Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_605_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
69 Мб
Скачать
Diagnostic Procedures
Figure 5.23 (A) Normal blood flow of the kidney; (B) complete interruption of blood inflow into the kidney due to thrombosis. All: Courtesy: Dr. E. Matevossian, Klinikum
rechts der Isar.
121
In visceral medicine, it is of outstanding importance to assess the blood flow in the hepatic artery and portal vein (e.g., in case of portal hypertension or after kidney transplantation).
A continuous wave Doppler (CW Doppler) continuously generates and receives US waves. A CW Doppler uses two transducers. One transducer continuously transmits and one transducer continuously receives signals.
In contrast to CW Doppler systems, PW Doppler Systems use a single transducer for transmission and reception. Pulsed wave (PW) Doppler systems transmit short pulses of US into the tissue. The pulse travels for a given time until it is reflected back. It then returns to the transducer over the same time interval, but at a shifted frequency. A timing mechanism controls the range gating that samples the returning Doppler shift data from a given region. Only Doppler shift data from inside that area is dis­played. The transducer alternates transmission and reception of US. Doppler signals can be acquired from a known depth. Today, pulsed Doppler systems for spectral Doppler measurements are typically used in combination with US B-mode imaging, which is known as duplex US. Duplex scanners use arrays of elements to produce both the B-mode image and the Doppler spectrum. This facilitates accurate anatomical location of the blood flow under investigation.
An advantage of CW Doppler systems is their capability for measuring movements with high velocity, which is needed in applications like cardi­ography, where high blood flow rates occur. A CW Doppler cannot pro­vide range resolution because it is unable to separate Doppler signals that arise from different points along the transmitted US beam. If two blood vessels intersect the US beam, it will not be possible to separate velocities
122 Biomedical Engineering in Gastrointestinal Surgery
at the different points along the beam, so it cannot be used to produce color flow images.
One main advantage of pulsed Doppler is its ability to provide Doppler shift data selected from a small segment along the US beam, referred to as the “sample volume.” The location of the sample volume can be controlled by the operator.
Recent Developments and Current Research
There are recent developments in measuring blood flow via low­peak frequency-modulated continuous wave (FMCW) and stepped frequency-modulated continuous waves (s tep-FMCW) US Doppler systems. Low-peak FMCW us es a new demodulation technique: the Doppler signals are demodulated with a reference FMCW signal to adjust delay times so that they are equal to propagation times between the transmitter and the receiver. Doppler signals can be obtained from a selected position, as with a sample volume in PW Doppler systems
[49].
Step-FMCW leads to a high SNR and a high range resolution com­pared to traditional pulse-echo signals. In step-FMCW ultrasonic ranging, the phase and magnitude differences at stepped frequencies are used to sample the frequency domain. Step-FMCW features lower peak power, wider dynamic range, lower noise figure, and simpler electronics in com­parison to pulse-echo systems
[50].
Technologies using the Doppler-effect with pulsed US waves are widely matured. Recent developments are not based on developments of the CW Doppler technology but mostly on combinations of Doppler visualizations with other modifications of US, like elastography.

