DICOM PS3.16 2020a - Content Mapping Resource​

Page 1261​

Code Value​

Code Meaning​

Definition​

Notes​

113846​

Target Exposure Index​

The target value used to calculate the Deviation Index​

 

 

 

as defined in IEC 62494-1; see PS3.3 definition of​

 

 

 

Exposure Index Macro.​

 

113847​

Deviation Index​

A scaled representation of the accuracy of the Exposure​

 

 

IndexcomparedtotheTargetExposureIndexasdefined​

 

 

in IEC 62494-1; see PS3.3 definition of Exposure Index​

 

 

Macro.​

 

113850​

Irradiation Authorizing​

The clinician responsible for determining that the​

 

 

 

irradiatingprocedurewasappropriatefortheindications.​

113851​

Irradiation Administering​

Thepersonresponsiblefortheadministrationofradiation.​

113852​

Irradiation Event​

An irradiation event is the loading of X-Ray equipment​

 

 

 

caused by a single continuous actuation of the​

 

 

 

equipment's irradiation switch, from the start of the​

 

 

 

loadingtimeofthefirstpulseuntiltheloadingtimetrailing​

 

 

edge of the final pulse. Any automatic on-off switching​

 

 

 

of the irradiation source during the event is not treated​

 

 

as separate events, rather the event includes the time​

 

 

 

between start and stop of irradiation as triggered by the​

 

 

user. E.g., a pulsed fluoro X-Ray acquisition shall be​

 

 

 

treated as a single irradiation event.​

 

113853​

Irradiation Event UID​

Unique Identifier of an Irradiation Event.​

Retired. Replaced by​

 

 

 

(113769, DCM, "Irradiation​

 

 

 

Event UID")​

113854​ Source of Dose Information​

Method by which dose-related details of an Irradiation​

 

 

 

Event were obtained.​

 

113855​

Total Acquisition Time​

Total accumulated acquisition clock time in the scope of​

 

 

the including report (i.e., the sum of the Irradiation​

 

 

 

Duration values for accumulated acquisition irradiation​

 

 

events).​

 

113856​

Automated Data Collection​

Direct recording of data by a relevant system.​

 

113857​

Manual Entry​

Recordingofdatabyahumanoperator,includingmanual​

 

 

transcription of electronic data.​

 

113858​

MPPS Content​

The data is taken from an MPPS SOP Instance.​

 

113859​

Irradiating Device​

A device exposing a patient to ionizing radiation.​

 

113860​ 15cm from Isocenter toward​

15cm from the isocenter towards the X-Ray source; See​

 

Source​

IEC 60601-2-43.​

 

113861​ 30cm in Front of Image Input​

30cm in front (towards the tube) of the input surface of​

 

Surface​

the image receptor; See FDA Federal Performance​

 

 

 

StandardforDiagnosticX-RaySystems§1020.32(d)(3).​

113862​

1cm above Tabletop​

1cm above the patient tabletop or cradle; See FDA​

 

 

 

Federal Performance Standard for Diagnostic X-Ray​

 

 

 

Systems §1020.32(d) (3).​

 

113863​

30cm above Tabletop​

30cm above the patient tabletop of cradle; See FDA​

 

 

 

Federal Performance Standard for Diagnostic X-Ray​

 

 

 

Systems §1020.32(d) (3).​

 

113864​ 15cm from Table Centerline​

15cm from the centerline of the X-Ray table and in the​

 

 

direction of the X-Ray source; See FDA Federal​

 

 

 

Performance Standard for Diagnostic X-Ray Systems​

 

 

 

§1020.32(d) (3).​

 

- Standard -​

Page 1262​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

113865​ Entrance exposure to a 4.2 cm​Standard breast means a 4.2 centimeter (cm) thick​ breast thickness​ compressed breast consisting of 50 percent glandular​

and 50 percent adipose tissue.​

 

See Department of Health and Human Services, Food​

 

and Drug Administration. Mammography quality​

 

standards; final rule. Federal Register. Oct. 28, 1997;​

 

