Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5772_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
58 Мб
Скачать

1 Physical and Technical Principles of Color Doppler Sonography

R. Kubale and G. Hetzel

Historical Development

Duplex sonography and its offshoot, color duplex sonography, are relativelynew methods that combine the pulsed echo tech­nique of sectional image formation with the Doppler evalua­tion of blood flow. It is based on B-mode ultrasound, which permits an accurate morphological description of gynecologi­cal and obstetric findings as a result of increasingly higher res­olution and faster frame rates. The first attempts to measure blood flow date back to Satomura 1970s that Pourcelot Doppler to investigate cerebrovascular diseases. Since 1977,
1
11
was able to use continuous wave (CW)
12
. It was not until the early

B-Mode Sonography

Physical Principles of Echo Production
The ultrasound wave is a density and pressure wave that prop­agates through a medium, its behavior essentially conforming to the laws of sound and optics. Sound attenuation, sound re­flection, and resonance phenomena can be utilized as means of identifying defects in the insonated medium. The pulse-echo principle is most commonly used in medicine and will be dis­cussed below in greater detail.
Acoustic impedance. The echoes that are displayed in an ultra­sound image are based on reflections that occur when the sound wave travels through media of different acoustic im­pedance. Acoustic impedance (Z) is a material-dependent (i.e. tissue-dependent) quantity that is the product of the density of the material ( being, we shall assume that this velocity is a material- and tissue-independent constant (equation 1) Processes and Approaches, p. 25).
ρ) and the velocity of sound in it (c). For the time
(see New Technical
authors have describe d Doppler ultrasound applications in the abdomen and especially in obstetrics and gynecology During the past 15 years, these applications have become widely adopted and established through the use of color-flow techniques and ongoing refinements in instrumentation
Below we shall explore the physical and technical prin­ciples that are essential for optimally utilizing and interpreting B-mode and color Doppler sonography and avoiding misinter­pretations.
basic frequency (harmonic energy). These frequencies, which create unwanted signals in the conventional B-mode image (see Principle of Echo Detection and Scanning Techniques, p. 3), can be utilized for specific “harmonic” imaging. This tech­nique, called tissue harmonic imaging (THI), makes it possible to examine even patients who are technically difficult to scan (see New Techniques of Signal Acquisition and Processing, p. 25).
The acoustic impedance of tissues has a value similar to that of water. The exact value varies with the composition of the tissue (Table 1. sues are markedly different from the impedances of air and bone.
Table 1.1 Impedance values and degree of reflection in biological tis­sues
Medium Sound
1). Note that the acoustic impedances of soft tis-
Reflection
)
factor R (relative to H
velocity c (m/s)
Specific density
ρ (g/cm
Impedance Z (g cm
3
)
–2s–1
3, 13, 16
10
.
O)
2
.
Z = c ρ (1)
Because the sound wave consists of alternating high and low pressure peaks, the tissue that transmits the wave is tran­siently compressed, and the sound velocity changes: the “peaks” at a higher pressure move faster than the “valleys” at a low pressure. This phase-dependent change in sound velocity distorts and steepens the original sine wave, analogously to a
2
water wave breaking on the shore. As a result, the reflected wave acquires a component that is a multiple of the emitted
Water 1496 0.997 1.491 0.000 Fat 1476 0.928 1.370 0.042 Muscle 1568 1.058 1.660 0.054 Liver tissue 1570 1.055 1.660 0.054 Bone 3360 1.850 6.200 0.614 Air 331 0.0012 0.3972 0.999
Impedance (Z) is defined as the product of the specific density (ρ) of the tissue and the sound velocity (c). An incident ultrasoundbeam is partially reflected at interfaces between tissues of different impedance. The degree of the reflec­tion is measured by the reflection factor R in relation to water.
Reflection. Reflection occurs at every interface between two media that have different acoustic impedances. The reflected portion of the ultrasound wave (the echo) increases with the magnitude of the acoustic impedance difference. The amount of reflection that occurs at the interface between two different tissues with impedances Z reflection factor R (Table 1.
R =
Z Z
tissue1
tissue1
Z + Z
tissue2
tissue2
and Z
tissue1
1) as given by equation (2):
is described by the
tissue2
(2)
If the impedance mismatch between the tissues is small, the transmitted ultrasound wave will retain enough energy to pro­duce additional echoes in deeper tissue layers. But when a large impedance mismatch exists, such as tissue–air or tissue– bone interfaces, almost all of the incident sound energy is re­flected. Objects located behind such an interface cannot be
visualized with ultrasound.
B-Mode Sonography
Principles of Ultrasound Instrumentation
Principle of Echo Detection and Scanning
Techniques
Ultrasound pulses are generated and processed by means of piezoelectric elements that are assembled into a transducer. In principle, a transducer is both a transmitter and a receiver of sound. An electric pulse (pulse length 0.5–2 wave trains of 1 duration) excites the active elements of the transducer. The re­sulting mechanical vibration propagates through the tissue.
