Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5319_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
92 Мб
Скачать
Chapter  • Experimental Methods of Structure Determination
13
. Fig. 13.8 A perforated pinhole mask can be used for adiffraction
experiment with alaser pointer. For this the displayed hole patterns (above) must be brought to the size of the wavelength of laser light. The diffraction patterns below were generated from these masks. The holes in the two left masks are all of the same size, which is compara­ble to having only one type of atom. The hole pattern changes from wide-meshed squares to askewed parallelogram (see red unit cells). The diffraction patterns reect the symmetry and distance of the holes to one another. In the third and fourth masks on the right, the distance
coordination to histidine or cysteine). These heavy at- oms dominate the diffraction pattern so that they betray their position in the crystal lattice. Another method takes advantage of anomalous scattering. This effect is based on the interaction of X-rays with the electrons of heavy atoms. As aresult, aspherical wave propagating towards an atom is reected with aphase shift. Simply stated, it is returned with adelay. This effect depends on the wave­length and can be used to determine the phase. The crys­tal is measured on asynchroton (aparticle accelerator that also produces electromagnetic radiation in abroad wavelength range, including X-rays) and the diffraction experiment is performed at several different wavelengths. Anomalous scattering requires that aheavy atom is pres­ent in the protein structure. This is already the case for metalloproteins. But an alternative approach can also be taken. Proteins produced in aspecial expression sys­tem (Sect.12.6) can be generated with selenomethionine instead of methionine. The heavier selenium acts as an anomalous scatterer in the diffraction experiment. Espe­cially for small molecules, there are methods that allow asimple reconstruction of the phase information from probability considerations in the intensity distribution among different reections, the so-called “direct meth-
between the repeating units is identical to those of the rst masks on the left. The composition of the motif in the repeating unit, however, varies. It is made up as clusters of multiple holes and can be compared to the different atoms in amolecule. The distance between the diffract­ed light reections (lower row) is identical for the rst, third, and fourth masks. The intensity of the diffracted radiation, however, varies from reection to reection. It contains information about the composition and the geometry of the individual motifs and, thus, the “molecules” which give rise to the diffraction patterns
ods.” Such methods are being developed also for pro­tein structure determination. Often an already solved, geometrically related protein structure can be used as astarting model for astructure determination (molec- ular replacement method). Since the structure prediction programs such as Alphafold or Rosettafold (Sect.20.6) have meanwhile reached aconvincing reliability, it is also possible to generate astarting model in this way. The model is translated and rotated in the elementary cell by computer simulations until acalculated diffraction pattern is obtained that matches the experimentally ob­served diffraction pattern of the unknown protein.
The phasing obtained at the beginning of the struc­tural analysis with these methods is only approximate and must be “rened.” For this purpose, the initial model is shifted and modied step by step until an optimal agreement with the experimental diffraction data has been achieved. Altogether the regeneration of the phas­ing information is not trivial. Even in the 1960s, phasing calculations kept one scientist busy for several years. The methodical progress and the increased performance of computers now allow this to be accomplished in afew minutes. Even today, however, this step can still be very challenging for proteins. It is becoming apparent, how-
. • Diraction Power and Resolution Determine the Accuracy of aCrystal Structure


. Fig. 13.9 View of a crystal structure of aldose reductase
(Sect.27.4). The electron density (the so-called 2F0–Fc density at the 1σ level) is displayed as ablue mesh on the predened contour level around atryptophan residue. In(a), the diffraction data were obtained at aresolution of 4 Å, and aFourier transform was used to calculate the electron density. The resolution increases from(a) 4 Å to(b) 3 Å, to(c) 2 Å, and to(d) 0.66 Å. The resolution in the last-shown contour
ever, that the structure determination of medium-sized proteins is becoming routine. Historically, the time span from crystallization to structure determination could be quite long. Urease is certainly acuriosity. It was the rst protein to be successfully crystallized. JamesB. Sumner accomplished this back in 1926. Its 3D structure, how­ever, was rst elucidated in 1995, that is, 70years later!
