Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5762_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
79 Мб
Скачать
16
a
C. F. Dietrich et al.
• Signal separation between microbubbles and tissue, based on the assumption that the latter behaves in a perfectly linear fashion (i.e., the tissue responses follow the trans­mitted waveform exactly).
A higher MI results in a better penetration but also
increases destruction of the microbubbles. The contrast agent dose balances the contrast enhancement intensity. In the early phase of CEUS, it prevents over enhancement of struc­tures with shadowing. In the contrast enhancement duration, a sufcient contrast agent may concentrate in the late phase.
In practice, this perfectly linear model is not completely
true since the transmitted waves become distorted as they are conducted through any medium and this produces harmon­ics, which are the basis for harmonic imaging widely used in B-mode scanning. It should be pointed out that microbubble harmonics are generated in a different way, by the fact that the microbubbles resist compression more strongly than expansion, so their response to a symmetrical ultrasound pulse is asymmetrical, generating harmonics.

2.2.2 Image Depth Penetration

neareld [6]. Most importantly bubble destruction (“the circle of disaster”) should be avoided. Regarding com­monly observed artifacts we refer to the respective para­graph below [5].

2.2.3 Focus

Usually the focus should be positioned at the distal border of the target lesion but this might vary in some scanners. For detection, a deeper location of the focus (at least two-thirds of the screen) is recommended (Fig.2.1).
2.2.4 Gain (Received Signal Amplication)
The gain should be usually set at or very slightly above the noise oor so that before microbubbles arrive, the image is dark. If the gain is set too low, sensitivity is too low and weak microbubble signals are not detected. If the gain is set too high signal saturation occurs possibly with acoustic shadow­ing (Fig.2.2).
The image depth penetration is determined by many factors

2.2.5 Background Signal (Noise)

including the manufacturer and transducer technology, transducer frequency, acoustic power (mechanical index), focus and other technology-dependent factors, and nally by the patient’s condition. Limited depth penetration can be overcome by lowering the transmit frequency with the dis­advantage of lower spatial resolution, eventually resulting in suboptimal imaging of small supercially located lesions. Increasing the MI may improve penetration but at the expense of microbubble destruction, especially in the
A dual-image display format is often recommended since the nonlinear image is almost black making it difcult to focus on the lesion of interest, which is especially necessary when examining small and difcult to detect focal liver lesions. In the dual-image display, a conventional B mode fundamental image and a bubble-only contrast image are displayed side­by- side [6]. In contrast, it is also possible to overlay the con­trast and B mode image, which doubles the screen size. For
b
Fig. 2.1 Focal zone set during liver contrast enhanced ultrasound (CEUS). Normal CEUS of the liver showed an appropriately placed focal zone at the bottom of the image (a). Poor quality CEUS of a nor-
mal liver showed focal zone set in the near eld (b), resulting in signi­cant loss of contrast signal in the far eld
b
a
2 Contrast Enhanced Ultrasound: How toPerform It inLiver Tumors?
c
17
Fig. 2.2 Gain adjustments during liver contrast enhanced ultrasound (CEUS). CEUS with appropriately adjusted gain demonstrates normal enhancement of liver parenchyma (a). While too high (b) or too low (c) set of gain was not suitable for appropriate imaging
quantitative studies, the dual-image display is advantageous

2.2.7 Frame Rate

since it is important to keep the transducer at the same place and avoid motion. It should be mentioned that the quality of the B mode image in dual-image displays is inferior to that obtained in non-contrast mode with the same settings [6].
For focal liver characterization with adequate visualization and recording the frame rate should be adjusted 10Hz. Too high frame rates can augment bubble destruction and too low a frame rate does not allow real-time imaging [6] (Table2.1).

