Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3860_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
15.09.2026
Размер:
11 Мб
Скачать
☆
14 A. R. Abadir and J. E. Silberzweig
https://t.me/med1917
stiffness and torque-control characteristics. Some cathe­ters incorporate a fine wire mesh (braid) in the wall to increase torque transmission to the tip.
Catheters are characterized by material, shape, outer diameter measured in French (1 mm ⫽ 3F), presence and number of side holes, length, maximal wire diameter accepted, and pressure rating. The most commonly used diagnostic catheters are 4Fr and 5Fr. The size influences the properties of the catheter, such as handling, maximal flow rate, and size of the hole made in the artery. Diag­nostic catheters (3Fr) and coaxial microcatheters (2Fr to 3Fr), designed to fit into standard sized angiographic catheters or larger-diameter “guiding catheters,” are used to access tortuous distal vessel branches or minimize ac­cess puncture size.
20,21
The catheter is commonly divided into three parts: hub, shaft, and tip. The hub is on the trailing end and is where syringes and injectors can be attached and wires passed. The shaft provides length to the catheter to en­able it to reach the target vessel. Catheter lengths are typically 65 cm for abdominal and pelvic arteriography and 90 to 100 cm for thoracic and carotid arteriography. The most common site for a catheter to fail during a power injection is at the junction of the hub and shaft because intraluminal pressure is highest there. Fortu­nately, this is external to the patient. The tip of the catheter is tapered to fit the guidewire and facilitate atraumatic percutaneous insertion.
Shape
Perhaps the most defining element of a catheter is its shape, as determined by the curves at its distal aspect. Some basic, all-purpose shapes are used in almost every case. Conversely, many catheters have been designed spe­cifically to work for individual vessels.
High-volume, high-flow-rate studies in large vessels are best performed with pigtail catheters. The leading end of such a catheter is looped in a circle, and there are side holes on the distal straight shaft to evenly disperse con­trast during injection. The pigtail shape resists catheter recoil, causing vessel injury, during high-flow contrast injection. (Pigtail shapes also are often used for drainage catheters because the distal circle helps prevent migra­tion of the catheter. The side holes in drainage catheters are larger and are in the inner aspect of the pigtail curve to maintain patency as the drained space surrounding the catheter collapses. Additionally, during catheter ex­change over a guidewire, the guidewire is less likely to exit a side hole located on the inner curvature of the loop.)
To catheterize a vessel selectively, the tip must be curved or angled in such a way as to engage the origin of the vessel of interest. Such catheters can be simple or complex, depending on the number of curves they con­tain. The primary curve is the curve closest to the cathe-
ter tip. Additional curves beyond the primary curve (sec­ondary, tertiary) help to force the catheter tip deeper into the selected vessel once its origin has been engaged by contacting the opposite aortic wall. Additional curves also may help to stabilize the catheter against the aortic wall during contrast injection.
Simple shapes have only one or two curves, such as the hockeystick-shaped catheter (e.g., Berenstein, Kumpe). Some arteries, such as the celiac axis or internal iliac artery, have steeply angled origins and may be difficult to catheterize. Two solutions to this situation are to ap­proach these arteries superiorly (from an axillary or bra­chial artery puncture) or to use a catheter that has a downward pointing tip. A downward pointing catheter tip from a femoral approach can be achieved by either inverting the curve of a traditional catheter on itself (e.g., making a Waltman loop from a braided cobra catheter) or by using a subset of complex catheters known as recur- vant catheters. A Waltman loop can be created by catheter­izing a branch vessel of the aorta, such as the renal artery. A guidewire then is positioned within the catheter so that the guidewire tip is at the origin of the branch vessel. Advancing the catheter and guidewire as a unit will form a loop in the aorta, drawing the tip out of the branch vessel. The tip is now directed inferiorly, ready for place­ment in a steeply angled vessel.
22
Recurvant catheters have a trailing curve that turns the catheter 180 degrees back on itself, and the leading curve (primary) serves to engage the orifice of the vessel. Exam­ples of these are the Simmons and Sos-Selective catheters (Fig. 2-4). The trailing curve must be reformed in a large vessel, such as the thoracic aorta or aortic bifurcation. Although a recurvant catheter may be used for abdomi-
FIGURE 2-4. Recurvant selective catheters. Simmons 1, Sim­mons 2, Simmons 3, Sos selective (
left to right
).
nal visceral catherization, the catheter shaft must be long
https://t.me/med1917
enough to reach the thoracic aorta to reform its trailing curve. The tip of the reformed catheter is brought supe­rior to the origin of the intended vessel and drawn slowly down until the tip engages. An interesting point to note about looped catheters and recurvant shapes is that catheter manipulations now have the opposite effect that they normally do, inserting the catheter at the groin withdraws the tip from the vessel and pulling on it ad­vances it further into the vessel.
