Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_103_библиотеки_им_акад_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
44 Мб
Скачать
17 Intraocular Pressure
209
17.10 Bioresonator Applanation Resonance Tonometer (ART)
Technology—Eklund et al. (2003) proposed a new tonometer based on the resonance principle. It estimates IOP by a simultaneous continuous sampling of the force needed to applanate the cornea and the corresponding area of contact [41]. It is commercially available as the Bioresonator ART (Medical sensors and Instruments, BioResonator AB, Sweden) in man­ual and automatic versions. In the manual (BioResonator ART manual) version, the sensor is pushed manually toward the cornea, and in the automatic version (BioResonator ART servo), this motion is servo-controlled(with a small motor that controls the sensor movement). The instrument gives the median of multiple IOP measurements and a quality score that denotes the standard deviation of the recordings.
A sensor and a transducer continuously mea­sure the contact force along with the area of con­tact. A cylindrical piezoelectric element in the resonance sensor has a known resonance fre­quency. When it comes in contact with the cornea, the acoustic impedance changes the resonance frequency with a shift in the frequency propor­tional to the area of contact between the sensor and the cornea. The IOP is calculated using a complex formula that considers the slope of the relationship between the force and frequency in a specic frequency shift with respect to the appla­nation area interval between 4.3 and 11.0mm2 [41]. The entire apparatus is slit-lamp mounted like in GAT, or can be handheld. It warrants all usage precautions that apply to GAT; i.e., local anesthesia must be applied before IOP measure­ments can be taken, and the probe must be thor­oughly sterilized before use.
The advantages of this technique are as fol­lows: (a) there is no need for the uorescein dye, (b) contact time with the cornea is less than that for GAT (hence, the chances of corneal abrasion damage are reduced) [42], (c) it is portable, and (d) it is self-calibrated.
The disadvantages of this technique are: (a) the patient needs to remain still during measure-
ment, (b) IOP may be overestimated at higher ranges and if the alignment is off-center [42,
43], (c) it needs adequate surface moisture on
the cornea to get good acoustic impedance (therefore, it is difcult to use in people with dry eye and ocular surface disorders) [42]. In addi­tion, ART is also affected by CCT and other cor­neal biomechanical factors [43].
17.11 Conclusion
There are numerous commercially available tonometers, each with advantages and disad­vantages. However, the IOP measurements of these are not interchangeable. Despite several recent developments and promising early reports, none of the alternative tonometers have gained widespread use in routine clinical prac­tice. The Goldmann applanation tonometer remains the gold standard despite some limita­tions in its use in certain clinical settings. Although an ideal tonometer does not exist, it is preferable to get consistent and reliable read­ings every time with the same type of equip­ment for each patient to diagnose and treat people with glaucoma.
Funding Hyderabad Eye Research Foundation, Hyderabad, India
Disclosure Ramyashri S: None; Aparna Rao: None; Sardar M Khan: None.
References
1. Da Silva F, Lira M. Intraocular pressure measure­ment: a review. Surv Ophthalmol. 2022;67:1319–31.
2. Okafor KC, Brandt JD.Measuring intraocular pres­sure. Curr Opin Ophthalmol. 2015;26:103–9.
3. Garcia-Feijoo J, Martinez-de-la-Casa JM, Morales­Fernandez L, etal. New technologies for measuring intraocular pressure. Prog Brain Res. 2015;221:67–79.
4. ElMallah MK, Asrani SG. New ways to mea­sure intraocular pressure. Curr Opin Ophthalmol. 2008;19:122–6.
5. Goldmann H, Schmidt T. Über Applanationstonometrie. Acta Ophthalmol. 1957;134:221–42.
210
R. S et al.
6. Stamper RL.A history of intraocular pressure and its measurement. Optom Vis Sci. 2011;88:e16–28.
7. Kotecha A, White E, Schlottmann PG, Garway-Heath DF.Intraocular pressure measurement precision with the Goldmann applanation, dynamic contour, and ocular response analyzer tonometers. Ophthalmology. 2010;117:730–7.
8. Brandt JD, Beiser JA, Kass MA, Gordon MO.Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology. 2001;108:1779–88.
9. Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol. 1993;38:1–30.
10. Punjabi OS, Kniestedt C, Stamper RL, Lin SC.Dynamic contour tonometry: principle and use. Clin Experiment Ophthalmol. 2006;34:837–40.