2.2.6 Dynamic Range

2.3 CEUS oftheLiver, Examination
The dynamic range is the range of signal intensities to be displayed. It should be set to optimize the enhancement pattern. A small dynamic range will decrease the signal lev­els (“grey levels”) in the image and increases visual con­trast but can limit the differentiation between different degrees of enhancement. A wide dynamic range increases the number of “greys,” allowing for better differentiation between different degrees of enhancement [6]. A large dynamic range allows to improve identication of the increased rim signal in patients with highly vascularized metastatic lesions. A narrow dynamic range is preferred for visualization of FLL with low perfusion. A wide dynamic range should be used in perfusion quantication studies to avoid signal saturation [6].
The pre-contrast examination preparations are important, which include the identication of the target focal liver lesion, the identication of the best position of the patient, and the optimal scan plane to minimize out-of-plane motion from respiration, usually longitudinal along the axis of the respiratory movements [6].
larger) should be inserted in the antecubital vein of left arm, while avoiding interaction of the injector with the right-sided examiner (Fig. 2.3). Some important inuencing factors should be avoided, e.g., avoid the side of the breast (or axil­lary) surgery to minimize the risk of worsening lymphedema.
Technique
For injection of contrast agents, the cannula (20 gauge or
18
a
Table 2.1 Ultrasound contrast agents in clinical use
Brand Shell material Gas core Denity Lipid Octauoropropane 1.1–3.3 12.0 Yes Left ventricular opacication Optison Sonicated albumin Octauoropropane 3.0–4.5 0.5–0.8 Yes Left ventricular opacication SonoVue
(Lumason)
Sonazoid Sucrose Peruorobutane 2.1 1.2 Yes Characterization of focal liver
Lipid Sulfur hexauoride 1.5–2.5 0.15–0.56 Yes Left ventricular opacication
Mean diameter (μm)
Concentration as prepared (×1 × 109)/mL
(C)FDA approved FDA approved indications
Characterization of focal liver lesions
lesions
C. F. Dietrich et al.
b
c
d
Fig. 2.3 Preparation for liver contrast enhanced ultrasound. Choose the most suitable contrast agent, SonoVue (a) or Sonazoid (b). Preparing for the contrast agents according to manual indications (c–e). The can-
nula (20 gauge or larger) should be inserted in the left arm, preferably the antecubital vein, to avoid interaction of the injector with the right­sided examiner (f)
e
2 Contrast Enhanced Ultrasound: How toPerform It inLiver Tumors?
19
f
Fig. 2.3 (continued)
Central line and port systems can be used as long as there is no lter requiring a high injection pressure, but be aware of a possible shorten contrast arrival time [6]. The catheter should be removed after the exclusion of any pseudoanaphy­lactic reaction. When multiple injections are anticipated, a three-way stopcock may be valuable and facilitates sequen­tial administration of the contrast agent and then the saline ush, without removal of either syringe. The author almost never uses a three-way stopcock.
The timer should be started at the time of the beginning
of the UCA injection. The application via a central venous line with a much shorter arrival time is a good reason for this [6].
The injection bolus for SonoVue™ is given at about
1–2ml/s to avoid high pressure with the risk of microbubble destruction. Immediately after injecting the contrast agent, a 5–10ml saline bolus should be given to ush the line with higher pressure >2ml/s. The contrast dose depends on the
Fig. 2.4 Doppler blooming artifact. Following contrast administration, there is a marked increase in Doppler signal throughout the liver with color pixels displaying well beyond the expected vessel margins, indi­cating blooming artifact
quality of the machine and the machine setting. It is suggested to use lower dosages between 0.4 and 4.8ml in small pediatric patients. Artifacts might appear in the early
Repeated injection is advised under the following circum-
stances [6]: phases of enhancement with a too high contrast agent dose, including acoustic shadowing, over-enhancement of small structures, and signal saturation (Fig.2.4). Meanwhile, too low a dose might cause the concentration of microbubbles to be subdiagnostic in the late phase, mimicking the detec­tion of wash-out [6].
• There are additional FLL, which require characterization.
• The initial injection failed and did not provide the full answer to the detection and/or characterization of a FLL to allow for assessment of missing information.
20
C. F. Dietrich et al.
• A wash-out region may be identied on sweeps of the liver in either the PVP or the LP to allow arterial enhance­ment characterization.
2.4 Improved Detection ofFocal Liver
Lesions
Conventional ultrasound is the most commonly used imag­ing modality for focal liver lesions, but is less sensitive in the detection of FLL while comparing with CECT, CEMRI, or intraoperative US. With the application of CEUS, it has dra­matically increased detection rate of FLL before operation, especially in liver metastases 10mm [718].
2.5 Characterization ofFocal Liver
Lesions
The contrast features of focal liver lesion (FLL) should be described in terms of the enhancement degree and enhance­ment phase. It is important to know in advance if the liver is normal or diseased (e.g., liver cirrhosis, brosis, or ste­atosis). This may affect the contrast enhancement features of the FLL and its surrounding liver parenchyma. Enhancement including isoenhancing, hyperenhancing, and hypoenhancing, which refers to the progressive inten­sity of the signal relative in FLL to the adjacent paren­chyma. “Wash-out” is dened by the reduction in enhancement degree which follows peak enhancement. Sustained enhancement refers to the continuation of the iso- or hyperenhancement in the FLL relative to the adja­cent parenchyma over time. Non- enhancing refers to the complete absence of enhancement [6].
The timing (early versus late onset, fast versus slow),
degree (complete, incomplete), and pattern should be described in comparison to the surrounding “normal” paren­chyma. The combined evaluation of the arterial contrast enhancement and portal venous and late wash-out of a lesion compared to the surrounding healthy liver parenchyma allows characterization of a FLL either as non-hepatic tissue (e.g., malignant, inammatory, or brotic) if wash-out is present or as benign if iso- or hyperenhancement can be observed in comparison to the surrounding liver parenchyma.
In addition, analyzing the arterial vessel architecture in
the early arterial wash-in phase allows further characteriza­tion, especially in benign focal liver lesions as hemangioma with peripheral nodular contrast enhancement and centripe­tal ll in [19] or as focal nodular hyperplasia with typical
vascularity [20]. The vascular pattern of hepatocellular ade­noma is more complex [21]. The characteristics are also valid for pediatric patients [22, 23].
The combined evaluation of the above diagnostic features makes it possible to characterize FLL in patients with liver cirrhosis as typical for HCC according to the Liver Imaging Reporting and Data System (LI-RADS) [24, 25].

