Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3585_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
102 Мб
Скачать
1.7 Retrograde Approach
109
a
b c
de
Fig. 1.156 CTO of the RCA. An epicardial channel runs from the
atrial circumex branch of the LCX to the RCA.At a bifurcation, over­lapping branches are difcult to separate in the RAO (a) and AP+CA
(b) projections. In the RAO+CA projection (c), good separation was obtained, and a guidewire could be advanced into the channel. The guidewire (d) and a microcatheter (e) successfully tracked the channel
110
Fig. 1.157 Multiple septal channels to the LAD. It was difcult to
identify or isolate the best channel because multiple channels crossed and overlapped each other
1 Mitsudo’s PCI Techniques forCTO
Fig. 1.159 A tortuous channel between septal branches arising in dif-
ferent directions, one from the posterior descending branch of the RCA and the other from the LAD
ab c
Fig. 1.158 Two apparently independent branches. (a & c) RAO. b.
AP+CR.Multiple channels are seen, but all of them overlap the main vessel for most of the route (a). You should isolate a channel in a differ-
ent projection that permits better separation from the main vessel (b) and then return to the original projection that gives the longest view of the channel (c)
1.7 Retrograde Approach
111
Fig. 1.160 A vessel that is
seen to not be a channel in a different projection. In the RAO+CA projection, the two vessels encircled by a dotted line appear to form channels between common donor and recipient vessels (a). In different projections, one of these vessels loses continuity (i.e., it is not a channel) (b & c)
abc
1.7.8 Guidewire Selection andHandling forEach Channel Type
For tracking a collateral channel, it is better for a guidewire to have a very low tip load, as soft a shaft as possible, and good slipperiness. Currently, a Sion guidewire is my rst choice.
Until the entrance of the channel is reached, a retrograde guidewire needs a tip curve of a sufcient size for introducing it into the channel (Fig.1.161). Once the retrograde guide­wire has reached the entrance of the channel, you advance it further with minimal force and attempt to track the chan­nel to reach the artery distal to the occlusion. However, the guidewire will often fail to track a small tortuous channel and reach the distal true lumen because the tip curve is too large or the wrong shape. Therefore, it is often necessary to advance a microcatheter into the channel for an appropriate distance and exchange the guidewire for another wire with a different tip curve.
It is very useful to perform tip injection through the microcatheter after withdrawal of the rst guidewire because it can conrm tortuosity of the channel and continuity with the distal true lumen without interference by overlapping vessels (Fig.1.162).
For retrograde tracking of a collateral channel, the tip of the guidewire should ideally be curved according to the tortuosity and branching of the channel (Fig.1.163), but a guidewire with a small tip curve is generally satisfactory (Fig.1.161c).
To prevent entry into a side branch at the vertex of a small acute curve when tracking a collateral channel, the guide­wire may need to have a tip curve that is smaller and more acute than the curve in the channel. A very small and tortu-
ous channel may be difcult to track with a guidewire that has a 0.014-inch tip. Such a channel can often be tracked by a tapered polymer-jacketed guidewire (e.g., an XT-R) if the tip is shaped into a very acute curve. However, a tapered guidewire has the potential to cause intimal injury, particu­larly during the phase of the cardiac cycle when the tip of the guidewire whips to and fro extensively. Accordingly, you should take great care when manipulating a tapered guide­wire to track a collateral channel.
To track a channel with multiple small corkscrew-type bends (Fig.1.164), it is recommended to make the tip ex­ible by forming a small curve using a guidewire inserter (Fig.1.165a & b) and to reshape the wire by using a 25-G shaping mandrel to obtain the optimal curve at the tip (Fig.1.165ac).
When tracking a collateral channel, I generally keep the following points in mind:
1. Use minimal force when advancing the guidewire into the
channel, and never push it hard.
2. Primarily view the channel in a projection perpendicular
to its longitudinal axis.
3. Always keep the tip of the guidewire parallel to the chan-
nel while advancing it, especially as it passes through an
acute bend.
4. Advance the guidewire with as little force as possible and
only when it is oriented in the correct direction.
5. Exchange the guidewire for another with more slipperi-
ness if the resistance is too great to advance it without
employing strong force.
