Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3585_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
102 Мб
Скачать
1.6 Antegrade Approach
89
If you can conrm that the guidewire has crossed the occlusion through the true lumen, you should advance a microcatheter over the guidewire, exchange the guidewire for another with a less stiff tip, and deliver the appropriate devices to dilate the lumen. If collateral imaging fails to provide clear conrmation that the guidewire has crossed the occlusion through the true lumen, you should check for free backow of blood from the microcatheter after with­drawing the guidewire and then perform tip injection. After conrming that the guidewire crossed the CTO through the true lumen, you should record the conrmation.
1.6.5 Microcatheter
You must always use a microcatheter to assist the guide­wire in crossing a CTO.Using a microcatheter allows you to exchange guidewires and to nely adjust the orientation, movement, and backup of the guidewire.
1.6.5.1 Selection oftheMicrocatheter
Which microcatheter should be chosen for antegrade PCI? The minimum requirement to be met may be the ability to improve the manipulability of the guidewire more. In partic­ular, friction between the microcatheter and guidewire needs to remain relatively constant, even after passing through a very tortuous vessel (Fig.1.131).
In addition, the microcatheter needs to have a tip that can easily enter an occlusion when advanced over a guidewire. To meet this requirement, the microcatheter should have a tip that is as small as possible and a shaft that is as slim as pos­sible. A microcatheter with superior crossability even with rotation is better. A desirable microcatheter can be used to advance a Conquest Pro 12 or another guidewire that has a high tip load without perforation of the microcatheter by the guidewire tip, even when the tip is located at a curve of the target vessel. There is no ideal microcatheter among the many currently available, so the optimal microcatheter should be selected depending on the morphology and hard­ness of the target lesion.
Fig. 1.131 CTO of an RCA with a very tortuous proximal segment. In this situation, using a soft microcatheter will cause greater friction with
the guidewire and reduce its manipulability
90
1 Mitsudo’s PCI Techniques forCTO
1.6.5.2 Position oftheMicrocatheter Tip
When manipulating a guidewire toward a CTO through a microcatheter, you should place the tip of the microcathe­ter near the entrance of the occlusion. Positioning the tip of the microcatheter about 5mm from the entry will improve backup for the guidewire. However, if the distance is too short, the microcatheter will bias the guidewire toward the larger curvature of a curved vessel and increase the risk of guidewire whipping. You may approximate the microcath­eter tip to the CTO entry if stronger backup is absolutely necessary, but you should usually position the catheter tip at 1 to 1.5cm from the entry to ensure good manipulability of the guidewire.
1.6.6 Device Delivery Strategy
After a guidewire has crossed a CTO through the true lumen, you can deliver devices along the guidewire. To reduce the risk of injury or perforation of the coronary artery segment distal to the occlusion during device delivery, you should exchange the CTO crossing guidewire for another guide­wire. Before performing guidewire exchange, you should rst advance a microcatheter along the guidewire.
Advancing this device is the only procedure during PCI for which an interventionalist may use a strong pushing force.
Because of good trackability, a microcatheter with a soft tip may be able to cross a tortuous lesion that is relatively hard, while a microcatheter with a stiff tip often fails to do so, since the latter type of catheter can be advanced into a hard CTO but cannot track a tortuous vessel.
The crossability of a microcatheter also depends on the diameter of its shaft. A microcatheter with a slimmer shaft can cross an occlusion more easily. One may think that a microcatheter with a stiffer shaft provides stronger backup. Considering the importance of reducing friction by rotating the microcatheter to facilitate its advancement, a stiffer shaft is better as it provides superior torque and greater resistance to twisting deformation. Nonetheless, a microcatheter may only be rotated to-and-fro several times.
