Добавил:
Sekretar
kiopkiopkiop18@yandex.ru
t.me/Prokururor I Вовсе не секретарь, но почту проверяю
Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз:
Предмет:
Файл:Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5545_Библиотеки_им_академика_М_И_Перельмана.pdf
X
- •Preface
- •Contents
- •1: Structure of Matter
- •2: Radioactive Decay
- •2.1 Spontaneous Fission
- •1.1.1 Radiation
- •1.2 The Atom
- •1.2.3 Nuclear Binding Energy
- •1.3 Nuclear Nomenclature
- •1.5 Questions
- •Suggested Readings
- •2.2 Isomeric Transition
- •2.2.1 Gamma (γ)-Ray Emission
- •2.2.2 Internal Conversion
- •2.2.2.1 Problem 2.1
- •2.2.2.2 Answer
- •2.3 Alpha (α)-Decay
- •2.4 Beta (β−)-Decay
- •2.5 Positron (β+)-Decay
- •2.6 Electron Capture
- •2.7 Questions
- •Suggested Readings
- •3.1 Radioactive Decay Equation
- •3.1.1 General Equation
- •3.1.2 Half-Life
- •3.1.3 Mean Life
- •3.1.4 Effective Half-Life
- •3.2 Units of Radioactivity
- •3.3 Specific Activity
- •3.4 Calculation
- •3.5 Successive Decay Equations
- •3.5.1 General Equation
- •3.5.2 Transient Equilibrium
- •3.5.3 Secular Equilibrium
- •3.6 Questions
- •Suggested Readings
- •4.5 Poisson Distribution
- •4.6 Gaussian Distribution
- •4.7 Chi-Square Test
- •4.8 Minimum Detectable Activity
- •4.10 Questions
- •Suggested Readings
- •5.1 Cyclotron-Produced Radionuclides
- •5.2 Reactor-Produced Radionuclides
- •5.2.1 Fission or (n, f) Reaction
- •5.2.2 Neutron Capture or (n, γ) Reaction
- •5.6 Radionuclide Generators
- •5.8 Questions
- •Suggested Readings
- •6.1.1 Specific Ionization
- •6.1.2 Linear Energy Transfer
- •6.1.3 Range
- •6.1.4 Bremsstrahlung
- •6.1.5 Positron Annihilation
- •6.2.1.1 Photoelectric Effect
- •6.2.1.2 Compton Scattering
- •6.2.1.3 Pair Production
- •6.2.1.4 Raleigh Scattering
- •6.2.1.5 Photodisintegration
- •6.3.2 Half-Value Layer
- •6.5 Questions
- •Suggested Readings
- •7: Gas-Filled Detector
- •7.1 Principles of Gas-Filled Detector
- •7.2 Ionization Chamber
- •7.2.1 Ion Chamber Survey Meter
- •7.2.2 Dose Calibrator
- •7.2.2.1 Constancy
- •7.2.2.2 Accuracy
- •7.2.2.3 Linearity
- •7.2.2.4 Geometry
- •7.2.3 Pocket Dosimeter
- •7.3 Proportional Counter
- •7.4 Geiger–Müller Counter
- •7.5 Questions
- •Suggested Readings
- •8.1 Scintillation Counter
- •8.4.3 Characteristic X-Ray Peak
- •8.4.4 Backscatter Peak
- •8.4.5 Iodine Escape Peak
- •8.2 Solid Scintillation Detector
- •8.2.1 NaI (Tl) Detector
- •8.2.2 Bismuth Germanate Detector
- •8.2.3 Barium Fluoride Detector
- •8.2.4 Lutetium Oxyorthosilicate Detector
- •8.2.5 Gadolinium Oxyorthosilicate Detector
- •8.2.6 Yttrium Oxyorthosilicate Detector
- •8.2.7 Yttrium Aluminum Perovskite Detector
- •8.2.8 Lutetium Yttrium Oxyorthosilicate Detector
- •8.2.9 Lanthanum Bromide Detector
- •8.3 Solid-State Detector
- •8.3.2 Cadmium–Zinc–Tellurium Detector
- •8.3.3 Cesium Iodide (CsI(Tl)) Detector
- •8.3.4 Solid Scintillation Counter
- •8.3.4.1 NaI(Tl) Detector
- •8.3.4.2 Photomultiplier Tube
- •8.3.4.3 Preamplifier
- •8.3.4.4 Linear Amplifier
- •8.3.4.5 Pulse-Height Analyzer
- •8.3.4.6 Display or Storage
- •8.4 Gamma-Ray Spectrometry
- •8.4.1 Photopeak
- •8.4.6 Positron Annihilation Peak
- •8.4.7 Coincidence Peak
- •8.5 Liquid Scintillation Counter
- •8.5.1 Quenching
- •8.6.1 Energy Resolution
- •8.6.2 Detection Efficiency
- •8.6.2.1 Intrinsic Efficiency
- •8.6.2.2 Photopeak Efficiency or Photofraction
- •8.6.2.3 Geometric Efficiency
