DICOM PS3.17 2020a - Explanatory Information |
Page 811 |
|
Attribute |
Tag |
Value |
Comment |
Columns |
(0028,0011) |
32 |
This case is a non-square |
|
|
|
image |
Bits Allocated |
(0028,0100) |
32 |
|
Pixel Aspect Ratio |
(0028,0034) |
1\1 |
|
Float Pixel Padding Value |
(0028,0122) |
-200 |
|
Float Pixel Padding Range Limit |
(0028,0124) |
-100 |
|
Float Pixel Data |
(7FE0,0008) |
(44 times) -150, -0.1356, 1.317, (28Forconvenience","isused |
|
|
|
times) -150, -0.986, 0.4402, -0.0251,toseparatethevalues,type |
|
|
|
0.6077, 0.2982, 0.0872, 2.6927, |
is OF |
|
|
2.1434, -0.5543, -0.3014, -150 etc. |
|
Table BBBB.2-2. Example data for the Dimension Organization Module
Attribute |
Tag |
Value |
Comment |
Dimension Organization Sequence |
(0020,9221) |
|
|
>Dimension Organization UID |
(0020,9164) |
1.2.3.5.6 |
Sample UID |
Dimension Organization Type |
(0020,9311) |
3D |
|
Dimension Index Sequence |
(0020,9222) |
1 |
|
Item 1 |
First Item describing Stack ID |
|
|
>Dimension Index Pointer |
(0020,9165) |
(0020,9056) |
|
>Functional Group Pointer |
(0020,9167) |
(0020,9111) |
|
>Dimension Organization UID |
(0020,9164) |
1.2.3.5.6 |
|
>Dimension Description Label |
(0020,9421) |
Stack ID |
|
Item 2 |
Second Item describing In-Stack Position Number |
|
|
>Dimension Index Pointer |
(0020,9165) |
(0020,9057) |
|
>Functional Group Pointer |
(0020,9167) |
(0020,9111) |
|
>Dimension Organization UID |
(0020,9164) |
1.2.3.5.6 |
|
>Dimension Description Label |
(0020,9421) |
In-Stack Position Number |
|
Table BBBB.2-3. Example data for the Pixel Measures Macro
Attribute |
Tag |
Value |
Comment |
Pixel Measures Sequence |
(0028,9110) |
1 |
|
>Pixel Spacing |
(0028,0030) |
3.75\3.75 |
|
>Slice Thickness |
(0018,0050) |
5 |
|
Table BBBB.2-4. Example data for the Frame Content Macro
Attribute |
Tag |
Value |
Comment |
Frame Content Sequence |
(0020,9111) |
|
|
>Frame Acquisition Number |
(0020,9156) |
1 |
|
>Dimension Index Values |
(0020,9157) |
1\15 |
|
>Stack ID |
(0020,9056) |
1 |
|
>In-Stack Position Number |
(0020,9057) |
15 |
|
>Frame Comments |
(0020,9158) |
... |
|
>Frame Label |
(0020,9453) |
... |
|
- Standard -

Page 812 DICOM PS3.17 2020a - Explanatory Information
Table BBBB.2-5. Example data for the Identity Pixel Value Transformation Macro
Attribute |
Tag |
Value |
Comment |
Pixel Value Transformation Sequence |
(0028,9145) |
|
Single item with fixed values |
>Rescale Intercept |
(0028,1052) |
0 |
|
>Rescale Slope |
(0028,1053) |
1 |
|
>Rescale Type |
(0028,1054) |
US |
|
Table BBBB.2-6. Example data for the Frame VOI LUT With LUT Macro
Attribute |
Tag |
Value |
Comment |
Frame VOI LUT Sequence |
(0028,9132) |
|
|
>Window Center |
(0028,1050) |
0 |
|
>Window Width |
(0028,1051) |
50 |
Covering -25 to 25 |
Table BBBB.2-7. Example data for the Real World Value Mapping Macro
Attribute |
Tag |
Value |
Comment |
Real World Value Mapping Sequence |
(0040,9096) |
|
|
>Double Float Real World Value First Value Mapped |
(0040,9214) |
-16.739 |
|
>Double Float Real World Value Last Value Mapped |
(0040,9213) |
21.434 |
|
>Real World Value Intercept |
(0040,9224) |
0 |
Identity |
|
|
|
transformation |
>Real World Value Slope |
(0040,9225) |
1 |
Identity |
|
|
|
transformation |
>Measurement Units Code Sequence |
(0040,08EA) |
|
|
>>IncludeTable8.8-1"CodeSequenceMacroAttributes" |
|
(UCUM, {t}, "t") |
|
>Quantity Definition Sequence |
(0040,9220) |
|
|
>>Include Table 10-2 "Content Item Macro Attributes |
|
(113068, DCM, |
|
Description" |
|
"Student's T-test") |
|
The Palette Color Lookup Table used is the Spring Color Palette (see Figure BBBB.1-3 Resulting Color LUT Spring).
