Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3597_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
87 Мб
Скачать
192 4
https://t.me/med1917
ECG
40 mm Hg
LA
LV
Hemodynamic Data
0
CF
1 cm = 3.9 mm Hg
9.46 cm
3.4 cm
9.46 • CF
3.4
2
= 10.85 mm HgMVG
Area DFP
Figure 4-7 Hemodynamic tracing used to calculate mitral valve area (MVA).
Shaded area, Diastolic mean valvular gradient (MVG) surrounded by the diastolic filling period (DFP). CF, Correction factor or scale factor; DFP, dia­stolic filling period; ECG, electrocardiogram; LA, left atrial pressure; LV, left ventricular pressure (scale 0 to 40 mm Hg); MVG, mean valvular gradient.
DFP
Step 2. Measure diastolic filling period (DFP).
DFP cm then convert to time
=
3 4
. ( )
cm sec/ cm second
× =
3 4 1 10 0 34
. .
Step 3. Convert planimetered area to mean diastolic pressure
gradient.
mm Hg/ cm
3 9 1
.
MVG cm
2
.
3 4
cm
.
10 85
.
mm Hg= × =9 46
Step 4. Compute mitral valve flow. For the mitral gradient (similar
to AVAs), DFP is in centimeters at this point, owing to scale factor.
Flow
CO
=
DFP HR
3500
=
77 2
2
=
×
0 34 80
128 7..= mL/min
mL/min
3500
s/beat bpm
. .
×
=
3500
×
0 34 80
Step 5. Compute MVA.
Mitral value flow
MVA
=
×
0 44 44 3 10 85
. . .
128 7
=
0 85 44 3 3 3
.
× ×
. . .
=
128 124 3
..
7
= cm
.
1 0
2
.
Notes on Mitral Valve Gradient
Obtaining an accurate PCWP is crucial (see earlier discussion about PCW accuracy; see Fig. 4-3, A). Use of direct LA pressure from trans­septal measurement is the most accurate method. Trans-septal cath­eterization will confirm pressure gradients, especially for suspected prosthetic mitral stenosis. However, if the PCWP-LV pressure trac­ings show no significant gradients, trans-septal catheterization is unnecessary.
Simplified Mitral Valve Gradient Calculation by Cui et al
Cui et al simplified estimation of the mitral valve gradient and thus simplified the calculation of MVA from hemodynamic tracings. Because the mean mitral valve gradient is the pressure difference between the mean left atrium pressure (MLAP) and mean left ven­tricular pressure (MLVP) during diastole (i.e., MLVG = MLAP MLVP), the computation of the mean mitral valve gradient depends on simply knowing the MLVP. The MLAP is easily obtained from the
=
https://t.me/med1917
4
Hemodynamic Data 193
LV
LA
LVEDP
Baseline
Figure 4-8 Cui method to calculate mitral valve gradient from hemody-
namic tracings. Left ventricular (LV) and left atrial (LA) pressures are super­imposed. The LV-LA gradient is triangular in shape, representing half of the rectangle delineated by A and B, diastolic filling period (DFP); C, left ventricu­lar end-diastolic pressure (LVEDP); D, mitral value closing; and E, mitral value opening. (From Cui W, Dai W, Zhang G: A new simplified method for calculat­ing mean mitral pressure gradient. Catheter Cardiovasc Interv 70[5]:754– 757, 2007.)
LV
E
LA
D
C
A
LVEDP
B
DFP
electronically meaned LA signal on the hemodynamic recorder. The area under the LV pressure during diastole is roughly a triangle with the three corners formed from the intersections of the DFP starting and ending points (mitral valve closure and opening) marking the vertical lines intersecting with the LV pressure line (rising diagonally across diastole; Fig. 4-8). This triangular area can be estimated from the rectangular area LVEDP × DFP divided by 2. Thus, the MLVP is equal to the LVEDP/2. From this key calculation, the mitral valve gradi­ent is therefore simplified as
MVG MLAP LVEDP/
2
There is strong correspondence between MVAs calculated by the Gorlin and Hakki formulae, both before and after mitral balloon val­vuloplasty, with an error of estimates at 1.6 mm Hg. The Cui mitral valve gradient slightly overestimated the gradient before but not after mitral valvuloplasty. The Hakki formula significantly underestimated mitral valve gradients after mitral balloon valvuloplasty. Unlike Hakki, the Cui mitral valve gradient was not affected by mitral regurgitation, Ao insufficiency, AF, or HR.