Once a microcatheter has become wedged in an occlu­sion or tight stenosis and cannot be rotated further, continued attempts to rotate it in the same direction will cause defor­mation of the shaft. It is very difcult to slide the guidewire through such a deformed microcatheter. To withdraw the microcatheter, it is also necessary to withdraw the guidewire that has successfully crossed the lesion.
a
1.6.6.1 Microcatheter
For the reasons mentioned above, a microcatheter with good crossability should be selected. The crossability of a microcatheter is a function of the characteristics of the lesion and the properties of the microcatheter tip. Here, I will discuss a phenomenon that is particularly notable with balloon catheters, but is also applicable to microcatheters. In brief, it is difcult to cross a hard (e.g., calcied) occlu­sion using a microcatheter with a soft tip. This is because the tip of a microcatheter tends to are out when blocked by a hard obstruction (Fig.1.132a) and it will be deformed if pushed forcibly (Fig. 1.132b). These changes substantially increase the prole of the tip and make it more difcult for the microcatheter to cross the occlusion. In contrast, if the microcatheter has a stiff tip that is not easily deformed, it can be advanced while deforming the lesion instead.
b
c
Fig. 1.132 Microcatheter with a small soft tip. When a microcatheter
with a small soft tip is pushed forward into a severely calcied hard occlusion, the tip may are out (a) or be deformed (b). If the microcath­eter crosses such a hard occlusion, its tip may be stretched when the catheter is withdrawn from the lesion (c). The appropriate microcathe­ter for a severely calcied hard lesion should have a small stiff tip
1.6 Antegrade Approach
91
1.6.6.2 Balloon Catheter
If a microcatheter fails to cross the occlusion, you should try to deliver a semi-compliant balloon catheter with the small­est diameter feasible. As is the case for microcatheters, a bal­loon catheter with a stiff tip is more suitable for crossing a hard straight occlusion, while a balloon catheter with a soft tip is desirable for a tortuous lesion.
If a balloon catheter that has been advanced with suf­cient backup fails to cross the occlusion, you should consider the next strategy. Before proceeding to the next step, you should withdraw the balloon catheter and inspect its tip. If the tip is even slightly ared, cut off about 0.3mm using sharp scissors (Fig.1.133), since the balloon catheter may cross the occlusion without its ared tip. Instead of using a new balloon catheter, I modify the rst catheter in this way to save limited medical resources (see Chap. 2.3.4 “Delivery of a Balloon Catheter for the Kissing Balloon Technique” [page 186]).
Fig. 1.133 Modifying a
balloon catheter with a ared tip. During attempts to penetrate a severely calcied hard occlusion with a balloon catheter or to enter a side branch through a stent strut, the tip of the catheter may become ared (a), and even slight aring can interfere with delivery. After the ared tip is cut off with scissors (b), the balloon catheter may become able to cross the lesion or pass the stent strut
ab
1.6.6.3 Anchoring
As mentioned in Sect. 1.3 (“Guiding Catheter”), anchoring is very effective for stabilizing a guiding catheter, which can easily become disengaged when a balloon catheter is pushed in through it. I use an Amplatz left-type guiding catheter (e.g., AL 1 ST) when performing PCI for CTO of the RCA, and anchoring is sometimes necessary to stabilize this cath­eter because its curve does not always t the size of the sinus of Valsalva. In this setting, the side branch used for anchor­ing may be the conus branch, sinoatrial node branch, right ventricular branch, or atrial branch. The anchoring balloon should be slightly oversized and should be inated at a low pressure of about 4 to 6atm.
1.6.6.4 Tornus Crossing Catheter
If the occlusion is too hard to cross with a balloon catheter, you can often use a Tornus crossing catheter instead. If the guiding catheter is unstable even when using a Tornus cath-
92
b
1 Mitsudo’s PCI Techniques forCTO
eter, you should maintain the anchoring balloon at its previ­ous position to stabilize the guiding catheter. An anchoring balloon with a diameter of 2mm or more can be used as a trapping balloon during delivery and removal of the Tornus catheter.
After delivering the Tornus catheter, you should exchange the guidewire for a 0.014-inch extra support-type guide­wire and try to advance a balloon catheter over the wire. (Alternatively, you can exchange the guidewire for a oppy RotaWire in order to perform rotablation.) If the balloon catheter fails to cross the lesion, you should exchange it for a Tornus catheter again and exchange the 0.014-inch guide­wire for a oppy RotaWire to start rotablation (see Chap.
5.1.3 “Tornus Catheter” [page 245]).