13.5 Diffraction Power and Resolution
Determine the Accuracy of aCrystal Structure
Apicture of the contents of the unit cell is the result of the Fourier transform. It is portrayed in terms of the electron density in space (. Fig.13.9). To which detail this electron density can be determined depends on the number of different wave fronts which are superimposed with correct amplitude and phase. The number of wave fronts is equivalent to the number of observed reections
density is so high that hydrogen atoms can be recognized as single den­sity peaks in the difference density map (positive is yellow, negative is violet of the so-called F0–Fc difference density at the 2σ level). The electron density is so clearly structured at 2 Å(c) that it is simple to t the indole building block in place. At 4-Å resolution(a), this assign­ment is problematic and can easily lead to errors
(see above). In the diffraction images with the laser beam (. Fig.13.8), it can be seen that the intensities decrease signicantly towards the rim. The maximum achievable resolution is, therefore, determined by the ultimately measurable reection still observed at the rim. It is gen­erated by the array of planes with the smallest mutual distance in the lattice that can still be observed in the diffraction experiment. In . Fig.13.7a, the spacing be­tween aset of planes is large and the occupation density with atoms is high. Therefore, the observed reection will be strong. In . Fig.13.7b, the planes are closer together, the occupation density with atoms, however, is lower, and the generated reection intensity becomes weaker. In . Fig.13.7c, the planes of an array are very close to one another, and the occupation with atoms is strongly thinned out. Consequently, only avery weak reection is expected here, possibly reaching the resolution limit.
For small organic molecules, this resolution is easily achieved in that the atoms are visible as distinct maxima in the electron density. If the crystal’s quality is dimin-
Chapter  • Experimental Methods of Structure Determination
13
ished due to lattice defects or disorder, the resolution is poorer. The resolution obtained from protein crystals is usually between 1.5 and 3 Å. This means, in the best case, aresolution is achieved that is on the order of magni­tude of abond length. The upper limit is in the range of the cross section of abenzene ring. Recently, resolu­tions of less than 1 Å have been increasingly achieved (. Fig.13.9). In such cases, many details can be seen, for example, individual hydrogen atoms or multiple spatial arrangements of side chains.
At higher resolution, the electron density max­ima are directly assigned to the atoms in the molecule (. Fig.13.10). In the beginning this assignment is crude; the phases used in the Fourier transform are only ap­proximate. The position of the detected maxima must still be optimized. This is dened as “renement of the structure.” For this, the experimentally observed diffrac­tion pattern is compared with the diffraction pattern that is calculated from the atomic positions of the prelimi­nary model. Such an initial model has to be optimized in aleast squares renement, which is an iterative mul­tistep process. Structural parameters such as atomic co­ordinates and parameters describing thermal motion are iteratively modulated in small steps, and at each rene­ment cycle the achieved agreement is quantied by the so-called R-factor. It determines the agreement between the diffraction data calculated from the obtained model with those from the original X-ray diffraction experi­ment. R-values range from zero for perfect agreement between the calculated and observed diffraction patterns to about0.6 for aset of measured diffraction data com­pared to aset of random data. In protein crystallogra­phy, an R-factor in the range of0.2 is considered ade­sirable target for data with aresolution of about 2.5 Å. For small organic molecules, renement typically results in values of R < 0.05.
If the measurement is very accurate, the density of a“pseudomolecule” with spherical atoms can be sub­tracted from the observed electron densities at the end of the structure determination. What remains is the electron distribution of the bonds between and lone pairs at the atoms in the molecule. This is, however, only possible with extremely high-resolution measurements. At lower resolution, as is the case in moderately resolved protein structure determinations, adirect assignment of the at­oms of the protein to the electron density maxima cannot be made (. Fig.13.10). More commonly, the course of the chains is tted to the electron density. Because pro­teins are constructed from 20different amino acids that prefer to take on typical geometries, the interpretation of the electron density is simplied (. Fig.13.10e,f).
Electrons scatter X-rays. Therefore, the number of electrons around an atom determines how well it is de­tectable in the resulting density. Hydrogen atoms have only one electron in their shell. As aresult, they are of­ten undetectable or inaccurately located in the electron
density. Hydrogen atoms can usually be identied in the density of small molecule crystal structures, but this will only be possible in protein structures if the resolu­tion is less than 1 Å. This is unproblematic as long as the hydrogen atoms are in positions that correspond to spatially xed positions on arigid molecular scaffold, e.g., hydrogen atoms on phenyl rings. It is more difcult when the hydrogen atom is on aconformationally exible group or groups that can be protonated or deprotonated. It would be ideal to know whether acarboxyl group is ionized or exists as afree acid and in which direction the hydrogen atom is oriented. This information can only be gleaned indirectly from the structure of the protein, through an accurate analysis of the spatial orientation and interaction geometry of the surrounding hydrogen bonding partners.
