54 Handling catheters
Diameter
e diameter of the catheter should be as small as
possible to accomplish the task at hand. Most angi-
ography is performed with a 4-Fr or 5-Fr catheter
over 0.035-inch guidewires.
Length
e catheter must be long enough to reach the
target site and still have enough length outside
the patient for appropriate manipulations. Most
commonly used catheters range from 65 cm to
100 cm in length but shorter and longer catheters
are available for specic purposes. In general, the
shortest catheter that will perform the task should
be used (Table 6.1). e catheter length is typically
selected based on the distance from the access site
to the target. Keep in mind, that the length selected
must be supported by the guidewire being used.
Sometimes, a longer guidewire must be placed to
support the chosen catheter.
Head shape
Catheter head shape determines function. A brief
glance at a catheter helps to determine its intended
purpose. e straight ones are generally exchange
catheters. e ones with large rounded heads are
ush catheters, especially if there are multiple side
holes. e angled ones are selective catheters of
some kind, many with specic purposes.
Special features
Special features of catheters include various coat-
ings, radiopaque tips, and graduated measurement
markers.
CATHETER HEAD SHAPE
DETERMINES FUNCTION
e potential for constructing dierent catheter head
shapes is unlimited. Although hundreds are currently
being marketed, most catheter usage in most prac-
tices is based on the consistent and well-developed
use of just a few types. Each specialist has functional
favorites. Flush and selective catheters have divergent
purposes and substantially dierent appearances.
General catheter types are shown in Figure 6.1.
ese include ush, exchange, and selective catheters.
Selective catheters may have either a simple curve or
a complex curve. Some of the catheters commonly
used in endovascular practice are listed in Table 6.2.
Flush arteriographic catheters, such as the pig-
tail or tennis racket, are useful for higher pressure
Table 6.1 Catheter length purposes
Length (cm) Purpose
15–20 Dilator
65 Abdominal aortography with
lower extremity runoff
65–80 Selective renal and visceral
catheterization
90 Arch aortography, thoracic
aortography
100 –120 Selective cerebral catheter
65 –120 Exchange catheter
A B C D
Fig. 6.1 Representative examples of the major
categories of catheters as detailed in Table 6 . 2 .
(A) Flush catheters are used for aortography. They
have an end hole and multiple side holes for high-
pressure, high-volume contrast administration. This
flush catheter example is a tennis racket catheter.
A variation on this is the pigtail catheter, which is
often used for arch aortography. (B) The straight
catheter is used as an exchange catheter, to place
a different guidewire into the artery. (C) Selective
catheters have an end hole and a specially shaped
tip. Selective catheters may have a simple curve, for
directing a guidewire, as with this TegT catheter.
(D) Selective catheters may also have a complex
curve, as in this C2 Cobra. These catheters have
a larger head that must be reformed in the aorta
before being used to cannulate an artery. They
have specifically designed shapes and are used to
enter aortic side branches such as carotid or renal
arteries. Complex curve catheters are often used to
cannulate the innominate or common carotid arter-
ies in patients with a bovine arch or a tortuous arch.
https://t.me/med1917

Catheter head shape determines function 55
Table 6.2 Catheters for general endovascular practice
Catheter type Catheter name Function Length (cm) Caliber (Fr) Comments
Flush Pigtail Aortography 65, 90, 100 4, 5
Tennis racket Aortography 5
Omni
®
Flush Aortography: select aortic bifurcation,
select renal artery
65 4, 5
Exchange Straight Exchange guidewires, general
arteriography
70, 90, 100 4, 5
Selective: simple curve
Short bent tip TegT Direct guidewire through lesion or into
branch (30°)
70, 100 5 Other examples: Berenstein,
Vert, DAV
Kumpe More angled than TegT(45°) 40, 65 5
Long bent tip Multipurpose A Direct guidewire, longer tip (45°) 65, 100 5
Multipurpose B More angled than Multipurpose A (70°) 100 5
Hook shape RIM Tight curve 65 5 Other examples: Hook,
ChuangA, B, D, and E,
Shepherd Hook
Selective: complex curve
Cerebral Simmons Reshape to enter difficult arch
branches, direct guide
100 5 Other examples: see Chapter 9,
Table 9.2
Vitek Reshape to enter difficult arch
branches
100, 125 5
Renal C2 Cobra Directs guidewire into side branch at
90-degree angle
65, 80 4, 5 Other examples: SosOmni
®
2,
C3 Cobra
Renal double curve Directs guidewire laterally and
inferiorly acute angle
80 5 Other examples: renal curve
andrenal curve 2
Abbreviation: RIM, Roche Inferior Mesenteric.
https://t.me/med1917

