Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5760_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
89 Мб
Скачать
6.1 · Abdominal Aorta, Visceral andRenal Arteries
D2
orthogonal
D1
axial
D1 > D2
425
. Fig. 6.31 Sources of variation in measuring the maximum abdomi-
nal aortic aneurysm (AAA) diameter in a patient with an elongated aorta and anterolateral deviation of its course. A maximum anteroposterior (AP) diameter of 62.6mm (D1) is measured in the transverse abdominal view (right image), while the maximum orthogonal diameter measured perpendicular to the long axis of the aorta is 45.5mm (left image). At the site of the AP diameter measurement in the transverse view, the orthogonal diameter is 39.7mm (D3). The AP diameter is represented by the white line, the orthogonal diameter by the yellow line. The AP diam­eter measured in the right image corresponds to the CT-based diameter measurement in the axial plane (i.e., in the plane used for measuring AAA size when CT is performed without multiplanar reconstruction)
6
. Fig. 6.30 Pitfall in measuring abdominal aortic aneurysm (AAA)
diameter. Patients with an aortic aneurysm often have an elongated aorta with an outward curve to the left. When transverse images are obtained with the transducer in the normal abdominal position, this can lead to overestimation of the diameter because the aorta is being measured in an elliptical plane (D1). Overestimation of aneurysm size by measurement in the wrong scan plane can also lead to overestimation of the rupture risk. The correct diameter of the aneurysm is measured by rotating the transducer clockwise until a round image of the aorta comes into view (D2). With the transducer in this position, the true transverse diameter is measured orthogonally (perpendicular to the vessel axis)
ultrasound, this error can be avoided by rotating the trans­ducer from the transverse abdominal position until it is per­pendicular to the long aortic axis (conrmed when the oval shape of the aorta becomes more rounded) (
. Figs.6.28, 6.29,
and 6.30). When the elongated aorta deviates anteriorly, the diameter is best measured in the sagittal plane (. Fig.6.31).
Without this standardized approach, it is not possible to take accurate and reproducible serial measurements of aortic diameter. is must be taken into account when interpreting discrepant results obtained with the same modality or with dierent modalities.
Studies show good intraobserver and interobserver agreement for sonographic measurement of the orthogonal AAA diameter, which is a prerequisite for obtaining mean­ingful results in serial measurements. Such a standardized approach is necessary to capture true size increases over time and avoid errors resulting from poor methodology (Sun 2006; AbuRahma 2006; Collins etal. 2007; Stavropoulos and Charagundla 2007).
Once again, adequate aortic diameter measurement requires clockwise rotation of the transducer from the posi­tion showing the largest aortic diameter to the view allowing diameter measurement perpendicular to the long aortic axis (change from elliptical to rounded shape of the aortic cross­section) (
. Fig.6.30).
In summary, the following procedure is recommended for the sonographic evaluation of the aorta and reliable AAA diameter measurement:
5 Identify the aorta in the transverse plane and image its
course from the suprarenal segment to the bifurcation
5 Identify the renal artery origins and aortic bifurcation to
determine AAA extent in relation to these structures
5 Evaluate the entire aorta for possible elongation: may
require transverse, oblique, or even longitudinal trans-
ducer positions
5 Identify the widest (axial) aortic diameter
5 From this position, rotate the transducer for measure-
ment of AAA diameter perpendicular to the vascular axis
(rounded rather than elliptical shape of the aortic cross-sec-
tional area; may require a longitudinal view in some cases)
5 Use the leading-edge method to measure AAA diameter 5 Document ndings and measurements
New techniques such as 3D ultrasound and ultrasound/ CT image fusion have the potential to improve the accuracy of aneurysm size determination in the future, especially in serial examinations of patients with AAA (Bredahl et al. 2013; Pster 2014).
426
Chapter 6 · Visceral andRetroperitoneal Vessels
6.1.6.3.4 Comparison ofUltrasound
andComputed Tomography
No consistent picture emerges from studies comparing ultra­sound and CT (Singh et al. 2004); discrepancies are oen attributable to the study design (Beales etal. 2011). Investiga-
conditions. In view of the problems discussed here, there appears to be an urgent need to standardize measurements and ensure that we measure AAA diameter accurately and reproducibly in serial examinations. A consensus should specify how, where, and when to measure AAA diameter.
tors oen fail to describe details of the ultrasound technique used to determine abdominal aortic aneurysm (AAA) size (e.g., transducer position, plane in which diameter is mea­sured). In a study reviewing the methodology of maximum AAA diameter measurement (Long etal. 2012), only 40% of the studies included (n= 23) specied the plane of acqui­sition and 30% the caliper positions, and only 10% of the
6
studies used the leading-edge method. Long etal. dened a quality score for the description of methodology in the stud­ies reviewed and found a mean quality score of 2.5 (of a total of 4). Surprisingly, they found an even lower mean quality score of 1.6 for the description of methodology in guidelines for screening programs. Most of the guidelines they reviewed did not specify the axis of measurement or caliper positions (outer or inner diameter) (Long et al. 2012; Lederle et al.
1995), and these issues are rarely discussed (Moll etal. 2011). Overall, the review of Long et al. conrms that a range of dierent methods are in use for measuring AAA diameter.
