Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5772_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
58 Мб
Скачать
Physical and Technical Principles of Color Doppler Sonography
ab
Fig. 1.26 Vibration artifacts associated with high-grade stenosis and AV fistula. a Patient 1: Vibration artifact in the lesser pelvis due to a high-grade stenosis (angiography demonstrated a 90% transplant renal artery stenosis).
can lead to spotty color flow (Fig. 1.27) or can mimic thrombo­sis. Technicalproblems such as too large a Doppler angle or cal-
1
cifications in the scanned region can be recognized in the B­mode image and avoided. It should be noted that acoustic shadowing varies with the insonation angle and that angling the color box can produce an oblique shadow (Fig. 1.
False-Positive Flow
Image clutter, pulsation artifacts, or gross motion artifacts (es­pecially in power Doppler) can mimic the presence of in­traluminal flow or tissue perfusion.
Pulsation and motion artifacts. Image clutter (flash artifact) may be caused by setting the gain too high or the PRF too low. Soft-tissue pulsations or motion artifacts can produce color
28).
b Patient 2: Coarse, continuous vibration artifact located to the right of the midline, anterior to the aorta. c Patient 2: Angiogram demonstrates an AV fistula between the gastroduodenal artery and a branch of the superior mesenteric vein, causing immediate opacification of the portal vein.
signals that are misinterpreted as flow. Transmitted cardiac pulsations can produce an apparent flow pattern, especially in cysts located near the diaphragm. Confetti-like artifacts as­sociated with high-grade stenoses and AV fistulae are caused by the vibration of perivascular soft tissues and structures in response to rapid jet currents.
Slice-thickness artifacts and “phantom vessels.“ Slice-thick­ness artifacts are particularly common in older scanners. They can cause diagnostic errors in color duplex scanning, as flow information from neighboring vessels may be projected into an occluded or thrombosed vessel. “Ghost vessels” or “phantom vessels” can appear behind strong reflectors such as the dia­phragm or pleura (Fig. 1. not only contains color signals but can even yield a Doppler spectrum.
c
28), mimicking a second vessel that
22
a
Fig. 1.27 Color Doppler artifacts and pitfalls. I: Effect of color priority. a Longitudinal scan of an artery with optimum settings (Table 1.3)
shows complete filling of the vessel lumen.
b
b When color priority is decreased, the intraluminal color signal be-
comes spotty.

New Technical Processes and Approaches

b
a
Fig. 1.28 Color Doppler artifacts. II: Shadowing, motion artifacts, and mirror-image artifact. a Calcific plaque with a heavy acoustic shadow. The shadow runs per­pendicularly in the B-mode image, but the shadow in the color-flow image is oblique owing to the angle of the color box. b Longitudinal scan of the posterior tibial artery, with motion artifacts (flash artifacts) induced by movement of the transducer. c Mirror-image artifact along the pleural apex. At the top of the image is the subclavian artery with the origin of the vertebral artery. Below
the subclavian is a “phantom vessel” caused by the reflection of color
Doppler signals from the pleura, which is oblique to the beam.
New Technical Processes and Approaches
Technical advances in ultrasonography have been felt most strongly in the areas of transducer technology and manufac­ture; computer hardware with the capabilities of digital high­speed electronics; and new approaches to signal acquisition, filtering, and processing.
New Developments in Transducer Technology
Transducers with an increasing density of piezoelectric el­ements are offering higher resolution along with new, flexible scanning formats based on the steering capabilities of ar rays (see Principle of Echo Detection and Scanning Techniques, p. 4).
Trapezoid scan. When a linear transducer is operated as a phased array, a sector field can be added to each side to expand the linear scan into a trapezoid scan. This provides a broader field of view and makes it possible to evaluate even hard-to­scan vascular segments such as the origin of the vertebral artery or common carotid artery from the brachiocephalic trunk.
c
2 D and multi-D arrays. Other innovations in transducer tech-
nology are two-dimensional (2 D) and multidimensional (multi-D) arrays. Unlike conventional array designs (one-line arrays), the 1.5D array consists of multiple parallel rows of el­ements (n m matrix). The rows are driven in the transmit and receive modes such that the slice thickness (elevation plane) can be kept optimally thin in both the near and far fields (Fig. 1.
