Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_103_библиотеки_им_акад_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
44 Мб
Скачать
436
O. R. Marmoy
32.2.1 Recording Equipment
32.2.1.1 Electrodes
Electrodes to record the VEP are typically Cl-Cl­chloride, gold cup, or surface electrodes applied to the occipital scalp. According to the interna­tional 10–20 system, at a minimum, the electrode should be placed at the mid-occiput (Oz) and is typically referred to as the mid-frontal region (Fz), with a ground placed around the central scalp (Cz), to record a single channel VEP. As these are a minimum requirement, additional electrodes placed laterally (i.e., O1/O2, or even T5/T6) can be used to identify the contributions from each hemisphere to a stimulus. The latter is useful in evaluating hemianopic or bi-temporal/ bi-nasal eld decits [3]. Furthermore, in pediat­ric practice, an electrode placed inferiorly to the Oz over the inion (Iz) provides larger amplitudes for those under 8 years of age for a standard check width (50) and for those under 12years of age for a small check width (12.5) [4]. An illus­tration of these electrode positions is seen in Fig.32.1.
Before applying the electrodes, the applica­tion site should be prepared using an abrasive gel. This is typically performed using a cotton tip, which can also be used to part the hair, depending on the location. Electrode impedance (measured by passing a low voltage signal through the elec­trodes to observe their respective resistance) should then be measured to ensure that these are balanced (typically within 20%) and maintained below 5 kOhms. It is the author’s practice to wrap a Coban strip around the electrode sites following electrode application to minimize movement (including coiling of wires) and potential shift of electrodes during testing to minimize external noise intrusion.
32.2.1.2 Signal Acquisition
The VEP is a time-locked subtraction of the background electroencephalographic signal. As such, this is an analog signal arising from cortical neurons. Modern systems require digitization of the signals to enable processing on software and potential post-hoc analysis or modication. To achieve this, the incoming ongoing electro-
Fig. 32.1 Illustration of electrode positions for the VEP.Electrodes are placed according to positions dened in the international 10–20 system. The left illustration shows the side view of the head, with active electrodes posi­tioned at Oz (middle of the occiput), Iz (inion), with the
reference electrode placed anteriorly over Fz (mid- frontal) and ground around the center of the head (Cz). The poste­rior view of the head shows the multichannel VEP electrode array, with the Oz and Iz electrodes, but also lateral elec­trodes placed over the left (O1) and right (O2) occiput
32 The Visual Evoked Potential
437
graphic signal is sampled at a high rate (at a mini­mum, twice the frequency of the signal of interest according to the Nyquist-Shannon theorem) in the order of <1kbps. All signals will have a form of differential amplication of the signal prior to signal analysis.
Differential ampliers work in a manner whereby the difference between two inputs is cal­culated. Therefore, a large signal in one electrode and a minimal signal in another will produce a large amplitude output. An example of this would be the Oz electrode being an active electrode over the visual cortex and the Fz electrode being a relatively inactive electrode over the anterior scalp; therefore, the VEP produced is due to a dif­ference in signal strength between these elec­trodes. Signals of the same size and temporal prole, when they enter both electrodes, are said to be in common mode, which typically results in minimal signal output and highlights the impor­tance of balanced electrode inputs. The amplied signals are then subject to ltering, typically used to attenuate frequency bands, allowing one to eliminate unwanted signals (i.e., noise). For the VEP, this is typically set as a 0.3–100 Hz band­pass lter as per the ISCEV (International Society for Clinical Electrophysiology of Vision) standards.
The size of the incoming signal can be auto­matically rejected in some software programs by determining a rejection criterion based on ampli­tude. For most instances, this is 100–200μV (sig­nals above this are automatically rejected).
As the VEP is a time-locked signal of the background electroencephalogram (EEG) sig­nal, averaging is required to improve the signal– noise ratio. Typically, 100 trials are used to constitute an average response, which is then repeated to demonstrate reproducibility. The number of trials may be reduced in people with poorer xation and compliance but should be repeated to conrm the reproducibility of the response.
32.2.1.3 Display andSoftware
Requirements
Display requirements for the analysis of the VEP should allow analysis of the entire waveform on a high-resolution monitor. The software should
enable measuring the peak-time and amplitude of the VEP, ideally with enabled post-hoc manual or automatic rejection of trials to improve the sig­nal–noise ratio.
