Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3835_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
15.09.2026
Размер:
13 Мб
Скачать
☆
10 Therapeutic IVUS and Contrast Imaging 237
https://t.me/medicina_free
Fig. 10.3 Representative histology images of porcine coronary arteries treated with rapamycin microbubbles and IVUS. a Artery section treated with rapamycin microbubbles only. b Artery section treated with rapamycin microbubbles and therapeutic intravascular ultrasound. “I” and “M” denote vessel intima and media, respectively. Scale bars are 500 µm. Source Kilroy et al. (2015)
flow) direction. Clearly, for maximum ARF effect, the transducer should be as long in the axial direction as possible consistent also with the desire to minimize the rigid length near the device tip. A rigid tip creates challenges when negotiating tortuous vessels. Using a similar ARF optimized IVUS transducer, enhanced delivery of a model drug was demonstrated in ex vivo and in vivo arteries (Kilroy et al. 2014a). The long duration, moderate pressure, and ARF pulses were used to affect model drug delivery from the microbubbles accumulated on the vessel. It was observed that drug delivery was restricted to within the −6 dB beamwidth (Fig. 10.4). This demonstrates the potential for ARF to assist in providing a spatially controlled drug delivery effect to a vessel wall. Using both axial and rotational translations, it fol­lows that it is possible, in principle at least, to “paint” the vessel wall according to a therapy plan individualized to any unique asymmetric lesion.
Microbubbles in IVUS Imaging
Shell-stabilized gas-filled microbubbles are widely used as the basis of a contrast agent in ultrasound imaging. Comprehensive reviews of microbubble design and clinical applications are available in the literature (Cosgrove and Lassau 2010; Unnikrishnan and Klibanov 2012). As a blood flow tracer, in their simplest usage, microbubbles provide a bright echo signal that delineates the
238 J. A. Hossack
https://t.me/medicina_free
Fig. 10.4 a Microscope image of DiI delivery along the vessel wall. Gray-scale bar measures the pixel intensity in arbitrary units of fluorescence. The dotted yellow lines denote the −6-dB beam width of the ultrasound transducer (1.3 mm). b Fluorescence intensity increase over background in the artery plotted along the circumference in an ex vivo artery where delivery was performed in phosphate-buffered saline solution. Plot axes were truncated to show that the fluorescence intensity increase occurs along the entire vessel wall. A maximum fluorescence intensity increase of 10× was measured in this artery with a mean of 5.3 ± 2.9 fold. This result was produced with a sonoporation pulse with a peak negative pressure of 2 MPa and pulse repetition frequency of 1 kHz. The region from 0 to 2 mm has a lower fluorescence intensity increase because the catheter was not centered and may have blocked the flow of microbubbles to this region. Source Kilroy et al. (2014a)
extent of the vasculature and the blood filled chambers of the heart. In this context, microbubbles have been used for assessing myocardial perfusion defects following myocardial infarction (Wei et al. 1998). They have also been used to assess enhanced perfusion and angiogenesis in cancer applications (Hohmann et al. 2003; Zhao et al.
2010). In additional to having a role in assessing anatomy (via enhanced of vessel
boundaries) and function (i.e., blood flow/perfusion), microbubbles are now finding applications using ligand-based molecular targeting to vascular endothelial growth factor receptor-2 (VEGFR-2) in cancer settings (Pochon et al. 2010). Microbubbles have received regulatory approval in most advanced countries for a range of car­diovascular, abdominal imaging, and molecular imaging applications. In particular, in the USA, microbubbles have been approved for better delineating the chamber boundaries in the heart and, in April 2016, were FDA approved for the character­ization of liver lesions (FDA 2016). Initial clinical trials have been completed on VEGFR-2 targeted microbubbles in the USA and Europe and it seems probable that
10 Therapeutic IVUS and Contrast Imaging 239
https://t.me/medicina_free
molecular targeted microbubbles will achieve regulatory approval for clinical usage in coming years.
