Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_4597_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
15.09.2026
Размер:
17 Мб
Скачать
☆
Tracheostomies
192
6.6
Montgomery Ventrach (left) and Montgomery valve (right). Photo courtesy of Boston Medical Products. All rights reserved.
6.7
Passy-Muir valves. Clockwise from far left: PMV PMV 007, PMV PMV PMA courtesy of Passy­Muir.
TM
2000,
TM
005, PMVTM
TM
2020-S,
TM
2020, and
TM
2000. Photo
closure of the valve, which is accompanied by a noticeable click. This valve was discontinued in 2000 (based on personal communication, Rachel Proctor, Olympic Medical, May 21, 2009). The Shiley Phonate speaking valve (Figure 6.8) is bias-closed with a silicone diaphragm. It has a hinged cap that makes the diaphragm available for cleaning. One model has a port to attach to an oxygen source; however, this model was not included in the study. The Montgomery Ventrach has a bias-open silicone diaphragm and can be used with ventilator­dependent patients. The Passy-Muir PMV
TM
PMV dependent patients. The PMV ing, and the others are used with rubber nondisposable tubing. All the Passy­Muir valves are bias closed (Zajac et al., 1999). The PMV version of the PMV without 15-mm adapters (Fitsimones, 2003).
the 1999 study were higher than in the earlier work because they were associ­ated with patients on mechanical ventilation. Except for minor differences, all the valves exhibited similar low resistance—even those with a bias-closed de­sign. There were some differences in manufacturing specifications between the
2000 (clear), and PMVTM 2001 (purple) can also be used for ventilator-
TM
2000, was designed for use with metal tracheostomy tubes
Each of these valves was tested using different flow rates. The flow rates in
TM
007 can be used with disposable ventilator tub-
TM
005 (white), PMVTM 007 (aqua),
TM
2020, a modified
Chapter 6 Phonation With a Tracheostomy
6.8
Shiley Phonate with oxygen port (SSVO). Covidien.
6.9
Eliachar speaking valve. Hood Labs (Pembroke, MA).
193
1993 and 1999 studies. The researchers measured the thickness of the PMVTM 007 and found a thinner diaphragm than in the earlier study. This could explain the finding of reduced resistance; however, they were unable to confirm a de­sign change with the company (Zajac et al., 1999).
These six valves were also tested in vivo with the use of a mouthpiece; patients repeated the sound /pa/, and the researchers tested for leakage of air through the valve. They found that both the Olympic Trach Talk and the Mont­gomery Ventrach showed significant air loss during expiration when the valves should have been closed (Zajac et al., 1999).
The Eliachar speaking valve (Figure 6.9) is made of silicone and is designed for use with the Hood stoma stent. It has a low-profile appearance and flap design that provide low resistance. The Hood speaking valve (Figure 6.10) and the TRACOE Phon Assist are both flap valves. The Phon Assist has an oxygen supply port, but its most unique feature is an adjustable airflow mechanism that patients can use to dial up their desired level of airflow.
The Shikani-French valve uses a different mechanism than the other dia­phragm valves—a unidirectional flow ball valve. It has lower resistance, a lower profile, and is more easily hidden under clothing than the other valves (Shi­kani, French, & Sievens, 2000). Originally called the Hopkins valve, the Shikani­French valve has been adapted for use with plastic as well as metal tracheostomy tubes. Figure 6.11 shows the mechanism of the Shikani-French ball valve. In the
6.10
Hood speaking valve. Hood Labs (Pembroke, MA).
6.11
Mechanism of Shikani­French valve. Reprinted with permission from “Preserving Oral Communication in Individuals With Tracheostomy and Ventilator Dependency,” by D. C. Tippett and A. A. Siebens, 1995.
American Journal of Speech Language Pathology, 4
Copyright 1995 by the American Speech­Language-Hearing Association. All rights reserved.
(2), 55–61.
Chapter 6 Phonation With a Tracheostomy
resting state, the ball lies on the wall of the air chamber (A). On inspiration, the ball moves toward the opening of the cannula, allowing air to enter the cannula (B). Airflow pushes the ball back, but it is prevented from entering the can­nula by a wire stop. On expiration (C), the ball seals the port, forcing air into the upper airway. Because of this design, less pressure is required for valve closure, which is almost instantaneous. This is especially important in patients with very low tidal volumes and vital capacities. The valve is more compact than the other valves and can be used in the standard versions of all commercially available tracheostomy tubes. The valve can also replace the inner cannulas of these tubes and provide lower resistance (Shikani et al.).