This can be described as follows through the Palette Color Lookup Table:
Red has a constant value of 255
Green has a linear segment that starts at 0 and ends at 255
Blue has a linear segment that starts at 255 and ends at 0
Using the Segmented Color Lookup Table all three can be described by a discrete segment with length 1 to specify the starting value (0,1,value) followed by a linear segment of length 255 with the end-value (1,255,end-value).
Table BBBB.2-8. Example data for the Palette Color Lookup Table Module
Attribute |
Tag |
Value |
Comment |
Red Palette Color Lookup Table Descriptor |
(0028,1101) |
256\0\8 |
|
GreenPaletteColorLookupTableDescriptor |
(0028,1102) |
256\0\8 |
|
Blue Palette Color Lookup Table Descriptor |
(0028,1103) |
256\0\8 |
|
Segmented Red Palette Color Lookup Table |
(0028,1221) |
0,1,255,1,255,255 For convenience "," is used to |
|
Data |
|
|
separate the values, type is OW |
- Standard -

DICOM PS3.17 2020a - Explanatory Information |
Page 813 |
||
Attribute |
Tag |
Value |
Comment |
Segmented Green Palette Color Lookup |
(0028,1222) |
0,1,0,1,255,255 |
For convenience "," is used to |
Table Data |
|
|
separate the values, type is OW |
Segmented Blue Palette Color Lookup Table |
(0028,1223) |
0,1,255,1,255,0 |
For convenience "," is used to |
Data |
|
|
separate the values, type is OW |
300 |
|
LUT Spring |
|
|
|
|
|
|
|
|
|
||
250 |
|
|
|
|
|
|
200 |
|
|
|
|
|
|
|
|
|
|
|
|
R |
150 |
|
|
|
|
|
G |
|
|
|
|
|
|
B |
100 |
|
|
|
|
|
|
50 |
|
|
|
|
|
|
0 |
|
|
|
|
|
|
0 |
50 |
100 |
150 |
200 |
250 |
300 |
Figure BBBB.1-3. Resulting Color LUT Spring
The values specifying the range to be mapped to the Color LUT are given by the Minimum Stored Value Mapped and the Maximum Stored Value mapped.
Table BBBB.2-9. Example data for the Stored Value Color Range Macro
Attribute |
Tag |
Value |
Comment |
Minimum Stored Value Mapped |
(0028,1231) |
-16.739 |
Corresponds to the first LUT entry |
|
|
|
(255,0,255). |
Maximum Stored Value Mapped |
(0028,1232) |
21.434 |
Corresponds to the last LUT entry |
|
|
|
(255,255,0). |
Table BBBB.2-10. Example data for the Parametric Map Frame Type Macro
Attribute |
Tag |
Value |
Comment |
Parametric Map Frame Type Sequence |
(0040,9092) |
|
|
>Frame Type |
(0008,9007) |
DERIVED\PRIMARY\FMRI\T_TEST |
|
- Standard -

Page 814 |
DICOM PS3.17 2020a - Explanatory Information |
- Standard -

DICOM PS3.17 2020a - Explanatory Information |
Page 815 |
CCCC Populating The Simplified Echo
Procedure Report Template (Informative)
This Annex provides guidance to understand and populate the TID 5300 “Simplified Echo Procedure Report” and its sub-templates. For implementers familiar with the TID 5200 “Echocardiography Procedure Report”, which is largely replaced by TID 5300, some re- lationships and differences are also explained.
CCCC.1 Structure Overview
Measurements in this template (except for the Wall Motion Analysis) are collected into one of three containers, each with a specific sub-template and constraints appropriate to the purpose of the container.
•Pre-coordinated Measurements
•Are fully standardized measurements (many taken from the ASE practice guidelines).
•Each has a single pre-coordinated standard code that fully captures the semantics of the measurement.
•The only modifiers permitted are to indicate coordinates where the measurement was taken, provide a brief display label, and indicate which of a set of repeated measurements is the preferred value. Other modifiers are not permitted.
