Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5814_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
02.09.2026
Размер:
21 Мб
Скачать
138 Diagnostic ultrasound: imaging and blood ow measurements
(a)
(b)
Figure 4.44 Portable scanners from three different manufacturers: (a) Fuji Sonosite (courtesy of Sonosite), (b) pocket scanner with a 3 MHz element-phased array (courtesy of Siemens), and (c) Vscan phased array system with color Doppler (courtesy of GE).
139Chapter four: Gray-scale ultrasonic imaging
(c)
Figure 4.44 (Continued) Portable scanners from three different manufacturers: (a) Fuji Sonosite (courtesy of Sonosite), (b) pocket scanner with a 3 MHz element­phased array (courtesy of Siemens), and (c) Vscan phased array system with color Doppler (courtesy of GE).
Another advancement is the elimination of the cable that connects an ultrasonic probe to the imaging console by adopting wireless technol­ogy. Figure4.45 shows a wireless portable scanner developed by Siemens where the wireless radio frequency (RF) signal is transmitted at 8.7 GHz.
Figure 4.45 A wireless scanner capable of color Doppler. (Courtesy of Siemens.)
140 Diagnostic ultrasound: imaging and blood ow measurements
(a)
(b)
Although ultrasound computed tomographical scanners have only found limited success in breast imaging, a new breast scanner very similar to x-ray mammography has recently been introduced. A unique feature of this scanner is the concave curved linear array probe that conforms better to the breast curvature. The linear array is mechanically scanned linearly to acquire multiple slices of B-mode images for later 3D recon­struction. It was approved by the FDA for dense breast scanning as a com­plementary tool for x-ray mammography. Figure 4.46(a) shows how the
Figure 4.46 (a) Automated breast ultrasonic scanner and (b) corresponding coro­nal image. (Courtesy of GE.)
141Chapter four: Gray-scale ultrasonic imaging
linear mechanical scanning of the linear array is performed on a patient and (b) a coronal B-mode breast image acquired by such a scanner, which shows a lesion at the lower right-hand corner. A coronal view, which is not possible with a conventional ultrasound scanner, can be obtained follow­ing 3D reconstruction.
References and Further Reading Materials
Flax SW and O’Donnell M. Phase-aberration correction using signals from point
reectors and diffuse scatterers: Basic principles. IEEE Trans Ultrasonics Ferroelect Freq Cont 1988; 35: 768–778.
Greenleaf JF. Computerized tomography with ultrasound. Proc IEEE 1983; 71:
330–337.
Karaman M and O’Donnell M. Synthetic aperture imaging for small scale systems.
IEEE Trans Ultrasonics Ferroelect Freq Cont 1995; 42: 429–442.
Nock L and Trahey GE. Phase aberration correction in medical ultrasound using
speckle brightness as a quality factor. J Acoust Soc Am 1989; 85: 1819–1826.
O’Donnell M. Coded excitation system for improving the penetration of real-time
phased array imaging system. IEEE Trans Ultrasonics Ferroelect Freq Cont 1992; 39: 341–351.
Ophir J and Maklad NF. Digital scan converters in diagnostic ultrasound imaging.
Proceedings of the IEEE 1979; 67: 654–664.
Shung KK, Smith MB, and Tui BWN. Principles of medical imaging. San Diego:
Academic Press, 1992.
Shung KK and Thieme GA. Ultrasonic scattering by biological tissues. Boca Raton,
FL: CRC Press, 1993.
Thomenius KE. Evolution of ultrasound beamformers. In Levy M, Schneider SC,
and McVoy BR (eds.), Proceedings of the 1996 IEEE Ultrasonics Symposium, New York, 1996, pp. 1615–1622.
