Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_4597_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
15.09.2026
Размер:
17 Мб
Скачать
☆
6.14
Airflow of fenestrated tracheostomy tubes on inspiration and expiration. (A) Cuffless fenestrated tracheostomy tube. Note inspiration around and through tube and expiration around and through tube, with boost of air up to vocal cords. (B) Cuffed fenestrated tracheostomy tube with cuff inflated. Note inspiration and expiration through (but not around) the tube. A small amount of air is directed to the vocal cords on expiration. (C) Cuffed fenestrated tracheostomy tube with cuff deflated. Note inspiration is through the tube and around the deflated cuff. A small amount of air is directed to the vocal cords on expiration.
Chapter 6 Phonation With a Tracheostomy
Several authors have reported cases of subcutaneous emphysema with the use of fenestrated tracheostomy tubes (Mozart & Stuart, 2001; Orme & Welham,
2006). Orme and Welham reported a case of massive subcutaneous emphysema shortly after the percutaneous placement of a fenestrated tracheostomy tube. Later these clinicians were directed to the manufacturer’s instructions, which directed that a fenestrated tube should not be used in the early postoperative period and that the position of the fenestration should be checked once placed. Prior to the maturation of the stoma, air can become displaced into the pretra­cheal tissues with the use of a fenestrated tube.
Talking Tracheostomy Tube (Cuff Inflated)
A talking tracheostomy is a tube that allows a ventilator-dependent patient to speak; unlike other methods, it does not require cuff deflation. This tube was designed for the ventilator-dependent patient who is at high risk for aspiration or cannot otherwise tolerate cuff deflation. An additional air supply line pro­vides an extra source of air independent of ventilation. This supply line carries air from a regulated compressed source and terminates with an opening just above the cuff. When connected to a source of compressed air or oxygen, oc­clusion of the air control port allows gas flow to be directed to the vocal cords. Figure 6.15 shows airflow with a talking tracheostomy tube.
For optimal use of the talking tracheostomy, airflow must be adjusted. Start­ing at 4 liters per minute, the flow can be increased up to 12–15 liters per min­ute until optimum voice quality is reached. However, one author (Hess, 2005) cautioned that high flow rates could be associated with a greater risk of airway injury. At these high flow rates, humidification is important to prevent drying of laryngeal tissues (Kazandjian & Dikeman, 2008). Because of the continuous independent flow of air, talking tracheostomy tubes allow speech throughout the entire respiratory cycle (Tippet & Siebens, 1995).
Tippet and Vogelman (2000) suggested that the adjustment of airflow is a delicate balance between patient comfort and clarity of speech. A flow rate be­tween 2–15 liters per minute is compatible with speech. A rate of 5 liters per minute is necessary to produce a whisper, and 8–10 liters per minute is neces­sary for speech with normal volume. Flow rates above 10 liters per minute have been reported to be uncomfortable and may necessitate warming and humidify­ing the airflow to make it more comfortable and reduce the drying effect.
Talking tracheostomy tubes have some disadvantages. Speaking is depen­dent on a caregiver and a source of compressed gas. The air supply line fre­quently becomes clogged with secretions and prevents airflow above the cuff. In this case, it is acceptable to flush the air supply line with 5 ml of sterile water (Shahvari, Kigin, & Zimmerman, 1977). A source of suction should be provided to remove the secretions from above the cuff. The accumulation of secretions is a persistent complication of speaking tracheostomy tubes, and deep subglottic suctioning above the cuff may be necessary before beginning speaking trials (Tippet & Vogelman, 2000).
The production of speech with talking tracheostomy tubes requires some trial and error in order to produce optimal voice quality. Patients need time to adjust to the new sensation of airflow and practice occlusion of the airflow port. Theoretically, patients can speak throughout the entire respiratory cycle
203
204
Tracheostomies
6.15
Illustration of airflow with talking tracheostomy tube. Reprinted with permission from “Preserving Oral Communication in Individuals With Tracheostomy and Ventilator Dependency,” by D. C. Tippet and A. A. Siebens, 1995.
(2), 55–61. Copyright 1995 by the American Speech-Language-Hearing Association. All rights
4
reserved.
American Journal of Speech Language Pathology,
with this tube; however, Leder and Traquina (1989) reported that some patients have difficulty speaking through inspiration because it feels unnatural. They recommended that tube occlusion be timed for the expiratory phase of the ventilator.
