tibial orientation at 90° exion are the combination of KA, which resurfaces the patient’s pre-arthritic knee without releasing ligaments, and an insert with a medial 1:1 ball-in-socket and a lateral at articular surface along with PCL retention.
A. J. Nedopil et al.
Denition andAdvantages ofKinematically Aligned Total Knee Arthroplasty
Kinematically aligned (KA) total knee arthroplasty resurfaces the patient’s pre­arthritic knee to restore the articular surfaces, limb alignment, and joint line obliq­uity. Using components that restore the articular surface geometry and soft tissue constraints of the native knee, these steps align the knee’s three kinematic axes with the axes of the femoral and tibial components [1–3]. Resurfacing the pre-arthritic knee retains the native resting lengths of the collateral ligaments and the posterior cruciate ligament (PCL), thus eliminating the need for their release [4].
The high long-term implant survival rates at 7, 13, and 16years after KA TKA are comparable to or higher than survival rates after mechanically aligned TKA [5–7]. Furthermore, radio-stereometric analysis (RSA) of tibial baseplate migration after unrestricted kinematically aligned TKA with posterior cruciate ligament (PCL) retaining components has shown that the risk of tibial baseplate loosening at 1 and 2years is low and comparable to or less than mechanically aligned TKA, even up to 9° of varus alignment [8–10]. The high Forgotten Joint Scores (FJS) and Oxford Knee Scores (OKS) and the restoration of native knee medial and lateral tibial compartment forces can encourage surgeons to explore the advantages of resurfacing the patient’s pre-arthritic knee, thus avoiding costly and time- consuming physical therapy, and enabling early functional recovery [1, 11–17].
Performing theKA TKA Surgical Technique While Preserving thePCL withManual Instruments andCaliper Measurements Offers Greater Accuracy than Robotics
Because KA resurfaces the knee by matching the resection thickness of the distal and posterior femoral condyles to the thickness of the femoral component, KA TKA is a surgery based on millimeter measurements and not angular measurements.
10 Kinematic Approach
119
Surgical Steps forResurfacing thePre-Arthritic Femur withManual Instruments andFour Caliper Measurements
The surgeon relies on four caliper measurements of the femoral resections to verify that the femoral component resurfaces the patient’s pre-arthritic articular surface (Fig.10.1). After inspection of the distal and posterior femoral condyles, partially worn cartilage is removed to expose subchondral bone. Worn cartilage is compen­sated with augments that reduce the thickness of the resected condyle by 2mm [18] (Fig.10.2). After the caliper measurement of the resected condyle, each femoral resection is corrected so that its thickness equals that of the condyle of the femoral component after compensating 2mm for cartilage loss and 1mm for the kerf made by the saw blade to limit deviations from the pre-arthritic distal and posterior femo­ral articular surfaces to within 0±0.5mm [19].
Accuracy withManual Instruments
Two studies that used manual instruments to perform KA TKA reported that the most inaccurate distal and posterior medial femoral resections deviated from
0.9mm over-resection to a −1.5mm under-resection [20, 21]. These studies also reported the effect of surgeons’ experience and the learning curve of performing KA TKA accurately, which was negligible. Three studies that used robotic instrumenta­tion to perform mechanically aligned TKA reported the most inaccurate distal and posterior medial femoral resections deviated from a 2.3 mm over-resection to a
−0.9mm under-resection [22–24]. These deviations due to inaccuracy can be dev- astating because 1mm or more over-resections from the medial pre-arthritic femo­ral articular surface worsen the FJS (100 best, 0 worst) and OKS (48 best, 0 worst) by a minimum of 31 and 7 points [25].
Surgical Steps forResurfacing thePre-Arthritic Tibia withManual Instruments andFour Caliper Measurements
To set the proximal tibia parallel to the patient’s pre-arthritic varus-valgus (V-V) orientation and medial tibial slope, the surgeon rst sets the tibial resection guide 8mm distal to the tibial articular surface and parallel to its V-V orientation after visually compensating for cartilage and bone wear. Next, they set an angle wing, inserted through the tibial guide’s medial slot, parallel to the patient’s pre-arthritic slope. When stabilizing the tibial resection guide to the bone the surgeon adjusts the I-E orientation of the guide to match the anterior-posterior axis of the oval-shaped lateral tibial plateau. After performing the proximal tibial resection, two caliper measurements verify that the tibial resection sets the tibial component’s V-V