Studies comparing ultrasound and CT show good
agreement of the two modalities (correlation coecient of
0.91) (Manning etal. 2009) with the majority of investiga­tors concluding that ultrasound underestimates maximum aneurysm diameter compared with CT (mostly measured in AP plane) (Jaakkola et al. 1996, Sprouse et al. 2003, Manning etal. 2009, Long et al. 2012). Conversely, other authors point out that axial CT scans, without orthogonal reformation, oen measure the diameter in oblique slices of the AAA, which tend to overestimate aneurysm diameter when there is relevant elongation of the aorta. is is clearly apparent from two studies performed by Sprouse etal. (2003 and 2004). In the earlier study, CT yielded larger diameters than ultrasound in 95% of cases. Specically, AAA diam­eters by CT were 5.69±0.89cm versus 4.74±0.91 cm by ultrasound (and the dierence was signicant, p<0.05). In the later study, Sprouse etal. (2004) found good agreement between ultrasound and CT diameter measurements with a mean dierence of only 0.8mm when CT measurements were taken in reformatted slices allowing true orthogonal measurement. Comparison of axial and orthogonal AAA
6.1.6.3.5 Abdominal Aortic Aneurysm
Screening: Rupture Risk
Accurate and reproducible (orthogonal) measurement of aortic diameter is essential for preventing both excessive diagnostic testing (stressful for patients) and excessive treat­ment (surgical risk, complications, and postoperative mor­bidity) in screening and surveillance programs (. Figs.6.36 and 6.37). Abdominal aortic aneurysm (AAA) is a non- malignant condition and it is therefore even more impor­tant not to put patients at risk through poor measurement methodology and operating on aneurysms that would never rupture during their lifetime. In addition, the low risk of rupture of <3–5%/year for AAA <5.5cm must be weighed against the surgical risk and perioperative morbidity. Patients undergoing endovascular aneurysm repair (EVAR) also face several risks (endoleaks in up to 10% of cases, risk of stent gra limb occlusion of up to 5%) and require regular follow-up examinations with radiation and contrast medium exposure.
Clinically, AAA rupture is suggested by ank and back pain (occasionally abdominal pain), a palpable pulsating tumor, and shock. Gray-scale imaging depicts a hypoechoic structure of variable extent and comprising inhomogeneous or layered portions around the aorta in the retroperitoneum (
. Fig. 6.75b (Atlas)). In patients with a contained aneu-
ryms rupture, color duplex ultrasound demonstrates para­vascular ow signals at the site of the leak. Leakage must be dierentiated from other hypoechoic periaortic structures such as retroperitoneal brosis, horseshoe kidneys, or lym­phoma, which may occur in conjunction with an aneurysm (see 7 Sect. 6.1.6.3.8; . Figs.6.38, 6.39, and 6.40; . Figs.6.85,
6.86,
and 6.87 (Atlas);). An examiner performing ultraso­nography in emergency patients must pay special attention to such accompanying conditions as they have important implications.
If there is may appear uid-lled. Fistula connections to the vena cava can be demonstrated by color duplex.
perforation into the duodenum, the intestine
diameter measurements by CT revealed signicantly larger mean diameters when measurements were taken in axial slices compared with orthogonal slices (58 mm versus
54.7mm, p<0.05). e overestimation of axially measured diameters increased with aortic angulation. ese results clearly illustrate that selection of the correct slice for AAA diameter measurement aects ultrasound (. Figs.6.29 and
6.30) and CT alike. Awareness of this pitfall is important
both in clinical routine and in the setting of clinical studies (Long etal. 2012).
CT is generally considered the gold standard for measur­ing AAA diameter because it is not examiner-dependent and less susceptible to errors resulting from poor examination
6.1.6.3.6 Aortic Dissection
Aortic dissection is only amenable to percutaneous ultra­sound diagnosis if the intimal ap extends into the abdominal aorta. In the abdominal aorta, sonography is a valid method for assessing the extent of dissection, and spectral Doppler evaluation is helpful in identifying extension into arteries arising from the aorta or intermittent obstruction of blood ow at the origin of these arteries by the intimal ap moving synchronously with the heart (see . Fig.6.88 (Atlas)).
In a dissecting aneurysm, splitting of the arterial wall with tearing of the intima is suggested on gray-scale ultra­sonography by the presence of a ap in the vessel lumen;
2
ab
6.1 · Abdominal Aorta, Visceral andRenal Arteries
this ap can be identied by its hyperechoic reection and typical undulating motion. Dissection is conrmed in the color mode by dierent ow velocities and directions in the true and false lumens (. Fig. 6.33). e power mode and contrast-enhanced ultrasound (CEUS) will help demonstrate slow ow in the false lumen and dierentiate it from partial thrombosis. e color-coded ow directions contribute to the identication of the entry and re-entry sites. Knowledge of the relationships of the origins of the visceral and renal arteries to the true and false lumens determines the thera­peutic management. As the dissected ap may extend into the groin, the iliac arteries must be included in the examina­tion (see . Figs.6.88, 6.89, and 6.90 (Atlas)).
Spectral Doppler imaging at the sites of aortic branch ori-
gins for identifying extension of the dissection into renal or
visceral arteries
as well as intermittent occlusion or stenosis of a branch artery by the intimal ap (dynamic blood ow reduction) is important for the therapeutic approach. If there is extension into an aortic branch, the Doppler waveform will depict ow in the true and false lumens and may also contain signals produced by oscillation of the intimal ap. If there is narrowing or intermittent occlusion of a branch origin by the intimal ap, spectral Doppler will reveal signs of stenosis or systolic deceleration (decreased systolic velocity or even zero ow) (
. Fig.6.88 (Atlas)). Dissection with intermittent ow
obstruction can lead to chronic ischemia and patients are at risk of acute ischemic events. Dynamic ow obstruction is dicult to detect with a morphologic imaging modality (CT, angiography) when the lumen is lled with contrast medium.
Conversely, with isolated dissection of the abdominal aorta, which is rare (Knabe etal. 2001), the proximal origin is typically not detectable by transcutaneous sonography (except with the transducer directed retrosternally from a jugular position). In these cases, transesophageal echocar­diography (Link 1999), computed tomography, or magnetic resonance imaging is required for diagnosis. e diagnostic accuracies in aortic dissection are 70% for ultrasound (Nien­aber etal. 1993), 98% for transesophageal echocardiography (Sommer etal. 1996), and 100% for MRI (Silverman 2000) and CT.
6.1.6.3.7 Follow-Up After Open Surgical
andEndovascular Aneurysm Repair
Diagnostic Algorithm
z
Follow-up (see . Fig.6.32) aer open surgical repair of an abdominal aortic aneurysm (AAA) or patch angioplasty of aortic stenosis must ensure early identication of complica­tions such as suture aneurysm, anastomotic stenosis, recur­rent stenosis, or abscess.