56 Handling catheters
injection (up to 1200 psi) performed while minimiz-
ing a jet eect that might destabilize arterial plaque
or thrombus. Most aortography and some peripheral
arteriography can be performed with these catheters.
e rounded catheter head can be converted to a
hook shape by positioning a guidewire partially into
the head of the catheter. Most aortography begins
with one of these ush catheters. e shape of the
catheter head may be altered dramatically, depend-
ing on the location of a guidewire placed within it.
Although this is true for all catheters that are not
straight, a catheter with a signicant curve may
assume the widest variety of congurations depend-
ing on the location of the guidewire (Figure6.2). It
is not unusual to have a momentary but irritating
challenge in removing a ush catheter over a guide-
wire. is is because the ush catheter usually has a
head shape that makes the passing guidewire form
a 180-degree loop with the tip of the wire coming
back toward the operator. When the rmer segment
of the guidewire reaches the catheter head, the cath-
eter head ips up straight. If the catheter head is in a
location that is too small in diameter to handle this
maneuver, or there is disease nearby that could be
dislodged by the ipping catheter head, consider
taking another approach. Another option is to pass
the guidewire through the catheter head but leave
the catheter head curved. Go ahead and pull the
catheter back over the guidewire and the curves at
the head of the catheter should straighten. Do not let
go of the guidewire as you walk down the wire when
doing the “pin and pull” maneuver. Sometimes, if
you let go of the wire, the loop in the catheter head
maintains itself and actually pulls the guidewire
back with it.
e straight catheter is useful, as a general
exchange tool, for interval arteriography and to
measure pressure proximal and distal to a lesion.
It can add stiness to a guidewire to allow it to
pass or probe a lesion more easily or uncoil a wire
that has piled up proximal to an unpassable lesion.
When a guidewire is in the correct location but it
is not a sti enough rail over which to pass a larger
endovascular treatment device, an exchange cath-
eter is placed and the indwelling wire exchanged
for an exchange guidewire or a very sti wire.
Selective catheters have either a simple or a com-
plex curved head shape. e dierence between
A B C D E
Fig. 6.2 Catheter head shape depends on guidewire position. The shape of the catheter head can be
significantly modified by the position of the guidewire. This is especially true for flush catheters and com-
plex curve selective catheters. (A) A guidewire is present within the shaft of a hook-shaped visceral cathe-
te r. (B) The hook of the catheter’s head splays out as the guidewire enters the curve. (C)As the floppy part
of the guidewire tip exits the catheter, the head of the catheter splays further. (D) The catheter straightens
as the firm portion of the guidewire shaft occupies the curved portion of the catheter head. (E)As the firm
portion of the guidewire exits the catheter head, the catheter straightens out completely.
https://t.me/med1917

Catheter head shape determines function 57
a simple curve and a complex curve catheter is
demonstrated in Figure 6.3. Every shaped cath-
eter has a primary curve. When a catheter has
only a single curve (primary curve) near its tip, it
is a simple curve catheter. is is usually a “hockey
stick” shape and it provides an angle of trajectory
for a guidewire tip passed through the catheter that
may be anywhere from just a few degrees to almost
90 degrees from the sha of the catheter itself
(Figures 6.3 and 6.4B). A complex curve catheter
has a primary curve and also a secondary curve,
and occasionally a tertiary curve. A complexcurve
A
B
Primary curve
Secondary curve
Primary curve
Fig. 6.3 Comparison of simple curve and complex curve catheters. (A) Simple curve catheters have
only a primary curve, which determines how they function. (B) Complex curve catheters have a
primary curve and a secondary curve and, sometimes, “a tertiary curve”. Complex curve catheters
go into the body in the straight configuration because they go in over a guidewire. When a complex
curve catheter is near its destination, the wire is withdrawn and the catheter head begins to take its
shape. The catheter head usually must be reformed or reshaped for the catheter head to take its
functional shape. This is shown in Chapter 9, Figure 9.4.
A B C
E F
D
Fig. 6.4 Selective catheters in action. (A)ACobra catheter is used for renal artery catheterization.
(B) A Berenstein catheter is used for cannulation of the superficial femoral artery. (C) A tightly curved
hook-shaped catheter is used for catheterization of the visceral arteries. (D)AHeadhunter catheter
can be used to enter the branches of the aortic arch. (E) A Simmons catheter is used to cannulate the
carotid and innominate arteries. (F)AShepherd Hook catheter is used to cross the aortic bifurcation.
https://t.me/med1917