Recurvant catheters need to be disengaged from their selective position before removal; other wise, the vessel origin may be avulsed or dissected. In addition, recurvant catheters should have their curves opened by a guidewire before being removed from the patient.
Side holes
Catheters may have side holes in addition to the end hole to provide a larger surface area for contrast to exit the catheter (or for fluid to enter the catheter in the case of drainage catheters). A higher rate of contrast injection thus may be achieved (the taper of a catheter substan­tially reduces the diameter of the end hole compared with the remainder of the catheter lumen). In addition, a high-velocity jet, which potentially may damage the vessel, is not formed during forceful contrast injection, as may be the case with an end-hole catheter.
Flush catheters are designed for high-rate, high-vol­ume injections. An example is the angiographic pigtail catheter that has side holes placed on the straight portion of the catheter, often several centimeters from the end, in a “flush” configuration (Fig. 2-5). The side holes serve to maximize the rate of contrast injection possible, center the catheter in the vessel, and minimize unwinding of the curved catheter tip. The number and size of holes are constrained by the need to maintain both the strength of the catheter and some output through the distal end. Flush catheters typically have 8 to 12 side holes. The lumen between the distal side hole and end of the cathe­ter potentially can form clot if sufficient flow is not pre­sent during flushing.
Selective catheters, such as the cobra, are available as an end hole catheter or with two small sideholes immedi­ately adjacent to the catheter tip. The presence of the side holes results in a diminished end hole jet during contrast delivery with a power injector and thereby de­creases the risk of intimal injury.
Situations exist, however, where the use of an end-hole catheter is advantageous during selective catheterization. One or two side holes projecting into the parent vessel also may mask a situation in which the tip of the catheter is occlusive or buried in the vessel wall by allowing blood return. Rapid injection into such a “wedged” vessel, or into the wall, may precipitate vessel damage or rupture.
Catheters and Guidewires 15
FIGURE 2-5. Flush catheters. Pigtail, Neff, Omni (
The pigtail and omniflush catheters have small side holes on the distal catheter shaft to distribute the contrast and injection force evenly ( nantly as a stabilizer and blunt leading edge. The Neff catheter side holes ( the shaft; the operator can pull the limb down the contralateral iliac artery for a slow selective injection rather than having to maneuver the catheter over the aortic bifurcation.
arrows
); the tip configuration serves predomi-
curved arrows
) are on the short limb rather than on
left to right
Side hole catheters should not be used for therapeutic embolization. The presence of a side hole increases the risk of entangling an embolization coil at the catheter tip. Use of side hole catheters is avoided in the coronary and cerebral circulation because of the increased risk of thrombus formation in the potentially stagnant blood in the side holes or tip with subsequent embolization.
■ Conclusion
In interventional radiology procedures, planning is im­perative. If the basic aspects of the case are anticipated, unexpected situations can be handled in a rational fash­ion. Experience will provide a set of solutions to particu­lar problems, but it does not substitute for a logical plan that avoids potential problems and equipment incom­patibilities. If properly performed, a percutaneous inter­vention can provide valuable diagnostic information and effect a good therapeutic result with minimal risk.
REFERENCES
1. Seldinger SI. Catheter replacement of the needle in percutaneous angiography. Acta Radiol 1953;39:368–376.
2. Schrader R, Steinbacher S, Burger W, et al. Collagen application for sealing of arterial punctures in comparison to pressure dressings: a randomized trial. Cathet Cardiovasc Diagn 1992;27:298–302.
3. Eisenberg RL, Fiske CE, Hedgcock MW. Catheter angiography through aortofemoral grafts: inadvertent catheterization of the na­tive iliac artery. AJR Am J Roentgenol 1979;32:852.
).
16 A. R. Abadir and J. E. Silberzweig
https://t.me/med1917
4. Kim D, Orron DE, Skillman JJ, et al. Role of superficial femoral artery puncture in the development of pseudoaneurysm and arte­riovenous fistula complicating percutaneous transfemoral cardiac catheterization [Comment]. Cathet Cardiovasc Diagn 1992;126: 327–328.
5. Altin RS, Flicker S, Naidech HJ. Pseudoaneur ysm formation and arteriovenous fistula formation after femoral artery catheterization: association with low femoral punctures AJR Am J Roentgenol 1989; 152:629–631.
6. Sreeram S, Lumsden AB, Miller JS, et al. Retroperitoneal hema­toma following femoral artery catheterization: a serious and often fatal complication. Ann Surg 1990;4:328–333.
7. Lechner G, Jantsch H, Waneck R, et al. The relationship between the common femoral artery, the inguinal crease, and the inguinal ligament: a guide to accurate angiographic puncture. Cardiovasc Intervent Radiol 1988;11:165–169.
8. Rupp SB, Vogelzang RL, Nemcek AA, Jr. et al. Relationship of the inguinal ligament to pelvic radiographic landmarks: anatomic cor­relation and its role in femoral arteriography [Comment]. J Vasc Interv Radiol 1993;4:834.