One study compared the Shikani-French ball valve (named Hopkins at the time of the study) with the Passy-Muir valve and found less inspiratory resis­tance and easier cleaning with the ball valve (Tippett & Vogelman, 2000).
The Tucker valve uses a different mechanism. The Tucker tracheostomy tube is composed of metal with a fenestrated flap on the inner cannula that opens on inhalation and closes on exhalation. This action forces air up through the upper airway (Tippett & Vogelman, 2000). Airflow during inspiration rotates the flap to an open position and then to a closed position during expiration (Tippett & Siebens, 1995).
195
The Clinical Use of Speaking Valves.
speaking valves are externally placed on the 15-mm or low-profile connector of a cuffless or deflated-cuff tracheostomy tube. Patients and their caregiv­ers require education to effectively use the valve and to alleviate any anxiety about shortness of breath. They also need to know how to remove the valve in the event of acute dyspnea. Patients should not be left unsupervised with the valve in place until their ability to tolerate it has been established (Tippett & Vogelman, 2000).
Initially, a schedule for using the speaking valve should be structured to tar­get continuous wearing during daytime hours. The manufacturer recommends that the Passy-Muir valve not be used while sleeping; however, one study es­tablished its potential usefulness for one night in seriously ill tracheostomy patients without any untoward effects such as apnea, desaturation, respiratory distress, or cardiac arrhythmias. Accordingly, recommendations for testing these valves for longer night duration were concluded (Gross, Atwood, Grayhack, & Shaiman, 2003).
Studies have shown that speaking valves provide more benefits than merely verbal communication. Lichtman and colleagues (1995) studied the potential secondary benefits of speaking valves. They found that patients accumulated significantly fewer secretions with speaking valves than without. Other re­ported benefits include more effective swallowing, more effective coughing, a restoration of intrinsic PEEP, an enhanced Valsalva maneuver, and the preven­tion of aspiration.
Several studies have examined the effect of speaking valves on aspiration. Dettelbach, Gross, Mahlmann, and Eibling (1995) studied 11 patients with tra­cheostomies and clinical evidence of aspiration. They found that aspiration was reduced or eliminated in all patients when a speaking valve was in place. Suiter and others (2003) studied swallow physiology as well as the incidence of as­piration with tracheostomized nonventilator-dependent patients under three
With the exception of the Tucker valve,
Tracheostomies
196
different conditions: cuff inflated, cuff deflated/cuffless tube, and with a Passy­Muir speaking valve. They found that cuff deflation significantly improved swallowing ability by increasing both pharyngeal transit duration as well as duration of hyoid excursion. In addition, they found that cuff status had no ef­fect on aspiration. However, there was significantly less incidence of aspiration with the speaking valve when compared to the cuff-inflation and cuff-deflation conditions. Eight out of 10 patients who aspirated thin liquids with cuff infla­tion or deflation were able to safely take liquids with the valve in place. The authors concluded that the speaking valve might restore subglottal air pressure lost with use of a tracheostomy tube and that patients who are unable to toler­ate thin liquids may be able to ingest them safely with the valve in place.
Elpern, Okonek, Bacon, Gerstung, and Skrzynski (2000) also studied the effect of the Passy-Muir valve on aspiration and obtained different results. In their study of 15 patients with tracheostomies and evidence of aspiration, 7 pa­tients aspirated thin liquids, 5 patients aspirated only when the valve was off, and 2 patients aspirated with and without the valve. They found that aspiration was less frequent with the valve in place and concluded that a speaking valve can reduce, but not eliminate, aspiration in patients with a tracheostomy.
Manzano and colleagues (1993) found that the use of a speaking valve de­creased the amount of secretions, improved the ability to cough, reduced the need for tracheal suctioning, and improved the sense of smell in patients. Al­though they did not specifically study patient mood, they also found that pa­tients appeared to be happier and interacted better with their families and nursing staff.