2.6 Artifacts

Knowledge of the basic physical and technical principles of ultrasound is needed to understand sonographic images and ndings and to be able to evaluate the possibilities and limi­tations of the method. Conventional and CEUS imaging are susceptible to multiple artifacts (imaging errors) since the properties assumed to be constant, such as straight-line sound propagation, attenuation (penetration), sound speed, acoustic eld characteristics (the narrower the acoustic eld, the better the suppression of side lobes not corresponding to wanted signals), acoustic attenuation, damping (due to reection, absorption, refraction, scatter, and interference) (Fig. 2.5), and other factors, often deviate from the actual properties of the sound beam. Knowledge of such artifacts helps to avoid errors.
The visualization of the contrast agent signals is based on an interaction between the emitted ultrasound wave and the microbubbles, which depends on the equipment settings (acoustic power, image rate, focal zone, etc.). Wrong equip­ment settings are often the reason for CEUS artifacts that can result in uncertain diagnoses or even misdiagnoses in extreme cases.
Perhaps the most important CEUS artifact is bubble destruction (Fig. 2.6). The MI plays a crucial role here. It balances the signal intensity and penetration on the one hand and the stability of the microbubbles on the other hand. The CEUS “circle of disaster” is characterized by the following criteria: microbubble destruction increase in contrast agent dose attenuation (shadowing) higher mechanical index additional microbubble destruction. The secret is to nd a good balance between the contrast agent dose and the equipment- specic settings.
Pseudoenhancement arises from nonlinear artifacts occurring in FLL that appear echogenic on conventional B mode ultrasound and eventually deep in location (Fig. 2.7). The presence of oscillating microbubbles in vascularized tissue between the transducer and the object of interest may create nonlinear echoes that can give the appearance of enhancement of a deep lesion relative to
2 Contrast Enhanced Ultrasound: How toPerform It inLiver Tumors?
21
Fig. 2.5 Mirror Image Artifact. Contrast enhanced ultrasound (CEUS) of the liver showed a mass in the posterior right lobe with peripheral hyperenhancement adjacent to the inferior vena cava (IVC). A mirror
image of the lesion and IVC was opposite the interface with the dia­phragm and lung base
Fig. 2.6 Near eld bubble burn-off artifact. A horizontal stripe of low signal in the near eld due to inhomogeneous microbubble destruction (bubble burn-off)
22
C. F. Dietrich et al.
Fig. 2.7 Contrast ultrasound enhancement (CEUS) of a hepatocellular carcinoma lesion immediately after transarterial chemotherapy and radiofrequency ablation. CEUS prior to contrast administration showed
background tissue. This pseudoenhancement typically occurs in the late portal venous phase and progresses over time in distinction to real enhancement, which always ini­tiates within the arterial phase [26]. This nonlinear propa­gation of the ultrasound beam increases with bubble concentration.

