Добавил:
Upload Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
ECG Interpretation Made Incredibly Easy (5th edition).pdf
Скачиваний:
111
Добавлен:
11.03.2016
Размер:
10.76 Mб
Скачать

A LOOK AT ICDS

197

 

Explain that the patient may be hospitalized for 24 to 48 hours to monitor his heart rhythm.

Provide pacemaker teaching if the patient had a pacemaker inserted. (For more information about pacemaker teaching, see “What to teach the patient,” page 190.)

A look at ICDs

An implantable cardioverter-defibrillator (ICD) is an electronic device implanted in the body to provide continuous monitoring of the heart for bradycardia, ventricular tachycardia, and ventricular fibrillation. The device then administers either paced beats or shocks to treat the dangerous arrhythmia. In general, ICDs are indicated for patients for whom drug therapy, surgery, or catheter ablation has failed to prevent the arrhythmia.

The procedure for ICD insertion is similar to that of a permanent pacemaker and may be inserted in a cardiac catheterization or electrophysiology laboratory. Occasionally, a patient who requires other cardiac surgery, such as coronary artery bypass, may have the device implanted in the operating room.

What it is

An ICD consists of a programmable pulse generator and one or more leadwires. The pulse generator is a small battery-powered computer that monitors the heart’s electrical signals and delivers electrical therapy when it identifies an abnormal rhythm. The leads are insulated wires that carry the heart’s signal to the pulse generator and deliver the electrical energy from the pulse generator to the heart.

Storing and retrieving information

An ICD also stores information about the heart’s activity before, during, and after an arrhythmia, along with tracking which treatment was delivered and the outcome of that treatment. Devices also store electrograms (electrical tracings similar to ECGs). With an interrogation device, a practitioner or technician can retrieve this information to evaluate ICD function and battery status and to adjust ICD system settings.

Automatic response

Today’s advanced devices can detect a wide range of arrhythmias and automatically respond with the appropriate therapy, such as bradycardia pacing (both singleand dual-chamber), antitachycardia pacing, cardioversion, and defibrillation. ICDs that provide

NONPHARMACOLOGIC TREATMENTS

198

Types of ICD therapies

Implantable cardioverter-defibrillators (ICDs) can deliver a range of therapies depending on the arrhythmia detected and how the device is programmed. Therapies include antitachycardia pacing, cardioversion, defibrillation, and bradycardia pacing.

Therapy

Description

 

 

Antitachycardia

A short series of rapid, electrical pacing pulses are used to

pacing

interrupt ventricular tachycardia (VT) and return the heart to

 

its normal rhythm.

Cardioversion

A lowor high-energy shock (up to 35 joules) is timed to the

 

R wave to terminate VT and return the heart to its normal

 

rhythm.

Defibrillation

A high-energy shock (up to 35 joules) to the heart is used to

 

terminate ventricular fibrillation and return the heart to its

 

normal rhythm.

Bradycardia pacing

Electrical pacing pulses are used when the heart’s natural

 

electrical signals are too slow. ICD systems can pace one

 

ventricle (VVI pacing) of the heart at a preset rate. Some sys-

 

tems will sense and pace both chambers (DDD pacing).

therapy for atrial arrhythmias, such as atrial fibrillation, are also available. (See Types of ICD therapies.)

How it’s programmed

When caring for a patient with an ICD, it’s important to know how the device is programmed. This information is available through a status report that can be obtained and printed when the practitioner or specially trained technician interrogates the device. This involves placing a specialized piece of equipment over the implanted pulse generator to retrieve pacing function. If the patient experiences an arrhythmia or the ICD delivers a therapy, the program information recorded helps to evaluate the functioning of the device.

Program information includes:

type and model of the ICD

status of the device (on or off)

A tiny computer in the ICD tracks information before, during, and after an arrhythmia.

A LOOK AT ICDS

199

 

detection rates

therapies that will be delivered: bradycardia pacing, antitachycardia pacing, cardioversion, and defibrillation.

Arresting developments

Follow these guidelines if your patient experiences an arrhythmia:

If the patient experiences cardiac arrest, initiate cardiopulmonary resuscitation (CPR) and advanced cardiac life support.

If the ICD delivers a shock while you’re performing chest compressions, you may also feel a slight shock. Wear gloves to prevent shock.