- •8.6.3 Dead Time
- •8.7 Gamma Well Counter
- •8.8 Thyroid Probe
- •8.8.1 Thyroid Uptake Measurement
- •8.9 Questions
- •Suggested Readings
- •9: Gamma Camera
- •9.1 Gamma Camera
- •9.1.2 Detector
- •9.1.3 Collimator
- •9.1.4 Photomultiplier Tube
- •9.1.5 X-, Y-Positioning Circuit
- •9.1.6 Pulse-Height Analyzer
- •9.2 Digital Camera
- •9.2.1 Solid State Digital Camera
- •9.3 Questions
- •Suggested Readings
- •10.1.1 Spatial Resolution
- •10.1.1.1 Intrinsic Resolution
- •10.1.1.2 Collimator Resolution
- •10.1.1.3 Scatter Resolution
- •10.1.2.1 Bar Phantom
- •10.1.2.2 Line-Spread Function
- •10.1.2.3 Modulation Transfer Function
- •10.1.3 Sensitivity
- •10.1.3.1 Collimator Efficiency
- •10.1.4 Uniformity
- •10.1.5 Pulse-Height Variation
- •10.1.6 Nonlinearity
- •10.1.7 Edge Packing
- •10.2 Gamma Camera Tuning
- •10.4 Contrast
- •10.4.1 Count Density
- •10.4.2 Image Noise
- •10.4.4 High Count Rate
- •10.4.6 Patient Motion
- •10.5.1 Daily Checks
- •10.5.1.2 Uniformity
- •10.5.2 Weekly Checks
- •10.5.3 Monthly Checks
- •10.5.3.1 High-Count Uniformity Calibration
- •10.5.3.2 Collimator Integrity
- •10.5.4 Annual, Semiannual, or As-Needed Checks
- •10.6 Questions
- •References and Suggested Readings
- •11.1.1 Central Processing Unit
- •11.1.2 Computer Memory
- •11.1.3 External Storage Device
- •11.1.4 Input/Output Device
- •11.1.7 Digital-to-Analog Conversion
- •11.1.8 Digital Image
- •11.2.1 Digital Data Acquisition
- •11.2.2 Static Study
- •11.2.3 Dynamic Study
- •11.2.4 Gated Study
- •11.2.7 Display
- •11.3.1 PACS
- •11.4 Questions
- •Suggested Readings
- •12: Single Photon Emission Computed Tomography
- •12.1 Tomographic Imaging
- •12.2 Single Photon Emission Computed Tomography
- •12.2.1 Data Acquisition
- •12.2.2 Image Reconstruction
- •12.2.2.1 Simple Backprojection
- •12.2.2.2 Filtered Backprojection
- •12.2.2.3 The Convolution Method
- •12.2.2.4 The Fourier Method
- •12.2.2.6 Iterative Reconstruction
- •12.3 SPECT/CT Scanner
- •12.4 Factors Affecting SPECT
- •12.4.1 Photon Attenuation
- •12.4.2 Attenuation Correction Methods
- •12.5 Partial-Volume Effect
- •12.5.2 Sampling
- •12.5.3 Scattering
- •12.6.1 Spatial Resolution
- •12.6.2 Sensitivity
- •12.6.3 Other Parameters
- •12.7.1 Daily Tests
- •12.7.2 Weekly Tests
- •12.7.2.1 Spatial Resolution
- •12.9 Questions
- •References and Suggested Readings
- •13: Positron Emission Tomography
- •13.1 Introduction
- •13.2 PET Radiopharmaceuticals
- •13.3.2 Block Detector
- •13.5 Coincidence Timing Window
- •13.6 PET/CT Scanner
- •13.7 PET/MR Scanner
- •13.7.2 MR Scanner
- •13.7.3 Commercial PET/MR Scanner
- •13.8 Mobile PET or PET/CT Scanner
- •13.9 Micro-PET Scanner
- •13.11 Data Acquisition
- •13.12 Image Reconstruction
- •13.13 Factors Affecting PET
- •13.13.1 Normalization
- •13.13.2 Photon Attenuation Correction
- •13.13.4 Random Coincidences
- •13.13.5 Scatter Coincidences
- •13.13.6 Dead Time
- •13.13.7 Radial Elongation
- •13.14.1 Spatial Resolution
- •13.14.2 Sensitivity
- •13.14.2.1 Noise Equivalent Count Rate
- •13.15.1 Daily Tests
- •13.15.1.1 Sinogram Check
- •13.15.2 Weekly Tests
- •13.15.2.1 Normalization
- •13.18 Questions
- •References and Suggested Reading
- •14.1 Background
- •14.5 Artificial Neural Network
- •14.7 Machine Learning
- •14.7.1 Decision Tree
- •14.7.2 Random Forest
- •14.7.3 Support Vector Machine
- •14.7.4 Computer Vision
- •14.8 Deep Learning