32.2.2 Stimuli
A VEP can theoretically be elicited to a variety of visual stimuli, for example, color, motion, contrast, luminance, or in some circumstances, an event related to particular stimuli. Despite this, the major clinical applications of the VEP relate to luminance and contrast through the ash (f-) and pattern reversal (pr-) or pattern onset-offset (po-) stimuli, respectively. Accordingly, there are ISCEV standards for these ash and pattern stimuli [3].
The prVEP, poVEP, and f-VEP show different waveforms, which are typically named according to their respective polarity (P for positive and N for negative), timing (i.e., P100= positivity at 100ms), or order of presentation (i.e., P1 or P2). These waveforms are illustrated in Fig.32.2.
For patterned stimuli, there should be an abrupt change of contrast at the stimulus onset without any signicant alterations in mean lumi­nance. This is easily achieved with the older visual display units (VDU), such as plasma dis­play panels or cathode-ray tube devices; how­ever, at the time of writing, these devices are obsolete and not widely manufactured. Unfortunately, common alternatives such as liq­uid crystal display monitors are usually unsuit­able as these have a detrimental luminance artifact with pattern reversals [5, 6]. Modern devices such as digital light projection (DLP) systems or organic light-emitting diode (OLED) devices may be able to overcome these issues, but these are still under review [7, 8]. Any transient luminance artifacts can be easily visualized by darkening the room and placing a piece of paper before the observer’s eyes, facing toward the stimulus, to monitor for any transient luminance alterations visually. Otherwise, these can be mea­sured formally with a photodiode. Any individu­als setting up laboratories should be aware of these potential VDU limitations and their con­founding effects on the PVEP or PERG.
438
O. R. Marmoy
Fig. 32.2 Typical waveforms of the pattern electroretino­gram (PERG), pattern reversal visual evoked potential (prVEP), pattern onset-offset VEP (poVEP), and ash
Fig. 32.3 The effect of check width on the pattern rever­sal VEP.The gure on the left shows the waveform altera­tions with check width, with the top (large check) showing the earliest peak-time, with the amplitude then showing an
The temporal frequency of the stimulus can either be transient or steady-state. The former is most common in clinical practice, as this allows the isolation of individual waveform compo­nents, which can be measured for their peak­time, amplitudes, and shape to identify dysfunction. Steady-state responses are those recorded at higher temporal frequencies (typi­cally >10Hz). These have been utilized in clini­cal practice with varying purposes; the most commonly used is a sweep VEP to provide an objective estimation of visual acuity [9].
The stimulus eld size should subtend at least 15° of the visual eld, typically viewed at around 1m, per the ISCEV VEP standards [3]. It is the author’s practice to have a larger eld size, which is advantageous for pediatric patients whose xa­tion and eye position may be more variable; it is
VEP (f-VEP). The waveforms are plotted by amplitude and time and labeled according to the conventional nomenclature consistent with the ISCEV standards
inverse U-shaped function with decreasing check width, with increasing peak-time. These data are plotted accord­ing to the check width in the right panel (amplitude in blue, peak-time in red)
increased to around 30° to allow for variable xa­tion. However, with increasing eld size, one must be more aware of the alterations in the prVEP morphology. Typically, with larger eld sizes, there are larger contributions from paramacular prVEP components, which is dis­cussed later in this chapter.
Check width is an important factor for per­forming the VEP. The ISCEV VEP standard species that the prVEP should be recorded with a large (1°) and small (0.25°) check width. Importantly, the amplitude typically shows an inverse U-shaped function with the peak ampli­tude around 15 min of arc (0.25°), while the amplitude shows a more linear increase with decreasing check width (Fig. 32.3). Expanding the number of check widths used in clinical prac­tice can reveal macular pathway dysfunction
32 The Visual Evoked Potential
439
where small check widths are affected rst [10] and can give more consistency to a response. Therefore, while a minimum of two check widths should be used, additional check widths are ben­ecial to characterize macular pathway function in practice.
32.3 Technique
A typical recording setup would have a patient seated comfortably in a chair, preferably with a high back and headrest. Electrodes would be applied with low, balanced impedances. During application, a discussion on how the test is per­formed and what is needed from the patient can be initiated, and any questions can be answered. It is essential to ensure that the patient is relaxed to minimize muscle artifacts (i.e., chewing, neck straining, etc.) and ensure their comfort as best possible (i.e., with neck pillows, adjustment of seating position, etc.).