The mechanical vibration is reflected from a target object, and the returning sound wave induces the transducer elements to
generate an electric signal. The time (t) between the emission of the pulse and the reception of the echo is a measure of the distance (z) of the transducer elements from the reflecting ob­ject (pulse-echo principle)
(equation 4):
z =1/2 ct (4)
µs
Refraction. Reflection is strongly angle-dependent and is sub­ject to the same laws as optical refraction. If the sound strikes the interface between different tissues at a perpendicular angle, most of the reflected wave will return to the transducer. But if the sound strikes the interface at an oblique angle, only part of the reflected wave will reach the transducer, and the
wave will deviate from its original direction as it continues on
through the second medium (refraction).
Scattering. If the sound beam encounters a rough surface or small reflectors whose diameters are significantly smaller than the wavelength of the sound, the beam will be reflected or “scattered” in various directions. This occurs not just at inter­faces but everywhere in the tissue. Because of scattering, ho­mogeneous soft tissue is depicted as an interference pattern
with a coarse or fine texture, depending on the transducer
frequency (Speckle).
Absorption and attenuation. In addition to losses due to reflec­tion and scattering, some of the ultrasound energy is absorbed in a frequency-dependent fashion as it overcomes cohesion and relaxation forces in the medium. The amplitude of the sound pressure declines exponentially with the thickness of the medium traversed—i.e., it is attenuated. Attenuation refers to the relationship of the initial sound pressure to its pressure after the sound has propagated for a given distance. It depends on the distance traveled by the sound pulse, the frequency of the transducer, and a material-specific constant. The following simplified formula
(equation 3)
is valid for soft tissues and for
frequencies from 0.2 MHz to 100 MHz:
where c is the velocity of sound and t is the time from pulse
transmission to echo reception.
B-mode image. A two-dimensional sectional image is pro­duced when a number of adjacent ultrasound beams are trans­mitted and received in the same plane. In the B-mode tech­nique (for “brightness mode”), the echoes are displayed at a brightness that is proportional to their amplitude. The bright­ness-modulated signals that are received (ultrasound lines) are temporarily stored in a matrix. The contents of this matrix are transferred to a monitor at correlative sites and assembled into a geometrically correct ultrasound sectional image (Fig. 1.
1).
Physical and Technical Principles
Attenuation (dB) = frequency (MHz) distance traveled (cm) (3)
Fig. 1.1 Production of an ultrasound image, illustrated for a linear array transducer. Transverse abdominal scan through the liver, kid­neys, pancreas, and spine. Whena pulse is transmitted, portions of the ultrasound energy are reflected from the organ surfaces and from in-
terfaces within the organ parenchyma. These echoes are assigned to
specific locations based on the transit time for a known sound velocity.
The sum of the brightness-modulated points yields a sectional image
in which the liver, aorta, pancreas, and kidneys can be evaluated.
3
Physical and Technical Principles of Color Doppler Sonography
The first ultrasound imagers were based on the manual movement of a single transducer over the patient (compound scan). These were initially replaced by mechanical transducers in which an element was rotated to produce a fan-shaped beam pattern.
Transducer arrays. Most scanners today employ transducer ar­rays consisting of individual elements arranged in a closely spaced row. Each array has a certain number of elements, in some cases more than 196.The elements are fired (excited) in groups, each of which transmits and receives an ultrasound line. Starting on one side of the array, the edge line is acquired first. Adding an element on one side of the group and turning off an element on the other side, the active group is shifted across the array. As the next group transmits and receives, the tissue is scanned in a sequential fashion. As this process is con-
a Linear b Convex
1
B
A
AB AB
B
A
tinued, a sectional image is produced. If the elements in the array are arranged in a straight row, the transducer is classified as a linear array.If the elements are arranged along a curve, the unit is called a curved or convex array (Fig. 1.
2).
Electronic beam steering. A phased array scanner can scan the tissue in a sector- or trapezoid-shaped pattern. All of the el­ements in the array are active (group width m equals the num­ber of elements n). A sector-shaped scan is produced by elec- tronic b eam steering (Fig. 1.
3). The wavefront is steered by im-
posing time delays on the excitation pulses that are applied to the elements in the array. In this process the individual narrow elements (of width approximately 0.5
λ; λ = wavelength) be-
have like “Huygen point emitters,” meaning that each element is the source of a spherical wave in relation to the scan plane. All of the spherical waves add together to form a wavefront that is angled away from the array as determined by the time delay (Fig. 1.
3). To accomplish electronic steering in an array,
the elements must be spaced sufficiently close together, i.e., there must be a sufficiently fine element pitch or high element density. The individual elements must be so narrow that their directional characteristic is broad enough to allow adequate lateral signal transmission. The returning echoes must not be extinguished by the width of an individual element.
Recent technical developments include transducers with a trapezoidal beam pattern. The “trapezoid scan” is produced by sweeping a linear-array image field and represents the special case of a sector scan with a virtual pivot point located behind the transducer surface (Fig. 1.
