Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3597_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
87 Мб
Скачать
292 6
https://t.me/med1917
Figure 6-17 Example of an implantable cardiac monitor. (EnSite, Velocity,
Quartet, SJM Confirm and St. Jude Medical are trademarks of St. Jude Medical, Inc. or its related companies. Reprinted with permission of St. Jude Medical, © 2015. All rights reserved.)
The Electrophysiology Laboratory and Electrophysiologic Procedures
Facilities, Personnel, and Equipment
In earlier days, surgeons in the operating room performed pacemaker implantation. As the device underwent progressive miniaturization, the procedure could be performed without a thoracotomy and moved to the catheterization laboratory. The skills required also evolved to include the ability to cannulate a central vein and place the lead against the appropriate myocardial location. With the advent of implantable defibrillators and resynchronization therapy devices, the operating physician needed to be adept at cannulating the coronary sinus and other coronary venous branches, manipulating “over the wire” leads, and comfortable inducing VF during a procedure. The need to perform high-quality coronary sinus venograms in multiple planes reinforced the shift from operating room to EP laboratory.
The support personnel required for an implant procedure are the same as for EPSs and ablations. They include a scrub nurse or techni­cian familiar with the operating physician’s preferences, a circulating nurse, and an individual responsible for electrical testing. It is useful to have a cardiovascular radiology technician. The presence of a nurse to administer conscious sedation depends on whether an anesthesi­ologist or nurse anesthetist is present during the procedure. In some institutions, deep sedation for testing defibrillation thresholds is administered by the electrophysiologist.
In the EP lab, a representative of the pacemaker or defibrillator manufacturer is often present for clinical support. A well-trained rep­resentative has experience and knowledge of the company’s products, and his or her support is very helpful in achieving a good procedural outcome. However, responsibility for all aspects of the procedure remains with the operating physicians. Consideration of procedural liability dictates no hands-on involvement of nonhospital employees participating in more than a consultative role.
Pacemaker Systems Analyzer
For implantation of pacemakers and defibrillators, a pacemaker systems analyzer (PSA) is crucial. Because the circuitry replicates that of the PG, the PSA will accurately predict the performance of this generator. During our implants, we use the PSA provided by the manu­facturer. The physician credentialing for pacemaker and defibrillator implantation procedures is controversial and in the recent past has also been contentious. Studies show that in patients undergoing defi­brillator implantation, as a group, electrophysiologists are associated with fewer complications and better outcomes than cardiologists or thoracic surgeons.
https://t.me/med1917
6
The Electrophysiology Laboratory and Electrophysiologic Procedures 293
Surgical Type of Sterile Environment
Early concerns about sterility are well founded. A defibrillator or pace­maker is a foreign body that will remain in place for several years, and hence, infection is a primary concern. Generally, the operating room is considered to be a highly sterile area and offers the best protection from infection. The catheterization laboratory is considered to be an intermediate sterile and high-traffic area but offers the advantage of high-quality radiography with high-resolution images, multiple pro­jections, and image magnifications, along with the fact that these labo­ratories are fully equipped with all of the necessary catheters, sheaths, and wires that may be required. The right combination for device implantation is a dedicated EP laboratory, in which the ventilation system meets standards for operating rooms, and a rigid protocol for aseptic techniques.
The importance of adequate lighting for these procedures cannot be overemphasized. Many such procedures are performed on patients in whom antiplatelet and antithrombotic agents are not interrupted. To prevent pocket hematomas, it is important to be able to visualize the PG pocket. Although not always available in every laboratory, a high-intensity headlamp is particularly useful, especially when inspect­ing the pocket for bleeding vessels.
An electrocautery-surgery device involves the use of a pen that delivers alternating current in the radiofrequency range and creates resistive heating of tissues; it can be used for making incisions or achieving hemostasis.
Preprocedural Nursing Considerations
Patient Preparations
The evening before and morning of device implantation procedures, patients are instructed to scrub and wash the neck, shoulders, and chest with a chlorhexidine-containing soap. Preoperative skin cleans­ing with chlorhexidine-alcohol is superior to cleansing with povidone­iodine for prevention of surgical site infection, especially with
Staphylococcus aureus
have their chest hairs clipped before the procedure, and all patients should fast for at least 6 hours before. Although the administration of prophylactic antibiotics is controversial, we administer intravenous (IV) antibiotics immediately before the procedure. The IV line is started on the same side of the planned procedure to facilitate venog­raphy, if necessary.