29).
The smallest elevation aperture (i.e., the center row of el­ements) is used for imaging in the near field, while the largest elevation aperture (i.e., all rows) is used for deeper scanning to help achieve good penetration depth and slice-thickness reso­lution. The greatest resolution increase occurs in the plane per­pendicular to the transducer (Fig. 1.
Further subdivision and miniaturization of the array lines has led to the design of 2 D arrays that can scan along as well as perpendicular to the conventional scan plane. This makes it possible to acquire data from a pyramid-shaped volume and
generate sectional images with any desired orientation.
29).
Physical and Technical Principles
23
Physical and Technical Principles of Color Doppler Sonography
a
a
b
Elevation
lateral
late
l
ra
xia
axial
l
1
Elevation
Fig. 1.29 Improved resolution with a multi-dimensional transducer array. Slice-thickness focusing of a conventional transducer with a single row of elements (a)and a multi-dimensional array (b). While the axial resolution remains unchanged, the resolution perpendicular to the transducer plane (elevation plane) can be significantly improved by focusing.
tions of the array’s various transmitted pulses are temporarily stored in digital form, digitally focused along with other adja­cent subapertures, and utilized for image formation. As an ex­ample, Fig. 1.
30 shows the principle and effect of increasing the
aperture by a factor of 2. Other factors can also be achieved.The more the aperture is enlarged, the better the lateral resolution and the greater the penetration depth and the ability to recog­nize details. The practical resolution limit is reached when all elements in the array are used to compute the information for every point in the ultrasound image.
Parallel processing. Another advantage of high-spee d digital electronics is the ability to parallel-process multiple ultra­sound lines from the information of one transmitted pulse. The practical advantages of parallel processing are a faster frame rate, especially in the two-dimensional Doppler techniques, and an improved ability to analyze rapid, dynamic processes and appreciate details.
Special filtering techniques. When special filtering techniques are applied, nonlinear tissue properties can be utilized to form the image. One cause of nonlinear tissue propertiesis the effect of ultrasound pressure on sound propagation velocity in the tissue, especially when high sound pressures are used. Because of this mechanism and the inherent differences in sound prop­agation velocity in different tissues, leading to different transit times, the pulse emitted by the transducer undergoes distor­tion in the tissue. If we analyze the frequency content of the re­ceived echoes, we find that they contain certain harmonic mul­tiples, or overtones, of the basic frequency (Fig. 1.
31).
Tissue harmonic imaging. Second-harmonic waves, or wave components at twice the transmitted frequency, are especially
24
New Techniques of Signal Acquisition and Processing
High-speed digital technology. The latest generation of ultra-
sound scanners can fully digitize the received echo signals while the signals are being acquired. They are also distin­guished by their high computer capacity. The high-speed dig­ital technology that is used in these systems combines the ad­vantages of optimum signal scanning and digital technology (precision, signal fidelity, stability, signal dynamics) with a very high temporal resolution for transit time and phase cor­rection. These optimal focusing conditions provide excellent lateral and axial resolution with very high signal dynamics, making it easier to appreciate diagnostic details.
Synthetic aperture. Other improvements result from enlarging the active surface area of the transducer (the aperture). This re­quires accurate, digital interim storage of the high-frequency echo signals from the individual array elements in unchanged form. This makes it possible to enlarge the aperture beyond the conventional theoretical maximum. In the synthetic aperture (e.g., SynAps) of the Sonoline Elegra system, echoes from por-
Digital memory
A
1
Parallel processing
Fig. 1.30 Synthetic aperture scanning: technical principle and effect. Each aperture consists of 128 elements, for example. With a digital in­terim memory, it is possible to combine the imaging information from multiple subapertures (two shown here: A aperture of 256 elements. The advantages are greater penetration depth and improved lateral resolution.
