Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5814_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
02.09.2026
Размер:
21 Мб
Скачать
199Chapter seven: Contrast media and harmonic imaging
Renal tumor
Perfusion Time (s)
y
OpticalImageofaBurstingBubble
34
27
20
13
7
0
Figure 7.18 Rell time of a kidney estimated by quantifying echogenicity of the organ following the injection of a contrast agent. The center of the kidney has a tumor. Its rell time is very different from those of the surrounding structures. (Courtesy of Prof. Kathy Ferrara at University of California at Davis.)
passively to an organ of interest, or it may actively seek out the targeted tissue by binding it with specic molecules that interact only with a cer­tain group of molecules in a tissue (Lanza et al., 2000; Ferrara et al., 2007). Figure7.19 shows a sequence of optical images on the left acquired by a high-speed camera as a bubble bursts, maybe induced by insonication. The electron micrograph on the right shows a bubble with an oil shell that
0.5 µm
5 µm
EMimageofdrugdeliver
vehiclewiththickoillayer
designedtocarrydrugs
0 0.4 0.8 1.2
1.6 2.0 2.4
Time (µsec)
Figure 7.19 Snapshots of the rupture of a bubble at different times, photo­graphed by a high-speed camera. (Courtesy of Prof. Kathy Ferrara at University of California at Davis.)
200 Diagnostic ultrasound: imaging and blood ow measurements
can be loaded with a drug to be released upon ultrasound insonication. In this case, the bubbles are passively carried by the blood stream to the site of interest. The encapsulated bubble may also be conjugated with a target­ing ligand, which targets a certain molecule on the endothelial surface so that the bubble actively seeks the molecule and binds with it (Figure7.15). The drug that the bubble carries is then released upon insonication.
References and Further Reading Materials
Averukiou MA. Tissue harmonic imaging. IEEE Ultrasonics Symp Proc 2000; 2:
1563–1572.
Chang PP, Shung KK. Interaction of ultrasound with contrast agents. In Thomsen
HH, Muller RN, and Mattery RF (eds.), Trends in contrast agents. Berlin: Springer, 1998.
Crum LA and Prosperetti A. Nonlinear oscillations of gas bubbles in liquids:
An interpretation of some experimental results. J Acoust Soc Am 1983; 73: 121–127.
de Jong N. Improvements in ultrasound contrast agents. Eng Med Biol Mag 1996;
15: 72–82.
de Jong N, Cornet R, and Lancee CT. High harmonics of vibrating gas-lled micro-
spheres. Part I. Simulations. Ultrasonics 1994; 32: 447–453.
Ferrara K, Pollard R, and Borden M. Ultrasound microbubble contrast agents:
Fundamentals and application to gene and drug delivery. Annu Rev Biomed Eng 2007; 9: 415–447.
Gessner R and Dayton P. Advances in molecular imaging. Mol Imaging 2010; 9:
117–127.
Goldberg BB, Liu JB, and Forsberg F. Ultrasound contrast agents: A review.
Ultrasound Med Biol 1994; 20: 319–333.
Lanza G, Hall C, Scott, M, Fuhrhop R, March J, and Wickline S. Molecular imaging
with targeted ultrasound contrast agent. IEEE Ultrasonics Symp Proc 2000; 2:
1917–1926. Leighton TG. The acoustic bubble. San Diego: Academic Press, 1994. Medwin H. Counting bubbles acoustically: A review. Ultrasonics 1977; 15: 7–13. Morse PM and Ingard KU. Theoretical acoustics. New York: McGraw Hill, 1968. Plesset MS. The dynamics of cavitation bubbles. J Appl Mech 1949; 16: 277–282. Rayleigh L. On the pressure developed in a liquid during the collapse of a spherical
cavity. Phil Mag 1917; 34: 94–98. Shi W and Forsberg F. Ultrasonic characterization of the nonlinear properties of
contrast microbubbles. Ultrasound Med Biol 2000; 26: 93–104. Wilson SR and Burns P. Microbubble-enhanced US in body imaging: What role?