The patient should wear their most recent spectacles for testing. If these are unavailable, trial frames can be used, or if unavailable, a note made in the report that the tests were performed uncorrected (and dene their correction). The pattern VEP can withstand a signicant degree of optical blur but is dependent on test strategy and other patient factors, the degree of which should be noted.
For children, alternating the stimulus checker­board with a video of their choosing will most likely result in better testing conditions and a happier child! The prVEP depends on xation and attention; therefore, this should be monitored closely for xation loss, deliberate mis-xation, or xation drift. One way this can be achieved is by simultaneously using camera systems (i.e., CCTV, IP cameras) to monitor the patient’s xa­tion during signal acquisition. Thus, in periods of xation loss, signal acquisition can be paused and resumed once xation has returned. In younger children, larger eld sizes are necessary to attract attention and account for some unavoid­able xation variance. In these circumstances, it is best to work with colleagues as a pair, one to “distract and interact” with the child and the other to focus solely on signal acquisition.
Fixation can be encouraged using small toys or ngers at the top of the stimulus screen. As the prVEP is dominated by the inferior eld, small obscurations in the superior eld have negligible impacts on the PVEP and enable more reliable xation on the stimulus. Intermittent averaging is required to allow for the patient’s attention. Prolonged presentation of a checkerboard stimu­lus will lead to tiredness, habituation, or defocus, even in the most enthusiastic of adults.
32.4 Clinical Application andInterpretation
The typical responses to ash, pattern reversal, pattern onset-offset stimuli are illustrated in Fig. 32.2. To interpret the VEP waveform, the acronym “PAST” (Peak-time, Amplitude, Shape, Transoccipital asymmetry) can be used to evalu­ate the response.
32.4.1 Peak-Time
The peak-time of the waveform signies the time taken for cortical activation following afferent pathway stimulation. The peak-time is most con­ventionally measured from the major positive peak of the respective VEP waveform, for exam­ple, the P100 of the prVEP, C1 of the poVEP, and P2 of the ash VEP [3]. A delay in peak-time (i.e., a peak-time exceeding the laboratory refer­ence range) can result from dysfunction of the physiological substrates underlying the response. For example, a delay of the prVEP P100 can be due to a dysfunction of the macular, retinal ganglion cells, optic nerve, or chiasmal/retro chi­asmal visual pathway. To elaborate on the dys­function site, a PERG is required to delineate between macular, retinal ganglion cell, or optic nerve/pathway dysfunction. While prVEPs to standard check widths are fairly resistant to an optical blur of <5 D, the interpretation of peak­time delays to smaller check widths must be cau­tiously made, as this may result from uncorrected refractive error, reduced contrast (i.e., from ante­rior segment disease such as cataract) or poor xation/blurring. Another important consider-
440
O. R. Marmoy
ation is the maturation of the VEP.The VEP to standard (large) check widths is typically within 10% of adult peak-times by 6months of age, but younger people may have longer peak-times due to the immaturity of their visual pathways.
Similarly, poVEPs can show shape-dependent changes with maturation [11]. This is an impor­tant consideration for interpreting the PVEP, as these factors should be excluded before attribut­ing a peak-time delay to macular pathway dys­function, including age-matched reference data. Simultaneous PERG and PVEP recordings can elaborate on an abnormal VEP, particularly for suspected optical blur or defocus, as the PERG is far more sensitive to uncorrected refractive error and contrast alterations [12].
Peak-time delays of the prVEP with normal amplitude most commonly signify conduction delay but are not specic to demyelinating dis­ease. In demyelinating disease, a conduction delay can be observed even following the resolu­tion of visual symptoms or, in some cases, may show abnormality as a “clinically silent” lesion, providing paraclinical evidence of demyelination in conditions such as multiple sclerosis [1]. Importantly, peak-time delay is not specic to demyelination but can be observed in various optic nerve diseases, albeit to varying degrees. Typically, with severe forms of demyelinating optic neuropathies, PERG N95 loss and ampli­tude reduction of the VEP following the acute phase are often associated with poorer visual out­comes (Fig.32.4).