5.4.5 US Elastography

US elastography is an imaging technique to evaluate the mechanical properties of soft tissue by applying strain on the organ examined and detecting differences in its tissue density and stiffness found useful in showing abnormalities of muscle and breast tissue, and in the detection of tumors
[53].
The stiffness of the liver correlates, e.g., with its content of fibrous tis­sue. A fibrotic liver is harder in the case of fibrosis, and even harder in cirrhosis than a healthy liver. Most malignant tumors are, likewise, harder than the surrounding tissue, a feature which was always used in medicine to detect pathological findings (manual palpation).
[51,52]. It has been
123Diagnostic Procedures
The following techniques of US elastography will be considered in detail below: acoustic radiation force impulse imaging (ARFI), shear wave elastography (SWE), and shear wave dispersion US vibrometry (SDUV).
5.4.5.1 Acoustic Radiation Force Impulse Imaging
ARFI imaging is a form of strain elastography which is mainly used for liver, thyroid, and breast imaging. In this technique, the tissue is excited internally by a focused US pulse. ARFI is based on the principle that as the US pulse passes a tissue, the displacement of soft tissues is larger than the displacement of hard tissues. As a result, this technique offers a qualitative color-coded or grayscale elastogram representing relative tissue stiffness
[51].
Characteristics of the Modification
ARFI transmits a brief acoustic radiation force (0.0031 ms) to gen­erate a localized displacement in tissue. This technique utilizes a single transducer for both transmitting the radiation force and tracking the resulting displacement of tissue.
To obtain displacement information for a spatial location through ARFI imaging, a reference line is first acquired by a conventional US pulse. Afterward, a radiation force impulse is induced in the same location to cre­ate a slight displacement. Thereafter, with the help of conventional US, a series of tracking lines are acquired for monitoring displacement and recov­ery of the tissue. To observe the recovery of tissue as it returns to its origi­nal configuration, approximately 46 ms of tracking are required. The repetition of this reference push-track procedure over other spatial locations allows the creation of a 2D image by aligning the displacements of each location in time relative to its pushing pulse
[54].
Strengths and Weaknesses
A general strength of ARFI is its ability to image deeper tissue, which is not reachable by external compression
[51]. Moreover, ARFI requires
merely slight hardware modifications to add an alternative imaging mode similar to Doppler or M-mode
[54] (Table 5.3).
Table 5.3 Key facts on ARFI Typical
applications
Oncology Endocrinology
Strengths and weaknesses
High depth
penetration
No spatial
resolution
Recent developments
n/a Real-time
Research potential and future trends
scanning
Heat management
124 Biomedical Engineering in Gastrointestinal Surgery
A weakness of ARFI imaging is its inability to identify the tissue’s composition. Nowadays, it is only possible to detect differences in the tissue’s softness or stiffness in comparison with adjacent tissue. A fur­ther limitation is often encountered as a result of the obscur ity of the acoustic force, as its influencing factor s—the acoustic intensity and the degree of attenuation to the acoustic pulse—are normally unknown.
Recent Developments and Current Research
Technical challenges fo r this imaging technique involve the implementation of real-time scanning into a clinical US unit and its heat management. The former may easily be achieved because the imaging sequence of ARFI fits with the software and hardware per­formances of c linical scanners. The latter, which is mainly influ­enced by ARFI pushes, may be addressed by using parallel beam­forming.
5.4.5.2 Shear Wave Elastography
SWE is a type of US elastography that uses shear waves to assess tissue elasticity and display it in a quantitative manner. Unlike acoustic compres­sive waves, which spread in the same direction as the particle compres­sion, the propagation of shear waves proceeds orthogonal to the stimulated displacement. Shear waves can result from an acoustic radiation force, mechanical punch, or external sources imaging
[56].
[55]. SWE is used for breast
Characteristics of the Modification
The attenuation of shear waves is approximately 10,000 times more rapid than conventional US
[51]. SWE provides a quantitative real-time
elastogram, whereby elasticity can be depicted as a superimposition of a color-coded image measured in kPa over a B-mode image. In the images, stiffer tissues appear in red while softer tissues are coded in blue. The image resolution remains around 1 mm
[56].
Strengths and Weaknesses
SWE is currently the only approach with the ability to provide quan-
titative and local elastic information in real time.
A limitation of this technique is the weakness of the generated shear waves due to dissipation, which occurs after spreading a few millimeters. Stronger shear waves need increased US power, which causes overheating in the hardware and concerns over acoustic power
[56] (Table 5.4).
Table 5.4 Key facts on SWE Typical applications
Strengths and weaknesses
Diagnostic Procedures
Recent developments
125
Research potential and future trends
Oncology, e.g.,
liver tumors
Detection of
liver cirrhosis
Figure 5.24 SWE: Focal nodular hyperplasia of the liver (A) as compared to normal tissue (B). Courtesy: Prof. K. Stock, Klinikum rechts der Isar.
Quantitative and
local elasticity information
Low power of
shear waves
n/a Increasing
acoustic power
Currently, the detection of liver diseases is the main clinical applica-
tion (
Fig. 5.24).
Recent Developments and Current Research
Future challenges must deal with creating an upswing in the shear waves’ amplitude in order to increase their ability to travel through tissue while still limiting the acoustic power to safe levels.
5.4.5.3 Shear Wave Dispersion Ultrasound Vibrometry
In contrast to other imaging techniques mentioned earlier, shear wave dispersion ultrasound vibrometry (SDUV) quantifies not only the elastic­ity of tissue, but also its viscosity. This technique utilizes shear wave prop­agation speed, which is measured in tissue at multiple frequencies within the range of hundreds of Hertz. One sample application of this technique is liver fibrosis staging.
126 Biomedical Engineering in Gastrointestinal Surgery
In general, SDUV is a fast imaging technique, especially when
repeated pulses are used. Accordingly, only 50200 ms are needed to achieve measurements
[57].
Unlike ARFI, which is based on transient shear waves, SDUV uses periodic shear waves and the dispersion of their velocity to qualify viscos­ity. The estimation of shear wave speed is based on the phase differences and multiple cycles of shear wave vibration.
Strengths and Weaknesses
The main benefit of SDUV is the quantification of elasticity with simultaneous consideration of viscosity. SDUV seems to be more benefi­cial than ARFI when tissue displacement or SNR is low. The shear wave propagation depends only on the material properties and not on US intensity and beam shape. Therefore, measurements are device­independent. In addition, the risk of interference due to shear wave echo is reduced. Finally, short acquisition times of SDUV (about 0.1 s per acquisition) “also allows fast acquisition of multiple measurements at dif­ferent locations within the organ of interest to get a comprehensive assess­ment of tissue state”
[57].
The limitation of SDUV rests in its ability to provide a single-point measurement. 2D imaging may be theoretically possible, but is time­consuming under the current SDUV technique (
Table 5.5).
Recent Developments and Current Research
One notable future development for SDUV is an extension in the field of applications. Another trend is an optimization in radiation force delivery to create shear waves that produce better information. Beyond this, there is a need to develop better methods for detection of shear waves and solving for the viscoelastic material properties of the tissue will further enhance the performance of SDUV.
Table 5.5 Key facts on SDUV Typical
applications
Examination
of liver fibrosis
Strengths and weaknesses
Quantification of
elasticity and viscosity of tissue
Long examination of
2D imaging
Recent developments
n/a Optimizing
Research potential and future trends
radiation force delivery
Extending field
of applications
127Diagnostic Procedures
Elastography in general is expected to remain in broad use in medical diagnostics. It is even predicted to gain further relevance in the future. It combines techniques which have high potential to detect cancerous dis­eases. Therefore, it is necessary to improve modifications like SWE, which is currently limited in its performance.