The accuracy of structure determination depends on the resolution of the data obtained from acrystal. Even if the structure of the protein is displayed on the computer screen like that of asmall organic molecule, its geometry is determined much less accurately. The margins of error for small molecule determinations are about 0.01 Å for bond lengths, 0.1° for bond angles, and 1–2° for dihedral angles (Chap.16). For protein structures, the errors are much larger and difcult to quantify. They depend on how the structure has been rened. Mostly the electron density does not allow the resolution of single atoms. Therefore, amino acids are placed in the electron density with idealized bond lengths and angles. Their geometry is left at the predened knowledge-based values for subse­quent renement. The assignment of atom types for the placement of side chains is partly based on assumptions. Knowledge-based values are used or an attempt is made to keep the hydrogen-bonding network consistent. These aspects must be considered when evaluating the accuracy of aprotein structure. The result of crystal structure de­termination is aspatially and temporally averaged image of an “average” molecule representing the entire crystal. It is often found that the electron density in some regions indicates only areduced occupancy of aside chain or part of abound ligand. In addition, alternative orienta­tions (conformations) may be seen. Sometimes the elec­tron density of entire regions is missing. This is indicative of spatial “disorder” and argues for adistribution over multiple orientations in the crystal. When describing the diffraction phenomena, we had seen that lattice planes, and thus the atoms arranged there, will contribute to the diffraction pattern only if they are found periodically as an array of planes at the same location over the whole crystal. This is not the case with disorder. The strict peri­odicity is lost and with it the contribution to the diffrac­tion pattern. Disorder in the crystal can also be dynamic, that is, the corresponding group moves back and forth between two or more arrangements as athermal motion. Alternatively, the disorder can be static, meaning that several orientations exist side by side in acrystal, but are
. • Diraction Power and Resolution Determine the Accuracy of aCrystal Structure


. Fig. 13.10 The crystal structure determination of organic mol-
ecules and proteins requires crystals with an edge length of approx.
0.01–0.3 mm. (a)In the X-ray beam, adiffraction pattern is obtained
(compare . Fig. 13.8), which in the past was registered on aphoto­graphic plate, today with an area detector on adiffractometer. The diffraction pattern of protein crystals shows amuch denser reection pattern. (b)Structures of small molecules are usually measured in the laboratory on an automated diffractometer. (c)Electron density indi­cates the positions of individual atoms. (d)Data from protein crys­tals are nowadays collected almost exclusively at synchrotron sources. (e)With approximated phases, aFourier transform is performed and the electron density in space is obtained, which is contoured according to apredened electron density level. (c,f)After structure renement, the density is interpreted and amodel of the diffracting molecule is tted. (g)Because of the complexity of their structure, proteins are usually represented by aribbon model that depicts the course of their polymer chain in space. (h)The spatial blurring of the electron density is associated with thermal motion of the atoms. It is represented for
randomly distributed. Because the structure is an aver-
aged picture, these arrangements are randomly scattered
throughout the crystal with different orientations. If part
of the molecule is completely disordered, i.e., scattered in
many orientations, the electron density is usually not visi-
ble. Today, diffraction data on protein crystals are mainly
collected at synchrotron radiation sources. Only in rare
cases are they still collected at an in-house facility using
radiation from an X-ray tube. In such atube, electrons are
emitted from alament and accelerated in ahigh-voltage
electric eld (1–100 kV). The electrons then collide with
ametal anode. As the electrons decelerate on the anode
material, X-rays are produced as characteristic brems-
strahlung. Electrons are expelled from the inner shell of
the metal atoms and electrons from an outer shell take
their place. The emitted radiation is therefore determined
by the energy difference between the shells and is specic
small molecules by ellipsoids encompassing 50% of the atomic popu­lation probability. (i)For proteins, thermal motion is indicated as so­called B-factors using acolor-coding scheme projected onto the fold­ing pattern. Red indicates high thermal motion, while blue indicates low thermal motion. (k)The spatial arrangement of the molecules in the crystal lattice shows adense packing of the protein molecules (here thrombin). However, large, at rst glance “empty” channels exist in the crystal packing between the molecules. These channels are occu­pied by alarge number of water molecules. Because of their extensive thermal motion and the resulting disorder, they are not detected in the electron density. However, small-molecule ligands can diffuse through these channels when the crystals are soaked with asolution of these ligands. (7 https://sn.pub/m1wZeO)
for the metal of the anode material. In asynchrotron, radiation is produced as electromagnetic waves when electrons accelerated to nearly the speed of light are forced to follow acurved path by magnetic elds. The radiation is emitted tangentially to the trajectory of the electrons, because physically achange in the direction of the velocity vector on the curved trajectory means an acceleration and thus leads to aspecial form of brems­strahlung. In this way, abroad spectrum of wavelengths can be produced. If desired, monochromatic radiation can be produced using mirrors and crystals, where the strong reection of acrystal (e.g. graphite) is used as the primary beam. The synchrotron beam is several orders of magnitude more intense than the beam from an X-ray tube. Measuring times of days on an instrument in one’s own laboratory can be reduced to seconds. However, to minimize radiation damage to the crystal samples, it is