While peak-time delay is most common, in some pathologies, one can expect an atypically early peak-time. Such abnormalities are usually associated with an alteration in waveform mor­phology, such as bid or double-peaked wave­form, discussed in the “shape” analysis below.
32.4.2 Amplitude
Reduced amplitude of a VEP suggests poor sig­nal strength; as such, this is associated with mac­ular pathway dysfunction. Reduced amplitude is also a relatively non-specic indication of macu­lar pathway abnormality. It is often observed in
conditions with reduced physiological signal generation; for example, this may be reduced in retinal disease, optic atrophy (whereby fewer axons are functioning), or optic nerve hypopla­sia. It is benecial to record pattern VEPs to larger check widths to observe the alterations with changes in spatial frequency. Recording to a range of check widths is also useful, as it can help with the monitoring of the disease, whereas those for standard or small checks may progres­sively diminish in amplitude; therefore, the larger check widths may sometimes be the only means by which disease progression can be monitored [13].
While reduced amplitudes are most com­monly observed in diseases of the macular path­way, in rarer circumstances, VEPs may be atypically large or “giant.” This may occur with normal peak-time of the response in patients with chronically raised intracranial pressure, perhaps due to thinning of the skull, reduced cerebrospi­nal uid thickness, or cortical hyperexcitability [14]. Similarly, atypically large VEPs can occur with cortical hyperexcitability in epileptic disor­ders such as neuronal ceroid lipofuscinosis bat­tens disease, where a “spike and wave” type paroxysmal VEP response is observed, often with an early peak-time [15]. In such circumstances, VEP testing may be halted to minimize the risk of pattern or ash-related seizure induction.
The pattern VEP can be a useful addition in monitoring retinal disease, particularly in those with rod–cone photoreceptor dystrophies. As eld size reduces from the periphery and encroaches upon the macula, the full-eld ERG can be reduced or extinguished, albeit with vesti­gial PERG responses. As the fovea is often more preserved during the later disease stages, the prVEP recorded to a range of check widths can provide information on the remaining integrity of the macular pathway and be used for monitoring disease progression. This is likely due to the expanded foveal representation of the PVEP due to cortical magnication, but it is particularly useful to apply this when the ERG or PERG is extinguished to provide a functional outcome measure in disease [16]. An example of this is shown in Fig.32.5.
32 The Visual Evoked Potential
441
Fig. 32.4 An example of the waveforms seen in a patient with left eye (LE) optic neuritis is associated with myelin oligodendrocyte glycoprotein antibody-associated disease (MOGAD). The normal waveforms of the pattern ERG, VEP, and ash VEP are shown in the top panels. The patient’s (unaffected) right eye (RE) shows normal PERG ndings, with normal prVEP to large and small check
32.4.3 Shape
The morphology of the VEP can be altered in dis­ease states and so can be used to delineate visual pathway dysfunction. The most common abnor­mality in the VEP is a broadened or bid VEP waveform.
A broadened prVEP can occur with macular pathway abnormalities, likely due to temporal dispersion of the afferent signal to the striate cor­tex. This has been reported in grading schemes of VEPs in craniosynostosis [13], alongside other pathologies. This has been proposed as a
widths and normal ash VEP. The patient’s affected LE shows N95 amplitude loss and shortening of P50 peak­time, associated with retinal ganglion cell dysfunction. The patient’s prVEP is reduced and delayed in the LE relative to the RE to both large and small check widths. The ash VEP is slightly simplied relative to the RE but is of similar amplitude and peak-time
waveform N75-N135 peak-time difference exceeding 90ms [13, 17]. The signicance of a broadened prVEP likely reects mild macular pathway dysfunction in the absence of amplitude or peak- time anomalies. Nonetheless, these changes are worth monitoring as they may sug­gest an early manifestation of visual pathway dysfunction.