Hypoechoic structures around a prosthesis, particularly at the sites of anastomosis, can be checked for the presence of ow using color duplex ultrasound to dierentiate suture aneurysm from postoperative hematoma and abscess. Suture aneurysms are pseudoaneurysms and hence are character­ized by to-and-fro ow (steam engine sound) in the Doppler waveform from the site of wall perforation (see . Fig.6.92 (Atlas)). An abscess suspected on clinical grounds can be
427
4
1
3
. Fig. 6.32a, b Complications of interventional and open surgi-
cal repair of abdominal aortic aneurysm (AAA). a Following surgical Y-prosthesis implantation: 1 suture aneurysm (typically at the upper anastomosis or, if the prosthesis extends into the femoral artery, at the lower anastomosis); 2 occlusion of an iliac limb; 3 anastomotic stenosis (distal anastomosis). b Types of endoleaks after endovascular aneurysm repair (EVAR): 1 type I endoleak, at the proximal or distal attachment site; 2 type II endoleak, from patent lumbar arteries; 3 type III endoleak, separation of modular components; 4 type IV endoleak, device failure (porosity)
2
3
1
conrmed by ultrasound-guided ne-needle aspiration biopsy.
Aer open surgical aneurysm repair, sonographic follow­up has sucient validity to identify typical complications such as suture aneurysm (
. Fig.6.34), anastomotic stenosis,
and iliac limb occlusion (in patients with a Y-stent gra).
e following complications of open surgical and endo­vascular AAA repair have therapeutic implications and
require special attention in sonographic follow-up
:
5 Aer open surgical repair:
5 Recurrent aneurysm– suture aneurysm 5 Anastomotic stenosis/iliac limb occlusion 5 Abscess/infection
5 Aer endovascular aneruysm repair (EVAR):
5 Endoleaks (types I, II, III) 5 Further sac growth 5 Stent gra migration/fracture (domain of conven-
tional X-ray)
5 Iliac limb occlusion 5 rombotic deposits as a source of embolism or
cause of luminal narrowing
Aer EVAR, patients should undergo imaging surveil­lance at 6-month intervals. AAA shrinkage indicates ade­quacy of the repair and rules out an endoleak (Giannoni etal. 1998; ompson etal. 1998). Conversely, even a small further increase in aortic diameter (B-mode) points to a therapeutically relevant endoleak (with increasing pressure
6
428
Chapter 6 · Visceral andRetroperitoneal Vessels
6
. Fig. 6.33 Aortic dissection of longer duration with sclerotic thickening of the dissection membrane (same patient as in . Fig.6.27). In this
patient, gray-scale imaging already shows the oating membrane (indicated by “d” in the color ow images) in the lumen and rules out intermit­tent obstruction of blood ow into arterial branches at their origins from the aorta. The waveforms shown along with the color ow images were obtained in the true lumen (wl) and false lumen () (center and right, respectively). The bright artifact apparent in both waveforms in early systole is due to oscillation of the dissection membrane. The membrane is also apparent in the corresponding CTA scan
. Fig. 6.34 Suture aneurysm
after open surgical stent graft repair for abdominal aortic aneurysm (AAA). The arrow in the color ow images (transverse ori­entation on the left, longitudinal orientation on the right) indicates the site of leakage. The CT scan was obtained for stent graft siz­ing prior to interventional closure of the leak
in the aneurysm sac) and should prompt a careful search for the site of leakage using color duplex ultrasound (CDUS). If no endoleak is detected, the search should proceed using contrast-enhanced ultrasound (CEUS) or computed tomog­raphy angiography (CTA).
Serial ultrasound follow-up should be supplemented by an annual plain X-ray examination of the abdomen to rule out stent fracture, which can be identied sonographically only if an endoleak is presdent (type IV).
In summary, the following algorithm is proposed for the follow-up of patients aer EVAR (in line with the general policy of stepwise diagnostic workup advocated throughout this book):
1. B-mode ultrasound (follow-up at 6-month intervals):
5 Development of AAA diameter over time
5 Diameter decreases → continue routine follow-up 5 Diameter remains constant or increases → search
for endoleak, successively using CDUS, CEUS, and CTA (as required)
CDUS (PRF, gain): indicated if B-mode measurement shows
2. constant or increasing sac diameter (orthogonal plane)
5 Type of endoleak 5 Types I and III: treatment 5 Type II (treatment required?):
5 Low-ow → follow-up (possibly at shorter intervals
of 3months)
5 High-ow → reintervention
3. CEUS: indicated if CDUS fails to identify an endoleak despite increasing AAA diameter
5 Same diagnostic accuracy as CTA 5 Superior in detecting small low-ow endoleaks (late
retrograde blood ow into aneurysm sac via patent lumbar artery)
4. CTA (gold standard): indicated to search for suspected endoleak (B-mode ndings) not detected by CDUS or CEUS
e basic idea of this stepwise approach is that the next test following in the recommended sequence of diagnostic
6.1 · Abdominal Aorta, Visceral andRenal Arteries
429
6
procedures should only be performed if it is expected to pro­vide therapeutically relevant information.
In addition, a systematic procedure is recommended to ensure reliable identication of endoleaks and other compli­cations aer EVAR (. Fig.6.32):
5 Gray-scale examination in transverse orientation to
evaluate the upper stent end and its relationship to the
renal artery origins.
5 Gray-scale measurement of the largest orthogonal diam-
eter of the residual aneurysm sac (see . Figs.6.77a and
6.82 (both Atlas)).
5 Transverse CDUS (low pulse repetition frequency) of the
aneurysm site and the stented segment from the renal
artery origins to the bifurcation, focusing on the origins
of the lumbar arteries and the inferior mesenteric artery
(. Figs.6.78 and 6.80 (both Atlas)).