Although simple, the Cui mitral valve gradient still has potential problems. HR changes will affect the shape of the triangular area under the LV pressure curve, and thus, tachycardia may cause a poten­tial overestimation of valve severity.
Tricuspid Valve Gradients
Because small gradients (5 mm Hg) across the tricuspid valve may lead to significant clinical symptoms, precise measurement of hemo­dynamics through two large-lumen catheters may be required. Match pressures through two catheters (or through the two lumens of a balloon-tipped catheter if correctly positioned) before placement in the RA and RV to avoid technical error. The Gorlin valve area formula has not been validated for the tricuspid or pulmonic valves.
Measurement of Cardiac Output
In the cardiac catheterization laboratory, CO is determined by one of two techniques: (1) Fick, with measurement of oxygen consumption (see earlier) or (2) indicator dilution (TD, using a PAC; see Chapter 1).
194 4
https://t.me/med1917
Table 4 -1
Hemodynamic Data
Cardiac Shunt Locations
Location
Atrial Septal Defects
Primum (low) RA, RV Secundum (mid) RA Sinus venosus (high) R A Partial anomalous pulmonary venous return
(pulmonary veins entering RA)
Ventricular Septal Defects
Membranous (high) RV Muscular (mid) RV Apical (low) RV Aorticopulmonary window (connection of aorta
to PA)
Patent ductus ar teriosus (normally closed
Ao- PA connection at birth)
Ao, Aortic; PA, pulmonary arter y; RA, right atrium; RV, right ventricle.
Earliest Step -Up Location (for Left-to -Right Shunts)
RA
PA
PA
Intracardiac Shunts
A shunt is an abnormal communication between the left- and right­heart chambers. The direction of blood flowing through the shunt is left to right, right to left, or sometimes bidirectional. In the absence of shunting, the pulmonary blood flow (right side of the heart) is equal to the systemic blood flow. Table 4-1 lists intracardiac shunt locations. A left-to-right shunt increases the amount of blood to the right side of the heart and increases pulmonary blood flow, now the sum of the systemic blood flow plus shunt flow. With a right-to-left shunt, the amount of blood shunted from the right side of the heart to the left is added to that normally ejected into the systemic circulation, making systemic blood flow greater than pulmonary blood flow by the amount of blood flow in the shunt (Fig. 4-9). Intracardiac shunts have been evaluated by oximetry, radionuclide perfusion, and Doppler echo flow measurements. Oximetry is the most common method used in the catheterization laboratory.
Oximetry Procedure: Diagnostic Saturation Run
A diagnostic “saturation run” uses 1- to 3-mL heparinized syringes to obtain blood from the superior vena cava (SVC), inferior vena cava (IVC), RA, RV, and PA in a rapid, organized manner. A standard balloon-tipped PA flotation catheter is satisfactory, but a large-bore end-hole or side-hole (multipurpose) catheter performs better rapid sampling. Heparinize saturation syringes with < prepare the labels and list of sample sites in advance (Box 4-10).
The saturation run begins after diagnostic hemodynamic data and CO have been obtained and before right-sided heart pull back. With the catheter positioned in the right or left PA, measure oxygen consumption (Fick method). On catheter pull back, one operator manipulates the catheter under fluoroscopic and pressure control while an assistant aspirates the blood samples at each location along the run. Each new sample is obtained after several milliliters of blood have been withdrawn and discarded so that blood left from the previ­ous catheter sampling location will not contaminate the new blood oxygen. The entire diagnostic run should take approximately 5 to 7 minutes.
0.5 mL. In addition,
Body
Lungs
L R
https://t.me/med1917
4
Hemodynamic Data 195
EPBF = PBF +
shunt flow
SVC
L R = PBF – EPBF
IVC
R
CS
L = SBF – EPBF
PV
Right heart
O2 cons.
PVO
– MVO
2
R L
EPBF EPBF
2
Shunt
R L
Shunt
L – R
Shunt
R L
R L
=
EPBF
Left heart
PA
Ao
R L
ESBF = SBF +
shunt flow
2
O
cons.
2
SAO2 – MVO
2
Figure 4-9
Lungs Body
PBF = SBF =
O2 cons.
PVO2 – PAO
Schematic diagram for right-to -lef t and left-to-right shunting across the hear t. Ao, Aorta; CS, coronary sinus; EPBF, ef fective pulmonar y blood flow; ESBF, ef fective systemic blood flow; IVC, inferior vena cava; L, left; MVO2, mixed venous oxygen saturation or content; O2 cons., oxygen consumption; PA, pulmonary ar tery; PAO2, pulmonary arterial oxygen satura­tion or content; PBF, pulmonary blood flow; PVO2, pulmonary venous oxygen saturation or content; PV, pulmonary vein; R, right; SAO2, systemic ar terial oxygen saturation or content; SBF, systemic blood flow; SVC, superior vena cava.