11. Kamppeter BA, Jonas JB.Dynamic contour tonom­etry for intraocular pressure measurement. Am J Ophthalmol. 2005;140:318–20.
12. Kanngiesser HE, Kniestedt C, Robert YC.Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma. 2005;14:344–50.
13. Willekens K, Rocha R, Van Keer K, etal. Review on dynamic contour tonometry and ocular pulse ampli­tude. Ophthalmic Res. 2015;55:91–8.
14. Moghimi S, Torabi H, Fakhraie G, et al. Dynamic contour tonometry in primary open angle glaucoma and pseudoexfoliation glaucoma: factors associated with intraocular pressure and ocular pulse amplitude. Middle East Afr J Ophthalmol. 2013;20:158–62.
15. Sullivan-Mee M, Billingsley SC, Patel AD, et al. Ocular response analyzer in subjects with and without glaucoma. Optom Vis Sci. 2008;85:463–70.
16. Terai N, Raiskup F, Haustein M, etal. Identication of biomechanical properties of the cornea: the ocular response analyzer. Curr Eye Res. 2012;37:553–62.
17. Asano S, Asaoka R, Yamashita T, et al. Visualizing the dynamic change of ocular response analyzer waveform using variational autoencoder in asso­ciation with the peripapillary retinal arteries angle. Sci Rep. 2020;10:6592. https://doi.org/10.1038/
s41598- 020- 63601- 8.
18. Aoki S, Murata H, Matsuura M, etal. The relationship between the waveform parameters from the ocular response analyzer and the progression of glaucoma. Ophthalmol Glaucoma. 2018;1:123–31.
19. Ting DSW, Pasquale LR, Peng L, et al. Articial intelligence and deep learning in ophthalmology. Br J Ophthalmol. 2019;103:167–75.
20. Nair M, Tagare S, Venkatesh R, Odayappan A. Articial intelligence in glaucoma. Indian J Ophthalmol. 2022;70:1868–9.
21. Youse S, Pasquale LR, Boland MV.Articial intel­ligence and glaucoma: illuminating the black box. Ophthalmol Glaucoma. 2020;3:311–3.
22. Chaurasia AK, Greatbatch CJ, Hewitt AW.Diagnostic accuracy of articial intelligence in glaucoma screening and clinical practice. J Glaucoma. 2022;31:285–99.
23. Yuan X, Zhang J.Real-time monitoring of intraocular pressure in glaucoma patients using wearable Mobile medicine devices. J Healthc Eng. 2022;2022:2271937.
https://doi.org/10.1155/2022/2271937.
24. Yang C, Huang X, Li X, etal. Wearable and implantable intraocular pressure biosensors: recent Progress and future prospects. Adv Sci (Weinh). 2021;8:2002971.
https://doi.org/10.1002/advs.202002971.
25. Shah JV, Albors GO, Williams J, etal. Clinical study of the IOPTx™ system—an electroceutical wear­able to lower intraocular pressure. Curr Eye Res. 2021;46:1531–8.
26. Beltran-Agulló L, Buys YM, Jahan F, etal. Twenty­four- hour intraocular pressure monitoring with the SENSIMED triggersh contact lens: effect of body posture during sleep. Br J Ophthalmol. 2017;101:1323–8.
27. Ittoop SM, SooHoo JR, Seibold LK, etal. Systematic review of current devices for 24-h intraocular pressure monitoring. Adv Ther. 2016;33:1679–90.
28. Szurman P, Gillmann K, Seuthe AM, et al. EYEMATE-SC Study Group. EYEMATE-SC trial: twelve-month safety, performance, and accuracy of a Suprachoroidal sensor for telemetric mea­surement of intraocular pressure. Ophthalmology. 2023;130:304–12.
29. Nakakura S.Icare® rebound tonometers: review of their characteristics and ease of use. Clin Ophthalmol. 2018;12:1245–53.
30. Liu J, De Francesco T, Schlenker M, Ahmed II.Icare home tonometer: a review of characteristics and clini­cal utility. Clin Ophthalmol. 2020;23(14):4031–45.
31. Fayed MA, Chen TC.Pediatric intraocular pressure measurements: tonometers, central corneal thickness, and anesthesia. Surv Ophthalmol. 2019;64:810–25.
32. Chui WS, Lam A, Chen D, Chiu R. The inu­ence of corneal properties on rebound tonometry. Ophthalmology. 2008;115:80–4.