1.6.6.5 Excimer Laser Coronary Angioplasty
(ELCA)
If a retrograde guidewire has crossed the occlusion, a loop can be formed between the antegrade and retrograde guiding catheters, which will almost denitely ensure successful deliv-
ery of the devices. If an antegrade guidewire has crossed the occlusion, a loop cannot be formed until guidewire exchange, so a balloon catheter or Tornus catheter may be unable to cross the lesion. In such cases, a laser catheter with a diameter of
0.9mm can be used to perform angioplasty (excimer laser cor­onary angioplasty: ELCA). However, ELCA is not effective for recanalizing a severely calcied lesion, and there is often no choice other than to perform rotablation for this purpose.
1.6.6.6 Rotablator
To perform rotablation before device delivery, you have to exchange the guidewire for a RotaWire through a microcath­eter with its tip xed at the CTO entry point. If the micro­catheter is advanced too far, its tip will become bent upon withdrawal of the guidewire from the catheter (Fig.1.134), and it will be impossible to orient the tip toward the entry point. Therefore, when exchanging guidewires before rotab­lation, it is important to only press the tip of the microcath­eter lightly against the CTO entry point.
a
Fig. 1.134 If a microcatheter is advanced into an occlusion, its tip
becomes bent on withdrawal of the guidewire. A microcatheter with a guidewire inside it is relatively straight (a). Upon withdrawal of the guidewire, the microcatheter (and its tip) becomes curved due to the release of accumulated force (b). When advanced through a microcath­eter with a slightly bent tip that has lost coaxiality with the occlusion, it will be difcult for a RotaWire to follow the correct route. A microcath-
c
eter with a slightly bent tip can be reused after withdrawing it and straightening the tip. However, a microcatheter with a severely deformed and folded tip (c) can hardly be returned to its original shape. Therefore, after its tip has been pushed into the occlusion, you should pull a microcatheter back slightly to release the accumulated force before withdrawing the guidewire from the catheter
1.7 Retrograde Approach
93
1.6.6.7 Double or Triple Wires
In the past, I would deliver a device to the occlusion through a lumen that was enlarged by using two or even three CTO crossing wires. However, I have not employed this strategy recently because the CTOs that I encounter have not required or permitted it.
1.6.7 From Balloon Ination toStent
Placement
After device delivery has been achieved, balloon ination may be performed. In general, you should dilate an occlu­sion by inating a small balloon (1.0 to 1.5mm in diameter) and then perform observation by IVUS.The ballooning and stenting procedures employed for CTOs are largely the same as those for non-occlusive lesions, except that special care is needed in case of crossing a false lumen and for preserving a small side branch when treating double CTOs of the main trunk and a larger branch. Please refer to IVUS-guided wir­ing in Sect. 1.7 “Retrograde Approach.”
1.7 Retrograde Approach
The “retrograde approach” is a term that has become widely used since 2005, when Dr. Osamu Kato proposed a break­through strategy that he referred to as “controlled antegrade and retrograde subintimal tracking (CART).” However, Dr. Kato and colleagues actually started to employ retrograde
guidewires around 2000 to track a collateral channel and reach the vessel distal to a CTO for use as a landmark when navigating an antegrade guidewire through the occlusion.
Here I summarize the strategies that I include in the cat-
egory of retrograde PCI for CTO.
1. Direct retrograde crossing: This involves retrogradely advancing a guidewire through a collateral channel to reach the lumen of the target coronary artery distal to a CTO and directly crossing the occlusion by using a retro­grade CTO guidewire with the support of a microcatheter for the proximal true lumen (Fig.1.135a).
2. Reverse CART technique: This involves navigation of a retrograde guidewire to a space within a CTO that was previously reached by an antegrade guidewire and enlarged by a balloon and using the retrograde wire to cross the occlusion (Fig.1.135b).
3. CART technique: This involves navigation of an ante­grade guidewire to a space in the distal part of an occlu­sion that was previously reached by a retrograde guidewire and enlarged by a balloon and using the antegrade wire to cross the occlusion (Fig.
4. Antegrade CTO crossing: This involves placing the tip of a retrograde guidewire in the distal part of a CTO and using it as a landmark while crossing the CTO ante­gradely (Fig.1.135d).
5. Kissing wire technique: This involves advancing an ante­grade or retrograde guidewire into a CTO and using it as a landmark to navigate another guidewire across the occlusion from the opposite direction (Fig.1.135e).
