Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3585_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
102 Мб
Скачать
262
5 Mitsudo’s Non-pushing PCI Techniques
Fig. 5.30 LAD bifurcation lesion. Because the diameter of the proxi-
mal MB was much larger than that of the distal MB, POT was per­formed after stenting. When a balloon was advanced into the SB, it
caused shortening of the proximal part of the stent (arrows), making it difcult to deliver a stent into the SB
Fig. 5.31 Stenting a calcied LAD bifurcation lesion. After stenting, a balloon was advanced into the SB and caused shortening of the stent
(arrows) due to malapposition
5.6 Stenting
263
Fig. 5.32 CTO of the LAD successfully recanalized by reverse
CART.After a guidewire had crossed the CTO through a false lumen, a stent was implanted in the false lumen. After stenting, a balloon was advanced into the SB and caused shortening of the stent (arrows) at the
gap between lesion and stent. Planned SB stenting was abandoned because shortening of the stent (i.e., strut deformation) prevented deliv­ery of a second stent to the SB
264
Fig. 5.33 Stent deformation by a balloon advanced into the SB or by
an IVUS catheter during withdrawal after stenting. A stent may be deformed if a balloon is advanced forcibly or an IVUS catheter is with­drawn forcibly
5 Mitsudo’s Non-pushing PCI Techniques
calcied, debulking of the lesion should often be per-
formed by scoring/cutting balloon angioplasty or
rotablation.
For a bifurcation lesion, such as a distal LM bifurcation lesion:
4. Perform POT immediately after implanting a stent.
5. If difculty is encountered when delivering a previously
used balloon into a stent, withdraw the balloon into the
guiding catheter and then rewrap it there by ination at
about 10atm.
6. and 7. To deliver a balloon into the SB for post-stenting
KBI, special techniques may be required.
1. Without using a guidewire that provides relatively strong
support, it is impossible to alleviate bias of the stent/bal-
loon toward the larger curvature of a bend in the target
vessel (Fig.
5.35). For example, the Sion guidewire is not
suitable for use when delivering a stent/balloon because it provides insufcient support. The Sion Blue Extra­Support wire, or another guidewire that provides equiva­lent support, is preferable.
2. The stent should be well apposed when it is initially placed. If there is malapposition of any strut, the stent can easily be deformed if the strut is hit by another device. At a large bend in the target vessel, a strut is likely to be hit by another device, even if the struts are all well apposed. In a tapered vessel, a stent with a diameter as close as pos­sible to that of the proximal segment should employed. In a tapered vessel with a bend, a stent with a diameter that ensures optimal apposition in the proximal segment of the vessel should be used.
Fig. 5.34 Stent deformation by other devices at a curved bifurcation
with a large-diameter proximal MB.A stent placed at such a bifurcation lesion is likely to show malapposition at its proximal part. When a bal­loon is introduced for POT, it may hit a strut and deform the stent
5.6.2.3 Practical Approach toPlacing aConformable Stent
The following strategies can be employed when placing a conformable stent:
1. In principle, use a guidewire that provides relatively strong
support.
2. In a tapered vessel, place a stent with a size matching the
diameter of the proximal segment of the vessel by inat­ing the SDB at a low pressure.
3. Prepare the lesion sufciently to allow optimal stent
expansion, even by low-pressure ination of the SDB.Particularly if the distal segment of the vessel is
In a tapered vessel, there are two methods of placing a stent with a diameter matching that of the proximal seg­ment. One method is to perform double stenting, while the other is carry out low-pressure implantation of a stent with a diameter matching the proximal segment, followed by distal and proximal post-dilatation. I generally employ the latter strategy. If the proximal vessel has a diameter of 3.5 mm, I choose to implant a 3.5-mm stent even if the distal vessel diameter is only 2.5mm. If there is no plaque at the distal edge of the stent, there is little risk that low-pressure expan­sion of the stent (about 6atm) will cause dissection.
If there is plaque at the distal edge of the stent and dis­section unfortunately occurs at the stent edge, this can be overcome by implanting a short additional stent (about 8mm long) with a diameter matching that of the distal vessel.