33. Vinciguerra R, Ambrósio R Jr, Elsheikh A, et al. Detection of keratoconus with a new biomechanical index. J Refract Surg. 2016;32:803–10.
34. Vinciguerra R, Elsheikh A, Roberts CJ, etal. Inuence of pachymetry and intraocular pressure on dynamic corneal response parameters in healthy patients. J Refract Surg. 2016;32:550–61.
35. Ruberti JW, Sinha Roy A, Roberts CJ.Corneal bio­mechanics and biomaterials. Annu Rev Biomed Eng. 2011;13:269–95.
36. Tonnu PA, Ho T, Sharma K, et al. A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol. 2005;89:847–50.
37. Tonnu PA, Ho T, Newson T, et al. The inuence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-con­tact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol. 2005;89:851–4.
38. Gopesh T, Camp A, Unanian M, et al. Rapid and accurate pressure sensing device for direct measure-
17 Intraocular Pressure
211
ment of intraocular pressure. Transl Vis Sci Technol. 2020;9:28. https://doi.org/10.1167/tvst.9.3.28.
39. Albert DM, Keeler R.The pressure: before and after Schiøtz. Ophthalmol Glaucoma. 2020;3:409–13.
40. Senthil S, Chary R, Ali MH, et al. Schiotz scleral intraocular pressure readings predict Goldmann applanation readings better than rebound tonometry. Cornea. 2019;38:1117–23.
41. Eklund A, Hallberg P, Lindén C, Lindahl OA. An applanation resonator sensor for measuring intra­ocular pressure using combined continuous force
and area measurement. Invest Ophthalmol Vis Sci. 2003;44:3017–24.
42. Mulak M, Czak WA, Mimier M, Kaczmarek R.A com­parison of intraocular pressure values obtained using a Goldmann applanation tonometer and a handheld version of applanation resonance tonometer: a pre­liminary report. Adv Clin Exp Med. 2018;27:481–5.
43. Salvetat ML, Zeppieri M, Tosoni C, Brusini P. Repeatability and accuracy of applanation reso­nance tonometry in healthy subjects and patients with glaucoma. Acta Ophthalmol. 2014;92:e66–73.
Gonioscopy
ArnavPanigrahi , ShikhaGupta , andVineyGupta
18
18.1 Introduction
Gonioscopy is necessary when testing for glau­coma. It enables the visualization and magnica­tion of the iridocorneal angle and helps identify its conguration, thereby aiding a proper diagno­sis. The primary purpose of this test is to distin­guish between angle closure and open-angle glaucoma and help diagnose the various open and closed angle varieties of secondary glaucoma by identifying the characteristics specic to these conditions. Currently, intraoperative gonioscopy is increasingly being used in angle-based surger­ies. Although a simple technique, it takes years of practice to master the art of gonioscopy. Hence, practicing this technique as frequently as possi­ble is advisable.
18.2 The Optics ofGonioscopy andLenses Used
Gonioscopy entails the visualization of angle structures. The main problem with the direct visualization of angle structures is that it is dif­cult to ascertain the position of the angle with respect to the eye due to the sudden rarecation of the refractive index when the light rays pass
A. Panigrahi · S. Gupta · V. Gupta (*) Dr. Rajendra Prasad Centre for Ophthalmic Sciences, AIIMS, New Delhi, India
from the cornea (μ=1.33) into the air (μ=1.0). Since the critical angle at the cornea–air interface is about 46°, only the rays that are incident on the cornea at an angle of <46° can emerge from the anterior chamber into the air, allowing the direct visualization of the angle. This is usually not pos­sible in eyes with normal anatomy (Fig.18.1a) but might be possible in pathological conditions like distorted corneal contour, keratoconus, and keratoglobus, where the angle structures may be directly visible without any aid. In normal sce­narios, this problem can be circumvented in two ways. One entails the use of a thick concavo­convex lens placed on the cornea, with a power higher than the corneal dioptric power, to prevent the phenomenon of total internal reection; this allows most of the rays to pass out of the cornea following refraction (Fig. 18.1b) (the principle behind direct gonioscopy). The second entails using a plano-concave lens with a coupling agent to nullify the difference in refractive index between the two media, hence avoiding total internal reection such that the emergent light can then be brought onto the central frame using plane mirrors placed at a specic angle (Fig. 18.1c) (the principle behind indirect gonioscopy).