Box 4 -10 Sample Sites for Oxygen Saturations During
Diagnostic Saturation Run
Right Side of the Heart
Left PA Right PA Main PA PA above pulmonary valve (PApV) RV below pulmonar y valve (RVpV) RV (mid) RV (apex) RV at tricuspid valve (RVTV) RA at tricuspid valve (RATV) RA (mid) SVC (high) SVC (low) RA (high) RA (low) IVC (high, just beneath heart, above hepatic vein) IVC (low, above renal vein, but below hepatic vein)
Left Side of the Heart
Arterial saturation, Ao If possible, cross ASD, pulmonar y vein saturation Patent foramen ovale or LA
Ao, Aortic; ASD, atrial septal def ect; IVC, inferior vena cava; LA, lef t atr ium; PA, p ulmonar y arter y; RA, right atrium; RV, right ventricle; SVC, superior vena cava.
196 4
= +
=
https://t.me/med1917
Table 4 -2
Hemodynamic Data
Oxygen Saturation Values for Shunt Detection
Level of Shunt
Atrial (SVC/IVC to right aor ta) Ventricular Great vessel
IVC, Inferior vena cava; SVC, superior vena cava.
= (3 SVC + 1 IVC)/4 and difference from PA should be 7% normally.
MVO
2
*Difference betwe en distal and proximal chamber; for example, for atrial septal defect (ASD).
Significant Step- Up Difference* O2% Saturation
755
Oxygen Step-Up: Evidence of a Shunt
A left-to-right shunt is suggested when an oxygen step-up or increase of oxygen content in that chamber or vessel exceeds that of a proximal compartment. A step-up in oxygen saturation at the PA by more than 7% above the RA saturation is indicative of a left-to-right shunt at the atrial level (Table 4-2). Similarly, the desaturation of arterialized blood samples from the left heart chambers and aorta suggests a right-to-left shunt. In determining the site of the right-to-left shunt, sequential sam­pling can be made from the LA, LV, and aorta.
Mixed venous blood is assumed to be fully mixed PA blood. With a left-to-right shunt, measure mixed venous blood one chamber proxi­mal to the step-up. In the case of an atrial septal defect (ASD), compute the mixed venous oxygen content from the weighted average of vena caval blood (i.e., as [3 × SVC + IVC]/4). When pulmonary venous blood
) per-
centage saturation is 95%.
2
Shunt Calculation
The Fick or left-sided indicator dilution methods of CO determination are used to measure systemic flow (see Fig. 4-9). Using the Fick method, the following formulae apply:
1. Systemic flow
O consumption mL/min
Q L/min
( )
s
=
2
arterial mixed venous O co
( )
2. Pulmonary flow, QP (L/min)
O consumption mL/min
pulmonary venous pulmonary arterial
( )
2
Thus, the effective pulmonary blood flow (EPBF) is
O consumption mL/min
Q
=
EPB
pulmonary venous mixed venous O
( )
2
Normally, the EPBF is equal to the systemic blood flow. In a left­to-right shunt, EPBF is increased (by the amount of the shunt) as follows:
EPBF systemic flow shunt flow left-to-right
In a right-to-left shunt, EPBF is decreased (by the amount of the shunt):
EPBF systemic flow shunt flow right-to-left
Shunt volume is determined by use of equations 1 and 2 (later).
The shunt fraction is the ratio of pulmonary to systemic flow (called Q
, where Q is flow, P is pulmonary, and S is systemic)
p/Qs
for a left-to-right shunt. Flow ratios (Q
( )
nntent
2
( )
OO content
2
( )
content
22
( )
( )
) >1.5 often require closure.
p/Qs
https://t.me/med1917
4
Hemodynamic Data 197
Example for Left-to-Right Shunt Atrial Septal Defect Calculation
Data obtained at catheterization are:
Location Saturation (%) Location Saturation (%)
Arterial 92 PA 81 SVC 71 PV 98 Mid R A 85 IVC 70 Low R A 68
Hemoglobin g
O consumption mL/min
2
==13
210
1. Compute O2 content.
Arterial O content mL O /g g/dL
2 2
0 98 1 36 14 1 10
= × × ×
. . .
mL O /L
188
=
2
Mixed venous O2 content (use estimate of mixed venous oxygen saturation):
3 1
SVC IVC
+
71 1 36 14 1 10 136
Pulmonary artery O content mL O /g g/dL
Pulmonary vein O content mL O /g g/dL
× × × =. .mL O /g g/dL mL O /L
for mixed venous oxygen saturation:
4
(. . . . )
71 71 71 70
+ + +
4
2 2
0 81 1 36 14 1 10= × × ×
2 2
2 2
( . . . )
=
11552mL O /L
0 98 1 36 14 1 10
= × × ×
( . . . )
18
=
882mL O /L
=
.