2). Only linear arrays that have a
high element density can produce a trapezoid scan (see New Developments in Transducer Technology, p.23).
d Trapezeiumc Sector
Fig. 1.2 Functional principles of different transducer types. a In a linear array, the elements are fired in offset groups at times A and B, producing a rectangular beam pattern. b In a convex array, the transducer has a curved surface resulting in a fan-shaped beam. c Sector scan with a phased-array transducer. All the elements are ac-
4
tive. d Trapezoid scan. This unit is operated like a phased array with a “vir­tual pivot point” (VD) located behind the transducer.
VD
Trans­mitter
Delay circuits Pulses at
τ
1
τ
4
τ
8
τ
n
t0−τ
time t
1
t0−τ t
Array
1
4
8
n
n
Effective aperture A
ϕ sound direction c=τ
c
n
c=sound
velocity
Wavefront
s
0
Fig. 1.3 Phased array with electronic beam steering. The phased
τ
array transducer usesa variable electronic time delay (
τ
direction of the beam. If the time delay
is greater than τn, the excita-
i
) to changethe
i
tion pulses in the transducer will reach element n of the array first, before reaching element 1. The superimposed wave segments will combined to form a wavefront that is steered away from the trans­ducer axis by the angle
, depending on the time delay.
B-Mode Sonography
Transducer Selection: Penetration Depth, Beam
Pattern, and Resolution
The selection of a transducer or scanning technique (see Prin­ciple of Echo Detection and Scanning Techniques, p. 4) is guided by the nature of the desired information and by ana­tomical considerations. Important decision-making criteria are the size of the region under study, the necessary scanning depth, and the necessary resolution.
Field-of-view width and penetration depth. Since the linear array produces a wide near field, this type of array is preferred for scanning superficial regions. The convex array has a broader far field with good near-field resolution, making it an ideal compromise for abdominal scanning. The phased array has a sector format that is particularly useful for difficult imag­ing conditions, as in echocardiography for scanning deeper structures. The trapezoid scan has advantages similar to the convex scan in terms of good near-field resolution and far-field
width. An additional advantage, as in the phased array, is a small, straight transducer face that is easier to couple to the body surface for numerous applications.
Penetration depth and transmission frequency. The penetra­tion depth is the maximum distance between the sound trans­mitter (transducer) and the deepest site in the tissue from
which a reflection (echo) can still be acquired. Owing to the frequency-dependent absorption of ultrasound in tissue, the penetration depth is not only tissue-dependent but also frequency-dependent. For this reason, the selected transmis­sion frequency is a tradeoff between penetration depth and desired resolution and thus depends on the nature of the study. High frequencies are well suited for the high-resolution imag­ing of superficial structures, while low frequencies are better for imaging deeper structures. The maximum penetration depth is inversely proportional to the transmission frequency.
ab
Pulse length
Resolution
Slice thickness
Fig. 1.4 Definition of axial and lateral resolution. a Two interfaces or point reflectors located at different depths along
the beam axis can still be recognized as separate structures if the dis­tance between them is greater than one-half the spatial length of the
emitted pulse. Despite depth-dependent and frequency-dependent
tissue absorption with an increasing shift of the echoes toward lower
frequencies, the axial resolution of the image remains fairly constant. b Lateral resolution depends on the varying width of the ultrasound beam with increasing depth. Two adjacent point reflectors in the beam can still be identified as separate points if the distance between
them is greater than the “6-dB beam width” at the depth of interest.
This parameter is useful for describing the spatial resolution of two points in the transducer planeand for estimating the effect of point re-
flectors or interfaces that are directly adjacent to the imaged surface in the longitudinal direction (scan plane–thickness phenomenon).
Scan plane
Physical and Technical Principles
Resolution. Resolution defines the smallest distance between
two point reflectors that can still be recognized as separate points. Axial resolution denotes resolution along the axis of the beam. Lateral resolution describes resolution at right angles to the beam axis (Fig.1.
4).
Axial resolution. Axial resolution is determined chiefly by the spatial length of the emitted ultrasound pulse. Short pulse lengths combine d with a broadband transducer (see below) provide the best axial resolution. Typical values for axial reso­lution are 2
λ (λ = wavelength). Consequently, the axial resolu-
tion is usually better than the lateral resolution and increases
with the ultrasound frequency.
Lateral resolution and beam pattern. Lateral resolution is one of the basic parameters for describing the quality of image for­mation and affects the diagnostic performance of the system as a whole. Generally it is in the range of 3–5 wavelengths and de­pends on the ultrasound beam pattern.
Figure 1.
5 shows a beam pattern in which all of the intensi-
ties are normalized to the maximum sound pressure. It is eas­ier to analyze resolution in this type of diagram. Owing to in­terference effects, even a flat transducer produces a con-
Beam
Near field
Focal zone
Far field
Fig. 1.5 Near field, far field, and focal zone of an ultrasound beam (the area insonated by an emitted ultrasound pulse). The near field has an inhomogeneous interference structure that greatly limits the image quality. Resolution is best in the focal zone, where the beam is con­stricted. As the beam diverges again in the far field, there is an increas­ing deterioration of lateral resolution.