The commonly used gonioscopes for slit-lamp angle visualization utilize indirect optics. The prototype for such a lens is the Goldmann single mirror gonioscope. It consists of a plano-concave lens at the base, the corneal interface, and a single
© The Author(s), under exclusive license to Springer Nature Singapore Pte Ltd. 2024 T. Das, P. Satgunam (eds.), Ophthalmic Diagnostics, https://doi.org/10.1007/978-981-97-0138-4_18
213
214
A. Panigrahi et al.
Fig. 18.1 The schematic diagram shows the transmission of light rays originating from the anterior chamber angle, with and without the aid of a goniolens. In the natural state, all the light rays undergo total internal reection, and none emerges from the eye (left). A direct (middle) or
a
Fig. 18.2 Indirect gonioprisms commonly used for angle evaluation: (a) the Goldmann 2-mirror lens; (b) Goldmann 4-mirror lens; (c) Latina selective laser trabeculoplasty lens; and (d) Posner gonioprism with a handle
b
plane mirror placed within the body of the gonio­scope at 62°, which reects the emergent light rays out of the cornea almost perpendicular to the lens. The base is larger than the corneal diameter (12 mm for the Goldmann-type lens). These lenses require a coupling agent to ensure homog­enous contact between the lens and the cornea, thus accounting for the sudden change in curva­ture at the corneoscleral junction. These lenses are considered indirect as the visualization of the structures occurs after reection by a plane mir­ror. The placement of the mirror is such that the image obtained is reversed but not crossed. For example, the inferior angle can be visualized by ensuring superior placement of the mirror, but the image on the right side of the mirror is still the right side of the concerned angle. Other lenses with similar builds are the Goldmann 2-mirror (Fig.18.2a), 3-mirror, 4-mirror lenses (Fig.18.2b) and the Allen Thorpe goniolens. The Goldmann 3-mirror lens has a mirror placed at 59° for the
indirect (right) goniolens allows the light rays to exit the eye, allowing the observer to visualize the angle struc­tures. The green lines indicate the placement of the mirrors
c
d
gonioscopic visualization of the opposite angle. The other two mirrors, placed at 67° and 73°, are for visualizing the equatorial and peripheral ret­ina, respectively. The Allen Thorpe lens uses 4 prisms instead of plane mirrors. The central lens in each of these variants can be used to evaluate the posterior pole. This allows for complete glau­coma evaluation, simultaneously assessing the disc and the angle. It can also be used to stabilize the globe, hence ensuring a more accurate exami­nation, especially in eyes with nystagmus. The Latina selective laser trabeculoplasty (SLT) gonio laser lens (Fig.18.2c) is a modication of a single mirror Goldmann-type goniolens spe­cially designed to deliver laser light during laser trabeculoplasty. It consists of a single plane mir­ror placed at an angulation of 63° and utilizes indirect optics to visualize angle structures. It has a 1× magnication; hence the size of the beam emitted by the laser machine and that visualized by the observer is the same. It may also
18 Gonioscopy
215
incorporate a special lid ange that prevents the lens from slipping due to the resistance caused by squeezing the patient’s eyelids. A special modi­cation includes the ve high-contrast bars imaged over the trabecular meshwork (TM). The bars are spaced at 400μm, providing a guide for deliver­ing 10 laser spots, ve corresponding to and ve between the aforementioned bars. The bar can be rotated 10 times to complete a full circle around the angle, allowing for the accurately spaced delivery of 100 laser spots to the TM.
A modication to conventional indirect goni­oscopy involves the use of goniolens with a base smaller than the cornea. This allows these lenses to use the precorneal tear lm as a coupling agent by excluding the steep curvature change at the corneoscleral junction from its contact surface. This obviates the need for external coupling agents like methylcellulose. The smaller diame­ter of these lenses also allows for diagnostic and therapeutic maneuvers during gonioscopy. Pressing the lens on the central cornea displaces the aqueous humor from the central anterior chamber (AC) toward the periphery, pushing the iris root posteriorly and opening up the angle structures. This maneuver, called “dynamic goni­oscopy” or “indentation gonioscopy,” helps dif­ferentiate between appositional and synechial closure. This maneuver can also be therapeutic as it helps decongest the AC by partially opening up the angle during an acute angle closure (AAC) attack. The prototype for such a lens is the Zeiss 4 mirror goniolens. It uses a plano-concave lens of 9 mm diameter as its base and 4 mirrors inclined at 64° as the primary reecting surface.