Anticoagulation Issues
Device implantation in the anticoagulated patient has been controver­sial but is less so nowadays. Of late, several investigations have dem­onstrated the safety of performing implantation procedures without interrupting warfarin (Coumadin). In fact, it is safer to continue Cou­madin than to bridge the patient with heparin before and after the procedure. We continue maintenance anticoagulation for implant pro­cedures, especially if the patient has AF or prosthetic valves. The use of heparin postimplant is associated with more bleeding.
The use of novel oral anticoagulants is relatively new, and little data are available regarding the risk of bleeding with implants in patients using these medications. For patients undergoing implanta­tion of a cardiac device, it is our practice to interrupt the use of these medications in the perioperative period.
after clean-contaminated surgery. Male patients
Pacemaker Troubleshooting
1. Failure to deliver output
2. Failure to capture
294 6
https://t.me/med1917
Box 6 -2 Gathering Information for Effective Pacemaker
1. Indication(s) for pacing
2. Implantation date of each component
3. Pacemaker/leads model
4. Current programmed parameters
5. Battery volt age/impedance
6. Lead impedance(s)
7. M easured parameters
8. Chest x- ray if necessar y
3. Oversensing
4. Undersensing
5. Pacemaker-mediated tachycardia (PMT)
6. Lead dislodgement
7. Pacemaker syndrome
8. Runaway pacemaker
understanding its “normal function” is essential, as is obtaining basic information about the patient, implanted pacemaker leads, and pro­grammed parameters (Box 6-2). Believe it or not, considering that so many special features are available in different pacemakers, it is very easy to misinterpret normal pacemaker functions. Obtaining a thor­ough history is crucial to determine the course of a problem.
mise determines the urgency for repairing a problem. If the patient is severely bradycardiac and a pacemaker programmer is unavailable, transcutaneous or temporary pacing may be lifesaving. If a pacemaker malfunction occurs shortly after implantation, consider poor lead placement, lead dislodgement, loose set screws, and lead reversal, rather than lead conductor/insulation fracture and battery depletion.
magnet will terminate the tachycardia. In very rare cases of a “runaway pacemaker” (caused by a major pacing circuit component failure), urgent surgical PG removal is necessary.
devices is tracking AF or flutter at an upper tracking rate. Magnet placement over the PG drops the pacing rate to a “magnet rate” for that specific device until programmed parameters are changed (i.e., acti­vating a mode switch feature or using nontracking modes, such as [DDI] or [VVI]). Tachycardia in sensor-driven pacemakers may simply be remedied by turning the sensor “off.”
The Electrophysiology Laboratory and Electrophysiologic Procedures
Troubleshooting
Before attempting to troubleshoot a pacemaker malfunction,
The presence or absence of symptoms or hemodynamic compro-
In most cases of pacemaker-driven tachycardia, application of a
The majority of pacemaker-driven tachycardia by DDD or VDD
Electrocardiogram Assessment
It is critical to understand the basic timing of a dual chamber pace­maker (Fig. 6-18).
1. Look for pacing spikes.
2. If spikes are present, measure rate interval, presence of evoked response (complete paced beats), fusion beats, and pseudofusion beats.
3. If absent, apply a magnet (mode will be switched to AAI/VOO or DOO) to see spikes.
Causes of Absent Pacing Output
1. PG output failure
2. Oversensing (can be diagnosed by applying a magnet)
AP APVP VP
https://t.me/med1917
6
The Electrophysiology Laboratory and Electrophysiologic Procedures 295
Base rate interval
V–A Interval
Paced AV delay
PVARP
A
TARP
B
C
D
E
F
Figure 6-18 A, Basic timing in dual chamber pacemakers. TARP, Total
atrial refractory period; V-A, ventricular-atrial. B, Atrial oversensing, confirmed by marker channels showing atrial sensing (AS) markers without P waves. C, Ventricular oversensing. Observe marker channel for ventricular sensing (VS) without a QRS. D, Loss of ventricular capture (arrow). E , Atrial under­sensing; P waves without markers. F, Pacemaker Wenckebach (WB). Occurs when the atrial rate increases but does not exceed the 2 : 1 block point.
Continued
296 6
https://t.me/med1917
G
H
The Electrophysiology Laboratory and Electrophysiologic Procedures
Atrial undersensing
Placing a magnet on the device
I
Figure 6-18, cont’d
maker pace in the atria and ventricle (DOO). Arrow, Placing of a magnet on the device. I, Example of DDIR pacing mode: The response to sensing is to inhibit atrial output. No SAV is started, the ventricle is paced at the lower rate, and if af ter a V– A interval there is no AS, then AP/VP and PAV occur. DDIR is not appropriate for this patient due to a complete heart block (CHB). AS, Atrial sensing; PMT, pacemaker-mediated tachycardia; SAV, sensed atrioventricular.