9. Spijkboer AM, Scholten FG, Mali WP, et al. Antegrade puncture of the femoral artery: morphologic study. Radiology 1990;176:57–60.
10. Mozersky DJ, Olson RM, Coons HG, et al. Doppler controlled needle director: a useful adjunct to angiography. Radiology 1973; 109:221–222.
11. Khangure MS, Chow KC, Christensen MA. Accurate and safe punc­ture of a pulseless femoral artery. Radiology 1982;144:927–928.
12. AbuRahma AF, Robinson PA, Boland JP. Safety of arteriography by
direct puncture of a vascular prosthesis. Am J Surg 1992;164: 233–236.
13. AbuRahma AF, Robinson PA, Boland JP, et al. Complications of arteriography in a recent series of 707 cases: factors effecting out­come. Ann Vasc Surg 1993;7:122–129.
14. Grollman JH Jr, Marcus R. Transbrachial arteriography: techniques and complications. Cardiovasc Intervent Radiol 1988;11:32–35.
15. Bakal CW, Friedland RJ, Sprayregen S, et al. Translumbar arch aortography: a retrospective controlled study of usefulness, tech­nique, and safety. Radiology 1991;178:225–228.
16. Yandow D, Wojtowycz M, Alter A, et al. Detection of retroperitoneal hemorrhage after translumbar aortography by computerized to­mography. Angiology 1980;31:655–659.
17. Quigley MJ, Sniderman KW, Yeung EY. Translumbar aortography: experience with a steerable pigtail catheter. Can Assoc Radiol J 1993;44:29–34.
18. Dotter CT, Judkins MP, Fische LH. Safety guide spring for percu­taneous cardiovascular catheterization. AJR Am J Roentgenol 1966;98:957–960.
19. Baum S. Catheters and injectors. In: Abrams, H, ed. Abrams’ Angiog- raphy, 2nd ed. Boston: Little, Brown and Company; 1971.
20. Wright KC, Wallace S, Charnsangavej C, et al. Flow-directed cathe­ter for superselective arterial catheterization: an experimental evaluation. Cardiovasc Intervent Radiol 1986;9:54–56.
21. Reidy JF, Ludman C. The safety of outpatient arteriography using 3F catheters. Br J Radiol 1993;66:1048–1049.
22. Waltman AC, Courey WR, Athanasoulis C, et al. Technique for left gastric artery catheterization. Radiology 1973;109:732–734.
S.I. Wahl and K. M. ZinnFilming and Injection Techniques
https://t.me/med1917
3
■■■
Filming and Injection Techniques
SAMUEL I. WAHL AND KENNETH M. ZINN
Selecting the proper filming technique and injection rate is critical in performing high-quality angiographic stud­ies. Many variables must be considered, and each exami­nation, or run must be tailored appropriately. The repre­sentative filming sequences and injection parameters provided herein are those typically used at our institu­tions, and they may need to be modified to suit the individual patient.
■ Image Acquisition
Currently, two basic filming techniques are used in angiography: conventional filmscreen imaging and digi­tal subtraction angiography (DSA) imaging. Regardless of the techniques used, it is essential that a fluoroscopic or scout view of the area of interest be obtained at the start of the study, before any invasive manipulation is done. This is especially important in the abdomen, where retained barium from a recent computed to­mography (CT) or gastrointestinal (GI) study could pre­clude obtaining an adequate examination. A scout film taken immediately after placing the patient on the table may also reveal an important element of an unknown clinical history. For example, we saw two patients in whom previously placed inferior vena cava filters were discovered on scout fluoroscopy when the patients were placed on the angiographic table for insertion of an in­ferior vena cava filter. In both cases, the patients were poor “historians,” and they had incomplete medical charts. Furthermore, it is essential that overlapping vas­cular segments be separated by additional views when necessary to answer a clinical question and that multiple
views be obtained whenever critical information is needed (Figs. 3-1 and 3-2).
Film-screen angiography
Conventional film-screen imaging is often referred to as cut-film imaging. It has been largely replaced by digital ac­quisition techniques; however, it is of both historical and current interest because it is the basis on which the sub­sequent technology was derived. Cut-film arteriography uses cut sheets of film stacked or placed in a roll within a delivery magazine or cassette called a rapid film changer. Rapid film changers carry between 20 and 30 14 ⫻ 14­inch films. They operate at maximum filming rates of either four or six films per second (Puck, AOT-S; Siemens­Elema-Schonander, Inc., Elk Grove Village, IL, U.S.A.). The number of films per second and the duration of film­ing during anindividual run areprogrammed into the sys­tem before each angiographic run. The most important consideration in determining the appropriate filming se­quence is the rate of blood flow in the vessel(s) being stud­ied as well as the particular pathology being investigated, with faster filming rates during the rapidly flowing arterial phase and slower filming during the capillary and venous phases. Filming near the contrast injection site generally mandates faster rates than filming downstream at a dis­tance from the catheter. The goal is to maximize the diag­nostic information obtained during each contrast injec­tion without excessive contrast dose or exposure to the patient and without a waste of x-rayfilm or radiation.