Transducer
Center line
Lateral cross sections throught the beam
Lateral
extent of a sound beam
Intensity at center line
Axial
5
Physical and Technical Principles of Color Doppler Sonography
stricted beam pattern with a natural focus. The lateral resolu­tion is best within the focal zone of the transducer. The depth of the focal zone is always frequency-dependent. The focal zone Z of a flat transducer (with a natural focus) is defined by equation (5):
Zf= kA2/λ (5)
where A is the aperture, or the active transducer length used for transmission and reception, and
λ is the frequency-depend-
ent wavelength.
The boundary between the focal zone of the transducer and the near field, which is marked by inhomogeneities (extinction phenomena), is defined as twice the value of the lateral resolu­tion at the focus. This near-fieldportion of the image is difficult or impossible to interpret because of artifacts. The far field begins at twice the value of the lateral resolution at the focus. The focus can be moved toward the transducer, thereby im­proving resolution, by the use of acoustic lenses, such as ones made of silicone rubber, by curving the transducer face (me­chanical focusing), or by electronic focusing. For a given frequency, it is advantageous to use the largest possible aper-
1
ture with suitable focusing. The advantages of a large aperture are greater penetration depth and better lateral resolution. The aperture can be reduced during reception by modifying the number of active elements (dynamic aperture) to minimize in­terference in the near field.
Resolution in the z axis. A third, frequently neglected, parame­ter is the resolution in the plane perpendicular to the trans-
a
Transmit focusing
Array elements
5
τ
3
3
τ
1
1
τ
0
0
1 3
s
5 s=path in the tissue Delay line
Focus
´ ´
´
b
Addition
Receive focusing
5
τ
3
3
τ
1
1
τ
0
0
1 3
5
Fig. 1.6 Transmit and receive focusing. a In transmit focusing, the excitation pulses to the array elements are delayed such that the outer crystals are fired before the inner ones, al­lowing the beam to be focused at different depths. b In receive focusing, echoes returning from the point designated as the focus arrive at element 0 first, where the return path is the short­est. If the echo-induced signals from that element are delayed by time
τ
, they can be time- and phase-matched to the signals from elements
0
5 and 5. If all of the echo-induced signals are delayed according to the
6
differences in their return times, all the elements will yield coherent signal components, so that adding the delayed signals will give an op­timum echo display for the focal point.
ducer (z-axis resolution). This parameter is described by the slice thickness from which reflectors and scatterers located outside the transducer plane can generate an echo in the trans-
f
ducer plane (see Artifacts and Pitfalls, p. 8). The slice thickness focus of a given transducer is dependent on the transmission frequency.
Signal Processing: Focusing, Preprocessing, and Postprocessing
Focusing. Every type of transducer that is to be connected to an
ultrasound imaging system requires special front-end circuitry that contains a transducer-specific interface. This circuitry in­cludes all the components that link the transducer elements to the processing channels. Most of the front-end electronics is for transmit and receive focusing. As noted above, the beam of a flat transducer converges to a natural focus. While this focus can be modified by acoustic lenses, true variable adjustment of the depth of the focal zone can be achieved only by electronic means. In all types of arrays, focusing in the scan plane is ac­complishing by means of variable electronic delay circuitry. Fo­cusing can be done in both the transmit and receive modes (Fig. 1.
6).
Receive focusing. In receive focusing, the echo signals that re­turn to the array first are delayed by the difference in arrival times from the more peripheral signals. The goal is to ensure a coherent display of all signals that are received from the varia­ble focal point.
Transmit focusing. The transmitted beam can be focused by delaying the excitation pulses in the array. Multizone focusing can be achieved only by scanning sequentially at different focal depths, since the focus of the wavefront can no longer be changed after the pulse has left the array. While transmit fo­cusing allows the user to define the number and location of the focal zones, multizone focusing results in a slower frame rate. For example, three focal zones triple the time required to generate an image, and so the frame rate is reduced by a factor of 3.
Dynamic focusing. Other focusing possibilities are available in the receive mode. By varying the delay times during reception
´ ´
´
of the echo signals, it is possible to keep the focus at the depth from which echoes are momentarily being received. This process, called dynamic focusing, does not reduce the frame rate and operates automatically throughout echo reception, without user input, according to an optimized algorithm. The electronic focusing described above is a feature of all electronic transducers.
Time gain compensation. Preprocessing refers to the part of the signal processing circuit in which scan data are manipu­lated before they reach the scan converter. It begins with the removal of unwanted effects caused by the sound attenuation and center frequency changes that occur with increasing depth. Echoes from deeper objects are attenuated more than shallower echoes on their way to the transducer. Time gain compensation (TGC) is a preprocessing technique that com­pensates for this attenuation by amplifying the echoes from
B-Mode Sonography
more distant reflectors as a function of their depth. Otherwise the echo signals from deeper sources would appear increas­ingly dark. Generally the TGC is set manually, because “homo­geneous image appearance” and “region of interest” are not quantities that can be preprogrammed. One way to optimize the setting is to take a subcostal transverse scan of the liver, for example, and then adjust the TGC curve until the brightness of the liver appears homogeneous. If scanning must be done through fluid (e.g., ascites) or through the uterus, the TGC should be reduced accordingly to avoid tissue effects of over-
whelming brightness.
