Добавил:
Upload Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Thermal Analysis of Polymeric Materials

.pdf
Скачиваний:
147
Добавлен:
24.03.2015
Размер:
12.15 Mб
Скачать

Appendix 10–Extreme DTA and DSC

827

__________________________________________________________________

Fig. A.10.4

To accomplish and exceed the fast calorimetry just suggested, one can turn to integrated circuit calorimetry. The measuring methods may be modulated calorimetry, such as AC calorimetry or the 3 -method [7,8], thin-film calorimetry as shown in Fig. A.10.4, or may involve standard heating or cooling curves as well as DSC configurations as illustrated in Chap. 4. Modern versions of such fast heating calorimeters are based on silicon-chip technology for measuring heat flow using thermopiles integrated in the chip and resistors for heating. Increasingly more numerous chips have become available based on free-standing membranes of SiNx, produced by etching the center of a properly coated Si-chip, as illustrated in Figs. A.10.5 and A.10.6, which contain each a schematic of the calorimeter.

Figure A.10.5 illustrates the performance of several integrated circuit thermopiles, ICT, by Xensor Integration. The calibration without sample shows a constant output voltage of the sensor, decreasing at higher frequency, depending on the membrane used. The phase-angle response shows an analogous change at higher frequency. A linearity-check revealed that dynamic heat capacity measurements should be possible over the frequency range of 1 mHz to 100 Hz, compared to the 0.1 Hz limit typical for TMDSC of Sect. 4.4. This allows to analyze nanogram samples deposited from solution on the position shown in Fig. A.10.5. Similar chips for the measurements of heat capacity of samples of below 1 mg, contained in small aluminum pans, showed time resolutions well below 1 s when using heat-pulses [9].

Finally, Fig. A.10.6 illustrates the change in the crystallization rate when going from slow cooling, measured by standard DSC, to fast standard DSC, and to, ultimately, a chip calorimeter with 5,000 K s 1 (300,000 K min 1) [10]. The cooling rates were varied over five orders of magnitude by using different calorimeters. The cooling rates in the figure from right to left are: for 4 mg in a standard DSC, 0.02, 0.03, 0.08, 0.2, and 0.3 K s 1; for 0.4 mg in a fast DSC: 0.3, 0.8, 3, and 8 K s 1; for

828 Appendix 10–Extreme DTA and DSC

__________________________________________________________________

Fig. A.10.5

Fig. A.10.6

0.1 g in an IC calorimeter as shown in the sketch: 0.45, 0.9, 2.4, and 5 kK s 1. In this chip-calorimeter, the SiNx film, supported by the Si-chip frame was about 500 nm thick. The cooling was achieved by overcompensating the heater power with a cooled purge gas. The heating experiments could also be used to study the reorganization, as discussed in Sect. 6.2.

Appendix 10–Extreme DTA and DSC

829

__________________________________________________________________

The principle of microcalorimetry is illustrated with Fig. A.10.7 ( TA™ of TA Instruments, Inc.). The tip of an atomic force microscope, AFM, is replaced by a Ptwire that can be heated and modulated, as is illustrated in detail with Fig. 3.96. A typical resolution is about 1.0 m with heating rates up to 1,000 K min 1. A temperature precision of ±3 K and a modulation frequency up to 100 kHz has been reached. The figure shows the control circuit for localized thermal analysis. In this case the probe contacts the surface at a fixed location with a programmed force, controlled by the piezoelectric feedback of the AFM. A reference probe is attached next to the sample probe with its tip not contacting the sample, allowing for

Fig. A.10.7

differential measurements. Equal dc currents are applied to both probes and are controlled by the temperature-feedback circuit to achieve a constant heating rate. The temperature of the sample, Ts, can be determined, after calibration from the resistance of the probe. Since the resistance of platinum has an almost linear dependence on temperature between 300 and 600 K, Rs, can be expressed as shown in Fig. A.10.7, where Rso and are the resistance and the temperature coefficient of the platinum wire at To. The resistance of all lead wires and the heavy Wollaston-portion of the sensor are neglected. Their resistance is small when compared to the V-shaped platinum tip. In addition to the dc current, an ac current can be superimposed on the tip to obtain a temperature-modulation. The dc difference of power between the sample and reference probes is determined by measuring the dc voltage difference between two probes after low-pass filtering (LPF) and the ac voltage difference is measured by the lock-in amplifier (LIA). This arrangement permits an easy deconvolution of the underlying, dc and the reversing ac effects (see also Sect. 4.4). An example which illustrates a qualitative local thermal analysis with a microcalorimeter is given in Fig. 3.97, the limits were probed in [12].