Film or acquisition sequences
Filming sequences are stated as the number of films ex­posed per second for a specified duration of seconds. The
17
18 S. I. Wahl and K. M. Zinn
https://t.me/med1917
A
B
FIGURE 3-1. Multiple views are needed to uncover overlapped vessels in this 73-year-old man with critical ischemia and a failed
left common femoral to above-knee vein-bypass graft. There is popliteal artery occlusion. The operating surgeon wishes to avoid the groin and is planning the proximal anastomosis from the proximal superficial femoral artery. (A) An anteroposterior (AP) view of the left femoral triangle and thigh. The arrow shows a patent proximal superficial femoral artery at the site of the proposed proximal anastomosis. (B) The oblique view demonstrates the occluded graft stump medially and the deep femoral artery laterally. The arrow highlights a tight stenosis at the origin of the superficial femoral artery, which was completely hidden in the AP view. This stenosis would result in compromised inflow into a graft placed distal to it and the graft would occlude. Ultimately, proximal anastomosis was placed at the common femoral artery as a result of this oblique angiogram.
A
FIGURE 3-2. The angiogram and physical examination must correlate. (A) An anteroposterior (AP) pelvic angiogram, part of an
aortofemoral runoff study. The patient has critical left lower-extremity ischemia, diffuse superficial femoropopliteal artery occlusive disease, and symmetric femoral pulses. Note the extensive significant disease in the right external iliac artery. (B) Right anterior oblique view of the left external iliac artery. The arrow highlights a significant lesion approximately 2 cm above the inguinal ligament, which was not seen in the AP view. Because the femoral pulses were palpated as symmetric, and there is extensive significant disease in the right external iliac artery, a hemodynamically significant lesion must be suspected somewhere in the left common external iliac artery, and multiple obliques are mandated until the etiology of the physical finding is found.
B
Filming and Injection Techniques 19
https://t.me/med1917
filming can be divided into segments, if so desired, to fol­low or parallel the blood flow. For example, a typical pro­grammed run for a routine abdominal aortogram might be three films per second for 4 seconds and then one film per second for 8 seconds. This would expose a total of 20 films over approximately 12 seconds. In our institutions, we would express this particular film sequence as “3 for 4 and 1 for 8.” Certain studies require that the filming con­tinue well into the venous phase. For example, a typical filming sequence for selective superiormesentric arteriog­raphy for GIbleeding or superior mesenteric portography might be one film/per second for 10 seconds and then one film every other second until the film changer has ex­posed the entire magazine of film (“1 for 10, then 1/2 out”). Thus, the entire run would give a total of 20 films lasting for a total of 30 seconds, which is adequate for late venous visualization(Table 3-1).
Filming should be tailored to the patient and the anat­omy. At one extreme, rapid flow through an arte­riovenous fistula mandates a minimal framing rate of three frames per second; four to six frames per second may be needed. On the other hand, for imaging an aneurysm in a patient with low cardiac output, one expo­sure every second or every other second may suffice.
A scout film obtained prior to any contrast injection is
used to verify proposed positioning and exposure tech­nique. Scout films usually are obtained after any signifi­cant patient repositioning as well as at the start of any procedure.
The initial exposed film of the run is referred to as the zero-second film. This exposure is obtained just prior to contrast opacification. It provides a necessary baseline for dynamic evaluation of the film run and allows the radi­ologist to analyze the run for any contrast-related arti­facts, for example, baseline calcium projected over an artery. Historically this film provided a “mask” for photo­graphically subtracting out the bony background struc­tures (after “polarizing” the film to reverse black and white) from the images so that only the vessels being studied are visualized. This process is called, for obvious reasons, subtraction imaging. Since the advent of DSA, photographic subtraction imaging is used infrequently, if ever; however, historically it provided the conceptual ba­sis for DSA. Zero-second and scout films also serve as a baseline for dynamic evaluation of the run (e.g., to differ­entiate extravasated contrast from preexisting calcifica­tions). This zero-second film differs from the scout film in that the zero film is included in the total filming run. In other words, it is the first of the sequence, whereas the scout film is obtained prior to the actual filmed run.