Dynamic frequency filtering. Another preprocessing step is dy­namic frequency filtering. The frequency-dependent attenua­tion of the sound, with a depth-dependent spectral shift toward lower frequencies, makes this a necessary step. Con­current dynamic frequency filtering compensates for this shift effect and improves the signal-to-noise ratio in deeper regions.
Multifrequency and broadband technique. For these tech­niques to provide useful image improvement, the transducer must have the capacity for broadband excitation and broad­band signal processing. Other advantages of broadband trans­ducers are a shortened pulse length for better axial resolution (see Fig. 1. this way a single transducer can be used to image superficial structures at high resolution as well as structures located at greater depths. The multifrequency technique and broadband technique significantly expand the performance range of transducers and simplify the examination.
Edge enhancement. Other steps are demodulation and low­pass filtering and the ability to modify the shape of the echo signal in preprocessing. Edge enhancement, for example, can be used to steepen the slope of a change in signal level, thus providing better differentiation of echoes across an interface. (Edge enhancement is accomplished by high-pass filtering of the demodulated echo signals and is applied only to line-of­sight data.) Tissues appear to have a more fine-grained texture, and interfaces perpendicular to the beam direction (e.g., vessel
wall boundaries) are more clearly defined. The effect, then, is
an apparent improvement of axial resolution.
Dynamic range. The dynamic range makes it possible to match the contrast level of the ultrasound image to diagnostic re­quirements. From the large range of echo intensities that can be handled by modern systems, the user can select the dy­namic range from the smallest to the brightest visible B-mode echoes that appears best for the current application. A smaller dynamic range is useful for defining contours, measuring the cranial biparietal diameter, and vascular imaging. A higher dy­namic range is useful for evaluating the structure of tissues like the placenta, as this will define even the faintest tissue echoes.
Scan converter. The function of the scan converter is to convert the data acquired by various scanning techniques (e.g., in polar coordinates) into a correct spatial display pattern (e.g., in car­tesian coordinates) on the viewing screen. Because data read­out from the scan converter follows a television format (video) or computer format (e.g., SVGA), the ultrasound information
4) and the capacity for multifrequency excitation. In
requires intermediate (digital) storage on or before reaching the scan converter.
Correlation of the B-mode image. “Correlation” of the B-mode image is an averaging process that is carried out in the image memory. The time-averaging process can be structured in such a way that when new information is entered into memory, a portion of the stored (old) information from the previous image is retained in memory, and the new value is added to it.
The effect is similar to “afterglow” (persistence) on the monitor screen. Correlation can be useful in slowly changing image sequences (e.g., in the abdomen, peripheral vessels, etc.),
where it improves the signal-to-noise ratio. With fast-moving processes (e.g., fetal cardiac imaging), setting the correlation too high can cause a smearing of structures with loss of image quality.
Postprocessing. In postprocessing, the gray-scale depiction of the image information on the monitor can be manipulated by altering the data transfer characteristics between the scan con-
verter and the monitor. This may be done, for example, to en­hance or suppress low-level echoes in the final image.
Analysis of B-Mode Information and Artifacts
Optimizing the Examination and Analyzing the B-Mode Information
Homogeneous display of echo texture. After selecting the
transducer that is appropriate for the desired information and the region under study (see Transducer Selection: Penetration Depth, Beam Pattern, and Resolution, p. 5), the user adjusts the transmission power output, focal depth, receiver gain, and TGC curve to obtain a homogeneous display of echo texture within the area of interest. It is good practice to adjust these parame­ters in, say, a subcostal oblique scan of the liver. If fluid is pres­ent in the near field (e.g., ascites), the TGC should be reduced to avoid effects of too much brightness.
The principal criteria for evaluating organs such as the liver, spleen, kidneys, uterus, and ovaries are size, shape, delinea­tion, and changes in the basic echo pattern. The change may be hypoechoic, hyperechoic, or isoechoic. These pattern changes are helpful in detecting lesions such as hepatic, uterine and ovarian tumors or morphological changes in the placenta.
Artifacts and Pitfalls
Ultrasound image formation is based on certain physical as­sumptions, some idealized, such as those of constant sound
velocity and attenuation or the assumption of straight-line sound propagation in the body. Deviations from these assump­tions lead to artifacts. Some of these phenomena are diagnosti­cally useful and furnish additional information on tissue prop­erties, while others can lead to misinterpretation.
“Echo enhancement.“ Acoustic enhancement behind cysts is a diagnostically useful artifact. With the assumption of constant attenuation, the TGC curve produces a depth-dependent signal amplification in order to achieve uniform echo amplitudes
Physical and Technical Principles
7
Physical and Technical Principles of Color Doppler Sonography
from different depths (see Signal Processing: Focusing, Prepro­cessing, and Postprocessing, p. 6). However, the low degree of absorption and reflection in fluids causes too much signal gain, resulting in enhancement of echo brightness behind the fluid (Fig. 1.
7).