5 Longitudinal CDUS evaluation of the stent ends with
spectral Doppler measurement to conrm patency, dem-
onstrate stenosis, and identify type I endoleaks using an
acute Doppler angle at the anchoring sites (
(Atlas)).
5 If CDUS reveals ow in the aneurysm sac, this must
be conrmed by spectral Doppler interrogation. is
is especially important in the early postinterventional
phase before complete thrombosis of the aneurysm has
occurred and movement of the stent can mimic ow
signals (pseudoendoleak). Such pseudoendoleaks are
dierentiated from true endoleaks by the demonstration
of to-and-fro ow in the spectral Doppler waveform
(especially with the sample volume at the site of leakage,
e.g., patent lumbar artery entering the aneurysm sac).
is ow pattern is characteristic of endoleaks, which
resemble pseudoaneurysms in terms of hemodynamics
(see . Fig.6.80 (Atlas)). Mirror artifacts can be ruled out
by insonation from dierent directions, supplemented
by waveform information.
5 Patients with type II endoleaks not requiring reinter-
vention (patent lateral branches, lumbar arteries, inferior
mesenteric artery) can be managed by shortening the
follow-up interval (see
(failure of xation) and type III endoleaks are reliably
detected by ultrasound.
Whether the ultrasound examination allows adequate evalu­ation aer AAA repair strongly depends on the individual acoustic window. Published data on the reliability of ultra­sound are inconsistent, especially with regard to the iden­tication of type II endoleaks (Ashoke etal. 2005; Sanford etal. 2006). Some authors showed CDUS to be sucient to rule out endoleaks with reported sensitivities of 77–96% and specicities of 90–94% (D’Audiret 2001; Golzarian et al. 2002; Sato etal. 1998; Sun 2006; AbuRahma 2006; Collins etal. 2007; Stavropoulos and Charagundla 2007); however, the patient populations investigated were small. e di­culty in detecting an endoleak is that it requires not only a good acoustic window and adequate machine settings (low
. Fig.6.32). Higher-ow type I
. Fig.6.79
pulse repetition frequency) but also great care in setting the Doppler angle correction cursor relative to the leak jet so as not to miss a type II endoleak. Moreover, motion and mirror artifacts in the excluded aneurysm sac can mimic ow in the color duplex examination, which must be dierentiated from an endoleak by spectral Doppler interrogation (for technical details and optimization of settings see
Therapeutic Relevance of CDUS-derived
z
Hemodynamic Information
Despite its inherent methodological limitations, color duplex ultrasound allows reliable evaluation for endoleaks aer EVAR. Doppler waveforms contribute therapeutically rele­vant supplementary hemodynamic information not provided by CEUS or CTA (Schäberle etal. 2014).
e diagnostic questions to be answered by the sonographic examination of patients with a suspected endoleak can be summarized as follows:
5 Is an endoleak present? 5 If yes, which type (type I, II, III, or IV or a combined
endoleak)?
5 If a type II endoleak is present, is it fed by the inferior
mesenteric artery or by a patent lumbar artery?
5 How should the endoleak be managed?
5 Follow-up only 5 If treatment is required, how urgent is the reinter-
vention?
Ȥ elective Ȥ urgent Ȥ emergency intervention.
kCombined Endoleaks
Type I endoleaks develop in the early postinterventional period and tend to occur in conjunction with other endole­aks, commonly type II.A second endoleak can relieve the intrasac pressure buildup that would otherwise result from blood entering the sac through a type I endoelak. CDUS pro­vides information on blood ow velocity and other ow char­acteristics at the entry point, which is relevant for estimating the acute risk of rupture associated with a type I endoleak. Like a pseudoaneurysm, a type I endoleak is characterized by to-and-fro ow with blood entering the sac during sys­tole and leaving it during diastole. A monophasic waveform showing ow into the aneurysm sac but little or no backward ow indicates a high acute rupture risk or that the blood leaves the sac through a second, paradoxical endoleak, which may be termed an exoleak. Such a constellation is suggested by the demonstration of untypical orthograde ow in the lumbar artery or the inferior mesenteric artery draining the aneurysm through the endoleak (see (both Atlas)). A Doppler waveform from the inferior mes­enteric artery or origin of a patent lumbar artery provides the necessary hemodynamic information (ow character and direction) to fully capture the ow situation. is informa­tion is essential for estimating the risk of rupture and to plan an individual reintervention strategy.
7 Sect. 6.1.2.1.2).
. Figs. 6.78 and 6.79
430
Chapter 6 · Visceral andRetroperitoneal Vessels
6
. Fig. 6.35a, b Low-ow type II endoleak. a Flow in the former aneurysm sac fed by a patent right-sided lumbar artery (arrow) displayed in
red (ow toward transducer) along with blue-coded ow in the stent graft (both iliac limbs). The waveform shows to-and-fro ow of very low frequency at the entry site, consistent with slow ow in the thin-caliber lumbar arteries and hence low ow volumes entering the aneurysm sac. These hemodynamic features suggest that there is no risk of rupture and that the endoleak is likely to close spontaneously. In addition, the anteroposterior aneurysm diameter has decreased from 5.7 to 5.1cm. With this constellation of ndings, no immediate reintervention is neces­sary. Instead, duplex ultrasound follow-up at 3-month intervals is indicated. b Angiogram with selective probing of the lumbar artery conrms a low-ow endoleak and blood ow in a small portion of the aneurysm sac. An endoleak with these features can be managed by watchful waiting. In this patient, coils were placed in the same angiography session
kHigh-ow and Low-ow Type II Endoleaks
e Doppler waveform from the endoleak jet provides relevant information on endoleak hemodynamics (see
. Figs.6.77, 6.78, 6.79, 6.80, and 6.81 (all Atlas)). A low-ow
endoleak fed by a patent lumbar artery can be managed by watchful waiting as long as the aneurysm sac does not expand (see . Figs.6.81 and 6.82 (both Atlas). Because these type II endoleaks tend to close spontaneously, the risk of rup­ture is low (spontaneous thrombosis in up to 50% of type II endoleaks according to Carter etal. (2000)). However, vali­dated criteria for the identication of low-ow endoleaks are not available for any imaging modality (Liewald etal. 2001; Parry etal. 2002; White etal. 2000). Time-intensity curves derived from CEUS allow quantitative analysis of endoleak ow dynamics. Again, validated data on the therapeutic rel­evance are not available. Methodologically, at least time-to­peak curves are required, while the sum of intensities alone is not sucient.