TABLE 3-1. Typical Filming and Flow Rates
Study Injection Rate Filming Projections Comments
Aortic arch 40 mL total @ 20 mL/s 3/s ⫻ 4 s, 1/s ⫻ 8 s 30–45 RPO⫹ AP Pigtail ~ 1.5 cm above valve Thoracic aorta 60 mL total @ 30 mL/s 3/s ⫻ 4 s, 1/s ⫻ 8 s 45 RPO ⫹ AP Include thoracic inlet to diaphragm Abdominal aorta 40 mL total @ 20 mL/s or 3/s ⫻ 4 s, 1/s ⫻ 8 s AP or biplane Catheter above celiac artery
30 mL total @ 15 mL/s filming
Pelvic 30 mL total @ 15 mL/s 2/s ⫻ 3 s, 1/s ⫻ 8 s AP Oblique as needed; IPO will open iliac
Internal iliac 25–40 mL total @ 2–4 mL/s 2/2 ⫻ 3 s, 1/s ⫻ 8 s Ipsilateral PO Ipsilateral ant. oblique with penis over
Subclavian/axillary 12–20 mL total @ 3–4 mL/s 2–3/s ⫻ 4 s, 1/s ⫻ 6s AP Brachial 12 mL total @ 3 mL/s 1–2/s ⫻ 8 s AP Check for possible high brachial bifur-
Radial/ulnar/hand 20–30 mL total @ 2–3 mL/s 1–2/s ⫻ 10–20 s Hand supinated Maximal vasodilation of extremity may
Celiac 40–50 mL total @ 5–8 mL/s 2/s ⫻ 5 s, 1/s ⫻ 10 s AP SMA 60 mL total @ 8 mL/s 1/s ⫻ 10 s, ½ s out AP Can use 30–60/mg intraarterial papa-
IMA 15–20 mL total @ 2–3 mL/s 2/s ⫻ 5 s, 1/s ⫻ 10 s AP Renal 12–15 mL total @ 4–8 mL/s 3/s ⫻ 3 s, 1/s ⫻ 6 s AP or 30 IAO For renal tumor evaluation, increase con-
Selective pulmonary
angiogram 40 mL total @ 20 mL/s 4/s ⫻ 4 s, 1/s ⫻ 4 s AP and 45 IPO Measure pulmonary artery pressure
IVC 60 mL total @ 30 mL/s 2/s ⫻ 3 s, 1/s ⫻ 8s AP
a
bifurcation
contralateral thigh after intracavenous injection for impotence evaluation
cation at the axillary level; catheter placed in axillary artery
be needed
verine to enhance venous phase; film covers 30/s to ensure visualization of SMV-portal vein
trast volume and delay filming for visualization of renal vein
prior to contrast injection
AP, anteroposterior; IAO, ipsilateral anterior oblique; IMA, inferior mesenteric artery; IPO, ipsilateral posterior oblique; IVC, inferior vena cava; RPO, right posterior oblique; SMA, superior mesenteric artery; SMV, superior mesenteric vein.
a
Flow rates and volumes are given for cut film or unsubtracted digital acquisition with full strength iodinated contrast material.
20 S. I. Wahl and K. M. Zinn
https://t.me/med1917
Digital subtraction angiography
Rather than exposing cut film at regular intervals, fluoro­scopically acquired images can be mapped sequentially into computer memory. The acquisition begins with “test shots,” which are equivalent to scouts, obtained to see whether the angiographic unit’s programmed exposure factors are correct. If the test shots are accepted, the actual acquisition series begins. As with conventional angiographic runs, a zero-second acquisition is obtained to act as the mask by slightly delaying the contrast injec­tion during the acquisition. From this image, some or all the background structures can be digitally subtracted. Precontrast background information (the “zero” expo­sure, or “mask”) can be subtracted out electronically from images obtained during contrast administration. Thus, underlying bone and soft tissue densities are sub­tracted out of each screen pixel (picture element) as con­trast passes through the image. This markedly enhances contrast sensitivity. Nonsubtracted images always should be viewed along with the subtracted ones; in this manner, underlying bone or soft tissue artifacts can be accounted for (Fig. 3-3). Originally, intravenous injections were used to image the arterial circulation (intravenous digital sub­traction angiography, or IV-DSA), but this still required a large amount of contrast, was subject to extensive motion artifact, and yielded only marginal spatial resolution; however, the use of intraarterial contrast injections (in­traarterial digital subtraction arteriography, or IA-DSA) allowed the use of smaller volumes of iodine while retain-
ing excellent contrast sensitivity; IA-DSA is also subject to less motion artifact. Over the last decade, technical ad­vances have markedly improved spatial resolution as
1–3
well.
Although decreased spatial resolution of DSA compared with film-screen arteriography still exists and can compromise the detailed evaluation of subtle small­vessel abnormalities, in clinical practice, the problem probably occurs infrequently. Thus, IA-DSA has replaced film-screen arteriography for most applications. The studies also can be interpreted directly from the com­puter monitor, thus eliminating the wait for processing a film run and significantly decreasing study time. Another advantage is that the images can be electronically post­processed to change the contrast and window level, inte­grate frames, and grade stenoses. Most angiographers prefer DSA to cut-film studies because the time required to complete the angiogram and the contrast load is sig­nificantly less than with DSA. Whereas spatial resolution is better than with cut films with DSA, DSA has a better chance of identifying extravasation (it is true that subtrac­tion can be performed on cut films, but this is tedious and expensive). We have found that DSA induces copious artifacts when searching for a GI bleeding source (as a result of breathing and bowel motion), and we use high­resolution nonsubtracted digital acquisition in these set­tings with full-strength, low-osmolar contrast.