Beam former
128 channels
A
1
Tot a l
*) into a large total
1,A1
A1*
New Technical Processes and Approaches
Transmitted waveform
Pressure wave faster than negative pressure component
Waveform that has passed through tissue
Waveform
Fig. 1.31 Effect of nonlinear sound propagation in tissue. A sound
wave typically begins as a sine wave. The tissue is compressed, however, by the positive component of the pressure wave, and so the sound velocity is temporarily increased. This leads to a more rapid propagation of the positive part of the pressure wave (right arrow). During the rarefaction phase of the pressure wave, the tissue can ex­pand again and the sound velocity decreases. As a result, the negative pressure component of the sound wave is conducted moreslowly. This phase-dependent change in sound velocity distorts and steepens the original sine wave, analogously to a wave breaking on the shore. The effect increases with increasing depth. The result is that the reflected
wave acquires a harmonic multiple of the basic transmitted frequency (“harmonic energy”). These frequencies, which are troublesome in a conventional B-mode image,can be specifically utilized for“tissue har­monic imaging” (THI). This technique can be used successfully even in patients who are technically difficult to scan.
f02f03f
0
Spectrum Depth
prominent in the returned spectrum. Tissue harmonic imaging (THI) involves isolating these components and suppressing the basic frequency components. This technique has already been used to obtain “clearer,” artifact-free images in patients who are technically difficult to scan.
Phase-inversion or pulse-inversion techniques. The phase-in-
version or pulse-inversion techniques can be used to suppress the fundamental transmission spectrum and use only the har­monic frequency components over a wide frequency band to form the image (wideband tissue harmonic imaging, ensemble tissue harmonic imaging). Instead of a single pulse, two identi­cal pulses are transmitted into the tissue per image line. The first is a standard ultrasound pulse, and the second is an exact copy with its phases inverted. The reflected and backscattered pulse pairs are subtracted (Fig. 1.
32). All of the unchanged,
“linear” signals cancel out. In this way all the signals from the tissue that are changed owing to nonlinear effects are selected out and can be used for image formation, independent of filter­ing techniques. The effect is cumulative and increases with penetration depth. The THI technique is most effective at mod­erate and greater depths. Further penetration past a depth of 12–15 cm is limited by absorption. This technique provides sig­nificantly greater image improvement than traditional filter­ing techniques for extracting the nonlinear (harmonic) com­ponents of the returned signal. The phase-inversion technique
Pulse 1
Pulse 2
Sum
Linear propagation
Fig. 1.32 Principle of the phase-inversion or pulse-inversion tech­nique. Two identical pulses whose phases are inverted (shifted 180) are transmitted into the body (top and center rows: pulse 1 and pulse
2). When the returning signals are added, the linear (unchanged) echoes cancel out (lower left). The signals that have been distorted due to nonlinear effects (right) add together and can be used for image formation. This process suppresses the fundamental (basic)
frequencies and the odd harmonic signal components while amplify-
ing the even harmonic components, especially the second harmonics.
Nonlinear propagation
has already proved successful clinically in enhancing the visu­alization of smaller lesions (Fig. 1.
33). When harmonic imaging
is combined with ultrasound contrast agents, even the small­est metastatic lesions can be identified (Fig. 1.
34).
Programmable image processor. By incorporating a special programmable image processor into the ultrasound system, it is possible to raise the computer capacity for image processing (several billion operations per second) to a level that was pre-
viously reserved for external workstations. In this way, various image processing tasks that formerly had to be done on sepa­rate systems (offline) can now be performed directly (online) and in real time. An example of this is adaptive contrast en­hancement. The result is ultrasound images that are better suited to the physiology of human vision, making it possible to conduct the examination even in a bright room.
Panoramic imaging. Another application based on the pro-
grammable image processor is panoramic imaging, first intro­duced as the SieScape technology in the Sonoline Elegra sys­tem. In panoramic imaging, an extended field of view is ob­tained by manually scanning a standard transducer over the region of interest without a position sensor. This information is first displayed on the monitor in standard fashion. Based on the level of agreement between the individual real-time images, correlational algorithms are used to determine the transducer position on the body. This positional information is then used to seamlessly add the new image to the previous image. Thus
when the transducer is moved along one plane or around an area, a real-time panoramic image is obtained. In this way the operator can freely define an image format that is optimum for the specific anatomy (Fig. 1.