hemodynamic parameters derived from Doppler
Two
waveforms
obtained in the endoleak may be helpful in pre­dicting the risk of rupture. On the one hand, the waveform shape provides information on resistance to blood ow. When a single endoleak is present, to-and-fro ow should predominate (due to the variation in pressure through the cardiac cycle, resulting in systolic ow into the aneurysm sac and diastolic ow back into the feeding artery). Consistent with this assumption, a higher spontaneous thrombosis rate was reported for endoleaks with bidirectional ow compared with endoleaks showing predominantly systolic inow (simi­lar to the ow prole of peripheral arteries) and little diastolic backward ow (Carter etal. 2000; Parent etal. 2002). Surpris­ingly, a more recent study identied bidirectional Doppler ow to be associated with a higher rate of sac growth (Bee­man etal. 2010). is observation is counterintuitive because
one would normally expect further aneurysm growth when inow through an endoleak is higher than outow (i.e., the waveform is monophasic), unless the blood leaves the sac via another route– an exoleak (see
. Fig.6.79 (Atlas)). An aneu-
rysm with blood entering the sac through one endoleak and leaving it through another is less likely to close spontaneously than a sac with bidirectional ow through a single endoleak. Correct placement of the sample volume for spectral Dop­pler interrogation is essential to capture blood ow direc­tions at the endoleak site (patent lumbar artery or inferior mesenteric artery) (see . Figs.6.36 and 6.82 (Atlas)). Else- where in a perfused residual aneurysm sac, eddy ow with circulatory movement of blood (like in a pseudoaneurysm) may give rise to a monophasic waveform and misinterpreta­tion of endoleak hemodynamics.
In addition, spectral Doppler can be used to measure ow velocity at the blood entry site (patent lumbar artery, inferior mesenteric artery) and thus estimate the ow vol­ume. A high systolic velocity suggests a large endoleak vol­ume or highly dynamic ow situation (
. Figs.6.35, 6.36, 6.81
(Atlas), and 6.82 (Atlas)). A study investigating intrasac ow velocities proposed a Doppler cuto >80 cm/s to identify endoleaks not expected to seal spontaneously (Arko 2003). In the author’s experience, this cuto is rather high and a cuto on the order of 50–40cm/s appears to better identify endoleaks requiring repair (. Fig.6.35b). However, Beeman et al. (2010) conclude that ow velocity in the aneurysm sac is not a relevant predictor of sac growth. Again, it is important to ensure that measurement is performed at the site of entry of a patent feeder into the aneurysm sac and with adequate Doppler angle correction (. Fig.6.36a). With a meticulous technique, the time-averaged velocity (inten­sity-weighted) during systolic inow and diastolic outow should be identical.
6.1 · Abdominal Aorta, Visceral andRenal Arteries
431
6
. Fig. 6.36a–j Sonographic endoleak characterization. a Endoleak in the upper portion of a stent graft following endovascular aneurysm
repair (EVAR) for abdominal aortic aneurysm (AAA). There is blue-coded ow in the former aneurysm sac adjacent to red-coded ow in the stent graft (S). To determine whether a type I or type II endoleak is present, a Doppler waveform is obtained from the origin of the inferior mesenteric artery (AMI), which shows high-frequency bidirectional ow with ow into the aneurysm sac during systole (S) and a PSV of 100cm. These nd­ings are consistent with a high-ow endoleak. b To-and-fro ow with a PSV of 105cm/s is conrmed along the course of the inferior mesenteric artery (AMI; systolic ow (S) toward the excluded aneurysm (toward transducer) and diastolic ow (D) away from transducer). While the waveform represents ow over time and thus captures its bidirectional nature, the color ow image selectively shows diastolic ow (blue-coded ow, away from transducer) and thus suggests orthograde ow. c Hemodynamic evaluation at an arbitrary site in the aneurysm rather than at the blood entry site is unsuitable for endoleak characterization. With the sample volume for spectral Doppler measurement placed away from the entry site (as done here for illustration), the PSV is 49cm/s and ow is in one direction, which is usually due to circular ow in the sac. d Waveform from a site adjacent to the Doppler sampling site in c: PSV of only 25cm/s but to-and-fro ow. e An additional endoleak is identied between the 2 iliac limbs of the stent graft, through which blood enters the inferior portion of the aneurysm sac (PSV of 1m/s). Flow through this endoleak is unidi­rectional into the aneurysm sac with the blood taking a meandering course in the sac and leaving through the inferior mesenteric artery.
f In the contrast-enhanced ultrasound (CEUS) examination, the large endoleak (EL) from the inferior mesenteric artery can only be suspected. g The second endoleak, fed by a patent lumbar artery between the iliac limbs, is visible in the same location as in the color ow image shown in e. This endoleak must be dierentiated from the bright wall reection (WR). h This CEUS image was obtained more inferiorly and is shown here
to underline the importance of always interpreting contrast-enhanced images along with the corresponding unenhanced gray-scale images in order not to mistake bright spots for endoleaks. When such spots are present without and with contrast enhancement (indicated by arrowheads), they do not represent ow and hence do not suggest an endoleak. i The CTA examination in this patient demonstrates an endoleak with contrast medium extravasation (arrow) in the upper portion of the stent graft without allowing clear identication of the source of the endoleak. j Contrast medium extravasation into the aneurysm sac (arrow) is also seen more distally, and there appears to be a meandering communication between the upper and lower stent portion. The CTA appearance does not allow clear identication of the number of endoleaks present in this patient
To-and-fro ow is not physiologic and only occurs in endoleaks and pseudoaneurysms. A bidirectional waveform thus dierentiates a true endoleak from mirror or pulsation artifacts.