1.135c).
94
a
b
c
Fig. 1.135 Techniques for the
retrograde approach. (a) An antegrade guidewire (landmark) is not always required to perform direct retrograde crossing. As I attempt antegrade PCI for CTO before switching to the retrograde approach whenever possible, I always leave the antegrade guidewire in position as a landmark for direct retrograde crossing. (b) Reverse CART technique. (c) CART technique. (d) Antegrade CTO crossing: A retrograde guidewire is used as a landmark for the distal true lumen. The tip of the retrograde guidewire is positioned just beyond the distal end of the occlusion, (1) and an antegrade guidewire is advanced toward it (2). (e) Kissing wire technique using a retrograde guidewire as a landmark for the distal true lumen. A retrograde guidewire is advanced into the occlusion (1), and then an antegrade guidewire is navigated toward the retrograde wire (2)
1 Mitsudo’s PCI Techniques forCTO
proximal distal
d
Antegrade guidewire Retrograde guidewire
e
1.7.1 Indications fortheRetrograde Approach
the proximal end of the lesion to serve as a landmark. PCI is contraindicated for occlusions resulting from coronary
ostial atresia. To rule out this possibility, you must perform As mentioned above, I consider antegrade PCI to be the “pri­mary intervention for CTO.” However, I think that retrograde PCI can be indicated if there is a collateral channel that can be tracked and if any of the following requirements are met:
CT angiography if there is no dimple at the proximal end of
an ostial occlusion, since a diagnosis of ostial atresia can
only be established by CT angiography (Fig.1.138).
3. If the distal end of an occlusion is at a branching point and both branches require recanalization despite having small
1. If a large dissection (false lumen) has been created in a CTO after failure of antegrade PCI (Fig.1.136): You may be lucky enough to nd a route that leads to the true lumen at the end of the occlusion, but an antegrade CTO crossing guidewire will readily enter the large false lumen in most cases. Therefore, you should consider early switching to retrograde PCI.
Before switching to retrograde PCI, you should advance a relatively soft antegrade CTO guidewire. This wire may possibly be able to cross the occlusion through the true lumen, and it can serve as a landmark to facilitate retrograde wiring if crossing fails.
2. If the proximal cap of a CTO is located at a large coronary ostium and is inaccessible from the ostium (Fig. 1.137): Such a CTO cannot be recanalized by antegrade PCI.If pos­sible, an antegrade guiding catheter should be positioned at
luminal diameters (Fig. 1.139): Even if an antegrade guidewire crosses such an occlusion through the true lumen, it often penetrates plaque at the branching point and enters the true lumen distal to the carina. In this situ­ation, the side branch may not be isolated even after dila­tion by a small balloon. In contrast, a guidewire advanced retrogradely to the branching point always enters plaque proximal to the carina, so the side branch will not be occluded by inating a balloon in the occlusion.
4. If an antegrade guidewire fails to enter the distal true lumen (Fig.1.140).
5. If an antegrade guidewire undergoes deviation into the subintimal space after being blocked by a hard obstruc­tion (Fig. 1.141): When an antegrade guidewire with a very high tip load (e.g., a Conquest Pro 12) is blocked by a hard obstruction within a CTO and cannot be advanced
1.7 Retrograde Approach
95
further through the true lumen, you should advance another guidewire retrogradely. If the retrograde guide­wire can reach the site of the obstruction, you may then bypass the obstacle relatively easily with the antegrade guidewire. Eventually, this may lead to successful cross­ing of the CTO by the reverse CART technique (reverse CART) with a short false lumen passage.
6. If an antegrade guidewire undergoes extravascular devia­tion while being advanced through a long occlusion with
a
anatomy unclear: In this situation, you should promptly switch to retrograde PCI.If a retrograde guidewire fails to enter the true lumen, you may try to advance an antegrade guidewire again using the retrograde guidewire as a land­mark (kissing wire technique). This often leads to suc­cessful intimal/subintimal tracking and allows reverse CART.