The superior contrast sensitivity of DSA allows consid­erable reduction in contrast density necessary to produce optimal images. Generally, the contrast agent can be di-
A
FIGURE 3-3. Comparison of the nonsubtracted and subtracted images is extremely important when interpreting digital subtrac-
tion arteriography. (A) Selective digital subtraction arteriogram of the left external iliac artery. The single arrowhead highlights a linear defect in more proximal external iliac artery, which might be interpreted as a guidewire-induced iatrogenic dissection. More distal defects (dual arrows) could be similarly interpreted. (B) The nonsubtracted image demonstrates that these are really substraction artifacts related to underlying bony cortex (arrowhead) and a hip prosthesis (dual arrows).
B
Filming and Injection Techniques 21
https://t.me/med1917
luted with saline by 25 to 50%. Injection rates can be reduced slightly compared with film-screen angiography, but total volumes can be reduced significantly. Moreover, one also has the option of using full-strength contrast in a nonsubtracted mode, which may be necessary for pul­monary arteriography, abdominal arteriography, and pe­ripheral angiography, where bowel gas, respiration, or other patient movement causes motion artifact and pixel misregistration.
Digital “road mapping” is a helpful utility that aids in catheter negotiation through tortuous vessels and in other aspects of critical catheter placement. The operator leaves an overlay of the vasculature on the monitor from a previous contrast injection, which can be used as a map during the real-time fluoroscopic maneuvering of the catheter or guidewire.
3,4
During the procedure, any fluoroscopic image can be saved or “grabbed” for future viewing. Also, the fluoro­scopic image remains on the monitor and can be magni­fied without requiring additional fluoroscopy. Postproc­essing adjustments always can be made to improve image quality. One new equipment option allows active rotation of the C-arm (up to 20 to 30 degrees/second) during contrast injection; this technique can obtain multiple projections during a single contrast run.
With all the advantages offered by DSA, only a few disadvantages are encountered. These stem primarily from sequential pixel malalignment resulting from pa­tient, cardiac, respiratory, and bowel motion. The quality of DSA images is degraded significantly by motion, and this problem is particularly common with abdominal and pulmonary angiography. These problems, however, can be reduced by giving clear patient instructions and by using proper positioning and mild restraints. Postprocess­ing the acquisition by “pixel shifting” or selecting a new subtraction mask often eliminates motion-related degra­dation. Glucagon and abdominal compression also can be used before the study to decrease bowel activity. Cur­rent systems, however, allow high-quality, high-resolution nonsubtracted digital images to be obtained in 1024 ⫻ 1024 pixel matrices. Nonsubtracted digital images retain many of the benefits of subtracted studies, including de­creased table time, and can be used where motion artifact significantly degrades the image; however, nonsubtracted digital images require contrast rates and volumes that approach those needed for cut-film arteriography.
The use of low osmolar contrast medium greatly re­duces pulmonary irritation and therefore eliminates coughing during pulmonary angiography. By simply us­ing a nonsubtracted mode, excellent pulmonar y arterial visualization is accomplished, even in the presence of respiratory motion. With state-of-the-art digital systems, measures used to reduce bowel and respirator y motion are often unnecessary.
■ Aortofemoral Arteriography
Aortofemoral arteriography presents a special challenge to filming because the target vascular territory is consid­erably longer than conventional film or fluoroscopic field. The most basic approach would be to perform a separate series over each vascular segment, working proximally from the aorta down through the tibial and pedal arteries. This approach would require four or five separate film runs and contrast injections. More com­monly used techniques for obtaining aortofemoral arte­riograms generally employ a single programmed power injection and either table or image intensifier movement or special long film cassettes to acquire images in a single series. The more commonly used devices for aor­tofemoral arteriography are the stepping table and “bo­lus chasing.” The long-leg changer (BC Medical Ltd., Montreal, Quebec, Canada), another such device, is no longer manufactured, but it is still in use.
A stepping table traditionally used a rapid film changer (Puck or AOT-S film changer 14 ⫻ 17-inch film) as the patient is moved in “steps” over several different stations, extending from the pelvis to the feet during a single contrast injection. A long-leg changer, on the other hand, is a ceiling-mounted x-ray tube that uses multiple (usually six) 14 ⫻ 51-inch film cassettes housed within a rotating film unit below the patient. The time interval between each cassette exposure can be tailored to the patient so as to ensure complete opacification from the distal aorta to the pedal vessels. Such large-field changers are advan­tageous for arteriography and venography of the lower extremities because the entire extremity can be evaluated on each exposure, and calculation of the timing for lower extremity arteriography is less crucial. The disadvantages are the need for an additional ceiling-mounted radio­graphic tube with a tube to a floor distance of between 87 and 110 inches and a maximum filming rate of one film every 2.5 to 3 seconds. Extremely long contrast columns (8 to 10 seconds) are imaged. Late-generation stepping tables can acquire images digitally.