Some difficulty has been reported with positioning the tube so its openings are unobstructed. Positioning is also an issue because of the weight of the ven­tilator circuit, which can pull the tube down and away from the stoma, resulting in an air leak. The weight of the ventilator tubing can also cause the rotation of
Chapter 6 Phonation With a Tracheostomy
the tube and the blockage of air vents coming in contact with the tracheal wall (Tippet & Vogelman, 2000). A malpositioned tube leaks air around the stoma and makes loud gurgling sounds. In addition, voice quality with this type of tra­cheostomy tube may not reach more than a whisper (Hess, 2005; Shahvari et al.,
1977). For these reasons, the use of talking tracheostomy tubes with rehabilita­tion patients may be limited (Tippett & Siebens, 1995).
The speaking tracheostomy tube has also been used to provide continuous suction above the cuff in patients at risk for aspiration from a large amount of oropharyngeal secretions. When connected to suction, the removal of up to 600 ml of secretions per day has been reported (Shahvari et al., 1977).
A waiting period of 3–5 days is recommended after the initial tracheotomy procedure before using a talking tracheostomy tube. The waiting period al­lows the tract to heal and prevents air leaks through the fresh tissue planes, which could lead to subcutaneous or mediastinal emphysema (Tippett & Vogelman, 2000).
205
Nomori Ventilator Talking Tracheostomy Tube.
oped a new talking tracheostomy tube specifically for patients on mechanical ventilation. The tube has slits along the sides of the shaft and an elastic cuff that inflates and deflates in synchrony with respirations. The cuff expands on inspi­ration with positive pressure from the ventilator and deflates on expiration.
Sixteen patients with a variety of underlying conditions used the proto­type tube with excellent results. All patients except one were able to speak with strong voice quality. The one patient who was not able to vocalize had a conspicuously bent trachea from long-standing tuberculosis. Two patients with particularly weak voice quality due to COPD and spinal cord injury used the Passy-Muir valve, which improved the loudness of their voices.
One of the most notable factors from this initial clinical trial is that there were no complications from this tracheostomy tube. It never became occluded with secretions, and no cuff problems developed (Nomori, 2004). Many of us eagerly anticipate its production for clinical use, hopefully in the near future. However, Hiroaki Nomori recently reported that the new tube has not yet been approved by the Japanese Ministry of Health and Welfare (based on personal communication, December 30, 2008).
Blom Tracheostomy Tube System.
gist, developed the Blom tracheostomy tube system; it has recently been in­troduced to the market and clinical trials are in progress. Its unique design is intriguing, and initial clinical trials will determine its clinical usefulness. The tube comes in four sizes: 4, 6, 8, and 10. The fenestrated aperture on the outer cannula is located only 1 mm above the cuff. This unique placement is intended to prevent contact with the tracheal mucosa when the cuff is inflated. In addi­tion, it has a subglottic suctioning port to reduce the incidence of ventilator­associated pneumonia when connected to continuous suction.
In addition to a standard inner cannula and a subglottic suctioning can­nula, the Blom tracheostomy tube system (Figure 6.16) also has two additional cannulas that can provide speech. The speech cannula is designed to allow speech to those patients who require a fully inflated cuff, while the low-profile cannula is designed for use with patients who can tolerate cuff deflation. The
Eric Blom, a speech and language patholo-
A Japanese physician devel-
206
Tracheostomies
6.16
Blom tracheostomy tube system. Note fenestration of outer cannula just above the cuff, speech cannula, and inner cannula with subglottic suctioning port. Reprinted with permission of Pulmodyne (Indianapolis, IN).
speech cannula is very soft and floppy, fits within the outer cannula, and has two unique valves—a flap valve and a bubble valve—that function at different parts of the respiratory cycle.
Figure 6.17 shows the airflow with the Blom tracheostomy tube system. The flap valve, located at the tip of the speech cannula, opens during inhalation, allowing air to flow through the tube and into the trachea. At the same time, the bubble valve, located along the dorsal aspect of the speech cannula, ex­pands into the fenestration and seals it, preventing air from escaping into the upper airway. On exhalation, the flap valve closes and the bubble valve col­lapses, exposing the fenestration and directing the airflow through it and up to the vocal cords. Another unique feature of this tracheostomy tube is the ex­haled volume reservoir, which prevents the ventilator from recognizing false low-volume readings as air is redirected through the upper airway rather than through the ventilator. This feature is likely to be appealing to practitioners who manage ventilators, because, to date, there are no described methods to prevent low-volume alarms as exhaled air is diverted from the expiratory limb of the ventilator. The manufacturer recommends that use of laser or electrosurgery electrodes should be avoided with this tube to prevent the release of toxic gases and fire in an oxygen-rich environment.
Chapter 6 Phonation With a Tracheostomy
207
6.17
Airflow with Blom tracheostomy tube on inspiration and expiration. As air is inspired through the tube, the flap valve (located at the distal tip of the tube) opens. The flexible bubble valve (located on the dorsal shaft of the tube) also expands, covering the fenestration in the outer cannula. On exhalation, the flap valve closes, allowing airflow around it and collapsing the bubble valve to allow air through the fenestration and to the vocal cords. Reprinted with permission of Pulmodyne (Indianapolis, IN).

