Is this artifact diagnostically useful?
The presence of shadowing is a useful diagnostic criterion for biliary or renal calculi and calcifications. For example, the definition of a gallstone within the gallbladder is that of an echogenic, mobile, shadowing mass. In some circumstances, shadowing may be more apparent than the causative structure itself. This improves the operator’s ability to find a subtle mass, for example a small breast malignancy or a foreign body. It is important to keep in mind that small structures such as small gallstones or kidney stones may not demonstrate shadowing if the size of the object is smaller than the beamwidth. To demonstrate or accentuate shadows from small structures, sonographers should use the highest frequency transducer possible and turn off spatial compounding.
Figure 3.9 (a): In this example, a small renal calculus casts an
acoustic shadow when interrogated in conventional imaging mode
without spatial compounding.
Figure 3.9 (b): With spatial compounding turned on, the shadow is
much less apparent.
Unfortunately acoustic shadowing may sometimes mask regions of pathology.
Figure 3.10 (a): A transverse view of the fetal chest is shown. The
sonographer noted an abnormal cardiac axis. The cause is not
evident in this image because an important diagnostic clue is hidden
in the shadow from the fetal spine.
Figure 3.10 (b): A different angle of approach revealed the fetal stomach (s) adjacent to the heart in this fetus with a congenital
diaphragmatic hernia.
Total acoustic shadowing: examples
Figure 3.11: Classic appearance of acoustic shadowing associated
with gallstones.
Figure 3.12: A skin calcification scanned in conventional mode
demonstrates total acoustic shadowing.
Figure 3.13: The mass in Figure 3.12 has been scanned in spatial
compounding mode. Now it casts a clean central shadow surrounded
by a dirty shadow.
In spatial compounding a smaller mass may produce multiple shadows.
Figure 3.14: A hypodermic needle is imaged in transverse view using
spatial compounding. The needle casts acoustic shadows along
multiple lines of sight. If spatial compounding was turned off, only
the middle vertical shadow would remain. This artifact has previously
been described as the “Sputnik” artifact in recognition of the
artifact’s similarity to the first artificial satellite.
Figure 3.15: The same image as Figure 3.14 with the multiple
shadows highlighted.
Shadowing can also be important in 3D imaging.
Figure 3.16: All artifacts that affect ultrasound imaging in 2D
(sectional) imaging also affect 3D imaging. Some artifacts can be
accentuated in 3D. It is also possible that the artifact may be
innocuous on multiplanar reformats (MPR), but be very pronounced
in volume renderings. This image demonstrates a surface volume
rendering of the fetal face. A large shadow is seen diagonally
crossing the face (highlighted by the arrows). The causative
structure is the fetal arm which is not seen in this projection of the
volume rendering.

3.4 Dirty shadowing

Also known as
Incomplete shadowing
Causative mechanisms
As with total acoustic shadowing, a structure of high attenuation is present within the field of view adjacent to a structure of normal or low attenuation. Ultrasound penetrates into the causative structure and produces some echoes but is highly attenuated in the process.
Which system assumptions have been breached?
The assumption that the attenuation rate is constant.
Typical appearance
Hypoechoic shadow beyond the causative structure. Always along the lines of sight.
Can this artifact be reduced, eliminated or circumvented?
If the dirty shadowing is minor, adjustments in TGC and gain may allow the operator to investigate the tissues beyond the causative structure. Some machines also allow the operator to change lateral gain compensation or the system may have a built-in automated gain compensation algorithm. If the dirty shadowing is severe, a new acoustic window must be used to interrogate the distal tissues. This may not always be technically possible.
Is this artifact diagnostically useful?
Most of the time, this artifact is a nuisance and is not diagnostically helpful.
In the example below a Mirena intrauterine contraceptive device (IUCD) casts a complete or dirty shadow depending on the beam dimensions at the location of the device.