Contraindications for Speaking Valves.
benefits in addition to speech, not all patients can tolerate them. Patients with poor pulmonary reserves can develop hypercarbia. Because some degree of air trapping can occur with speaking valves, patients with severe lung disease may not be able to fully exhale (Kazandjian & Dikeman, 2008). The contraindications for the use of a speaking valve are listed as follows:
■
Inability to tolerate cuff deflation (risk of aspiration) Hemodynamic instability ■ Severe pulmonary disease ■ Total laryngectomy ■ Vocal cord paralysis ■
While speaking valves provide many

Patients Who Require Intermittent Positive-Pressure Ventilation

It is clear that patients who require continuous positive-pressure ventilation need a cuffed tube; however, patients who use intermittent ventilation require cuff inflation only during that time. Intermittent positive-pressure ventilation is applied to patients during weaning from mechanical ventilation, for nocturnal ventilation, and for bronchial hygiene. As patients are weaned from continuous mechanical ventilation, they spend increasing amounts of time free from the
Chapter 6 Phonation With a Tracheostomy
ventilator at which time the cuff can be deflated. During the weaning process, altering the tracheostomy tube can accommodate the patient’s changing goals.
When patients do not require continuous cuff inflation, it is usually rela­tively easy to restore phonation when they are without the ventilator. This can be done by simply deflating the cuff and occluding the tube to allow air up to the vocal cords, while assessing for the adequate movement of air around the deflated cuff. However, cuffed tubes should never be capped. Thus, these pa­tients will require a different type of tracheostomy tube.
The Bivona TTS and the Arcadia CTS are ideal choices to accommodate a patient who requires intermittent cuff inflation. In contrast to standard low­pressure, high-volume cuffs, these tubes have low-volume, high-pressure cuffs. When inflated, they can exert high pressures against the tracheal wall, so care­ful attention to minimizing cuff pressure should guide practice. The significant advantage of this type of tube is its deflation characteristics, which allow it to lie flush against the shaft of the tube when deflated, creating no obstruction to airflow outside the tube. When deflated, the TTS and CTS tubes function as cuffless tubes, and they are the only cuffed tubes currently on the market that can be capped safely.
A TTS or CTS tube can be capped while the patient is free from the ventila­tor, and the cuff can be inflated when returning to the ventilator. This technique often speeds up the weaning process and frequently marks the beginning of freedom from the ventilator.
When inflating the cuff of the TTS or CTS tubes, the clinician must take care to ensure minimal pressure in the cuff. Because of the inflation character­istics of the high-pressure cuff, direct cuff pressure measurement will result in a high reading; thus, the minimal leak technique is recommended. In order to distribute pressure evenly and prevent a loss of cuff volume from the diffusion of gas, the cuff should be inflated with sterile water (not saline).
A CTS or TTS tube is also ideal for patients who require intermittent cuff inflation because of bronchial hygiene or administration of inhaled medi­cations. The cuff can be inflated during these maneuvers and deflated after completion.
In addition to concerns with the high-pressure cuff, the primary disadvan­tage of the TTS and CTS tubes is that they are single-cannula tubes. Patients who have large volumes of thick secretions may be at particular risk for tube obstruction because these tubes lack an inner cannula. In this case, measures should be taken to ensure optimal mobilization of secretions, or it may be nec­essary to change to a tube with an inner cannula.
197

Patients Who Require Continuous Mechanical Ventilation

Unless there are extenuating circumstances, phonation should not be attempted in patients who are hemodynamically unstable and on mechanical ventilation. In most cases, it is best to wait until patients are more stable and can tolerate cuff deflation because most options for phonation require cuff deflation or tube change.
There is competition between airflow requirements for speech and gas ex­change in the ventilator-dependent tracheostomy patient (Shea, Hoit, & Banzett,
Tracheostomies
198
1998). Normally, patients speak during expiration using air that has already been used for gas exchange. Competition occurs when patients “steal” air from al­veolar ventilation during the inhalation phase, producing sound by diverting air around the tracheostomy tube and up to the larynx. Speech during inhalation produces relative hypoventilation.