Summary

The many options available to restore phonation make it difficult at times to determine the best method for a particular patient. The best method is often the simplest, but numerous factors must be considered. The first factor in determin­ing the optimal choice for phonation is if the patient requires mechanical venti­lation. The second factor is whether the patient can tolerate cuff deflation. When continuous mechanical ventilation is not required, a cuffless tube with capping trials works well for the majority of patients. For patients who require intermit­tent mechanical ventilation, a TTS or CTS tube with capping trials allows the flexibility of a cuff and provides minimal resistance to airflow. Patients on con­tinuous mechanical ventilation who can tolerate cuff deflation can be managed with leak speech. Those patients who are on continuous mechanical ventilation and require cuff inflation usually provide the greatest challenge. Available op­tions are talking tracheostomy tubes and the Blom tracheostomy tube system. Fenestrated tracheostomy tubes are rarely used and must be fit individually to ensure the central placement of the fenestration within the airway.
208
Tracheostomies
Health professionals must recognize the importance of a patient’s voice to his or her well-being and for effective communication with the health care team. The restoration of a patient’s voice is often an emotional moment and can represent the beginning of a return to normal for the patient. It requires patience and an individual, methodical approach to determine the best strategy for each individual patient.

Key Points

Speaking valves are ideally used with a cuffless tracheostomy tube but ■ may be used with a completely deflated cuff as long as the patient’s abil­ity to breathe comfortably has been thoroughly assessed.
■
For patients who do not require continuous mechanical ventilation, a cuff­less tube with capping trials allows phonation for the majority of patients.
■
For patients who require intermittent mechanical ventilation, a TTS or CTS tube with capping trials allows the flexibility of a cuff and offers minimal resistance to airflow when deflated.
■
Ventilator-dependent patients who can tolerate cuff deflation can be managed with leak speech, but those who are on continuous mechanical ventilation and who require cuff inflation usually provide the greatest challenge.
References
Bard, M. C., Slavit, D. H., McCaffrey, T. V., & Lipton, R. J. (1992). Noninvasive technique for es-
timating subglottic pressure and laryngeal efficiency. Annals of Otology, Rhinology, and
Laryngology, 101(7), 578–582. Dettelbach, M. A., Gross, R. D., Mahlmann, J., & Eibling, D. E. (1995). Effect of the Passy-Muir
valve on aspiration in patients with tracheostomy. Head & Neck, 17(4), 297–302. Elpern, E. H., Okonek, M. B., Bacon, M., Gerstung, C., & Skrzynski, M. (2000). Effect of the
Passy-Muir tracheostomy speaking valve on pulmonary aspiration in adults. Heart and
Lung, 29(4), 287–293. Fitsimones, L. (2003). Tracheostomy and ventilator speaking valves: Benefits beyond speech.
Vital Signs. Retrieved October 27, 2009, from http://www.passy-muir.com/pdfs/article-
vitalsigns.pdf Fornataro-Clerici, L., & Zajac, D. J. (1993). Aerodynamic characteristics of tracheostomy speak-
ing valves. Journal of Speech and Hearing Research, 36, 529–532. Gross, R. D., Atwood, C. W., Grayhack, J. P., & Shaiman, S. (2003). Lung volume effects on pha-
ryngeal swallow physiology. Journal of Applied Physiology, 95, 2211–2217. Hess, D. R. (2005). Facilitating speech in the patient with a tracheostomy. Respiratory Care,
50(4), 519–525. Hoit, J. D., & Banzett, R. B. (1997). Simple adjustments can improve ventilator-supported speech.
American Journal of Speech Lang Pathology, 6, 87–96. Hoit, J. D., Banzett, R. B., Lohmeier, H. L., Hixon, T. J., & Brown, R. (2003). Clinical ventilator
adjustments that improve speech. Chest, 124, 1512–1521. Hoit, J. D., Shea, S. A., & Banzett, R. B. (1994). Speech production during mechanical ventilation
in tracheostomized individuals. Journal of Speech and Hearing Research, 37, 53–63. Holmberg, E. B., Hillman, R. E., & Perkell, J. S. (1988). Glottal airflow and transglottal air pres-
sure measurements for male and female speakers in soft, normal, and loud voice. Journal
of the Acoustical Society of America, 84(2), 511–529. Kazandjian, M. S., & Dikeman, K. J. (2008). Communication options for tracheostomy and
ventilator-dependent patients. In E. N. Myers & J. T. Johnson (Eds.), Tracheostomy: Airway
management, communication, and swallowing. San Diego, CA: Plural Publishing.