3. Lesion preparation: To achieve low-pressure stent expan-
sion, it is necessary to sufciently prepare (pre-dilate) the
5.6 Stenting
265
lesion. Wherever stent expansion is suboptimal, a post­dilatation balloon or another device may become caught on a strut, possibly leading to deformation of the stent.
The diameter of the distal vessel (particularly the diameter of the distal MB at a bifurcation) may be much smaller than the diameter of the stent implanted in the proximal segment. An oversized stent will exert little expansionary force after implantation. Many stents exert a maximum force of only 2 to 3kPa when expanded to their target diameter. If a stent is going to be expanded to a smaller diameter than the target, sufcient lesion prepa­ration must theoretically be performed to allow sustained stent expansion with low-pressure balloon ination (1.5–2.0atm).
4. POT: In a bifurcation lesion, POT should only be per­formed to dilate the proximal MB to the nal target diam­eter and thus improve apposition of the proximal struts. In a tapered non-bifurcation lesion, POT should also be per­formed before post-dilatation of the distal part of the stent if there is malapposition of the proximal struts.
5. Rewrapping: If a balloon is advanced into a stent for post­dilatation or KBI after being used for pre-dilatation and not rewrapped, considerable friction may be generated within the stent, and it can be impossible to move the bal­loon forward. If this happens, the balloon should be with­drawn into the guiding catheter and inated at a relatively low pressure of about 10atm, followed by rewrapping in a ne covering tube. After a balloon is rewrapped in this way, it can usually be delivered into a stent very easily.
In the era of bulky balloons, the interventionalist had to withdraw a balloon completely and rewrap it outside the patient. To rewrap the balloon in those days, it was usually returned to its protective sheath and inated at about 10atm. The abovementioned rewrapping technique copies this old method of balloon rewrapping, but per­forms it inside the guiding catheter. Inating and deat­ing a balloon in a smooth lumen with a relatively small diameter increases longitudinal folding of the balloon, which will then reduce friction during delivery of the bal­loon into a stent.
6. KBI or high-pressure MB dilatation: After wiring the SB through the MB stent and advancing a kissing balloon into the MB, you may sometimes encounter difculty when advancing a kissing balloon into the SB across the MB stent. If the stent struts have been crossed by the tip of the balloon, but not by its shoulders, you may immedi­ately perform KBI at a low pressure of about 6atm. This will often facilitate advancement of the balloon into the SB. Even if performing KBI once fails to completely deliver the balloon into the SB, several cycles of balloon ination/deation can lead to gradual advancement of the
SB balloon and eventually allow it to smoothly cross the stent into the SB.To avoid over-dilation of the proximal segment of the vessel, you must not allow the marker of the MB balloon to protrude beyond the proximal edge of the MB stent, and simultaneous ination of the two kiss­ing balloons should be conned within the stent.
High-pressure ination of the MB balloon alone can also be useful and is possible after withdrawal or suf­cient advancement of the SB balloon.
As is also the case in (7) below, you should never use the anchoring technique with the MB balloon delivered into the distal MB or any other balloon. This is because forcibly pushing a device that has been caught on a stent strut is very likely to cause deformation of the stent.
7. Checking and cutting off a damaged balloon tip, using a new balloon, or using the buddy wire/balloon techniques: If a balloon cannot be advanced toward the SB and if the site of obstruction appears to be distant from the balloon’s distal marker, the tip of the balloon may be blocked by a stent strut.
Possible reasons for this are (1) the balloon tip has become
frayed or (2) the angle between the tip of the balloon and the stent strut/vessel wall is too small or large.
As mentioned in (6) above, high-pressure ination of the
MB balloon alone, which is possible after withdrawal or suf­cient advancement of the SB balloon, can be useful and is worth trying rst. It may also be effective to withdraw the SB balloon into the guiding catheter or another site where it can be rotated freely and rotate it 180 degrees before readvance­ment. However, it is impossible to deliver a balloon into the SB across the struts of the MB stent if the tip has become frayed. Therefore, you should cut off the frayed tip of the balloon or exchange it for a new one.