830 Appendix 10–Extreme DTA and DSC

__________________________________________________________________

References for Appendix 10

1.Wunderlich B (2000) Temperature-modulated Calorimetry in the 21st Century. Thermochim Acta 355: 43–57.

2.White GK (1979) Experimental Techniques in Low Temperature Physics. Clarendon, Oxford, 3rd edn.

3.Gmelin E (1987) Low-temperature Calorimetry: A particular Branch of Thermal Analysis. Thermochim Acta 110: 183–208.

4.Lounasma OV (1974) Experimental Principles and Methods below 1 K. Academic Press, London; see also Bailey CA, ed (1971) Advanced Cryogenics. Plenum Press, New York.

5.Wu ZQ, Dann VL, Cheng SZD, Wunderlich B (1988) Fast DSC Applied to the Crystallization of Polypropylene. J Thermal Anal 34: 105–114.

6.Pijpers TFJ, Mathot VBF, Goderis B, Scherrenberg RL, van der Vegte EW (2002) Highspeed Calorimetry for the Study of the Kinetics of (De)vitrification, Crystallization, and Melting of Macromolecules. Macromolecules 35: 3601–3613.

7.Jeong Y-H (2001) Modern Calorimetry: Going Beyond Tradition. Thermochim. Acta

377:1–7.

8.Jung DH, Moon IK, Jeong Y-H (2003) Differential 3 Calorimeter. Thermochim. Acta

403:83–88.

9.Winter W, Höhne WH (2003) Chip-calorimeter for Small Samples. Thermochim. Acta

403:43–53.

10.Adamovsky SA, Minakov AA, Schick C (1003) Thermochim. Acta 403: 55–63. Figure modified from a presentation at the 8th Lähnwitz Seminar of 2004, courtesy of Dr. C. Schick.

11.Moon I, Androsch R, Chen W, Wunderlich B (2000) The Principles of Micro-thermal Analysis and its Application to the Study of Macromolecules. J Thermal Anal and Calorimetry 59: 187–203.

12.Buzin AI, Kamasa P, Pyda M, Wunderlich B (2002) Application of Wollaston Wire Probe for Quantitative Thermal Analysis, Thermochim Acta, 381, 9–18.

Appendix 11

831

__________________________________________________________________

Description of an Online Correction of the Heat-flow Rate

A basic analysis of the heat-flow rate in a DSC was derived in Sect. 4.3.6 with Figs. 4.69 and 4.70. To use these equations, special baseline calibrations are necessary, as illustrated in Fig. 2.29. With the development of data analysis by computer, it has become possible to make a number of corrections online, i.e., during the run. The Tzero™ method is the first of such programs [1]. Figure A.11.1 illustrates a conventional DSC baseline over an extensive temperature range of 450 K. Considerable improvement arises from the correction with the Tzero™ method before recording the data. The main correction concerns the asymmetry of the instrumentation which is not likely to change from run to run. An initial calibration sets the instrument parameters and leads to such a much improved performance. In this appendix the Tzero™ method of TA Instruments will be summarized [2].

To analyze the instrument performance, the DSC shown in Fig. 4.54 is used. Note that in the literature cited to this appendix the used quantities are represented by different symbols than used here [2,3]. The furnace or block temperature Tb is called often To (hence Tzero™ method). The heat-flow rate called dQ/dt = . is called q in [2], and should not to be confused with the here used rate of temperature change dT/dt = q. The temperature difference used in Chap. 4 and maintained here is written as T = Tr Ts, a positive quantity on heating of a sample run versus an empty pan, giving a positive differential heat flow into the sample. (In [2,3], in contrast, T is set equal to Ts Tr, making . proportional to T).

Next, the true sample and reference temperatures which determine the actual heatflow rates into the sample and reference pans are computed by modeling the DSC of Fig. 4.54 by an electric circuit which behaves analogously, as described in [4], for

Fig. A.11.1

832 Appendix 11–Online Correction of the Heat-flow Rate

__________________________________________________________________

example. The thermal resistances are equated to electrical resistances and the heat capacities with capacitors. The heat-flow rates are then given by electrical currents calculated from Ohms law and lead to similar answers as in Figs. 4.64–72.