68(208):55852-55994; see 900.2(uu).​

113866​ Copied From Image Attributes​Thedataiscopiedfrominformationpresentintheimage​

 

attributes. E.g., dose attributes such as CTDIvol​

 

(0018,9345).​

113867​ Computed From Image​

The data is computed from information present in the​

Attributes​

imageattributes.E.g.,byusingdosimetryinformationfor​

 

the specific irradiating device make and model, applied​

 

to technique information such as KVP and mAs.​

113868​ DerivedFromHuman-Readable​The data is derived from human-readable reports. E.g.,​

 

Reports​

by natural language parsing of text reports, or optical​

 

 

character recognition from reports saved as images by​

 

 

the irradiating device.​

113870​

Person Name​

The name of a specific person.​

113871​

Person ID​

An identification number or code for a specific person.​

113872​

Person ID Issuer​

The organization that issued a Person ID.​

113873​

Organization Name​

The name of an organization.​

113874​ Person Role in Organization​

The role played by a person in an organization.​

113875​ Person Role in Procedure​

The role played by a person in a procedure.​

113876​ Device Role in Procedure​

The role played by a device in a procedure.​

113877​

Device Name​

The name used to refer to a device; usually locally​

 

 

unique.​

113878​

Device Manufacturer​

Manufacturer of a device.​

113879​

Device Model Name​

Model Name of a device.​

113880​

Device Serial Number​

Serial Number of a device.​

113890​

All Planes​

All planes of a multi-plane acquisition equipment.​

113893​

Length of Reconstructable​

The length from which images may be reconstructed​

 

Volume​

(i.e., excluding any overranging performed in a spiral​

 

 

acquisition that is required for data interpolation).​

 

 

Value is distinct from (1113825, DCM, "Scanning​

 

 

Length"), which is the actual length of the table travel​

 

 

during the entire tube loading, according to IEC​

 

 

60601-2-44,andincludesoverranging.Alsodistinctfrom​

 

 

any actual Reconstructed Volume, which may depend​

 

 

on the slice thickness chosen for a particular​

 

 

reconstruction.​

113895​ Top Z Location of​

The Z location that is the top (highest Z value) of the​

 

Reconstructable Volume​

Reconstructable Volume. Specified as the Z component​

 

 

within the Patient Coordinate System defined by a​

 

 

specified Frame of Reference.​

113896​ Bottom Z Location of​

The Z location that is the bottom (lowest Z value) of the​

 

Reconstructable Volume​

Reconstructable Volume. Specified as the Z component​

 

 

within the Patient Coordinate System defined by a​

 

 

specified Frame of Reference.​

- Standard -​

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1263​

Code Value​

Code Meaning​

Definition​

Notes​

113897​ Top Z Location of Scanning​

The Z location that is the top (highest Z value) of the​

 

 

Length​

scanning length. Specified as the Z component within​

 

 

 

the Patient Coordinate System defined by a specified​

 

 

 

Frame of Reference.​

 

113898​ Bottom Z Location of Scanning​The Z location that is the bottom (lowest Z value) of the​

 

 

Length​

scanning length. Specified as the Z component within​

 

 

 

the Patient Coordinate System defined by a specified​

 

 

 

Frame of Reference.​

 

113899​

Exposed Range​

The range along the z axis of the total volume irradiated,​

 

 

 

per IEC 60601-2-44, Ed. 3, 203.115(b).​

 

 

 

The start and stop of loading corresponding to the outer​

 

 

 

edgeofthefullwidthhalfmaximumofthefree-in-airdose​

 

 

 

profile for the beam collimation used.​

 

113900​ Dose Check Alert Details​

Report section about cumulative dose alerts during an​

 

 

 

examination.​

 

113901​ DLP Alert Value Configured​

FlagdenotingwhetheraDLPAlertValuewasconfigured.​

 

113902​ CTDIvol Alert Value Configured​Flag denoting whether a CTDIvol Alert Value was​

 

 

 

configured.​

 

113903​

DLP Alert Value​

Cumulative Dose Length Product value configured to​

 

 