A somewhat similar challenge exists for contrast agents as for IVUS from a clinical translation perspective: Both IVUS and contrast agents take a previously non-invasive imaging modality and make it minimally invasive. This inevitably adds cost, very moderate level of pain and a small risk (primarily of infection) associated with the required needle-based veinous access. As in the case of IVUS versus tran­scutaneous ultrasound, the penalty involved in migrating into an invasive modality is rewarded with expanded research and clinical information. Nevertheless, the slow rate of adoption, at least in the USA, has been disappointing to those performing research in the field. However, IVUS benefits via the fact that it is minimally inva­sive from the outset so the addition of microbubbles creates only marginal increases to cost and risk.
The high echo signal obtained from microbubbles is a consequence of their very low density and high compressibility relative to adjacent tissues and structures. How­ever, because other targets in an ultrasound field can also give rise to a strong echo signal (e.g., vessel/blood interface, etc.), it is highly desirable to exploit the unique echo signal “signature” of a microbubble. The high compressibility of a microbub­ble gives rise to a nonlinear vibrational response to impinging compressional ultra­sound wave energy. As a consequence, harmonics of the transmitted waveform are present in the echo signal. Most often the first harmonic of the fundamental (trans­mitted) signal is used in signal isolation methods. It should be noted that in the literature this harmonic signal is usually referred to as the “second harmonic” to mean the first harmonic of the transmitted signal—i.e., centered at 2f
where f
o
is the fundamental, transmitted, ultrasound center frequency. However, higher har­monics and subharmonics are sometimes employed. Given the very high frequencies typically used in IVUS, it can be observed that achieving transducer and system bandwidth to encompass both transmitted and nonlinear harmonic signal presents an obvious technical challenge. A number of signal isolation methods have been devel­oped to isolate the nonlinear microbubble signal and thereby isolate it from adjacent tissue signal. The most common approaches involve multiple pulse excitation along a common acoustic beamline (Shen et al. 2005). These methods include “ampli­tude modulation” (AM) (Brock-Fisher et al. 1996), “pulse inversion” (PI) (Chapman and Lazenby 1997), and “contrast pulse sequences” (CPS) (Phillips 2001; Phillips and Gardner 2004). These methods all involve the use of the observation that serial signals acquired of an identical, stationary, target behave as a linear system when all components of the pulse-echo system are linear. More simply stated, in a linear system, the echo signal is a scaled replica according to the amplitude/phase of t he excitation (transmitted) signal. Thus, these acquired linear signals can be scaled and added, or subtracted, in order that linear origin signals cancel. In the case of ampli­tude modulation, a scaled replica of an initial pulse is used (e.g., “1,” “½”). The two received signals have a compensatory scaling applied to the, for example, smaller of the two signals such that when subtract, perfect cancelation occurs (i.e., in this case just described, the second pulse “½” is scaled by a factor of 2 resulting in a “1”). In the case of pulse inversion, the two transmitted signals are inverted replicas of each
o
240 J. A. Hossack
https://t.me/medicina_free
other (i.e., “1,” “−1”). Upon summing the received signals, the linear origin signals cancel. “Contrast Pulse Sequences” involves a combination of both AM and PI (e.g., ½, −1, ½). In t his CPS example, the three received signals can be summed, without scaling, to achieve cancelation of the linear origin signal. Notice that in all of these three methods, any nonlinear origin signal will not perfectly cancel. It should also be noticed that these nonlinear signal separation methods are sensitive, to varying degrees, to both subharmonics and superharmonics—not just the second harmonic. It is also possible to isolate nonlinear signals using conventional frequency filtering. Additionally, cancelation techniques are frequently combined with frequency filter­ing (Shen et al. 2005). It may be observed that these multi-pulse methods, especially the longer pulse methods (e.g., CPS), are susceptible to imperfect fundamental signal cancelation whenever there is tissue motion between successive pulses (Shen et al.
2005). In the context of IVUS, this is a concern even in the presence of the very s hort
inter-pulse intervals implicit in very short pulse-echo transit durations.