G, Atrial undersensing. H, Magnet makes the pace -
Placement of a magnet on the PG is the next step. If spikes are seen, oversensing is the issue. If spikes are still absent, either the pacemaker is not putting out a pulse or the pulse does not reach the heart.
Pulse Generator Output Failure Causes
1. Battery depletion
2. Component failure
3. Electrocautery or defibrillator over or near the PG
4. Therapeutic radiation therapy over or near the PG
Noncapture Causes*
1. Low-output/high-capture threshold/exit block
2. Lead dislodgement
3. Lead conductor or insulation fracture
4. Loose set screws
5. Severe metabolic derangement or drugs
*Artificial noncapture: Pacing in the refractory period due to undersensing.
https://t.me/med1917
6
The Electrophysiology Laboratory and Electrophysiologic Procedures 297
Undersensing Causes (Undersensing Causes Overpacing)
1. Poor lead positioning
2. Lead dislodgement
3. Lead conductor/insulation fracture
4. Infarction of tissue near lead tip
5. Safety pacing
6. (DVI) mode
7. Ectopies with short amplitude
8. Severe metabolic derangement
9. Defibrillation over or near pacemaker PG Action needed: increase pacemaker sensitivity, change the sens-
ing mode to unipolar mode, and, if necessary, repair or replace lead.
Oversensing Causes (Oversensing Causes Underpacing)
1. Myopotentials
2. Cross talk
3. Lead conductor/insulation fracture
4. Electromagnetic interference
5. T-wave oversensing and far-field sensing
6. Loose set screw To repair, decrease the sensitivity. For T-wave oversensing and
far-field sensing, we recommend prolonging the refractory period.
The sensing mode can be switched from the unipolar to bipolar
setting. As is the case with unipolar sensing, the “larger antenna” is capable of causing oversensing. If leads are defective, lead repair or replacement is curative.
Pacemaker Syndrome
Pacemaker syndrome is seen in patients with normal sinus rhythm who have an inhibited VVI pacemaker or in those with dual chamber pacemakers if the atrial lead does not appropriately sense or capture the atrium.
Once atrial contribution/kick is lost, cardiac output diminishes
and the patient may experience dyspnea, fatigue, palpitations, chest/ neck/throat pulsations, and even syncope (as a result of atrial contrac­tion against closed tricuspid and mitral valves [Cannon A waves]).
In patients with VVI pacemakers, we recommend allowing more
time to be spent in intrinsic normal sinus rhythm by lowering the pacing rate. If this is not possible, upgrade to a dual chamber pace­maker. In patients with dual chamber pacemakers who have problems with atrial leads, reprogramming atrial lead parameters should resolve the issue; if not, atrial lead replacement is recommended.
Pacemaker-Mediated Tachycardia
Premature ventricular contraction (PVC), premature junctional con­traction (PJC), or loss of atrial sensing or capture can cause PMT. With retrograde conduction of a ventricular ectopy to the atrium after com­pletion of the postventricular atrial refractory period (PVARP), the retrograde “P” is sensed by the pacemaker, which in turn starts an AV
298 6
https://t.me/med1917
The Electrophysiology Laboratory and Electrophysiologic Procedures
delay after which a ventricular paced beat is forced. This beat can continue at the maximum rate of the upper tracking rate (PMT).
PMT can be terminated with the following actions:
1. Application of a magnet over the PG
2. Retrograde V–A conduction block
3. Programming a longer PVARP so that the retrograde P falls within the PVARP
Different manufacturers have different PMT remedies, such as PVARP extension for one beat after a PVC or DVI pacing for one beat after PVC.
Cross Talk
Cross talk occurs when the ventricular channel senses the atrial pacing spike and interprets it to be an intrinsic ventricular event, in which case ventricular output is inhibited. In patients who are pacemaker dependent and have no underlying escape mechanism, cross talk causes asystole, which may be fatal (on the ECG, one sees an atrial spike followed by P waves without a ventricular output/pacing spike).
In some cases, simply programing the atrial impulse to lower amplitude or decreasing ventricular sensitivity can remedy the problem. At times, increasing the blanking period is necessary.
Safety pacing is another remedy for cross talk. Allow a brief period of ventricular sensing early after atrial pacing spike delivery (this interval follows the blanking interval). If an event is sensed during this interval, the pacemaker will force a ventricular paced beat with a short AV delay of 100 to 120 msec (Fig. 6-19).