120
A. J. Nedopil et al.
Fig. 10.1 The image shows the intraoperative verication worksheet the surgeon completes for each kinematically aligned total knee arthroplasty. It includes the patient number, surgeon name, sex, age, BMI, date of surgery, right or left knee, type of primary deformity (varus, valgus, or patel­lofemoral), condition of ACL and PCL, target thickness of femoral resections, initial and corrected caliper-measured thickness of each femoral and tibial resection, and the implanted component sizes
10 Kinematic Approach
121
Fig. 10.2 The composite of images shows the manual instruments used to make the distal and posterior femoral resections. The features include an offset distal referencing guide with two holes (orange squares) for compression screws (upper left and right), a posterior referencing guide set at 0° with small, medium, and large width posterior feet with two holes (orange squares) for com­pression screws and removable shims to compensate for 2mm of distal and posterior cartilage wear (lower left), and a washer, available in 1 and 2mm (shown) thicknesses, to correct for an over-resection of wear (lower left), and a washer, available in 1 and 2mm (shown) thicknesses, to correct for an over-resection of a distal femoral condyle (lower right)
orientation, and the inspection of the medial tibial resection in conjunction with two caliper measurements veries that the tibial resection matches the patient’s pre­arthritic slope. When necessary, differences between the medial and lateral thick­nesses of the tibial resection are ne-tuned by recutting the tibia to correct the difference. Equalization of these two thicknesses is the initial verication that the V-V orientation of the tibial resection is correct. When necessary, differences between the medial anterior and posterior thicknesses are ne-tuned by recutting the tibia to correct the difference, which sets the posterior slope [26].
122
A. J. Nedopil et al.
Protecting thePCL by Inserting aPin Anterior tothePCL Fossa
To maintain the PCL’s integrity, the surgeon can drill a pin anterior to the PCL fossa, which blocks the saw from posterior translation. To verify the PCL’s integrity, the surgeon should ex the knee to 90°, inspect, and palpate the PCL to conrm its maintained tension (Fig.10.3).
Three Surgical Steps forBalancing theKA TKA that Use aSpacer Block andtheInsert Goniometer
At 90° exion, the surgeon chooses the spacer block thickness that restores a medial pivot. They insert the spacer block between the posterior femoral condyles and the proximal tibia and gently pivot the spacer block at its handle. If the spacer block pivots medially, i.e., the medial portion of the spacer block is pinched between the medial condyle and tibia while its lateral portion translates anteriorly posteriorly, the surgeon extends the knee to examine the extension space. If the spacer block pivots laterally or centrally, the surgeon increases the thickness of the spacer block until a medial pivot is restored. They then proceed to examine the extension space using the same spacer block thickness.
At full extension, the surgeon conrms that the spacer block thickness restores a rectangular extension space. They apply a V-V moment to the knee and look for lift- off of the medial and lateral femoral and tibial resection from the spacer block.
Insert Goniometer
Medial 1:1 Ball-In-
Socket
Articular Surface
0˚
Screw-Home
Tibial Screw-Home (Extension)
Fig. 10.3 The composite of a right knee with valgus deformity demonstrates an additional tech­nique to preserve the PCL’s insertion when resecting the proximal tibia (left) and the conrmatory step to ensure its integrity (right)
Axis of Rotation
-15˚
Internal
Orientation
Lateral Flat
Articular Surface
90˚ Flexion
10 Kinematic Approach
123
The tibial resection is ne-tuned by removing additional tibia until the extension space becomes rectangular [27]. The surgeon refrains from releasing collateral liga­ments and the PCL.Hence, when a KA TKA appears too tight, the surgeon read­justs the tibial resection so that the tibial component’s orientation resurfaces the knee’s pre-arthritic articulation.
The nal step of balancing the KA TKA occurs with trial components at full extension and 90° exion. The surgeon uses the insert goniometer to determine the optimal insert thickness after matching an anatomically shaped tibial baseplate to the tibial resection surface [28, 29]. The design of the insert goniometer consists of markings in front of the medial compartment that are calibrated to the axis of rota­tion and dene the orientation of the insert in relation to the femoral component. The insert goniometer measures tibial orientation relative to a reference line, laser etched vertically on the center of the trial femoral component’s medial condyle. Combining an angular scale on the medial insert and a vertical line on the medial femoral condyle enables the surgeon to measure tibial orientation in extension and at 90° exion intraoperatively during passive knee positioning (Fig.10.4).