0 71
2. Compute systemic flow (equation 1).
225
mL O /min
( )
188 136
2
mL O /L
225
= = =
52
2
.
4 3
Q L/min
S
3. Compute pulmonary flow (equation 2).
225
mL O /min
( )
188 155
2
mL O /L
225
= = =
33
2
.
6 8
Q
P
4. Compute.
6 8
.
Q
P
= =
1 6
Q
S
4 3
.
.
The LR shunt is 6.8 L/min 4.3 L/min or 2.5 L/min.
If absolute flows are not required, the QP/QS ratio can be deter-
mined using saturations only as follows:
QQSAO MVO
P
S
2
=
PVO PAO
2 2
2
where SAO2 is systemic arterial oxygen saturation, PVO2 is pulmonary venous oxygen saturation, MVO and PAO
is pulmonary arterial oxygen saturation.
2
is mixed venous oxygen saturation,
2
Using saturation data from the example of left-to-right shunt:
98 71
Q
P
=
Q
S
= =
98 812717
1 6.
198 4
https://t.me/med1917
Hemodynamic Data
Example for Right-to-Left Shunt
Data obtained at catheterization are:
Location Saturation (%) Location Saturation (%)
Arterial 89 L A 88 SVC 81 PA 82 Mid R A 83 PV 96 Low R A 82 IVC 70
Hemoglobin g
O consumption mL/min
2
==15
195
1. Compute O2 content.
Arterial = (0.89 ×
15 g/100 mL × 1.36 mL O2/g × 10) = 182 mL O2/L
Mixed venous (estimated mixed venous) saturation = (.81 + .81 + .81 + .70) 4 = 0.78
Mixed venous = (0.78 × Pulmonary arterial = (0.82 ×
=
167 mL O2/L
Pulmonary venous = (0.96 ×
=
196 mL O2/L
15 g/dL × 1.36 mL O2/g × 10) = 159 mL O2/L
15 g/dL × 1.36 mL O2/g × 10)
15 g/dL × 1.36 mL O2/g × 10)
2. Compute systemic flow (equation 1).
O consumption
arterial mixed venous O content
( )
2
2
=
ml O /min
195
2
(( )182 1592− ml O /L
3. Compute pulmonary blood flow (equation 2).
O consumption
2
( ) tt
pulmonary venous pulmonary arterial O conten
4. Compute Q
=
( )
=
( )
mL O /min
195
( )
196 167
.
EPB
pulmonary venous mixed venous O content
195
mml O /min
2
196 159
mL O /L
2
mL O /L
O consumption
2
5 3
.−=
2
.
=
6 7
2
L
2
L/min=
2
Shunt calculations:
Left-right shunt Q Q L/min
Right-to-left shu
= = =6 7 5 3 1 4. . .
P EPB
nnt Q Q L/min
= = =8 5 5 3 3 2. . .
S EPB
Limitations of the Oximetric Technique
1. Because of its low sensitivity, oximetry may fail to detect small (<1)
shunts.
2. The application of the Fick principle to calculate blood flow pre-
sumes a steady state during the diagnostic run and measurement of oxygen consumption (i.e., timely collection of saturation sample within a period during which oxygen consumption and CO are stable).
3. The oximetry method also assumes that complete mixing is
achieved instantly and that blood samples obtained are representa­tive of blood in the respective compartment.
4. The rate of systemic blood flow is important in detecting a
shunt by oximetry. A high systemic flow tends to equalize the AVO difference across a given vascular bed. In the presence of elevated
2
https://t.me/med1917
systemic blood flow, the mixed venous oxygen saturation is higher than normal and intrachamber variability caused by streaming is reduced. In contrast, when systemic blood flow and mixed venous oxygen saturation are low, a larger step-up must be detected before a significant left-to-right shunt is diagnosed.