In the sketch A of Fig. A.11.2 the equivalent electrical circuit for a conventional DSC measurement is drawn. The heat-flow rate into the sample calorimeter (pan + sample) is represented by .s, and .r is the heat-flow rate into the empty pan which is the reference calorimeter. The heat-flow rate into the sample itself should then be

. = .s .r and matches Eq. (3) of Fig. 4.69 when assuming the thermal resistances from the furnace to the measured temperatures are equal (Rspl = Rrpl = R):

(1)

Including also the heat capacities of the sample and reference platforms, Cspl and Crpl, into the model leads to the bottom sketch in Fig. A.11.2. The heat-flow rate into the sample needs now a four-term heat-flow-equation making use of a second temperature difference Tb = Ts Tb besides T = Tr Ts:

(2)

The second and third terms express the imbalance of the thermal resistances and heat capacities outside the calorimeters. The fourth term, the effect of different heating rates between reference and sample calorimeter. This last term is of importance when a transition occurs in the sample and does not follow the assumptions made for Figs. 4.71 and 4.72 [3]. (Note the differences to [2] from changed symbols and signs,

. = q, T = T, Tb = To, qs = dTs /dt, as well as from expanded subscripts).

Fig. A.11.2

Appendix 11–Online Correction of the Heat-flow Rate

833

__________________________________________________________________

The use of the corrections B in Fig. A.11.2 needs two calibration runs of the DSC of Fig. 4.54. The heat capacities of the calorimeter platforms, Cspl and Crpl, and the resistances to the constantan body, Rspl and Rrpl, must be evaluated as a function of temperature. First, the DSC is run without the calorimeters, next a run is done with sapphire disks on the sample and reference platforms without calorimeter pans. From the empty run one sets a zero heat-flow rate for .s and .r. This allows to calculate the temperature-dependent time constants of the DSC, written as s = CsplRspl andr = CrplRrpl, and calculated from the equations in the lower part of Fig. A.11.2. For the second run, the heat-flow rates are those into the sapphire disks, known to be mcpq, as suggested in Figs. 4.54 and 4.70. The heat-capacity-correction terms are zero in this second calibration because no pans were used. From these four equations, all four platform constants can be evaluated and the DSC calibrated.

As the next step, the effect of the pans must be considered. The differences between sample and reference pan cause similar heat-flow problems as the platforms and can be assessed by the diagram of Fig. A.11.3. The dash-dotted lines indicate the

Fig. A.11.3

sample and reference platforms from which .s and .r, contained in Eq. (3), enter into

the calorimeters. The heat-flow rate into the sample, .sample, however, is modified by the heat resistance between pan and platform, Rpn, and the heat capacity of the pan

Cspn. The actual sample temperature inside the calorimeter pan, in turn, is not Ts, but the temperature Tspn. For the reference side, one assumes the same thermal resistance Rpn, but because of the possibly different mass of the reference pan, a different Crpn. Assuming further that the reference calorimeter is empty, as is the usual operation

procedure, there is no .reference.

Using the two measured temperature differences T and Tb listed in Fig. A.11.3 and inserting them into the heat-flow rate expressions of sketch B of Fig. A.11.2 yields the two heat-flow rates of Fig. A.11.3 and their difference, expressed in

834 Appendix 11–Online Correction of the Heat-flow Rate

__________________________________________________________________

Eq. (2), above. To obtain the actual heat-flow rate into the sample, .sample, one has to compute the heat-flow rates into the two pans. Since both pans are out of the same

material Cspn = mspn cpan and Crpn = mrpn cpan, where cpan is the specific heat capacity of the pan material and m represents the corresponding mass. Since all .r goes into the

empty pan, one can use its value to compute .sample:

(3)

Insertion of the measured values for .s and .r, contained in Fig. A.11.3, leads to an

overall equation for .sample. The needed sample and reference pan temperatures in Eq. (3) can be calculated with some simple assumptions about the contact resistance

between pan and platform Rpn [2]: Tspn = Ts .s Rpn and Trpn = Tr .r Rpn. The commercial software includes typical values derived from data given in [5],

considering geometry, purge gas, and pan construction.