“Shadowing by a strong reflector.“ Another potentially useful artifact is the distal shadowing that occurs behind strong re­flectors. It may be caused by substances such as bone, stones, and air, which have an acoustic impedance very different from that of adjacent soft tissues (Table 1. impedance mismatch, much of the ultrasound energy is re­flected back toward the transducer. If the “reflector” is in­sonated at an approximate right angle, it produces a very bright echo (Fig. 1. the impression of a renal stone or similar lesion. However, if the reflector is insonated at a very oblique angle relative to the scan plane, most of the energy is reflected away.Shadowing oc­curs, but the reflector is not represented by a bright echo in the B-mode image.
Other shadowing effects are seen at the edge of circular ob­jects. If the beam intercepts an oblique interface at a tangential angle, deflection of the beam may occur, producing a distal
1
acoustic shadow. The interfaces are only partially defined.
Reverberations. Reverberations are artifacts caused by multi­ple reflections between two highly reflective interfaces. The prolonged transit time leads to faulty spatial encoding, creat­ing a cascade of parallel echo bands spaced at uniform inter­vals. Reverberation artifacts can be eliminated by insonating the reflecting surface at an oblique angle.
Comet-tail artifact. A comet-tail or resonance artifact is also caused by multiple reflections. When reflective interfaces are located very close together with weakly attenuating tissue be-
7). The associated distal shadow reinforces
1). Because of this large
tween them, they will generate many closely spaced echoes that form a long, bright stripe located distal to the actual reflec­tor. This can occur with minute air bubbles, for example, or when structures with a high sound velocity, such as cholesterol crystals or metallic foreign bodies, are encountered by the beam, setting up multiple internal reflections.
Mirror-image artifacts. Another special case is the mirror­image artifact. When an object is located in front of a strong re­flector, it will be “insonated” by that reflective interface. The returning echo is reflected back again and, because of its pro­longed transit time, forms a second, virtual image of the object behind the strong reflector (Fig. 1.
Slice-thickness artifacts. Slice-thickness artifacts appear as areas situated at right angles to the scan plane that appear to fill in the echo-free space within small, fluid-filled objects. The artifact can be eliminated by changing the beam direction or using a different probe.
Side-lobe artifacts. Signals from reflectors located outside the scan plane can produce “ghost echoes” due to side-lobe effects. Side lobes are sound pressure peaks that are located outside the main beam and are directed obliquely outward from it. These artifacts have become rare, however, as a result of refine­ments in equipment.
Refraction and diffraction artifacts. Refraction and diffraction artifacts occur at interfaces between media that have very different ultrasound velocities. An example is the interface be­tween bone (higher sound velocity) and silicone or other plas­tics (lower sound velocity).
7).
Fig. 1.7 B-mode artifacts: examples from abdominal imaging. a Hepatic cyst with posterior acoustic enhancement due to overamplification of the ultrasound signals. b Gallbladder with a stone (bright echo with a poste­rior acoustic shadow) and edge shadows (arrows). c Comet-tail artifact caused by multiple reflections from air. d Hepatic calcification with a mirror-image artifact projected on the opposite side of the diaphragm.
a
b
8
c
d

Duplex and Color Doppler Sonography

Physical Principles of Motion Detection
Duplex and Color Doppler Sonography
The techniques of Doppler shift acquisition are classified as
one-dimensional or two-dimensional.
Doppler effect. The oldest approach to motion detection is the Doppler technique. It is based on the Doppler effect, first de­scribed mathematically by the physicist Christian Johann Doppler (1803–1853). He observed that the light emitted from stars that were moving toward the earth was shifted toward the blue end of the spectrum, i.e., toward higher frequencies and shorter wavelengths (“blue shift”). Conversely, a red shift
was observed when the earth and star were moving apart. The same principle occurs in acoustics. The sound of an automobile engine changes its pitch when it moves relative to an observer.
The change is most noticeable at the moment when the vehicle is passing the observer.
Frequency shift. When the Doppler effect is applied in medical diagnostics, the moving objects are red blood cells that are moving toward or away from the transducer at various veloci­ties (Fig.1. from the moving blood cells show a frequency shift of
8). The echo signals reflected back to the transducer
f rela-
tive to the original transmitted frequency f. This shift depends on the magnitude and direction of the blood flow velocity and is described mathematically by the equation (6):
f =(2fv cos Θ)/c (6)
where transmitted frequency (f) and the frequency of the reflected signal. It is proportional to the velocity v of the blood cells and to the cosine of the angle beam and the vessel axis (Fig. 1. to the sound velocity in the medium being scanned (approxi­mately 1540m/s) and is usually within the audible range of 500–20,000 Hz in medical examinations. The factor 2 in the equation relates to the fact that the Doppler effect is operative twice in the pulse-echo technique: once when the pulse travels from the transducer to the moving blood cells, and again when the reflected signal returns to the receiver.
f represents the detectable Doppler shift between the
Θ between the Doppler sampling
8). It is inversely proportional
One-dimensional or spectral techniques. In the one-dimen­sional techniques, the vessel is scanned by one beam and the flow velocities are analyzed along the path of that beam. The information acquired is displayed as a time-dependent Dopp­ler spectrum. The spectral sampling region in one-dimensional techniques may cover the full depth of the beam (continuous-
wave or CW Doppler) or only a specific, selected site along the
path of the b eam (pulsed-wave or PW Doppler) (Fig. 1.