If a balloon with an intact tip is difcult to deliver, you
may advance a very small balloon past the obstruction and pre-dilate the stent cell.
Since the highly conformable PROMUS Element stent
is prone to proximal deformation, the design was modied when developing the PROMUS Premier, which has four or ve links in the three struts at its proximal edge. This modication has made the proximal 3mm of the stent very rigid and thus increased the risk of fracture at the inter­face between the rigid and more exible parts of the stent. Hence, the PROMUS series cannot escape one of the great­est defects of the rst-generation DES.It is important for interventionalists to consider ways of making the best use of currently available conformable stents to achieve implan­tation without deformation and not to employ stents with a new design that may apparently make stenting easier, but is actually inferior.
266
Fig. 5.35 Using a guidewire that provides strong backup. A guidewire
or balloon catheter tends to track along the larger curvature of a bend in a vessel. Upfront use of a guidewire with strong backup may allow advancement of a balloon catheter along a different route. It can also be effective to rst advance a balloon with good deliverability on an out­wardly biased guidewire and then advance an optimally sized balloon along a guidewire with strong backup
5.6.3 Using aChild Catheter
After sufcient preparation of the lesion, you should repeat­edly advance and pull back the balloon catheter. If you notice any difculty in advancing or withdrawing the balloon cath­eter or a Crusade or IVUS catheter, you should consider using supplementary techniques for stent delivery.
The supplementary techniques for stent delivery include (1) the buddy wire technique (Fig.5.36a), (2) the buddy bal­loon technique (Fig. 5.36b), (3) the slip-through technique (Fig.5.36c, d), (4) the balloon trapping technique (Fig.5.36e), and (5) the child-in-mother guide technique (Fig.5.36f, g).
In recent years, use of a child catheter (the GuideLiner or equivalents such as Guidezilla and GuidePlus) has become
5 Mitsudo’s Non-pushing PCI Techniques
widespread because a child catheter can be manipulated sim­ply and reliably. However, the GuideLiner has a larger pro­le than a balloon or stent, and its tip becomes biased toward the greater curvature of a bend as it is advanced through a vessel, i.e., the same side where resistance to devices occurs. In other words, pushing a GuideLiner too vigorously may cause injury to the vessel (Fig.
5.37a). Advancing a bal-
loon catheter rst (Fig. 5.37b) or inating a small-diame­ter (2 mm) balloon so that half of it protrudes from the Guideliner (Fig. 5.37c) may facilitate advancement of the catheter. However, it is safer to use a balloon inated inside a pre-dilated distal segment or in an implanted stent as a coaxial anchor. Then the GuideLiner catheter can be advance reliably while slightly pulling back the anchoring balloon catheter (Fig.5.37d).
When advancing a Guideliner across a proximally implanted stent into a distal segment, it is necessary to take great care to prevent stent deformation by the child catheter. A stent can readily be deformed if the tip of the GuideLiner even slightly comes into contact with a strut. Use of a coax­ial anchoring balloon enables safer advancement of the GuideLiner.
When attempting to deliver a stent through a previously implanted stent, it may be impossible to advance the new stent from a GuideLiner catheter. If this happens, you should advance the GuideLiner and just keep it at the point of obstruction. With the support of this child catheter, you may then be able to advance the stent by applying gentle force. This method probably succeeds because the Guideliner lls the gap between the two stents (Fig.
5.38a) or because repo-
sitioning the guiding catheter corrects the bias of the catheter (Fig.5.38b).
The sleeve of the GuideLiner can block advancement of a balloon/stent or may cause deformation of a stent. Usually, the tip of a balloon or stent comes into contact with the rim of the GuideLiner sleeve (Fig. 5.39a, b). To prevent this, the GuideLiner should be pushed in a little before advancing a device into it to ensure that the GuideLiner passes the tip of the guiding catheter and that the stent is advanced along the medial side of the sleeve (Fig.5.39c).