2.6.1 Long Liver Enhancement

Prolonged innocuous liver enhancement has been very rarely observed over the past decade after the bolus injection of microbubble contrast agents. It appears as a heterogeneous enhancement in the liver most often observed during the per­formance of the CEUS examination and often around 2min and lasting up to 5h after contrast injection on both B-mode and contrast-specic modes (Figs. 2.8 and 2.9). It is not destroyed by sonication at high MI. The enhanced signals can also be observed in the portal and superior mesenteric veins, though not in the systemic circulation. It is very simi-
the lipiodol deposition inside the lesion devoid of signal with a few echogenic foci
Fig. 2.8 Prolonged heterogeneous liver enhancement. B mode ultra­sound through the liver following contrast administration (SonoVue) showed patchy, heterogenous areas of increased echogenicity. This appearance may last for several hours. However, it is likely not clini­cally signicant and should not be confused for pathology
2 Contrast Enhanced Ultrasound: How toPerform It inLiver Tumors?
Fig. 2.9 Prolonged heterogeneous liver enhancement. B mode ultra­sound through the liver following contrast administration (Sonazoid) showed wavey, heterogenous areas of increased echogenicity. This appearance may last for several hours. However, it is likely not clini­cally signicant and should not be confused for pathology
lar to US ndings of free portal venous gas that have been observed in end-stage oncological (e.g., gastrointestinal and urological) diseases within the last hours to days before death, in severe enterocolitis in newborns and adults as well as in some asymptomatic patients [27].

2.7 Safety

Ultrasound contrast agents are safe with a very rare inci­dence of adverse events. Laboratory checks to assess thyroid, liver, or renal function before administration is not necessary since there are no hepato-, cardio-, or nephrotoxic effects. The reported incidence of severe side effects is lower than that with current CT contrast agents and is comparable to those with MR contrast agents [6]. It has been reported that life-threatening anaphylactic reactions in abdominal applica­tions had a rate of 0.001%, with no death in a series of more than 23,000 abdominal patients [28].

