Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3676_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
43 Мб
Скачать
416
https://t.me/med1917
Fig. 14.57 80 KVp technique with 25 mL contrast: 3D Volume rendered image of aorta with reference levels
T.A. Hameed
Fig. 14.58 Tricuspid aortic valve with calcification
Fig. 14.59 80 KVp technique with 25 mL contrast. 3D
Volume rendered image of aorta and iliac arteries
14 Transcatheter Aortic Valve Replacement Planning
https://t.me/med1917
417
14.7.2 CT Technique
Timing run 1: 1 mL@3.5/s, saline push 20 mL@3.5/s:
22 s
Timing run 2: 2 mL@3.5/s, saline push25mL@3.5/s:
31 s
CT scanner: 4 cm long detector array, 128- channel
MDCT
Scan coverage: Aortic arch to common femoral
arteries (Single scan acquisition) KVp 80 mAs 1260 Scan time 25 s Scan delay 28 sec
Contrast Injection: 25 mL (10 mL@3/s +
15 mL@2/s) + saline push 30 mL@2/s
Images demonstrate diagnostic quality exami­nation with enhancement density > 200 HU in the aortic root and iliac arteries (Figs. 14.50 and
14.51). Images acquired in standard locations
and planes demonstrate aortic annulus with effec­tive diameter of 28.4 mm. There is a trileaflet valve with moderate calcifications..
The patient was treated by implantation of 29 mm Sapien 3 valve via transfemoral approach (Fig. 14.60). Chest radiograph after TAVR dem­onstrates the prosthetic valve in place (Fig. 14.61).
Fig. 14.60 Fluoroscopy image during placement of 29 mm Sapien 3 valve via transfemoral approach
Fig. 14.61 Chest radiograph with transcatheter heart valve in place
418
https://t.me/med1917
T.A. Hameed
14.7.3 Diagnosis
CTA for pre-TAVR planning with low contrast dose with good diagnostic quality.
14.7.4 Discussion
The risk of CIN must be considered relative to the potential benefits of the contrast enhanced exami­nation. The risk of CIN is higher with larger con­trast dose and the CTA examination may thus be obtained with as low a contrast dose as possible. The goal is to attain adequate contrast enhance­ment in the aorta and iliac arteries. For a given vol­ume of contrast, the attenuation is higher with 80 KV compared to 120 KV. The contrast enhance­ment in blood vessels also depends on the volume and concentration of iodinated contrast. Use of bolus timing run helps in determining contrast arrival time and expected peak enhancement to determine appropriate time to scan (scan delay). With standard bolus triggering technique, there is a variable scan delay for the scanner to get ready to scan, usually around 4–5 s, requiring a larger vol­ume of contrast to be injected. Bolus timing run to determine the time of contrast arrival can be used to plan the start of injection more accurately with­out the need for scan delay after bolus triggering.
14.7.5 Pearls and Pitfalls
The contrast injection duration is selected to be close to the scan duration. In order to achieve higher attenuation in the aorta, the time to begin CT scanning after contrast injection (scan delay) should be near the end of contrast bolus train, i.e., contrast bolus arrival time + injection duration. Due to overall higher attenuation achieved with 80 KVp, injection rate can be slower (~2 mL/s) compared to 4–5 mL/s otherwise used with 120 KVp imaging. Biphasic injection (3–3.5 mL/s followed by 2 mL/s) can be useful in achieving a higher initial attenuation which is sustained by a slower, longer injection for scan duration.
14.8 Case 7
14.8.1 History
An 85-year-old with severe symptomatic aortic stenosis and decreased renal function; for pre­TAVR evaluation CT with low contrast dose.
14 Transcatheter Aortic Valve Replacement Planning
https://t.me/med1917
419
14.8.2 Findings
ECG-gated helical CTA examination of the chest, abdomen, and pelvis performed with total con­trast volume of 34 mL with 80 KV technique (Figs. 14.62, 14.63, 14.64, 14.65, 14.66, 14.67,
14.68, 14.69, 14.70, 14.71, 14.72, 14.73, 14.74,
and 14.75).
Scout image demonstrates a large patient with transverse diameter of greater than 500 cm at the level of pelvis (Fig. 14.62).
The examination was performed with 80 KVp on 8 cm long MDCT detector array (256-slice 128-channel) CT scanner. Two bolus timing runs using 1 and 2 mL in ascending aorta and in femo­ral arteries demonstrate time of contrast arrival at 17 s and 23 s, respectively (Figs. 14.63 and
14.64).
Two separate sets of CT images through the chest (ECG gated) (Figs. 14.65, 14.66, 14.67,
14.68, 14.69, and 14.70) and abdomen/pelvis
(Figs. 14.71, 14.72, 14.73, 14.74, and 14.75) (without ECG gating) were acquired using 15 mL for chest and 16 mL for the abdomen and pelvis, respectively. CTA images through the chest are of adequate quality showing the aortic annulus area of 584 mm2 (Fig. 14.70). There is moderate calci-
fication in the aortic valve (Fig. 14.69). 3D VR image shows no significant tortuosity of thoracic aorta (Fig. 14.70).
The CTA images through lower abdomen and pelvis (Figs. 14.71, 14.72, 14.73, and 14.74) demonstrate good diagnostic quality with normal size of aorta and iliac arteries without significant plaque or tortuosity (Fig. 14.75).
