Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_37_библиотеки_им_акад_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
15.09.2026
Размер:
17 Мб
Скачать
☆
CHAPTER 9 Diagnostic Imaging in the Neonate
https://t.me/medicina_free
209
using US to explore the ocean floor of the North Atlantic. Medical ultrasonography has its roots in sound navigation and ranging (sonar) developed during World War II.
In medical ultrasound, a transducer (essentially a piezoelectric crystal) converts electrons into a mechanical vibration that creates high-frequency sound waves within the body. The same transducer serves as both the transmitter of the sound wave and the receiver of the reflected sound. Within the body, these high-frequency sound waves propagate through the soft tissues until they meet a reflective surface that reflects some of those sound waves back to the transducer. The percentage of the sound beam reflected relates to the difference in the acoustic impedance of the material being evaluated. When the acoustic impedances of materials are similar, as is the case with the musculature of the abdominal wall (e.g., liver, kidney), most of the sound is transmitted, and a small percentage is reflected at each interface. As the sound wave travels through the abdominal wall to the liver, the abdominal wall–liver interface reflects a portion of the beam and transmits most of the sound through the liver to the liver–kidney interface. The small difference in acoustic imped­ance between the liver and kidney causes reflection of some of the beam and transmission of most of it to the posterior abdominal wall. This allows the visualization of multiple interfaces that are deeper than the first structure encountered. If the velocity of the sound beam in tissue is known, the distance to the reflective surface can be estimated by mea­suring the time it takes for the pulse to travel the distance to and from the object imaged.
Most of the tissues in the body have similar acoustic impedances; however, air has extremely low impedance, and bone has extremely high impedance. This means that there is a big dif-
ference in the acoustic impedance between these substances and the organs most commonly imaged. For this reason, both bone and air reflect nearly all of the sound that reaches them. This is why a cou­pling gel is used on the skin surface to eliminate the air gap between the transducer and the skin. This also explains why imaging through the liver gives a good acoustic window to deeper struc­tures, but bowel gas obscures imaging lower in the abdomen. For ultrasonographic imaging of the
brain in a neonate, the anterior fontanel serves as the acoustic window because the bone of the skull acts as a reflective surface that limits
through-transmission of US to deeper struc­tures. The mastoid fontanel serves as a window
to the posterior fossa. Bulk fluids within the body, such as urine in the urinary bladder, bile in the gallbladder, or cerebrospinal fluid in the ventricles, have no internal interfaces and therefore are seen as solid black on conventional ultrasonography. Cysts have a sharp posterior wall and have increased through-transmission because the sound wave pen­etrates the fluid without any reflections to block transmission of the sound.
Doppler ultrasonography takes advantage of the physical principle that the US reflection from a moving object distorts the wavelength, with the distortion related to the velocity of the object being measured. This is the principle that causes the pitch of a train’s whistle to change from high to low as the train passes an observer. It is the same principle used by radar guns to monitor the speed of a car or to measure the velocity of a pitcher’s fastball. In fact, this same principle is responsible for the “red shift” observed by astronomers in determining that we live in an expanding universe. The Doppler evalua­tion in medical ultrasonography uses the distortion of the wavelength caused by moving red cells to identify flowing blood.
One of the major advantages of US imaging is the lack of ionizing radiation. Although most
diagnostic imaging utilizes low doses of ionizing radiation, any radiation exposure is a concern and should be avoided when possible. The portability of US equipment has made it a valuable adjunct to diagnostic imaging in the neonatal intensive care setting.
Clinical Utility in the Neonatal Intensive Care Setting
Ultrasonography has had a major effect on the evaluation of the neonatal brain. Most of the
early work focused on intracranial hemorrhage, which was a common occurrence in preterm neo­nates. Even though the incidence has decreased,
intracranial hemorrhage remains an issue for which US imaging is extremely well suited.
Ultrasound equipment has improved tremendously, and with the addition of color and pulse Doppler technology, great strides have been made in the refinement and sophistication of intracranial imag­ing. Numerous complex structural abnormalities can be recognized, and screening for developmental
UNIT TWO Support of the Neonate210
https://t.me/medicina_free
abnormalities can largely be accommodated with cranial ultrasonography. Because bone reflects most of the sound, limiting through-transmission, the open fontanel is the window to the brain. As the fontanel closes over time, ultrasonography becomes less and less useful for intracranial imaging.
Renal imaging offers another major role for
ultrasonography in the neonatal unit. The kid-
neys are well visualized ultrasonographically, either through a posterior approach or, more commonly, by using the liver or spleen as soft-tissue acoustic windows to the kidneys. US is an excellent way
to evaluate hydronephrosis, which is now fre-
quently picked up on routine prenatal evaluations.