Chapter  • Experimental Methods of Structure Determination
13
. Fig. 13.11 If zinc chloride is reacted with 2,4,6-tris(4-pyridyl)-
1,3,5-triazine (top center), crystals of aporous structure with cavities are formed(a). Adrop of solvent, in which an organic test molecule to be investigated is dissolved, is added to the crystals and the test molecule can diffuse into the huge cavities of the crystal lattice. If the process occurs slowly and in thermodynamic equilibrium, the test compound will place itself in aregular fashion in the empty cavities of the lattice(b). Its geometry can then be determined together with the lattice of the crystalline sponge (c). Only very small amounts of
necessary to measure the structures at about 100 K in acryogenic nitrogen gas stream. At this temperature, many motions in the crystal are frozen and mostly static disorder is observed. Nevertheless, it has been shown that the structures determined correspond well to the situa­tion at room or body temperature. These conclusions can be drawn by comparing the results with the analogous determination from NMR spectroscopy (Sect.13.7) and molecular dynamics simulations (Sect.15.7). Neverthe­less, detailed studies of crystal structures collected at low and ambient temperatures have shown, as expected, that the conformer distributions of side-chain rotamers can differ at these different temperatures and this is reected in the determined structures. It should be noted, however,
test substance are required for the method. For example, substances formed during drug metabolism can be characterized after their chro­matographic separation(d). The molecule shown in(c) is the glucuro­nidated metabolite of gembrozil. (7 https://sn.pub/aJLQMB)
that this is also where the largest differences to NMR structures and those from molecular dynamic (MD) sim­ulations are observed.
Crystallography of small molecules is still the most powerful analytical method for characterizing the chem­ical composition and especially the stereochemistry of organic compounds. This method allows the absolute conguration of molecules to be determined with con­dence on the basis of so-called anomalous dispersion. It requires the presence of an electron-rich atom that exhibits such an anomalous dispersion contribution in the molecule under investigation. For small molecules, phosphorus or sulfur atoms may be sufcient for these studies. For elements such as chlorine, bromine, or zinc,
. • Electron Microscopy: Topographic Images Reveal Macromolecular Structures


the effect is stronger. However, aprerequisite for all these crystallographic determinations is the growth of asingle crystal. Unfortunately, this is often not trivial. What can be done if the compound under investigation will not crystallize? In many cases, there is simply not enough material to grow acrystal. Apromising alternative has emerged in recent years: Diffusion in crystalline sponges! The concept dates back to Makoto Fujita’s group at the University of Tokyo, Japan. For example, when zinc chloride is reacted with 2,4,6-tris(4-pyridyl)-1,3,5-tri­azine, crystals with aporous cavity structure are formed.
. Fig.13.11a shows asection of the crystal packing of
this structure, which contains large cavities. These crys­tals can be placed in adrop of organic solvent containing asolution of the substance to be studied. The substance can then diffuse into the cavities of this crystal lattice. If this diffusion is slow and in thermodynamic equilib­rium, the guest molecules will be taken up and arranged regularly in the lattice. Everything else then proceeds as in aroutine structure determination. The absolute con­guration can also be determined in this sponge. The zinc atoms help in this process. They are the necessary anomalous scatterers in the crystal lattice. The method requires only very small amounts of substance and the crystallizability of the substance under investigation is not required. The method can greatly assist medicinal chemists in the analysis of their synthesis products.
However, this process can provide other extremely valuable assistance in drug development. Once adrug molecule has been developed to the stage of aclinical candidate, it is necessary to study in detail how the sub­stance is chemically modied in the human body. In Sect.27.6, we will learn how enzymes, particularly in the liver, chemically modify drugs, or “metabolize” them. This is the process of converting adrug molecule into aform that can be more easily eliminated from the body via the urine. Because the metabolic process creates active ingredients that are potentially new to the human body, it is important to determine exactly what substances are being formed. What is their stereochemistry and are they toxic to the body? But before these questions can be an­swered, the degradation products must rst be structur­ally characterized. This is where the crystalline sponge method comes into play. So-called homogenates can be used to simulate the metabolism in the liver in the lab­oratory. Any metabolites formed are then separated by chromatography. Usually, only avery small amount of the substance is available for subsequent analysis. Mass spectrometry would be one option for analysis. However, this does not provide the exact topology and stereochem­istry. If, on the other hand, the individual fractions are allowed to diffuse into the crystalline sponges, the chem­ical structure, including stereochemistry, can be obtained in the best case! . Fig.13.11d shows an example of the metabolic degradation of gembrozil, alipid-lowering drug from the class of brates (Sect.28.6). The four me-
tabolites formed could be diffused into the crystalline sponge as guests. . Fig.13.11b,c shows the structure of the metabolite glucuronidated at the acid function. Only atiny amount was needed. Crystallography per­formed on metabolites diffusing into crystalline sponges certainly has immense potential to massively assist in the difcult unraveling of drug metabolism.