- •14.8.1 Convolutional Network
- •14.8.2 Recurrent Neural Network
- •14.8.3 Generative Adversarial Network
- •14.8.4 Transfer Learning
- •14.9 Radiomics
- •14.10 Natural Language Processing
- •14.11 Large Language Model
- •14.12 Generative Artificial Intelligence
- •14.13.1 Prompt
- •14.13.2 Token
- •14.13.3 Hallucination
- •14.13.4 Deepfake
- •14.13.5 Overfitting
- •14.15 Chatbot
- •14.18 Legal Implication
- •14.20 Questions
- •References
- •15.1 Introduction
- •15.2.1 Scheduling
- •15.2.2 Image Acquisition
- •15.2.3 Image Processing
- •15.2.4 Interpretation
- •15.2.5 Reporting
- •15.3.1 Oncology
- •15.3.2 Cardiovascular Disease
- •15.3.3 Bone Scintigraphy
- •15.3.4 Thyroid Imaging
- •15.5 Drug Development
- •15.6 Questions
- •References and Suggested Reading
- •16: Internal Radiation Dosimetry
- •16.1 Radiation Unit
- •16.1.1 Roentgen
- •16.1.2 Rad
- •16.1.3 Gray
- •16.1.4 Rem
- •16.1.5 Radiation Weighting Factor
- •16.1.6 Quality Factor
- •16.1.7 Sievert
- •16.2 Dose Calculation
- •16.2.1 Radiation Dose Rate
- •16.2.2 Cumulative Radiation Dose
- •16.2.3 Factors Affecting Ã
- •16.2.4 The S Values
- •16.4 Pediatric Dosage
- •16.5 Questions
- •References and Suggested Readings
- •17: Radiation Biology
- •17.1 The Cell
- •17.2.1 DNA Molecule
- •17.2.2 Chromosome
- •17.5 Cell Survival Curves
- •17.6 Factors Affecting Radiosensitivity
- •17.6.1 Dose Rate
- •17.6.2 Linear Energy Transfer
- •17.6.4 Chemicals
- •17.7 Radiosensitizer
- •17.7.1 Oxygen
- •17.7.2 Pyrimidine
- •17.7.3 Others
- •17.8 Radioprotector
- •17.9 Apoptosis
- •17.13.1 Hematopoietic Syndrome
- •17.13.2 Gastrointestinal Syndrome
- •17.13.3 Cerebrovascular Syndrome
- •17.14.1 Somatic Effects
- •17.14.1.1 Carcinogenesis
- •17.14.1.3 Dose–Response Relationship
- •17.14.1.5 Leukemia
- •17.14.1.6 Breast Cancer
- •17.14.1.7 Other Cancers
- •17.14.1.10 Nonspecific Life-Shortening
- •17.14.1.11 Cataractogenesis
- •17.14.2 Genetic Effects
- •17.14.2.1 Spontaneous Mutation
- •17.14.2.2 Doubling Dose
- •17.14.2.3 Genetically Significant Dose
- •17.17 Questions
- •References and Suggested Readings
- •18.1 Introduction
- •18.2 Radiation Protection
- •18.2.3 Occupational Dose Limits
- •18.2.4 ALARA Program
- •18.2.5.1 Time
- •18.2.5.2 Distance
- •18.2.5.3 Shielding
- •18.2.5.4 Activity
- •18.2.6 Personnel Monitoring
- •18.2.6.1 Film Badge
- •18.2.6.2 Thermoluminescent Dosimeter
- •18.2.6.3 Optically Stimulated Luminescence Dosimeter
- •18.3 Radiation Regulations
- •18.3.1 License
- •18.3.1.1 General License
- •18.3.1.2 Specific License of Limited Scope
- •18.3.1.3 Specific Licenses of Broad Scope
- •18.3.2 Radiation Safety Committee
- •18.3.3 Radiation Safety Officer
- •18.3.4.3 Supervision
- •18.3.4.4 Mobile Nuclear Medicine Service
- •18.3.4.5 Written Directives
- •18.4 Bioassay
- •18.6 Radioactive Waste Disposal
- •18.6.2 Release into Sewerage Systems
- •18.6.4 Other Disposal Methods
- •18.7 Radioactive Spill
- •18.8 Recordkeeping
- •18.10 Dirty Bombs
- •18.11 Types of Accidental Radiation Exposure
- •18.12 Protective Measures in Case of Explosion of a Dirty Bomb
- •18.13 Verification Card for Radioactive Patients
- •18.14 Radiation Phobia
- •18.15 European Regulations Governing Radiation
- •18.16 Questions
- •References and Suggested Readings
- •Index

13.10 Dual- andTriple-Head Gamma Cameras inPET Imaging
227
13.8 Mobile PET or PET/CT Scanner