35). Panoramic imaging has been
combined with power Doppler flow imaging to produce a color-encoded panoramic imaging process marketed as Color SieScape.
Physical and Technical Principles
25
Physical and Technical Principles of Color Doppler Sonography
a
Fig. 1.33 Effect of tissue harmonic imaging (THI),
illustrated for renal cysts and the gallbladder. a Longitudinal scan through the right kidney of a
patient with small renal cysts that are barely detectable
in the fundamental mode (left). The THI mode (right) improves the delineation of the cysts and gives an arti­fact-free view of the cyst lumen. b Enlarged view of a gallbladder with a polyp. In the fundamental mode (left) the lumen contains spurious scattered echoes, and there are accentuated semicircu-
lar artifacts in the adjacent liver as a result of reverbera-
tions with a curved array. In the THI mode (right), the gallbladder lumen is free of artifacts and the other arti­facts are less pronounced.
1
b
Fig. 1.34 Effect of contrast harmonic imaging in the detection of small hepatic metastases following the injection of 300 mg/ml Levovist. After 2.5 minutes the contrast agent has become concen­trated in the reticuloendothelial system (RES). The transmit power is increased, causing the contrast microbubbles to vibrate and burst. This “acoustic emission” produces hyperechoic enhancement of the liver parenchyma. Foci that lack a RES do not enhance, and remain hypoechoic. As a re­sult, even small metastases thatare difficult or impossible to detect on plain images can be identified by their sharp contrast with the sur­rounding hepatic tissue.
26
New Technical Processes and Approaches
a
Fig. 1.35 Panoramic images of the breast. a Breast with a postoperative hematoma. A standard transducer is
scanned over the region of interest. Initially this information appears on the monitor in the usual form. Based on the degree of agreement between the individual real-time images, image-comparison algo-
3 D Imaging. If the transducer is not moved just in the selected plane as in panoramic imaging but is moved or angled perpen­dicular to that plane, a three-dimensional (3 D) ultrasound data set is acquired. These movements may be accomplished by a mechanical apparatus within a special transducer (e.g., a sweep mechanism in a convex array) or by free manual move­ments of the transducer. Manual movements may involve three-dimensional position detection using position sensors or a method based on uniform, predefined movement patterns.
The “3-Scape” real-time 3 D imaging technique is based on the latter method and employs panoramic technology with stan­dard transducers to enable freehand volume acquisition by, say, moving the transducer perpendicular to the scan plane (sweeping is also possible). In 3-Scape imaging, the volume
b
rithms are used to assemble the separate images into a panoramic scan. In this way the entire breast can be displayed in one image, and
the location of the hematoma can be seen.
b The exact size of a silicone breast implant can also be determined
following augmentation mammoplasty.
data are collected and computed during the scanning process.
While the transducer is still moving, this data set is used to
generate a perpendicular guide plane that can be use d for orientation. The three-dimensional data set can then be dis­played as a complete volume in a maximum intensity projec­tion (MIP), for example, or surface projection algorithms can be used to construct a surface-rendered image. The data set can also be manipulated to display arbitrary planes of section
within the 3 D volume that are not accessible with conven­tional B-mode imaging (Fig. 1.
36).
If flow information (in the power mode) is also obtained during data acquisition, the 3 D technique can provide excel­lent survey views for the complete documentation of organs,
vascular distributions, and tumor blood supply.
Physical and Technical Principles
Fig. 1.36 Quadruplet gestation imaged by real-time 3 D ultrasound with arbitrary plane selection. The three-di­mensional data set can be manipulated to display the sectional planes that give the best view of an embryo or other region of interest. The images in this case reveal a quadruplet gestation.