Fig. 14.62 Scout radiograph—transverse diameter of greater than 500 cm at the level of pelvis
420
https://t.me/med1917
T.A. Hameed
Fig. 14.63 80 KVp, 1 mL contrast. Timing run—in ascending aorta
Fig. 14.64 80 KVp, 2 mL contrast. Timing run—in femoral arteries
14 Transcatheter Aortic Valve Replacement Planning
https://t.me/med1917
Fig. 14.65 Oblique coronal view for selection of aortic annulus plane
421
Fig. 14.67 KVp 80, mAs 1650, ECG gated: 15 mL con­trast for chest. Aortic valve annulus
Fig. 14.66 Oblique sagittal view for selection of aortic annulus plane
Fig. 14.68 Aortic annulus area measured with annulus area of 583.75 mm sq
422
https://t.me/med1917
Fig. 14.69 Aortic Valve with moderate calcification
T.A. Hameed
Fig. 14.71 80 KVp, 16 mL contrast for abdomen and pelvis: axial image, infrarenal aorta, metallic hardware in spine
Fig. 14.70 80 KVp, 15 mL contrast. 3D VR: thoracic aorta
Fig. 14.72 Axial image: Adequately enhanced common iliac arteries
Fig. 14.73 Axial image: Adequately enhanced external iliac arteries
14 Transcatheter Aortic Valve Replacement Planning
https://t.me/med1917
Fig. 14.74 Axial image. Adequately enhanced femoral arteries
423
Fig. 14.75 KVp 80, mAs 1500, Non-gated. 16 mL con­trast. 3D VR. Abdominal aorta without tortuosity and mild to moderately tortuous iliac arteries
424
https://t.me/med1917
T.A. Hameed
14.8.3 Diagnosis
CTA with adequate quality for pre-procedure evaluation of TAVR with low contrast dose.
14.8.4 Discussion
The patient is of large size and has metallic hard­ware in the spine. Factors affecting contrast enhancement in aorta include patient size and there is decreased enhancement with higher body mass index (BMI). Use of imaging with 80 KV will increase the contrast enhancement for a given contrast dose. However, low KV is associ­ated with higher noise, which is to be compen­sated by increasing the tube current by several fold and this can become a limiting factor partic­ularly in large patients. ECG-gated imaging requires even higher tube current for optimal image quality compared to non-gated imaging. In this case with large patient size and imaging with 80 KVp, there is high noise in the ECG-gated images. However, as CTA utilizes high contrast resolution, diagnostic quality can still be obtained with relatively high noise compared to low con­trast resolution needed for imaging of viscera or other extravascular soft tissues.
For the second set of images through the abdo­men and pelvis, a high tube current (1500 mAs) was possible due to non-gated imaging over a shorter distance (abdomen and pelvis compared to combined chest, abdomen, and pelvis) which is able to provide good image quality with low noise (<25 HU) even with large patient size and
metal in spine. Adequate density of contrast enhancement in the arteries was achieved with only 15–16 mL contrast due to 80 KV and opti­mal timing using bolus timing run.
14.8.5 Pearls and Pitfalls
Image noise is decreased with higher tube current for a given tube voltage. Newer image recon­struction techniques utilizing iterative techniques and iterative model based reconstruction are very helpful in reducing noise. The image noise can also be reduced with the use of a smooth filter (soft tissue) compared to a high resolution filter (cardiac), which may not be necessary in case of imaging of large vessels such as aorta.
14.9 Case 8
14.9.1 History
A 77-year-old with failed prosthetic aortic valve and decreased renal function for pre-TAVR eval­uation CT with low contrast dose.
14.9.2 Findings
CTA performed with a total of 28 mL intravenous contrast dose using two separate image acquisi­tions demonstrates adequate image quality for assessment of aortic root including the biopros­thetic valve and aorta and iliac arteries.
14 Transcatheter Aortic Valve Replacement Planning
https://t.me/med1917
425
14.9.3 Technique
Two timing bolus runs using 1 mL contrast each were performed. Two separate contrast injections using 17 mL and 9 mL contrast were used for two separate CT imaging through chest (ECG gated) and abdomen/pelvis (non-gated), respectively.
Timing run 1: 1 mL@3.5/s, saline 20 mL@3.5/s: 18 s Timing run 2: 1 mL@3.5/s, saline 25 mL@3.5/s: 23 s
CTA Chest:
Helical ECG gated KVp 80 mAs 1525 Scan time 5 s 17 mL (7 mL@3.5/s, 10 mL@2/s) + NS 30 mL@2/s
CTA Abdomen and Pelvis:
Helical non-gated KVp 80 mAs 615 Scan time 5 s Contrast: 9 mL (3 mL@3/s, 6 mL@2/s) + NS
20 mL@2/s
The images demonstrate adequate image qual­ity with density of greater than 200 HU in the aortic root (Figs. 14.76, 14.77, 14.78, 14.79,
14.80, 14.81, 14.82, 14.83, 14.84, 14.85, 14.86,
and 14.87). The density in iliac arteries is less than 200 HU but adequate for diagnosis. There is bioprosthetic valve in place. The virtual ring drawn at the superior margin of the posts shows a relatively spacious sinus with ring to left coro­nary artery ostium distance of 7 mm placing this at low risk for coronary occlusion.
Fig. 14.76 80 KVp technique: Bolus timing run with 1 mL at ascending aorta with attenuation curve showing peak enhancement at 18 s