13.6 Electron Microscopy: Topographic
Images Reveal Macromolecular Structures
In addition to X-rays, beams of electrons and neutrons can also be used for diffraction experiments to determine the structures of molecules. Electron beams have the great advantage over X-rays and neutrons that they can be bent by magnets. Therefore, it is possible to build con­verging (convex) lenses for them, comparable to alight microscope, to show amagnied image. With X-rays, this is practically impossible or only very inefcient to real­ize. This leads to the described phase problem of X-ray crystallography. The operation that performs the task of a“converging lens” has to be replaced by aFourier trans­form. Only then does the electron density of the mole­cules, and thus their spatial structure, become accessible with X-rays. In the electron microscope, amagnetic lens can be used to perform this Fourier transform directly and then even obtain an image of individual molecules as “shadows.” However, for along time this approach did not provide the necessary resolution to reveal the desired details in these structures.
In recent times, new developments—especially in the eld of detectors for the registration of electron beams—have enabled electron microscopy to make ade- cisive breakthrough in the determination of the structure of huge macromolecular complexes. The method does not require crystallized proteins; single molecules can be studied. Particularly in the case of larger assemblies of huge protein complexes, crystallization is often the bottle neck for timely structure determination.
Cryo-electron microscopy (cryo-EM) examines mol­ecules as individual particles, similar to the way acom­puter tomograph in medicine scans apatient from all sides. The intact protein samples are exposed to an electron beam in ahigh vacuum (. Fig.13.12). Prior to this, they must be ash-frozen in avitreous water en­vironment. They are then exposed to the electron beam in many orientations at the temperature of liquid nitro­gen. Thus “vitried,” thousands of projection images of the protein are taken. The next step is the reconstruc­tion of the 3D structure of the studied molecule from these usually very noisy projections. This is done using
image-processing algorithms. The result is acomposite
Chapter  • Experimental Methods of Structure Determination
13
. Fig. 13.12 Schematic representation of the workow during struc-
ture determination with cryo-electron microscopy. Protein molecules are ash-frozen in avitreous water coat and deposited on agraphite grid under high vacuum(a). An intense electron beam (sketched by turquoise cylinder) sweeps across the grid, creating many thousands of shadow images as projections of the molecules deposited on the grid(b). The extremely noisy shadow projections(c) are sorted in the
spatial image of the charge density of the microscopi­cally observed object. Since electron beams have amuch shorter wavelength than light waves, structures down to the 2–3 Å range can be resolved in favorable cases. By averaging alarge number of measured shadow projec­tions, the signal-to-noise ratio of the projections can be signicantly improved. By iteratively evaluating individ­ual projections, reconstructing the 3D structure, and sub sequently improving the alignment and superposition of the raw shadow images, the representation of the molec­ular complex under investigation is continuously rened.
In 2017, Jacques Dubochet, Joachim Frank, and Richard Henderson were awarded the Nobel Prize in Chemistry for their contributions to the development
of the cryo-EM technique. Dubochet helped the tech­nique achieve abreakthrough by cryopreparing the pro- tein samples in the form of avitried embedding in an amorphously solidied water shell about 100 nm thick. At the same time, this step allows the samples to be mea­sured at low temperature, since the high-energy electron
computer in terms of similar orientations(d), compiled, and then av­eraged(d, e). As a result, the molecule under investigation emerges more and more strongly from the background noise(e). The same pro­cedure is followed with the many thousands of projections (f), and an averaged spatial image of the protein molecule is generated from these images, similar to the way in which acomputer tomograph is generated in medicine(g)
beam would destroy the proteins very quickly at room temperature. Joachim Frank developed the algorithms to discover and extract the recurring patterns from the noisy projected images of randomly distributed molecules, ul­timately reconstructing amodel of the spatial structure of the molecules under study. Richard Henderson helped cryo-EM make an experimental breakthrough. Step by
-
step, he was able to increase the image resolution of the bacteriorhodopsin he studied (Chap.29) until the folding pattern of this protein was nally visible.