Another sonographic parameter that can help the exam­iner in identifying patients requiring reintervention aer EVAR is the (anteroposterior) aneurysm diameter in the
time-motion mode. A pulsatile diameter variation of the aneurysm sac indicates an endoleak that should be treated (see . Fig.6.81 (Atlas)).
To the best of our knowledge, no published studies have investigated the possible diagnostic role of pulsation-related diameter variation of the residual aneurysm sac. In a small pilot study of 18 patients with type II endoleaks aer EVAR
432
Chapter 6 · Visceral andRetroperitoneal Vessels
conducted by the author, all 6 high-ow endoleaks showed a diameter variation of >2mm through the cardiac cycle in the time-motion mode. Four of the 6 high-ow endoleaks led to sac growth within 6months and were treated, while the other two showed a diameter increase >4mm aer another 3 months. Nine low-ow endoleaks showed no relevant pulsation (i.e., <2mm variation through the cardiac cycle) and no sac growth over the next 6months (measurement tolerance ±2mm). Four of these 9 endoleaks closed spon­taneously. However, 3 low-ow endoleaks showed relevant pulsation of 2–3mm. Pulsation of the residual aneurysm sac is typically not observed when no endoleak is present.
e diagnostic and therapeutic relevance of sac pulsation
6
in the time-motion mode might depend on the type of stent gra (structure of the prosthesis) used for EVAR.In this pilot study, all patients were treated with Endurant (Medtronic) or Zenith (Cook Medical) stent gras.
Atlas)). ese endoleaks, which may rupture, have higher ow velocities at the site of leakage and are detected with sensitivities and specicities of 95–100% (Karthikesalingam etal. 2012). Overall, most published data suggest that CDUS has adequate diagnostic accuracy for detection of type I and III endoleaks and identication of therapeutically relevant type II endoleaks aer EVAR.An occasional patient requires supplementary CEUS or CTA to resolve inconclusive CDUS ndings. While complications of CEUS are extremely rare, it must be borne in mind that it is an invasive examination and should only be performed by qualied examiners.
While a variety of diagnostic parameters exist, sac size
remains the most important criterion in deciding about the
management of patients with type II endoleaks after EVAR
(surveillance versus intervention). Shrinkage of the residual aneurysm sac rules out a relevant endoleak, and the ques­tion regarding the best imaging modality becomes negligible. Conversely, when the sac expands, this points to a relevant
kPreinterventional Identication of the Feeding Artery
When interventional closure of a type II endoleak is planned, it is helpful to know whether a patent lumbar artery or the inferior mesenteric artery is the culprit and whether it is a right-sided or le-sided lumbar artery. e side is dicult to identify using CEUS, and CTA at most provides some indi­rect clues. A Doppler waveform from the blood entry site with demonstration of to-and-fro ow allows identication of the side and also dierentiation between an inferior mesenteric artery endoleak and a lumbar artery endoleak. A waveform showing unidirectional rather than bidirectional ow sug­gests that a second endoleak functioning as an exoleak is present and should be searched for (see . Fig.6.79 (Atlas)).
endoleak, which must be identied using the full armen­tarium of imaging modalities. Since aneurysm diameter measurement is subject to some uncertainty, especially when insonation conditions are poor, an unchanged orthogonal diameter±0.5cm should prompt a CDUS evaluation to rule out an endoleak. Patients with inconclusive CDUS ndings should have a CEUS examination next. If the diagnosis still remains unclear, CTA is indicated.
Multiphase CTA is considered the gold standard for fol-
low-up aer EVAR.However, caution is required regarding the use of potentially nephrotoxic contrast agents in these patients, who are typically elderly and have multiple comor­bidities including renal insuciency. CDUS, which was ini­tially used as an alternative to CTA in the follow-up of EVAR
Follow-Up of EVAR: CDUS Versus CEUS and CTA
z
While studies report adequate validity for color duplex
ultrasound (CDUS)
in identifying type II endoleaks that require intervention, contrast-enhanced ultrasound (CEUS) appears to be superior, especially in the detection of small endoleaks and in identifying endoleaks in patients with poor insonation conditions, and has even been reported to be comparable to computed tomography angiography (CTA) (Karthikesalingam etal. 2012). However, in all cases where CDUS identies an endoleak that is conrmed by to-and-fro ow in the Doppler waveform, CEUS or CTA is not necessary before interventional endoleak closure (Carter et al. 2000; Chaer etal. 2009). Misinterpretation of mirror or pulsation artifacts in color ow imaging can be avoided by moving the transducer around to view the suspected endoleak from a dierent angle. Demonstration of to-and-fro ow conrms an endoleak, ruling out mirror or pulsation artifacts. When an endoleak has been identied by CTA or CEUS, CDUS can contribute therapeutically relevant information (see above). Spectral Doppler (possibly with echo enhancer) is also indis­pensable to identify steno-occlusive lesions at vessel origins in patients with a branched stent gra.
Studies consistently conrm that, with adequate
insonation conditions, CDUS allows sucient identica-
type I and III endoleaks (. Figs.6.78 and 6.79 (both
tion of
(. Figs.6.35 and 6.36), shows discrepant results in terms of sensitivity and specicity compared with CTA. However, many investigators conclude CDUS follow-up to be sucient or even equal to CTA, reporting sensitivities of 80–100% and specicities of 74–100% (Beeman et al. 2009; Chaer et al. 2009; Schmieder et al. 2009; Mirza et al. 2010; Sato etal. 1998; Wolf etal. 2000; Zannetti etal. 2000; Parent etal. 2002; McLaerty etal. 2002; ompson etal. 1998; Fletcher etal.