 

trigger an alert; see NEMA XR 25-2010 Dose Check​

 

 

 

Standard.​

 

113904​

CTDIvol Alert Value​

Cumulative CTDIvol value configured to trigger an alert;​

 

 

 

see NEMA XR 25-2010 Dose Check Standard.​

 

113905​

Accumulated DLP Forward​

A forward estimate of the accumulated DLP plus the​

 

 

Estimate​

estimatedDLPforthenextProtocolElementGroup;see​

 

 

 

NEMA XR 25-2010 Dose Check Standard.​

 

113906​ Accumulated CTDIvol Forward​Aforwardestimateatagivenlocationoftheaccumulated​

 

 

Estimate​

CTDIvolplustheestimatedCTDIvolforthenextProtocol​

 

 

 

Element Group; see NEMA XR 25-2010 Dose Check​

 

 

 

Standard.​

 

113907​

Reason for Proceeding​

Reason provided for proceeding with a procedure that​

 

 

 

is projected to exceed a configured dose value.​

 

113908​ Dose Check Notification Details​Reportsectionaboutdosenotificationsduringaprotocol​

 

 

 

element.​

 

113909​

DLP Notification Value​

Flag denoting whether a DLP Notification Value was​

 

 

Configured​

configured.​

 

113910​

CTDIvol Notification Value​

Flag denoting whether a CTDIvol Notification Value was​

 

 

Configured​

configured.​

 

113911​

DLP Notification Value​

Dose Length Product value configured to trigger a​

 

 

 

notification for a given protocol element.​

 

113912​

CTDIvol Notification Value​

CTDIvol value configured to trigger a notification for a​

 

 

 

given protocol element.​

 

113913​

DLP Forward Estimate​

A forward estimate of the DLP for the next Protocol​

 

 

 

Element Group; see NEMA XR 25-2010 Dose Check​

 

 

 

Standard.​

 

113914​

CTDIvol Forward Estimate​

A forward estimate of the CTDIvol for the next Protocol​

 

 

 

Element Group; see NEMA XR 25-2010 Dose Check​

 

 

 

Standard.​

 

113921​

Radiation Exposure​

Theamountofionizingradiationtowhichthepatientwas​

 

 

 

exposed.​

 

- Standard -​

Page 1264​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

113922​

Radioactive Substance​

Type, amount and route of radioactive substance​

Retired. Replaced by​

 

Administered​

administered.​

(440252007, SCT,​

 

 

 

"Administration of​

 

 

 

radiopharmaceutical").​

113923​

Radiation Exposure and​

Exposuretoionizingradiationandassociatedpreventive​Retired. Replaced by​

 

Protection Information​

measures used to reduce the exposure of parts of the​(73569-6, LN, "Radiation​

 

 

body like lead apron or eye, thyroid gland or gonad​

Exposure and Protection​

 

 

protection.​

Information")​

113930​ Size Specific Dose Estimation​The Size-Specific Dose Estimate is a patient dose​

 

 

 

estimate that takes into account the size of the patient,​

 

 

such as described in [AAPM Report 204] or [AAPM​

 

 

 

Report220]byusinglineardimensionsmeasuredonthe​

 

 

patient or patient images or estimated from patient age.​

113931​

Measured Lateral Dimension​

The side-to-side (left to right) dimension of the body part​

 

 

being scanned (per [AAPM Report 204]).​

 

113932​

Measured AP Dimension​

The thickness of the body part being scanned, in the​

 

 

 

antero-posterior dimension (per [AAPM Report 204]).​

 

113933​

Derived Effective Diameter​

The diameter of the patient at a given location along the​

 

 

Z-axis of the patient, assuming that the patient has a​

 

 

 

circular cross-section (per [AAPM Report 204]).​

 

113934​ AAPM 204 Lateral Dimension​ The Size Specific Dose Estimation is computed using​

 

 

 

Table 1B (32cm phantom) or Table 2B (16cm phantom)​

 

 

of [AAPM Report 204].​

 

113935​ AAPM 204 AP Dimension​

The Size Specific Dose Estimation is computed using​

 

