Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3764_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
31.08.2026
Размер:
22 Мб
Скачать
78 PART III Wires Technology
Figure 9.11 Sasuke.
OTW port. This system provides augmented sup­port and higher penetration force due to the stabil­ity of the catheter over the workhorse wire and enables use of less extreme secondary curves of the CTO wire, as the angle of approach is more co­axial, and can help prevent eccentric cap penetra­tion. Elimination of side load can restore the force that would be lost due to wire buckling compared with standard microcatheters. (2) Side branch access. Once a lesion is crossed and luminal wire position is secured, a DLMC can ensure that a secondary wire advanced through the OTW port to access side branches will follow the same luminal path, avoiding sub-intimal or abluminal wiring. This is of particular value for side branches beyond the distal cap of CTOs, or when re-wiring side branches through stents as part of bifurcation procedures. (3) Drug delivery. Treatment of no reflow with med­ications targeted at the distal coronary bed can be per­formed through the OTW port, without the risk of losing wire position.
(4) Antegrade wiring. In addition to the original application of accessing proximal caps that are angu­lated with side branches at the origin, the Re-Cross catheter (IMDS) has been specifically engineered to facilitate AWE and ADR approaches.
Sasuke
The Sasuke dual lumen microcatheter (Figure 9.11) is tapered, with a soft 4.0 mm radio-opaque tip which makes it trackable, with a 38cm hydrophilic coating and stainless-steel shaft which enhances deliverability. The OTW exit port is set back 6.5 mm from the tip, which is not truly directional, but rotation of the cath­eter is possible for closer alignment of the exit port.
NHancer Rx
The NHancer Rx (Figure 9.12) is also tapered and hydrophilic, with a radiopaque atraumatic tip. Similarly, the OTW exit port is set back 6.5 mm from the tip, and deliverability is aided by a removable stylet in the OTW port.
Figure 9.12 NHancer RX.
Figure 9.13 ReCross.
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
CHAPTER 9 Microcatheters: Characteristics and Use 79
ReCross
The ReCross microcatheter (IMDS) (Figure 9.13) is a new device which differs slightly in its construction and application. There is no monorail and it is deliv­ered OTW through the tip which connects to the white hub at the back. There are 2 exit ports set back 8 mm (blue) and 12
mm (white) positioned at opposite sides of the catheter, separated by 180 degrees. The white hub therefore allows the wire to exit at the tip, and through the side exit port 12
mm back from the tip, whereas the blue hub wire only exits from the 8 mm side port. This unique design facilitates both antegrade wiring and antegrade dissection re-entry techniques. As an antegrade tool it is a deliverable MC that allows parallel wiring whilst maintaining the original wire position and facilitating rapid wire exchange and can similarly be used for intraplaque redirection. In dissection re-entry, the 2 juxtaposed exit ports allow co-axial wire exit to direct at the true lumen. In doing so, the redundant lumen can simulta­neously facilitate STRAW.
Directional
Angled SuperCross
The Angled SuperCross microcatheters (Teleflex) (Figure 9.14) are designed to overcome sharp angula­tion, useful for wiring side branches particularly in the context of complex disease. Specific scenarios include retroflexed circumflex, diagonal branches in tortuous LAD segments, and retrograde approaches to grafted vessels with significantly distorted anatomy when the 90° and 120° catheters are most useful. A platinum/tungsten radio-opaque coil at the tip ena­bles good visibility and torque response. Once the ves­sel has been accessed it can be exchanged by balloon trapping.
Venture Rx
The Venture Rx rapid exchange microcatheter (Figure
9.15) has an adjustable tip capable of up to a 90° angle to aid wiring in tortuosity and side branch access. An 8 mm platinum tip provides good visibility,
Figure 9.14 Angled SuperCross.
80 PART III Wires Technology
Figure 9.15 Venture RX.
and multi-layered shaft construction enables good torque control and kink resistance. Manipulation of the tip deflection knob and torque handle at the proxi­mal end allows precise angle adjustment according to the task.
Acknowledgments
With thanks to Optima Education and Vascular Perspectives for reference and illustration support. A special thank you to Professor James Spratt, Dr Colm
Hanratty, Dr Simon J Walsh, Dr Simon Wilson who are mentors and authors of the antegrade iBook which I have learned from and use still on a regular basis.
