Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5814_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
02.09.2026
Размер:
21 Мб
Скачать
1.2
Normalized Pulse Echo Signal
Normalized Pulse Echo Amplitude
(b)
1
0.8
0.6
0.4
0.2
0
050 100 150
1.4
1.2
1
Distance
(a)
77Chapter three: Ultrasonic transducers and arrays
Figure 3.33 The spatial resolutions of an ultrasonic transducer in the axial and lateral directions are determined by the pulse duration and beam width: (a) full width with main lobe separation, (b) half-main lobe width separation, (c) less than half-main lobe width separation, and (d) full width half maximum.
For a rectangular array element, beam widths on the x-z plane and y-z plane are
dx = 2f#xλ and dy = 2f#yλ where f#x = z0x/b and f#y = z0y/h are the f# values on the x-z plane and y-z
plane, respectively.
0.8
0.6
0.4
0.2
0
40 60 80 100 120 140
Distance
1.4
Normalized Pulse Echo Signal
150
Normalized Pulse Echo Signal
(d)
78 Diagnostic ultrasound: imaging and blood ow measurements
1.2
1
0.8
0.6
0.4
0.2
0
40 60 80 100 120 140
Distance
(c)
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
50 100
Distance
Figure 3.33 (Continued) The spatial resolutions of an ultrasonic transducer in the axial and lateral directions are determined by the pulse duration and beam width: (a) full width with main lobe separation, (b) half-main lobe width separation, (c) less than half-main lobe width separation, and (d) full width half maximum.
79Chapter three: Ultrasonic transducers and arrays
Df
The depth of focus Df, that is, within this region the intensity of the beam within –3 dB of the maximal intensity at the focus for a circular aperture and a rectangular aperture, is found to be linearly related to the wavelength (Kino, 1987; McKeighen, 1998).
Df
fc fr
2
7.2 and 7.1
#
2
(3.39)
#
The depth of eld can be dened alternatively as the region between axial distances where the beam widths become 2d.
From these relationships, it is clear that an increase in frequency that decreases wavelength improves both lateral and axial resolutions by reducing the beam width and the pulse duration if the number of cycles in a pulse is xed. Unfortunately, these improvements are achieved at a cost of a shorter depth of focus.
The axial and lateral resolution of a transducer can be improved from an increase in the bandwidth with backing or matching and focusing. The spectrum of an ultrasonic pulse varies as it penetrates into tissue because the attenuation of the tissues is frequency dependent. It is known that the center frequency and bandwidth of an ultrasonic pulse decrease as the ultrasound pulse penetrates deeper. In other words, the axial resolution of the beam worsens as the beam penetrates deeper into the tissue. In com­mercial scanners, pulse shape and duration are maintained by time gain compensation and some form of signal processing.
3.5.5 Focusing
Better lateral resolution at a certain axial distance can be achieved by acoustic focusing. However, an improvement in the lateral resolution or focusing at a certain range is always accompanied by a loss of resolution in the region beyond the focal zone, as illustrated in Figure3.34(a).
The general principles of focusing are identical to those in optics. The two most often used schemes, a lens and a spherical or bowl-type transducer, are illustrated in Figure3.34(a), where zf and Df are, respec­tively, focal distance and depth of focus, and (b). The acoustic lens shown in Figure3.34(a) is a convex lens, which means that the sound velocity in the lens material is less than the medium into which the beam is launched. The convex lens is preferred in biomedical ultrasonic imag­ing in that it conforms better to the shape of the body curvature. This is illustrated in Figure3.35. Some of the lens materials frequently used in medical applications can be found in Table3.4. The most important requirements are that it should have an acoustic impedance similar to the last matching layer and be slightly attenuative to absorb reverberations
80 Diagnostic ultrasound: imaging and blood ow measurements
If Sound Velocity in Lens < Sound Velocity in Water
If S
r
Piezoelectric element
Tr
(b)
am
Convex lens
2a
D
ansducer
Spherically focused element
f
z
f
Focused beam
(a)
Nonfocused be
Figure 3.34 Two modes of focusing that have been used to focus ultrasonic beams: (a) focusing with a lens and (b) self-focusing.