27
Physical and Technical Principles of Color Doppler Sonography
References
1 Bonnefous O, Pesque P: Time domain formulation of pulse Doppler ul-
trasound and blood velocity estimation by cross correlation. Ultrason. Imaging 8 (1986) 73–85
2 Campbell S, Diaz-Recasens J, Griffin DR: New Doppler technique for
assessing uteroplacental blood flow. Lancet 1 (1983) 675–678
3 FitzGerald DE, Drumm JE: Non-invasive Measurement of human Fetal
Circulation using Ultrasound: A new Method. BJM 2 (1977) 1450–1451
4 Gosling RG, King DH: Ultrasound Angiology. In Macus AW, Adamson J
(eds.): Arteries and Veins. Churchill-Livingstone, Edinburgh 1975
5 Haerten R: Verfahren der Farbdoppler-Sonographie: Ein Methoden-
vergleich. Ultraschall Med. 14 (1993) 225–230
6 Kasai C, Namekawa K, Koyano A, Omoto R: Real Time two-dimensional
Blood-flow Imaging using autocorrelation technique. IEE Tran. Soc. Son. Ultrason. SU 32 (1985) 458–463
7 Kay SM: Modern Spectral Estimation: Theory and Practice. Prentice
Hall 1988
8 Klews PM: Physik und Technik der farbkodierten Duplexsonographie.
In Wolff K-J, Fobbe F (eds.): Farbkodierte Duplexsonographie. Thieme, Stuttgart 1993
9 Liu D, Kim J, Schardt M: Modified autocorrelation method compared
with maximum entropy method and rf cross-correlation method as mean frequency estimator for Doppler. IEEE Ultrasonic Symposium 1991, 1285–1290
1
10 Maulik D, Nanda NC, Saini VD: Fetal Doppler Echocardiography:
methods and characterization of normal and abnormal hemodynam­ics. Am. J. Cardiol. 53 (1984) 572–578
11 Pourcelot L: Applications clinique de l’examen Doppler transcutane. In
Pourcelot L (ed.): Velocimetric Ultrasonore Doppler. Iserme, Paris 1994
12 Satomura S: Ultrasonic Doppler Method for the Inspection of Cardiac
Functions. J. Acoust. Soc. Am. 29 (1957) 1181–1183
13 Seitz KH, Kubale R: Duplex-Sonographie der abdominellen und retro-
peritonealen Gefäße. VCH, Weinheim 1987
14 Soldner R: Physikalische Grundlagen der sonographischen Bildge-
bung, Ultraschall-Gewebe-Interaktion und Sicherheitsaspekte. In Bogdahn U, Becker G, Schlachetzki F (eds.): Echoverstärker und trans­kranielle Duplex-Sonographie. Berlin 1998
15 Stuart B, Drumm J, FitzGerald DE, Diugnan NM: Fetal blood velocity
waveforms in normal pregnancy. Br. J. Obstet. Gynaecol. 87 (1980) 780–785
16 Taylor KJ, Burns PN, Woodcock JP, Wells PN: Blood flow in deep
abdominal and pelvic vessels: ultrasonic pulsed Doppler analysis. Radiology 154 (1985) 487–493
17 Thompson RS, Trudinger BJ, Cook CM: Doppler ultrasound waveforms
in the fetal umbilical artery: quantitative analysis technique. Ultra­sound Med. Biol. 11 (1985) 707–718
18 Weiser HF, Birth M: ViszeralchirurgischeSonographie. Springer,Berlin
Heidelberg New York 2000
19 Hetzel G: Neue technische Entwicklungen auf dem Gebiet des Ul-
traschalls. Der Radiologe 10 (2003) 777-792
28

2 Safety Aspects of Doppler and Color Doppler Sonography

H.-D. Rott
Doppler sonography, with its ability to define and evaluate the fetal blood supply, has been an important addition to diagnos­tic ultrasound in obstetrics. Because these techniques require power outputs and ultrasound intensities that are considera­bly higher than in B-mode imaging, the essential safety that

Mechanisms of Tissue Effects

The biological effects of ultrasound are based largely on heat­ing and cavitation. These effects are dependent on different sound field parametersand tissue properties and can therefore occur separately from each other. Other primary physical ef­fects have no bearing on the clinical safety of ultrasound.