References

1. Sidhu PS, Cantisani V, Dietrich CF, Gilja OH, Saftoiu A, Bartels E, Bertolotto M, et al. The EFSUMB guidelines and recommen­dations for the clinical practice of Contrast-Enhanced Ultrasound (CEUS) in non-hepatic applications: update 2017 (short version). Ultraschall Med. 2018;39:154–80.
2. Sidhu PS, Cantisani V, Dietrich CF, Gilja OH, Saftoiu A, Bartels E, Bertolotto M, etal. The EFSUMB guidelines and recommendations for the clinical practice of Contrast-Enhanced Ultrasound (CEUS) in non-hepatic applications: update 2017 (long version). Ultraschall Med. 2018;39:e2–e44.
3. Claudon M, Dietrich CF, Choi BI, Cosgrove DO, Kudo M, Nolsoe CP, Piscaglia F, etal. Guidelines and good clinical practice rec-
23
ommendations for Contrast Enhanced Ultrasound (CEUS) in the liver– update 2012. Ultraschall Med. 2013;34:11–29.
4. Claudon M, Dietrich CF, Choi BI, Cosgrove DO, Kudo M, Nolsoe CP, Piscaglia F, etal. Guidelines and good clinical practice rec­ommendations for Contrast Enhanced Ultrasound (CEUS) in the liver– update 2012: a WFUMB-EFSUMB initiative in coopera­tion with representatives of AFSUMB, AIUM, ASUM, FLAUS and ICUS.Ultrasound Med Biol. 2013;39:187–210.
5. Dietrich CF, Ignee A, Greis C, Cui XW, Schreiber-Dietrich DG, Hocke M.Artifacts and pitfalls in contrast-enhanced ultrasound of the liver. Ultraschall Med. 2014;35:108–25. quiz 126-107
6. Dietrich CF, Averkiou M, Nielsen MB, Barr RG, Burns PN, Calliada F, Cantisani V, et al. How to perform Contrast-Enhanced Ultrasound (CEUS). Ultrasound Int Open. 2018;4:E2–E15.
7. Itabashi T, Sasaki A, Otsuka K, Kimura T, Nitta H, Wakabayashi G. Potential value of sonazoid-enhanced intraoperative laparo­scopic ultrasonography for liver assessment during laparoscopy­assisted colectomy. Surg Today. 2014;44:696–701.
8. Muhi A, Ichikawa T, Motosugi U, Sou H, Nakajima H, Sano K, Sano M, etal. Diagnosis of colorectal hepatic metastases: compari­son of contrast-enhanced CT, contrast-enhanced US, superpara­magnetic iron oxide-enhanced MRI, and gadoxetic acid-enhanced MRI.J Magn Reson Imaging. 2011;34:326–35.
9. Cantisani V, Ricci P, Erturk M, Pagliara E, Drudi F, Calliada F, Mortele K, etal. Detection of hepatic metastases from colorectal cancer: prospective evaluation of gray scale US versus SonoVue(R) low mechanical index real time-enhanced US as compared with multidetector-CT or Gd-BOPTA-MRI. Ultraschall Med. 2010;31:500–5.
10. Larsen LP, Rosenkilde M, Christensen H, Bang N, Bolvig L, Christiansen T, Laurberg S. Can contrast-enhanced ultrasonog­raphy replace multidetector-computed tomography in the detec­tion of liver metastases from colorectal cancer? Eur J Radiol. 2009;69:308–13.
11. Piscaglia F, Corradi F, Mancini M, Giangregorio F, Tamberi S, Ugolini G, Cola B, etal. Real time contrast enhanced ultrasonog­raphy in detection of liver metastases from gastrointestinal cancer. BMC Cancer. 2007;7:171.
12. Konopke R, Kersting S, Bergert H, Bloomenthal A, Gastmeier J, Saeger HD, Bunk A.Contrast-enhanced ultrasonography to detect liver metastases: a prospective trial to compare transcutaneous unenhanced and contrast-enhanced ultrasonography in patients undergoing laparotomy. Int J Colorectal Dis. 2007;22:201–7.
13. Larsen LP, Rosenkilde M, Christensen H, Bang N, Bolvig L, Christiansen T, Laurberg S.The value of contrast enhanced ultraso­nography in detection of liver metastases from colorectal cancer: a prospective double-blinded study. Eur J Radiol. 2007;62:302–7.
14. Dietrich CF, Kratzer W, Strobe D, Danse E, Fessl R, Bunk A, Vossas U, etal. Assessment of metastatic liver disease in patients with primary extrahepatic tumors by contrast-enhanced sonography versus CT and MRI.World J Gastroenterol. 2006;12:1699–705.
15. Konopke R, Kersting S, Saeger HD, Bunk A. Detection of liver lesions by contrast-enhanced ultrasound– comparison to intraop­erative ndings. Ultraschall Med. 2005;26:107–13.
16. Quaia E, D’Onofrio M, Palumbo A, Rossi S, Bruni S, Cova M.Comparison of contrast-enhanced ultrasonography versus base­line ultrasound and contrast-enhanced computed tomography in metastatic disease of the liver: diagnostic performance and con­dence. Eur Radiol. 2006;16:1599–609.
17. Forner A, Vilana R, Ayuso C, Bianchi L, Sole M, Ayuso JR, Boix L, etal. Diagnosis of hepatic nodules 20mm or smaller in cirrho­sis: prospective validation of the noninvasive diagnostic criteria for hepatocellular carcinoma. Hepatology. 2008;47:97–104.
18. Dong Y, Zhang XL, Mao F, Huang BJ, Si Q, Wang WP.Contrast­enhanced ultrasound features of histologically proven small (</=20 mm) liver metastases. Scand J Gastroenterol. 2017;52:23–8.
24
C. F. Dietrich et al.
19. Dietrich CF, Mertens JC, Braden B, Schuessler G, Ott M, Ignee A. Contrast-enhanced ultrasound of histologically proven liver hemangiomas. Hepatology. 2007;45:1139–45.
20. Dietrich CF, Schuessler G, Trojan J, Fellbaum C, Ignee A. Differentiation of focal nodular hyperplasia and hepatocel­lular adenoma by contrast-enhanced ultrasound. Br J Radiol. 2005;78:704–7.
21. Dietrich CF, Tannapfel A, Jang HJ, Kim TK, Burns PN, Dong Y. Ultrasound imaging of hepatocellular adenoma using the new histology classication. Ultrasound Med Biol. 2019;45:1–10.
22. Sidhu PS, Cantisani V, Deganello A, Dietrich CF, Duran C, Franke D, Harkanyi Z, et al. Role of Contrast-Enhanced Ultrasound (CEUS) in paediatric practice: an EFSUMB position statement. Ultraschall Med. 2017;38:33–43.
23. Chiorean L, Cui XW, Tannapfel A, Franke D, Stenzel M, Kosiak W, Schreiber-Dietrich D, et al. Benign liver tumors in pediatric patients – Review with emphasis on imaging features. World J Gastroenterol. 2015;21:8541–61.
24. Lyshchik A, Kono Y, Dietrich CF, Jang HJ, Kim TK, Piscaglia F, Vezeridis A, etal. Contrast-enhanced ultrasound of the liver: tech-
nical and lexicon recommendations from the ACR CEUS LI-RADS working group. Abdom Radiol (NY). 2018;43:861–79.
25. Wang JY, Feng SY, Xu JW, Li J, Chu L, Cui XW, Dietrich CF. Usefulness of the contrast-enhanced ultrasound liver imaging reporting and data system in diagnosing focal liver lesions by inex­perienced radiologists. J Ultrasound Med. 2020;39:1537–46.
26. Yu H, Jang HJ, Kim TK, Khalili K, Williams R, Lueck G, Hudson J, etal. Pseudoenhancement within the local ablation zone of hepatic tumors due to a nonlinear artifact on contrast-enhanced ultrasound. AJR Am J Roentgenol. 2010;194:653–9.
27. Cui XW, Ignee A, Hocke M, Seitz K, Schrade G, Dietrich CF. Prolonged heterogeneous liver enhancement on contrast­enhanced ultrasound. Ultraschall Med. 2014;35:246–52.
28. Piscaglia F, Bolondi L, Italian Society for Ultrasound in M, Biology Study Group on Ultrasound Contrast A.The safety of Sonovue in abdominal applications: retrospective analysis of 23188 investiga­tions. Ultrasound Med Biol. 2006;32:1369–75.
Improved Detection ofFocal Liver Lesions withContrast Enhanced Ultrasound
Bei-JianHuang, YiDong, andWen-PingWang
3
Abbreviations
CEUS Contrast enhanced ultrasound CT Computed tomography FLL Focal liver lesion HCC Hepatocellular carcinoma IOUS Intraoperative ultrasonography IO-CEUS Intraoperative contrast-enhanced ultrasound MRI Magnetic resonance image