Nowadays, under optimal conditions, complexes can be determined down to atomic details of their spatial structure. Only recently did Holger Stark’s group at the MPI in Göttingen succeed in determining acryo-EM structure of apoferritin with aresolution of 1.25 Å! This is certainly still the exception. When comparing the reso- lution limits of diffraction methods and cryo-EM, some caution is required because the resolutions are dened and calculated differently. Referring to the closest lat­tice plane distance, as in diffraction techniques, does not
. • Electron Microscopy: CT on Single Molecules and Diraction on Two-dimensional Crystals


work in the cryo-EM technique. In order to reliably eval­uate the quality of the data collected at the many pro­jections in athree-dimensional volume, an evaluation in Fourier space is performed. This is done by dividing the dataset into two equal parts and calculating the Fourier shell correlation. From the inside to the outside, the data are processed in individual shells and it is determined how well the Fourier components of the transformed data correlate in both parts of the dataset. The shell in which the correlation falls below apredetermined thres­hold is specied as the resolution limit. The analysis is, therefore, more of ameasure of the internal consistency of the two splitted datasets.
Cryo-EM is already the method of choice for study­ing multiprotein complexes, such as the spliceosome, and for addressing mechanistic questions about the function of individual proteins in acomplex. Conventional struc­ture determination in drug design typically requires res­olutions of 2 Å or better. Structures must be able to be determined without signicant effort on ligand series. In contrast, each additional EM structure virtually means acompletely new structure determination. For these rea­sons, classical X-ray crystallography is likely to remain the workhorse for many years to come. One advantage of cryo-EM is certainly that the molecules are in afrozen water environment. This is certainly closer to physiolog­ical conditions than acrystalline assembly. In addition, the method is better at capturing different states of apro­tein. For example, if the protein exists as amonomer and adimer side by side, this will be recorded. However, the computational cost of structure determination increases dramatically with increasing resolution.
An alternative, which was initially pursued inten­sively, is to use crystalline material for the electron mi­croscopic electron microscopic studies. Since the aver­aged image of several aligned molecules is examined in aperiodic crystalline array, astronger signal is observed. The converging lens step (Fourier transform) is omitted and the reections in the diffraction space are measured on the crystalline sample as in X-ray diffraction. Elec­trons penetrate only slightly into the crystalline sample material, but are scattered much more strongly by the molecules in the crystal. Therefore, much smaller crys­tals can be used, and even crystals that are wafer-thin in one direction and consist of only one or afew molecular layers are sufcient.
Despite the lower radiation exposure, the electrons still result in considerable destruction of the samples. It is important to remember that the crystals used are only about abillionth of the sample quantity of acrystal used for X-ray diffraction. The data for an X-ray structure can often be measured on one single crystal. The electron mi­croscope, on the other hand, requires many hundreds of the tiny crystals, often only 5 µm in size. They are also ash-frozen under high vacuum and exposed directly to the electron beam. The images are also very noisy and
have to be averaged over many images. To obtain detailed resolution perpendicular to the two-dimensional crystal plane, crystals must be measured in many orientations. AFourier transform is used to obtain acharge density distribution of the molecules, similar to X-ray diffraction. Its interpretation or renement is done in the same way as in X-ray experiments. The phases required for the Fourier transform can be determined in the electron microscope by direct imaging in the “converging lens mode.” In the group of Tamir Gonen at the University of California, Los Angeles, USA, the idea was developed to minimize radiation damage by reducing the intensity of the elec­tron beam. Thus, agreater number of reections can be collected on individual crystals. In this way, the micro-ED method was successfully applied to aresolution range of about 1 Å by electron diffraction. The method can also be extended to the structure determination of small organic molecules. Since only very small crystals are required, the method works even with materials that look almost like an amorphous powder to the naked eye. To demonstrate its applicability, the Gonen group took nished drugs and crushed the tablets. The resulting powder was still crystalline enough to be used in an electron microscope to determine the structure of the drug molecules in the sample. It was also possible to nd sufciently large crys­tallites of the precipitant that remained in aask after asubstance was recovered following evaporation of the solvent. They still allowed the structure to be determined using the micro-ED method. This method may represent a breakthrough in crystallographic structure analysis, since in many cases the supposed bottleneck of single crystal growth may prove to be irrelevant.