IVUS Specific Applications of Microbubbles for Imaging
Microbubbles have long been recognized in IVUS imaging as making a contribu­tion toward better delineating the boundary between the lumen and the wall of a blood vessel (Vavuranakis et al. 2005; Masuda et al. 2001). The increased contrast obtained when using microbubbles affords improved detection of in-stent neoin­tima and reduces observer variability when assessing stenosis and neointima area (Masuda et al. 2001). It is now established that plaque microvascularity and plaque vulnerability are closely correlated (Naghavi and Falk 2010;Howardetal.2015). Consequently, the use of microbubbles with IVUS has been used to assess the vasa vasorum that develops in coronary atherosclerotic lesions (Vavuranakis et al. 2005,
2008; Goertz et al. 2007a; Carlier et al. 2005) [As an aside, although IVUS is nec-
essary for assessing the vasa vasorum in the coronary arteries, a transcutaneous ultrasound approach for assessing the vasa vasorum in carotid arteries is feasible (Song and Zhang 2015)]. Microbubbles, even when using a non-optimized IVUS device, yield enough vascular contrast enhancement to enable assessment of vas­cular density (Carlier et al. 2005). It is possible to measure intra-plaque perfusion density by contrasting the gray-scale image intensity acquired before, during, and after microbubble infusion (Vavuranakis et al. 2008). Using this approach, it has been observed that increased microbubble image signal between the intima-media bound­ary and the adventitia, following infusion, was associated with increased plaque vascularity (Fig. 10.5). Thus, it i s believed that a combination of IVUS plaque mor- phological assessment (e.g., identifying cap thickness, calcification, lipid core) and vasa vasorum density imaging using microbubble-enhanced IVUS provides a pow­erful instrument and method for identification of plaques susceptible to rupture—i.e., “vulnerable plaques” (Carlier et al. 2005).
Efforts have been made to improve the sensitivity and specificity of IVUS instru­mentation for microbubble detection. In this context, the very high frequencies com-
10 Therapeutic IVUS and Contrast Imaging 241
https://t.me/medicina_free
Fig. 10.5 Depiction of qualitative representation of enhancement. Unprocessed images are dis­played a before, b during, and c after injection of microbubbles. Corresponding processed images are displayed in d–f. Enhancement is graded from minimal (blue) to maximal (red). Values are a percentage of the maximum grey level intensity difference (255). Arrows indicate points of intense, stable enhancement at the media–adventitia border. Diffuse points of enhancement are present nearby. Source Vavuranakis et al. (2008)
242 J. A. Hossack
https://t.me/medicina_free
monly used in IVUS present significant technical challenges. Super harmonics occur at very high frequencies (e.g., 40–80 MHz) and require very high signal bandwidth. It needs to be emphasized that this infers very high signal bandwidth at every “link” in the processing chain—transmitter electronics, transducer, transducer interconnecting cable, receiver electronics, and digitization. The mismatch between commonly used IVUS frequencies and the natural resonant frequencies of commonly used microbub­bles (<5 MHz typically), has motivated the consideration of subharmonic detection. These challenges have motivated development of dual-frequency IVUS transducers and methods to suppress echoes originating from tissue and preserve the nonlinear signal originating from microbubbles (Goertz et al. 2007b). IVUS imaging using sub­harmonic and second-harmonic signals have been evaluated in vessel phantoms and have yielded up to 15 dB of microbubble-specific contrast enhancement (Goertz et al.
2006a, 2007b). These approaches have also been in tested a fully developed rabbit
aortic plaque (in vivo) (Goertz et al. 2006a, b). Harmonic imaging provided a suc­cessful means of imaging the structure of the vasa vasorum with an ~8 dB increase in sensitivity relativeto standard, fundamental frequency-based, IVUS imaging (Goertz et al. 2006a, b) (Fig. 10.6). These early successes provided evidence that specialized microbubble imaging is feasible at IVUS imaging frequencies and motivated further development of newer techniques that improve microbubble-specific detection in t he vasculature. These new methods include radial modulation (Yu et al. 2014), superhar­monic (Ma et al. 2014), ultraharmonic imaging (Maresca et al. 2014), and chirp-based imaging (Shekhar and Doyley 2013; Shekhar et al. 2016). For example, Maresca et al. recently demonstrated the use of conventional band-limited IVUS transducers with chirp transmit sequences for ultraharmonic contrast-enhanced imaging of microvas­culature not readily visualized by conventional IVUS imaging (Maresca et al. 2013,
2014) (Fig. 10.7). The data provided by contrast-enhanced IVUS can also be used
to extend the quality and accuracy of existing methods for characterizing the vessel wall. For example, Huntzicker et al. (Huntzicker et al. 2016) have recently demon­strated that in model-based IVUS elastography, the additional information gained in relation to the geometry of detected microvessels can be used to improve the accuracy of assessments of stress distribution in the vicinity of the fibrous cap on a vulnerable plaque.