US has a role in the evaluation of a jaundiced patient because it is ideal for evaluating cystic structures, such as the gallbladder, and can readily
identify dilated biliary ducts. Jaundice caused by biliary obstruction from a choledochal cyst, for instance, can be diagnosed readily with US. Because of the reflectivity of bone and bowel gas, US imag­ing is much more effective in the upper abdomen, in which the liver and spleen serve as the acoustic windows, or in the pelvis, in which the urinary bladder can function as the window.
Although US is limited by bone, it has a signif-
icant role in the evaluation of the hips in the neonate. Because the capital femoral epiphysis of the
newborn is cartilage, the hip can be well imaged in a neonate. Maternal estrogen causes ligamentous laxity. This changes significantly during the first weeks of life; therefore, the accuracy of hip ultrasound
examinations improves after the first 3 to 4 weeks of life. Ultrasonography is very good for the detec-
tion of developmental dysplasia of the hip and can be used to evaluate the degree of femoral head coverage, the acetabular angle, and any instability of the hip.
The use of US to provide the localization
of central lines is becoming common practice.
The types of lines frequently used for infants in the NICU include umbilical venous catheters (UVCs), umbilical arterial catheters (UACs), and PICCs.
Many complications can occur if the tips of these lines are not in the correct position, including pleural effusions, cardiac arrhythmias, cardiac tamponade, thrombosis, liver hematoma, liver injury, and portal hypertension. The gold stan-
dard for assessing the position of lines and tubes has been radiographs of the chest, abdomen, or both.
The use of US has many advantages over radi­ography, including real-time assessment of the
line; no radiation exposure; the ability to place and manipulate the catheter under US guidance; reduced insertion time; and fewer malpositions, manipulations, and complications.* Bedside US
can be accomplished with appropriate training. Numerous recent studies have suggested that the use of US for line placement has good sensitivity and specificity compared with radiographs.
†
Focused Discussion: Cranial Ultrasonography
Ultrasonography is an ideal tool for evaluating the brain in a newborn. In general, an ultrasono­graphic examination is the first step in the imaging evaluation for any neurologic question. A routine
screening US is recommended at between 7 and 14 days of age in all premature infants born under 32 weeks’ gestation and should be repeated at 36 to 40 weeks of gestational age.32
Structural abnormalities, intracranial hemorrhage, sequelae of anoxic or ischemic events, and infec­tion are all well assessed via ultrasound. The most common approach is through the anterior fontanel, but additional information can be gained with axial imaging through the squamosa of the temporal bone. The posterior fossa can be evaluated through the posterior lateral fontanel. Familiarity with the normal anatomy is essential. Coronal and parasagit­tal views are obtained. Normal structures can be
easily recognized; their absence or deformity can define developmental abnormalities of the brain.
The ventricular size and configuration are assessed. Characteristic ventricular configurations can define lobar or semilobar holoprosencephaly. In addition, the ventricular configuration can suggest septo-op­tic dysplasia or agenesis of the corpus callosum. The corpus callosum can be visualized directly; abnor­malities of the corpus callosum are commonly asso­ciated with Chiari malformation and other structural abnormalities of the brain, such as Dandy-Walker malformation. Dilation of one or more of the ven-
tricles can be an indication of a pathologic con­dition. An obstruction of the flow of cerebrospinal
fluid (CSF) in the region of the Sylvian aqueduct manifests with a disparity in ventricular size. The lateral and third ventricles are enlarged, whereas the fourth ventricle remains normal in size. Dilation of
*
References 2, 10, 12, 16, 17, 18, 19, 23.
†
References 2, 12, 16, 17, 18, 19, 24, 28.
CHAPTER 9 Diagnostic Imaging in the Neonate
https://t.me/medicina_free
211
one of the lateral ventricles, especially when associ­ated with an area of porencephaly, is indicative of an in utero destructive event. Seizures or apnea may
indicate an anoxic or ischemic event in a neo­nate. Certain structural abnormalities can suggest a
specific diagnosis; for instance, periventricular nod­ules and cortical tubers define tuberous sclerosis. US examination is less sensitive than computed tomog­raphy (CT) and magnetic resonance imaging (MRI) in defining subtle areas of gray-matter heterotopia or focal pachygyria, examples of developmental abnormalities associated with seizures. US imaging
is very sensitive to intracranial hemorrhage, and areas of increased echogenicity can be demon­strated in areas of edema. Intracranial hemorrhage
is an important concern in a premature neonate (Fig.
9.8) and is classified into four grades; each grade is
prognostically significant.34 Grade I hemorrhage usually has a good outcome, whereas the progno­sis with grade IV hemorrhage is frequently poor.