b
c
Fig. 1.136 CTO of the distal RCA (Segment #3) (a). (b) An antegrade
guidewire has reached the distal true lumen, but has tracked through a false lumen on its way to the exit. After completion of PCI, the posterior descending branch was still occluded. (c) Imaging immediately before
reattempting PCI showed a large false lumen in Segment #3. A guide­wire could easily track the posterolateral branch, so the posterior descending branch was approached retrogradely, and the CTO was suc­cessfully recanalized without side branch occlusion
96
ab
1 Mitsudo’s PCI Techniques forCTO
a b
Fig. 1.137 In-stent occlusion in the proximal RCA.A stent previously implanted from the ostium of the RCA precluded sufcient backup by the
guiding catheter and diagnostic contrast imaging
Fig. 1.138 Coronary ostial atresia. (a) CAG. (b) CT angiogram. RCA angiography provides good visualization of the LCA.CT angiography
shows no ndings suggestive of LCA ostial atresia
ab
1.7 Retrograde Approach
Fig. 1.139 CTO of the distal RCA (Segment #3). Antegrade PCI was
attempted, but the CTO could not be crossed. Subsequently, retrograde PCI via the LCX was successful
97
Fig. 1.141 Antegrade guidewire blocked by a hard obstruction (a). (b)
After a retrograde guidewire has entered the occlusion in the true lumen, it may become easier to direct the antegrade guidewire toward the true lumen
Fig. 1.140 CTO of the LAD
(a) (Segment #7). After failure to cross the CTO antegradely (b), the retrograde approach via the RCA was successful (c)
abc
98
ab c
1 Mitsudo’s PCI Techniques forCTO
1.7.2 Collateral Channels
Whether the retrograde approach can be performed is depen­dent on the availability of accessible collateral channels. General knowledge of the collateral channels that usually exist may help you to discover obscure channels on CAG.In addition, understanding the anatomy of these channels facili­tates selection of the optimal uoroscopy angles for viewing the longitudinal axis of each channel, as well as assisting with assessment of the accessibility of each channel and devising the optimal tracking strategy.
Here, I present the collateral channels that I have success-
fully tracked when performing retrograde PCI, although this list may not include all types of channels that could poten­tially exist.
Collateral channels that can be used for the retrograde
approach to CTOs of the RCA, LAD, and LCX are presented in Figs.1.142, 1.143, and 1.144, respectively.
Diagnostic CAG images and control CAG images
obtained immediately before switching to retrograde PCI should be reviewed to identify every possible collateral channel and determine their accessibility.
Continuity of a channel may be obvious in a certain view, but not in another view. Performing tip injection of contrast medium with a microcatheter will eventually lead to a con­clusion regarding channel continuity. However, the anatomy of collateral channels generally follows certain rules. An epi­cardial channel does not cross another channel without com­munication, i.e., it does not jump over a nearby vessel to join another more distant vessel. As a result, collateral channels that connect the anterior and posterior sides of the heart run on the surface of the atria or through the atrial septum, or through the ventricular septum, or on the surface of the apex. Even if two channels running through the ventricular septum appear to cross each other, they will actually run on the left and right ventricular sides of the septum and not in the same plane (i.e., overhead crossing). You must not miss such chan­nels that appear to cross each other on CAG.A guidewire may be able to track a channel that is invisible on CAG and reach the lumen of the target vessel distal to the occlusion. This is particularly true for septal channels, and it should be remembered that a guidewire may be advanced beyond the visible end of a channel (Fig.1.145).
de f
Fig. 1.142 Collateral channels available for the retrograde approach to
CTO of the RCA. (a) RAO. (1) LCX→ACCAVB. (2) SB→AVB. (3) PL→ FWEp→RCAPL. (4) SB4PD. (5) LAD apex4PD. (b) RAO.RVBSB and RVBLAD. (c) RAO. (1) Dg→Apex→RVB. (2) Dg → Apex→4PD. (d) LAO. (1) LCX → ACC → AVB. (2) LCXFWEpAVB. ACC atrial circumex channel, AVB atrioven-
tricular branch, FWEp free wall epicardial channel. (e) LAO. Intracoronary collateral; Right atrial branch (SNA) → Kugel anastomosisAVB (AVNA). SNA sinoatrial node artery, AVB atrioven­tricular branch, AVNA atrioventricular node artery. (f) RAO.Intracoronary collateral; RVB  RVB.RVB: right ventricular branch