3.1 Introduction

• Conventional ultrasound is the most frequently used imaging modality as the rst-line imaging of abdominal organs, including the liver, but is reported to be less sensi­tive than CECT, CEMRI, or intraoperative ultrasound in the detection of focal liver lesions (FLL).
• CEUS has dramatically increased the capability of con­ventional ultrasound for detection of FLL, especially those invisible on conventional ultrasound.
• CEUS has a considerably higher sensitivity of up to 80–90% in detecting liver metastases, comparable to that of CECT and CEMRI [1].
• CEUS is of particular useful in detecting liver metastases 10mm.
3.2 Reasons forFocal Liver Lesions Not Detected onCEUS
• Despite of its high detection rate and diagnostic accuracy
in FLLs, CEUS still faces the challenges in the detection of some indistinctive lesions, especially when the diame­ter of lesion is less than 10mm.
• In the background of liver cirrhosis, some early or recurrent
HCCs may be isoechogenic with indistinctive margins, or show hyperechoic similar with those of cirrhosis nodules.
• After molecular targeted therapy for colorectal liver
metastasis, some lesions’ volume may shrink and become isoechoic. These lesions are so-called occult tumors since they cannot be detected on CEUS or even on CECT.
• In the condition of sever fatty liver, hemotherapy-induced
steatohepatitis, hepatic sinus obstruction, liver structure changes after repeated surgical procedures and local abla­tive treatment or residual or recurrent lesions located adjacent to treatment area, FLLs might be difcult to be detected by ultrasound.
• Specic tumor location, such as too deep or close to dia-
phragm, subcapsular tumors affected by rib occlusion or abdominal wall reverberation, may result in difculties in tumors detection.
3.3 Detection ofLiver Primary Malignancies
• CEUS have improved detection and characterization of
HCC.Homogeneous hyperenhancement during the arte­rial phase and mild wash-out are indicative for HCC in liver cirrhosis.
• The incident rate of recurrent HCCs ranging from 45.2%
B.-J. Huang (*) · Y. Dong · W.-P. Wang Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai, China e-mail: huang.beijian@zs-hospital.sh.cn;
dong.yi@zs-hospital.sh.cn
© The Author(s), under exclusive license to Springer Nature Singapore Pte Ltd. 2021 W.-P. Wang et al. (eds.), Contrast-Enhanced Ultrasound Imaging of Hepatic Neoplasms,
https://doi.org/10.1007/978-981-16-1761-4_3
to 60.0% after HCC hepatectomy. The CEUS enhance­ment pattern of recurrent HCCs including hyper- or iso­enhancement during arterial phase, with no wash-out in portal or late phases (Fig.3.1).
25