Of these points, (3) and (4) should especially be followed if it becomes difcult to advance the tip of the guidewire
112
a
b
c
ab
1 Mitsudo’s PCI Techniques forCTO
in the correct direction. Obtaining images from different uoroscopy angles often provides new information. Part of the collateral channel that appears straight in one projection may be seen to be tortuous in another projection orthogonal to the rst (Fig.1.166), or it may be found that the guide­wire has entered a small side branch. Moreover, the actual branching point and direction of a collateral channel may be different from those expected initially (Fig.1.167). Under uoroscopic guidance in the optimal views for assessing channel tortuosity and branching, you should optimize the
Fig. 1.161 Optimal guidewire tip curves for channel
selection. (a & b) To advance a guidewire from the main vessel into a collateral channel, the wire should initially have a tip curve that is of an appropriate size for the diameter of the main vessel. (c) After reaching the entrance of the collateral channel, the guidewire should be reshaped to create a smaller tip curve in order to track the channel with a smaller diameter
tip curve of the guidewire by viewing the reference image and advance the wire in the desired direction with minimal force while performing rotation to help it pass through the channel.
In order to advance the guidewire when it has become dif­cult to move forward through the channel, you will need to determine the correct course for the wire by viewing images in other projections and intentionally orient its tip in that direction. You should always remember never to forcibly push a guidewire that has become difcult to advance.
Fig. 1.162 Tip injection. Injecting contrast medium from the guiding
catheter provides little information on the anatomy of the target channel because many other overlapping vessels are also visualized (a).
However, tip injection from a microcatheter clearly visualizes the target channel (b)
ab
ab
ab
1.7 Retrograde Approach
Fig. 1.163 Epicardial channel from
the posterolateral branch of the RCA.The tip of a microcatheter is indicated by the arrow. The guidewire needs to be reshaped to have appropriate tip curves for the diameter and tortuosity of the branch/channel to be tracked. Initially, the guidewire needs a relatively large curve at the tip to select the right branch (a). When advanced further, the guidewire needs a small tip acute curve (b). Imaging by tip injection provides clear information about these requirements
113
Fig. 1.166 Variation of channel appearance in different uoroscopy
projections. A channel that appears linear in one projection (a) may be tortuous in another projection (b)
Fig. 1.164 Corkscrew-type collateral channel. This channel could not
be tracked by a regular guidewire, but was tracked by a broken-tipped guidewire (shown in the next gure)
a
b
Fig. 1.167 Variation of channel appearance in different uoroscopy
projections. A channel that appears to have no branches in one projec­tion (a) may show a branch in another projection (b)
c
Fig. 1.165 Making a broken-tipped Sion guidewire. (a) A regular
curve is formed using the tip of a guidewire introducer. (b) The curve is altered to a J-shape. (c) The curve is made less acute again by using a 25-G shaping mandrel
114
Column 8 Sion Guidewire
1 Mitsudo’s PCI Techniques forCTO
As of March 2015, the Sion is the best guidewire for the abovementioned collateral channel tracking technique. A Sion guidewire can successfully track a tortuous collateral channel as the shaft of its several centimeters long tip has optimal exibility and slipperiness. To track a tortuous channel with tight bends, the guidewire needs to be exible up to the tip if possible. Unfortunately, the actual tip of every guidewire is not exible as it has a brazed part about 1-mm-long. If the tip of a guidewire is broken by bending it excessively (see Fig.1.165b), only 0.6–0.7mm of the tip remains inexible, so the guidewire becomes able to track even a corkscrew-type channel with tight bends. Simply bending the tip of the guidewire into a J-shape would rather increase the risk of causing vascular injury. Therefore, the bent tip is straightened again, and the angles are smoothed to form a natural curve at the tip (see Fig.1.165c), creating a safe and effective guidewire for collateral channel tracking. Figure1.168 shows an example of successfully tracking a tortuous channel with a broken-tipped Sion guidewire.
A new guidewire with a completely exible tip has been proposed. This wire would have no core at its tip, and consequently the spring coil would be much more exible, so it would have an overwhelming advantage over existing guidewires for tracking a non-branching tortuous channel. However, if the channel has a side branch at the vertex of one of its curves, would it be possible to control the guidewire and prevent entry into the side branch? In such circum­stances, a guidewire with a core at its tip usually seems to have better performance. As is known by every interven­tionalist and is expected from the reason a vessel becomes tortuous, a tortuous vessel (especially an epicardial artery) often has a side branch at the vertex of one of its curves. Recognizing such a side branch and navigating the guidewire through the channel across the branching point are the keys to successful collateral channel tracking.