Wagner RF, Smith SW, Sandrik JM, and Lopez H. Statistics of speckle in ultrasound
B-scans. IEEE Trans Soncis Ultrasonics 1983; 30: 156–163.
chapter ve
c
2cos
Doppler ow measurements
As was discussed earlier in Chapter 2, the Doppler effect provides a unique capability for ultrasound to measure blood ow (Evans and McDicken, 2000; Jensen, 1996). Upon insonication by an ultrasound beam, the echoes scattered by blood carry information about the veloc­ity of blood ow. Blood ow measurements are frequently performed in a clinical environment to assess the state of blood vessels and func­tions of an organ. Ultrasonic Doppler instruments allow a measurement of instantaneous blood ow velocity. Combined with pulse-echo instru­ments, instantaneous ow rate in a blood vessel as a function of time and cardiac output can be measured noninvasively with ultrasound. At pres­ent, very few clinical options are available to do so. Figure5.1 shows an ultrasound beam of frequency f insonifying a blood vessel, making an angle of θ relative to the velocity v. Here it is assumed that blood ows in a vessel with a uniform velocity v. The returned echoes are Doppler shifted. The Doppler shift frequency fd is related to the ultrasound frequency f by Equation (2.43):
v
f
=
d
θ
f
where c is the sound velocity in blood and may be assumed to be 1540 m/s. The Doppler-shifted frequencies happen to be in the audio range for blood ow velocities in the human body for an ultrasound frequency between 1 to 15 MHz.
Conventionally, two different approaches have been used for ultra­sonic Doppler ow measurements: continuous-wave (CW) and pulsed­wave (PW) Doppler.
5.1 Nondirectional CW owmeters
A CW system is shown in Figure 5.2. A probe consisting of two piezo­electric elements, one for transmitting the ultrasound signal and one for receiving echoes returned from blood, is excited by an oscillator. The Doppler-shifted echoes are amplied, demodulated, and band pass ltered to remove the carrier frequency and other spurious signals. Suppose that the ultrasound signal generated by the oscillator is given by
143
144 Diagnostic ultrasound: imaging and blood ow measurements
Bt
dd
1 2
()
shift
am
R elemen
n
Transducer
Skin surface
Blood vessel
v
θ
Doppler equatio
2vcosθ
=
f
d
c
f
Figure 5.1 An ultrasound beam is incident upon a blood vessel and makes an angle of θ relative to the direction of blood ow.
Acos(ωt), where A denotes signal amplitude and ω (the angular frequency) = 2πf. The demodulated signal would be
gAtB tA
(, )cos()cos[()]
dd
{cos[(2)]cos
t
}ωω =ω ω+ω= ω+ω+ ω
d
where the echoes are represented by Bcos[(ω + ωd)t] and ωd = 2πfd. The magnitude of constant B is determined by the scattering strength of blood. Much work has been done to better understand the relationship between the Doppler power generated by blood and hematological
Master
Transducer
eceiving
t
Transmitting element
oscillator
Amplifier
Speaker
Doppler sound Average Doppler
Coherent demodulation
Demodulator
Filter
Zero crossing
counter
Figure 5.2 Block diagram of a CW Doppler owmeter.
Spectrum
analyzer
Spectrasonogr
Time domain Frequency domain
Transmitted signal
f
0
t
Received signal
f
0
f
c
t
Demodulated signal
f
t
fc = Frequency of clutter signal e.g., from blood vessel wall
f
c
f
d
145Chapter ve: Doppler ow measurements
f
d
f
f
Figure 5.3 Doppler signals in the time and frequency domain showing the effect of demodulation.
and hemodynamic factors (Shung et al., 1992; Mo and Cobbold, 1993). Doppler power from blood has been found to be related to ow distur­bance, hematocrit, and the degree of red blood aggregation, which is in turn affected by the concentration of plasma proteins such as brino­gen and local shear rate. The output of the demodulator contains both the ultrasound carrier frequency and the Doppler shift, illustrated in Figure 5.3, where the signals in the time and frequency domains are shown, respectively, on the left and right. The carrier signal can be readily removed by band pass ltering by setting the cutoff frequency of the band pass lter at the high end to be much lower than the carrier frequency. A problem in ultrasonic Doppler blood ow measurement is that the blood vessels that produce large reected echoes are slow mov­ing as well. In Doppler terminology, these large, slow-moving echoes are called clutter signals, shown in Figure5.3 as fc. The cutoff frequency of the band pass lter at the low end has to be designed to minimize the interference of these clutter signals. The design of this band pass lter in the low-frequency region, which performs the function of high pass,
146 Diagnostic ultrasound: imaging and blood ow measurements
0
High pass or wall filter response
f
c
Figure 5.4 Clutter rejection lter or wall lter is used to suppress large echoes produced by slow-moving blood vessel walls. fc = frequency of clutter signal (e.g., from blood vessel wall).