Radiology 2010; 257: 24–39.
chapter 8
Intracavity and high-frequency (HF) imaging
Conventional ultrasonic imaging systems typically use frequencies from 2 to 15 MHz. Scanners at lower frequencies have the advantage of a larger depth of penetration but suffer from poorer resolution. To improve spa­tial resolution, one obvious strategy would be to increase the frequency. The axial resolution is determined by the pulse duration or bandwidth of the pulse. For a xed number of cycles per pulse, an increase in fre­quency would result in a reduction in wavelength and thus pulse dura­tion. The relationship between frequency (wavelength) and lateral spatial resolution is given by Equation (3.29). These relationships are graphically illustrated in Figures8.1 and 8.2. As ultrasound frequency is increased to 50 MHz, and an axial resolution and lateral resolution of better than 20 and 100 μm for an f# of 2.9 can be achieved, respectively. The price to be paid is an increase in attenuation. The effect of attenuation coefcient of a few types of tissues of clinical interest is shown in Figure8.3. At 50 MHz, the depth of penetration for most tissues would be limited to 4–5 mm. Although, as was discussed, the depth of penetration may be increased slightly by introducing novel signal processing methods such as coded excitation, the range of frequencies that can be applied to a certain organ is limited.
8.1 Intracavity imaging
Intracavity imaging such as transesophageal, transrectal, and transvagi­nal imaging is a partial solution to achieving improvements in spatial res­olution. Since imaging organs like the heart, prostate, and uterus/ovary from the body surface does not usually allow the utilization of frequen­cies higher than 5 MHz because they are deep-lying organs, probes may be modied to be inserted through open cavities of the body to be placed closer to these organs to allow higher frequencies to be used.
8.1.1 Transesophageal cardiac imaging
Phased arrays consisting of more than 48 elements at frequencies from 5 to 7.5 MHz can be mounted on the tip of an endoscope with a diameter
201
202 Diagnostic ultrasound: imaging and blood ow measurements
Axial Resolution vs. Bandwidth
100
80
60
m)
40
axial
R
20
0
0255075 100 125 150
Bandwidth (MHz)
Figure 8.1 Calculated axial resolution as a function of bandwidth.
less than 10 mm for imaging the heart from the esophagus. The endo­scope allows the manipulation of the position and direction of the phased array. During scanning, the transesophageal probe is inserted into the esophagus and the tip is positioned against the wall of the esophagus under local anesthesia. A majority of the probes are capable of biplane imaging; i.e., two orthogonal images are produced. More advanced ver­sions can produce images in any direction by mechanically rotating the array. The most advanced probes have 2D arrays, allowing 3D imaging
Laterial Resolution vs.
300
250
200
(µm)
150
laterial
R
100
50
0
0255075 100 125 150
Center Frequency
f# = 2.9
f# = 0.7
Center Frequency f0 (MHz)
Figure 8.2 Calculated lateral resolution as a function of ultrasound center freque nc y.
203Chapter 8: Intracavity and high-frequency (HF) imaging
Depth (mm)
At 50 MHz α α α α
≈ 2.5 dB/mm
blood
≈ 1.1 dB/mm
cornea
≈ 1.7 dB/mm
iris
≈ 10 dB/mm
skin
0.9
0.8
0.7
0.6
0.5
0.4
0.3
Trasnfer Function
0.2
0.1
Transfer Function vs. Depth at
1
0
0 0.5 1 1.5 2
50 MHz
Cornea Skin section Blood Iris
Figure 8.3 The effect of ultrasound attenuation of four different types of tissues plotted as a function of depth at 50 MHz. The transfer function is given by e
βz
where β is the attenuation coefcient and z is the propagation depth.
of the heart in real time, which will be discussed in Chapter 8 in more detail. Transesophageal imaging of the heart yields better images of not only the whole heart because of the higher frequencies, but also the base of the heart, which cannot be adequately accessed by transthoracic imaging. An additional benet is that transesophageal imaging allows continuous monitoring of the cardiac functions, which has proven valu­able in anesthesiology during surgery. A photo of such a probe is shown in Figure8.4. Commercial catheters (10 French units, 1 F = 0.33 mm outer diameter) mounted near the tip on the side of a linear array at a frequency of 8 MHz are also available for intracardiac imaging. The catheter can be guided to the heart with a guide wire via a peripheral artery.