Fig. 1.168 Successful tracking of a tortuous channel by a broken-tipped Sion guidewire
ab
cd
1.7 Retrograde Approach
115
1.7.9 Conrmation ofChannel Penetration
artery have similar anatomy. In this case, advancing a micro-
catheter along the guidewire without conrming that the wire When you think that the retrograde guidewire has reached the lumen of the target coronary artery distal to the CTO, you should perform retrograde imaging via other collateral channels to conrm that the tip of the wire is in the distal true lumen. A retrograde guidewire quite often enters a vein or a cardiac ventricle (Fig.1.169a & b), probably via a pre­existing stula. In particular, you may think a guidewire has entered the artery lumen distal to the CTO although it has actually entered a vein if the vein and the target coronary
has reached the distal true lumen will cause arteriovenous
disruption at the stula, resulting in pericardial hemorrhage.
A guidewire that has tracked a collateral channel may reach the lumen of a small vessel with very similar anatomical fea­tures to the distal true lumen. Although you may be almost certain that a retrograde guidewire has reached the distal true lumen without performing angiography, you should never fail to perform contralateral imaging to conrm this before advancing a microcatheter along the guidewire.
Fig. 1.169 Conrmation of retrograde guidewire crossing by angiogra-
phy. The guidewire seems to advance toward the septum in the RAO view (a), but it shows considerable deviation toward the LCX in the
LAO view (b). The guidewire appears to have reached the posterior descending artery via a septal branch (c), but tip injection demonstrates that it has actually entered a vein running parallel to the artery (d)
116
1 Mitsudo’s PCI Techniques forCTO
1.7.10 Advancing aMicrocatheter into theDistal True Lumen
After a retrograde guidewire has crossed a collateral channel, you should advance a microcatheter into the true lumen dis­tal to the CTO.A microcatheter can often be advanced eas­ily through a relatively large and less tortuous channel, but frequently will not track a small tortuous channel so well. In the latter situation, the microcatheter will often be easier to advance if it is rotated and pushed simultaneously. However, most microcatheters have a thin shaft and cannot tolerate too much rotation, while the microcatheters with a shaft thick enough to tolerate rotation often fail to penetrate a channel because of their large diameter.
To enable a thin microcatheter to track a collateral chan­nel, you should advance it while alternately performing clockwise and counterclockwise rotation a limited number of times. For example, you can perform three clockwise rota­tions and then three counterclockwise rotations to reverse the torque back to neutral, after which you can perform an additional three counterclockwise rotations (six counter­clockwise rotations in total). This may be followed by six clockwise rotations and then another six counterclockwise rotations and so on. Such rotation is performed to reduce friction, and not for achieving a large screw effect as with a Tornus microcatheter.
If this maneuver does not help the microcatheter to pass through the channel, you may introduce a small-diameter balloon catheter. For this purpose, a small-diameter bal­loon with a small tip prole and a soft tip is naturally better. Whether the balloon should be inated to dilate the chan­nel after it has been advanced depends on the risk of car­diac tamponade due to coronary artery perforation following
balloon ination. A ventricular septal branch can be safely dilated by a balloon, but the epicardial part connecting the septal branch with the true lumen distal to the CTO should not be dilated by the edge of the balloon, and the body of the balloon should be used (Fig.1.170). Dilating an epicardial channel with a balloon is associated with a risk of cardiac tamponade, and forcibly pushing a balloon catheter into the epicardial channel further increases this risk. Rather than attempting to force a large balloon catheter into an epicardial channel, it is safer to pass a balloon catheter with a smaller tip prole through the channel without pushing hard and then inate the balloon at a low pressure within the channel.
If the balloon catheter fails to cross the collateral channel, you should search for another trackable channel while keep­ing the guidewire that crossed the rst channel in position.
If there is no other trackable channel, you should position the tip of the retrograde guidewire near the distal end of the occlusion as a continuous landmark to assist antegrade wir­ing. Such a continuous landmark will considerably increase the probability of successful antegrade crossing of the CTO.
If a microcatheter crosses the collateral channel, you should advance the retrograde guidewire and then the micro­catheter to near the distal end of the CTO.If there is a side branch near the distal end, it is better to introduce the ret­rograde guidewire into the side branch. This is because if the retrograde guidewire remains near the distal end of the occlusion, it will often go forward together with the micro­catheter when the microcatheter is advanced. This may lead to knuckling of the guidewire tip so that it readily deviates into the subintimal space near the distal cap of the occlusion (Figs.1.171 & 1.172) and will preclude retrograde intralumi­nal crossing of the CTO.