•Post-coordinated Measurements
•Are measurements for which DICOM has not established pre-coordinated codes, but that are performed with enough regularity to merit configuration and capturing the full semantics of the measurement. For example these measurements may include those configured on the Ultrasound System by the vendor or user site. Some of these may be variants of the Pre-coordinated Measure- ments.
•A set of mandatory and conditional modifiers with controlled vocabularies capture the essential semantics in a uniform way.
•A single pre-coordinated code is also provided so that when the same type of measurement is encountered in the future, it is not necessarytoparseandevaluatethefullconstellationofmodifervalues.Sincethismeasurementhasnotbeenfullystandardized, the pre-coordinated code may use a private coding scheme (e.g., from the vendor or user site).
•Adhoc Measurements
•Are non-standardized measurements that do not merit the effort to track or configure all the details necessary to populate the set of modifiers required for a post-coordinated measurement.
•The measurement code describes the elementary property measured.
•Modifiers provide a brief display label and indicate coordinates where the measurement was taken. Other modifiers are not per- mitted.
CCCC.2 Use Cases
CCCC.2.1 Use Case 1: Store and Extract Specific Measurement
TheuserwishestoperformmeasurementsontheUltrasoundSystem,storethemtothePACSandlaterhaveaspecificmeasurement (sayABC)automaticallydisplayedintheoverlayorautomaticallyinsertedintoareportpageonthereviewsystem.Thisdoesnotrequire the receiver to understand any of the semantics of the measurement.
CCCC.2.1.1 Configuration
TheUltrasoundSystemisconfiguredtoencodeaparticularmeasurementusingaspecificpre-coordinatedcode(andcodemeaning).
- Standard -

Page 816 |
DICOM PS3.17 2020a - Explanatory Information |
In the case of measurements from the Core Set, it is a well-known pre-coordinated code (i.e., the code is in CID 12300 “Core Echo Measurements”), the full semantics are well-known and the measurement will be recorded in TID 5301 “Pre-coordinated Echo Measurement” . Likely most, if not all, of the Core Set measurements come pre-configured on the Ultrasound System.
In the case of vendor-specific or site-specific measurements, it is a pre-coordinated code managed by the site or the vendor which is entered and persisted on the Ultrasound System. Since the code is not well-known, the measurement will be recorded in TID 5302 “Post-coordinated Echo Measurement” along with the modifiers describing its semantics.
The receiver (i.e., the PACS display package or the reporting package) is configured to associate the specific pre-coordinated code with a location on the overlay or a slot in the report.
The form of the user interface for these capabilities is up to the implementer.
CCCC.2.1.2 Operation
The user takes measurements on the Ultrasound System, including measurement ABC. All these measurements are recorded in the SimplifiedAdultEchoSRobject.IfmultipleinstancesofmeasurementABCareincluded,oneofthemmaybeflaggedbytheUltrasound System by setting the Selection Status for that instance to the reason it was selected as the preferred value.
The Ultrasound System stores the SR object to the PACS.
ThePACSorthereportingpackageretrievestheSRobjectandscansthecontentslookingformeasurementswiththepre-coordinated code for measurement ABC. If multiple instances are found, the receiver takes the one for which the Selection Status has been set.
The receiver renders the measurement value to the display or report, annotating it with the recorded Units, Code Meaning, and/or Short Label as appropriate.
Note that in this use case the receiver handles the measurement in a mechanical way. As long as the measurement can be unam- biguously identified, the semantics do not need to be understood by the receiver.
CCCC.2.2 Use Case 2: Store and Process Measurements
The user wishes to perform measurements on the Ultrasound System, store them to the PACS and later perform processing of some or all of the measurements on a CVIS (Cardiovascular Information System) or other system. Processing may include incorporating measurements into a database, performing trend analysis, plotting graphs, driving decision support, etc. One measurement taken at end systole may be compared to the "same" measurement that is taken at end diastole, etc. Measurements at the same Finding Site might be collected together.
CCCC.2.2.1 Configuration
As in Use Case 1, the Ultrasound System is configured to encode each measurement using a specific pre-coordinated code (and code meaning).
Again, measurements from the Core Set use a well-known pre-coordinated code and are recorded in TID 5301 “Pre-coordinated Echo Measurement” while vendor-specific or site-specific measurements use locally managed codes and are recorded in TID 5302 “Post-coordinated Echo Measurement” along with the modifiers describing their semantics.
