Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3835_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
15.09.2026
Размер:
13 Мб
Скачать
☆
32 T. Ma and Q. Zhou
https://t.me/medicina_free
compared to 5 mm with the 35-MHz transducer, was observed at 90 and 120-MHz, respectively. At the same time, the CNRs of the images at 90 and 120-MHz dropped to 2.1 and 1.6, respectively. Figure 2.14d shows the phantom image generated by the 150-MHz transducer. Although the resolution of t he image was further improved, the image could only penetrate less than 500 µm in depth and provided weaker contrast for the strong acoustic reflectors, such as surface of polyimide tube and boundary of graphite rod. Thus, it appears that transducer at such frequency is not suitable for IVUS imaging applications.
Despite the decreased penetration depth and CNR, IVUS transducers of 90 and 120-MHz are still capable of providing valuable high-resolution information around the inner lumen area. The fused phantom images captured by 35/90-MHz catheter and 35/120-MHz catheter are shown in Fig. 2.15a, b, respectively. The image co- registration in each imaging pair is accomplished by simply rotating one of the IVUS images by 180°. These results suggest that the multi-frequency catheters can achieve a better characterization of the artery tissue by integrating the deep imaging depth of lower frequency transducer and high-resolution of higher frequency transducer.
To further investigatethe feasibility of the multi-frequency catheters in a setup that is closer to the clinical settings, tissue-mimicking phantom imaging was performed with the presence of blood. The center lumen of the phantom was filled with fresh swine blood and the polyimide tube was still filled with DI water. As expected, a slightly decreased penetration depth of 4 mm and a reduced CNR value of 3.2 were obtained in the image at 35-MHz (Fig. 2.16a) due to the strong attenuation of the blood in the lumen. As shown in Fig. 2.16b–d, the much stronger attenuation of the higher frequency ultrasonic waves in the blood dramatically limits the imaging depth and degrades the image quality. These results indicate that to improve imaging depth with a high-frequency transducer may require the temporal removal of luminal blood during the in vivo imaging, similar to the flushing mechanism in OCT, which is an intrinsic drawback of multi-frequency IVUS imaging.
Fig. 2.15 Fused tissue-mimicking phantom images captured by a 35/90-MHz multi-frequency IVUS catheter and b 35/120-MHz multi-frequency IVUS catheter. White color: 35-MHz ultrasound image; orange color: 90-MHz ultrasound image; green color: 120-MHz ultrasound image; dynamic range: 45 dB; scale bar: 1 mm (Ma et al. 2015b)
2 Advances in Multi-frequency Intravascular Ultrasound (IVUS) 33
https://t.me/medicina_free
Fig. 2.16 Tissue-mimicking phantom images in the presence of blood at a 35-MHz, b 90-MHz, c 120-MHz, and d 150-MHz. Dynamic range: 45 dB; scale bar: 1 mm (Ma et al. 2015b)
Fig. 2.17 IVUS images of the human coronary artery at a 35-MHz, b 90-MHz, and c 120-MHz.
Dynamic range: 50 dB; scale bar: 1 mm (Ma et al. 2015b)
Human cadaver coronary artery images acquired at 35, 90, and 120-MHz are dis­played in Fig. 2.17. The 35-MHz transducer provides more complete morphological information of the coronary artery owing to the deep penetration depth (Fig. 2.17a). With the 90-MHz transducer, the three-layer structure (intima, media and adven­titia) pointed by the red arrow is identified in Fig. 2.17b because of the improved axial resolution. However, it is blurred and not apparent in Fig. 2.17b. In Fig. 2.17c, although the 120-MHz transducer can only image through the intima layer due to the limited penetration depth, an improvement in both axial resolution and lateral reso­lution is achieved, indicated by a reduced speckle size in the image of intima layer. The fused image pairs acquired by the 35/90 and 35/120-MHz integrated catheters were shown in Fig. 2.18a, b respectively. The integrated i mage pairs, combining the advantages of the high-resolution superficial image of lumen area offered by the high-frequency IVUS transducer (90 or 120-MHz) and the deep penetration image of the entire artery wall provided by the low-frequency transducer (35-MHz), yield a more comprehensive visualization of the artery wall and plaque volume.