Largely because of the low patient volume and high cost, many community hospitals cannot afford PET or PET/CT scanners, but can take advantage of mobile PET
or PET/CT that provides PET scanning services to different locations. PET or PET/
CT scanners and necessary accessories are installed in sturdy vans, along with
nuclear pharmacy facilities. The mobile unit moves to different clients’ facilities on
different days depending on the schedule. The patient’s schedule and delivery of
PET tracers must be well coordinated to provide efcient services. The owner of the
mobile unit must have a license from the appropriate authorities to operate mobile
PET/CT and also a letter of agreement between the client and the licensee to provide the service. The van must meet the Department of Transportation’s overload
regulations, and the rules and regulations of re safety and security of local
authorities.
13.9 Micro-PET Scanner
For research animal imaging, clinical PET scanners with a large bore give poor
spatial resolution. Micro PET scanners with a smaller bore and, hence, smaller in
overall size (to be tted in small rooms) have been developed by several manufacturers. The typical bore diameter is about 16 cm. The spatial resolution can be
obtained as small as 1mm with the use of LSO detectors. These scanners are useful
for drug evaluation in animals.
13.10 Dual- andTriple-Head Gamma Cameras inPET Imaging
Conventional dual-head and triple-head gamma cameras (Fig. 12.2) can be utilized
as PET cameras by connecting the appropriate heads with a coincidence circuitry.
The typical time window is ~12ns for dual-head and ~10ns for triplehead cameras.
In SPECT mode, the cameras are used with collimators, whereas in PET mode, collimators are removed, and therefore they can be used in either mode as needed.
These cameras are attractive to community hospitals and third world countries
because of their low cost with the scope of PET imaging.
These cameras suffer from the disadvantage of low sensitivity due to low detection efciency of NaI(Tl) crystal for 511-keV photons. To improve sensitivity,
thicker crystals of sizes 1.6 to 2.5cm have been employed in some cameras, with a
resultant increase in coincidence photopeak efciency of only 3–4%, with concomitant degradation of spatial resolution. There is a signicant camera dead time loss
and pulse pile-up of counts in PET mode in the absence of a collimator because the
number of detectors is only two or three, unlike thousands in PET cameras. Overall,
the spatial resolution of a multihead camera is poorer than that of a dedicated PET
scanner. The use of these cameras is fading due to popularity of PET cameras.

228
X
AB
ABCD
BC
ABCD
13 Positron Emission Tomography
13.11 Data Acquisition
In PET imaging, two 511-keV annihilation photons are detected in coincidence by
two opposite detectors along a straight line, called the line of response (LOR). In a
full ring system, data are collected in 360° simultaneously, whereas in the partial
ring system, the rings are rotated around the patient for 360° data acquisition. There
are three steps in PET data acquisition. First, the location of the detector pair in the
ring is determined for each coincident event. Next, the pulses are analyzed by PHA
to check if they are within the energy window set for 511 keV.Finally, the position
of the LOR is determined in polar coordinates to store the data in computer memory.