IV
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
PART IV
Wires Technique
10
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
CHAPTER 10
CTO Wires: Engineering 101 and Principles of Wire Manipulation
Rahul Kurup1 & Luiz Fernando Ybarra2,*
1
South West Sydney Local Health District (Campbelltown and Liverpool Hospitals) and Sydney Medical
School University of Sydney, NSW, Australia
2
London Health Sciences Centre, Schulich School of Medicine & Dentistry, Western University, London, ON, Canada * Corresponding author
Wire engineering 101
There have been significant technological advances in the equipment used for percutaneous coronary interven­tion (PCI), and this is particularly evident in the current generation of coronary guidewires. Understanding wire technology and the construction of coronary guidewires allows for better usage of the wire and increased technical success rates. This is particularly important given that the predominant mode of procedural failure in chronic total occlusion (CTO) PCI using the contemporary “hybrid” approach has been the inability to cross the lesion with a guidewire [1]. A lack of understanding of the properties and utility of wires available can also lead to increased complications and prolonged procedural times.
Broadly speaking, the components of a coronary guidewire (Figure 10.1) include shaft/body, transition segment, and tip.
The wire shaft is arguably the least critical component of the guidewire but plays a role in transmitting operator hand movements and overall wire utility. The core, which is attached to the distal end of the shaft and extends to the tip of the guidewire, is often the stiffest part of the wire that gives stability, support, and steerability. Most cores of contemporary guidewires are made from stainless steel, nitinol (nickel-titanium alloy), or a combination of the two (i.e., stainless steel body with Nitinol distal tip seen in the Runthrough guidewire from Terumo). Nitinol, known for being a “memory metal” due to its good shape retention, is more durable, flexible, and kink resistant compared to stainless steel. Stainless steel, however, has greater torque transmission with better support and push. Another important parameter of the core that affects the performance of guidewires is the core diam-
eter. A larger diameter core improves support and torque transmission whereas smaller diameter cores are more trackable.
The tip of the guidewire is often where most varia­tion between wires exist and therefore plays the largest role in determining overall performance and clinical utility. The distal tip is covered with radiopaque spring-coils (typically platinum) to allow fluoroscopic visualization. Variations in the tip construction can include the tip core style, tip core taper, spring-coils, coatings, and covers. The clinical implications of these variations are shown in Table 10.1.
Variations in overall guidewire construction can also influence wire characteristics such as shaft support, tip stiffness, torque transmission, wire lubricity, and track­ability. Increased wire shaft support, which can be quantitatively measured, leads to the increased ability to deliver devices such as balloons and stents over the wire. Highly supportive wires, such as the Grandslam, Wiggle, and Ironman wires, can facilitate device delivery through tortuous arterial anatomy.
Tip stiffness is defined as the load in grams (gram­force), measured using an electronic scale, needed to buckle the tip of the guidewire to a fixed length (Figure 10.2). As each company has its own specific measurement methodology, the advertised tip load may not be exactly comparable between wires of dif­ferent companies. Stiffer wires can penetrate through caps but have an increased risk of exiting the vessel architecture resulting which can result in vessel dis­section, false lumen creation, and perforation.
Newer technology has also improved wire torque transmission which is the ease and efficiency of rota­tional force transmission applied by the operator at the
Chronic Total Occlusions: A Guide to Recanalization, Third Edition. Edited by Ron Waksman and Shigeru Saito. © 2024 John Wiley & Sons Ltd. Published 2024 by John Wiley & Sons Ltd.
83
84 PART IV Wires Technique
Figure 10.1 Components of a coronary guidewire.
Table 10.1 Variation in distal tip components and its clinical implications.
Component Variation Clinical implication
Tip Type 2-piece core with shaping
ribbon (wire core does not extend right to the tip) 1 piece core (core-to-tip) • Increased tip force transmission
Tip Core Taper Longer taper • Increased steering/trackability
Short taper • Easier to knuckle
Coatings Hydrophilic • Increased lubricity and trackability
Hydrophobic • Greater tactile feedback
Covers Polymer (over coils vs over
core)
• Increased shaping ability
• Easy to prolapse
• Increased safety due to softer tip
• Greater tactile feedback
• Less supportive
• Increased push force transmission
• Increased support
• Less tactile feedback
• Needs to be activated by water
• Less trackable compared to hydrophilic coatings
• Increased lubricity and improved vessel tracking (useful for negotiating tortuous lesions and finding microchannels in total occlusions)
Figure 10.2 Measurement of guidewire tip stiffness in gram-force (gf). Image reproduced from Asahi Intecc website.