The surgeon selects the ideal insert thickness by observing changes between dif­ferent insert thicknesses in the external orientation of the tibia in extension and internal tibial orientation in 90° exion as measured by an insert goniometer. The surgeon records the extension and exion tibial orientations for each insert
Right Knee with Valgus Deformity
A pin (cyan arrow), inserted
anterior to the PCL insertion,
protects the PCL
Fig. 10.4 The composite shows left trial components, which measure the tibial internal-external orientations relative to the femur in extension and at 90° exion. The markings in the front of the insert’s medial compartment are calibrated to the axis of rotation within the medial socket (top). The intersection of the vertical line along the femoral component’s medial condyle with the insert’s markings denes the tibial orientation, which measured 0° in extension (bottom left) and −15° internal at 90° exion (bottom right)
Inspection and palpation of the
PCL (framed by cyan lines)
confirms its integrity
124
thickness. The optimal thickness is the insert that provides the greatest orientations in extension and/or 90° exion [28].
A. J. Nedopil et al.
Restoring ACL Stability Requires aMedial 1:1 Ball-in-Socket Insert Conformity toRestore aMedial Pivot
The kinematics of the knee in the axial plane during extension largely depend on the intact anterior cruciate ligament (ACL) and the geometry of the articular surfaces [30, 31]. During knee exion-extension, the medial femoral condyle is a sphere (=ball) that rotates on the medial tibial plateau and the medial meniscus (=socket) to produce a combination of exion-extension and I-E rotation [31]. Because the medial condyle hardly translates from anterior to posterior during a exion motion, the medial compartment of the knee is analogous to a constrained ball-in-socket joint. The ACL aids in the medial condyle’s anterior-posterior (A-P) stability. The stable position of the medial condyle contributes to the isometry of the medial col­lateral ligament (MCL) during range of motion [32]. A medial 1:1 ball-in-socket insert conformity restores the stable position of the femoral component’s medial condyle during gait, downhill walking, and stair descent in contrast to low­conforming PCL retaining or posterior stabilized insert conformities [33, 34]. In vivo uoroscopic evaluation of KA TKA with PCL retention and a 1:1 ball-in­socket insert conformity reported restoration of close to native tibiofemoral kine­matics. In contrast, a low-conforming PCL retaining insert exhibits a loss of a medial pivot indicative of abnormal tibiofemoral kinematics [1, 35]. Restoring a native medial pivot with a 1:1 medial ball-in-socket insert conformity improves patient-reported outcomes when implanted with KA principles [12, 36].
Restoring I-E Tibial Rotation Requires anIntact PCL, aMedial 1:1 Ball-In-Socket Insert withaLateral Flat Articular Surface that Mimics theMorphology oftheNative Knee
Tibial External Orientation inExtension Tightens theLateral Collateral Ligament
While the medial femoral condyle does not translate, the lateral condyle exes but also translates from anterior to posterior during a exion motion. The convex shape of the lateral tibial plateau and the mobile lateral meniscus permit this coupled motion of the lateral condyle. The coupling of a stable medial condyle with a lateral condyle that translates from anterior to posterior during exion translates into inter­nal tibial rotation during exion. The anterior position of the lateral condyle (i.e.,
10 Kinematic Approach
external tibial orientation) during knee extension increases the distance between the lateral collateral ligament’s femoral origin and its insertion on the bula head, thus tightening the lateral collateral ligament (LCL) [32]. Hence, the external tibial rota­tion during terminal knee extension (=screw-home) is essential to tension the LCL and to restore a rectangular extension space.
125
Tibial Internal Orientation at 90° Flexion Reduces theQ-Angle
As the native knee exes, the PCL tension drives the tibia’s internal rotation. This, in turn, reduces the Q-angle and optimizes the tension in the retinacular ligaments, which enables optimal patellofemoral tracking during knee exion [37–39]. Ideally, a TKA should restore this native tibial internal orientation at 90° exion, promoting patellofemoral tracking. As the native knee moves from extension to 90° exion, the tibia internally rotates by an average of 17°–18° [37, 40–42].