Because each detector is connected to many opposite detectors in coincidence,
which detector pair detected a coincidence event must be determined. As in gamma
cameras, the position X, Y of each detector in the ring is determined by
CD
AD
(13.1)
(13.2)
where A, B, C, D are the pulses from the four PM tubes attached to the block, as
shown in Fig.13.1.
Next, the four pulses (A, B, C, and D) are summed up to give a Z pulse, which is
then checked by the PHA if its amplitude is within the energy window set for the
511-keV photons. If it is outside the window, it is rejected; otherwise, it is accepted
for storage.
The last step in data acquisition is the storage of the data in the computer. Unlike
conventional planar imaging, where individual events are stored in a (X, Y) matrix,
the coincidence events in PET imaging are stored in the form of a sinogram.
Consider an annihilation event occurring at the * position in Fig.13.13a. The coincidence event is detected along the LOR indicated by the arrow between the two
detectors. It is not known where along the line of travel of the two photons the event
occurred, because they are accepted within the set time window (say, 12 ns) and
their exact times of arrival are not compared. The only information we have is the
positions of the two detectors in the ring that registered the event; that is, the location of the LOR is established by the (X, Y) positioning of the two detectors. Many
coincident events arise from different locations along the LOR and all are detected
by the same detector pair and stored in the same pixel, as described below.
For data storage in sinograms, each LOR is dened by the distance (r) of the
LOR from the center of the scan eld (i.e., the center of the gantry) and the angle of
orientation (ϕ) of the LOR (i.e., the angle between r and the vertical axis of the
eld). A matrix of an appropriate size is chosen, dened by the r, ϕ coordinates,
rather than by X-, Y-coordinates used in SPECT data acquisition, and counts in each

ab
13.11 Data Acquisition
Fig. 13.13 PET data acquisition in the form of a sinogram. (a) Each LOR datum is plotted in
(r
, ϕ) coordinates. (b) Data for all r and ϕ values are plotted to yield the sinogram indicated by the
shaded area (only a part is shown). (Reprinted with the permission of the Cleveland Clinic
Foundation)
229
LOR are stored in the corresponding pixel in the matrix. If we plot the distance r on
the x-axis and the angle ϕ on the y-axis, then the coincidence event along the LOR
(r, ϕ) will be assigned at the cross-point of r and ϕ values (Fig.13.13b). In a given
projection, adjacent detector pairs constitute parallel LORs (at different r values in
Fig.13.13a) at the same angle of orientation. The plot of these LORs will be seen
as a horizontal row at angle ϕ. Similarly, LORs from different projections (i.e., at
different angles ϕ) for the same r values can be plotted, which will give a vertical
line. When all projections around the eld of view are considered, the plot of the
LORs at different projection angles and r values will result in the shaded area in
Fig.13.13b, which is called the sinogram. A typical normal sinogram is shown in
Fig.13.14.
The sinogram represents a single slice of data for a transverse FOV obtained
from a single ring of the PET scanner. PET data are acquired directly into a sinogram in a matrix of appropriate size in the computer. Each pixel corresponds to a
particular LOR characterized by (r, ϕ) containing all coincidence counts detected by
the detector pair along the LOR.Data can be collected in both static and dynamic
imaging using either the frame mode or the list mode, described in Chap. 11.
Because PET scanners are axially xed, whole-body imaging is accomplished
by the use of a computer-controlled bed-table that moves along the axis of the scanner. The whole-body scan of the patient is obtained at different axial positions of
the bed.