CCCC.2.2.2 Operation
The user again takes measurements on the Ultrasound System which are recorded in the Simplified Adult Echo SR object and if multiple instances of a measurement are included, one of them may be flagged by the Ultrasound System by setting the Selection Status for that instance to the reason it was selected as the preferred value.
The Ultrasound System stores the SR object to the PACS.
ThereceivingdatabaseorprocessingsystemretrievestheSRobjectandparsesthecontents.ThecontentsofTID5301“Pre-coordin- ated Echo Measurement” have known semantics and are processed accordingly.
On first encounter, measurements in TID 5302 “Post-coordinated Echo Measurement” will likely have unfamiliar pre-coordinated codes (since the pre-coordinated code in Row 1 of TID 5302 is not taken from CID 12300 “Core Echo Measurements”, but rather was likely produced by the vendor of the Ultrasound System). Depending on the sophistication of the receiver, parsing the modifiers may
- Standard -

DICOM PS3.17 2020a - Explanatory Information |
Page 817 |
provide sufficient information for the receiver to automatically handle the new measurement. If not, the measurement can be put in an exception queue for a human operator to review the values of the modifiers and decide how the measurement should be handled. In between those two possibilities, the receiver may be able to compare the modifier values of known measurements and provide the operator with a partially categorized measurement.
In any case, once the semantics of the measurement are understood by the receiver, the corresponding pre-coordinated code can be logged so that future encounters with that measurement can be handled in an automated fashion.
The receiver may also make use of the Selection Status values or may database all the provided measurement values or allow the human to select from the provided set.
Note that in this use case the receiver handles the measurements based on the semantics associated with the measurement.
CCCC.3 Differences of Note Between TID_5200 and TID 5300
CCCC.3.1 Report Sections
In TID 5200 “Echocardiography Procedure Report”, containers and headings were used to facilitate the layout of printed/displayed reports by collecting measurements into groups based on concepts like anatomical region. Further, TID 5200 permitted Ultrasound Systems to add new sections freely, TID 5300 “Simplified Echo Procedure Report” does not. Section usage was a source of problem- aticvariabilityforreceiversofTID5200.TID5300constrainsthis.Whensuchgroupingsareuseful,forexamplewhenprintingreports, it makes more sense to configure it in one place (in the receiving database/reporting system) rather than configuring such groupings independently(andpossiblyinconsistently)oneachultrasounddeviceinadepartment.Receiversmaychoosetogroupmeasurements based on Finding Site or some other logic as they see fit. This avoids the problem of trying to keep many Ultrasound Systems in sync. SRobjectsareconsideredacquisitiondata/evidence.IfthefindingsaretranscodedintoCDAreports,sectionswilllikelybeintroduced in the CDA as appropriate.
CCCC.3.2 Finding Observation Type
The Finding Site is the location at which the measurement was taken. While some measurements will be an observation of the structure of the finding site itself, other measurements will be an observation of something like flow, in which case the Finding Site is simplythelocation,nottheactualthingbeingobserved/measured.Toclarifythisdistinction,FindingObservationTypewasintroduced in TID 5302 “Post-coordinated Echo Measurement” . For example, when the measurement is a peak velocity and the Finding Site is a valve, to distinguish between a measurement of the velocity of the blood through the valve, and a measurement of the velocity of thevalvetissue,theFindingObservationTypewouldbesetto"HemodynamicMeasurement"or"BehaviorofFindingSite"respectively.
CCCC.4 Usage Guidance
CCCC.4.1 Finding Site
Modifiers are not permitted on the Finding Site in TID 5302 “Post-coordinated Echo Measurement” since such modifiers resulted in different ways of encoding the same concept. TID 5302 requires the use of a single anatomical code that fully pre-coordinates the location details of the measurement. CID 12305 “Basic Echo Anatomic Sites” has proven to be sufficient to encode the ASE Core Set of measurements. Implementers are strongly recommended to using codes from that list. If there is a truly significant location detail that needs to be captured, e.g., to identify a specific segment of the atrial wall, or a specific leaf of a valve as the location of the measurement, then the implementer may introduce a new code (CID 12305 is extensible) or better yet, new codes can be added to CID 12305 through a DICOM Change Proposal.