All 4 quadrants are visible simultaneously, with a rotation of merely 11° allowing for the areas between the adjacent mirrors to be viewed. The Posner (Fig.18.2d) and Sussman lens are other prototypes used for indentation gonioscopy. A comparison between the various types of indirect gonioprisms is given in Table18.1.
Direct gonioscopy utilizes a thick convex lens placed on the cornea to nullify the rarecation of the refractive index when the light emerges from the cornea into the air. This prevents total internal reection and allows the direct visualization of the angle. These may come with or without han­dles. Since a single lens is used, and there is no provision of plane mirrors to make the emergent light perpendicular to the cornea, the patient’s head must be tilted away from the observer to allow the emergent rays to reach the observer’s cornea. These lenses can be used for diagnostic purposes but are more commonly used as surgi­cal goniolens for the intraoperative visualization of angle structures during procedures like goni­otomy and gonioscopic-assisted transluminal tra­beculotomy (GATT). Koeppe lens (Fig. 18.3a) and Swan Jacob lens (Fig.18.3b) are prototypes for direct goniolenses. Other examples are the Barkan lens, Thorpe lens, and Mori lens. The Mori upright surgical goniolens incorporates a two-mirror design that redirects the oblique emergent rays to the coaxial position perpendicu­lar to the cornea, hence obviating the need for tilting the patient’s head during surgery. This allows for 360° visualization of angle structures while the head is supine, without requiring intra­operative maneuvers like head tilting.
Table 18.1 Comparison of indirect gonio prisms
Lens type Diameter of corneal contact 12mm 12mm 9mm Overall diameter 15mm 18m 9mm Rim size 1.5mm 3mm None Mirror angulation 62° 59° 64° Mirror height 17mm 12mm 12mm Distance from central cornea 3mm 7mm 5mm Radius of curvature 7.4mm 7.4mm 7.85mm Coupling uid Required Required Not required Dynamic gonioscopy Manipulation Manipulation Indentation
Goldmann Single mirror
Goldmann Three mirrors
Zeiss Four mirrors
216
a
Fig. 18.3 The direct goniolenses commonly used in routine glaucoma practice: (a) Koeppe lenses of varying sizes and (b) a Swan Jacob lens with a handle
A. Panigrahi et al.
b
18.3 Techniques Used inGonioscopy
If indentation gonioscopy is performed, gently pressing the goniolens against the cornea opens up the angle structures, allowing one to differen-
18.3.1 Indirect Lenses
tiate between appositional and synechial closure. Contrary to traditional belief, similar maneuvers
There is no dened standardized technique for performing indirect gonioscopy. Here is a simple method for performing gonioscopy using an indi­rect handheld gonioprism without a handle. The readers are advised to experiment with different gazes and gonioprisms to nd the technique most suitable for them.
Explain the procedure thoroughly to the patient and take verbal consent. Anesthetize the eye with 0.5% proparacaine eye drops. Ask the patient to sit upright at the slit lamp, with the forehead pressed rmly against the headrest. If a Goldmann-type lens is used, put 2–3 drops of coupling agent (generally methylcellulose) on the corneal surface of the lens. Indentation gonio-
are also possible with the Goldmann-type gonio­lenses, although the process is far more compli­cated. To open up the angle visible in the superior mirror, which is the inferior angle, the patient is asked to look superiorly, and gentle pressure is applied on the superior rim of the contact surface of the goniolens. This maneuver displaces the uid in the superior angle inferiorly, thereby opening the inferior angle if appositional closure is present. This “opening up” is visualized in the superior mirror. Hence, to open the angle seen in one of the mirrors of the Goldmann goniolens, the patient is asked to look toward the direction of the mirror, and the pressure is applied on the sclera adjacent to the mirror.
lenses do not require coupling agents for per­forming gonioscopy. Ask the patient to look down and gently elevate the upper eyelid. Dock
18.3.2 Direct Lenses
the corneal surface of the gonioscope onto the inferior fornix and rmly appose it against the cornea to prevent the formation of air bubbles in the interface, which impedes clear visualization of angle structures. Rotation of the lens and inter­mittent pressure on the goniolens against the cor­nea also helps remove air bubbles if present. Ask the patient to look straight ahead and direct an oblique beam of 3–4mm in length at an angle of 30°–45° onto the mirror of interest. Move the slit lamp back and forth until the angle structures are visible, and rotate the lens to visualize the whole angle in the primary position.