Finally, new digital “bolus chasing” techniques are an alternative to both a traditional stepping table or a long­leg changer. With bolus chasing, the operator can con­trol table movement manually while visualizing contrast opacification in real time. The operator then can alter the rate of the table movement to match the speed and position of the contrast bolus (i.e., bolus chasing), ac­quiring images longitudinally. Digital subtraction is ac­complished by duplicating the table movement and frame rate after the contrast has dissipated to obtain a “mask.” “Knee-arrival time” is computed by fluoroscopi­cally observing a small test bolus of contrast arrive at the popliteal artery prior to the full injection; then it is used by the system’s microprocessor to optimize the contrast
22 S. I. Wahl and K. M. Zinn
https://t.me/med1917
injection rate and volume and to coordinate it with an appropriate acquisition sequence.
Opacification of extremities can be enhanced, if neces­sary, by inducing reactive hyperemia after the release of pressure cuffs from the calves, warming the extremity, or using vasodilators such as intraarterial nitroglycerin, tola­zoline, or papaverine. Of these, transcatheter nitroglyc­erin, 200 lg, is preferred by some for augmenting visuali­zation of the lower extremities.
1
Spot films using 70-mm or 105-mm cameras are some­times used in angiography, but the small size of the frames makes fine detail difficult to appreciate, and these are now considered “legacy” devices by many. Cineradiog­raphy is found primarily in cardiac angiography suites, but the high radiation doses and the cumbersome small­field film viewing make it insufficient for abdominal and extremity work. Most peripheral arteriograms do not re­quire the extremely rapid frame rate necessary in cardiac angiography.
■ Flow Rates and Contrast Injectors
As stated previously, selecting the proper contrast flow rate for angiography is an extremely important part of a high-quality angiogram. Several programmable auto­mated injection devices are commercially available that provide predictable and reproducible contrast volumes delivered at a specific flow rate for angiographic studies. For example, a routine (digital) abdominal aortogram may require a total contrast volume of 30 mL at a flow rate of 15 mL/second (i.e., a 2-second injection).
Power injectors must be programmed prior to each run, depending on the vascular territory being studied,
avoiding the inconsistent, unknown rates of hand injec­tions. Furthermore, power injectors can deliver higher flow rates than hand injections. It is imperative that, prior to a power-injected run, a hand-delivered test injection be performed to ensure proper catheter position and to confirm the flow rate required.
The primary injection parameters programmed into the power injector prior to the run are the flow rate (mL/sec), the total contrast injection volume (mL), and the linear rise (Fig. 3-4). The flow rate is selected on the basis of vessel diameter and blood flow observed fluoro­scopically with a hand injection. Total contrast volume depends on the desired column length and the size of the territory to be opacified. For example, a 1.5- to 2-second contrast column issufficient to fill thefield of a thoracic or abdominal aortogram. Image acquisition must be delayed so that at least one noncontrast zero-second image is ac­quired for a maskbefore the arrival ofcontrast in the field. When the area of interest is at the catheter injection site, the injection typically is delayed 0.5 to 1 second after the start of image acquisition. If imaging is remote from the site of injection, however, a variable delay in filming may be programmed to occur after the start of the contrast injection. For example, with a catheter injection in the common iliac artery and filming over the foot, a long filming delay (e.g., 7 to 10 seconds) can aid in extending the series to allow acquisition of a good foot arteriogram. A catheter-specific maximum pressure limit is set to avoid catheter rupture during power injection. (Note that pres­sure is not a primary parameter.) Pressure generated dur­ing injection is related to the viscosity of the contrast me­dium, the programmed flow rate, the catheter luminal diameter, and catheter material and length. For example, a typical nylon, 90-cm, 5F pigtail catheter might have a
A
FIGURE 3-4. The injection settings for a flush aortogram are shown on the control panel from a Mark V power injector (Medrad,
Inc., Pittsburgh, PA, U.S.A.). (A) The pressure limit is set 999 pounds per square inch (psi) just below the catheter pressure limit. (B) By using the “status function” button, the interventionalist can view the actual injection parameters after the run is completed. The arrow demonstrates that the combination of flow parameters, contrast viscosity, and catheter factors actually generated a pressure of 760, considerably less than the set pressure limit.