Summary
We have successfully developed and prototyped multi-frequency IVUS imaging sys­tem catheters with three different frequency combinations prepared by PMN-PT and
34 T. Ma and Q. Zhou
https://t.me/medicina_free
Fig. 2.18 Fused IVUS images of human coronary artery captured by a 35/90-MHz multi-frequency IVUS catheter and b 35/120-MHz multi-frequency IVUS catheter. White color: 35-MHz IVUS image; orange color: 90-MHz IVUS image; green color: 120-MHz IVUS image; dynamic range: 50dB;Scalebar:1mm(Maetal.2015b)
LiNbO3single-crystals. The multi-frequency IVUS catheter, with a clinical compat­ible size of 0.95 mm in diameter, is featured by the back-to-back arrangement of a conventional IVUS transducer and a high-frequency IVUS transducer to achieve accurate co-registration of two IVUS images. The performance of the high-frequency IVUS transducer at different frequency ranges (90, 120, and 150-MHz) was evaluated and compared to find the optimal frequency range for the high-frequency transducer in the multi-frequency IVUS catheter, considering imaging depth, imaging resolu­tion, and CNR. Tissue-mimicking phantom imaging with and without the presence of blood shows that the multi-frequency catheter carries the complementary strengths of the deep imaging depth of the conventional IVUS and the high-resolution of the high-frequency transducer. On the other hand, these results also point out a potential weakness of the multi-frequency catheter, which is flushing of the luminal blood is required to ensure the functionality of this technique. The in vitro human cadaver coronary artery imaging demonstrates the capability of the multi-frequency catheter to provide more comprehensive visualization of the vascular structure and to facil­itate the assessment of the vulnerable plaque. Compared to other multi-modality intravascular imaging techniques, the multi-frequency IVUS imaging capitalizes the advantage of cost-effectiveness because only a moderate modification of the current commercial IVUS system is needed. Besides, the apparent clinical utility of this ultrasound-only technology can be promptly explored during the translational stage since most of the interventional physicians are familiar with the ultrasound technol­ogy.The future translation of the multi-frequency IVUS imaging into clinical use will not only consolidate the leading status of the IVUS technology in the interventional cardiology practice, but also prosperously lead to patient benefits.
2 Advances in Multi-frequency Intravascular Ultrasound (IVUS) 35
https://t.me/medicina_free
Some Advances in High-Frequency IUVS
Introduction
The performance of the multi-frequency IVUS transducers used in Section “Multi-
-frequency IVUS Imaging” is summarized in Fig. 2.19, including center frequency,
imaging depth, imaging contrast, and imaging resolution. The results are in agree­ment with the classic ultrasound theory that increasing the center frequency of the transducer results in better imaging resolution and decreased imaging depth. Even though different piezoelectric materials were selected to fabricate the IVUS trans­ducers at different center frequencies, there was still some electrical impedance mis­match between the IVUS transducers and electrical system due to the fact that the transducer aperture size was set to be 0.5 mm. Future implementation of a matching circuit into the system will potentially solve this issue and improve the overall per­formance of multi-frequency IVUS catheter. There is a significant gap between the conventional IVUS and the competitive technology OCT, indicated by the red dot, regarding the image resolution and penetration depth. Combining a higher frequency IVUS transducer in a frequency range of 90–120-MHz with the conventional IVUS catheter is a more feasible and simpler solution to reduce the controversy caused by insufficient resolution and fill the gap between the conventional IVUS and OCT. However, the requirement of temporal removalof luminal blood with the aim of elim­inating the stronger attenuation effect for higher frequency ultrasound is a weakness of this technology that needs to be addressed in the future. Although further increas­ing the center frequency of the transducer to 150-MHz or higher could theoretically
Fig. 2.19 Summaries of IVUS transducers’ performances at different center frequencies in this study, including imaging depth, imaging contrast, and imaging resolution. Red dot: OCT; dashed blue dot: 1-3 composite IVUS transducer (Ma et al. 2015b)
36 T. Ma and Q. Zhou
https://t.me/medicina_free
reach the resolution level of OCT, the extremely shallow imaging depth and reduced imaging contrast will dramatically degrade the image quality. Thus, it may not be worthwhile to use the transducer with such high frequencies for IVUS applications. As discussed in Section “Multi-frequency IVUS Imaging”, owing to its broadband characteristics, the 90-MHz transducer exhibits a better axial resolution than the 120­MHz transducer. This suggests that the integration of a broader bandwidth transducer is favorable to further renovate the multi-frequency IVUS catheters.