 

 

Table 1C (32cm phantom) or Table 2C (16cm phantom)​

 

 

of [AAPM Report 204].​

 

113936​ AAPM 204 Sum of Lateral and​The Size Specific Dose Estimation is computed using​

 

 

AP Dimension​

Table 1A (32cm phantom) or Table 2A (16cm phantom)​

 

 

of [AAPM Report 204].​

 

113937​ AAPM 204 Effective Diameter​The Size Specific Dose Estimation is computed using​

 

 

Estimated From Patient Age​

Table 1D (32cm phantom) or Table 2D (16cm phantom)​

 

 

using an effective diameter estimated from the patient's​

 

 

age using Table 3 of [AAPM Report 204].​

 

113940​

System Calculated​

Values calculated from other existing parameters.​

 

113941​

In Detector Plane​

A segmented region of the detector surface within the​

 

 

 

irradiated area (but might not be near the center of the​

 

 

detector).​

 

113942​

X-Ray Reading Device​

AdevicethatcreatesdigitalimagesfromX-Raydetectors​

 

 

(Direct, Indirect or Storage).​

 

113943​

X-Ray Source Data Available​ ParametersrelatedtotheX-Raysource(generator,tube,​

 

 

etc) are available to the recording application.​

 

113944​

X-Ray Mechanical Data​

Parameters related to the X-Ray Mechanical System​

 

 

Available​

(Stand,Table)areavailabletotherecordingapplication.​

113945​ X-Ray Detector Data Available​Parameters related to the X-Ray Detector are available​

 

 

to the recording application.​

 

113946​

Projection Eponymous Name​ Describes the radiographic method of patient, tube and​

 

 

detectorpositioningtoachieveawelldescribedprojection​

 

 

or view.​

 

113947​

Detector Type​

Type of Detector used to acquire data. E.g., Images.​

 

113948​

Direct Detector​

Detector that directly transforms the input signal to pixel​

 

 

values.​

 

- Standard -​

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1265​

Code Value​

Code Meaning​

Definition​

Notes​

113949​

Indirect Detector​

Detectorthattransformsanintermediatesignalintopixel​

 

 

 

values. E.g., a scintillator-based detector.​

 

113950​

Storage Detector​

Storage detector that stores a signal that is later​

 

 

 

transformed by a reader into pixel values. E.g., a​

 

 

 

phosphor-based detector.​

 

113951​

Film​

Film that is scanned to create pixel values.​

 

113952​

Table Mount​

The cassette/detector is mounted in the patient table.​

 

113953​

Unmounted Detector​

The cassette/detector is not mounted.. E.g., a cassette​

 

 

 

placed underneath the patient.​

 

113954​

Upright Stand Mount​

The cassette/detector is mounted in an upright stand.​

 

113955​

C-Arm Mount​

The cassette/detector is mounted on a c-arm.​

 

113956​

CR/DR Mechanical​

Method of mounting or positioning a CR/DR cassette or​

 

 

Configuration​

detector.​

 

113957​ Fluoroscopy-Guided Projection​An integrated projection radiography system capable of​

 

 

Radiography System​

fluoroscopy.​

 

113958​

Integrated Projection​

A projection radiography system where the X-Ray​

 

 

Radiography System​

detector, X-Ray Source and gantry components are​

 

 

 

integrated and the managing system is able to access​

 

 

 

details of each component.​

 

113959​

Cassette-based Projection​

A projection radiography system where the X-Ray​

 

 

Radiography System​

detector, X-Ray Source and gantry components are not​

 

 

 

integrated. E.g., cassette-based CR and DR systems.​

 

113961​

Reconstruction Algorithm​

Description of the algorithm used when reconstructing​

 

 

 

the image from the data acquired during the acquisition​

 

 

 

process.​

 

113962​

Filtered Back Projection​

An algorithm for reconstructing an image from multiple​

 

 

 

projectionsbyback-projectingthemeasuredvaluesalong​

 

 

 

the line of the projection and filtering the result to reduce​

 

 

 

blurring.​

 