Chapter 6 Phonation With a Tracheostomy
Leder, S. B. (1990). Importance of verbal communication for the ventilator-dependent patient.
Chest, 96(4), 792–3. Leder, S. B., & Ross, D. A. (2000). Investigation of the causal relationship between tracheostomy
and aspiration in the acute care setting. Laryngoscope, 110, 641–644. Leder, S. B., & Traquina, D. N. (1989). Voice intensity of patients using a Communi-Trach I
cuffed speaking tracheostomy tube. Laryngoscope, 99, 269–274. Lichtman, S. W., Birnbaum, I. L., Sanfilippo, M. R., Pellicone, J. T., Damon, W. J., & King, M. L.
(1995). Effect of a tracheostomy speaking valve on secretions, arterial oxygenation,
and olfaction: A quantitative evaluation. Journal of Speech and Hearing Research, 38(3),
549–555. MacBean, N., Ward, E., Murdoch, B., Cahill, L., Solley, M., Geraghty, T., et al. (2008, September).
Optimizing speech production in the ventilator-assisted individual following cervical spi-
nal cord injury: A preliminary investigation. International Journal of Language and Com-
munication Disorders, 1–12. Manzano, J. L., Lubillo, S., Henriquez, D., Martin, J. C., Perez, M. C., & Wilson, D. J. (1993). Verbal
communication of ventilator-dependent patients. Critical Care Medicine, 21(4), 512–517. Mozart, M. J., & Stuart, H. (2001). Subcutaneous emphysema caused by a fenestrated tracheos-
tomy tube. Anaesthesia, 56(2), 191–192. Nomori, H. (2004). Tracheostomy tube enabling speech during mechanical ventilation. Chest,
125(3), 1046–1051. Orme, R. M., & Welham, K. L. (2006). Subcutaneous emphysema after percutaneous tracheos-
tomy—time to dispense with fenestrated tubes? Anaesthesia, 61(9), 911–912. Passy, V., Baydur, A., Prentice, W., & Darnell-Neal, R. (1993). Passy-Muir tracheostomy speaking
valve on ventilator-dependent patients. Laryngoscope, 102(6), 653–658. Saul, A., & Bergstrom, B. (1979). A new permanent tracheostomy tube-speech valve system.
Laryngoscope, 89(6, Pt. 1), 980–983. Shahvari, M.B.G., Kigin, C. M., & Zimmerman, J. E. (1977). Speaking tracheostomy tube modi-
fied for swallowing dysfunction and chronic aspiration. Anesthesiology, 46(4), 290–291. Shea, S. A., Hoit, J. D., & Banzett, R. B. (1998). Competition between gas exchange and speech
production in ventilated patients. Biological Psychology, 49, 9–27. Shikani, A. H., French, J., & Sievens, A. A. (2000). New unidirectional airflow ball tracheostomy
speaking valve. Otolaryngology—Head and Neck Surgery, 123(1), 103–107. Suiter, D. M., McCullough, G. H., & Powell, P. W. (2003). Effects of cuff deflation and one-way
tracheostomy speaking valve placement on swallow physiology. Dysphagia, 18, 284–292. Tippett, D. C., & Siebens, A. A. (1995). Preserving oral communication in individuals with tra-
cheostomy and ventilator dependency. American Journal of Speech and Language Pathol-
ogy, 4(2), 55–61. Tippett, D. C., & Vogelman, L. (2000). Communication, tracheostomy and ventilator dependency.
In D. C. Tippett (Ed.), Tracheostomy and ventilator dependency. New York: Thieme. Toremalm, N. G. (1968). A tracheotomy speech valve. Laryngoscope, 78(12), 2177–2182. Zajac, D. J., Fornataro-Clerici, L., & Roop, T. A. (1999). Aerodynamic characteristics of speaking
valves: An updated report. Journal of Speech, Language, and Hearing Research, 42, 92–100.
209
This page intentionally left blank
Care of the Tracheostomy Patient
Linda L. Morris
7
The tracheostomy patient requires vigilant care, especially in the first several postoperative days. Meticulous care and maintenance of the tracheostomy tube can prevent many complications. This chapter discusses the care required for the patient with a tracheostomy, including maintaining the tube, cleaning the inner cannula and stoma, caring for the cuff, mobilizing secretions, and caring for other tracheostomy appliances such as the T-tube. In addition, consider­ations for patients at home will also be reviewed, including managing home emergencies.

Maintenance of the Tracheostomy Tube

General care of the tracheostomy patient includes ensuring the presence of ev­eryday and emergency supplies at all times, as well as complete familiarity with care of the tube. Bedside supplies, listed in Table 7.1, should include an extra tracheostomy tube of the same size and one size smaller, an obturator of appro­priate size and type, a suction source and appropriately sized suction catheters, a
211