Therefore, one direction to advance IVUS technology is to implement the state­of-the-art composite materials. The newly developed micromachined PIN-PMN-PT single-crystal 1-3 composite transducer at 40-MHz for IVUS application, providing an improvedaxial resolution of 43 µm and a comparable penetration depth to conven­tional IVUS transducers, was investigated (Li et al. 2014a; Yuan et al. 2008). For the possible implementation of 1-3 composite material in multi-frequency IVUS imag­ing, the proposed 1-3 composite transducers in a frequency range of 60–90 MHz, indicated by the blue dashed dot in Fig. 2.19, serves as a more promising substitute for the high-frequency transducers in the multi-frequency IVUS catheters so that the enhanced resolution can be achieved without excessively sacrificing imaging depth. Moreover, it is the fact that lateral resolution of the multi-frequency IVUS catheter system is almost three to four times worse than the axial resolution, which is also the technical barrier and natural shortcoming of all other IVUS systems. The images quality of ultrahigh-frequency IVUS in the deeper region because of the further downgraded lateral and contrast-to-noise ration. Some beamforming techniques that combine virtual source synthetic aperture (VSSA) focusing and coherence factor weighting (CFW) are reported to have the possibility to improve the lateral reso­lution of IVUS imaging. Furthermore, as discussed previously, a further increase of operating frequency will allow for the improvement of axial resolution, while the penetration depth of IVUS will be sacrificed. Thus, an imaging method that improves the penetration depth of high-resolution IVUS would also be of major clinical importance. Modulated excitation imaging is known to allow ultrasound waves to penetrate further. Qiu et al. recently reported an ultrasound system specif­ically for modulated-excitation-based IVUS imaging (Qiu et al. 2017). The system incorporates a high-voltage waveform generator and an image processing board that is optimized for IVUS applications. The results show that the modulated excitation IVUS system was able to provide increases of 86.7% in penetration depth and 9.6 dB in the signal-to-noise ratio for 60-MHz IVUS. In this section, we will introduce three advances of ultrahigh-frequency IVUS technology including the implementation of broadband 1-3 composite IVUS transducer to enhance axial resolution, utilization of virtual source synthetic aperture focusing techniques to improve lateral resolution, as well as the modulated excitation method to increase the penetration depth.
2 Advances in Multi-frequency Intravascular Ultrasound (IVUS) 37
https://t.me/medicina_free
Development of PIN-PMN-PT 1-3 Composite IVUS Transducer
Imaging resolution is inversely proportional to the frequency bandwidth of the trans­ducers. Transducers with higher center frequency and broader bandwidth can provide better image resolution (Foster et al. 2000b). In our previous work, 80-MHz IVUS transducers were investigated and could provide axial resolution of 35 µm and pen­etration of 2 mm (Li et al. 2011). The fine resolution is critical for detecting certain vulnerable plaques such as thin-cap fibroatheroma (TCFA), which has a cap thick­ness less than 65 µm (Kolodgie et al. 2001). The drawback of working at 80-MHz is the shallow penetration depth resulting from the strong attenuation in blood and vessel wall. An alternative method to improve resolution without sacrificing penetra­tion depth is to use transducers with broader bandwidth at relatively lower frequency (40-MHz). Moreover, a broadband transducer is advantageous for multi-frequency signal processing to differentiate blood and vessel (Li et al. 2008).
To achieve optimal imaging performance, the piezoelectric material of an IVUS transducer must be carefully chosen. Conventionally, Pb(Zr,Ti)O (PZT)-based ceramics, such as PZT-5H, have been used for fabricating IVUS transducers. Recently, binary-relaxor-based ferroelectric single-crystal Pb(Mg
1/3Nb2/3
high-frequency ultrasonic transducers (Rhee 2007; Yuan et al. 2006). Compared with PZT ceramics, PMN-PT crystal demonstrated improved sensitivity and band­width, resulting from its higher piezoelectric coefficient (d electromechanical coupling coefficients (k and Shrout 1997b; Tian et al. 2007a; Zhang and Shrout 2010). Ceramic or crystal 1-3 composite material exhibits even broader bandwidth than monolithic materials because of composite’s very high coupling coefficient and low acoustic impedance (Smith 1989). However, composites for transducers operating above 20-MHz cannot be made using traditional dice-and-fill technique because of the thickness limitation of the dicing saw blade, which is no less than 10 µm (Cheng et al. 2003; Zhang et al.
2011; Zhou et al. 2011a). Mechanical dicing can also produce highly stressed and
damaged surface layers, thereby degrading the properties of 1-3 composites (Lee et al. 2010). For 1-3 composites, the frequency of the first lateral mode is empirically expressed as
)-PbTiO3(PMN-PT) has also been investigated for fabricating
~ 2000 pm/V), and
33
~ 0.58, k33~ 0.9) (Luo et al. 2000;Park
t
3
where flis the frequency of first lateral mode resonance, VTis the shear wave velocity of filler, and d However, previously reported micromachined 1-3 composites were fabricated from PMN-PT crystal. The binary PMN-PT crystal has relatively low coercive field (E
2.5 kV/cm), depoling temperature (T 130 to 178 °C) (Park and Shrout 1997a; Tian et al. 2007b; Ye and Dong 2000; Zhang
V
T
=
f
√
l
is the kerf width (Jiang et al. 2006;Tianetal.2007a).