A
V
H
3
V
RBB
A
RB
2
V
1
200 msec
A
V
4
SN
A
V
5
Figure 6-19 Intracardiac electrograms recorded from various positions on
a multipolar catheter positioned across the tricuspid valve. Numbers 1 to 5 refer to the intracardiac location of the catheters along with the correspond ­ing electrograms; 1, most-distal location recording a large ventricular elec­trogram and no atrial electrogram; 5, most-proximal location displaying a large atrial electrogram with a small ventricular electrogram. The His poten­tial is observed when the catheter is in the area of the tricuspid annulus and the atrial and ventricular electrograms recorded are approximately of equal size (position 3). A , Atrium; Ao, aorta; AVN, atrioventricular node;
CS, coronary sinus; H, His; HB, His bundle; MS, membranous septum; PA, pulmonary artery; RB, right bundle; RBB, right bundle branch; SN, sinus
node; V, ventricle. (From Grossman W: Cardiac catheterization and angiog­raphy, ed 2, Philadelphia, 1974, Lea & Febiger.)
Ao
MS
CS
PA
HB
AVN
https://t.me/med1917
6
The Electrophysiology Laboratory and Electrophysiologic Procedures 299
Table 6 -2
Possible Components of Evaluation Before Electrophysiologic Testing*
Procedure Purpose
Histor y and physical
examination
Neurologic evaluation Perform if history and physical suggest to
Electroencephalogram Rule out seizure disorder Computed tomography/
magnetic resonance imaging
Carotid ultr asound Identify significant cerebrovascular
12-lead ECG Identify previous myocardial infarction
24- to 48 -hour ambulator y
ECG
Event recorder Correlation of symptoms with ECG events Head-up tilt-table testing Diagnose vasovagal/vasodepressor
Echocardiogram and
radionucleotide ventriculography
Stress test (with or without
perfusion scanning)
Cardiac catheterization Definition of coronary anatomy
ECG, Electrocardio gram; LV, left ventricular; Q–T, inter val from star t of Q wave to end of T wave; RV, right ventricular.
*Selected procedures may var y dep ending on clinical presentation.
Identify signs and symptoms of cardiac or
neurologic disease
Identify factors known to exacer bate
arrhy thmias
Determine details of syncopal events
rule out neurologic disease
Identify focal lesion
disease
Identify intraventricular conduction delays Identify prolonged Q –T interval Identify preexcitation syndromes Correlation of symptoms with ECG events Quantitation of ambient ectopy Identify diurnal variation in arrhythmia
syncope Assessment of LV and RV size and function Detection of valv ular pathology
Detection of reversible ischemia Assessment of ef fects of catecholamines
on arrhythmia induction
Clinical Evaluations of the Patient Before Electrophysiology Procedures
The operating physician must comprehensively evaluate the patient before the study and plan the procedure on the basis of the specific needs of the individual patient. Whenever possible, review ECG docu­mentation of the clinical event. Some or all of the procedures listed in
Table 6-2 may be included in this evaluation.
Any potentially reversible arrhythmogenic factors, such as elec­trolyte abnormalities or decompensated congestive HF, should be cor­rected before the study is performed.
Discontinue all antiarrhythmic medications for at least five half­lives before the baseline study. For supraventricular tachycardia (SVT) studies, discontinue medications influencing AV nodal conduction (e.g., β
-blockers, digoxin, and calcium channel blockers).
Premedication
The patient should have nothing by mouth after midnight except for essential cardiac medications, which may be taken with a small
300 6
https://t.me/med1917
The Electrophysiology Laboratory and Electrophysiologic Procedures
amount of water. The patient may be lightly sedated with IV diazepam or midazolam. In long diagnostic studies or in catheter ablation cases, more intensive levels of sedation are usually required. IV fentanyl and midazolam achieve somnolence in the patient during lengthy ablation procedures. However, excessive sedation may influence the ability to induce arrhythmias in some individuals.
Arterial and Venous Access
Routine diagnostic EPS involves stimulation and recording of electri­cal activity from the right side of the heart. Venous access may be obtained from the femoral, subclavian, internal jugular, or antecubital veins. A routine initial diagnostic EPS usually involves insertion of at least three catheters, most commonly via the femoral veins. These large veins easily permit the introduction of three 5-F to 8-F catheters per vein. The number of catheters required and venous access selected depend on the type of study being performed and the data being col­lected. Mapping of a left-sided bypass tract may necessitate catheter­ization of the left side of the heart for precise mapping.