113963​

Iterative Reconstruction​

An algorithm for reconstructing an image from multiple​

 

 

 

projections by starting with an assumed reconstructed​

 

 

 

image,computingprojectionsfromtheimage,comparing​

 

 

 

the original projection data and updating the​

 

 

 

reconstructedimagebaseduponthedifferencebetween​

 

 

 

the calculated and the actual projections.​

 

113964​ At Surface of Patient​

A point at the surface of the patient within the irradiated​

 

 

 

area where the X-Ray beam enters the patient (i.e.​

 

 

 

towards the tube).​

 

113970​ Procedure Step To This Point​ The period of time from the start of a Procedure Step​

 

 

 

until the time point established by the context of the​

 

 

 

reference.​

 

113980​

Water Equivalent Diameter​

The diameter of a cylinder of water having the same​

 

 

 

X-Ray attenuation as the patient for a specified​

 

 

 

reconstructed slice (e.g., as described in [AAPM Report​

 

 

 

220]).​

 

- Standard -​

Page 1266​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

113981​

Water Equivalent Diameter​

The Size Specific Dose Estimation is computed using a​

 

 

Representative Value​

singlerepresentativevalueofWaterEquivalentDiameter.​

 

E.g., computed as per [AAPM Report 220] and used as​ the index into Table 1D (32cm phantom) or Table 2D​ (16cmphantom)of[AAPMReport204](i.e.,asdescribed​ in the Appendix of [AAPM Report 220]).​

The single value used may be a mean of the values​ across the entire scan range, or may be a value at a​ single location sufficiently representative of the body​ region.​

113982​ Water Equivalent Diameter​ The Size Specific Dose Estimation is computed using​ Integrated Across Scan Range​Water Equivalent Diameter values for a sample of slices​

across the entire scan range.​

E.g., computed as per [AAPM Report 220] and used as​ the index into Table 1D (32cm phantom) or Table 2D​ (16cmphantom)of[AAPMReport204](i.e.,asdescribed​ in the Appendix of [AAPM Report 220]).​

113983​ WaterEquivalentDiameterFrom​The Size Specific Dose Estimation is computed using​ Raw Data​ Water Equivalent Diameter values derived from Raw​

Data rather than reconstructed slices.​

E.g., used as the index into Table 1D (32cm phantom)​ orTable2D(16cmphantom)of[AAPMReport204](i.e.,​ as described in the Appendix of [AAPM Report 220]).​

113984​ WaterEquivalentDiameterFrom​The Size Specific Dose Estimation is computed using​ Localizer​ Water Equivalent Diameter values derived from a​

Localizer image.​

113985​ Series or Instance used for​ Water Equivalent Diameter​ estimation​

E.g., used as the index into Table 1D (32cm phantom)​ orTable2D(16cmphantom)of[AAPMReport204](i.e.,​ as described in the Appendix of [AAPM Report 220]).​

Unique identifier of the Series or Instance(s) used for​ Water Equivalent Diameter estimation, whether it be a​ Series of reconstructed single slice images or one or​ moreEnhancedMulti-frameimagesoraRawDataSeries​ or Instance.​

113986​

Z value of location of Water​

The Z location used for Water Equivalent Diameter​

 

Equivalent Diameter estimation​estimation at a single location whether it be computed​

 

 

using a reconstructed slice or Localizer or Raw Data.​

 

 

Specified as the Z component within the Patient​

 

 

Coordinate System defined by a specified Frame of​

 

 

Reference.​

113987​

AAPM 220​

A report describing methods of calculation of diameters​

 

 

of cylinders of water having the same X-Ray attenuation​

 

 

as reconstructed CT slices of patients described in​

 

 

[AAPM Report 220].​

114000​

Not a number​

Measurement not available: Not a number (per IEEE​

 

 

754).​

114001​

Negative Infinity​

Measurement not available: Negative Infinity (per IEEE​

 

 

754).​

114002​

Positive Infinity​

Measurement not available: Positive Infinity (per IEEE​

 

 