B
Filming and Injection Techniques 23
https://t.me/med1917
maximum pressure limit of 1050 pounds per square inch (psi). Excess pressure generally causes rupture at the catheter hub, where pressure ishighest. Atthe setpressure limit, a power injectortypically willterminate the injection so that rupture is avoided (Fig. 3-4).
Linear rise is an additional primary injection parame­ter. It is the time required from the onset of the injection (zero flow) to reach the maximum flow rate specified by the injector. Linear rise is used to avoid catheter recoil and should depend on the catheter type and position. A typical linear rise of 0.2 seconds is usually sufficient for multi-sidehole catheters during flush aortography. A longer linear rise (0.4 to 0.8 seconds) decreases the like­lihood of catheter recoil (and selective vessel damage) and should be used during selective arteriography.* For example, if an end-hole catheter is selectively placed in a vessel, the sudden recoil produced by the maximum rate of injection without a gradual rise potentially could dis­lodge the catheter or, worse, damage the vessel. Nonse­lective flush catheters with sideholes may allow the use of a shorter linear rise (0.2 seconds) as the contrast is gen­erally distributed equally in all directions from the multi­ple holes maintaining the catheter’s position. Generally, selective catheters with sideholes should be used with medium to long linear rise times. Contrast injectors also usually include a contrast syringe heater to reduce con­trast viscosity and to allow flow rates of up to 40 mL/sec. The 4 to 5 Fr catheters commonly used today are subject to considerably more recoil for a given flow rate than the 6 to 7 French catheters that were common a few years ago; thus, we have tended to use longer linear rise times when in selective positions.
■ Contrast Agents
Our central tenet is to administer the minimal amount of contrast necessary to obtain complete diagnostic infor­mation or complete an interventional study. Increasing volumes may add additional risk to patients with poor cardiac output or renal failure. Low-osmolality contrast agents are preferred in most vascular and interventional radiology departments because they decrease patient dis­comfort and reduce the incidence of adverse side effects, such as contrast-induced renal failure compared with conventional, higher osmolar agents. They also appear to reduce minor contrast reactions, such as urticaria, nau-
* This concept can be explained easily compared with sitting idle at a traffic light. When the light turns green, and you abruptly push your foot against the accelerator, a considerable amount of recoil will be placed on your head forcing it back­wards; if you gradually depress the accelerator in a progressive linear fashion, however, much less recoil will be encountered.
sea, and vomiting. Although rare, severe and fatal reac­tions are not eliminated by using low-osmolar contrast agents. The efficacy of steroid prophylaxis with low osmo­lar contrast is controversial. Institution-specific protocols should be consulted and followed.
6–8
The occurrence of clinically significant contrast-re­lated renal dysfunction usually is associated with preexist­ing renal compromise. Advanced age and diabetes con­tribute to risk.
9
Patients with multiple myeloma may incur renal damage as a result of tubular precipitation in high­risk groups; so imaging examinations that do not require iodinated contrast material should be considered if they can give clinically useful data. It is essential that patients who are high risk for renal failure receive low osmolar contrast. Some angiographers use furosemide and man­nitol infusions to maintain renal blood flow and prevent renal toxicity, but benefit from this protocol remains con­troversial.
10–13
Full-strength iodinated contrast for angiography gen­erally ranges from 300 to 370 mg/mL iodine. Dilute contrast for injection may be half to two-thirds strength when using digital acquisition. The newer contrast agents do not have anticoagulant properties typical of standard high-osmolar contrast; thus, it is essential that low-osmo­lar agents not be allowed to contact blood for long peri­ods. Thus, care must be taken not to allow blood to dwell
5
in power-injector syringes, tubing, or catheters prior to injection.
Despite the low incidence of contrast reactions using low-osmolar, iodinated contrast agents, the potential for problems in patients with a history of prior allergic reac­tions and in those with renal insufficiency remains. For this reason, carbon dioxide (CO viable alternative contrast agent.
) was developed as a
2
14
Carbon dioxide is a nonallergic, highly dissolvable gas that is excreted by the lungs, avoiding the nephrotoxicity of iodinated agents. By temporarily displacing blood, CO
decreases the radiographic density within the vessel
2
lumen, essentially the opposite radiographic effect of tra­ditional contrast agents. CO
can be used as an intravas-
2
cular contrast agent for both diagnostic and interven­tional studies; however, because of its potential for neurotoxicity, it is not used for studies performed above the diaphragm. Special delivery techniques have been designed for CO with air can be fatal.
because it is colorless, and admixture
2
15
One must be careful to use small, controlled injection volumes and correct patient posi­tioning to avoid gas trapping in nondependent structures such as the main pulmonary artery. The trapped gas can obstruct pulmonary blood flow, leading to cardiac fail­ure. If the injection volume, duration, and interval are adequate, unlimited quantities of CO
can be used. CO
2
angiography has been used during TIPS (transjugular intrahepator portosystemic shunt) procedures with excel­lent results; however, its contrast resolution is inferior to
2