CCCC.4.2 Measured Property
ThecodesinCID12304“EchoMeasuredProperties”havealsoproventobesufficienttoencodetheASECoreSetofmeasurements. It is expected that the majority of vendor-specific or site-specific measurements can also be encoded using these properties, but it is understandable that some additional codes may be needed. When introducing new codes, implementers should be careful not to in- troduce elements of the other modifiers, such as Finding Site or Cardiac Cycle Point, into the Measured Property. For example, do not introduce a property for Diastolic Atrial Length to be used for the left and right atria, rather for such a measurement, use Prop- erty=Length, Cardiac Cycle Point=End Diastole and Finding Site=Left Atrium or Right Atrium respectively.
- Standard -

Page 818 |
DICOM PS3.17 2020a - Explanatory Information |
CCCC.4.3 Image View
Implementers may use codes for image views beyond those listed in DCID 12226 “Echocardiography Image View” as needed, but note that Image View is only recorded if it is significant to the interpretation of the measurement. Inclusion of the Image View will likely isolate the measurement from other measurements of the same feature taken in different views.
CCCC.4.4 Cardiac Cycle Point
Note that (111973004, SCT, "Systole") is used here to refer to the entire duration of ventricular systole, while (416430001, SCT, "End Systole") is used to refer to the point in time where the aortic valve closes (or in the case of the right ventricle, the pulmonary valve). Therefore, a Vmax measurement for systole would mean the maximum velocity over the period of systole, and a Vmax measurement for end systole would mean the maximum velocity at the time point of end systole.
CCCC.4.5 Measurement Method
This distinguishes between two measurements that convey the same concept, but are obtained or derived in a different way. As with the Image View, this is only recorded if it is significant to the interpretation of the measurement.
CCCC.4.6 Selection Status
This is used to flag the preferred value when multiple instances of the same measurement are recorded in the SR object. Using this to communicate the value preferred by the operator or the Ultrasound System is very useful for receivers that lack the logic to make a selection themselves. In cases where there is no need or value in sending multiple instances of the same measurement, the issue can be avoided by only sending a single instance of any given measurement in the SR object.
CCCC.4.7 Additional Modifiers
The concept modifiers in the template are sufficient to accurately encode all the best practice echo measurements recommended by the ASE. Although TID 5302 “Post-coordinated Echo Measurement” is extensible and adding new modifiers is not prohibited, the meaning and significance of such new modifiers will generally not be understood by receiving systems, delaying or preventing import of such measurements. Further, adding modifiers that replicate the meaning of an existing modifier is prohibited.
CCCC.5 Example
Nest |
Relationship |
Value Type |
Concept Name |
Example Value |
|
|
CONTAINER |
AdultEchocardiographyProcedureReport |
|
> |
CONTAINS |
CONTAINER |
(125301, DCM, "Pre-coordinated |
|
|
|
|
Measurements") |
|
>> |
CONTAINS |
NUM |
(79969-2, LN, "Interventricular septum |
1.00 (cm,UCUM,"cm") |
|
|
|
diastolic dimension") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"IVSd (2D) " |
>> |
CONTAINS |
NUM |
(79991-6, LN, "Left ventricular ejection |
70.3 (%,UCUM,"%") |
|
|
|
fraction biplane (MOD) ") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LV EF (MOD) " |
>> |
CONTAINS |
NUM |
(79996-5, LN, "Left ventricular end |
118 (ml,UCUM,"ml") |
|
|
|
diastolic volume biplane (MOD) ") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LV EDV (MOD) " |
>> |
CONTAINS |
NUM |
(80001-1,LN,"Leftventricularendsystolic35.0 (ml,UCUM,"ml") |
|
|
|
|
volume biplane (MOD) ") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LV ESV (MOD) " |