Direct goniolenses are primarily used for intraop­erative gonioscopy. The patient’s head is tilted away from the observer, and the globe is pushed toward the nasal bridge (to visualize the nasal angle), with the observer seated in a temporal position. A surgical visco-dispersive ophthalmic viscoelastic device (OVD), like HPMC 2%, is placed on the contact surface of the surgical goni­olens. The lens is held with a handle and gently placed on the nasal limbus without any undue pressure not to induce any corneal folds. This allows clear visualization of the nasal angle, the
18 Gonioscopy
217
site of most angle surgeries, without apparent image distortion.
Direct goniolenses, like the Koeppe lens, may also come without a handle. They are held at the equator between the thumb and index nger of the right hand for right eye examination, the left hand for the left eye, and the head tilted toward the angle being visualized. The space between the lens and the eye is lled with a coupling agent. The light source is provided either by the operating microscope or a handheld slit lamp.
18.4 Structures Seen
onGonioscopy
Figure 18.4 shows an open angle in a normal eye, where all the angle structures are visible. From anterior to posterior, the angle structures seen are Schwalbe’s line (SL), anterior trabecular mesh­work (ATM), posterior trabecular meshwork (PTM), scleral spur (SS), and ciliary body band (CBB). The condensation of collagen bers forms the SL and marks the termination of Descemet’s membrane (DM). Identication of the SL is important because the labeling of the angle structures begins from the SL and contin­ues posteriorly toward the CBB.Hence, identify­ing the SL is crucial, especially in cases with featureless angles, where none of the angle struc­tures are seen. In these cases, the SL usually pro-
Fig. 18.4 Goniophotograph of a normal wide open angle. Structures seen, from anterior to posterior, are Schwalbe’s line (SL), anterior trabecular meshwork (ATM), posterior trabecular meshwork (PTM), scleral spur (SS), and ciliary body band (CBB)
vides a landmark from which the point labeling of the angle structures can begin. The identica­tion of the SL, although difcult in these cases, can be aided by the “corneal wedge” technique as follows: direct a narrow beam of light from the slit lamp obliquely on the cornea; the beam is split into an outer part (corresponding to the cor­neal epithelium) and an inner part (corresponding to the corneal endothelium); a fusion of both these parts marks the termination of the DM, which is the landmark for the SL.
The trabecular meshwork (TM) can be divided into two parts: the anteriorly placed non­functional anterior TM (ATM) and the posteri­orly placed functional TM (PTM). ATM forms the structural scaffold and has almost no outow function owing to the lack of Schlemm’s canal (SC) behind it. It is lightly pigmented or non­pigmented in appearance. Its posterior extent marks the site of an incision during ab interno angle surgeries like goniotomy and GATT.The PTM, on the other hand, forms the site of maxi­mum resistance to aqueous outow. Due to the SC behind it, it marks the site of conventional outow, also known as pressure-dependent out­ow or trabecular outow. It also indicates the incision site for inserting minimally invasive glaucoma drainage devices, like iStent and Hydrus implants. It is relatively easier to identify due to its pigmented nature. This might not be the case in pediatric eyes, where pigmentation is sparse, and a featureless angle may also be a pos­sibility. The PTM in the superior angle usually exhibits less pigmentation than in the inferior angle. Any change in the intensity of pigmenta­tion, or its distribution, may provide a clue to underlying pathologies. Figure 18.5a shows a densely pigmented TM in a patient with pigmen­tary glaucoma.
Another important landmark to identify in the angle is the SS.It is a ridge formed by the con­densation of the collagen tissue of the sclera, and it is the site of attachment of longitudinal bers of the ciliary muscle. Identifying the SS is rela­tively easy; the white band is immediately poste­rior to the PTM. Identifying the SS is very important, as it allows us to differentiate between open and closed angles. Due to the attachment of
218
ab
Fig. 18.5 Goniophotograph of a patient with (a) pigment dispersion syndrome, showing a densely pigmented trabecu- lar meshwork of the superior angle; (b) angle recession, showing an irregularly widened ciliary body band
A. Panigrahi et al.
the ciliary muscle, contraction due to light­induced miosis pulls the SS posteriorly, which may artefactually open the angle. Hence, a pre­liminary examination of the angle structures is advised under low illumination to avoid the abovementioned phenomenon and falsely label the angle as open. Occasionally, the scleral spur might be covered by ne projections of uveal tis­sue known as the iris processes. These do not cross the SS, are found in about one-third of healthy individuals, and are not associated with an increased risk of glaucoma.