58 Handling catheters
catheterheadneeds to bereshaped inside the arte-
rial system before it can be used to selectively can-
nulate a side branch (Figures 6.3 and 6.4E). e
complex curve catheter head assumes a relatively
straight conguration while it is being passed
over the guidewire. When it arrives at its destina-
tion, the guidewire is withdrawn and the complex
curve catheter attempts to shape itself. Specic
maneuvers must be performed to achieve the cor-
rect shape. is is discussed in detail in Chapter9.
e larger the head of a complex curve catheter,
the more space in the vascular system is needed to
reshape the head. e largest of the complex curve
catheter head shapes is the Simmons 3.
e bent tip Berenstein catheter is a simple
curve selective catheter, but it can be used simul-
taneously for general applications. is catheter is
useful for directing a guidewire through a critical
lesion or into a branch vessel. e bend at the tip
of the catheter confers directionality to the guide-
wire tip. e angled Glidecath
®
is similar but with
a soer catheter tip. e soness of the tip slightly
blunts the degree of angulation as the wire leaves
the catheter head. e vertebral catheter (Vert)
provides slightly more angulation and the catheter
tip is a little bit stier. Hook-shaped catheter heads
are useful for turning an acute angle at a tight cor-
ner. Complex curve catheter heads have a curve
in one direction, then back in the other direction.
ese are used primarily to select out aortic side
branches for cannulation. Examples include the
Simmons and the H3 Headhunter cerebral cath-
eters. Many head shape congurations have been
designed for selective catheterization, each with a
dierent purpose (see Chapter 9 for a detailed dis-
cussion of selective catheterization).
e rate at which contrast may be administered
through a catheter varies with the catheter caliber,
its length, and the number of side holes (Table 6.3).
DESCRIBING CATHETER BEHAVIOR
●
Trackability. e ability of the catheter to fol-
low the guidewire through tortuous vessels and
around corners without pulling the wire out
of its intended location or without halting its
advance over the wire.
●
Pushability. e description of how a force
applied by the operator at the hub of the cath-
eter relates to the forward movement of the tip
(the leading edge) of the catheter.
●
Crossability. e facility with which a cath-
eter follows the guidewire across a lesion or
through a diseased arterial segment.
●
Steerability. e steering responsiveness of the
catheter tip to handling maneuvers performed
at the hub.
●
French. e scale used to size catheters
(1 Fr = 0.33 mm).
HANDLING CATHETERS
Aer the guidewire has been placed into the major
ow stream of the arterial segment of interest, the
next challenge is to pass the catheter into the cor-
rect location. e catheter should be suitable for
the diameter sizing of the guidewire. In general,
the catheter will follow the guidewire when it is
advanced incrementally and the wire is pinned to
ensure that it does not also advance. Do not advance
the catheter unless the guidewire is pinned. If work-
ing over a hydrophilic guidewire, the wire must be
pinned with a pincer grasp so that it is not acciden-
tally advanced with the push forward on the cath-
eter. e hydrophilic guidewire must also be wet
to have pushability of the catheter. e catheter is
advanced by grasping it a certain distance from the
hub of the sheath, usually with the dominant hand,
and giving it a push. Do not advance the catheter
by grasping on the catheter hub. Grasp the catheter
along its sha and at a distance from the hub of the
sheath that can be supported by the overall stiness
of the combined guidewire and catheter. If the n-
gers are too far from the place where the catheter
goes into the sheath, the catheter and guidewire
will just bend, instead of the catheter sliding over
Table 6.3 Contrast media flow rates through
flush and selective catheters
a
Type
Caliber
(Fr)
Length
(cm)
Maximum
flowrate
(mL/second)
b
Flush
(multiple
side holes)
4 65 19
4 100 15
5 65 32
5 100 27
Selective
(end hole)
5 65 15
5 100 11
a
Data from Cook Medical, using 100% Oxilan 350
at25°C.
b
Maximum pressure = 1200 psi.
https://t.me/med1917