P
2d
2
P
~ 60 to 95 °C), and Curie temperature (Tc~
R/T
~
c
38 T. Ma and Q. Zhou
https://t.me/medicina_free
et al. 2008). Higher Ecwould allow transducers to be driven in a higher electrical field and higher T
for a broader temperature usage range (Tian et al. 2007b; Zhang
R/T
et al. 2008). During the fabrication process, transducers are frequently incubated at elevated temperatures (40–90 °C) for epoxy curing, wax bonding, electrode sputter­ing, etc., which can cause PMN-PT depoling. Depoling can make testing transducer during fabrication inconvenient. In addition, an IVUS catheter must go through a sterilization process before entering the market. Some sterilization processes, such as steam sterilization, work at relatively high temperature (90 °C). After sterilization, it is difficult to repole the IVUS transducer without taking the catheter out of the ster­ilized pouch. Therefore, thermal instability of PMN-PT may degrade transducers’ performance during fabrication and sterilization processes. By adding lead indium niobate (PIN) into the PMN-PT system, the ternary crystal PIN-PMN-PT was found to have superior electrical and thermal stability (E to 130 °C, T
~ 160 to 200 °C) to overcome the aforementioned drawbacks, while
c
maintaining the excellent piezoelectric properties (d
~ 5.5 to 6 kV/cm, T
c
~ 900 to 1900 pm/V, k33~0.83
33
R/T
~ 120
to 0.92) (Tian et al. 2007b; Zhang et al. 2008). PIN-PMN-PT has been investigated for high-frequency ultrasonic applications and demonstrated reduced temperature dependence of performance and minimal depolarization during fabrication (Chen et al. 2012; Sun et al. 2009). Composites based on PIN-PMN-PT crystal inherit the improvements of the thermal and electrical properties (Tian et al. 2010). Considering the merits of higher usage temperature range and coercive field, PIN-PMN-PT could be more advantageous than PMN-PT to fabricate micromachined 1-3 composite for high-frequency IVUS applications. In this paper, we report the use of micromachined PIN-PMN-PT 1-3 composite material for IVUS application. Miniature side-looking needle transducers and flexible IVUS catheters were built from the composite to oper­ate at 40-MHz. Ex vivo experiments were conducted to demonstrate the superiority of the composite for IVUS application.
PIN-PMN-PT crystal 1-3 composite was fabricated using inductively coupled plasma (ICP)-enhanced DRIE technique by H. C. Materials Corp. (Tian et al. 2010). Polished PIN-PMN-PT crystal plate was first coated with Cr/Au as the seed layer for subsequent nickel electroplating. Photolithography was used to lay out the etching pattern over the Cr/Au layer. A 4- to 5-µm-thick nickel mask was then electroplated with the inverse pattern of the photoresist. Crystal parts with the patterned nickel mask were loaded in ICP-enhanced plasma (Oxford Instruments, PlasmaLab 100, Abingdon, Oxford, UK) for DRIE. Following etching, kerfs in the etched crystals were filled with epoxy (EPO-TEK 301, Epoxy Technology Inc. Billerica, MA). After epoxy cures, the composite was lapped from both sides to the final thickness. The composite plate was sputtered with Cr/Au electrodes on both sides and then deposited with a thick conductive backing material. A cross-sectional micrograph of the composite on top of a thick conductive backing material is shown in Fig. 2.20. The thickness of the composite is around 27 µm and the kerf width is less than 5 µm. DRIE produces a near vertical etched profile with a side wall angle greater than 85°.
Side-viewing needle-type IVUS transducers and flexible catheter-type IVUS transducers were fabricated based on the PIN-PMN-PT 1-3 composite. The mea­sured f
was 41-MHz and BW was 86%. BW is comparable to the reported PMN-PT
c
2 Advances in Multi-frequency Intravascular Ultrasound (IVUS) 39
https://t.me/medicina_free
Fig. 2.20 PIN-PMN-PT 1-3 composite with conductive backing (Li et al. 2014b)
composite transducer (77–90%) (Yuan et al. 2006, 2008; Sun et al. 2010; Jiang et al.