230
Fig. 13.14 A typical
normal sinogram
indicating all detectors are
working properly
13 Positron Emission Tomography
Fig. 13.15 An illustration of the time of ight (TOF) technique in PET. The time difference
(t1–t2) in the arrival of two 511keV photons is related to Δx. Using TOF information, the location
of annihilation of the positron is determined within a spatial range of (c. Δt)/2, where c is the
velocity of light
13.11.1 Time ofFlight Method
The time of ight (TOF) PET technique is based on the measurement of time difference in the arrival of the two 511 keV annihilation photons at the detectors, as
illustrated in Fig.13.15. Suppose two detectors are equidistant x from the center of
FOV (CFOV) and a positron is annihilated in the patient at position at a distance Δx
from the CFOV.One of the 511 keV photons will travel x+Δx and the other will
travel x−Δx. Since the photons travel at speed of light (c), the difference in time
(Δt) of arrival of the two photons at the detectors is 2 Δx/c. Note that the photons
from the center of FOV arrive at the detectors simultaneously (Δx=0). The uncertainty in the annihilation location is usually much smaller than the diameter D of the
patient (with good timing resolution). The signal-to-noise ratio (SNR) increases

13.11 Data Acquisition
231
with improved timing resolution and is proportional to D/Δx i.e., 2D/(c. Δt). For a
40-cm subject and with a timing resolution of 0.6 ns, the SNR (sensitivity) increases
by a factor of 4.4. With fast electronics and scintillators, and also a shorter time
window, TOF PET scanners can measure the time difference Δt fairly accurately
and provide high-resolution images and better localization of the lesion. Furthermore,
the TOF method reduces the scanning time.
Decades ago, the TOF PET scanners were introduced using fast detectors like
cesium uoride (CsF) or barium uoride (BaF2) to allow high count rates, but spatial resolution and sensitivity were poorer than those of conventional PET scanners
with BGO, because of poor light production. With the advent of improved scintillators like LSO, LYSO and LaBr3, along with more efcient and reliable PM tubes,
TOF is reemerging as the accepted technique in commercial PET scanners. Philips
Healthcare markets Big Bore Gemini TF PET scanners based on the TOF technique.
Current TOF scanners offer better spatial resolution and sensitivity because of
increased light production in the detector material and the shorter timing resolution.
Moreover, this technique improves the image quality in heavy patients, compared to
the conventional PET scanners, which leads to a loss of counts at high count rates
as well as an increase in scattered counts.
13.11.2 Two-Dimensional Versus Three-Dimensional
Data Acquisition
Coincident counts detected by a detector pair are called the prompts which
include true, random, and scatter events described later (Fig.13.16). To eliminate random and scatter events, annular septa (~1mm thick and radial width of
7–10 cm) made of tungsten or lead are inserted between rings in multiring PET
scanners (Fig.13.17a). The septa act as parallel hole collimators in gamma cameras. They mostly allow direct coincidence events to be recorded from a given
ring and prevent random and scatter from other rings. This mode of data collection is called two- dimensional (2-D) acquisition. The use of septa reduces the
contribution of scattered photons from 30% to 40% without septa to 10–15%. To
improve sensitivity, detector pairs in two adjacent or nearby rings are also connected in coincidence. Such cross-coincidence connections can be made at most
among ve adjacent rings. Coincidence events detected by the detectors connected in the same ring are called the direct plane events, whereas those detected
by detectors interconnected between different rings are called the cross plane
events. Although the cross plane events increase the sensitivity, they degrade the
spatial resolution as a trade-off. The overall sensitivity in 2-D acquisition is
2–3% at best.
To improve further the sensitivity of PET scanners, the three-dimensional (3-D)
acquisition is employed in which the septa are retracted, or they are not included in
the scanner (Fig.13.17b). In this mode, all events detected by detectors in coincidence in all rings are counted, including random and scatter events, and the sensitivity in the 3-D mode increases four- to eightfold over 2-D acquisition. The incidences

232
ab
a
13 Positron Emission Tomography
Detector 1
b
c
Fig. 13.16 (a) True coincidence events. (b) Random coincidence events detected by two detec-
tors connected in coincidence along the dotted line. The two 511-keV photons originate from different positron annihilations. (c) Scattered coincidence events. One or both of the 511 keV photons
from the same annihilation event may be scattered with little loss of energy and may fall within the
PHA window and also within the coincidence time window to be detected as a coincidence event
by two detectors
Fig. 13.17 (a) 2-D data
acquisition with the septa
placed between the rings
so that true coincidence
counts are obtained
avoiding random events
and scatters. Detectors
connected in the same ring
give direct plane events.
However, detectors are
connected in adjacent rings
and cross plane data are
obtained as shown. (b)
When septa are removed,
the 3-D data acquisition
takes place, which includes
random and scatter events
along with true events.
(Reprinted with the
permission of the
Cleveland Clinic
Foundation)

13.13 Factors Aecting PET
233
of random and scatter can be reduced by having a smaller angle of acceptance; that
is, a detector is connected to a fewer number of opposite detectors.