Handling catheters 59
the wire. Each catheter and guidewire combination
has a certain body to it based on the overall stiness
and strength of the combined structures. e stier
the guidewire–catheter combination, the farther
down the catheter may be grasped with each incre-
mental grab and push by the operator. An innate
sense will be developed for this aer undertaking
many cases. If the catheter is grasped too far down
the sha, it will bend instead of advance. If it is
grasped in increments that are too small, this easy
part of the case may take all day.
e catheter will not advance easily over the
oppy tip portion of the guidewire (Figure 6.5).
When the catheter reaches the oppy portion of
the wire, the catheter head begins to take its shape
and pushability is dramatically decreased. e
guidewire may be advanced to ensure that a rm
segment of the wire is available for the catheter to
advance.
If the intended catheter cannot be advanced
over the guidewire and pushability is lost, con-
sider a stier wire to give additional support to
the catheter. is situation is oen encountered
in tortuous arteries, especially if they are calci-
ed. If the catheter cannot be advanced over the
guidewire through a lesion, the lesion itself may be
providing the resistance; either the stenosis is so
severe that the guidewire itself occludes the resid-
ual lumen, or the wire is subintimal.
When the catheter cannot be advanced into
position because of signicant vessel tortuos-
ity or because of the long distance between the
catheter head and the entry site, a steady and
gentle withdrawal of the guidewire while the
catheter is advanced decreases friction at the level
of the catheter head and usually accomplishes
the requisite advancement if the distance is not
too far. is is called the “push-pull” technique.
e risk of using this maneuver is potentially los-
ing guidewire access. Do not advance the cath-
eter beyond the end of the guidewire or allow
the catheter head to reform during advancement.
If the guidewire–catheter combination tends to
buckle freely, the operator can choose to insert a
stier wire. Another option is to leave the guide-
wire in place, remove the catheter, insert a long
sheath to decrease friction, and reinsert the cath-
eter through the sheath.
Technique: Catheter handling
1. Prepare the catheter by ushing and wiping
with heparin–saline solution.
2. Choose a catheter size, length, and head shape
that will most easily accomplish the proposed
task.
3. Place the guidewire in the correct location.
is is probably the most important factor in
successful catheter placement.
4. Whenever a guidewire is in place across the
access site without a catheter, hold pressure
at the arterial entry site until the appropri-
ate dilator, sheath, or catheter is passed to
tamponade bleeding from the arteriotomy.
A B C D
Fig. 6.5 Advancing the catheter over the stiff portion of the guidewire. (A) The catheter is advanced
into the arterial entry site over the floppy portion of the guidewire, which lacks the shaft strength to
support advancement of the wire. (B) The catheter buckles and advancement stalls. (C) The guidewire
is advanced so that the floppy portion is well inside the vasculature. (D) The catheter is advanced over
the stiffer portion of the guidewire.
https://t.me/med1917

60 Handling catheters
Bleeding into the subcutaneous tissue during
exchanges leads to postprocedure ecchymosis.
Pressure in the right place helps to minimize
the likelihood of hematoma formation.
5. Use a mosquito clamp to dilate the skin entry
site to the appropriate size for the intended
sheath or catheter.
6. Pass the sheath or catheter through the
skin and so tissue over the sti portion of
the guidewire, not on the oppy segment
(Figure6.5).
7. Consider using sequential arterial dilators, a
stier guidewire, or a hemostatic access sheath
if the catheter cannot be advanced across the
entry site.
8. Advance the catheter incrementally, a few
centimeters at a time.
9. Handle the catheter near its entry site when
advancing and at its hub when steering.
10. Maintain hand control of the guidewire.
Anassistant should pin the guidewire so that
it does not simultaneously advance.
11. If the catheter will not cross a lesion and
arteriography is the goal, use a dierent
approach and do not force it. If therapy is
the goal (e.g.,balloon angioplasty or stent
placement), make sure the guidewire is in
the correct location across the lesion. en
predilate the lesion with a small diameter
angioplasty balloon.
12. If the catheter will not advance because of
long distance and/or multiple turns from
the leading edge to the operator’s hand, and
there is plenty of guidewire in place ahead
of the catheter tip, gently and steadily with-
draw the wire while advancing the catheter
to reduce friction. Continuous fluoros-
copy is required during this maneuver to
ensurethat the guidewire is not withdrawn
too far.
13. If the catheter has a longer distance to travel
and cannot be advanced, it is usually due to
accumulated friction. Place a long sheath to
decrease friction, use a stier guidewire, or
choose an alternative entry site.
14. Aer the catheter is in place, aspirate and
ush with heparin–saline solution. Do not
inject anything unless there is appropriate
blood return.
15. Pu contrast to verify placement and ensure
that the catheter tip is in a safe position before
high-pressure injection is performed.
16. When removing or exchanging the catheter,
maintain guidewire placement by “ walking
along” the wire while the catheter is
being pulled (Figure 6.6). Intermittent
fluoroscopy checks should be performed
to ensure that the guidewire position is
maintained.
17. If the end of the catheter becomes knotted,
pass a sti guidewire into the catheter to
untangle it.
18. If there is any doubt about the location of
the catheter, make sure it is intraluminal.
Check for blood return. Twirl the catheter
and visualize the catheter head moving using
A
B
Fig. 6.6 “Walking along” the guidewire to
remove a catheter. When a catheter is exchanged,
the guidewire position is maintained while the
catheter is removed. The usual way to do this is
to use two hands and the “pin and pull” tech-
nique. The guidewire is pinned with one hand
and the catheter pulled with the other hand.
The alternative is to use one hand to remove
the catheter. (A)The catheter hub is grasped
between the thumb and the forefinger and the
guidewire is grasped between the fourth and fifth
fingers of the same hand. (B) The catheter hub
is withdrawn while the guidewire position is held
constant. Thefourth and fifth fingers then grasp
the guidewire again at a further distance along its
length and the maneuver is repeated. This creates
the appearance of the hand “walking along” the
guidewire.
https://t.me/med1917