,
8.1.2 Transrectal and transvaginal imaging
Probes at frequencies higher than 5 MHz are available for most imag­ing systems for insertion into the rectum or vagina for better imaging of the prostate and uterus/ovary. A linear array or curved linear array is mounted on the side or at the tip of a probe. A full bladder, which used to be recommended for transabdominal obstetrical imaging of a fetus, may now be replaced by transvaginal imaging. A photo of several transrectal and transvaginal probes is shown in Figure8.5.
204 Diagnostic ultrasound: imaging and blood ow measurements
Control housing
Flexible shaft
ew
Enlarged vi of the tip
Figure 8.4 A photo of a transesophageal probe. (Courtesy of Oldelft B.V., Delft, The Netherlands.)
8.1.3 Endoluminal imaging
Catheter-based imaging systems have also been used to image the gas­trointestinal tract, including the colon, esophagus, and stomach (Liu and Goldberg, 1999). A few manufacturers have developed specialized ultra­sonic imaging systems to accomplish the same by mounting ultrasonic transducers/arrays on the end of an endoscope. The frequency of the ultrasound probe ranges from 7 to 20 MHz, with a uid-lled balloon at the tip. Linear arrays and radial arrays (elements mounted on the circum­ference of an endoscope) have been introduced recently.
Figure 8.5 A photo showing several types of ultrasound probes, including linear arrays, linear curved arrays, Doppler probes, and transrectal and transvaginal probes (long and slender ones in the upper half of the photo). (Courtesy of Sound Technology, Inc., State College, PA.)
205Chapter 8: Intracavity and high-frequency (HF) imaging
8.2 Intravascular imaging
Imaging of the wall of blood vessels for the purpose of estimation of the degree of stenosis and characterization of atherosclerotic plaques has been pursued for many years with a variety of imaging modalities (Pandian, 1989; Liu and Goldberg, 1999). X-ray angiography has been the gold standard in the past for assessing stenosis. The drawbacks of x-ray are that (1) it is a form of ionizing radiation involving the injection of a contrast agent, and (2) it is a 2D projection image of a 3D struc­ture. More than two views are necessary to have a more accurate assess­ment of the stenotic vessel. As a result, its role is being challenged by both magnetic resonance imaging and ultrasound (Shung et al., 1992). Plaque composition characterization is of clinical importance in that it has been hypothesized that vulnerable plaques consisting of a lipid core with a brous cap are most likely to rupture, causing the formation of clots and serious clinical consequences, such as stroke and heart attack. Imaging options for characterizing plaque composition are quite limited. Fiberoptic angioscopy, in which an optic ber is introduced via catheter­ization to the site of interest for the visualization of plaque surface, and transcutaneous ultrasound have been used. The former procedure, which involves injection of saline for ushing out light opaque blood, can only visualize the lesion surface, whereas the latter suffers from poor resolu­tion. Intravascular ultrasound and optical coherent tomography (OCT) are possible alternatives to alleviate these problems (Liu and Goldberg, 1999; Bouma and Tearney, 2002). Intravascular ultrasound scanners typi­cally are operated in the frequency range from 20 to 60 MHz, depending upon the imaging catheter used. There are two different types of imag­ing catheters on the market today. In one, a single-element transducer at a frequency from 30 to 60 MHz of 1.75 mm diameter, making an angle of 10° relative to the direction normal to the long axis of the catheter, is mounted near the tip of the catheter. The transducer is mechanically rotated at a very high speed (~1800 rpm) by a shaft. These catheters have a size of 3.5 to 6 F. In another, a 64-element array at 20 MHz having a
1.5λ pitch is mounted around the circumference of a 3.5 F catheter (1.2 mm outer diameter). The elevational width of the array is 0.7 mm. Also mounted on the catheter with the array are several integrated circuit chips that perform the functions of low-noise broadband preamplica­tion and multiplexing. A synthetic imaging approach in which 1 element transmits and 14 elements receive is used to form the image. Figure8.6 shows a catheter with a mechanically rotated transducer (top) and a catheter with an array wrapped around the circumference (bottom). The array catheter, although it has a higher frame rate, yields an image qual­ity slightly inferior to that of the mechanically rotated type, primarily because of its lower frequency.