of signals inside the lens. As illustrated in Figure3.36, the focal length zf of a lens is given by
R
c
=
z
f
(3.40)
11/
n
where Rc is the radius of curvature and n = c1/c2, c1 is the velocity in the lens and c2 is being that in the medium. A popular material for convex
Piezoelectric element
Figure 3.35 Convex and concave lenses.
ound Velocity in Lens > Sound Velocity in Wate
Convex lens
Concave lens
81Chapter three: Ultrasonic transducers and arrays
R
c
2a
Figure 3.36 Focusing geometry.
z
f
z
lens is RTV silicon rubber, which has a velocity of 1010 m/s, acoustic impedance of 1.5 MRayl, and attenuation of 7 dB/cm-MHz. For a sili­con rubber lens in water and a focal distance of 4 cm, Rc can be readily calculated to be 2.12 cm from Equation (3.40). Concave lenses made of polyurethane or polystyrene have also been used. For concave transduc­ers a suitable ller material is needed to make the transducer face at. Polyurethane has been shown to t this need. The focal region formed by an acoustic lens is usually ellipsoidal. Its dimension depends on the relationship between wavelength and the diameter of the lens. In gen­eral, the bigger the diameter, the smaller the focal point. Ultrasonic imag­ing is diffraction limited because the beam cannot be properly focused in the region close to the transducer and beyond the near-eld and far-eld transition point. For a circular piston transducer of radius a, z0 = a2/λ. The
f# is a/(2λ), which is determined by the ratio of radius to wavelength. For
a ratio of radius to wavelength = 10, f# = 5. This means that the beam can­not be focused beyond an f# of 5. The only way to obtain focusing at a dis­tance greater than this is to either increase the aperture size or decrease the wavelength.
3.5.6 Protection circuits for transducers
Large signals in the order of 100 peak-to-peak volts are needed to drive a transducer. The receiving electronics must be protected from these high voltages (Lockwood et al., 1991; Cobbold, 2007). Figure 3.37(a) shows a protection scheme that is commonly used. Rs, Cs, and Lc represent source output resistance, capacitance, and inductance for a matching network. The load resistance Rl should be much larger than the input resistance of the receiving electronics, Ri, which is in general 50 Ω. When the source is on, both diodes D1 and D2 appear to be shorted. During transmission and reception the circuit behaves like the circuit shown in Figure3.37(b)
82 Diagnostic ultrasound: imaging and blood ow measurements
D1
V
(c)
R
s
s
C
s
L
c
R
l
Preamplifier
R
i
D2
Transducer
(a)
V
R
s
s
C
s
R
L
l
c
Transducer
(b)
C
s
L
c
R
l
R
i
Transducer
Figure 3.37 (a) A commonly used transducer protection circuit. (b) and (c) Circuit during transmission and reception.
83Chapter three: Ultrasonic transducers and arrays
Zx
Zx
Z
x
Z
in
0
Z
l
Figure 3.38 An electrical cable terminated by load impedance Zl.
and (c). At high frequencies (>20 MHz), the cable length may be changed or tuned to match to Ri. The input electrical impedance to a cable shown in Figure3.38 is given by
ZZ x
=
ZZ
0
in
ZZ x
tanh
0
l
tanh
0
l
(3.41)
where γ is a complex number = α + jβ, α = attenuation coefcient of the cable, β = propagation constant for an electromagnetic wave, and Z0 = characteristic electrical impedance of the cable. Assuming a lossless cable, Equation (3.41) becomes
Zj
=
ZZ
0
in
Zj
tan
0
l
tan
0
l
At 50 MHz the electromagnetic wavelength is 600 cm, and the cable needs to be considered. If Zl is zero, when x is λ/4 = 150 cm in this case,
βx = π/2, tanβx = ∞, and Z
= ∞. It means that the electrical impedance
in
from the transducer looking into the receiving electronics is ∞ or an open circuit. Another observation is that if Zl = Z0, Z
= Z0 for all x. If the trans-
in
ducer output impedance is Z0, the load is matched to the transducer inde­pendent of the cable length.