Heating
Ultrasound energy that is transmitted into tissue is partially re­flected, partially scattered, and partially absorbed and con-
verted to heat. The degree of heating depends on different properties of the ultrasound field and the exposed tissue. A key factor is the spatial peak time average intensity (I pulsed Doppler flowmetry, however, the power output of the device is a more important factor than field intensity sound frequency is also a factor, since higher frequencies are absorbed more strongly, and therefore the applied energy is concentrated over a smaller volume.
4, 23
). In
SPTA
. Ultra-
has been established for diagnostic ultrasound in general can­not be assumed for all applications. Consequently, the sonog­rapher should be familiar with potential bioeffects and their physical mechanisms in order to avoid any risks that may be associated with the examination.
sion in different tissues and organs varies substantially, and so the cooling component in different tissues can only be roughly estimated. With brief ultrasound exposure, perfusion con­tributes little to tissue cooling
Hyperthermia. An increase in temperature may be a risk, since cell division is inhibited above 30 C, and temperatures higher than 41 C can cause cell death with sufficiently prolonged ex­posure. A key factor besides the amount and duration of tissue heating is the sensitivity of the affected tissue fetal brain is extremely sensitive to temperature increases. Al­though brain tissue has a low absorption coefficient, secondary heating from the more strongly absorbing cranial bone must also be taken into that whole-body hyperthermia above 41⬚C can be teratogenic, depending on the duration of exposure
account
5
.
3
. The embryo-
4
. Animal experiments have shown
11, 2 4
.
Cavitation
Physical and Technical Principles
Absorption coefficient. Sound absorption in biological tissues
increases with the protein content of the tissue body fluids such as urine and amniotic fluid absorb almost no sound energy, and their absorption coefficient
0.002 and 0.003 dB cm kidney, and muscle have approximately the same absorption (
α = 0.4–0.6 dB cm
sorber ( risk. The secondary heating of soft tissues in close proximity to bone should also be considered when evaluating risk
Acoustically, the human embryo is equivalent to soft tissue. By the end of the first trimester, however, absorption increases owing to early mineralization of the bone. Thereafter it con­tinues to increase during the rest of the fetal period.
Heat conduction and perfusion. Heat conduction and blood flow contribute to the elimination of tissue heat. They are the major factors that determine the final temperature that is reached, especially on prolonged exposure. Whereas heat con­duction can be calculated to a good approximation, the perfu-
α =5–10dBcm
–1
MHz. Soft tissues such as brain, liver,
–1
MHz). Bone is by far the strongest ab-
–1
MHz), placing it at greatest thermal
19
. Water and
α is between
21, 25
Noninertial and inertial cavitation. Cavitation refers to the sound-induced formation and dynamic behavior of cavities and gas bubbles, which can produce a variety of physical, chemical, and biological effects in the medium cavitation occur: noninertial cavitation, which describes the prolonged resonant oscillations of preexisting gas micro­bubbles that do not collapse; and inertial cavitation, in which preexisting microbubbles (“cavitation nuclei”) rapidly expand in the rarefaction phase of the ultrasound wave and collapse again in the subsequent compression phase. This process can
.
generate local pressure amplitudes greater than 1000 MPa and local temperature peaks in excess of 1000 C. This may b e as­sociated with electrical arcing (“sonoluminescence”) and the formation of free radicals (OH properties that are indistinguishable from the radicals pro­duced by ionizing radiation stable and transient cavitation.
Sound-pressure amplitudes. Cavitation is a threshold effect.
