by the MB crossover technique (a) or the MB/SB full coverage technique (b), the optimal stenting strategy must meet the following requirements: (1) complete coverage, (2) complete apposition, (3) complete expansion, and (4) optimal stent design
Fig. 2.17 A tortuous
bifurcation lesion
2 Stenting ofBifurcation Lesions
Fig. 2.18 A bifurcation
lesion with extreme SB angulation
Column 13 Good Versus Bad Fracture
[Bad fracture]
1. When one of the links is fractured in a stent with low conformability (e.g., a multi-link stent), both edges of the stent will become more mobile, and this will cause static or dynamic edge effects (Fig.2.19).
2. When the long link in a stent is fractured, the stumps of the link may cause stimulation of the intima or malapposi­tion can occur at the fractured link.
2.2 Optimizing General-Purpose Stents forBifurcation Lesions
3. A stent with at least one link can be fractured at a strut rather than at a link (Fig.2.20). If strut fracture occurs, the stumps of the strut will stimulate the intima. In addition, the fractured strut loses its radial strength due to loss of continuity of the supporting arch. Thus, the fractured strut can easily be deformed by vascular recoil, narrowing the lumen that was previously expanded by balloon ination of the stent. This mechanism may be involved in some cases of in-stent occlusion (restenosis) at a fractured strut.
[Good fracture]
When can stent fracture be considered good? A good fracture can occur if a stent is designed with fracturable short
links between the strut peaks and valleys (Fig.2.21). If fracture occurs at a short link, it is unlikely that the stumps of the fractured link will stimulate the intima or that the strut will lose radial strength. These outcomes are even more unlikely if the stent has few links and high conformability.
I have long sought for a stent that can undergo good fracture. The Driver stent initially appeared to be suitable, but
it was found that fracture occurred at the struts and not at the links. However, I have found that the Nobori stent (par­ticularly the 3.5-mm stent) is more likely to be fractured at a link than at a strut. This may account for the low restenosis rate after implanting a Nobori stent, although this type of stent is liable to fracture if used at the RCA ostium. A modi­ed Nobori stent that is much more likely to undergo fracture at a link than a strut is needed. Moreover, a stent with no links or one designed with disappearing links could provide a good solution to the problem of stent fracture.
165
Fig. 2.19 Bad fracture. The parts of the stent proximal and
distal to the fracture have shifted horizontally in opposite directions
Fig. 2.20 Another example of bad fracture. The stent strut
has been fractured
Fig. 2.21 Good fracture. The stent has fractured at a short
link
166
cell After disconnection
2 Stenting ofBifurcation Lesions
[Stent “recoil”]
I often feel that something is wrong when I hear someone referring to stent deformation by external force as stent
recoil. This is because recoil means spontaneous movement of an object to return to its original shape. Thus, stent recoil describes the tendency of a stent that has been expanded by a balloon to shrink slightly due to its inherent elastic­ity after deation of the balloon. Partial luminal loss after balloon deation when implanting a balloon-expanded stent cannot be referred to as stent recoil because this phenomenon arises from recoil of both the vessel and the stent. A stent will show minimal deformation if it is implanted in a soft lesion, but severe deformation will occur after implantation in a hard lesion that has not been prepared sufciently.
In the POBA era, sudden luminal loss at the end of intervention was referred to as acute recoil, while late lumen loss
attributable to restenosis was called chronic recoil. Analogous with vascular luminal loss, narrowing of the stent lumen may therefore have also been referred to as recoil. While “recoil” may reasonably be used to describe luminal loss due to spontaneous narrowing of the target vessel after POBA, should it be used to describe narrowing of the stent lumen? It is clear that stent luminal loss is largely due to deformation caused by external forces, rather than spontaneous motion of the stent. In particular, chronic stent deformation (late stent luminal loss) is never due to stent recoil. If the term “stent recoil” is used to describe stent deformation, it does not accurately express the essence of this phenomenon. It may lead to classication of some stents as prone to “recoil” and possibly to an inappropriate focus on stent-related factors when investigating the mechanism of stent deformation. The vascular factors potentially involved in stent lumen loss should also be considered, in order to elucidate the mechanism applying external force to the stent and to devise countermeasures. In this sense, I think that we should pay careful attention to use of the term “recoil.”
