Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3597_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
87 Мб
Скачать
Right Left
https://t.me/med1917
3
Coronary Angiography and Ventriculography 123
Secondary
curve
Primary
curve
Secondary curve
Primary curve
Tertiary curve
Figure 3-16 Judkins right (JR) and Judkins lef t (JL) coronary catheters,
identif ying primary, secondar y, and tertiary cur ves of each catheter. These curves are designed to facilitate entr y into the ostia of each coronary ar tery.
Tertiary curve
A
Figure 3-17 Judkins lef t (JL) coronary catheter position. A, Correct align-
ment. B, Incorrect position and overinsertion. (From King SB, Douglas JS Jr: Coronar y arteriography and angioplasty, New York, 1985, McGraw-Hill.)
catheter in the same patient angles upward and a larger (5 cm) cath­eter angles downward into the coronary cusp. When the coronary ostium cannot be seated correctly or placed in a stable position, the catheter should be replaced with a better-fitting catheter rather than aggressively manipulated into the coronary artery (Fig. 3-17). A slight rotation of the catheter may be necessary to improve alignment of the catheter tip with the LM coronary trunk.
B
124 3
https://t.me/med1917
Coronary Angiography and Ventriculography
A
C
Figure 3-18 Judkins right (JR) coronar y catheter position. A, Correct align-
ment. B, Tip subselectively in conus arter y. C, Tip wedged in proximal ste­nosis. D, Tip impinging on lateral vessel wall. (From King SB, Douglas JS Jr: Coronary arteriography and angioplasty, New York, 1985, McGraw-Hill.)
The JR coronary catheter is sized by the length of the secondary curve and is available in 3.5-, 4-, and 5-cm sizes. In most cases, the 4-cm catheter is adequate. The JR catheter is advanced into the ascending aorta (usually with LAO projection) with the tip directed caudally (Fig. 3-18).
The RCA can be entered in most cases by one of two maneuvers:
1. Advance the catheter into the right coronary cusp and rotate the
catheter 45 to 90 degrees clockwise as the tip is pulled back 2 to 3 cm. As the right coronary orifice is engaged, the fluoroscope shows rotation of the tip toward the right coronary cusp and down­ward motion of the catheter.
2. Advance the catheter tip to 2 to 4 cm above the valve. When the
catheter is rotated clockwise for 45 to 90 degrees, the tip rotates toward the right cusp and descends approximately 1 to 2 cm, engaging the right coronary ostium from above.
If the coronary ostium is not engaged, the maneuvers are repeated, starting at a slightly different level each time. A brief contrast media injection into the right coronary cusp may help the operator direct the catheter. A slight but firm push-pull motion on the catheter is neces­sary to translate rotational motion at the hub down to the tip.
If stored rotational energy is not released by a small counter rota­tion after seating, the catheter may spring out of the right coronary ostium when the patient takes a deep breath. After seating, the opera­tor checks for pressure damping associated with ostial stenosis or conus branch cannulation.
B
D
Amplatz-Type Catheters
The left Amplatz-type catheter is a preshaped half-circle with the tapered tip extending perpendicular to the curve (Fig. 3-19). Amplatz catheter sizes (left 1, 2, and 3 and right 1 and 2) indicate the diameter of the tip curve. In most normal-sized adults, no. 2 left and no. 1 right
https://t.me/med1917
3
Coronary Angiography and Ventriculography 125
A
B
Figure 3-19 Catheterization of the coronary arteries (Amplatz technique,
left anterior oblique [LAO] projection). A, Catheterization of the left coronary artery (LCA). 1, The lef t coronar y catheter is advanced until the secondary curve rests in the noncoronary posterior aortic cusp and its tip points to the left coronary ostium. 2, The catheter is gently advanced and retracted until the left coronary ostium is engaged. B, Catheterization of the right coronary ar tery (RCA). 1, The right coronary catheter initially may point to the left coronary sinus. 2, It is withdrawn slightly and rotated clockwise until the tip points toward the RCA and the secondar y curve rests against the left aortic cusp. 3, The RCA is engaged as the catheter is advanced and withdrawn. (From Tilkian AG, Daily EK: Cardiovascular procedures: Diagnos ­tic techniques and therapeutic procedures, St Louis, 1986, Mosby.)