The smallest catheter in the GuideLiner series is 5.5 Fr. If an even smaller child catheter is needed, a 4-Fr Heartrail Kiwami OTW catheter can be used.
a
b
c
d
e
f
g
a
b
c
d
a
b
a
5.7 IVUS
267
Fig. 5.38 Failure to advance a stent through a previously implanted
stent. After advancing a child catheter to the point of obstruction, it can be easier to advance the other stent through the previously implanted stent, probably because the wall of the child catheter lls the gap between the two stents (a) or because repositioning the guiding catheter corrects the direction of the stent (b)
b
c
Fig. 5.39 Obstruction of a balloon/stent (a) or stent deformation (b)
Fig. 5.36 Supplementary techniques for stent delivery. (a) Buddy wire
techniques. (b) Buddy balloon technique. (c, d) Slip-through technique. (e) Balloon trapping technique. (f, g) Child-in-mother guide technique
overcome by using a GuideLiner catheter. As it is advanced, the Guideliner catheter deviates outward in the guiding catheter and orients the device inward, thereby overcoming the obstruction (c)
5.7 IVUS
It is well-known that an IVUS catheter must not be forcibly pushed forward if it becomes difcult to advance. An IVUS
Fig. 5.37 Advancing a child catheter. (a) If there is a lesion on the
larger curvature of a bend in the target vessel, attempts to forcibly advance a child catheter may injure the vessel, necessitating additional intervention. (b) Advancing a balloon rst can reduce luminal irregular­ity and ensure smooth advancement of the child catheter. (c) Advancing and inating a small-diameter (2.0 mm) balloon rst can facilitate advancement of the child catheter. (d) Inating a balloon in the stent landing zone and then pulling the balloon catheter back slightly will change the orientation of the child catheter toward the smaller curvature of the bend, helping the catheter to advance beyond the lesion
catheter typically has a short OTW region and the shaft is not stiff. Accordingly, the shaft can easily be bent by pushing the catheter, leading to disconnection of the inner wiring.
If it becomes impossible to advance an IVUS catheter, you should suspect that its tip has been trapped by an irregu­larity on the larger curvature of the vessel wall. Therefore, you should pull the catheter back slightly and then immedi­ately push it forward a little. Like a blocked stent or balloon catheter, this maneuver may allow the IVUS catheter to slip past the obstruction and advance again. If it becomes almost impossible to advance an IVUS catheter further after progres­sive escalation of difculty in moving forward, it should be pulled back enough to eliminate deection and then slowly advanced again. You should check whether the tip of the catheter can move forward even slightly, as is the case with a stent. If the tip moves forward a little, you should continue to push it with the same force until the shaft becomes deected slightly in the proximal coronary artery. Then the maneuvers of pulling back and pushing forward are repeated.
If the hard transducer of the catheter becomes caught at a bend in the vessel, you should slide the core shaft back slightly
268
cd
a
b
c
5 Mitsudo’s Non-pushing PCI Techniques
to increase the exibility of the catheter tip and then follow the above procedure (Fig.5.40a, b). After the catheter tip has passed the bend, you should slide the core shaft (i.e., transducer) for­ward again to restore the backup of the catheter (Fig.5.40c).
If all of these attempts fail, it is necessary to enlarge the lumen by (1) balloon angioplasty (upfront or repeated) or (2) by debulking the lesion with a Rotablator. After fail­ing to deliver the IVUS catheter although the lesion is suf­ciently dilated or stented, supplementary techniques can be employed, such as (1) the buddy wire technique, (2) the buddy balloon technique, or (3) the trapping technique. I prefer to use a Guideliner to increase the backup of the IVUS catheter.
5.8 Anchoring Technique
The term “anchoring technique” covers several different concepts, which are reviewed here in relation to non-pushing PCI. “Anchoring” means taking measures to stabilize a guid­ing catheter or another device or to ensure smooth advance­ment by inating and xing a balloon in the artery lumen. In contrast, xing a guidewire by inating a balloon and then advancing a device along the xed wire is referred to as “trapping.”