To summarize, the correction for asymmetry of the measuring platforms in Fig. 4.54 due to their thermal resistances and their heat capacities with Eq. (3) eliminates the main effect of the baseline curvature with temperature, as is shown in Fig. A.11.1. Further corrections include the mass differences of the pans and the thermal resistance between calorimeters and measuring platforms.

Remaining problems are the temperature gradients within the sample and their changes on heating or cooling, deformations of the sample pan and the accompanying changes in contact resistances which occur when samples expand or shrink. Also, the position of the calorimeters should be fixed and must not be altered during measurement, as can happen by vibrations or mechanical shocks. Changes of heat transfer due to radiation can be caused by fingerprints on calorimeters, platforms, and furnace cavity since fingerprints are easily converted at high temperatures to carbon specks with higher emissivity and absorptivity. Once cleaned, it is helpful to use dust-free and clean nylon gloves and to touch the calorimeters with tweezers only. The purge gas must not change in flux and pressure from the calibration and must not develop turbulent flow or convection currents. Finally, the room temperature must not fluctuate and set up changing temperature gradients within the DSC. Avoidance of all these potential errors is basic to good calorimetry.

References to Appendix 11:

1.Waguespack L, Blaine RL (2001) Design of A New DSC Cell with Tzero™ Technology. Proc 29th NATAS Conf in St. Louis, MO, Sept 24–26, Kociba KJ, Kociba BJ, eds 29: 721–727.

2.Danley RL (2003) New Heat-flux DSC Measurement Technique. Thermochim Acta 395:201–208.

3.Höhne G, Hemminger W, Flammersheim HJ (2003) Differential Scanning Calorimetry, 2nd edn. Springer, Berlin.

4.Holman JP (1976) Heat Transfer, 4th edn. McGraw-Hill, New York, pp 97–102.

5.Madhusudana CV (1996) Thermal Contact Conductance. Springer, New York.

Appendix 12

835

__________________________________________________________________

Derivation of the Heat-flow-rate Equations

The heat flow across any surface area, A, is given in Fig. A.12.1 by the heat-flow rate per unit area dQ/(Adt) in Eq. (1), a vector quantity in J m 2s 1. It is equal to the negative of the thermal conductivity,1 in J m 1s 1K 1, multiplied by the temperature gradient (dT/dr). Equation (2) represents the differential heat flow into the volume

Fig. A.12.1

V and can be derived from the definition of the heat capacity dQ/dT = mcp. The symbols have the standard meanings; is the density and cp, the specific heat capacity per unit mass, so that m = V .

Standard techniques of vector analysis allow to equate the heat flow into the volume V to the heat flow across its surface. This operation leads to the linear and homogeneous Fourier differential equation of heat flow, given as Eq. (3). The letter k represents the thermal diffusivity in m2 s 1, which is equal to the thermal conductivity 1 divided by the density and specific heat capacity. The Laplacian operator is 2 = 2/ x2 + 2/ y2 + 2/ z2, where x, y, and z are the space coordinates. In the present cylindrical symmetry, the Laplacian, operating on temperature T, can be represented as d2T/dr2 reducing the equation to one dimension. Equations (1) (3) form the basis for the further mathematical treatment of differential thermal analysis, as is given, for example by Ozawa T (1966) Bull. Chem. Soc. Japan 39: 2071.

The superposition principle is obeyed by the solutions of Eq. (3), i.e., the combined effect of a number of causes acting together is the sum of the effects of the causes acting separately. This allows the description of phase transitions by evaluating several separate solutions and adding them, as suggested in Fig. A.12.2. For a modeling of the glass transition in a DSC with Ri = 4 mm, see Fig. A.12.3.

836 Appendix 12–Derivation of the Heat-flow-rate Equations

__________________________________________________________________

Fig. A.12.2

Initially one finds the steady state of the glass with Eq. (4) of Fig. 6.65. At time t' the heat flux into the glass stops and it approaches the constant temperature T = qt' [= T1(Ri) at t'] with the transient T2 of Eq. (6), but at the same time t' there also begins a new transient, T3, for the liquid which reaches the new steady state at time , as described by Eq. (5) with t = t*. Note, that ks refers to the thermal diffusivity of the glass, and ks' to that of the liquid.

Fig. A.12.3