Grade I hemorrhage is confined to the germinal matrix in the caudothalamic groove. This is the last fetal germinal matrix to mature and is prone to hemorrhage in preterm babies. Grade II intracranial hemorrhage has intraventricular blood. Grade III hemorrhage is associated with ventricular dilation as the intraventricular clot enlarges the lateral ventricles. Grade IV hem­orrhage is defined by parenchymal extension. It
has been hypothesized that grade IV hemorrhage
may be the result of venous infarction that occurs from obstruction of the septal veins by the swollen germinal matrix hemorrhage. White-matter injury
of prematurity (periventricular leukomalacia) is a consequence of anoxic or ischemic injury to the brain that manifests as increased echogenicity in the deep periventricular white matter of the cen­trum semiovale. This may progress to cavitation and is then called cystic leukoencephalomalacia. US
can detect changes of leukoencephalomalacia, which is usually apparent within 2 weeks of birth.
33
COMPUTED TOMOGRAPHY
CT was initially developed in 1972 in Middlesex, England, by Electric and Musical Industries (EMI), an industrial research company. After signing the Beatles in 1962, the company sold its computer business; however, it kept a researcher named Godfrey Hounsfield and funded his independent research through the revenue generated by the Beatles’ success. Hounsfield imagined that he could determine what was inside a box by taking x-rays of the box at all angles. He then worked to build a computer that could reconstruct a slice of an object from the data of these x-rays acquired at various angles. He shared the Nobel Prize in Physiology and Medicine with Allan Cormack, who developed the theoretical mathematics for the invention of CT.
A B
FIGURE 9.8 Coronal (A) and parasagittal (B) ultrasound images from a cranial ultrasound study demonstrate an echogenic clot (arrows)
within the dilated right ventricle. The intraventricular clot with ventricular dilation defines a grade III hemorrhage.
UNIT TWO Support of the Neonate212
https://t.me/medicina_free
Early CT scanners produced images by acquiring x-ray data in multiple positions in an axial plane, moving to a different level between acquiring slices. The initial EMI scanner solved mathemat­ical equations representing the attenuation of the different x-ray beams, like a giant Sudoku puzzle. The complexity of mathematics involved as well as limitations in computers at the time contributed to the limited speed of the scanner; it took approxi­mately 10 minutes to acquire and reconstruct each slice, and the resolution was limited to an 80 × 80 matrix. For comparison, today’s fastest dual-source CT scanners can acquire 64 slices in about 0.07 second in the fastest scanning mode with a matrix of 1024 × 1024.
On modern multidetector helical CT scanners, image acquisition takes place as the patient is car­ried on a mobile table through a rotating gantry containing an x-ray source on the opposite side of the gantry from x-ray detectors. A ring of data is acquired as the patient is moved through the gantry; the path of image acquisition is in effect a helix, resembling a coiled spring, hence the term helical or spiral CT. Multidetector arrays of 4, 8, 16, 64, and even 320 elements allow rapid acquisition of multiple slices in a single rotation of the tube. This renders high-resolution isovoxel data sets (each volume element has the same width, height, and depth). Because the data are acquired in a contin­uous helix in an isovoxel data set, as if peeling an apple from top to bottom in one peel, the data can be reconstructed in any plane as well as rendered in three-dimensional (3D) format, just as you could imagine putting an apple peel back together to look like an intact apple. Reconstruction of the acquired data has undergone and continues to undergo revi­sions. In the most commonly employed method, filtered back-projection lines of gray representing the attenuation of the patient are layered on top of each other to create images, with different filters applied to accentuate different aspects of the image. This renders a cross-sectional slice that can show all of the structures within that slice. For instance, a slice through the upper abdomen may show the liver, spleen, pancreas, both kidneys, and the spine, each separated by a plane of fat and each with a subtly different density.
New technologies on recently introduced CT scanners include volumetric scanning, dual-source and dual-energy CT scanners, and technology designed to reduce the radiation dose to patients
by limiting the patient’s exposure during the scan or allowing for improved reconstruction of scans obtained at lower radiation doses. Volumetric scanning achieved with up to 320 detectors can allow for an area up to 16 cm to be scanned in a single rotation of the scanner gantry. Dual-source, dual-energy scanners can leverage two tubes and two detector arrays, either for an increased scan­ning speed of up to 0.07 second for 64 slices with both tubes at the same energy or improved tissue differentiation using different energies. Given the
recent attention to radiation effects from CT in both the medical literature and the lay press, there has been a renewed focus on technology to reduce the dose in CT scans. This includes
both improving the mechanics of the CT scanners themselves as well as developing more advanced reconstruction algorithms that allow scans obtained at lower doses to produce excellent-quality images.