1 2
1 2 3
(modified) Amplatz catheters give satisfactory results. In the LAO pro­jection, the tip is advanced into the left aortic cusp. Further advance­ment of the catheter causes the tip to move upward into the LM aortic trunk. It may be necessary to push Amplatz catheters down to move the tip up and out of the ostium to disengage the catheter from the LM ostium. If the catheter is pulled instead of being advanced, the tip moves downward and into the LM or CFX artery. Unwanted deep can­nulation of the CFX artery might tear this branch or the LM aortic trunk. The incidence of coronary dissection is higher with Amplatz catheters than with Judkins-type catheters.
The AR mod catheter has a smaller but similar hook-shaped curve. The catheter is advanced into the right coronar y cusp. As with JR catheters, the catheter is rotated clockwise by 45 to 90 degrees. The same maneuver is repeated at different levels until the RCA is entered. After coronary injections, the catheter may be pulled, advanced, or rotated out of the coronary artery.
Multipurpose Catheters
MP catheters are those that are primarily straight or slightly angled with an end hole and two side holes placed close to the tapered tip. An MP catheter can be used for left and right coronary injections. Extra care is required if this catheter is used for ventriculography.
126 3
https://t.me/med1917
Coronary Angiography and Ventriculography
Special-Purpose Femoral Angiography Catheters
The right coronary vein graft catheter is similar to a JR catheter but has a wider, more open primary curve that allows cannulation of a vertically oriented coronary artery vein graft. The left coronar y vein graft catheter is similar to the JR catheter but has a smaller and sharper secondary curve that allows easy cannulation of LAD and left CFX vein grafts, which are usually placed higher and more anterior than right coronary grafts with a relatively horizontal and upward takeoff from the aorta.
The internal mammary artery (IMA) graft catheter has a peculiar hook-shaped tip configuration that facilitates the engagement of IMA grafts, especially in patients with a vertical origin of the IMA.
Preformed Coronary Catheters from the Radial Artery Approach
Coronary Catheters from the Radial Artery Approach
Cannulation of the LCA using a preformed catheter is easier than can­nulation using an MP catheter. JL and JR coronary catheters can be used through the left or right arm with satisfactory results. Compared with the femoral technique, a JL3.5 catheter may be necessary with the left radial technique.
An Amplatz catheter can be used effectively from either the right or the left arm. This catheter is manipulated in a fashion similar to that described for the femoral approach. The coronary catheters can be removed without a guidewire, but the pigtail ventriculography catheter should be removed over a guidewire to straighten the pigtail loop and prevent it from kinking or lodging in the subclavian or axillary artery.
A special consideration from the arm approach is negotiating a sharp caudal turn near the origin of the right carotid arter y. To facili­tate passage into the central aorta, the patient should be instructed to take a deep breath to pull down the great vessels and heart while the operator advances the guidewire and catheter. The manipulations to seat the preformed radial catheter are similar to those for the Amplatz and MP techniques (see Fig. 3-19, C).
Radial Universal Catheters
Catheters used for coronar y angiography from the radial approach are called generically “universal catheters” because they can engage both the left and right coronary ostia. Among the most common is the “Jacky Catheter” (see Fig. 3-15, B). This catheter is positioned much like Amplatz maneuver catheters are positioned. The advantages of the Jacky and other universal catheters are reduced catheter exchanges with reduced potential for radial/brachial artery spasm, radiation exposure, and procedure time. The Jacky has a less acute terminal curve than the Tiger (TIG) catheter and has a tendency to sit more coaxially with the LMCA and engage the RCA with fewer movements.