Handling catheters 61
uoroscopy to be certain that it moves freely.
Pu contrast to see if it rapidly enters the ow
stream rather than staining the wall of the
artery or remaining stagnant.
19. It is common to have diculty identifying
the catheter as it is being placed. Consider
using a catheter with a radiopaque tip. When
advancing any catheter over a guidewire, there
is a slight bend of the wire at the location of
the catheter head and this imprint can be
monitored uoroscopically as it is advanced.
20. Remove the catheter over a guidewire to
straighten the catheter head. is prevents
injury to the endoluminal surface.
https://t.me/med1917

https://t.me/med1917

63
7
Guidewire and catheter passage
THE GOAL OF THE PROCEDURE
DETERMINES THE COURSE OF
THE GUIDEWIRE–CATHETER
APPARATUS
e guidewire–catheter combination is the endo-
vascular workhorse. Before you go to any trouble,
be as precise as possible about establishing the goal
of the procedure and anticipating the likely scope of
the arteriography or intervention. e operator must
decide what guidewire–catheter combination should
be used in any given case and decide how to get the
selected combination into the position required.
Aer percutaneous access has been achieved (see
Chapter 2), an initial sheath is placed (see Chapter 3).
Guidewires and catheters have become familiar tools
(see Chapters 4–6), and the next step is guidewire
and catheter passage to the desired location. e goal
of the intended procedure should be well proscribed
prior to the puncture. Whether arteriography or a
therapeutic intervention is planned will inuence
the puncture site choice and the approach. Options
should be considered for converting the procedure
from a strategic one to a therapeutic one if the need
should arise. Access for endovascular therapy is dis-
cussed in Chapter 14.
When arteriography is the goal, the small-
est acceptable access sheath should be placed.
Angiographic catheter head placement at specic
locations is required for visualization of the desired
anatomic area. Selective catheterization for arteri-
ography or therapy requires specic steps, which are
discussed in Chapter 9. Details of catheter head place-
ment for arteriography are discussed in Chapter 10.
ese maneuvers have been separated into a stepwise
approach for illustrative purposes. As the operator
gains facility, one step leads to the next in a ow of
events that lead toward the goal of the procedure.
Initial catheter placement
Bony landmarks or other anatomic signs are usu-
ally adequate to establish the correct location for
catheter head placement (Fig u re 7.1). e catheter
head may be advanced or withdrawn slightly with-
out reintroducing the guidewire, since the ush
catheter tip is usually rounded. Hand-powered
injection of contrast to establish the position is
usually worthless through a 5-Fr catheter into the
aorta. Hand-powered contrast injection may be
used to visualize smaller, lower-ow arteries. Aer
the catheter is in place, aspirate, remove bubbles,
ush with heparin–saline, and connect to the
injector, uid meniscus to uid meniscus. en
have the technician aspirate through the injector
and inspect the line for air bubbles.
Fluoroscopy for catheter placement
Table 7.1 summarizes some general concepts about
when to use uoroscopy for guidewire and catheter
placement. e guidewire should be advanced so
that its oppy portion extends beyond the loca-
tion desired for catheter head placement. Aer
the guidewire is appropriately placed, uoroscopy
stops while the catheter is prepared and advanced
over the guidewire outside the patient. Fluoroscopy
is initiated again when the catheter is in the artery.
e progress of the catheter is monitored using
uoroscopy until it is in the desired location. Some
catheters have a bright tip or radiopaque tip to help
visualize them as they are being placed. If not, the
catheter is typically visualized as an enhanced
shadow advancing over the wire as the added prole
of the wire and the catheter is apparent in compari-
son to the shadow of the wire alone. Fluoroscopy is
used as the guidewire is withdrawn to ensure that