PCL Retention Promotes Internal Tibial Rotation During Flexion
A well-known consequence of PCL excision is an increased exion space laxity, which is inconsistent and unpredictable between knees [43–45]. In addition, PCL excision causes a loss of about half of the internal tibial rotation from extension to 90° exion [39]. Trying to compensate for the loss of internal tibial rotation by reducing the posterior slope does not solve the problem [46]. In a cadaveric study, four surgeons performed a KA TKA with a medial ball-in-socket and lateral at tibial articular design. Before and after PCL excision, they compared the internal­external orientation of the tibia in extension and at 90° exion. With retention of the PCL, the mean internal tibial orientation was 15°, whereas with excision of the PCL, the mean internal tibial orientation decreased to 7°. This implies that retention of the PCL helped the prosthetic knee achieve mean internal tibial orientation (15°) that was comparable to the native knee (18°) (Fig.10.5). The loss of PCL-driven internal tibial rotation can, in turn, predispose the knee to patella maltracking and anterior knee pain.
126
A. J. Nedopil et al.
Right KA TKA with 1:1 Medial Ball-in-Socket
and Lateral Flat Articular Surface
a
25
20
15
10
5
0
Internal Tibial Rotation from
Extension to 90˚ Flexion (˚)
Fig. 10.5 The bar chart illustrates the difference in internal tibial rotation during cadaveric KA TKA, which occurs from a exion motion from 0° to 90°, before (a) and after PCL excision (b), which caused a loss of about half of the internal tibial rotation
KA TKA with PCL Retention (A)
KA TKA with PCL Excision (B)
15
b
7
An Insert withMedial 1:1 Ball-In-Socket Conformity andaLateral Flat Articular Surface Restores I-E Tibial Rotation
The sagittal plane conformity of the insert on the medial side determines passive knee kinematics. Medial Stabilized (MS) inserts from different manufacturers have varying levels of spherical conformity in the medial compartment. A study of 21 patients undergoing caliper-veried KA TKA with PCL retention compared the degree of internal tibial rotation between inserts with a 1:1 medial ball-in-socket or spherical conformity and one with a less than spherical medial ball-in-socket con­formity. Both inserts had an identical lateral at articular surface [47]. The mean tibial orientation for both insert conformities was similar in extension. In 90° and 120° exion, the 1:1 ball-in-socket conformity demonstrated 2° and 3° more inter­nal tibial orientation, respectively (Fig.10.6). Inserts with less than spherical con­formity allow the tibia to move anteriorly, thereby reducing the tension in the PCL.PCL tension is required for optimal internal tibial rotation. However, whether this minor increase in internal tibial rotation is clinically relevant remains to be seen.
An equally important condition for restoring native knee kinematics, in addition to a medial 1:1 ball-in-socket conformity, is the sagittal plane geometry of the lat­eral compartment of the insert. Posterior stabilized (PS), cruciate retaining (CR),
10 Kinematic Approach
127
a
b
c
KA TKA with 1:1 medial ball-in-socket conformity (A)
KA TKA with less than spherical medial conformity (B and C)
25
20
15
10
5
0
Internal Tibial Rotation from
Extension to 90˚ Flexion (˚)
Fig. 10.6 The bar chart illustrates the difference in PCL-driven internal tibial rotation during KA TKA, which occurs during a exion motion from 0° to 90° and 120°, between an insert with a lateral at articular surface and a 1:1 medial ball-in-socket conformity (a) and a less than spherical medial conformity (b and c). The insert with less than spherical medial conformity reduces internal tibial rotation by 2° to 3°
19
17
90˚ 120˚
Knee Flexion Angle
23
20
and ultra-congruent (UC) insert geometries that have an anterolateral and postero­lateral rim cause a mechanical stop of the tibial screw-home during knee extension and block internal tibial rotation during knee exion. A study quantitated the reduc­tion in tibial rotation caused by an anterolateral and posterolateral rim. Two sur­geons treated 23 patients undergoing KA with PCL retention [48]. They determined the difference in tibial rotation between inserts with a lateral at articulation surface (F), low-congruent (LC), and ultra-congruent (UC) inserts. Compared to the lateral at surface, the LC and UC surfaces decreased the magnitude of tibial rotation by reducing the external tibial orientation in extension and internal tibial orientation at 90° exion (Fig.10.7). Thus, a medial ball-in-socket insert with a lateral at articu­lar surface allows the PCL to drive tibial rotation during knee extension and exion. This axial plane motion is less likely with other insert geometries with lateral con­cavities, where the posterolateral rim functions as a “chock-block” [39].