Although some operators perform ventriculography through uni­versal catheters because there are side holes near the catheter tip, we do not recommend this practice (see end of the end-hole left ventricle [LV] gram, page 95 and figure 3-22). Certainly, a hand injection for ventriculography may be inadequate. In addition, the Jacky catheter tip frequently points directly toward the anterior wall of the LV, causing ventricular ectopy and possible perforation.
Saphenous Vein Graft and Internal Mammary Artery Angiography Catheters
The right coronary vein graft catheter is similar to a JR catheter but has a wider, more open primary curve that allows cannulation of a
https://t.me/med1917
3
Coronary Angiography and Ventriculography 127
vertically oriented coronary artery vein graft. The left coronar y vein graft catheter is similar to the JR catheter but has a smaller and sharper secondary curve that allows easy cannulation of LAD and left CFX vein grafts, which are usually placed higher and more anterior than right coronary grafts with a relatively horizontal and upward takeoff from the aorta.
The IMA graft catheter has a peculiar hook-shaped tip configura­tion that facilitates the engagement of IMA grafts, especially in patients with a vertical origin of the IMA.
Ventriculography Catheters (Pigtail, Halo, Multipurpose)
The pigtail catheter is the safest choice for ventriculography and is the most commonly used ventriculography catheter, with a preshaped tip making nearly a full circle ~1 cm in diameter. There are 6 to 12 side holes on the straight portion of the catheter above the curve. To enter the LV, the pigtail catheter is advanced to the aortic valve. The loop is positioned to the left in the RAO projection (resembling a “6”), and the catheter is pushed against the valve to make a U shape that facili­tates entry into the ventricle during deep inspiration. Inside the ven­tricle, the catheter can be placed in front of the mitral valve with the loop directed toward the apex, away from the valve (in the RAO posi­tion). A slight rotation, advancement, or withdrawal may be necessary to find a “quiet” position (one that does not cause frequent premature ventricular contractions) (Fig. 3-20). An angled (145 degrees) pigtail catheter may be helpful for this purpose, especially for horizontally oriented hearts.
A Halo catheter is a novel 5-French (5-F) catheter with a perpen­dicular helical tip with an inwardly and upwardly directed tip. The side holes are located on the helix and produce equivalent left ven­triculograms with minimal ectopy because the contrast jets are directed inward and not to the myocardium (Fig. 3-21). It is excellent for measuring distal LV chamber pressure in hypertrophic cardiomy­opathy because, in contrast to a pigtail catheter, there are no holes along the shaft to create a falsely low distal intraventricular pressure gradient.
3
Figure 3-20 Method of lef t ventricle (LV) catheterization. 1, Having crossed
the aortic valve, the pigtail catheter is in position. 2, The catheter is with­drawn 2 to 3 cm and rotated 70 to 90 degrees counterclockwise. 3, The coiled loop is in the inflow tract of the mitral valve. 4, If the catheter moves excessively in this position, it should be advanced until it is stable. (From Judkins MP, Judkins E: Coronary arteriography and left ventriculography: Judkins technique. In King SB III, Douglas JS Jr, editors: Coronar y ar teriog­raphy and angioplasty, New York, 1985, McGraw -Hill.)
1
2
4
128 3
https://t.me/med1917
Figure 3-21 Halo ventriculography catheter. Side holes are contained only
within the spiral, which is directed inward to reduce premature ventricular contractions.
Coronary Angiography and Ventriculography
Incorrect Incorrect
Correct
A
Figure 3-22
(MP) end- hole catheter in the right anterior oblique (RAO) projection. Correct positioning requires the catheter tip to be free in the center of the ventricle, near the tip of the papillar y muscles, and aiming toward the apex. If it is touching the inferior or inferolateral wall (left), it should be rotated clockwise. If it is touching the anteroseptal wall (right), it should be rotated counter­clockwise. For ventriculography, use of end-hole catheters should be aban­doned in favor of pigtail catheters for improved safety. (From King SB, Douglas JS Jr: Coronary arteriography and angioplasty, New York, 1985, McGraw-Hill.)