5.4.6 3D/4D Ultrasound

3D US is a volumetric imaging technology that provides a 3D view of internal str uctures. Dynamic volumetric imag ing, also known as “4D US” or “real-time 3D US,” extends the visualization with a time frame so that it is able to display motion instead of a static 3D data set.
3D data are usually acquired as a large number of consecutive tomographic images through movement of an US transducer array. Each tomographic image has to be gathered along wit h its positional information to constr uct a 3D data set. Accurate positional informa­tion is obtained through an electromagnetic position sensor, an electric gyro attached to t he probe, or by defining previous movement (
Tab le 5.6 ).
Static 3D images can be acquired manually by moving a 2D trans­ducer across a ROI, or automatically through the use of a 3D transducer that sweeps a 2D array of beams across the ROI. 3D/4D US requires rapid automatic sweeps of multiple adjacent 2D cross-sections.
Software
The software is the core of volumetric imaging technologies, espe­cially 4D visualizations, which need highly optimized algorithms. For applications like scanning a heart, a gated technique is applied to avoid distortion of a 3D data set due to movement. Tomographic images are rearranged according to the phase of the cardiac cycle and a 3D data set is constructed with only tomographic images at the same phase of the
Table 5.6 Key facts on 3D/4D US Typical applications
Obstetrics,
cardiovascular medicine
Visceral medicine
Strengths and weaknesses
Vague depiction
of internal structures
Recent developments
USCT/Warm
bath US
Research potential and future trends
Optimizing 3D US,
Increasing image quality
128 Biomedical Engineering in Gastrointestinal Surgery
cardiac cycle. The heart can be seen beating three-dimensionally by reconstructing many 3D data sets into a single cardiac cycle.
Strengths and Weaknesses
3D images provide examiners with an abundance of information, reducing the amount of interpretation needed and limiting the probability of misdiagnoses.
Compared to common US modifications, the amount of data involved is much higher. The depiction of one ROI demands up to 20 GB of data storage. Despite the increasing capability of computers, the processing time of data sets is still a limiting factor.
Recent Developments and Current Research
US travels through soft tissue at an average speed of 1540 m/s, which limits 3D scanning speed. The parallel receiving technique uses one broad US beam that is transmitted; its echoes are received as plural ultrasonic beams. In a 2D array probe, a high degree of parallel receiving is used and high-speed 3D scanning is possible. As a result, the profound advancements in 3D/4D imaging are mainly due to a general evolution of electronics and transducer arrays from linear systems to 1.25D, 1.5D,
1.75D, and 2D arrays and the latest matrix phase transducers, which are a current field of research
[58].
Currently, there are plans to make 4D US available through handheld devices, which has already been achieved in high-end devices
[59].
Today, the relevance of 3D/4D US systems is still low due to high purchasing costs and technological performance issues. In the future, the relevance of these systems is expected to rise as a result of technical improvements and mass-market adoption.