The critical field parameter that determines whether cavita-
–,H+
,H2O2, etc.) with mutagenic
15, 25
. A continuum exists between
25
. Two types of
29
Safety Aspects of Doppler and Color Doppler Sonography
tion occur is the negative sound-pressure amplitude p–(by consensus, this amplitude is stated as a positive value in MPa). Under certain physical conditions, however, the signs of the pressure values may change when an ultrasound wave is re­flected, so that the reflected wave acquires a negative pressure amplitude equal to the positive pressure amplitude of the inci­dent wave. As a result of this, the positive pressure amplitude may also be a relevant quantity. The cavitation threshold in­creases at higher frequencies.
Cavitation nuclei and cavitation threshold. Human tissue is fairly resistant to cavitation because generally it does not con­tain cavitation nuclei. Such nuclei are presumed to exist, however, following infusions, administration of echo contrast agent, gas gangrene infections, and open injuries. The negative sound-pressure amplitude necessar y for cavitation to develop is unknown. In shock-wave lithotripsy, cavitation is known to occur at negative pressures that exceed 10 MPa, but modern sonographic techniques employ fields whose negative peak pressure does not exceed 5 MPa (Fig. 2. large scale would seriously affect imaging because the induced bubbles would cause extreme increases in scattering and ab­sorption. So far these effects have not been observed in ultra-
2
sound examinations. Cavitation, then, is unlikely to be a risk factor in the ultrasound techniques in current use. It should be added, however, that echo contrast agents greatly lower the threshold for cavitation (see Ultrasound Contrast Agents, p. 31).
1a). Cavitation on a
a
12
(MPa)
p
8
4
0
b
100
)
2
10
(Wcm
SPTA
I
1
0.1
0.01
0.001
Fig. 2.1 Minimum (min), maximum (max) and mean values of the negative peak pressure p
Max Mean Min
Imag VagM Col
CW
Imag VagM M Col
(a) and I
PD
SPTA
Ther
Litho
CW Ther
PD Litho
intensity (b) of various European ultrasound scanners in various modes. The corresponding values of therapeutic and lithotripsy devices are also shown for comparison. Imag: B-mode imaging; VagM: M-mode with a vaginal probe; M: con­ventional M-mode; Col: color Doppler; CW: continuous-wave Doppler; PD: pulsed Doppler; Ther: physiotherapy; Litho: lithotripsy. (After ref­erence 6.)
30

Risk Assessment of Various Ultrasound Techniques

The risks associated with various ultrasound techniques de­pend chiefly on the likelihood of a biologically significant temperature rise and the potential for inducing cavitation. The field parameters of current diagnostic instruments that have a bearing on these effects are reviewed in Fig. 2.
1. Recent
measurements have essentially confirmed these values while also showing that newer systems tend to have higher field in­tensities but do not have higher negative peak pressures
14
.
Duplex Sonography
Duplex sonography combines B-mode imaging with pulsed Doppler flowmetry. B-mode provides the sectional image for locating the site of interestand positioning the Doppler sample volume, while flowmetry yields the desired information on blood flow velocity.
Heating. Biological risks have been ruled out for B-mode imag­ing. Flowmetry requires the additional use of a Doppler beam of greater pulse length and higher intensity (I sity of this stationary Doppler beam may fall within the upper range of therapeutic intensities. Animal experiments have
). The inten-
SPTA
shown that Doppler pulses can produce a biologically signifi­cant temperature rise in soft tissues, and that this effect can no longer be considered harmless can become particularly hazardous when bone located below the Doppler sample volume is exposed to the ultrasoundbeam, since bone has a high absorption capacity and undergoes rapid heating. It was found, for example, that when fetal guinea-pig brain was insonated within the intact skull, the temperature rise was greater near the bone on the opposite side of the skull than at the center of the brain conceivable that instruments with a high power output and prolonged scan time (30 seconds) can induce temperature rises during flowmetry that cannot be considered harmless.
Cavitation. Because the negative peak pressure is no greater in Doppler ultrasound than in B-mode and does not exceed 5 MPa, cavitation should not occur.
4, 23
(Fig. 2.2). The heating effect
4
. Based on these findings, it is
8
7
6
5
4
Temperature rise (°C)
3
2
1
0
0 60 120 180 240 300 360
Fig. 2.2 Temperature rise over exposure time at the center of iso­lated guinea-pig brain. Ultrasound field parameters: frequency 3.3 MHz, pulse duration
µs, PRF 4 kHz.