754).​

- Standard -​

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1267​

Code Value​

Code Meaning​

Definition​

Notes​

114003​

Divide by zero​

Measurement not available: Divide by zero (per IEEE​

 

 

 

754).​

 

114004​

Underflow​

Measurement not available: Underflow (per IEEE 754).​

 

114005​

Overflow​

Measurement not available: Overflow (per IEEE 754).​

 

114006​

Measurement failure​

Measurement not available: Measurement failure.​

 

114007​

Measurement not attempted​

Measurementnotavailable:Measurementnotattempted.​

 

114008​

Calculation failure​

Measurement not available: Calculation failure.​

 

114009​

Value out of range​

Measurement not available: Value out of range.​

 

114010​

Value unknown​

Measurement not available: Value unknown.​

 

114011​

Value indeterminate​

Measurement not available: Value indeterminate.​

 

114201​

Time of flight​

Measures the time-of-flight of a light signal between the​

 

 

 

camera and the subject for each point of the image.​

 

114202​

Interferometry​

Interferometry is a family of techniques in which waves​

 

 

 

are superimposed in order to extract depth information​

 

 

 

about the scanned object.​

 

114203​

Laser scanning​

Laserscanningdescribesthegeneralmethodtosample​

 

 

 

or scan a surface using laser technology.​

 

114204​

Pattern projection​

Projecting a narrow band of light onto a​

 

 

 

three-dimensionally shaped surface produces a line of​

 

 

 

illumination that appears distorted from other​

 

 

 

perspectivesthanthatoftheprojector.Itcanbeusedfor​

 

 

 

an exact geometric reconstruction of the surface shape.​

 

114205​

Shape from shading​

Atechniqueforestimatingthesurfacenormalofanobject​

 

 

 

by observing that object under different lighting​

 

 

 

conditions.​

 

114206​

Shape from motion​

Atechniqueforestimatingthesurfacenormalofanobject​

 

 

 

by observing that object under different motions.​

 

114207​

Confocal imaging​

An optical imaging technique used to increase optical​

 

 

 

resolution and contrast of a micrograph by using point​

 

 

 

illuminationandaspatialpinholetoeliminateout-of-focus​

 

 

 

light in specimens that are thicker than the focal plane.​

 

 

 

It enables the reconstruction of 3D structures from the​

 

 

 

obtained images.​

 

114208​

Point Cloud Algorithmic​

Point cloud that was calculated by an algorithm.​

 

114209​

Turntable Scan Method​

Scanning the object from different views by placing it on​

 

 

 

a rotating table.​

 

114210​

High resolution​

Higher resolution with a longer acquisition time.​

 

114211​

Fast mode​

Lower resolution with a shorter acquisition time.​

 

114213​

Iterative Closest Point​

An algorithm employed to minimize the difference​

 

 

 

between two clouds of points. It iteratively revises the​

 

 

 

transformation(translation,rotation)neededtominimize​

 

 

 

the distance between the points of two point clouds.​

 

114215​

Freehand​

Human controlled minimization of the distance between​

 

 

 

the points of two point clouds.​

 

114216​

Checkerboard​

Scanning the object from different views by placing it in​

 

 

 

front of a checkerboard pattern.​

 

120999​

Device Description​

Description in free form text describing the device.​

 

121000​

Unique Device Identifiers​

UniqueDeviceIdentifiers(UDIs)oftheentireequipment.​

 

- Standard -​

Page 1268​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

121001​

Quotation Mode​

Type of source for observations quoted from an external​

 

 

 

source.​

 

121002​

Quoted Source​

Reference to external source of quoted observations.​

 

121003​

Document​

Documentary source of quoted observations.​

 

121004​

Verbal​

Verbal source of quoted observations.​

 

121005​

Observer Type​

Type of observer that created the observations.​

 

121006​

Person​

Human observer created the observations.​

 

121007​

Device​

Automated device created the observations.​

 

121008​

Person Observer Name​

Name of human observer that created the observations.​

 

121009​ PersonObserver'sOrganization​Organizationorinstitutionwithwhichthehumanobserver​

 

 

Name​

is affiliated for the context of the current observation.​

 