Another type of radiation that can be used for diffrac­tion experiments on biomolecules is abeam of neutrons. Neutrons are produced either in areactor by nuclear s­sion of uranium isotopes or from aspallation source. In the latter case, heavy metals such as mercury emit out neutrons after they have been bombarded with high-en­ergy electrons in anuclear conversion reaction. Aneutron beam behaves much like an X-ray beam in adiffraction ex­periment. However, the strength with which the chemical elements contribute to the scattering power in areection is completely different to X-rays. For example, hydrogen is astrong scatterer and the Hisotope can be easily dis­tinguished from the Disotope (deuterium). An element like sulfur is aweak scatterer, and the metal vanadium is practically invisible to neutrons. Therefore, this metal is often used as amaterial for sample containers for neutron diffraction experiments. For biological samples, neutrons have the incredible advantage of being very powerful at visualizing geometries involving hydrogen atoms. This is especially true for the study of hydrogen bonds and pro­tonation states of acidic and basic groups. But also the orientation of water molecules and their dynamic behav­ior are visualized (. Fig.4.10). Why not do all structure determinations with neutrons? Unfortunately, the inten-
13
Chapter  • Experimental Methods of Structure Determination
sity of aneutron beam is much, much weaker than that of an X-ray beam, which means that the crystals required for neutron diffraction must be much larger, with edge lengths of several millimeters. Crystals of this size can be successfully grown from very few proteins.
13.7 Structures in Solution: The Resonance
Experiment in NMR Spectroscopy
Many atomic nuclei have an angular momentum, or spin. The nuclei that occur in biological systems that have anuclear spin are the hydrogen isotope 1H, the carbon isotope 13C, the nitrogen isotope 15N, the uorine isotope
19
F (in biology rarely found), and the phosphorus isotope
31
P (for the sake of simplicity, only nuclei with aspin of ½ will be considered here). Just as atop would, these nuclei rotate about their axes. As long as no magnetic eld is applied, the tops orient in all possible spatial di­rections. In amagnetic eld, they are forced into align­ment (. Fig.13.13). If atoy top is spun, it will move in the gravitation eld. This eld has, as the magnetic eld, one preferred direction. If the alignment of the rotation axis of the top and the direction of the gravitation eld, which is oriented towards the center of the Earth, are not exactly the same, the top will wobble. The end of the rotation axis performs acircular movement, an arc, with avery precise rotational velocity. It depends on the mass and geometry of the top. In physics this movement is known as precession.
Atomic nuclei with aspin behave in avery similar way. In contrast to the macroscopic top, they obey the laws of quantum mechanics. This means that the rota­tion axes that their precession movement takes on can only adopt very specic angles with respect to the applied eld direction. The result for the 1H, 13C, 15N, 19F, and
31
P nuclei is that the rotation axis for the precession arc can only be parallel or antiparallel to the direction of the eld (so-called spin ±½ particles). The orientation in the direction of the eld is energetically somewhat more favorable than the rotation antiparallel to the direction of the eld. Statistically, therefore, more nuclear spins in the substance sample will align with the direction of the eld. If an additional magnetic eld is applied to the outer magnetic eld and its frequency corresponds to the precession frequency of the nuclear spin, the occupancy of “parallel” to “antiparallel” spinning nuclei can be re­versed and aresonance absorption for the sample can be registered. After aparticular time span, the original situation is restored (relaxation) for entropic reasons.
The rotational velocity of the top’s axis for precession movements is characteristic for each type of nucleus. It further depends on, and is additionally modulated by, the composition of the chemical environment in which the nucleus resides. Acarbon atom of aphenyl ring has adifferent resonance frequency than that of an aliphatic
. Fig. 13.13 Atomic nuclei with a rotational momentum behave
like aspinning top. In the absence of an external magnetic eld, they orient in all possible directions randomly (left). Upon application of amagnetic eld, they orient their rotation axes parallel or antipar­allel to the direction of the eld (right). The precession movement is oriented in an arc around the applied eld direction. The two orienta­tions, parallel or antiparallel, with respect to the direction of the eld are energetically different. Because of this, there is asmall difference in occupancy between the two states. By applying an electromagnetic eld with afrequency corresponding to the rotational movement of the precession of the axis of the top, the occupation can be inverted. This resonance absorption, the exact frequency of which depends on the type of nucleus and its immediate chemical environment, is regis­tered with aspectrometer
chain. The relative position of the resonance absorption in relation to astandard reference is also called the chem- ical shift. Furthermore, the individual nuclei can perceive the spin orientation of the neighboring nuclei. An align­ment in the same direction as aneighboring nucleus is energetically different from that of an antiparallel orien­tation. This inuence also modulates the rotational speed of the spin on the observed nucleus. The information transfer regarding the orientation or the magnetic state of the nuclei in the vicinity can be transmitted over sev­eral bonds. This transfer can even occur through space without any direct covalent connection.
