Figure 3.17 (a): Longitudinal section.
Figure 3.17 (b): Transverse section.
A wide range of benign or malignant masses can produce dirty shadowing.
Figure 3.18 (a): Spiculated malignant breast tumor.
Figure 3.18 (b): Infiltrative testicular seminoma.
Figure 3.18 (c): Ovarian dermoid containing hair.
Gas is a common cause of dirty shadowing.
Figure 3.19: In this image, bowel gas casts a dirty shadow (ds) over the lower pole of the right kidney. In order to evaluate this area and
exclude an exophytic renal mass, the operator must use a different
approach.
Figure 3.20: This large attenuating ovarian dermoid could easily be
mistaken for a loop of bowel. The diagnostic clues which allowed
differentiation of this ovarian neoplasm from bowel included: 1) the
large diameter of the mass and 2) the lack of peristalsis.

3.5 Dropout

Causative mechanisms
Dropout, like all forms of acoustic shadowing, is caused by an attenuating structure in the field of view. Ultrasound penetrates the causative structure and produces some echoes but is attenuated in the process. Dropout may also occur with the use of small apertures (focus close to the transducer) because the ultrasound beam will have wide divergence in the far field resulting in rapid loss of intensity.
Which system assumptions have been breached?
Assumption that attenuation rate is constant.
Typical appearance
Hypoechoic or anechoic region beyond the causative structure. Always along the lines of sight.
Can this artifact be reduced, eliminated or circumvented?
Some forms of drop-out can be eliminated with the use of a more favorable acoustic window. Other strategies for reducing dropout include: 1) increasing gain or TGC settings, 2) reducing the operating frequency of the transducer or switching to a lower frequency transducer to improve penetration and 3) switching from harmonic to fundamental frequency imaging.
Is this artifact diagnostically useful?
Most of the time, this artifact is a nuisance and is not diagnostically helpful.
Figure 3.21: Diminished penetration is seen in this patient with
advanced steatosis.
Figure 3.22: Figures 3.22 and 3.23 demonstrate the same normal
liver scanned through the subxiphoid epigastric window. In this
image, penetration is compromised due to a poor acoustic window
(the xiphoid is partially obstructing the acoustic window).
Figure 3.23: Adjustment of the acoustic window has yielded an
improved diagnostic image.
Figure 3.24: Dropout artifact is seen in the liver of a patient with
chronic liver disease when the examination is performed with a 6C2
transducer at 4MHz. As shown in Figure 3.25, switching to 2.5MHz
eliminates dropout but the trade-off is reduction in spatial resolution.
Figure 3.25: Using a lower frequency has eliminated the dropout but
the spatial resolution is reduced.
Dropout can also be due to a loss of transducer contact with the patient or absence of acoustic couplant (gel). This cause is usually quite obvious as there is complete lack of any diagnostic information in the image from the transducer surface down.
Figure 3.26: When spatial compounding mode is on, some echo
information may be gathered in the zone of drop-out from adjacent
apertures which are in good contact with the patient. The double
arrow in this image highlights the zone of no transducer contact.
In the images below the dropout artifact is seen below the level of the portal vein in a normal patient. The region of the portal vein is attenuating due to the presence of fibrous tissue which binds together the portal triad.