Although there has been considerable success using existing commercial IVUS transducers with conventional levels of signal bandwidth (<80% −6 dB pulse-echo), there exists a high level of interest in creating transducers true multiple frequency capability. Although 50% −6 dB bandwidth (i.e., approximating a two-cycle pulse) is sufficient to achieving λ axial resolution—and lateral resolution is only able to approach λ resolution in near ideal focusing conditions (i.e., f/number ~1, at focus), higher bandwidth has immediate value if attempting to assess sub or super harmonic signals. Additionally, the higher bandwidth will improve any signal analysis depen­dent on spectral qualities of the received signal. For example, “virtual histology,” using radiofrequency analysis (Nair et al. 2002, 2007) has encountered significant limitations in terms of reported accuracy (Thim et al. 2000) and these are probably attributable at least in part to limited signal bandwidth. Higher transducer band­width can be achieved using high coupling coefficient single crystal materials, more
10 Therapeutic IVUS and Contrast Imaging 243
https://t.me/medicina_free
Fig. 10.6 In vivo contrast imaging results in an atherosclerotic rabbit aorta. a Fundamental mode before agent injection, where “C” is the catheter and “VC” is the vena cava. b Fundamental mode 10 seconds after injection where changes in adventitial enhancement are not evident, except for in a region at 4 o’clock and within the vena cava. c Harmonic mode before injection shows the tissue signals to be largely suppressed. d At 10 seconds after injection, the harmonic mode shows significant adventitial enhancement, consistent with the detection of adventitial microvessels. Scale of images is 12 mm across. The dynamic range of the fundamental and harmonic images are 40 and 25 dB, respectively. Source Goertz et al. (2006b)
complex matching layer schemes or cMUT technology. However, in the IVUS field, the small dimensions (physical fabrication difficulties, electrical matching challenges, cable loss effects, etc.) present more extreme practical challenges to the transducer designer. It also worth reiterating that due to the importance of tis­sue/contrast signal arising immediately adjacent to the transducer, using an offset transmitter/receiver approach is not a viable option. In one approach, multiple pas­sive layers are used or provide 20 and 40 MHz operation from a single element transducer (Vos et al. 2005). Stacked, high- and low-frequency element-based, trans­ducer designs have been proposed. Generally, the two components of the stack will
244 J. A. Hossack
https://t.me/medicina_free
Fig. 10.7 Images acquired in the tissue-mimicking phantom using the chirp reversal sequence and the pulse inversion sequence. a IVUS chirp image without contrast. b IVUS chirp image with contrast. c IVUS chirp reversal contrast image. d 22 MHz conventional IVUS image without contrast. e 22 MHz conventional IVUS image with contrast. f Ultraharmonic IVUS contrast image. Intensity scales in contrast modes are displayed relative to their respective B-mode intensity. In both the chirp reversal (c) and ultraharmonic (f) contrast modes, small model vessels (0.5 mm vessel to left) can be identified above background. In the ultraharmonic example, (f), a 0.2 mm vessel to bottom is readily visible. Source Maresca et al. (2013)
mechanically cross couple making for a challenge in obtaining a “clean” spectral per­formance at each of the two desired center frequencies. In one design, low-frequency (35 MHz) and high-frequency (90–150 MHz) transducer elements were stacked to provide for conventional resolution and imaging depth performance at 35 MHz and very fine, but shallow, imaging performance at the higher frequency (Ma et al. 2015). In a second stacked design, a low-frequency transducer element (6.5 MHz) was placed under a high-frequency transducer element (30 MHz) (Fig. 10.8) (Ma et al.