6.25 Upper curve: total power output 1120mW, I Lower curve: total power output 260 mW, I (After reference 4.)
SPTA
SPTA
1120 mW 260 mW
Time (s)
2.5 W/cm.
2.9 W/cm
Color Doppler
In color Doppler systems, blood flow velocities are determined not just in a single sample volume but over a larger “region of interest” as in B-mode imaging. These velocities are color-en­coded according to their direction and magnitude and are su­perimposed over the B-mode image.
Heating. Because the Doppler beam is not stationary as in du­plex sonography but is swept across the region of interest, the absorbed energy is distributed over a larger tissue volume, re­sulting in a substantially lower time average intensity. A useful rule of thumb is that the I
intensities in color Doppler are 10
SPTA
times higher than in conventional B-mode imaging, while the intensities in duplex scanning are 100 times higher (Fig. 2.
1b).
There are considerable differences, however, among different scanners and modes of operation. For example, as the region of interest is increasingly narrowed in its lateral dimension, the thermal effect is increased. When the region of interest is nar­rower than 2 mm, the thermal conditions are practically the same as in duplex scanning. As a general rule, however, the thermal risks associated with ordinary clinical use are small.
Cavitation. As in duplex scanning, cavitation should not occur in color Doppler imaging.
Power Doppler
The ultrasound exposure in power Doppler imaging is the same as in color Doppler. The processing mode is different, however, as power Doppler displays signal amplitudes rather
Risk Assessment of Various Ultrasound Techniques
than blood cell velocities. This type of signal processing can de­tect even slow flow velocities, providing a more accurate dis­play of vessel volumes. The thermal effect corresponds to that of color Doppler.
Color Velocity Imaging (CVI)
While this technique does not employ the Doppler principle, it does bear some similarities to color Doppler. The velocity and direction of blood flow are determined by the computer com­parison of successive B-mode echo images. The principle is that B-mode can detect acoustic inhomogeneities (“streaks”) in flowing blood and can track their motion. CVI requires a large computer capacity but uses low ultrasound intensities and is thermally harmless.
Transvaginal Scanning
A frequent concern in transvaginal scans is excessive exposure of the embryo owing to the proximity of the endovaginal probe. The opposite should be true, however, since the short range leads to less absorption, and therefore a lower power set­ting can safely be used. However, manufacturers have offset this advantage by using higher frequencies, which, while im­proving image resolution, also cause greater sound absorption that necessitates higher intensities. As a result, the average in­tensities in transvaginal B-mode scanning are somewhat higher than in abdominal scanning, but they are slightly lower in transvaginal pulsed Doppler
7
. The thermal exposure to the embryo or fetus is approximately the same as in transabdomi­nal sonography.
Ultrasound Contrast Agents
Gas-containing ultrasound contrast agents are being used with increasing frequency. They appear to be immunologically safe, and microbubble sizes on the order of 10 capillary blood flow.
Cavitation. At the same time, ultrasound contrast agents con­siderably lower the threshold for cavitation. It is unknown
whether administration of echo contrast agent is sufficient to
induce cavitation by diagnostic ultrasound. Stable cavitation
would increase the echogenicity of the contrast agent, but local hemolysis would also be expected to occur. So far, however, there have been no clinical reports of hemolytic effects tial cavitation would also be associated with free radical for­mation. This would contrast with the similar effect of ionizing radiation in the distribution pattern in the tissue: presumably the radicals would form chiefly in the intercellular compart­ment owing to the lower viscosity and would form intravascu­larly only at sites where the contrast agent is excited by ultra­sound. It is unclear at present whether this is apt to produce mutagenic effects. Since no experimental data have yet been published on this issue, a definitive risk assessment cannot be made. On the whole, the risk does not appear to be particularly
13
high
. There are no concrete objections to using ultrasound
µm do not affect
18
. Iner-
Physical and Technical Principles
31