The CBB forms the posterior-most part of the angle and lies adjacent to the iris root. It is wide in eyes with high myopia, trauma, posteriorly subluxated lens, pseudophakia, or aphakia. It is narrow in hyperopic eyes and those with an ante­rior insertion of the iris. A unilateral abnormally widened CBB, especially with the history and other signs and symptoms of ocular trauma, with or without an associated cyclodialysis cleft (Fig. 18.5b), is usually pathognomonic of post­traumatic angle recession.
cal and commonly used. Scheie [1] used a system based on the visibility of angle structures, allot­ted with a higher number of narrower angles (range: I–IV). Emphasizing the complexity of angle structures, Spaeth [2, 3] proposed a system incorporating multiple parameters limited to the angle. These include iris insertion (range A–E), iris conguration, iris angulation (range 10°–40°) (Fig. 18.6), and pigmentation of PTM (range 0–4). Various combinations of these parameters give a more accurate representation of the angle [4]. Becker limited his classication to 2 impor­tant angle parameters: the width of the trabecular zone (the space between the SL and SS, repre­sented by numbers) and the distance between the SS and iris root (represented by letters). Shaffer­Kanski has devised the most clinically relevant and practically useful system. It is a modication of the system originally developed by Shaffer [5] (Table18.2). It incorporates two parameters: iris angulation and visibility of structures. Combining both parameters can help the clinician estimate the risk of angle closure. The readers are directed to consult additional sources for a detailed dis­cussion of these nomenclature systems.
18.5 Identifying Occludability
It is important to document the ndings clearly
and concisely, which, even if elaborate, must Numerous systems have been devised to grade the occludability of the angle based on the num­ber of structures seen on the gonioscopy. The nomenclature systems currently in clinical use have been developed by Scheie, Spaeth, Becker, and Shaffer-Kanski; the latter is the most practi-
impart as much information as possible and be
easily understandable by fellow ophthalmolo-
gists. This helps avoid repeat examinations and
hence saves time in clinics. A simple example of
a diagrammatic representation of gonioscopy
ndings is given in Fig.18.7. Divide the eye into
TM + SS
SS/CB
18 Gonioscopy
219
Fig. 18.6 Varying degrees of angle recess and occlud­ability as classied according to the Spaeth system: the left and middle panels show occludable angle recesses of
Table 18.2 Classication and clinical application of anterior chamber angle
Grade Angle width Description Risk of closure 4 45°–35° Wide-open Impossible 3 35°–20° Wide-open Impossible 2 20° Narrow Possible 1 <10° Extremely
narrow Slit Slit Narrow to slit Probable 0 Closed Closed
Probable
SS
10° and 20°, respectively. The right panel shows a wide open angle with a 30° angle recess
four quadrants using two oblique lines intersect­ing at an angle of 90°. The four quadrants corre­spond to the part of the eye being examined, as seen by the examiner. The deepest part of the angle seen under the baseline conditions in each quadrant is marked. If an indentation or “manipu­lation” gonioscopy is performed, the quadrants are designated with doubled-ended arrows, and the part seen after the maneuver mentioned above is also marked. Additional information regarding the iris insertion, conguration, and angle pig­mentation can be added per the Spaeth nomen­clature. Special details, like PAS, cyclodialysis, and new vessels, can be drawn at their location to give a complete diagrammatic representation of the angle.
18.6 Look Beyond What YouSee
PAS
Fig. 18.7 Schematic representation of angle ndings in the right eye of a patient. The double-headed arrows rep­resent a dynamic gonioscopy being performed. The fur­thest structure seen on the baseline condition and maximal illumination and manipulation are marked for each corre­sponding quadrant. Any special ndings, like inferotem­poral peripheral anterior synechiae (PAS) at 8 clock hours in this gure, are also drawn
SS/CB
Although gonioscopy seems simple, mastering it requires optimal vigilance and years of practice. Gonioscopic ndings, on their own, may be of limited importance; but in conjecture with the relevant history and associated anterior segment ndings, may provide a lot of information. Let’s consider this example:
Case report: A 42-year-old male with prolifera-
tive diabetic retinopathy in both eyes was referred to the glaucoma clinic because of high IOP in both eyes. Anterior segment eval­uation showed the presence of neovasculariza-