>> |
CONTAINS |
NUM |
(80007-8, LN, "Left ventricular internal |
5.00 (cm,UCUM,"cm") |
|
|
|
diastolic dimension - 2D") |
|
>>> |
HAS PROPERTIES |
CODE |
(121404, DCM, "Selection Status") |
(121410, DCM, "User chosen |
|
|
|
|
value") |
- Standard -

|
|
DICOM PS3.17 2020a - Explanatory Information |
Page 819 |
|
Nest |
Relationship |
Value Type |
Concept Name |
Example Value |
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LVIDd (2D) " |
>> |
CONTAINS |
NUM |
(80007-8, LN, "Left ventricular internal |
5.50 (cm,UCUM,"cm") |
|
|
|
diastolic dimension - 2D") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LVIDd (2D) " |
>> |
CONTAINS |
NUM |
(80007-8, LN, "Left ventricular internal |
6.00 (cm,UCUM,"cm") |
|
|
|
diastolic dimension - 2D") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LVIDd (2D) " |
>> |
CONTAINS |
NUM |
(80011-0, LN, "Left ventricular internal |
3.00 (cm,UCUM,"cm") |
|
|
|
systolic dimension - 2D") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LVIDs (2D) " |
>> |
CONTAINS |
NUM |
(80031-8, LN, "Left ventricular posterior1.00 (cm,UCUM,"cm") |
|
|
|
|
wall diastolic thickness") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LVPWd (2D) " |
>> |
CONTAINS |
NUM |
(80068-0, LN, "Mitral valve area |
4.82 (cm2,UCUM,"cm2") |
|
|
|
(Planimetry) ") |
|
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"MV Area (Planim) " |
> |
CONTAINS |
CONTAINER |
(125302, DCM, "Post-coordinated |
|
|
|
|
Measurements") |
|
>> |
CONTAINS |
NUM |
(LVSIMOD,99CompanyName,"Left |
39 (ml/m2,UCUM,"ml/m2") |
|
|
|
Ventricle Stroke Index (MOD) ") |
|
>>> |
HAS CONCEPT MOD |
CODE |
(125306, DCM, "Measurement Type") |
(125313, DCM, "Indexed") |
>>> |
HAS CONCEPT MOD |
CODE |
(363698007, SCT, "Finding Site") |
(87878005, SCT, "Left |
|
|
|
|
Ventricle") |
>>> |
HAS CONCEPT MOD |
CODE |
(125305, DCM, "Finding Observation |
(44324008, SCT, |
|
|
|
Type") |
"Hemodynamic |
|
|
|
|
Measurements") |
>>> |
HAS CONCEPT MOD |
CODE |
(125307, DCM, "Measured Property") |
(90096001, SCT, "Stroke |
|
|
|
|
Volume") |
>>> |
HAS CONCEPT MOD |
CODE |
(370129005, SCT, "Measurement |
(125207, DCM, "Method of |
|
|
|
Method") |
Disks Biplane") |
>>> |
HAS CONCEPT MOD |
CODE |
(399264008, SCT, "Image Mode") |
(399064001,SCT,"2DMode") |
>>> |
HAS CONCEPT MOD |
CODE |
(125308, DCM, "Measurement Divisor")(8277-6, LN, "Body Surface |
|
|
|
|
|
Area") |
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LV SI (MOD) " |
>> |
CONTAINS |
NUM |
(29469-4, LN, "Left Atrium |
3.0 (cm,UCUM,"cm") |
|
|
|
Antero-posterior Systolic Dimension") |
|
>>> |
HAS CONCEPT MOD |
CODE |
(125306, DCM, "Measurement Type") |
(125316, DCM, "Directly |
|
|
|
|
measured") |
>>> |
HAS CONCEPT MOD |
CODE |
(363698007, SCT, "Finding Site") |
(82471001,SCT,"LeftAtrium") |
>>> |
HAS CONCEPT MOD |
CODE |
(125305, DCM, "Finding Observation |
(125311, DCM, "Structure of |
|
|
|
Type") |
the Finding Site") |
>>> |
HAS CONCEPT MOD |
CODE |
(125307, DCM, "Measured Property") |
(81827009, SCT, "Diameter") |
>>> |
HAS CONCEPT MOD |
CODE |
(370129005, SCT, "Measurement |
(122675, DCM, |
|
|
|
Method") |
"Anterior-Posterior") |
>>> |
HAS CONCEPT MOD |
CODE |
(399264008, SCT, "Image Mode") |
(399064001,SCT,"2DMode") |
>>> |
HAS CONCEPT MOD |
CODE |
(272518008, SCT, "Cardiac Cycle Point")(416430001, SCT, "End |
|
|
|
|
|
Systole") |
- Standard -

Page 820 |
|
DICOM PS3.17 2020a - Explanatory Information |
|
|
Nest |
Relationship |
Value Type |
Concept Name |
Example Value |
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"LA Dimen (2D) " |
> |
CONTAINS |
CONTAINER |
(125303, DCM, "Adhoc Measurements") |
|
>> |
CONTAINS |
NUM |
(385673002, SCT, "Interval") |
15.0 (ms,UCUM,"ms") |
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"MV Jet Duration" |
>> |
CONTAINS |
NUM |
(1483009, SCT, "Angle") |
27.0 (deg,UCUM,"deg") |
>>> |
HAS PROPERTIES |
TEXT |
(125309, DCM, "Short Label") |
"MV Leaf Angle" |
- Standard -