Externally defibrillating the patient is safe as long as the paddles aren’t placed directly over the pulse generator. The anteroposterior paddle position is preferred.

Be alert

Watch for signs of a perforated ventricle, with resultant cardiac tamponade. Ominous signs include persistent hiccups, distant heart sounds, pulsus paradoxus, hypotension accompanied by narrow pulse pressure, increased venous pressure, distended jugular veins, cyanosis, decreased urine output, restlessness, and complaints of fullness in the chest. Report any of these signs immediately to the practitioner and prepare the patient for emergency surgery.

Assess the area around the incision and report swelling, tenderness, drainage, redness, unusual warmth, or hematoma.

What to teach the patient

Explain to the patient and his family why an ICD is needed, how it works, potential complications, and what they can expect. Make sure they also understand ICD terminology.

Discuss signs and symptoms to report to the practitioner immediately.

Advise the patient to wear a medical identification bracelet indicating ICD placement.

Educate family members in emergency techniques (such as dialing 911 and performing CPR) in case the device fails.

Explain that electrical or electronic devices may cause disruption of the device.

Warn the patient to avoid placing excessive pressure over the insertion site or moving or jerking the area until the postoperative visit.

NONPHARMACOLOGIC TREATMENTS

200

Tell the patient to follow normal routines as allowed by the practitioner and to increase exercise as tolerated.

Remind the patient to carry information regarding his ICD at all times and to inform airline clerks when he travels as well as individuals performing diagnostic functions (such as computed tomography scans and magnetic resonance imaging) about the device.

Stress the importance of follow-up care and checkups.

That’s a wrap!

Nonpharmacologic treatments review

Pacemaker

A device that electrically stimulates the myocardium to depolarize

Atrial and ventricular stimulation

Atria: spike followed by a P wave and the patient’s baseline QRS complex and T wave

Ventricles: spike followed by a QRS complex and T wave

Atria and ventricles: first spike followed by a P wave, then a spike, and then a QRS complex

Pacemaker codes

First letter identifies the heart chambers being paced.

Second letter signifies the heart chamber sensed by the pacemaker.

Third letter refers to the pacemaker’s response to the intrinsic electrical activity of the heart.

Fourth letter refers to the pacemaker’s rate modulation.

Fifth letter refers to the pacemaker’s location or absence of multisite pacing.

Pacemaker modes

AAI = single-chambered pacemaker; paces and senses the atria

VVI = paces and senses the ventricles

DDD = fires when the ventricle doesn’t respond on its own; paces the atria when the atrial rate falls below the lower set rate; senses and paces both atria and ventricles

Evaluating pacemakers

Determine the pacemaker’s mode and settings.

Review the patient’s 12-lead ECG.

Select a monitoring lead that clearly shows the pacemaker spikes.

Interpret the paced rhythm.

Look for information that tells which chamber is paced and about the pacemaker’s sensing ability.

Troubleshooting pacemakers

Failure to capture: indicated by a spike without a complex

Failure to pace: no pacemaker activity on the ECG

A LOOK AT ICDS

201

 

Nonpharmacologic treatments review (continued)

Undersensing: spikes where they don't belong

Oversensing: won’t pace when the patient actually needs it

Useful for atrial tachycardia, atrial fibrillation and flutter, ventricular tachycardia, AV nodal reentry tachycardia, and Wolff- Parkinson-White syndrome

Biventricular pacemaker

Pacemaker has three leads: one to pace the right atrium, one to pace the right ventricle, and one to pace the left ventricle.

Both ventricles contract simultaneously, increasing cardiac output.