A single-element transducer can be translated or steered mechani­cally to form an image. Linear translators do not allow movements permitting generation of images at a rate higher than a few frames per second, although there are sector scanning devices that allow steering the transducer within a limited angle at a rate of 30 frames per second. Early real-time ultrasonic imaging devices almost exclusively used this type of transducer, which is called mechanical sector probe. A typical mechanical
84 Diagnostic ultrasound: imaging and blood ow measurements
Single elemen transducer
l
Mechanical sector
Mechanical drive
Oil
t
Figure 3.39 Detailed construction of a mechanical sector probe.
Cable
Probe housing
Coupling ge
Skin
Tissues
sector probe is shown in Figure3.39. The transducer is housed in a dome bathed in some form of oil to facilitate the transmission of the ultrasonic energy from the transducer to the housing. Mechanical sector probes that suffer from poor near-eld image quality because of reverberations between the transducer and the housing and xed focusing capability have now been largely replaced by linear arrays.
3.6 Arrays
Arrays are transducer assemblies with more than one element. These elements may be rectangular in shape and arranged in a line, called lin­ear array or 1D array, shown in Figure 3.40(a), or square in shape and arranged in rows and columns, called two-dimensional (2D) array, shown in Figure3.40(b), or ring shaped and arranged concentrically, called annu­lar array, shown in Figure3.40(c).
A linear switched array (sometimes called a linear sequence or simply a linear array) is operated by applying voltage pulses to groups of ele­ments in succession, as shown in Figure3.41, where the solid line and the dashed line indicate, respectively, the rst and second beams. In this way, the sound beam is moved across the face of the transducer electronically producing a picture similar to that obtained by scanning a single-element transducer manually. The amplitude of the voltage pulses can be uniform or varied across the aperture as shown in the gure by arrows of vary­ing length. As mentioned previously, amplitude apodization or varying the input pulse amplitude across the aperture is sometimes used to sup­press side lobes at the expense of worsening the lateral resolution. If the electronic sequencing or scanning is repeated fast enough (30 frames per second), a real-time image can be generated. Linear arrays are usually
Pitch
X (Azimuth)
(a)
ts
(b)
Annuli
(c)
Kerf
Kerf width
Array element
Y (Elevation)
Array elemen
85Chapter three: Ultrasonic transducers and arrays
Z
Figure 3.40 (a) A linear array. (b) A 2D array. (c) An annular array.
1 cm wide and 10 to 15 cm long with 128 to 256 elements. Typically 32 or more elements are red in a group. For the sake of achieving as good a lateral resolution as possible, the irradiating aperture size must be made as large as possible. The aperture size is in turn limited by the require­ment of maintaining a large number of scan lines. This point will become
86 Diagnostic ultrasound: imaging and blood ow measurements
ts
Array elemen
Direction of electronic scanning
Figure 3.41 An image is formed by a linear array by electronically sweeping the beam. A group of elements are red simultaneously to form one beam.
clearer in Chapter 4. Figure3.42 shows the detailed construction of a lin­ear array consisting of a backing material, a layer of piezoelectric mate­rial sandwiched between two electrodes, and two matching layers. Here a concave lens is used to focus the imaging plane in the elevational direc­tion or the slice thickness of the imaging plane. This is a problem of cru­cial importance in 2D imaging with 1D arrays because the slice thickness cannot be controlled throughout the depth of view. The slice thickness is the thinnest only at the focal point of the lens and becomes worse closer to the array or beyond the focal point. Figure3.43 illustrates how a large slice thickness can cause serious image artifacts, including reduction in contrast. The top panel shows that scatterers outside of the cyst region can cause the echogenicity within the cyst to increase in the eld of view where the slice thickness is large. The lower panel shows that although the location of the cyst farther from the transducer is not on the same imaging plane as the cyst closer to the transducer, an ultrasonic image will not be able to tell the difference.
In Figures 3.40(a) and 3.44, the space between two elements is called a kerf and the distance between the centers of two elements is called a pitch. The kerfs may be lled with acoustic isolating mate­rial or simply air to minimize acoustic cross talk. The kerfs are often cut into the lens and backing to minimize acoustic cross talk between adjacent elements through the backing, the lens, and matching layers. The size of a pitch in a linear array ranges from λ/2 to 3λ/2, where λ is the wavelength in the medium into which ultrasound is launched