2014). The lower-frequency transducer insonated microbubbles near their resonance
while emitted superharmonics were detected using the high-frequency element. This “transmit low, receive high” (TLRH) technique permits much high-resolution imag­ing of microbubbles than is possible when limited to imaging near the microbubble resonance frequency (Hu et al. 2010). Using this approach, an in vitro microbub­ble SNR of 12 dB has been demonstrated by Goertz et al. (2006a). More recently, the stacked element design has been extended to use lateral-mode transmission to yield yet higher versatility and performance (Wang et al. 2016). The transmit low, receive high imaging approach has been extended recently by Lindsey et al. (2016) to include adaptive (minimum variance) beamforming to achieve further refinements to imaging resolution. In vitro, combining minimum variance and phase coherence factor processing improved the spatial resolution by 41.7% while improvements of
10 Therapeutic IVUS and Contrast Imaging 245
https://t.me/medicina_free
Fig. 10.8 a Design of the dual-frequency IVUS transducer. b Prototype transducer housed on the tip of a 20-gauge hypodermic needle. Source Ma et al. (2014)
2.2 dB in contrast to tissue ratio and 37.2% in spatial resolution were achieved in vivo (Lindsey et al. 2016).
Molecular Imaging in the IVUS Field
Ultrasound molecular imaging is based on the detection of microbubbles that have adhered to a vessel wall resulting from a binding event between molecule specific
246 J. A. Hossack
https://t.me/medicina_free
ligand on the microbubble surface and a matched molecular receptor expressed on the vessel surface. Single microbubble origin signal to noise does not present a major challenge—certainly for the in vitro case (Klibanov et al. 2004). However, reliable separation of signals arising from tissue, freely circulating (non-adhered) microbub­bles and adhered microbubbles (“true” molecular signal) is challenging. The under­lying approaches used in conventional (non-IVUS) applications have been largely mapped over to use in IVUS. The same targeting and signal separation methods can be used—except that the signal bandwidths create more challenges. The potential of ultrasound molecular imaging to elucidate the molecular status of cardiovascular (and other) disease and thus enable the earlier detection, or staging, of disease has been recognized for many years now (Lindner 2004).
Early in vitro reports have proven that the signals from molecular targeted, and adhered, microbubbles possess sufficient signal to noise when using both funda­mental mode 40 MHz (Phillips et al. 2012) and when using nonlinear signal-based sensing approaches (Goertz et al. 2007a). An early (1999) in vivo IVUS molecular imaging experiment involved echogenic liposomes (ELIP) targeted to intercellular adhesion molecule-1 (ICAM-1), a Yucatan mini-swine model with atherosclerotic carotid arteries and 20 MHz IVUS (Demos et al. 1999). In 2000, Lanza et al. studied perfluorocarbon nanoemulsions for detection of tissue factor expression in swine carotid arteries. A more extensive study by Hamilton et al. (2004), also using ELIPs, reported specific detection of ICAM-1, vascular cell adhesion molecule-1 (VCAM-
1), fibrin, fibrinogen, and tissue factor in swine carotid arteries subsequent to angio­plasty balloon injury. IVUS imaging (20 MHz) indicated that all artery segments producing increased IVUS signal intensity were matched with histology consis­tent with increased expression of the target molecule. More recently, Phillips et al. (2012) presented in vitro 40 MHz IVUS imaging of VCAM-1 targeted microbub­bles (Fig. 10.9). Together, these in vivo studies suggest IVUS-based identification of the molecular-specific information may enable early characterization of endothelial injury following angioplasty balloon injury and identify atheroma components that may guide the eventual choice of therapeutic intervention.
Integrated IVUS and Microbubble Generation Devices
Usually, when IVUS and microbubbles are combined, the microbubbles are intro­duced via a separate intravenous injection or from a lumen and port on the catheter device itself. Of course, using the device to dispense microbubbles directly increases the local concentration of microbubbles allowing for greater potential for efficacy with minimal systemic-related risk. Recently, it has been proposed that microbubbles actually be formed in the catheter device itself (Dhanaliwala et al. 2013). Microbub­bles generated using a flow focusing microfluidics device havebeen tested in contrast­enhanced imaging (Hettiarachchi et al. 2007; Kaya et al. 2010; Dhanaliwala et al.
2013), molecular imaging (Talu et al. 2007; Seo et al. 2010) and therapeutic delivery