Shea and others (1998) asked five patients with long-term tracheostomies to perform two speaking tasks. The patients read a standard text and spoke extemporaneously while the investigators measured speech ventilation vol­umes. All the subjects had either a fenestrated tracheostomy tube or a deflated cuff. They performed both tasks under normal conditions with 100% oxygen and under experimentally induced hypercapnic conditions using a mixture of carbon dioxide and oxygen. Normal individuals at rest will hyperventilate dur­ing speaking. The investigators noted an average loss of 14% of alveolar ven­tilation when speaking. During induced hypercapnia, which caused shortness of breath, all the subjects could still speak adequately. Two patients adapted to hypercapnia by reducing the amount of air used for speaking. One patient increased the airflow used per syllable during hypocapnic speech, which is a maladaptive strategy. The investigators concluded that there is complete an­tagonism between the airflow requirements for speaking and gas exchange during the ventilator’s inspiratory phase. In their experiment, the behav­ioral drive to speak was modified but never fully suppressed by increasing respiratory drive.
Leak Speech (Cuff Deflated)
Leak speech involves creating a leak in the seal between the cuff and the tra­chea so that some air can be diverted up to the vocal cords. The ventilator provides the power behind the voice, and ventilator settings are adjusted to accommodate the leak. Leak speech is possible only when the patient uses a deflated cuff, a cuffless tube, or a fenestrated tracheostomy (MacBean et al.,
2008). In addition, leak speech is possible only when airflow is sufficient to generate adequate subglottal pressure. This pressure is usually limited to the inspiratory cycle and the beginning of the expiratory cycle. For this reason, leak speech is characterized by bursts of short phrases and long pauses (Hoit et al., 2003; MacBean et al.).
Leak speech is beneficial because it improves patient quality of life by re­storing communication, and the required cuff deflation reduces pressure against the trachea. However, it also involves trial and error until the optimal settings are achieved. Optimal leak speech requires ventilator adjustments to produce the best voice quality without discomfort or shortness of breath.
Inspiratory speech is something of a paradox because normal speech com­mences during expiration. Competition of tidal volume use between speech and ventilation varies with glottic resistance, which can be consciously controlled by the patient. When the glottis is narrowed, more tidal volume is delivered to the lungs; when there is little laryngeal resistance, less tidal volume is delivered to the lungs (Tippet & Vogelman, 2000).
Typically, the first step toward leak speech is to assess the size of the trache­ostomy tube with the cuff deflated. One must ensure the patient can move enough air comfortably around the deflated cuff. If the patient is unable to
Chapter 6 Phonation With a Tracheostomy
move enough air, it may be necessary to downsize the tracheostomy tube. Fol­lowing cuff deflation, the tidal volume should be increased up to 50% more than for resting ventilation, and the inspiratory time can also be increased to per­mit comfortable and longer speech. During initial trials, the patient will require some coaching to speak on inspiration when the ventilator breath is delivered. Inspiratory speech often requires some practice because it is the opposite of normal.
Some patients report anxiety or discomfort with leak speech because of the flow of air through the upper airway; this is especially common in those who have not experienced this sensation for a long time. This is often resolved by coaching the patient to time their voicing with inspiration. It may also help to instruct patients to push down, hold their breath, or tighten their throat. Patients can be taught to widen or narrow their glottic aperture in order to increase or decrease the volume delivered to the lungs (Tippet & Vogelman, 2000).
If patients become breathless during a leak speech trial, tidal volume or flow rate can be increased. Another strategy is to add PEEP. The addition of PEEP has several advantages, including lengthening the duration of tracheal pressure for voicing into the exhalation phase. Adding PEEP can allow patients to speak through 80% of the respiratory cycle (Hess, 2005; Hoit et al., 2003; Hoit & Banzett, 1997; Hoit, Shea, & Banzett, 1994; MacBean et al., 2008; Shea et al., 1998). The addition of 4–15 cm H produce higher quality speech, and allow more syllables per ventilator cycle. Patients using leak speech with PEEP have been observed to speak through the entire ventilator cycle without any pauses for breathing (Hess).