To measure an NMR spectrum (nuclear magnetic res­onance), asolution of the substance has to be placed in astrong magnetic eld. In addition, avariable electro­magnetic eld is applied to the sample. The frequencies at which the nuclei in the sample have resonance, meaning when they ip from parallel to antiparallel, are recorded. The resulting spectrum discloses information about the composition and the chemical environment around the studied nuclei. It contains information about the spatial structure of the molecules under investigation. Based on the work of Richard Ernst, who received the Nobel Prize in Chemistry in 1991, multidimensional NMR techniques have been developed over the last 30years. By applying suitable measurement conditions and selective electro­magnetic elds, information about the mutual inuence of resonance frequencies between individual nuclei is separated and analyzed. This bidirectional transfer of in­formation about the magnetic state of neighboring nuclei
. • From Spectra to Structure: Distance Maps Evolve into Spatial Geometries


is evident from the signal form in the multidimensional spectra, and it is registered in the form of cross peaks.
In 1965, Anet and Bourn performed decoupling ex­periments in which they saturated aparticular hydro­gen atom in the spectrum by irradiating it separately at the corresponding resonance frequency. They observed that the resonance intensity of aneighboring hydrogen atom, which was not covalently bonded to the saturated Hatom, increased by 45%. This, of course, provides im­portant information about the spatial proximity of the atoms. This so-called nuclear Overhauser effect (NOE) is often used in NMR experiments to elucidate the structure and conformation of biomolecules and their interactions.
Only the hydrogen isotope 1H occurs in nearly 100% natural abundance. Therefore, it can be assumed that for statistical reasons, two 1H nuclei will always be adjacent to each other in amolecule. In contrast, the 13C and 15N
isotopes are scarce. As aresult, statistically they are only
very rarely found in the direct vicinity of one another. Data on the mutual inuence of the magnetization of these nuclei are required for the spectra. Therefore, it is necessary to enrich the proteins with the appropriate iso­topes. For this, bacteria are fed with isotopically labeled substrates such as glucose or ammonium chloride and will then produce proteins that are isotopically enriched. It is even necessary to produce deuterated proteins for the structural investigation of very large proteins. To­day, by using numerous spectroscopic techniques, spectra from proteins of more than 800 amino acids have been successfully interpreted. The following questions can be addressed by NMR analysis:
Which atomic nuclei occur in which chemical environ-
-
ment?
What is in the immediate, covalently connected neigh-
-
borhood of these nuclei? Information about the spa-
tial orientation of atoms in the vicinity is also con-
tained within these spectral parameters.
Which geometric relationships are given between dif-
-
ferent segments of the polypeptide chain? This results
from information transfer about magnetic states of
nuclei that are not directly connected by covalent
bonds.
13.8 From Spectra to Structure: Distance
Maps Evolve into Spatial Geometries
This last-mentioned observation, which results from the nuclear Overhauser effect (NOE), yields intramolec­ular distances of spatially neighboring but not directly covalently bound atoms. The entire connectivity, that is, alist of all covalent bonds within amolecule, and alist of the recorded intramolecular noncovalent NOE distances are applied to generate the structure for the molecule (. Fig.13.14). For this purpose, so-called dis- tance–geometry calculations are used to create the spatial coordinates of the atoms.
Often, distance geometry produces several equally
good structural models that satisfy the experimentally
NMR studies are usually performed in solution. In this way, dynamic processes can be observed by changes in the spectra. Fluorine atoms are isotopically pure as 19F
nuclei and they are rarely used by Nature. As anucleus, it can be observed very well in the NMR spectrometer. Therefore, it is auseful probe to study the properties of
biomolecules. Fluorine-containing amino acids can be introduced into proteins under special expression con­ditions. The 19F resonances can then be used to study the structural and dynamic properties of proteins. An example is given in Sect.21.14.
. Fig. 13.14 A multidimensional NMR spectrum contains informa-
tion about the spatial vicinity of atomic nuclei in amolecule (here, the trypsin inhibitor from bovine pancreas). It is expressed in cross peaks. This provides information about the distance between noncovalent­ly bonded atoms in amolecule. The individual signals of the spectra are assigned to atoms in the molecule (e.g.,A and B). The positions that these atoms have in the polypeptide chain are known from the sequence of the protein (above left). The intensity of the cross peak indicates which spatial distance is found between nucleiA andB in the folded polypeptide chain (above right). Just as was done forA andB, the many other cross peaks are evaluated and translated into distance conditions. With the totality of this distance information, it is possible to fold the polypeptide chain in space
Соседние файлы в папке Библиотека им академика М.И. Перельмана