2.2.1 Temporary Link Stents
2.2.1.1 Concept ofthe“Temporary Link Stent”
I use the term “temporary link stent” to describe a stent with intact link(s) until it is expanded in a vessel, after which the link(s) are fractured or lost because of vascular movements or with the passage of time. A link facing the SB ostium can be easily fractured by application of horizontal force during KBI, and the stumps of the link will not stimulate the vessel. After loss of the links, it is desirable that the stent does not break up into rings and that continuity of the struts is maintained.
2.2.1.2 Classication ofTemporary Link Stents
1. Bioresorbable link stent: The links of the stent are made of a bioresorbable (drug-eluting) polymer and disappear from several months to about 1year after implantation.
2. Separable link stent: The links are made of metal but are designed to be easily disconnected by KBI and by vessel movements. If the stent is fractured at a link, the stumps of the link do not stimulate the vessel, so it can be inten­tionally fractured at a link to ensure high conformability while preventing vascular injury.
A stent with disconnectable links (Fig. 2.22) has been
developed, but it has not been marketed in Japan so far.
cell
Disconnection
of link
Fig. 2.22 A stent with disconnectable links. Balloon ination in the
SB disconnects a link, producing a larger cell and resolving SB jailing
1236
2.2 Optimizing General-Purpose Stents forBifurcation Lesions
167
2.2.1.3 Helical Coil Stents
Loss of links in a temporary link stent means that the individ­ual strut rings are independent of each other in the vessel lumen and malapposition may allow movement or shift of the stent struts. To preserve the integrity of a stent after loss of all links, it has to have a helical coil structure. Helical coil stents have very high conformability. Even if overlapped, such stents rarely become rigid and are unlikely to undergo strut fracture.
However, if a helical coil stent deployed in an aorto-ostial
lesion is not optimally apposed (even temporarily), longitu­dinal deformation can readily occur, and the stent may some­times be extruded into the aorta. If used for an aorto-ostial lesion, a helical coil stent should have several proximal struts with a durable 2-link pattern.
2.2.1.4 Optimal Design ofaStent forUse
withKBI
The optimal stent for deployment with KBI should be designed as follows:
1. It should permit sufcient enlargement of a strut cell for side branch access.
2. It should have few links (Figs.2.23 and 2.24) to permit sufcient enlargement of a cell while causing less vascular damage near the SB ostium and to minimize the risk of SB jailing or stent malapposition.
3. It should have disconnectable links (refer to Fig.2.21).
4. It should provide an adequate scaffold for SB shoulder support (Fig.2.24b).
5. Its struts should be uniformly expandable in every direc­tion by KBI.
A laser-cut stent, polished and bordered, generally has
struts that are square in the cross-sectional view. A strut with a square cross section can be bent easily by application of force in a direction parallel to a side of the square, but not in a direction that is at an angle to the side. Such a strut is most difcult to bend at an angle of 45 degrees to the side of the square.
A strut with a rectangular cross section can easily be bent
by application of force in the direction of the shorter side, but not in other directions, especially the direction of the longer side. A lamentous strut with a round cross section can be bent in every direction by application of identical force (Fig.2.25).
In principle, a stent with cylindrical struts is considered
to be best able to t into the two branches. On the other hand, cylindrical struts have a smaller contact area with the vessel wall and are prone to excessive deformation when contacted by a balloon. Thus, designing a stent with cylin­drical struts involves a trade-off between these two charac­teristics, although the actual clinical performance of stents with different strut shapes for bifurcation stenting is uncertain.
At least, it can be said that a stent with rectangular struts
is not suitable for bifurcation stenting. A stent with cylindri­cal struts is the best option, while a stent with rectangular struts (chamfered squares in cross section) is second best.