121010​ Person Observer's Role in the​Organizational role of human observer for the context of​

 

 

Organization​

the current observation.​

 

121011​ Person Observer's Role in this​Procedural role of human observer for the context of the​

 

 

Procedure​

current observation.​

 

121012​

Device Observer UID​

Unique identifier of automated device that created the​

 

 

 

observations.​

 

121013​

Device Observer Name​

Institution-provided identifier of automated device that​

 

 

 

created the observations.​

 

121014​ Device Observer Manufacturer​Manufacturer of automated device that created the​

 

 

 

observations.​

 

121015​ Device Observer Model Name​Manufacturer-providedmodelnameofautomateddevice​

 

 

 

that created the observations.​

 

121016​ DeviceObserverSerialNumber​Manufacturer-provided serial number of automated​

 

 

 

device that created the observations.​

 

121017​

Device Observer Physical​

Location of automated device that created the​

 

 

Location During Observation​

observations whilst the observations were being made.​

 

121018​ Procedure Study Instance UID​Unique identifier for the Study or Requested Procedure.​

 

121019​

Procedure Study Component​ Unique identifier for the Performed Procedure Step.​

 

 

UID​

 

 

121020​

Placer Number​

Identifier for the Order (or Service Request) assigned by​

 

 

 

the order placer system.​

 

121021​

Filler Number​

Identifier for the Order (or Service Request) assigned by​

 

 

 

the order filler system.​

 

121022​

Accession Number​

Identifier for the Order (or Service Request) assigned by​

 

 

 

the department information system.​

 

121023​

Procedure Code​

Type of procedure scheduled or performed.​

 

121024​

Subject Class​

Type of observation subject.​

 

121025​

Patient​

A patient is the subject of observations.​

 

121026​

Fetus​

Fetus of patient is the subject of observations.​

 

121027​

Specimen​

Specimen is the subject of observations.​

 

121028​

Subject UID​

Unique Identifier of patient or fetus who is the subject of​

 

 

 

observations.​

 

121029​

Subject Name​

Name of patient who is the subject of observations.​

 

121030​

Subject ID​

Identifier of patient or fetus who is the subject of​

 

 

 

observations.​

 

- Standard -​

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1269​

Code Value​

Code Meaning​

Definition​

Notes​

121031​

Subject Birth Date​

Birth Date of patient who is the subject of observations.​

121032​

Subject Sex​

Sex of patient who is the subject of observations.​

 

121033​

Subject Age​

Age of patient who is the subject of observations.​

 

121034​

Subject Species​

Species of patient who is the subject of observations.​

 

121035​

Subject Breed​

The breed of the subject.​

 

121036​

Mother of fetus​

Name of mother of fetus that is the subject of​

 

 

 

observations.​

 

121037​

Fetus number​

 

Retired. Replaced by​

 

 

 

(11951-1, LN, "Fetus ID").​

121038​

Number of Fetuses​

 

Retired. Replaced by​

 

 

 

(55281-0, LN, "Number of​

 

 

 

Fetuses").​

121039​

Specimen UID​

Unique Identifier of specimen that is the subject of​

 

 

 

observations.​

 

121040​

Specimen Accession Number​ Accession Number of specimen that is the subject of​

Retired.​

 

 

observations​

 

121041​

Specimen Identifier​

Identifierofspecimenthatisthesubjectofobservations.​

121042​

Specimen Type​

Coded category of specimen that is the subject of​

Retired. Replaced by​

 

 

observations​

(371439000,SCT,"Specimen​

 

 

 

Type")​

121043​

Slide Identifier​

Identifierofspecimenmicroscopeslidethatisthesubject​Retired. Replaced by​

 

 

of observations​

(111700, DCM, "Specimen​

 

 

 

Container Identifier")​

121044​

Slide UID​

Unique Identifier of specimen microscope slide that is​ Retired.​

 

 

the subject of observations​

 

121045​

Language​

The language of the content, being a language that is​

 

 

 

primarily used for human communication. E.g., English,​

 

 

French.​

 