proximal end of the guidewire to the distal end of the guidewire (Figure 10.3). For instance, the Asahi Intecc
company, which has been monumental in guidewire technology advancement, has incorporated novel designs into their contemporary guidewires. The Sion TECC technology, which involves the addition of Asahi’s pro­prietary ACT ONE rope coil over the twist and stainless steel core wire (Figure 10.4), facilitates more efficient torque transmission to the distal guidewire tip. Another example of improved guidewire design is the Inner Coil Technology (ICT) from Boston Scientific which involves the addition of a stainless steel inner coil to the distal tip of the stainless steel core (Figure 10.5). This improves wire torqueability, shape retention and durability.
The addition of coatings and polymer covers over guidewires influence the lubricity and trackability, which is defined as how easily a wire navigates the coro­nary artery, particularly tortuous vessels, and very tight occlusions. The coatings applied over the guidewire can
CHAPTER 10 CTO Wires: Engineering 101 and Principles of Wire Manipulation 85
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
Figure 10.3 Torque Transmission. Image reproduced from [2] Vascular Perspectives Ltd.
Figure 10.4 Asahi ACT ONE wire technology. Image modified and reproduced from Asahi Intecc website.
Figure 10.5 Inner Coil Technology (ICT) wire technology.
Image provided courtesy of Boston Scientific. © 2022 Boston Scientific Corporation or its affiliates. All rights reserved.
either be hydrophilic or hydrophobic. Hydrophilic coatings (e.g., polyethylene oxide or polyvinyl pyrrol­idone) require water to become activated and become
slippery with extremely low friction. On the other hand, hydrophobic coatings (i.e., polytetrafluoroethyl­ene (PTFE) or silicone) are inherently slippery and do not require water for activation. In addition to guide­wire tips, many wire shafts also are coated with silicone or PTFE to improve movement within catheters.
Polymer covers, which are generally variants of poly­urethane, can be added on top of hydrophilic coatings to further enhance wire lubricity. Hydrophilic, polymer­jacketed wires are often preferred to navigate subtotal occlusions, highly tortuous vessels, or difficult to access side branches. The trade-off between increasing lubricity of a wire is decreased tactile feedback (Figure 10.6) which
Figure 10.6 Guidewire lubricity vs tactile feedback.
86 PART IV Wires Technique
is the operator’s ability to feel the interaction between the wire and the vessel architecture. Therefore, hydrophilic, polymer-jacketed wires can more easily enter the subinti­mal space and cause dissections and perforations.
In addition, dedicated wires used for specific case sce­narios (i.e., externalization), are also needed.
“Workhorse” guidewires, which are used in most PCI cases, refer to low tip load guidewires that have a focus on safety. Some of the limitations of
CTO guidewires
There is a plethora of specialty coronary wires cur­rently available for commercial use and this offers a range of options for CTO operators (Table 10.2). Although it would be a luxury to have every wire available in the lab, the practical reality is that this is limited by cost, redundant wire design, and wire familiarity. Therefore, most CTO operators choose to work with a subset of wires that they are familiar with.
Table 10.2 Selected Coronary Guidewires Used in CTO Interventions.
Name Manufacturer Core
material
Sion Blue Asahi Intecc Stainless
steel
Minamo Asahi Intecc Stainless
steel
Samurai Boston
Scientific
Hi-Torque BMW Universal II Runthrough NS Extra Floppy Sion Asahi Intecc Stainless
Prowater Asahi Intecc Stainless
Hi-Torque Turntrac Specter Teleflex Nitinol/
Marvel Boston
Runthrough NS floppy
Samurai RC Boston
Abbott Nitinol 0.6 Spring coil
Terumo Nitinol/
Abbott Nitinol 0.8 Spring coil
Scientific Terumo Nitinol/
Scientific
Stainless steel
Stainless steel
steel
steel
Stainless steel Stainless Steel
Stainless steel Stainless Steel
Tip
Tip style Tip coating Tip Load (g)
0.5 Spring coil
Core-to-tip
ACT ONE cable
0.5 Spring coil
Core-to-tip
ACT ONE cable
0.5 Spring coil over
inner coil (ICT)
Core-to-tip
Shaping ribbon
0.6 Spring coil
Core-to-tip
0.7 Spring coil Core-to-tip ACT ONE cable
0.8 Spring coil Core-to-tip
Core-to-tip
0.9 Spring coil Shaping ribbon
0.9 Spring coil Core-to-tip
1.0 Spring coil Core-to-tip
1.2 Spring coil over inner coil (ICT) Core-to-tip
“workhorse” guidewires that affect its utility in CTO PCI include insufficient tip stiffness, lubricity and push that are often needed to traverse CTO caps which can be fibrotic and calcified. A major factor for the increased success rates seen in CTO PCI is improved wire technology resulting in stiffer and more supportive wires with greater torque response and wires with modified coatings and tapered tips.