These recent advances in CT technology have resulted in spectacular images and an explosion in CT utilization. The radiation dose in CT, however, is significantly higher than that in routine radiography. CT now accounts for more than 60% of the radiation exposure from medical imaging in the United States.
Trailing this growth in the utilization of CT has been greater understanding of and concern for the effects of ionizing radiation in medical imaging. Evaluation of the risk associated with the low doses of ionizing radiation used in medical imaging is a complex and evolving topic. The majority of stud­ies estimating risk from low doses of radiation have been based on models in which survivors of large known doses of radiation, such as from the atomic bombs in Japan, have been followed longitudinally to assess the risk of developing cancer. These mod­els assume that by extrapolating the effects of large radiation doses to low doses, the risk for low doses can be assessed in what is called a linear no-thresh­old model. Although the accuracy and assumptions in this type of model have been questioned, with some investigators claiming that there is in fact no risk, or even benefit, with low doses of radiation,
more recent studies have suggested that there is a small increased risk of cancer associated with clinical medical imaging.
5,6,13
A retrospective
cohort study that looked at the incidence of leukemia and brain tumors in pediatric patients who had undergone CT demonstrated an asso­ciation between the radiation received from
CHAPTER 9 Diagnostic Imaging in the Neonate
https://t.me/medicina_free
213
CT scans and leukemia and brain tumors.25 Although there was a significant increase in rel­ative risk suggested in this study, the absolute risk remained low.
Medical research suggests that the radiation dose currently used in diagnostic CT is associated with an increase in the risk for radiation-induced malignancy. Neonates, infants, and children are more susceptible to the effects of radiation than adults. This stems from the fact that for a given exposure to radiation, the smaller the patient, the greater is the effective dose because more of the radiation penetrates the patient. Also, longer
life expectancy puts younger patients at an increased risk because there is a longer period of time for a radiation-induced complication to develop.
Ionizing Radiation in Perinatal Medicine
The effects of ionizing radiation on the fetus of a pregnant patient also present complex and important concerns. Although the use of ionizing
radiation generally should be avoided in patients who are pregnant, there still exist indications for which the benefit of performing the examination may outweigh the risk to the fetus. Additionally, on rare occasions, a CT may have been performed on a patient who was not known to be pregnant before the examination. Although a comprehensive analy­sis of the risk and radiation exposure of a pregnant patient undergoing CT should be considered in consultation with a medical physicist, general guide­lines have been provided by the American College of Radiology in conjunction with the Society for Pediatric Radiology, as briefly described in the fol­lowing paragraphs:
Before conception, there has been no docu­mented genetically heritable risk in the human population. In the first 2 weeks after conception,
the only potential risk is felt to be the loss of pregnancy. Doses associated with radiographic
procedures have not been clearly associated with increased risk, although this is difficult to determine because approximately half of all conceptions are lost in this period, often without recognition of the loss by the woman.
Radiation exposure between 2 and 15 weeks after conception has more complex risk impli­cations. In general, radiologic procedures outside of
the abdomen and pelvis, including the head, neck,
1
and chest, should result in only a very low dose to the fetus from scatter radiation. In a patient who
is known to be pregnant, the study should be optimized to limit the dose to the fetus even further. When imaging of the abdomen and pelvis
is indicated or has been performed, consideration of the risk to the fetus takes on greater importance. In
centers that carefully manage their CT radiation dose, as do most children’s hospitals, the dose to the fetus is thought to be below the level associated with any developmental abnormality.
However, before counseling takes place, verification of the dose by a qualified medical physicist is rec­ommended. If the dose is determined to be low,
the majority of the risk associated with radiation exposure to the fetus is a small increased risk of cancer in later life, and termination of preg­nancy would not be indicated. Doses associated with increased developmental disorders are uncommon in routine practice and usually occur in circumstances that have important implica­tions for the pregnancy, such as in patients who require complex cardiology or interventional radiology procedures.
The effects of radiation exposure to the fetus
more than 15 weeks after conception are even smaller, with risk to the developing nervous system
occurring only at very high doses, usually beyond what would be typically encountered even with multiple radiology procedures. Therefore, after
2 weeks’ gestation, the predominant concern with low-dose diagnostic imaging is the small increase of developing cancer over a lifetime.
Although concern for the deleterious effects of radiation is important, it should always be viewed in the context of the patient as a whole. CT can be a powerful tool in evaluating the pediatric patient. Caution is the key: (1) image
only when indicated; (2) limit the scan to the region of concern; and (3) be cognizant of dose and use as low peak kilovoltage (kVp) and milliampere-second (mAs) as possible, and use dose-reduction technologies while maintaining
diagnostic-quality examinations.