For patients requiring beat-to-beat assessment of the hemody­namic effects of an induced arrhythmia, an arterial line may be placed. To completely evaluate a patient with Wolff-Parkinson-White (WPW) syndrome, the physician may have to access the LV for stimulation or recording of intracardiac electrical activity. If arterial catheterization is necessary, administration of IV heparin is mandatory. Patients undergoing prolonged studies involving venous access or those with a history of venous thromboembolism should also receive heparin.
Study Protocol
Although details of the protocol vary depending on the indication for the EPS and the information being obtained, most studies involve the recording and measurement of spontaneous intracardiac events and observation of the effects of programmed electrical stimulation. An initial study usually takes approximately 2 hours and depends on the complexity of the case. Possible components of the initial comprehen­sive EPS are listed in Box 6-3.
Box 6 -3 Possible Components of Comprehensive Initial
Electrophysiologic Study*
Measurement of basic intervals Determination of sinus node function Determination of atrial, AV nodal, His-Purkinje, and ventricular conduction
and refractoriness Identification of presence of dual AV nodal pathways Identification of presence, location, and electrical proper ties of accessor y
AV pathways Attempts to induce SV T Attempts to induce V T Determination of mechanism of induced arrhythmias Mapping of origin site(s) of induced arrhy thmias Determination of effect of IV antiarrhythmic drugs on induced tachycardia Determination of efficacy of anti-tachycardia pacing for induced tachycardia
AV, Atrioventricular; I V, intravenous; SVT, supraventricular tachycardia; VT, ventricular tachycardia.
*The actual procedure varies depending on the individual case; not all par ameters are assessed in all cases.
https://t.me/med1917
6
The Electrophysiology Laboratory and Electrophysiologic Procedures 301
Positioning of Catheters
The RA is easily accessible by any venous access. The most common site for stimulation and recording is the high posterior lateral wall in the region of the sinus node. The potential is most easily found with a catheter introduced by the femoral approach and passed into the RA across the tricuspid valve and into the RV. The catheter is with­drawn across the tricuspid valve with the application of a slight degree of clockwise torque that tends to keep the catheter in contact with the septum (see Fig. 6-21). A hexapolar or octapolar catheter in the His position may be used so that electrical activity from multiple electrode pairs can be recorded, and the one showing the most stable and con­sistent His potential may be displayed. If multiple attempts to record a His potential with one catheter are unsuccessful, a differently shaped or steerable catheter with a deflectable tip may be used. In most patients, a His potential can be recorded successfully.
Usually, the LA is approached indirectly by recording in the coro­nary sinus. Cannulation of the coronary sinus os can be achieved from the femoral, left subclavian, or right or left internal jugular approaches. The appropriate position of the coronary sinus catheter is verified fluo­roscopically in the LAO and right anterior oblique positions. The cath­eter curves upward toward the left shoulder in the LAO projection and is posteriorly oriented in the lateral projection. On the ECG, atrial and ventricular electrograms are recorded, with the timing of the LA elec­trograms appearing later than the high RA electrogram. Direct record­ing of electrical activity from the LA is possible in patients with a patent foramen ovale or atrial septal defect or by a trans-septal approach.
In most baseline studies, a catheter is placed in the RV apex. For ventricular stimulation protocols, a second ventricular catheter is sometimes placed in the RV outflow tract, although most operators prefer to reposition the catheter from the apex to the outflow tract. The order in which the catheters are placed in the right side of the heart varies among operators, and no particular order is mandatory. In
ommended that the RV apical catheter be positioned first to ensure adequate ventricular pacing in the event of catheter-induced trauma to the right bundle, which may result in complete heart block (CHB) while the His catheter is being positioned.
Catheterization of the LV may be necessary in patients with VT or preexcitation syndromes. The LV is most commonly accessed by the retrograde arterial approach for mapping procedures. Fluoroscopy that permits views in multiple planes is essential to ensure accurate positioning of the catheter. Stimulation may also be performed from the LV in cases in which patients with clinically documented sustained VT have arrhythmias that are noninducible with use of the standard protocol from the RV. To avoid inadvertent cannulation of the coronary arteries, the operator safely prolapses the catheter across the aortic valve. In patients with atrial septal defects, patent foramen ovale, or undergoing trans-septal punctures, the LV may be approached through the LA and mitral valve.
Measurement of Conduction Intervals
After the catheters are positioned, basic conduction intervals are mea­sured, including the basic sinus cycle length (A-to-A interval), P-wave duration, A–H interval, His spike duration, and H–V interval (Fig. 6-20). Measurements from the surface ECG, including the P–R, QRS, and Q–T intervals, are also recorded. Conduction interval measurements and refractory period measurements should be made at a paper speed of