13.12 Image Reconstruction
Image reconstruction of 2-D PET data is accomplished by the same ltered backprojection and iterative methods that have been described in detail under SPECT in
Chap. 12. In PET, the LORs in a sinogram are back-projected by the Fourier method.
In the iterative method, the projections are estimated by determining the weighted
sum of the activities in all pixels along an LOR across the estimated image and then
compared with the measured projection.
The reconstruction of images from 3-D data is complicated by a very large volume of data, especially in a multiring scanner. The direct application of the ltered
backprojection and iterative methods to these data is difcult, and so the 3-D sinogram data are rebinned into a set of 2-D equivalent projections by assigning axially
tilted LORs to transaxial planes intersecting them at their axial midpoints. This
method is called the single slice rebinning method (SSRB). In another method,
called the Fourier rebinning (FORE) method, rebinning is performed by applying
the Fourier method to each oblique sinogram in the frequency domain. This method
is more accurate than the SSRB method because of the more accurate estimate of
the source axial location. After rebinning of 3-D data into 2-D data, either the ltered backprojection or iterative method is applied.
13.13 Factors Affecting PET
As in gamma cameras, PET acquisition data are affected by photon attenuation,
variation in detection efciency of the detectors, scatter coincidences, partial volume effect, and dead time. These factors are already discussed under SPECT and
therefore only subtle points pertinent to PET data will be highlighted here. In addition, PET data are affected by some unique factors such as random coincidence and
parallax error (radial elongation), which will be discussed below.
13.13.1 Normalization
There are 10,000–32,000 detectors arranged in blocks, and coupled to several hundred PM tubes in modern PET scanners. Practically, as in gamma cameras, the
detection efciency varies from detector pair to detector pair due to variation in the
gain of PM tubes and the location of the detector in the block, resulting in nonuniformity of the PET data. Data are corrected for this factor by using what is called
normalization or uniformity correction. In the normalization of PET data, all detectors are exposed uniformly to a 511-keV photon source (e.g.,
68
Ge source), without

234
F
A
i
ii
Pe ee e
ab
D
Pe
n
ii
13 Positron Emission Tomography
an object in the eld of view, and data are collected in the 2-D or 3-D mode. The
normalization factors Fi are calculated for individual pixels as
mean
=
i
A
(13.3)
where A
is the mean of all pixel counts and Ai is the count in the ith pixel. The
mean
observed count Ci in the ith pixel from the patient is then normalized by
where C
normi
,
is the normalized count in the ith pixel. The normalization data collec-
norm, i
(13.4)
tion requires a long time (~6–8h) and is normally carried out overnight. These
factors are obtained weekly or monthly, and most vendors offer algorithms to obtain
them routinely.
13.13.2 Photon Attenuation Correction
Chang method: The two 511-keV photons in PET can traverse different thicknesses
of tissues before detection and are attenuated to a different degree similar to situations discussed under SPECT.If the two photons traverse a and b thicknesses of
tissues of an organ (Fig.13.18), then the attenuation correction P for each pixel (i.e.,
each LOR) is given by
ab
where μ is the linear attenuation coefcient of 511-keV photons in tissue and D is
the total thickness of the organ. When photons traverse various organs, differences
in linear attenuation coefcients and organ thicknesses must be taken into consideration. Then Eq. (13.5) becomes
D
i
1
(13.5)
(13.6)
As in SPECT, Eq. (13.5) has been employed to correct for attenuation in brain
PET imaging, based on the assumption of uniform density of tissue and a constant
μ for 511-keV photons in tissue (Chang method). However, the method tends to
cause artifacts due to underestimation of attenuation in the thorax area.
Transmission scan method: The transmission scan method was widely used for
attenuation correction of PET emission scan data before PET/CT was introduced.
The technique is similar to the one described in Chap. 12 under Attenuation
Correction Methods. In this method, normally, a 68Ge source is used to obtain the
transmission scan in dedicated PET imaging. The source is placed in a holder
mounted at the edge of the scanner bore, and the holder is rotated by a motor so that
data detected by all detector pairs can be acquired. Normally, a blank scan is
obtained at the beginning of the day without any object or patient in the scanner.
These blank scan data are used for all subsequent patients for the day. Next a

Det
1
13.13 Factors Aecting PET
235
D
ector 2
Fig. 13.18 Two 511-keV annihilation photons traverse thicknesses a and b of tissues of an organ.
