Figure 6.10 (b)
Figure 6.10 (c): The image in Figure 6.10 (a) is particularly degraded
by reverberation. A small mass in the superficial portion of the liver
would not be detectable.

6.2 Comet tail artifact

Causative mechanisms
Similar to reverberation, two or more strong tissue reflectors are interrogated at normal incidence (90°) and the acoustic pulse becomes partially trapped by the reflectors causing it to reflect back and forth. What separates comet tail artifact from large scale
reverberation is the small focal nature of the reflectors and therefore narrow width of the artifact. The comet tail artifact is often seen in adenomyomatosis/cholesterolosis of the gallbladder where cholesterol crystals in the mucosal folds of the gallbladder form reflective interfaces that promote the artifact’s formation.
Which system assumptions have been breached?
The assumption that each reflector generates an echo only once.
Typical appearance
Small stalactite-like artifacts projecting along lines of sight into otherwise anechoic regions.
Can this artifact be reduced, eliminated or circumvented?
Comet tail artifacts are usually persistent and are difficult to eliminate. Fortunately due to their small size, they rarely pose a major problem for adequate assessment of the target structure. The usual methods of reducing reverberation often fail to reduce this artifact completely. Spatial compounding tends to reduce the length of the comet tail. With large comet tail artifacts, a different angle of approach can help change the artifact’s position and allow visualization of structures previously obscured by the artifact.
Is this artifact diagnostically useful?
Since this artifact is characteristically associated with certain pathologies such as adenomyomatosis/cholesterolosis and colloid thyroid nodules, it is diagnostically useful in these specific circumstances.
Figure 6.11: Classic appearance of adenomyomatosis of the
gallbladder wall with numerous comet tail artifacts.
Figure 6.12 (a): Colloid cysts in the thyroid often demonstrate comet
tail artifacts.
Figure 6.12 (b): A second example of comet tail artifact due to
colloid cysts in the thyroid.
Figure 6.13 (a): Sludge-filled gallbladder with floating cholesterol
crystals which occasionally line up in a favorable configuration to
cause comet tail artifacts.
Figure 6.13 (b): A second example.
Comet tail artifacts can be associated with sutures, prosthetic material and foreign bodies.
Figure 6.14 (a): In this image, numerous small comet tail artifacts
are associated with a carotid endarterectomy patch. Note the artifact
is slightly larger in this non-compounded image than in the
compounded image below.
Figure 6.14: The comet tail artifact is less pronounced in this
compound image.
Figure 6.15: Wider comet tail artifact and dirty shadowing are
associated with a surgical clip in this patient post cholecystectomy.

Chapter 7: Ringdown

The ringdown artifact is classically associated with gas. It can also be seen with metal such as foreign bodies and needles. Ringdown is a prominent artifact which appears as streams of high amplitude echoes all the way to the bottom of the image.
Causative mechanisms
For ringdown to occur, high amplitude echoes need to be produced continuously after the incident ultrasound pulse strikes the causative structure. This can occur when the ultrasound beam encounters a cluster of gas bubbles. Gas bubbles act as elastic bodies and they can vibrate or resonate in the ultrasound beam. Ringdown is also sometimes seen in association with metal (foreign bodies, needles, surgical staples) because these structures continue to ‘ring’ after the acoustic pulse has gone through.
Which system assumptions have been breached?
The assumption that each reflector produces an echo only once.
Typical appearance
Individual streaks or entire curtains of high amplitude echoes emanating from the causative structure along the line of sight. If associated with bowel gas, peristalsis will change the location of the artifact in the image.
Common misconceptions
Students of ultrasound often confuse reverberation and ringdown. The two are quite distinct artifacts in their causation and