f
d
f
also called clutter rejection lter, has been problematic due to the fact that the magnitude of clutter signals is several orders higher than those from blood and may mask those from slow-moving blood (Figure5.4). A lter with a very steep slope or a method that carries out some forms of echo cancellation may be used (Jensen, 1996). The signal after band­pass ltering can be processed in different ways. It may be heard with a speaker since the Doppler shift is in the audible range. Alternatively, a zero-crossing counter can be used to estimate the mean Doppler fre­quency, or a spectrum analyzer can be used to display the spectrum. The zero-crossing counter estimates the number of zero-crossings of a signal. The number of zero-crossings, N, and the mean frequency, fm, of a signal are given, respectively, by
2
fPfdf
()
N
0
2
=
Pfdf
0
()
(5.1)
fP fdf
()
0
f
=
m
Pfdf
(5.2)
()
where P(f) is the probability density function at frequency f. For a pure sinusoidal signal of frequency fm, N = 2fm. Complication arises if the sig­nal is not sinusoidal, as in the case of Doppler ow measurements, where
147Chapter ve: Doppler ow measurements
=−
r
Cross-sectional view
of the vessel
R
r
Figure 5.5 Laminar blood ow in an artery has a parabolic ow prole.
v(r)
v
max
the blood ow is not uniform. For blood ow in a vessel, the velocity is related to radial distance r, shown in Figure5.5, by the following equation (Nichols and O’Rourke, 1990):
n
r
(5.3)
R
where v
vr v
() 1
max
is the peak velocity, n is an index indicating the nature of ow,
max
and R is the radius of the blood vessel. For parabolic ow, n = 2 and N =
1.15 f
, where f
max
is the maximal Doppler frequency.
max
The spectrum is usually displayed in the format shown in Figure5.6(a), where the vertical axis indicates Doppler frequency or velocity, the hori­zontal axis indicates time, and the gray scale indicates the intensity of the Doppler signal at that frequency or velocity. Figure5.6(b) illustrates how one vertical line in the Doppler sonogram is calculated and the complete Doppler sonogram obtained. At each instant of time, the vertical line dis­played represents the Doppler spectrum calculated at that time within a 5 to 10 ms time window. For a timescale of many seconds, the Doppler spectrum obtained within a very short window at a certain time can be represented by a line. From the Doppler spectrum, the mean frequency or other frequencies, e.g., median frequency, where the Doppler power spectrum is split into two equal halves, and mode frequency, where the Doppler power is the highest, can be readily estimated.
Doppler owmeters have been used to noninvasively assess vascular disorders. Flow disturbances near a stenosis cause the Doppler spectrum to broaden because blood ow velocity uctuates. There is, however, a caveat that must be recognized to avoid misdiagnoses: transit time spec­tral broadening. This is illustrated in Figure5.7(a) for a single scatterer traversing an ultrasound beam at velocity v. A nite time is needed for the scatterer to traverse the beam. In the time domain, there is a nite time duration, which is dened by Δt (–3 dB time duration from peak value),
148 Diagnostic ultrasound: imaging and blood ow measurements
(a)
Echo amplitudefromamovingstructure
city
(b)
t
1
An FFTisperformedonthewindowedwaveformandacolorscaleisusedto
Figure 5.6 (a) Spectrasonogram of CW Doppler signals produced by a mitral valve regurgitation jet. The top image is a B-mode apical four-chamber view of the heart. The dotted line indicates the direction of the Doppler beam. (Courtesy of GE Medical Systems.) (b) How a spectrasonogram is calculated.
4μs
denotethesignalamplitude
Time
Frequency or velo
t
1
Time