Clinical Utility in the Neonatal Intensive Care Setting
Cranial imaging is the most common use of CT in most neonatal intensive care settings.
CT adds significant specificity to the abnormalities
UNIT TWO Support of the Neonate214
https://t.me/medicina_free
recognized with ultrasonography. Concern about ionizing radiation and the fact that CT equipment is generally not portable makes obtaining a CT more difficult than obtaining a sonogram. CT is more
accurate in assessing the nature of extra-axial fluid collections and is very helpful in further defining structural abnormalities of the brain, particularly those associated with the abnormal distribution of gray or white matter. It is also
very good for evaluating intracranial hemorrhage and infection. Exquisite bone detail defines cranio­facial anomalies, choanal atresia and stenosis, and abnormalities of the petrous bone associated with hearing loss.
Focused Discussion: Computed Tomography Angiography of the Chest
Chest CT angiography (CTA) is becoming much more frequent in the NICU. CTA is often
used to evaluate abnormalities detected during intrauterine US, chest x-ray, and echocardiogram examinations, such as congenital lung lesions and potential surgical lesions such as vascular rings iden­tified on chest x-ray films.
Through carefully controlled administration of intravenous (IV) contrast, CTA can supple­ment, and in many cases replace, the need for traditional catheter angiography.15 Uses include
describing the vascular supply to congenital lung lesions such as bronchopulmonary sequestrations, defining vascular and airway anatomy for surgical planning in patients with vascular rings, and surgical planning in patients with complex cardiac disease.
Congenital lung lesions are a spectrum of devel­opmental abnormalities of the lung that include bronchial atresia, congenital pulmonary airway malformations, and bronchopulmonary sequestra­tion, among others. Often suggested on prenatal sonography and, in some cases, characterized on fetal MRI, CTA is often used in the postnatal
period to further characterize the lesion and plan surgical intervention.11 Bronchial atresia can
be diagnosed on CTA when an atretic bronchus is identified, often in association with a dilated, mucus-filled segment or “bronchocele/mucocele” distal to the atretic airway. Although often found in isolation, studies have identified bronchial atresia in up to 77% of other congenital lung lesions.20 Cystic pulmonary airway malformations (CPAMs) have been traditionally defined by the size of the
7,11
cysts within the lesion, with type 1 having cysts greater than 2 cm, type 2 with cysts less than 2 cm, and type 3 appearing solid; more recent alter­native classification systems have been proposed, and the pathology of type 3 has been disputed. CPAMs can have a multilobar and rarely bilateral distribution. CTA can define the cyst size and
distribution as well as provide a roadmap for surgical planning. Rare malignant potential for
type 1 CPAMs has been reported, as has recurrent infection. Bronchopulmonary sequestrations were traditionally defined as portions of the lung with a systemic arterial supply and then further charac­terized as either extralobar or intralobar. Extralobar sequestrations most often appear as a wedge-shaped mass with systemic arterial supply with a pleural margin, often on the left and below the otherwise normally formed lung. Intralobar sequestrations are thought to be developmental abnormalities of the lung composed of isolated nonfunctional lung tissue, sometimes accompanied by cystic change within an otherwise-normal lobe. CTA provides excellent identification and characterization of both the vascular supply and associated parenchymal change in these lesions.
Although vascular rings are often initially suggested on chest x-ray or upper GI, CTA and MR angiography have largely replaced other techniques in the characterization of vascular
7,14
rings.
CTA has the advantages of easy availability, improved depiction of the airway and lung paren­chyma, and short scanning times, which in some cases can eliminate the need for sedation. Disadvantages include the use of ionizing radiation and iodinated contrast. MR angiography is a viable alternative in some cases in which the need for a detailed evalua­tion of the airway and lung parenchyma justifies the need for sedation.
In embryonic development, the ventral and dor­sal aorta are connected by six pairs of aortic arches. Normally there is regression of some of these arches to result in the normal left-sided aortic arch. The sixth arch forms the ductus arteriosus and proximal right and left pulmonary arteries. Failure of regres­sion or inappropriate development of these primi­tive vascular structures leads to the development of vascular rings and slings.
Vascular rings are formed when vessels, either patent or atretic, encircle the trachea and esoph­agus, with the potential to result in esophageal or tracheal compression. The most common
CHAPTER 9 Diagnostic Imaging in the Neonate
https://t.me/medicina_free
A B
FIGURE 9.9 Axial (A) and volume-rendered three-dimensional (B) images of a computed tomography angiogram of the chest demon-
strate a double aortic arch with hypoplastic left arch (long arrow, A and B) with atresia of the posterior aspect of the left arch (short arrow, B). The volume-rendered image (viewed from behind) better demonstrates the spatial relationships of the arches to each other as well as to the encircled trachea (star).