“Workhorse”
Characteristics
Radiopacity
Hybrid 3cm Hydrophobic silicone
coating from tip to
1.5cm, then hydrophilic coating. Dark color shaft
Hybrid 3cm Hydrophilic with
hydrophobic silicone distal tip
Hydrophilic 4cm ICT – Inner Coil
Technology
Hydrophilic 3cm
Hybrid 3cm Hydrophilic coat with
silicone distal tip
Hydrophilic 3cm Flexible shaft for
increased trackability
Hydrophilic 3cm
Hybrid 3cm “Responsease”
Parabolic Core Grind
Hydrophilic 3cm
Hydrophilic 3cm
Hybrid 3cm Hydrophilic coat with
silicone distal tip
Hydrophilic 4cm Increased flexibility vs
Samurai
CHAPTER 10 CTO Wires: Engineering 101 and Principles of Wire Manipulation 87
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
Table 10.2
(Continued)
Very low tip load
Suoh 03 Asahi Intecc Stainless steel 0.3 Spring coil
Core-to-tip ACT ONE cable
Polymer Coated, Non-tapered
Name Manufacturer Core
material
Whisper LS/ MS/ES Fielder FC Asahi Stainless
Sion Black Asahi Stainless
Pilot 50 Abbott Stainless
Choice PT Boston
PT2 Intermediate Mongo/ Gladius MG 14
Pilot 200 Abbott Stainless
Raider Teleflex Stainless
Abbott Stainless
Steel
steel
steel
Steel
Stainless Scientific Boston Scientific Asahi Intecc Stainless
Steel
Nitinol 2.9 Spring coil Hydrophilic
steel
steel
Steel
Tip Load (g)
1.0/1.0/1.3 Spring coil
0.8 Core-to-tip Hydrophilic
0.8 Spring coil
1.3 Spring coil
2.1 Core-to-tip Hydrophilic
3.0 Spring coil
3.9 Spring coil
4.0 Spring coil
Tip style Tip coating Tip
Core-to-tip
Core-to-tip ACT ONE cable
Core-to-tip
Core-to-tip ACT ONE cable
Core-to-tip
Core-to-tip
Hydrophilic 3cm Retrograde wiring
High shaft flexibility
Characteristics
Radiopacity
Hydrophilic over polymer
over polymer Hydrophilic over polymer
Hydrophilic over polymer
over polymer
over polymer Hydrophilic over polymer
Hydrophilic over polymer Hydrophilic over polymer
3cm
3cm
3cm
3cm
35cm No coils
2cm
3cm Short ACT
ONE- Creates smaller knuckle
3cm
10cm
Polymer Coated, tapered tip
Name Manufacturer Core
material
Fielder XT -R Asahi Intecc Stainless
Steel
Bandit Teleflex Stainless
Steel
Fielder XT Asahi Intecc Stainless
Steel
Fielder XT-A Asahi Intecc Stainless
Steel
Tip Load (g)
0.6 Tapered
0.8 Tapered
0.8 Tapered
1.0 Tapered
Tip style Tip coating Tip
(0.010”) Spring coil Core-to-tip ACT ONE cable
(0.008”) Core-to-tip Spring coil
(0.009”) Spring coil Core-to-tip
(0.010”) Spring coil Core-to-tip ACT ONE cable
Hydrophilic over polymer
Hydrophilic over polymer
Hydrophilic over polymer
Hydrophilic over polymer
Characteristics
Radiopacity
16cm
10cm
16cm
16cm
(Continued)