9).
Two-dimensional techniques. Two-dimensional techniques
yield a spatial distribution of Doppler information that is ac­quired from multiple sampling sites. Multiple beams are trans­mitted, the received signals are selectively analyzed according to their depth, and the result is superimposed on the B-mode image in color-encoded form. The size of the analyzed area may cover the entire B-mode image, or it may be confined to a
window called the “color box.“
The image may depict the mean velocity in the volume el­ement and the flow direction (color duplex sonography, CDS), or it may represent the sum of the squares of the amplitudes as a measure of the total number of moving red cells or the inten­sity of the blood flow (power Doppler imaging). Other tech­niques are based on the velocity-related position change in characteristic echoes (time-domain correlation or CVI)
1
or on the digital amplitude subtraction of two pulses emitted at different times (bi-flow technique)
18
. The latter technique
yields a display with high spatial resolution, similar to digital subtraction angiography. At present, both of these techniques can be used only for superficial scanning under favorable in­sonation conditions. They have little or no practical impor­tance in gynecology, so they need not be discussed here.
Technical Principles and Equipment Settings
One-Dimensional Techniques of Motion Detection (CW and PW Doppler)
Physical and Technical Principles
Transducer
Θ
Fig. 1.8 Principle of Doppler sonography.
v
Doppler effect
f=2(f/c)v cos
f=Doppler frequency
f=transmission frequency
c=sound velocity
v=flow velocity
=beam-vessel angle
Θ
CW Doppler. Continuous wave(CW) Doppler employstwo sep­arate crystals, one of which continuously transmits sound
waves while the other continuously receives the reflected echoes (Fig. 1.
9a). Because of its continuous operation, CW
Doppler cannot provide depth discrimination or spatial locali­zation of an echo. The advantage of the technique is that even
Θ
very high flow velocities can be definitively analyzed (see below).
PW Doppler. Pulsed wave (PW) Doppler makes it possible to sample flow velocities of selected areas. The same crystals are used for transmitting and receiving, and short pulses are trans­mitted into the body as in B-mode imaging. A designated time interval is allowed for the pulse to travel to the desired Doppler
9
sampling site and return to the transducer, at which point the
gate for echo reception is briefly opened (Fig. 1.
9b). The ex-
Physical and Technical Principles of Color Doppler Sonography
1
a. CW Doppler
TM: Transmitter R: Receiver
Transducer
TM
b. PW Doppler
Transmit time: T Reception time: t
Transducer
Amplitude
Time
R
R
T
t
R
Transmit
T
Receive
PRF
2T t
max
1
=
T
Amplitude
Time
Fig. 1.10 Principle of the fast Fourier transform (FFT). Just as the human ear is able to break down a musical chord into its separate notes (e.g., G, B, and D), the FFT is designed to resolve a frequency
Amplitude
++
Time
Amplitude
Frequency
Amplitude
Time
spectrum into its constituent frequencies. The result is displayed as an
t
R
t
amplitude–frequency distribution (bottom graph).
10
Fig. 1.9 One-dimensional techniques: principles of CW and PW Doppler.
a CW Doppler. The transducer contains two crystals, each of which transmits and receives continuously. This instrument can detect even
high Doppler frequency shifts with no aliasing or other artifacts. The
disadvantage is that all the vessels insonated by the beam yield one
mixed signal, makingit impossible to assign thesignal to a specific ves-
sel.
b PW Doppler. The same piezoelectric crystals are used to transmit
and receive, and are driven in pulses. The echoes are received at a selected time interval (range gate), which depends on the transit time
T that the signal takes to reach the transducer from the desired depth.
The reception time t
determines the size of the sample volume that
R
will be analyzed.
aminer selects the size and depth of the range gate on the basis of visual guidance in the B-mode or color Doppler image.
Demodulation. The first step in obtaining velocity information from the echo signal is demodulation, which separates the Doppler signal (Doppler frequency
f) from the reference or
transmitted frequency. Demodulation is carried out in a phase detector by combining the echo signal with a reference signal, followed by rectification and low-pass filtering.
Spectral analysis. The echo returning from the blood to the transducer contains a mixture of frequencies that depend on the velocity distribution of the individual blood cells. In the next step, therefore, the frequencies contained in the Doppler signal are analyzed according to their distribution and ampli­tudes. This spectral analysis is accomplished by a mathemati­cal algorithm called the fast Fourier transform (FFT). Just as the human ear can resolve a musical chord into its constituent
notes (Fig. 1.
10), the FFT processor can break down the de-
modulated signal into its underlying (harmonic) Doppler com­ponents and their amplitudes.