215
symptomatic vascular ring is a double aortic arch (Fig. 9.9).14 In most cases the right arch is dominant
and higher than the left, although the arches can be the same size and one arch can be atretic. The second most common vascular ring is a right-sided aortic arch with aberrant left subclavian artery and left-sided ligamentum arteriosum. Less common vascular rings include left-sided aortic arch with aberrant right subclavian artery and right-sided liga­mentum arteriosum and right-sided aortic arch with mirror-image branching and left-sided ligamentum arteriosus as well as circumflex aortas. In several of these conditions, the sidedness of the ligamentum arteriosus or other atretic structures determines if a complete vascular ring is present. The atretic structure itself usually cannot be visualized. Instead, the presence of a dimple on the aorta or pulmonary artery, diverticulum of the subclavian artery, or position of the descending aorta on the side opposite the arch (also referred to as the three “Ds”) determines the position of the atretic segment.
Ultrasonography remains the first-line diag­nostic tool for the evaluation of the kidneys, liver, and spleen, but when a pathologic con­dition of the abdomen is a concern and a good acoustic window for US imaging is not available,
CT is frequently the examination of choice. CT can be performed with significantly less seda­tion than that necessary for MRI. CT eliminates
many of the artifacts, including those of vascular flow, respiratory motion, and even bowel peristalsis, that limit the utility of MRI. Skeletal lesions are
well visualized with CT. Ultrasonography is the method of choice in the evaluation of congen­ital hip dysplasia, but CT can be very helpful in evaluating the position of the femoral heads after reduction and treatment of congenital hip dysplasia when the patient is immobilized in a plaster cast.
CT can be very helpful in identifying the organ of origin of a specific pathologic condition.14 This is, of course, the first step in narrowing a differential diagnosis. The addition of IV contrast can define the presence and extent of tumor and infection or abscess. CT can also evaluate for adenopathy, as well as the presence of tumor thrombus in renal arteries and the inferior vena cava, which may affect the surgical approach to renal and hepatic neoplasms. The findings on CT often lead to a specific diagnosis. The multiplanar, cross-sectional rendition of anatomy, which allows structures to be distinguished from one another, and the improved
UNIT TWO Support of the Neonate216
https://t.me/medicina_free
contrast resolution make CT a useful tool when clinically indicated.
Focused Discussion: Intracranial Blood
Noncontrast CT is extremely sensitive and specific for the detection and localization of intracranial blood (Fig. 9.10). Blood from acute
hemorrhage has a density on noncontrast CT (measured in Hounsfield numbers) higher than any normal structure except bone and calcium. The increased contrast resolution of CT allows the dif­ferentiation of gray matter from white matter and lends itself to a detailed structural evaluation of the brain. The ventricles are low in density (0 Hounsfield units, equal to water), the white matter is denser, and the gray matter is even more dense still, followed by blood from acute hemorrhage and, finally, calcium and bone. Blood from acute hemorrhage is visu-
alized as white as soon as a clot is formed, and this high density will slowly decrease over time.
For instance, the blood in a subdural hematoma after 2 to 3 weeks will become lower and lower in density until it is indistinguishable from water and CSF. MRI can differentiate blood from a chronic
subdural hematoma for a longer time than CT because the protein within a chronic subdural hematoma modifies the signal on MRI for an extended period. Although all intracranial blood
does change density over time, the compartment in which the blood is found affects the rate of change to some extent. Therefore, the timing of an event
responsible for the blood cannot be precisely determined based on density alone.
The location of the blood is the next issue. CT is
the most accurate imaging method for detection of subarachnoid blood. The presence of subarach-
noid blood postpartum is common, even after a relatively nontraumatic birth. Unfortunately, on rare occasions, subarachnoid blood can cause vasospasm of vessels near the skull base, which can result in relative ischemia or hypoperfusion of the peripheral cortex. Areas of edema can be detected by looking for the loss of the normal gray–white differentiation or by finding a focal area of brain edema charac­terized by relatively low density resulting from the addition of low-density water to an otherwise-nor­mal area of the brain.