Fig. 1.170 Balloon dilation
of a collateral channel that could not be crossed by a microcatheter. When dilating the part of the channel indicated by a dotted line, great care should be taken not to cause perforation as this may lead to pericardial bleeding
ab
1.7 Retrograde Approach
Fig. 1.171 Advancing a retrograde microcatheter. The retrograde
guidewire should be introduced into a side branch near the distal end of the occlusion (a). If a retrograde microcatheter is advanced without rst doing so, the guidewire will deviate into the subintimal space near the distal end of the occlusion (b)
117
wires may also be used, depending on the hardness of the lesion. If it becomes trapped in plaque or the vessel wall, a guidewire from the Gaia series is more likely to be deformed or even fractured than a 0.014-inch guidewire or a Conquest Pro guidewire. Therefore, you should take special care to prevent trapping while retrogradely manipulating such a guidewire. If it becomes impossible to move a retrograde guidewire or advance it further, you should immediately check if it can be pulled back and should then withdraw the wire slightly before advancing it again. It is important to pre­vent the guidewire from being trapped.
If a retrograde guidewire becomes trapped in the tunica media, you should follow the same procedure that was sug­gested for a trapped antegrade guidewire, i.e., rst advance the tip of the microcatheter toward the point of trapping while pulling the guidewire back slightly, and then rotate the wire alternately clockwise and counterclockwise an identi­cal number of times while continuing to pull it back with minimal force. This maneuver should be performed with patience. If the guidewire can be pulled back into the micro­catheter, you should then direct it toward the proximal end of the occlusion from a point just distal to the site of trapping. It is important to be aware of the potential for deformation of the guidewire tip by trapping and withdrawal (rotation). You should withdraw the guidewire from the microcatheter to determine whether its tip has been deformed. If the tip becomes deformed, you should stop using the guidewire and exchange it for a new one (see Sect. 1.9. “Troubleshooting” [page 144]).
Fig. 1.172 CTO of the distal RCA (Segment #3). A retrograde guide-
wire that tracked a collateral channel to reach the distal end of the occlusion was introduced into a side branch before advancing a retro­grade microcatheter
1.7.11 Selection of (I) Guidewire, (II) CTO Penetration Strategy, and(III) Guidewire Handling Technique
The proximal cap of a CTO is generally harder than the distal cap. As I always start PCI for CTO via the antegrade approach and assess the hardness of the proximal cap before switch­ing to retrograde PCI, I make my rst attempt at retrograde crossing with a guidewire that is less stiff than that employed for the antegrade approach. I commonly use a 0.014-inch guidewire with a tip load of about 3.0g for the retrograde approach and prefer a guidewire that is easier to control with less probability of whipping. Tapered CTO crossing guide-
1.7.12 Direct Crossing Technique
Since I attempt antegrade PCI for CTO rst whenever pos­sible, I rarely perform direct crossing of an occlusion via the retrograde approach. Using the antegrade guidewire as a landmark theoretically makes direct retrograde crossing a subcategory of the kissing wire technique. In fact, manipula­tion of the guidewire for direct crossing is done in the same way as in the kissing wire technique (see the next subsection).
When antegrade PCI is performed for a bifurcation CTO, the antegrade guidewire may be unable to catch a dimple at the proximal end of the occlusion. In this case, a retrograde guidewire may be able to successfully cross the lesion (direct crossing) if it is manipulated toward the tip of the antegrade wire near the branching point. On the other hand, the ante­grade guidewire often starts to cross the occlusion if it is manipulated when the tip of the retrograde guidewire is near the proximal end of the CTO, even if it is in the subintimal space. Thus, PCI is often accomplished successfully by the kissing wire technique or reverse CART (Fig.1.173).
Some CTOs can only be recanalized by the direct cross­ing technique. A good example is a CTO at a large coronary ostium where an antegrade guiding catheter cannot catch a
118
1 Mitsudo’s PCI Techniques forCTO
dimple at the proximal end of the occlusion. Only retrograde PCI should be attempted for such a CTO, and if a retrograde guidewire crosses the occlusion, it should be introduced into an antegrade guiding catheter by using a snare (Fig.1.174).
a b
The optimum snaring point may be at the ostium of the target coronary artery or in the innominate artery (Fig.1.175). To perform snaring, you should use a snare ranging in size from 10 to 15mm.
Fig. 1.173 Successful recanalization of a CTO by the kissing wire
technique. After failure of antegrade PCI, the strategy was switched to the retrograde approach (a). After a retrograde guidewire had almost
Fig. 1.174 CTO at the RCA
ostium. A retrograde guidewire was captured with a snare and introduced into an antegrade guiding catheter
reached the proximal end of the occlusion, an antegrade guidewire was manipulated again, and the lesion was crossed successfully (b)