O of PEEP can extend speech duration,
2
199
Leak Speech With a Speaking Valve.
speaking valve rather than PEEP. Only three speaking valves are approved by the FDA for use with ventilator-dependent patients—the Montgomery Ventrach and Passy-Muir models PMV
Speaking valves allow the ventilator to deliver a breath through the tra­cheostomy tube while the exhaled air is redirected around the tube—and so, it is essential that the cuff be completely deflated. Manzano and others (1993) used the Passy-Muir valve in their series of ventilator-dependent patients to determine its usefulness, advantages, and disadvantages during mechanical ventilation. Because cuff deflation is vital to the use of the speaking valve, they recommended that a cuffless tracheostomy tube be used while the valve is in place. The presence of a deflated cuff makes aspiration possible, and because of this, it is necessary to closely monitor tidal volume peak inspiratory pres­sure, thoracic movements, the elimination of secretions, and continuous airflow through the mouth and nose.
Figure 6.12 shows the placement of the valve within the ventilator circuit.
The speaking valve can be applied one of two ways. For patients with a closed suction system, the valve should be applied to the side port to allow the suction catheter to pass easily into the tracheostomy tube (Hess, 2005). For patients who are not using a closed suction system, the valve can be applied directly between the ventilator tubing and the 15-mm connector. An HME is not functional with a speaking valve because no exhaled gas will pass through when it is in place. Some patients report discomfort when initially using a speaking valve due to drying of the pharyngeal membranes or decreased tone
TM
005 and PMVTM 007 (MacBean et al., 2008).
Leak speech can also be used with a
200
Tracheostomies
6.12
Proper placement of speaking valve within ventilator circuit. Valve is positioned between the tracheostomy tube and ventilator circuit to allow intake of air into the tube.
from weakness and lack of airflow through the upper airway. This effect can be mitigated by slow cuff deflation over several minutes (Hess).
In a study of 15 ventilator-dependent patients, Passy, Baydur, Prentice, and Darnell-Neal (1993) found that all the subjects using the Passy-Muir valve showed marked improvement in speech intelligibility, speech flow, elimination of speech hesitancy, and speech time. MacBean and colleagues (2008) studied the use of leak speech in two patients with cervical spinal cord injuries. They found that the use of PEEP in addition to a speaking valve provided the best quality speech.
Figure 6.13 shows airflow when a speaking valve is used with a ventilator. With a speaking valve, air will be redirected through the upper airway upon ex­halation and not returned to the ventilator. Therefore, the exhaled tidal volume alarm must be adjusted accordingly. In addition, the tidal volume and high­pressure alarms must be carefully monitored to avoid excessive air trapping (Kazandjian & Dikeman, 2008).
Most critical care ventilators will not easily tolerate a speaking valve; how­ever, the Puritan Bennett 760 has a speaking valve mode (Hess, 2005). The Respironics Esprit and the new Respironics V200 ventilator also have speak­ing modes in which the exhalation valve is closed, allowing gas flow past the vocal cords. In speaking modes, exhaled gas pressure is measured to evaluate whether the tube has become disconnected or obstructed.
Chapter 6 Phonation With a Tracheostomy
6.13
Airflow with valve and ventilator circuit. (A) Normal ventilator circuit. Note inflation of the cuff with inflow of air from the ventilator and exhaled air through exhalation valve. (B) Ventilator circuit with speaking valve. Note deflation of the cuff, with all exhaled air passing through the upper airway and bypassing the ventilator.
201
Cuffed Fenestrated Tracheostomy Tube (Cuff Inflated)
Patients who are ventilator dependent will usually require a cuffed tracheos­tomy tube. In order for these patients to phonate, one option is a cuffed fenes­trated tracheostomy tube. The proper fit of the fenestration within the airway is vital for the proper functioning of the tube as well as preventing complications such as granulation tissue. See chapter 4 for a discussion of the proper fitting of a fenestrated tracheostomy tube.
Figure 6.14 shows airflow through fenestrated tracheostomy tubes. View A shows inspiration and expiration through a cuffless fenestrated tracheos­tomy tube. Air moves through the upper airway and through the tube. In View B, when the cuff is inflated on a cuffed fenestrated tube, inspiration and expi­ration occur primarily through the tube; however, a small amount of air moves through the upper airway. In View C, the cuff is deflated on a cuffed fenestrated tube. Inspiration is primarily through the tube; however, a small amount of air is inspired though the upper airway. On exhalation, air must pass around the deflated cuff and though the fenestration to reach the upper airway.