2008), but almost doubled that of a PMN-PT crystal transducer (45%) (Zhou et al. 2007b). Tungsten wire targets with 6-µ m OD were imaged to determine axial and
lateral resolutions of the transducer, as shown in Fig. 2.21a. For comparison, the wire targets were also imaged by a 40-MHz PMN-PT single-crystal transducer (two
2
matching layers; one backing layer; 0.5 × 0.4 mm
aperture; unfocused; 47% BW (Zhou et al. 2007b), as shown in Fig. 2.21b. The axial and lateral resolutions were determined from the −6 dB envelope width from the wire located at 1.3 mm, which is close to the transducer’s natural focus (1.1 mm). For the PIN-PMN-PT composite
Fig. 2.21 Wire phantom images from a PIN-PMN-PT composite transducer; and b PMN-PT single-crystal transducer (Li et al. 2014b)
40 T. Ma and Q. Zhou
https://t.me/medicina_free
transducer, axial and lateral resolutions were 43 and 226 µm, respectively. For the PMN-PT single-crystal transducer, axial and lateral resolutions were 62 and 278 µm. As shown in Fig. 2.21, the wire phantom images of the composite transducer were much thinner than those of the single-crystal transducer because of the improved axial resolution. Images are displayed with 50 dB dynamic range.
An ex vivo IVUS image from the composite transducer is shown in Fig. 2.22a. For comparison, an image of the artery from the PMN-PT crystal transducer of the same cross section is shown in Fig. 2.22b. Both images are capable of differentiating the thickened intima (I), media (M), and adventitia (A) layers, as well as visualizing the calcified plaques (Ca, high-intensity areas followed by acoustic shadow). However, because of the improved axial resolution of composite transducer, Fig. 2.22a displays higher clarity (finer speckles) of all the three layers. The boundaries of IM and MA (denoted by green arrows) were identifiable by the composite transducer in Fig. 2.22a but are blurred in Fig. 2.22b and even unidentifiable at 9 to 11 o’clock. The improved resolution is also helpful to accurately determine the percentage of calcified plaque over the entire lumen circumference.
The micromachined PIN-PMN-PT single-crystal 1-3 composite material demon­strated improved effective electromechanical coupling coefficient k
(eff) compared
t
with monolithic single-crystal. The usage temperature of the composite is at least 30 °C higher than PMN-PT. The acoustic impedance is as low as 20 MRayl. Using the composite, needle-type and flexible-type IVUS transducers were fabricated and tested. 41-MHz transducers with 86% bandwidth were achieved, which resulted in 43-µm axial resolution. The characterization results of PIN-PMN-PT crystal com­posite showed superior piezoelectric properties which were comparable to that of PMN-PT-based crystal composite, as well as improved thermal property compared with that of PMN-PT based composite. PIN-PMN-PT crystal can be an alternative approach for fabricating high-frequency composite. Ex vivo IVUS imaging was con-
Fig. 2.22 IVUS image of human coronary artery at 40-MHz from a PIN-PMN-PT composite transducer; and b PMN-PT single-crystal transducer. I = intima; M = media; A = adventitia; Ca = calcified plaque. Green arrows denote the boundaries of IM and MA (Li et al. 2014b)
2 Advances in Multi-frequency Intravascular Ultrasound (IVUS) 41
https://t.me/medicina_free
ducted to demonstrate the superiority of improved axial resolution. The composite transducer was able to identify the three layers of a human coronary artery specimen with high definition, whereas the image from a single-crystal transducer showed lower clarity of the layered structures.
IVUS with Virtual Source Synthetic Aperture Focusing and Coherence Factor Weighting
As a widely recognized clinical tool for intravascular imaging and the irreplaceable technical basis of most multi-modality intravascular imaging techniques, however, IVUS itself still requires further improvement in the overall performance including spatial resolution and contrast-to-noise ratio (CNR) over the whole range of interest. Several approaches that have been reported to improve the IVUS image quality are illustrated in Fig. 2.23 by simulating the two-way beam pattern of single-element IVUS transducers with a 0.5-mm active aperture size in PZFlex (Weidlinger Asso­ciates, Cupertino, CA). Figure 2.23a shows the two-way beam pattern of a conven­tional flat 40-MHz IVUS transducer that is naturally focused at 1.7 mm. Increasing
Fig. 2.23 Simulated two-way beam patterns and −6 dB beamwidth contour (black dashed line) of IVUS transducers with 0.5 mm active aperture size: a 40-MHz IVUS transducer with flat aperture, b 80-MHz IVUS transducer with flat aperture, c 40-MHz IVUS transducer with mechanical focusing at 3 mm. d Simulated beam pattern from (a) with expected −6 dB beamwidth contour (green dash–dot line) after VSSA&CFW processing. The x-axis is lateral direction, and the z-axisisaxial direction (Yu et al. 2017)