2000). A review of the literature reports a mean sensitivity of 95% and mean specicity of 97% (McLaerty etal. 2002). Some investigators also point out advantages of CDUS over CTA such as the real-time evaluation of sac hemodynamics. is improves the detection of small lumbar artery endoleaks compared with CTA, which only displays the hemodynamic situation at individual points in time (e.g., venous phase). Other investigators reported inadequate diagnostic accuracy of CDUS for endoleak detection aer EVAR (AbuRahma 2006; Schuster et al. 2009). A more recent meta-analysis including 25 studies with a total of 3975 paired scans found a pooled sensitivity of 0.74 and a pooled specicity of 0.96 for all types of endoleaks (predominantly type II endoleaks) compared with CTA (Karthikesalingam etal. 2012).
Several studies conrm that the sonographic detection and localization of endoleaks (especially of the more dif­cult to detect type II endoleaks) can be improved by the
6.1 · Abdominal Aorta, Visceral andRenal Arteries
433
. Fig. 6.37a–c Follow-up of a patient with a branched stent graft. a Ultrasound shows occlusion of the right iliac limb. The left iliac limb is pat-
ent but there is localized luminal narrowing (arrow) at its origin due to thrombus formation. b Thrombus formation (T) in the main stent graft body extending to the iliac limb origins (S). c Stenosis of the superior mesenteric artery (A.MES.S) at the end of the branched graft segment. The intrastenotic peak systolic velocity (PSV) is 350cm/s (see sample volume) with a prestenotic PSV of 125cm/s (not shown). The PSV ratio calculated from these velocities is 3, consistent with 60–70% stenosis. (The main stent graft body in the aorta (A) is patent; the absence of color ow signals is due to the obtuse angle of insonation and high PRF)
6
administration contrast microbubbles (Böhm et al. 2000; Henao et al. 2006; McWilliams et al. 2002; Bendick et al. 2003; Heilberger etal. 1997; Clevert etal. 2008; Sarlon etal. 2009; Giannoni etal. 2007).
CEUS in Endoleak Detection
z
Following injection of the microbubble contrast agent (for protocol details see 7 Sect. 6.1.2.1.2), the examiner begins by searching for type I and III endoleaks at the stent gra ends (where the microbubbles arrive rst). Type II endoleaks, which are fed by patent lumbar arteries or the inferior mes­enteric artery, enhance later due to a longer transit time of the contrast microbubbles (and enhancement of these arter­ies may even persist into the venous phase). Following the search for endoleaks (for 2–3min aer microbubble admin­istration), the next 2–5min with persisting enhancement can be exploited to search for steno-occlusive lesions of the renal and mesenteric artery origins using color duplex utlrasound, which is especially important in patients with branched stent gras.
In patients with an initially inconclusive contrast­enhanced scan, a second microbubble injection can be given to then focus on sites with suspicious or inconclusive nd­ings aer the rst contrast bolus. If the results remain incon­clusive, CTA should follow.
Published data consistently show CEUS to be comparable to CTA in detecting endoleaks and even superior in char­acterizing them (Iezzi etal. 2010; Pster etal. 2009; Mirza et al. 2009). A meta-analysis of 11 studies with 981 paired scans comparing CEUS with CTA yielded a pooled sensitiv­ity of 0.96 and a pooled specicity of 0.85 with follow-up and angiography oen showing CEUS to be more accurate than CTA (Karthikesalingam etal. 2012). A small type II endoleak fed by retrograde ow in a patent lumbar artery enhances relatively late (longer transit time of microbubbles through long collateral pathways), and contrast medium extravasa­tion may be missed, even during the venous phase of CT.e dynamic real-time CEUS examination (. Fig.6.36) is more exible and thus better able to detect such late-enhancing endoleaks (Jung etal. 2008; Pster etal. 2009).
In a study investigating contrast harmonic imaging (CHI)
ultrasound (see
7 Sect. 1.1.5) in 50 patients with suspected
endoleaks following EVAR, CHI and CTA concordantly detected endoleaks in 30 cases and found no endoleak in 20 cases. In one patient, a combined type I/II endoleak was mis­classied as type II.In another patient, a type II endoleak was initially only detected by CHI and later conrmed by follow­up CTA.us, CHI had 99% sensitivity, 93% specicity, 99% negative predictive value, and 95% positive predictive value. Time-intensity curve (TIC) analysis was used in this study to evaluate perfusion of the aneurysm (Pster etal. 2009). Stan­dardized signal analysis aer echo enhancer administration can be used to compare enhancement in the stent gra and in the aneurysm sac; the intensity of enhancement varies with the amount of blood and can thus serve to quantify the endoleak, identifying at-risk aneurysms that warrant treatment.
Other Complications After EVAR
z
Ultrasound is clearly inferior to radiological imaging modali­ties in the detection of stent fracture and stent migration. Color duplex imaging will identify these complications only if they cause an endoleak (. Fig.6.84 (Atlas)). is is why sur­veillance programs for patients aer EVAR should include an annual plain X-ray examination of the site of the former AAA.
Further complications aer EVAR are occlusion of a stent
gra limb with peripheral ischemia (
. Fig.6.37a) and throm-
bus development in the main stent gra body, which may be a source of peripheral embolism (. Fig.6.37b), especially when the thrombus extends to the attachment site of a stent gra limb. Patients with a complex stent gra are addition­ally at risk of organic ischemia due to stenosis (. Fig.6.37c) or occlusion of the branched artery (renal artery, superior mesenteric artery).