A bid VEP is a waveform which adopts a P-N-P morphology, opposite to the typical con­guration of N-P-N.This can be associated with optic nerve or demyelinating disease, and it is the author’s experience that these changes are due to
442
O. R. Marmoy
Fig. 32.5 Electrophysiological ndings in a patient with retinitis pigmentosa. The full-eld ERGs show severely reduced DA a - and b-wave amplitudes, with preserved but signicantly reduced amplitudes of the LA cone­mediated responses indicative of severe rod–cone dystro­phy. The pattern ERG was reduced, and the amplitude did not increase substantially between the 15° and 30° elds, indicating distal retinal (cone-driven) dysfunction affect-
ing the 15°–30° eld. Despite these ndings, the prVEPs and fVEPs are within the reference range. The ultra­wideeld (Optos) fundus autouorescence shows pig­mentary changes in the periphery (red arrow) alongside a hyperuorescent ring around the macula. The macular OCT shows IS/OS loss around the macular periphery, with relative foveal sparing. The preservation of the prVEPs is likely due to the relative foveal sparing
Fig. 32.6 The prVEPs in a healthy participant were recorded with increasing central scotoma size (0–20°). The waveform alterations can be observed with increasing scotoma size to exhibit a bid morphology of the wave-
an enhancement of the paramacular prVEP com­ponents and a reduction of the major P100 peak [12]. Accordingly, the nding of a bid prVEP often suggests reduced central eld sensitivity or central scotoma, meaning that the paramacular retinal generators predominate the resultant
form with two positive peaks (red arrows). These peaks correspond to the paramacular components of the half­eld prVEP (right panel, red arrow)
VEP. This is illustrated in Fig.32.6 in a healthy individual; an increasingly larger scotoma alters the prVEP morphology to adopt a bid wave­form. As can also be seen, this waveform is simi­lar to the paramacular components (p80) observed contralaterally in the half-eld prVEP. As such,
32 The Visual Evoked Potential
443
Fig. 32.7 Electrophysiological ndings in a patient with Leber's hereditary optic neuropathy. The pattern ERG shows early P50 peak-time and N95 amplitude loss indi­cating severe retinal ganglion cell disease. The prVEP shows an abnormal morphology (red asterisk) with atypi­cal bid morphology. The half-eld prVEPs show attenu-
recording half-eld prVEPs in patients with bid prVEP waveforms helps one understand if these components truly arise from a paramacular origin or are caused by an alteration in the macular VEP waveform (Fig.32.7).
32.4.4 Transoccipital Asymmetry
Transoccipital asymmetries can be investigated when one extends the electrode montage to include a multichannel recording. This typically uses additional lateral electrodes, “O1” or “O2,” according to the international 10–20 system. An asymmetry across the occiput can dene transoc­cipital asymmetries. While individual laboratory reference ranges will determine the limits for a signicant transoccipital asymmetry (which may depend on the check width and the eld size), it is the author’s experience that a transoccipital asymmetry (i.e., a difference between the lateral electrodes) for an amplitude of >30% is consid­ered suspicious, and those with amplitudes >50% are considered signicant. Similarly, a peak-time difference exceeding 6ms across the transoccipi­tal array is also considered signicant.
A transoccipital asymmetry is typically sug­gestive of a chiasmal, post-chiasmal, or relative eld defect, depending on the pattern of abnor­mality and if conrmed with appropriate testing. The localization of the lesion can be achieved
ation of the ipsilateral major ip100 component (red asterisks), yet the contralateral paramacular components remain and correspond to the P80/P140 peaks observed in the full-eld prVEP. This patient was later detected to have a central scotoma on visual eld testing
with half-eld VEPs although knowledge of the underlying neural substrates generating the responses is crucial to determine the site of the dysfunction accurately.