Anchors can be non-coaxial or coaxial (Fig.5.41). When a non-coaxial anchor is used, the device can only be advanced by pushing it forward (Fig.5.41a). With a coaxial anchor, it is not always necessary to push the device forward, and it may be advanced almost spontaneously if the obstruction can be overcome by advancing a child catheter while pulling the anchor balloon back slightly (Fig.5.41b).
I refer to inating a balloon in a coronary artery or inside a guiding catheter for the purpose of xing a parallel guide­wire as “trapping.” Trapping may be used to push forward a device or to avoid an obstruction and allow spontaneous advancement of the device (Fig.5.41c, d).
Fig. 5.40 Failure to deliver an IVUS catheter. (a, b) It can be difcult
to deliver an IVUS catheter into a hard and tortuous lesion. If this prob­lem is encountered, slide back the core shaft and then slowly advance the entire catheter. This will increase the exibility and trackability of the catheter tip and may facilitate successful delivery of the catheter. (c) After the catheter tip has passed a bend in the vessel, slide the core shaft forward and then slowly advance the entire catheter
ab
Fig. 5.41 Anchoring technique. (a) Ination of a balloon in the SB to
facilitate advancing a device. (b) Ination of a balloon in the distal ves­sel to facilitate advancement of a child catheter. (c) Trapping with a
balloon to avoid an obstruction and allow spontaneous advancement of a device. (d) Trapping a guidewire to facilitate advancement of a microcatheter
ab
5.8 Anchoring Technique
269
5.8.1 Anchoring technique
The original anchoring technique (Fig.5.41a) employs ina­tion of a balloon in the SB to obtain strong backup for a guiding catheter when advancing a device into the MB.This technique is not required if the tip of the guiding catheter remains coaxial to the coronary artery ostium and if the shaft of the catheter can be brought into close contact with the contralateral wall of the aorta or aortic sinus or if the catheter can achieve the optimal power position.
Due to the vascular anatomy, it is sometimes impossible for any guiding catheter to achieve the optimal power position (Fig.5.42). Under such circumstances, it is often very dif­cult to engage a guiding catheter in the target coronary artery ostium, and the catheter may become disengaged and slide out of the ostium during PCI.If an SB is available, you should advance a guidewire into the SB for anchoring along with
engaging a guiding catheter at the coronary artery ostium. The anchoring technique is very useful to obtain strong backup for a device, as well as for stabilizing a guiding catheter.
The anchoring balloon should have a diameter one size (0.5 mm) larger than the lumen and a length of 15 mm, and it should be inated at a low pressure of about 4atm. Conversely, an undersized balloon inated at high pressure often fails to be an effective anchor.
When using the anchoring technique to increase backup for a device, it is important to be focused on pushing the device carefully to advance it. As stated in Sect. 5.1 of this chapter (page 243), the anchoring technique is useful for advancing a balloon into a severe native stenosis. However, a device should never be advanced with the aid of an anchor when pushing is not allowed. For example, the anchoring technique is typically contraindicated when delivering a bal­loon into the SB after stenting of the MB.
1
2
Fig. 5.42 Optimal power position. A guiding catheter achieves the
optimal power position when the tip is coaxial with the coronary artery ostium and the shaft is brought into close contact with the contralateral wall of the aorta (a). If a coronary artery shows extreme angulation at the ostium (b), the guiding catheter cannot achieve the optimal power
position. This is because keeping the tip of the catheter coaxial to the ostium decreases backup from the aortic wall (1), while ensuring suf­cient backup from the aortic wall results in loss of coaxiality between the catheter tip and the ostium (2)
270
5 Mitsudo’s Non-pushing PCI Techniques
5.8.2 Coaxial Anchoring
Coaxial anchoring with an OTW or monorail balloon cath­eter is often used to advance a child catheter (Fig.5.41b). If the guiding catheter has a small diameter, coaxial anchoring can be a very useful technique to assist stenting after deep engagement of the catheter. However, you should not use coaxial anchoring to push in a guiding catheter and instead should pull the shaft of the anchoring balloon back slightly to avoid the obstruction to catheter advancement. A balloon that has been used to pre-dilate the distal lumen is generally employed for coaxial anchoring. The distal end of the bal­loon should not be located beyond the distal edge of the stent. Sometimes, slightly pushing the anchoring balloon together with the child catheter for the last few millimeters cannot be avoided. If the child catheter still cannot be advanced, you should slightly push the tip of the child catheter into the site of obstruction and then gently insert a stent. The stent can often be delivered smoothly in this manner (Fig.5.43).