4
Hemodynamic Data 199
Angiography
Angiography is a qualitative method used to localize either left-to-right or right-to-left shunts. The shunt can be detected by injection of x-ray contrast medium into the closest proximal chamber. The left anterior oblique view with cranial angulation puts the interatrial and interven­tricular septae on face (i.e., on edge), which provides an ideal view for detection of contrast medium passage across the atrial and ven­tricular septal defects (see Chapter 3 for angiographic views to visual­ize an intracardiac shunt).
Equipment Used for Hemodynamic Study
Pressure Manifold and Setup
The optimal set of transducers, tubing, and manifolds for any labora­tory is that which is cost effective, familiar, accurate, and simple to use. Several varieties of disposable manifolds exist (Fig. 4-10) that can be coupled to transducers positioned either on the manifold or at the side of the catheterization table. For research studies, special transducer-tipped micromanometer pressure catheters and pressure sensor guidewires are used to obtain high-fidelity pressure recordings. High-fidelity recordings are not necessary for routine clinical hemo­dynamic studies; accurate measurements can be obtained with fluid­filled systems if appropriate precautions in the setup are taken.
Some clear plastic manifolds have several ports: (1) pressure and zero line, (2) saline flush, (3) contrast media at the third, and (4) closed waste line (to minimize contamination of personnel and laboratory). A three-port manifold combines a saline flush and waste port with a one-way valve.
Figure 4-10 Pressure manifold used for coronary angiography. At the end
of the manifold, the connection of the manifold to the catheter has a swivel connector. 1, Stopcock to pressure; 2, saline flush line; 3, contrast line; 4, waste line with a one-way valve.
4
1 2 3
200 4
https://t.me/med1917
200 mm Hg
100 cm/sec
Figure 4-11 Hemodynamic tracing showing the effect of an air bubble in
a pressure line. Note the change in hemodynamic waveform after the bubble is flushed: The ringing artifact (sharp points at arrows) is eliminated and fine detail of precise pressure restored. ECG, Electrocardiogram; LV, lef t ventricle.
Hemodynamic Data
ECG
LV
0
Bubble
in line
After flushing
For best pressure waveforms, tubing should be short and stiff, with the transducer as close to the catheter as possible. Tubing length from the catheter to the transducer should also be minimized; longer tubing contributes to more resonating or “ringing” artifacts (Fig. 4-11).
The zero level is set at mid chest (i.e., the measured anteroposte­rior diameter of the patient divided by 2 and added to the position of the table). When the transducer is raised above zero level, pressure is artificially lower. When the transducer is lower than zero, pressure is artificially higher (Fig. 4-12). When abnormally low pressures are seen initially, recheck the zero for proper positioning (at mid chest) and check for air bubbles or loose connections.
Physiologic Recorder
The physiologic recorder system is now a digital system that processes the hemodynamic and electrocardiographic signals from the trans­ducers and pacing electrode catheters used in electrophysiologic studies. The typical physiologic monitor/recorder is a multichannel unit that can process, display, and record ECG signals, pressure trac­ings, and direct current (DC) inputs from external sources (e.g., TD). The number of channels determines how many individual signals are displayed and recorded simultaneously. For routine cardiac catheter­ization, one ECG signal and two to three pressure channels are nor­mally recorded. In certain complex cases, such as electrophysiology studies and cases with complex congenital or valvular heart disease, it is common to use four to 18 channels.
Most recorders have an electronic calibration that allows the operator to input an electronic pressure standard. This feature pro­vides a convenient means for simulating pressure signals to calibrate the display. An external standard pressure reference can be input from a mercury manometer and the transducer calibrated to match the input pressure.
Display Settings for Hemodynamic Data
The laboratory has many choices when setting the display for hemo­dynamic waveforms. These settings include timing lines, signal display sweep speed, and pressure scale(s). The typical setting for a routine case is 1-second time lines (i.e., one line each second). Use a faster time line sequence for special studies or when a faster paper speed is used. For most procedures, the sweep speed of the monitor is
https://t.me/med1917
4
Hemodynamic Data 201
A
B
Figure 4 -12 A, To demonstrate the effect of zero position, a transducer
is connected to a fluid-filled tube and its stopcock is open to air. By lifting either the tubing (as shown in the left hand) or the transducer, one can see the zero move. B, The effect of first raising the tubing shows how zero offset is affected by height on the table. The blue line goes up and then raising the transduce to the same level produces the same effect. Set the zero at bedside and keep the transducer fixed throughout the study.
25 mm/sec. Slower sweep speeds (10 and 5 mm/sec) are used to examine changes in hemodynamics, such as respiratory variations in RV/LV pressure, over longer periods of time. Faster speeds can help in observing timing of diastolic and systolic pressures (e.g., LV and Ao; LA and LV).