6.1.6.3.8 Aortitis: Retroperitoneal
Fibrosis– Inammatory Abdominal Aortic Aneurysm
Inammation of the aorta is rare compared with atherosclero­sis. Aortitis may occur with or without dilatation and is com­plicated by obstruction, rupture and dissection. Underlying
434
Chapter 6 · Visceral andRetroperitoneal Vessels
causes include endocarditis, local perivascular foci, and sep­ticemia. e aorta is the most common site of bacterial infec­tion. e dierentiation of bacterial and nonbacterial aortitis
Normal
Normal
VC
Ao
can be dicult, dilatation is possible, and patients are at risk of aortic or aneurysm rupture (. Fig.6.91 (Atlas)). Develop-
Vertebra
ment of an eccentric or saccular aneurysm (mycotic aneu­rysm) is observed very early in the course of infection, and
Aortitis
infectious aortitis is rarely detected before an aneurysm has
Aortitis
formed (Narang and Rathlev 2007; Caspary 2016).
e aorta is involved in several large-artery vascular dis­eases including giant cell arteritis (. Fig6.40b) and Takaya­su’s arteritis as well as Behçet’s disease and Cogan’s syndrome.
Ormond’s
Ormond’s
Infl. aneurysm
Infl. aneurysm
Aortic involvement is typically asyomptomatic or presents
6
with chest or back pain. Morphologic ultrasound ndings include concentric wall thickening and sometimes dilatation of the aorta. Magnetic resonance angiography (MRA) and contrast-enhanced ultrasound (CEUS) may show increased aortic wall perfusion as a sign of local inammation. e risk of aneurysm rupture or dissection is higher in patients with underlying vasculitis than in patients with atherosclerotic aneurysm.
Especially in Behçet’s disease, dilatation is mostly eccen­tric, the increase in diameter can progress rapidly, and other complications such as concomitant supercial phlebitis or deep vein thrombosis may be present.
Occasionally, aortic inammation is found in association with rheumatoid arthritis, systemic lupus erythematosus, sarcoidosis or inammatory bowel diseases (Caspary 2016).
Isolated aortitis (possibly IgG4-related disease) is found in chronic periaortitis, retroperitoneal brosis (Ormond’s disease), and in inammatory aortic aneurysm. Inamma­tory aortic aneurysm (
. Fig. 6.39) may sometimes develop
from degenerative aortic aneurysm in the presence of peri­vascular inammation (Ketha et al. 2014) and aer stent gra repair. In Ormond’s disease (idiopathic in 70–80% of cases), inammation extends beyond the periaortic area (. Fig6.40a) and spreads within the retroperitoneal tissue.
Aortic wall thickening can be dierentiated from perivas­cular, retroperitoneal conditions in the B-mode examination by assessing the course of arterial branches arising from the aorta. Retroperitoneal lymphomas are circumscribed round lesions but may become conuent. Retroperitoneal brosis
. Fig. 6.38 Evaluation of the course of the proximal inferior mes-
enteric artery as it arises from the aorta for the dierentiation of aortic wall thickening and retroperitoneal perivascular pathology (see . Fig.6.85 (Atlas)). The topmost diagram illustrates the normal anatomy of the aorta (Ao) and vena cava (VC). The diagram labeled “Aortitis” illustrates the situation in giant cell arteritis: the inferior mesenteric artery arises from the aorta and then courses outside the hypoechoic, concentrically thickened wall of the aorta. In this condi­tion, the mesenteric artery takes the shortest path possible through the abnormal tissue (inammatory aortic wall). The diagram labeled “In. aneurysm” illustrates the ndings in inammatory abdominal aortic aneurysm (AAA). In this condition, the inferior mesenteric artery also passes through the aneurysmatically dilated and thickened abdominal wall, taking the shortest path possible, to then descend toward the left lower abdomen outside the thickened structure. If ath­erosclerotic lesions are present, they can help in dierentiating inam­matory wall thickening from thrombotic deposits in the aneurysm sac: atherosclerotic plaques make the intima appear bright, thus allowing dierentiation of thrombotic deposits extending into the lumen from inammatory wall thickening on the other side of the intima. The bottom diagram (labeled “Ormond’s”) illustrates the ndings in retro­peritoneal brosis. The proximal segment of the inferior mesenteric artery arising from the nonthickened aortic wall courses through the hypoechoic abnormal tissue before emerging out of it and passing to the left lower abdomen. In retroperitoneal brosis, the abnormal tissue pushes a longer proximal segment of the inferior mesenteric artery against the wall of the aorta. In patients with more advanced retroperi­toneal brosis, which also encases the vena cava or even the ureter, the diagnosis and dierentiation from other conditions are easier than in early disease with brotic tissue surrounding only the aorta. Therefore, the abnormal course of the proximal inferior mesenteric artery is the decisive criterion for making the dierential diagnosis (Diagrams cour­tesy of K.Amendt)
is typically depicted as a more hypoechoic structure covering the aorta from anteriorly and involving the vena cava. It tapers o laterally and may also surround and compress the ureter.
Specically, the course of the inferior mesenteric artery can help in dierentiating wall thickening of the aorta and retroperitoneal brosis (
. Figs.6.38 and 6.40). In retroperi-
toneal brosis, the proximal inferior mesenteric artery is pushed against the aorta and thus courses between the aortic wall and the hypoechoic brosis for several centimeters before piercing through the hypoechoic brotic cap (see . Fig.6.85 (Atlas)). In inammation of the aorta (e.g., giant cell arteri­tis), on the other hand, the inferior mesenteric artery pierces the hypoechoic layer around the patent lumen of the aorta directly at its origin to then descend into the le lower abdo­men outside the hypoechoic thickening. Inammatory wall
thickening in aortitis and in inammatory aortic aneurysm is circumferential and spares the vena cava (see
6.40). e exible selection of scanning planes in the ultra-
. Figs.6.39 and
sound examination facilitates determination of the relation­ships between the aorta, hypoechoic thickening, perivascular structures, and other vessels, thereby contributing to the dif­ferential diagnosis.
ickening of the aortic wall due to vasculitis with aortic dilatation or inammatory aortic aneurysm can be dierenti­ated from intraluminal thrombi by its relationship to the inti­mal layer. While this is relevant for diameter measurement, dierentiation is only possible when atherosclerotic plaque is present, which makes the intima appear brighter. e normal intima is not visible sonographically (. Figs.6.38 and 6.39).