However, attenuation of the two photons depends on the total thickness D of the organ regardless
of a and b
*
b
a
Detector
transmission scan is obtained with the patient in the scanner for each patient. The
ratios of the counts in each pixel (i.e., each LOR) between the blank scan and the
transmission scan are calculated for each patient. The emission scan is taken with
the patient in the same position as in the transmission scan to minimize errors in
correction factors, and then each LOR datum is corrected for attenuation by applying the corresponding ratio. It takes 20–40min for acquisition of the transmission
scan, depending on the source strength. Normally, the transmission scan is performed before the emission scan to avoid interference of radiation from the administered radioactivity. Other approaches include post-injection transmission scanning
(transmission scan after the emission scan) and simultaneous emission/transmission
scanning, but each method suffers from various disadvantages of its own.
CT transmission scan method: In PET/CT, the CT transmission scan is utilized
for attenuation correction, which takes less than a minute, thus improving the patient
throughput, and it has become the method of choice. As mentioned in the SPECT/
CT technique, typically at the beginning of the day, a blank CT scan without the
patient in the scanner is obtained, which is later used for subsequent patient studies
for the day. Next, the CT transmission scan of each patient is taken with the patient
in the CT scan eld before the emission scan, and an attenuation correction map is
generated from the ratios of counts of each pixel (i.e., LOR) of the blank scan and
the transmission scan. Because the PET and CT units are xed on the same gantry,
the patient remains in the same position on the table, which is then moved to the
PET scan eld for the emission scan. Factors from the attenuation map are

236
13 Positron Emission Tomography
subsequently applied to each LOR in the patient’s emission scan. The CT transmission method provides essentially noiseless PET images.
Attenuation depends on photon energy; therefore, correction factors derived
from ~70-keV CT x-ray scans must be scaled to the 511-keV photons of PET by
applying a scaling factor dened by the ratio of the mass attenuation coefcient of
511-keV photons to that of 70-keV photons. This factor is assumed to be the same
for all tissues except bone, which has a slightly higher mass attenuation coefcient.
As mentioned in the SPECT/CT section, the respiratory motion of the thorax during
scanning and intravenous contrast agents affect the CT attenuation factors. Use of
breath hold and water-based contrast agents helps mitigate these effects.
13.13.3 Attenuation Correction inPET/MR
Unlike in PET/CT imaging, MR does not have radiation to be attenuated and also a
transmission source cannot be installed because of the space constraint in PET/MR
scanners, so a new approach is needed to apply attenuation correction (AC) to PET
images in PET/MR.An inherent difculty arises from the fact that air and bone do not
produce any MR signal, whereas soft tissues do. Several approaches have been proposed mostly for brain imaging, although methods for torso imaging are evolving. A
common technique for AC in PET/MR is segmentation of MR images. In this method,
a transmission scan is obtained using a 68Ge source or a CT scan to generate an attenuation map that is coregistered with the MR image (commonly T1-weighted images,
which are best for delineating anatomy). The MR image is then segmented in different
tissue types such as bone, uid, air in paranasal sinuses, and brain tissues. Appropriate
linear attenuation coefcient values (μ) at 511 keV are then assigned to these tissues.
In another plausible approach, an atlas-based method is used, in which an atlas
comprises a template MR image together with a corresponding attenuation label
image. A template MR image is generated from the average of co-registered MR
images from multiple subjects (atlas). The label image could be a MR image segmented into different tissue classes (e.g., air, bone, and soft tissue) or a coregistered
attenuation map from a PET transmission scan or a CT scan with continuous attenuation values. The template MR image is coregistered with the MR image of a patient,
and when the transformation is applied to the atlas attenuation image, a patientspecic attenuation map is obtained. Although these methods offer reasonable attenuation correction for brain imaging, it is quite complicated and challenging to apply
them in whole body imaging. The issue is further complicated by truncation effects
extending beyond the MR eld, presence of MR surface coil in PET imaging eld, etc.
13.13.4 Random Coincidences
Random coincidence events occur when two unrelated 511-keV photons, arising
from two different positron annihilation events, are detected by a detector pair
within the same time window (Fig.13.16b). Random coincidences are largely minimized in 2-D acquisition by septa, whereas in 3-D acquisition in the absence of
Соседние файлы в папке Библиотека им академика М.И. Перельмана