In the past, the MP and other end-hole-type catheters have been used for ventriculography. This and other end-hole catheters are no longer recommended for ventriculographic studies because the high­pressure contrast jet may produce ventricular tachycardia (VT), con­trast injection in the myocardial tissue (contrast staining), or ventricular perforation (Fig. 3-22).
A, Catheterization of the left ventricle (LV) with a multipurpose
Common Problems in Accessing and Cannulating Coronary Arteries and Grafts
Left Coronary Artery
Short  Left  Main,  Separate  Ostia  for  Left  Anterior  Descending  and Circumflex Arteries.  In patients with a short LM artery segment
or separate ostia, it may be necessary to cannulate the LAD and CFX arteries separately. Using a smaller JL catheter (i.e., JL3.5, not JL4) often permits selective cannulation of the LAD artery. Slight with­drawal and clockwise rotation or use of a larger JL5 catheter also favors cannulation of the CFX artery. An Amplatz-type catheter is especially useful to cannulate the CFX artery separately but must be used with care to avoid artery dissections.
B-1a
https://t.me/med1917
B-1b
B-1c
B-1d
Figure 3-22, cont’d
in the RAO projection followed by power contrast injection with an ACIST medical injector, 13 mL/sec to maximal volume of 45 cc (less was injected when staining was noted). The right frame shows beginning of myocardial staining. 1b, Continued injection frames more staining of the LV. 1c, Dense fixed staining of contrast into the myocardium. When viewed in the left anterior oblique (L AO) projection (see B, 2a frames), one can see that this catheter was positioned into the lateral wall of the LV. 1d, Persistent con­trast stain after the end -hole ventriculogram.
B, 1a, Lef t to right, A 5- F Jacky catheter positioned
(Continued)
130 3
https://t.me/med1917
B-2a
Coronary Angiography and Ventriculography
B-2b
Figure 3-22, cont’d
the 5 -F Jacky catheter recorded during the power contrast injection as in
B, 1a. First and second frames show the beginning of myocardial staining. 2b, Continued deep penetration of contrast media into the myocardium
without perforation into the pericardial space with a dense fixed staining of myocardium. Fortunately, there was no clinical sequalae associated with this procedure. An echocardiogram, per formed urgently before the patient left the laboratory, showed no pericardial effusion or wall motion abnormal­ity. (B, Reprinted with permission from the Cath Lab Digest, copyright HMP Communications.)
High  Left  Coronary  Artery  Takeoff.  An unusually high origin of
the LMCA from the aorta can usually be cannulated with use of an MP catheter or an Amplatz-type catheter (e.g., AL 2). A long, tapered-tip MP catheter may be used to cannulate the high-origin LM trunk through the brachial approach.
Wide  Aortic  Root.  In patients with a relatively horizontal or wide
aortic root with upward takeoff of the LMCA, a large-curve JL5 or JL6, AL coronary, or MP catheter may be required.
B, 2a, Left to right, Simultaneous RAO LV gram with
Right Coronary Artery
The origin of the RCA shows more variation than that of the LCA. When encountering difficulty in locating the ostium, a contrast injection low into the right coronar y cusp helps direct the catheter. If the RCA is not seen with this flush injection, it may be totally occluded or may origi­nate anteriorly on the aorta or from the left sinus of Valsalva. In this case, the orifice is usually located above the sinotubular ridge. An AL catheter or a left bypass graft catheter can be used successfully to engage the RCA orifice located anteriorly or in the left cusp. Minimal anterior displacement of the RCA from the right coronary sinus is more common. In this case, the JR catheter tip may not be directed toward the right but looks foreshortened in the familiar LAO view. Directing
https://t.me/med1917
the catheter tip to the right in the usual fashion using the lateral view permits easy cannulation of the anteriorly directed right coronary orifice. Rarely, an aortogram is necessary to confirm the presence of the RCA.