5.8.3 Trapping
Trapping is a method of advancing a device by the same mechanism as coaxial anchoring. For trapping, a parallel guidewire is xed and used to facilitate advancement of a device such as a stent (not a child catheter), which differenti­ates trapping from coaxial anchoring.
Usually, a trapping balloon has passed through the lumen and thus reduced the space available for advancing the device. If a retrogradely advanced balloon is used to trap a guidewire, the available space will become larger, permitting smoother advancement of the device.
A retrograde guidewire that has crossed a CTO is difcult to trap stably in the lumen proximal to the lesion. To advance a retrograde microcatheter, the retrograde guidewire often needs to be trapped very tightly. If it is not trapped, the ret­rograde guidewire may move back through the CTO into the distal coronary artery segment, necessitating another attempt to retrogradely cross the lesion.
After a retrograde guidewire has successfully crossed a CTO to reach the lumen proximal to the lesion, it should be introduced into the antegrade guiding catheter and trapped there, allowing a retrograde microcatheter to be reli­ably advanced across the CTO into the antegrade guiding catheter.
Fig. 5.43 Failure to advance a child catheter with coaxial anchoring.
The child catheter should be advanced to the site of obstruction, after which it will ll the gap and facilitate delivery of a stent
5.9 Guidewire Loop
A guidewire loop can be created by trapping, snaring, or externalization (Fig.5.44). When snaring is used, the guide­wire may escape from the snare, leading to the loss of the loop. It is safer to combine trapping with snaring (Fig.5.44d). In most cases, a guidewire loop is created after a retrograde guidewire has crossed a CTO (see Chap. 1).
ab cd
5.10 Removing anIVUS Catheter
271
Fig. 5.44 Making a guidewire loop. (a) Trapping. (b) Snaring. (c) Externalization. (d) When snaring is employed, the guidewire may escape from
the snare, so it is safer to combine trapping with snaring
5.10 Removing anIVUS Catheter
pulling harder on the catheter can also lead to it becoming tightly stuck. What happens when an IVUS catheter is caught
You should not forcibly push a wire/catheter/device to advance it. Likewise, you should not use excessive force to withdraw an IVUS catheter or another device if difculty is encountered when withdrawing it.
during withdrawal? Is it reasonable to pull the catheter if there only seems to be a little difculty in withdrawing it?
One way that an IVUS catheter can be caught during withdrawal is illustrated in Fig. 5.45. Advancement of a device or balloon may be blocked when the gap between the tip of the device and the guidewire or between the balloon
5.10.1 If theIVUS Catheter Becomes Stuck during Withdrawal
tip and balloon membrane results in the tip being caught by irregularities on the larger curvature of a bend in the ves-
sel (Fig.5.45a). Conversely, a device may become stuck on After stent implantation, it is usual practice to employ IVUS to conrm stent apposition. After completing observation, the IVUS catheter can usually be withdrawn easily by slowly pulling it back after rst sliding its imaging core forward to the limit. If the guidewire used with the IVUS catheter has a low tip load, it is better to push slightly on the guidewire while slowly withdrawing the IVUS catheter.
During withdrawal, the IVUS catheter may sometimes become stuck on the stent. In some cases, the catheter can still be withdrawn by exerting a slightly stronger force, but
irregularities on the smaller curvature of a bend during with­drawal (Fig.5.45b).
During withdrawal of an IVUS catheter, it may be caught at the exit port (Fig.5.46a). To stop the exit port from col­liding with the vessel wall during withdrawal of the IVUS catheter, you should push the guidewire while pulling the catheter back, in order to direct the catheter toward the larger curvature at a bend in the vessel. It is important to conrm that the IVUS catheter has not been caught by irregularities of the vessel wall (Fig.5.46b).