Candidates

For treatment of patients with:

class III and IV heart failure, with both systolic failure and ventricular dysynchrony

symptomatic heart failure despite maximal medical therapy

QRS complex greater than 0.13 second

left ventricular ejection fraction of 35% or less

Benefits of biventricular pacing

Improves symptoms and activity tolerance

Reverses left ventricular remodeling and diastolic function, and reduces sympathetic stimulation

Radiofrequency ablation

Invasive procedure that uses radiofrequency energy to destroy heart tissue or conduction pathway responsible for arrhythmia

Types of ablation

Targeted ablation

Pulmonary vein ablation

AV nodal ablation (with pacemaker insertion)

Accessory pathway ablation

Implantable cardioverter-defibrillator

Implanted device that monitors for bradycardia, ventricular tachycardia, and ventricular fibrillation

Provides shocks or paced beats to break arrhythmia

Types of therapy

Antitachycardia pacing: bursts of pacing interrupt ventricular tachycardia

Cardioversion: shock timed to the R wave to terminate ventricular tachycardia

Defibrillation: shock to terminate ventricular fibrillation

Bradycardia pacing: pacing when bradycardia is present

Programming information

Type and model

Status of the device (on or off)

Therapies to be delivered: antitachycardia pacing (bradycardia pacing, cardioversion, defibrillation)

NONPHARMACOLOGIC TREATMENTS

202

Quick quiz

1. When using a transcutaneous temporary pacemaker, the energy level should be set at:

A.the highest milliampere setting the patient can tolerate.

B.the lowest milliampere setting that ensures capture of the myocardium.

C.a milliampere setting midway between the setting that causes capture of the myocardium and the setting at which symptoms first appear.

D.the milliampere setting that provides a heart rate of 80 beats per minute.

Answer: B. Select the lowest milliampere setting that causes capture of the myocardium. Higher energy levels will be too irritating for the patient.

2.Failure to capture is represented on the ECG as:

A.no pacemaker activity.

B.spikes where they shouldn’t be.

C.a spike on a T wave.

D.a spike without a complex.

Answer: D. A spike without a complex indicates the pacemaker’s inability to capture or stimulate the chamber.

3. The first letter in the five-letter coding system for pacemakers identifies the:

A.chamber in which the pacemaker senses intrinsic activity.

B.heart chamber being paced.

C.pacemaker’s response to the intrinsic electrical activity.

D.pacemaker’s response to a tachyarrhythmia.

Answer: B. The first letter identifies the heart chamber being paced, the second letter signifies the heart chamber that senses intrinsic activity, the third letter shows the pacemaker’s response to that activity, the fourth letter describes the rate modulation, and the fifth letter shows the location or absence of multisite pacing.

4. A patient with a biventricular pacemaker will have which lead configuration?

A.A lead in both ventricles only

B.A lead in each atria and each ventricle

C.A lead in the right atrium and each ventricle

D.A lead in both atria and the left ventricle

QUICK QUIZ

203

 

Answer: C. A biventricular pacemaker has three leads: one to pace the right atrium, one to pace the right ventricle, and one to pace the left ventricle.

5. After ablation of the AV node, the patient may need a pacemaker because:

A.impulses can no longer be conducted from the atria to the ventricles.

B.the sinoatrial node no longer initiates an electrical impulse.

C.accessory pathways now conduct impulses from the atria to the ventricles.

D.the AV node begins to fire at its intrinsic rate.

Answer: A. After AV nodal ablation, a pacemaker is needed because the electrical impulses from the atria are no longer conducted through the AV node to the ventricles.

6. Which therapy is the patient with an ICD receiving when the device delivers a series of small, rapid electrical pacing pulses?

A.Bradycardia pacing

B.Defibrillation

C.Cardioversion

D.Antitachycardia pacing

Answer: D. Antitachycardia pacing consists of a series of small, rapid electrical pacing pulses used to interrupt ventricular tachycardia and restore normal sinus rhythm.

Test strip

Time to try out a test strip. Ready? Go!

7. In the following ECG strip, the pacemaker is pacing and sensing the ventricles with 100% capture. The mode of response can’t be evaluated because of lack of intrinsic activity. You would determine that the patient has:

A.a VVI pacemaker.

B.a DDI pacemaker.

C.an AAI pacemaker.

D.an AOO pacemaker.

Answer: A. The patient has a VVI, or demand, pacemaker, which inhibits pacing when it senses ventricular activity.

NONPHARMACOLOGIC TREATMENTS

204

Scoring

If you answered all seven questions correctly, all right! You’ve taken the lead when it comes to nonpharmacologic treatment methods.

If you answered six questions correctly, terrific! Keep up the good pace.

If you answered fewer than six questions correctly, no sweat. Go with your impulse to reread the chapter and you’re sure to do better next time.

Соседние файлы в предмете [НЕСОРТИРОВАННОЕ]