The shape of a collection of blood is important in
evaluating intracranial hemorrhage. Subarachnoid
FIGURE 9.10 A single axial image near the vertex demonstrates a high-den-
sity lenticular mass (arrows) in the extra-axial space over the right cerebral cortex. The lenticular configuration is that of an epidural fluid collection and the high-density characteristic of blood from a chronic hematoma.
blood assumes a configuration that follows the arach­noid space. Therefore, it is most frequently seen in the suprasellar cistern, the ambient cistern, the Sylvian fissure, or the interhemispheric fissure, or layering on the tentorium. The most sensitive locations for identifying subarachnoid blood are in the region of the quadrigeminal plate cistern, the posterior aspect of the third ventricle, and the interpeduncular cistern. Subdural hematomas occur most frequently over the convexities or along the interhemispheric fissure. Those over the convexity can be differentiated from epidural hematomas by their crescentic configuration as opposed to the lenticular configuration of an epi­dural hematoma. The dura is the periosteum of the inner table of the skull; therefore, an epidural hema­toma is limited by the adhesion of the periosteum to the skull and hence the lenticular configuration. This also explains why epidural hematomas are most often associated with higher-pressure arterial bleeding and why subdural hematomas frequently are associated with venous bleeding. Another key to differentiating the compartment is the relationship to cranial sutures.
An epidural hematoma will not cross a suture line because of the anatomic limitation of the dura by the suture. A similar limitation by the dural attachment at suture lines helps distinguish a cephalohematoma from a caput succedaneum.
The direct sagittal imaging of MRI has revealed the high prevalence of subdural blood in the posterior fossa.
CHAPTER 9 Diagnostic Imaging in the Neonate
https://t.me/medicina_free
217
MAGNETIC RESONANCE IMAGING
Background
MRI is a modality that images protons or hydro­gen ions within the body. The rapid development
of magnetic resonance was a result of, in part, the transfer of sophisticated reconstruction algorithms used in CT and the computer power developed in other fields, such as the 3D graphics used in anima­tion, cartography, and seismology. These technologic advances allow tremendous amounts of information to be manipulated quickly enough to make image reconstruction a reality. MRI is essentially hydro-
gen imaging, and because the human body is 98% water, much hydrogen is available to image.
MRI is performed by placing a patient in a strong magnetic field, which varies slightly from the head to the foot. Each proton acts as a small magnet, and just as the needle on a compass orients itself in one direction when placed next to a magnet, the protons in the body align when placed into a strong magnetic field. This alignment of protons is essential to create an environment that has a net electromagnetic field. Without the alignment of protons by the magnetic field, the random orientation of protons would have no measurable net field effect when stimulated and therefore would create no signal to image.
Once the patient is in the magnetic field, a radiofrequency (RF) pulse is delivered. In current imaging systems, the pulse wave has the frequency of an FM radio wave. Less than 1 in 1 million hydrogen ions will absorb any energy, and only cer­tain RFs will allow the transfer of energy from the RF pulse to a hydrogen ion.
An analogy of this energy transfer can be seen on a schoolyard playground. Visualize a child on a swing. When pushing the swing in rhythm or reso­nance with the natural frequency of the motion of that swing, the swing will absorb the energy, and the child will swing higher and higher with each push. This natural rate of harmonic motion depends on the length of the rope on the swing and the mass of the swing and the child. If you were to push at a rate that was not synchronous with the swing’s natural rhythm, pushing would not allow the energy to assist in propelling the swing higher and higher, and in fact, you would disrupt the normal rhythm of the swing.
In a famous TV commercial for Memorex in the 1970s, the playback of Ella Fitzgerald’s voice caused a goblet to break, demonstrating the absorption of resonance frequency by the crystal in the goblet. The absorbed energy caused the goblet to shatter. In MRI, the FM RF energy is used to stimulate hydrogen ions or protons in the body.
Because the field strength of the magnet used for imaging varies slightly from one end of the patient to the other and the resonance frequency depends on the field strength of the magnet, one can selectively stimulate various locations within the patient. By changing the RF slightly, a differ­ent specific group of protons is stimulated. Protons stimulated by an RF pulse absorb energy and move to an unstable higher-energy state. They give up the absorbed energy as an RF pulse or “echo” of the pulse they received. The echo is received by an antenna, just as with a radio receiver, and converted to an image. The signals or echoes received are the result of T1 and T2 relaxation times, which are sim­ply physical parameters that describe the environ­mental interactions that influence the signal released from a proton. Such spin-echo pulse sequences are frequently used sequences in routine MRI.
An analogy of a spinning top can help to explain the T1 and T2 relaxation times that result in spin­echo imaging. Each hydrogen ion has a dipole moment (a positive pole and a negative pole) and therefore acts like a small magnet within the powerful magnetic field of the imaging magnet. These protons spin or precess with a precessional frequency that is related to the field strength of the magnet. Electromagnetic energy can be transferred to these protons if the energy is delivered at the res­onance frequency. Once an RF pulse of resonance frequency is delivered, a small number of protons will absorb this energy and move to a higher-en­ergy state. The T1 relaxation time reflects the time it takes for these excited protons to give up their higher energy and return to baseline.