Fig. 2.23 Design of 6-, 3-,
2-, and 1-link stents. A stent with fewer links is less likely to cause link-induced SB jailing
168
2 Stenting ofBifurcation Lesions
Fig. 2.24 Stents expanded by KBI exvivo. (a)
Expanding a 3-link stent results in enlargement of a strut cell at the SB ostium to an elliptical hole about 3mm in diameter with no zigzag strut. Hence, the lateral border of the hole () is proximal to the stent carina, with the average distance being about 2mm or slightly more. (b) Expanding a 2-link stent results in enlargement of a strut cell at the SB ostium to about 3mm in diameter while retaining the zigzag structure of the struts. This allows the stent to provide a more adequate scaffold for the SB shoulder (arrow) compared with a 3-link stent
a
Ca 2mm
b
Fig. 2.25 Cross-sectional views of stent struts. A strut with a rectangu-
lar cross section can easily be bent if force is applied in the direction of the shorter side, but not in the direction of the longer side. A strut with
2.3 Optimal Stenting Techniques forBifurcation Lesions
2.3.1 Need forKBI Pre-dilatation andPractical Approach
Although many interventionalists question the use of KBI for pre-dilatation, I often perform this technique for pre­dilatation of a bifurcation lesion with plaques around the
a square cross section can be bent more easily if force is applied in a direction parallel with its sides, but not in a direction at 45 degrees to a side. A cylindrical strut can be bent in any direction by the same force
carina (including the SB ostium), provided the diameter of the proximal MB is signicantly larger than that of the dis­tal MB.The objectives of performing KBI in this situation are (1) to assess the responsiveness of the proximal MB to balloon angioplasty (Figs.2.26 and 2.27) to reduce the risk of plaque shift after stent implantation and the risk of occlusion/severe stenosis of the SB ostium by carina shift during alternating balloon ination (Fig. 2.27). For pre­dilatation, I rst perform KBI at a pressure of only 6–8atm.
a
b
c
a
b
c
g
2.3 Optimal Stenting Techniques forBifurcation Lesions
If such low- pressure KBI cannot dilate the lesion suf­ciently, I perform alternating balloon ination at escalating pressures until sufcient dilatation of the SB ostium and the distal MB is achieved. I then repeat low-pressure (about 10 atm) KBI and check if there is any residual indentation.
Even if alternating balloon ination seems to have removed all indentations in the MB and SB balloons and to have sufciently dilated both branches, one or more indenta­tions may reappear in the body of either or both balloons during KBI. Such pre-dilatation failure typically occurs in patients who have an LAD-diagonal branch (Dg) ostial lesion with hard plaque at the shoulder of the branch. In such lesions, alternating balloon ination often causes carina shift (Fig. 2.28) that obscures balloon indentations, so that it appears that the SB ostium has been dilated sufciently. For optimal preparation of these lesions, I have often used a scor­ing balloon in recent years. The Lacrosse NSE is my favorite scoring balloon. In general, I combine scoring balloon angio­plasty and POBA as required to prepare bifurcation lesions according to the size of the MB/SB and the pattern of steno­sis (Fig.
2.29). IVUS guidance is absolutely required for
optimizing the combination of these angioplasty procedures.
169
Fig. 2.26 Need for KBI in pre-dilatation. At a bifurcation stenosis (a),
alternating balloon ination appears to have sufciently dilated the MB and the SB since there are no indentations when the balloon is inated in either branch (b). However, alternating ination has actually caused carina shift and deformation of the SB shoulder, resulting in transient elimination of indentations in the body of the SB balloon so that it appears as if the SB ostium has been sufciently dilated (b). During KBI, one or more indentations may reappear in the SB balloon, but such indentations can be eliminated by raising the KBI pressure to 8atm. As a result, both the MB and SB can be dilated sufciently as optimal preparation for stenting with KBI (c)
d
e
h
f
Fig. 2.27 Need for KBI in pre-dilatation of true bifurcation lesions. If
pre-dilatation of a bifurcation lesion (a) is performed by alternating bal­loon ination, the carina shifts toward the SB after ination of the MB balloon (b) and then shifts back toward the MB after ination of the SB balloon (c). After balloon deation, it is found that only minimal dilata-
tion of the lesion has been achieved (d). Following minimal pre­dilatation of a bifurcation lesion, placing a stent in the MB can cause plaque shift that leads to severe stenosis or occlusion of the SB (e). With pre-dilatation by KBI (f), the lesion will be optimally dilated (g), mini­mizing the risk of SB occlusion (h)