Wide  Aortic  Root.  In a patient with a horizontal and wide aortic
root, cannulation of the right coronary orifice and right coronary cusp may require an Amplatz or MP catheter.
High  Right  Coronary  Artery  Takeoff.  A relatively high origin of
the RCA may require an AL or AR mod catheter. The most common coronary anomaly (see page 137, nary Anomalies) is the CFX artery originating from the RCA or right coronary cusp and coursing posteriorly and downward. This location may be cannulated easily with use of an AR catheter.
3
Coronary Angiography and Ventriculography 131
Angiography of  Common Coro-
Assessment of Coronary Stenoses
The degree of an angiographic narrowing is estimated as the percent­age lumen reduction of the most severely narrowed segment com­pared with the adjacent angiographically normal vessel segment, seen in the worst x-ray projection. Because the operator uses visual estima­tions, an exact evaluation is impossible; in general, there can be a ± 20% variation between readings of two or more experienced angiog­raphers. Stenosis severity alone should not always be assumed to be associated with abnormal physiology (blood flow) and ischemia. Moreover, CAD is a diffuse process, and thus, minimal luminal irregu­larities on angiography may represent significant albeit nonobstruc­tive CAD at the time of angiography. The stenotic segment lumen is compared with a nearby lumen that does not appear to be obstructed but that may have diffuse atherosclerotic disease (Fig. 3-23). This explains why postmortem examinations and intravascular ultrasound (IVUS) imaging describe much more plaque than is seen on angiog­raphy. The percent diameter is estimated from the angiographically normally appearing adjacent segment. Because coronary arteries nor­mally taper as they travel to the apex, proximal segments are always larger than distal segments, often explaining the large disparity among several observers’ estimates of stenosis severity.
Percent diameter is the common measurement reported to convey stenosis severity. Recall that area of stenosis is greater than diameter stenosis. Estimates of narrowing assume that the lumen is circular, but the lumen is more often eccentric. Although the practice has been to report “exact” measurements (even when using only the naked eye), realistically and for practical purposes, there are only four categories of lesion severity:
1. Minimal or mild CAD, narrowings <50%
2. Moderate, stenosis between 50% and 70%
3. Severe, stenosis between 70% and 95%
4. Total occlusion (100%)
Technical note: Stenosis anatomy should not be confused with abnormal physiology (flow) and ischemia, especially for lesions 40% to 70% narrowed. For nonquantitative reports, the length of a stenosis is simply mentioned (e.g., LAD proximal segment stenosis diameter 25%, long or short). Other features of the coronary lesion may not be appreciated by angiography and may require IVUS or optical coher­ence tomography (OCT) (Fig. 3-24).
Several different angiographic pathologies commonly seen during coronary or peripheral vascular angiography include thrombus, dis­section, aneurysms, and coronary fistualae.
A thrombus is defined as a lucency within a vessel or chamber surrounded by at least three sides of contrast in an appropriate clinical setting (e.g., acute coronary syndromes). In the coronar y vessels, a thrombotic occlusion may totally block the flow of contrast and have a meniscus appearance (Fig. 3-25, A).
132 3
https://t.me/med1917
Coronary Angiography and Ventriculography
Proximal Distal
A
Diameter decrease
Angiographic
view (diameter)
Histologic
view (area)
Cross-sectional
decrease
B
Figure 3-23 A, Diagram of coronary arter y with stenosis (top) and corre-
sponding intravascular ultrasound (IVUS) images demonstrating diffuse nature of coronary artery disease (CAD). B, The percent narrowing is com ­pared only with the “normal”-appearing angiographic lumen, which may not be normal at all.
Extent
Severity
50%
50%
75%
75%
75%
95% 98%
88%
88%
0 25%
50% 25% 0
Composition
Complication
Figure 3-24
angiographic imaging and can be evaluated by intravascular ultrasound (IVUS).
Angiographic features that may not be appreciated on cine-