121046​

Country of Language​

The country-specific variant of language. E.g., Canada​

 

 

for Candadian French.​

 

121047​

Language of Value​

The language of the value component of a name-value​

 

 

pair.​

 

121048​ Language of Name and Value​Thelanguageofboththenamecomponentandthevalue​

 

 

component of a name-value pair.​

 

121049​ Language of Content Item and​ThelanguageofthecurrentContentItem(nodeinatree​

 

Descendants​

of content) and all its descendants.​

 

121050​ Equivalent Meaning of Concept​A precoordinated coded concept or text meaning for the​ Name​ namecomponentofaname-valuepairthatisequivalent​ tothepost-coordinatedmeaningconveyedbythecoded​

name and its concept modifier children.​

121051​ Equivalent Meaning of Value​ A precoordinated coded concept or text meaning for the​ value component of a name-value pair that is equivalent​ tothepost-coordinatedmeaningconveyedbythecoded​ value and its concept modifier children.​

121052​

Presence of property​

Whether or not the property concept being modified is​

 

 

present or absent.​

121053​

Present​

Retired. Replaced by​

 

 

(52101004, SCT, "Present")​

- Standard -​

Page 1270​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

121054​

Absent​

 

Retired. Replaced by​

 

 

 

(272519000, SCT, "Absent")​

121055​

Path​

Asetofpointsonanimage,thatwhenconnectedbyline​

 

 

segments, provide a polyline from which a linear​

 

 

 

measurement was inferred.​

 

121056​

Area outline​

Asetofpointsonanimage,thatwhenconnectedbyline​

 

 

segments, provide a closed polyline that is the border of​

 

 

adefinedregionfromwhichanarea,ortwo-dimensional​

 

 

measurement, was inferred.​

 

121057​

Perimeter outline​

Asetofpointsonanimage,thatwhenconnectedbyline​

 

 

segments, provide a closed polyline that is a​

 

 

 

two-dimensional border of a three-dimensional region's​

 

 

intersection with, or projection into the image.​

 

121058​

Procedure reported​

The imaging procedure whose results are reported.​

 

121059​

Presence Undetermined​

Presence or absence of a property is undetermined​

Retired. Replaced by​

 

 

 

(373068000, SCT,​

 

 

 

"Undetermined")​

121060​

History​

 

Retired. Replaced by​

 

 

 

(11329-0, LN, "History")​

121061​ Device Observer Manufacturer​Manufacturer-provided Class UID(s) of device that​

 

 

Class UID​

created the observations.​

 

121062​

Request​

 

Retired. Replaced by​

 

 

 

(55115-0, LN, "Request")​

121064​

Current Procedure Descriptions​

Retired. Replaced by​

 

 

 

(55111-9, LN, "Current​

 

 

 

Procedure Descriptions")​

121065​

Procedure Description​

A description of the imaging procedure.​

 

121066​

Prior Procedure Descriptions​

 

Retired. Replaced by​

 

 

 

(55114-3, LN, "Prior​

 

 

 

Procedure Descriptions")​

121068​

Previous Findings​

 

Retired. Replaced by​

 

 

 

(18834-2, LN, "Previous​

 

 

 

Findings")​

121069​

Previous Finding​

An observation found on a prior imaging study.​

 

121070​

Findings​

 

Retired. Replaced by​

 

 

 

(59776-5, LN, "Findings")​

121071​

Finding​

An observation found on an imaging study.​

 

121072​

Impressions​

 

Retired. Replaced by​

 

 

 

(19005-8, LN, "Impressions")​

121073​

Impression​

An interpretation in the clinical context of the finding(s)​

 

 

on an imaging study.​

 

121074​

Recommendations​

 

Retired. Replaced by​

 

 

 

(18783-1, LN,​

 

 

 

"Recommendations")​

121075​

Recommendation​

A recommendation for management or investigation​

 

 

 

based on the findings and impressions of an imaging​

 

 

 

study.​

 

121076​

Conclusions​

 

Retired. Replaced by​

 

 

 

(55110-1, LN, "Conclusions")​