T2 relaxation times relate to a second parame­ter of physical interactions. Although the protons are rotating at a frequency proportionate to the magnetic field in which they exist, they are not in phase. In other words, there is no net direction of polarity from all of these spinning magnets. Once the RF pulse perturbs or stimulates these protons, they begin to spin synchronously and therefore cre­ate a net magnetic field. This spinning net magnetic field generates an electromagnetic wave that can be
UNIT TWO Support of the Neonate218
https://t.me/medicina_free
picked up by the RF antenna of the imaging sys­tem as an “echo” of the original RF pulse delivered. (The principle of a spinning magnet inducing an electromagnetic pulse is the basis for the turbines of hydroelectric generators.)
Because its immediate electromagnetic environ­ment affects each proton differently, these protons will remain synchronous in their precession for a very short period. As they move out of phase or synchrony, the net magnetic field that was created dissipates; therefore, the signal received by the RF antenna diminishes. The T2 relaxation time indi­cates the time it takes the protons to go from a state of synchronous rotation, when maximal signal is created, to random, out-of-phase precession, with zero net magnetic field and hence no signal. The requirement of a net magnetic field to create a signal is used to evaluate flowing blood without the need for contrast. An RF pulse saturates the protons in the field being imaged. The saturated blood within the vessels of that field flow out of the field and are replaced with nonsaturated blood from an adjacent slice. Consequently, there is no signal from the vessel containing the blood flowing perpendicular to the slab being imaged.
Diffusion-weighted sequences have proven to be very sensitive in defining neonatal pathology. Diffusion weighting is the most sensitive technique in the identification of early anoxic-ischemic injury. Diffusion weighting takes advantage of the random Brownian motion of molecules in fluid and the restriction of Brownian motion by anatomic bar­riers or edema. Water within the ventricular system will diffuse homogeneously in all directions (i.e., no restricted diffusion) and will be a low signal on a diffusion-weighted sequence and high signal (white) on an apparent diffusion coefficient (ADC) map. An acute or subacute infarction, for instance, will cause restricted diffusion in the affected region as fluid rushes into cells due to failure of the cells’ homeostatic mechanisms, such as ion pumps. The restricted diffusion caused by acute cell death will be a high signal on the diffusion sequence and a low signal on the ADC map. Late findings after infarction will demonstrate facilitated diffusion with a high signal on both the diffusion and ADC map as the infarcted cells burst and therefore no longer restrict the movement of water molecules. Diffusion tensor imaging measures diffusion in at least six planes simultaneously. From that data, a map of the magnitude and direction of water movement can
be generated. The axons within white-matter tracts will allow diffusion in the direction of the axon but will restrict diffusion in any direction other than the course of the axon. This allows one to map the white-matter tracts and has been studied in relation to developmental abnormalities in the brain, the relationship of intracranial neoplasm to white-mat­ter tracts, and the plasticity of the developing brain in response to injury.
The key feature of the physics of MRI is that
images are acquired without ionizing radiation, which is particularly important in pediatrics. No
known harmful effect of either magnetic exposure or RF exposure at the levels used in MRI has been observed. However, MRI is still relatively new,
and one should be cautious in using MRI for fetal and newborn imaging. Energy deposition is a concern, and protocols have been established that limit patient exposure. Another concern is
the effect a magnetic field might have on electronic instrumentation, such as pacing devices and metallic surgical clips. The torque on metallic implants can be quite high, but this is rarely of clinical signifi­cance. However, the artifact caused by the distur­bance of the magnetic field can be significant. The most important and real safety concern is that of the magnetic-field attraction of ferromagnetic material. Pens, stethoscopes, or even oxygen canisters can act as projectiles when inadvertently brought too close to a magnetic field.
The main drawback of current MRI tech­nology is the time it takes to acquire an image. Motion-free imaging is necessary for optimal image quality, and because image acquisition in MRI takes minutes, sedation frequently is nec­essary. Respiratory and cardiac gating can help for
physiologic motion, but even physiologic motion can be problematic.
Focused Discussion: Practical Considerations
The physics of MRI is complex, and multiple
variables influence the signal received (Fig. 9.11). These influences variably affect the T1 and T2 relaxation times in spin-echo imaging. Imaging sequences tend to be called T1 or T2 sequences, depending on which physical parameter has the most influence on the appearance of the image. A helpful simplification of spin-echo imaging is that in T1-weighted spin-echo sequences, fluid is black,