5.4.7 Ultrasound Computed Tomography

Ultrasound CT (USCT) is a new digital imaging technique that creates reconstructed 3D images of inhomogeneous media, such as soft tissue. It attempts to solve the problem of the inv erse-scattering field. A few research groups have developed such systems to test its usability. One of them built a USCT that consists of a water-filled cylinder and contains 1920 transducers— 384 sending and 1536 receiving transducers—which are grouped in three rings on the cylinder surface. The water-filled cylinder can be moved in six different positions via an electric motor. The advantage of this method is the high spatial resolution and the high tissue contrast. There are different algo­rithms currently under research for evaluating the emerged data sets, such as
129Diagnostic Procedures
the synthetic aperture focusing technique (SAFT) algorithm developed and optimized by the Forschungszentrum Karlsruhe/Germany, the multifre­quency nonlinear 3D inverse-scattering algorithm, or new concepts like the 3D refraction corrected 360 degrees compounded reflection algorithm
[60,61]. The early diagnosis of breast cancer is still a major challenge in spite
of recent developments in research
[62].Thecurrentdiagnosticprocedure
only detects cancer that is already in a developed state. In order to diagnose early-stage cancer within the breast, an application using USCT is currently under development.
Future developments will focus on introducing this technique to clinical
use and to reduce the time needed to evaluate the examined data sets
[63].

5.5 NUCLEAR IMAGING SYSTEMS

Nuclear imaging systems use gamma rays, which, like X-rays, are a form of electromagnetic, indirectly ionizing radiation, but possess more energy due to higher frequencies
Images generated in nuclear medicine are a result of the selection and injection of a suitable radioactive tracer, the detection of the radiation, the use of tomographic reconstruction algorithms, and finally the conduc­tion of a series of corrections for the image acquisition. A radioactive biologically active substance is chosen in such a way that its spatial and temporal distribution in the body reflects a particular body function or metabolism. To avoid disturbances of vital functions while studying the distributed radiation, only small amounts of the tracers are administered to the patient. The gamma rays of positrons are emitted as the tracer decays. The distribution of the radioac­tive tracer is inferred from the detected radiation and mapped as a func­tion of time and/or space
Medical nuclear imaging assesses the functional aspects of organs, whereas other techniques describe their anatomical structure. With its current techniques, medical nuclear imaging offers high-resolution multi­dimensional images of organs in order to analyze complex structures and physiologic functions for—among other things—computer-assisted diagnosis, evaluation of treatments, and interventions
Nuclear imaging systems can be divided into three main categories: positron emission tomography (PET), single-photon emission tomogra­phy, and the hybrid systems (see
[64].
[65]. The tracer principle works as a basis
[65,66].
[65].
Section 5.8: Hybrid Systems).
130 Biomedical Engineering in Gastrointestinal Surgery

5.5.1 Gamma Camera

The gamma camera, also called scintillation camera, is the most com­monly used imaging device in nuclear medicine. It simultaneously detects radiation from the entire FOV and enables the acquisition of dynamic as well as static images of the area of interest in the human body general, the gamma camera consists of a collimator, a scintillation crystal, and photomultiplier tubes (PMTs).
Recent Developments and Current Research
Recent developments in gamma ray detection have addressed its implementation within multimodal or hybrid systems. The fusion of PET and single-photon emission computed tomography (SPECT) with MRI systems (see
Section 5.8: Hybrid Systems) has been a particular challenge
because PMTs and nuclear imaging electronic hardware are sensitive about magnetic fields. To overcome the problem of incompatibility, two approaches have been developed.
The first approach uses optical fibers to couple the scintillation crystals inside the magnet to either PMTs and electronics outside the fringe of the magnetic field, or to solid-state photosensors situated at the end of the magnet bore. However, such a connection tends to lose light signals and limits the axial extent of the PET detector array due to difficulties in connecting the fiber bundles to the scintillation crystals and then to rout­ing them out of the magnet
[65,68].
The other approach substitutes the PMT with avalanche photodiodes, which are magnetic field-insensitive, solid-state photon detectors. These photodiodes are directly coupled to the scintillation element
[67].In
[69].

5.5.2 Positron Emission Tomography

PET is a noninvasive nuclear imaging technique that can help detect anatomic, functional, and biochemical abnormalities in organs. Furthermore, measurements of body functions such as blood flow, oxygen usage, glucose metabolism, and tissue perfusion are possible. This wide range of a pplica tions makes PET highly attractive for diagnostic and interventional purposes i n cardiology, ne urology, and oncology
multiple rings of detectors. Similar to gamma cameras, the PET detec­tors comprise of scintillation cr ystals with coupled PMTs. The ring design utilizes the concept that two photons detected in close
[70].
A PET scanner consists of a dedicated camera system, including