When using a large eld, with standard (or large) check widths, with a mid-frontal reference electrode, prVEPs demonstrate a phenomenon known as paradoxical lateralization [18]. This phenomenon means that responses generated to the right half-eld stimulation (corresponding to the left hemisphere) are “paradoxically” observed over the right occipital electrode (Fig.32.6, right panel). This is likely due to the oblique orienta­tion of the cortical prVEP generators within the calcarine sulcus. Conversely, pattern onset VEPs do not exhibit paradoxical lateralization, yet pat­tern offset VEPs do, similar to prVEPs [19]. Abnormalities of the prVEP transoccipital distri­bution can therefore be elaborated upon using half-eld stimulation of the left or right-half eld of each eye. The abnormalities may be “crossed,” meaning that the distribution of the transoccipital asymmetry will be altered depending on which eye is stimulated (e.g., in chiasmal dysfunction or disproportion), or can be “un- crossed” in hom­onymous hemianopia (i.e., post- chiasmal lesions). Half-eld testing can be used to elabo­rate on full-eld prVEP distributions, to identify whether the transoccipital asymmetry is due to a relative decit of one half-eld or whether this is instead due to individual cortical architecture or
444
O. R. Marmoy
Fig. 32.8 A patient with genetically conrmed oculo­cutaneous albinism. Data for the pattern onset (poVEP) VEP, half-eld pattern reversal VEP (RHF prVEP/LHF prVEP), and ash VEP are presented for electrodes over­laying the left occiput (left panels), right occiput (middle panels), and a difference of these two (right panels). The RE (red) and LE (blue) data are illustrated. The pattern onset VEP shows that the major positive peak was largest over the contralateral occiput, which when compared
VEP generator orientation. If half- eld testing is not possible due to poor xation or cooperation, the poVEP can be used to corroborate any poten­tial eld defect. If the major pattern onset VEP positivity is largest over the same lateral elec­trode as the prVEP, this likely reects individual cortical architecture/VEP generator orientation, whereas if the laterality is opposite between
from each eye in the difference channel shows a phase reversal (red arrows). The half-eld prVEPs show a nor­mal distribution of the bi-temporal elds but an atypically symmetrical distribution of the bi-nasal elds. The fVEPs show a clear difference in lateralization between eyes, with a contralateral predominance of the major negativity (~70ms) and positivity (~120ms) from each eye, which is observed as a clear crossed asymmetry in the difference channel (red arrows)
prVEPs and pattern onset VEPs, this is sugges­tive of a relative decit of a half-eld [19]. For crossed asymmetries (i.e., where the direction of the lateralization depends on the stimulated eye), a virtual channel subtracting the difference between the right and left occipital electrode is useful to visualize the transoccipital symmetries between the eyes (Fig.32.8). Some centers also
32 The Visual Evoked Potential
445
perform a “chiasmal coefcient” calculation [20], in which the difference between the left and right electrodes from each eye are quantied for a signicant correlation (negative or positive). Such differences are usually utilized to investi­gate chiasmal misrouting associated with albi­nism although these may also be used to identify patients with chiasmal dysfunction (i.e., space­occupying lesions or chiasmal hypoplasia).
32.5 Pitfalls andPearls inRecording VEPs
The VEP can be an easily accessible and efcient technique for assessing macular or generalized pathway function and investigating visual path­way integrity. However, there are some circum­stances where one must be aware of the limitations or pitfalls of recording VEPs.
Perhaps one of the major misconceptions of the VEP is that a normal VEP is equivalent to normal sight. As the VEP only assesses the visual pathway from the retina to the primary striate cortex, disor­ders of higher visual processing (e.g., cerebral visual impairment) may not appear as abnormali­ties in the VEP.Similarly, the VEP cannot be used in isolation to determine normal acuity in people with functional disease. An otherwise normal patient with organically reduced visual acuity would be less likely to have a normal VEP.However, it should be investigated with a sweep VEP to esti­mate visual acuity (Fig.32.9) objectively.
An important technical consideration for recording the VEP is the importance of good xa­tion and attention. The VEP can be easily altered due to poor attention or xation on the stimulus. Therefore, it is often advantageous to simultane­ously monitor the patient’s xation with an infra­red camera to perform intermittent averaging (e.g., stimulus averaging is performed when the patient’s xation is veried by viewing corneal reections). This is particularly benecial in pediatric or non-compliant people so that any erroneous xation is accounted for when record­ing the VEP.
Lastly, an important note for recording the VEP is to remain aware of the principles of VEP signal acquisition. The VEP is an averaged response from the time-locked background EEG.As such, in people with abnormal poste­rior EEG activity (i.e., due to seizure disorders or encephalopathies), the signal-to-noise ratio of the VEP may be too poor to obtain a consis­tent or repeatable response. In such circum­stances, the VEPs elicited can be easily distinguished from erroneous EEG activity by recording non- stimulus trials. To perform this, one typically runs the VEP recording without presenting a stimulus (i.e., by turning off the monitor or facing the ash strobe down). If the non-stimulus and stimulus trials are compara­ble, this suggests that any recorded waveform is artifactual, whereas if these show a distinct dif­ference, then it provides condence in its authenticity.