The frequency resolution and temporal resolution of the spectrum depend on the number of flow samples (generally between 64 and 256) that are used in computing the FFT algo­rithm. The greater the number of samples that go into the com­putation, the higher the frequency resolution. With 128-point FFT, the information from 128 consecutive Doppler signals (transmitted pulses) is broken down into the underlying frequencies. With the next transmitted pulse, the calculation results in a consecutive complete spectrum with 128 frequen­cies.
Doppler spectrum. The result of the analysis is displayed either as an amplitude–frequency curve for a given point in time (Fig. 1.
10) or as a time-varying frequency spectrum (Doppler
spectrum) that covers one or more cardiac cycles. In the latter display,which is commonly used in medicine, the amplitude of the Doppler frequencies at each point in time is represented by the brightness of the pixels (Fig. 1.
11).
The Doppler spectrum forms the basis for a more detailed analysis, making it possible to determine and quantitate flow velocities and volumes and also to calculate resistance values that can demonstrate prestenotic or poststenotic changes in the vascular bed (see Analysis of Doppler Information and Arti­facts, p. 19).
Pulse repetition frequency. One disadvantage of pulsed Dopp­ler flowmetry is that the transit time of the pulses (T) in the ab­domen ranges from approximately 0.03 to 0.26 ms, depending on the scanning depth (e.g. 3–20 cm). This limits the shortest possible time interval between two successive pulse transmis-
Duplex and Color Doppler Sonography
Fig. 1.11 Time-varying frequency spectrum (Doppler spectrum). The spectrum was sampled from an artery over two cardiac cycles and plotted by FFT analysis. The horizontal axis is time, and the vertical axis is the Doppler frequencyshift or, after angle correction,the flow veloc­ity. The brightness of the image points in the spectrum is proportional
to the amplitude of the particular Doppler shift. In the first cardiac cycle the peak velocities (bright line at the top of the trace) and the mean velocities (dark line at the center) are shown as curves overtime course.
Relevant instrument settings: Doppler frequency: 5.5 MHz Pulse repetition frequency: 520 Hz
Wall filter: 50 Hz
Angle correction: 63
sequent scan at the same location (Fig.1.
12). The result is a cor-
relation vector that is characterized by its amplitude and phase difference. The phase difference at each image point in the color line supplies the mean frequency shift, and this value is directly proportional to the mean flow velocity. The variance can also be determined and displayed to indicate turbulence.
Kasai and Nameka first used the mathematical process of autocorrelation to calculate blood flow in 1982.An autocorre­lation A(t) of the function g(t) is described by
+
A(t)=
a
RF echo Digitized signal I/Q v
g′(τ) g(τ + t) dt (8)
Processing of demodulated signals
Quadrature phase detector
Flow analyzer
Fast Fourier transform Autocorrelation Autoregression
equation (8):
Flow
ana-
lyzer
Image
me-
mory
Physical and Technical Principles
sions. Hence, the rate at which the separate pulses are trans­mitted (pulse repetition frequency, PRF, equation 7)
cannot exceed 1/T (e.g., 25.6 or 3.8 kHz) without causing errors of depth discrimination.
PRF =
1
T
c
=
(7)
2s
In equation (7), s is the depth (m), PRF is the pulse repetition frequency (Hz), and c is the sound velocity (1540 m/s).
Two-Dimensional Techniques of Motion Detection
(Color and Power Doppler)
Color Duplex Sonography
Whereas pulsed spectral Doppler displays the velocity dis­tribution over time at a designated sampling site, color duplex scanning analyzes the flow velocities at a number of sites that are distributed over all or part of the sectional image. Color­flow imaging, unlike spectral Doppler, uses autocorrelation rather than Fourier analysis.
Autocorrelation. Recall that in FFT analysis each Doppler signal is sampled from the gate up to 256 times for each time interval in the spectrum, and the corresponding number of frequencies are calculated for that location. By contrast, autocorrelation in color Doppler imaging compares (correlates) all the demodu­lated Doppler signals along a color scan line that are sampled at a given time interval after transmission with those of the sub-
b
Autocorrelation method
I(Q)
T
=1/PRF
Time
1. First scan sequence
2. Second scan sequence
2 ... 16 Sampling pairs
∆Θ
∆Θ
1
2
Mean phase difference
∆Θ
mean flow velocity v
Fig. 1.12 Two-dimensional Doppler techniques. Principle of signal analysis (illustrated for autocorrelation). a Common principle of signal processing in all autocorrelation, auto­regression, and FFT methods. The high-frequency signal is demodu­lated in the quadrature phase detector. This demodulation is done
twice, with a phase shift of 90(I, Q). After it has been filtered by suita-
ble wall filters, the signal (which is digitized in modern systems) is fed
to the actual analyzer. A fast Fourier transform, autocorrelation, or au-
toregression is carried out, depending on the method used. The result is usually stored as color information in image memory. b In the most commonly used autocorrelation method, pairs of suc­cessive samplings are tested for their phase difference in a flow ana­lyzer. This difference is a measure of the flow velocity v. Time averag­ing the phase differences in the selected scan sequence yields the mean velocity v, which is color-encoded and superimposed on the B­mode image. (After Haerten 1993.
5
)
11