206 Diagnostic ultrasound: imaging and blood ow measurements
Ca
od
Mechanical Transducer
(b)
–Transducer that rotates on a drive shaft
(a)
Array/Solid State Transducer
Figure 8.6 (a) An intravascular imaging catheter with a mechanically rotated transducer. (b) A catheter with an array wrapped around the circumference. (Courtesy of Boston Scientic and Volcano Therapeutics.)
In addition to plaque characterization, intravascular ultrasound has been found to be useful in guiding the placement of a stent and moni­toring stent restenosis. Figure 8.7 shows a 40 MHz image of an artery, whereas Figure8.8 shows a 20 MHz image of a stent. At present, all intra­vascular ultrasonic imaging devices use side-looking catheters, which lack the capability of visualizing anatomic structures in front of the cath­eter. This capability is important in not causing any injury to the blood vessel itself, while being guided to the site of interest, such as perforation and dissection. Efforts are now underway in developing forward-looking
Figure 8.7 An image of an artery acquired at 40 MHz by a catheter with a rotating single-element transducer. (Courtesy of Boston Scientic.)
Plaque
Blo
theter
207Chapter 8: Intracavity and high-frequency (HF) imaging
Figure 8.8 A 20 MHz imaging catheter is placed at the mid-stent level. The right panel shows a longitudinal view of the stent. The echogenic stent structure is clearly seen. (Courtesy of Volcano Therapeutics.)
intravascular catheters (Degertekin et al., 2006), and a forward-looking device that is a variation of the rotating single-element type is now com­mercially available.
A more recent trend in intravascular imaging is in combining two or more modalities taking advantage of the merits of each (Li et al., 2010). This will be discussed in more detail in Section 9.4.2.
8.3 High-frequency imaging
Scanners operated at frequencies higher than 20 MHz have been devel­oped for applications in ophthalmology, dermatology, and small animal imaging. Typically these devices, called ultrasonic backscatter micro­scope or ultrasonic biomicroscope (UBM), obtain images by scanning a single-element ultrasonic transducer either in a sector format or linearly. The construction of a UBM is identical to that of a static B-mode scanner (Briggs and Arnold, 1996; Pavlin and Foster, 1995). Scanning can also be achieved by better utilizing the focus of the transducer by incrementally moving the transducer in the axial direction, called B-D (D stands for
208 Diagnostic ultrasound: imaging and blood ow measurements
Figure 8.9 A 256-element HF 30 MHz linear array of one-wavelength pitch fabri­cated with conventional dicing technology.
depth) mode scan. A composite image is formed by combing the focused segments of multiple images acquired as the transducer is moved in the axial direction. This mode of scanning improves lateral resolution by sac­ricing frame rate. Commercial UBMs can achieve a frame rate of 30 per second because the excursion range is extremely small. High-frequency scanners that utilize linear arrays have been commercially available for preclinical small imaging since 2009 (Foster et al., 2009). Extensive investi­gations are still being undertaken in developing HF arrays and associated imaging electronics (Ritter et al., 2002; Foster et al., 2009; Cannata et al., 2011; Hu et al., 2011) for preclinical and clinical applications. Conventional dicing technology may be used to fabricate linear arrays up to 50 MHz because dicing blades are only capable of dicing kerfs as small as 15 μm. For arrays of higher than 50 MHz, alternative technology such as micro­machining, which will be discussed in Chapter 9, laser dicing (Foster et al., 2009), or deep reactive ion etching (DRIE) (Yuan et al., 2008) may have to be exploited. A 256-element HF 30 MHz linear array of one-wave­length pitch fabricated with conventional dicing technology is shown in Figure8.9. The array with an aperture size of 2 × 6 mm had a bandwidth of 55% and a cross talk level of –27 dB. The design philosophy is similar to low-frequency arrays with one or more matching layers and a light back­ing. A laser- (excimer at 248 nm) diced 256-element 30 MHz array with a kerf width of 8 μm for a preclinical scanner (VEVO 2100) produced by Visualsonics (Toronto, Canada) is shown in Figure8.10. The correspond­ing images obtained by this array, shown in Figure8.11, clearly demon­strate the superior quality of the image, where all voids in the phantom are visualized, over that obtained with a UBM. This scanner is capable of performing conventional and color Doppler.
There are many clinical applications for high-frequency ultrasound. In ophthalmology, scanners at 20 MHz or slightly lower have been used