FIGURE 403 Three-dimensional (3D) reconstruction from spiral
computed tomography (CT) scan of abdominal aortic aneurysm (AAA).
Note tortuous iliac arteries (A). Degree of tortuosity may be underestimated
in direct anterior-posterior view, but on a more oblique angle (B), a more
significant degree of tortuosity is visible.
Many of the features necessary for an adequate aortic neck are
also necessary for the distal landing zone. Presence of thrombus,
calcification, and tortuosity can significantly hinder the iliac limb
seal. Ectatic or aneurysmal iliac arteries obviously affect the ability
of the graft to seal against the iliac limb. Most available endograft
systems require at least a 15-mm segment of iliac artery to be of
adequate caliber and free of significant disease to obtain a distal
seal. If this is not present, adjunct interventions can be performed
to assist in placing the device (i.e., iliac artery conduit placement,
coil embolization of the internal iliac artery [IIA]). Management of
these complicated situations is discussed in more detail later.
Endograft Design
Endograft design can greatly affect the ability of the device to be
placed in patients, particularly in those with complex anatomy.
Alterations in graft characteristics are what distinguish one manufacturer's device from another. Some key elements in endograft
design are outlined in the following discussions.
Delivery System
Standard endograft insertion involves placement of the device
through an arteriotomy in the common femoral artery, from where
the graft traverses the external iliac and common iliac arteries
(CIAs). The ability to deliver the endograft safely and effectively
in this fashion is a prerequisite for effective repair. Three factors
are important determinants of device delivery.
DELIVERY SYSTEM SIZE
With the placement of most endografts through the iliofemoral arterial
system, any site along this pathway can represent a size limitation, the
most common of which is the external iliac artery (EIA). Inadequate
diameter or presence of extensive calcifications can exclude standard endograft placement. It is intuitive that the size of the delivery
system cannot be larger than the size of the iliac arteries it traverses.
Most sheaths are sized based on inner diameter, so knowledge of the
outer diameter of the sheaths is necessary for safe graft placement.
Different manufacturers’ devices have different size measurements
for the delivery systems, so one device may be suitable for placement,
whereas another is not. Most delivery systems easily traverse an
iliofemoral segment of 7 to 8 mm in diameter (or a sheath that does
not exceed 21 F), although several designs that provide a lowerprofile system are currently in clinical trials in the United States.
FLEXIBILITY
Tortuosity, another anatomical variant, affects the ability to adequately deliver the endograft system. Tortuous iliac vessels can be
“straightened” with the use of stiff guidewires, but this is not always
25
possible or desirable. The ideal delivery system easily traverses
these arteries on the basis of an intrinsic degree of flexibility. Again,
different delivery systems have different abilities to track through
tortuous iliac arteries, and some may be more successfully placed
than others in this anatomical variant. Delivery systems composed
of long, flexible, tapered tips pass more easily than those with short,
stiff, blunt tips. In addition, other aspects of device construction,
such as metallic struts that provide columnar strength, increase
device rigidity and limit use in tortuous vessels.
25
DELIVERABILITY
A number of features have been noted to affect the deliverability
of endograft devices. As stated previously, long, flexible, tapered
tips pass more easily than short, blunt, stiff ones. This allows for
easier maneuverability through tortuous vessels, as well as past
sites of narrowing. Larger-caliber devices are also more difficult
to deliver, particularly in patients with smaller-diameter arteries.
25
Some delivery systems allow for placement of the endograft system through alternate sheaths, whereas other systems necessitate
use of the manufacturer's own delivery system. This can greatly
affect placement of specific endografts in specific anatomical
variants. A thorough understanding of the patient's arterial anatomy and the limitations of different endograft systems are important. The complexity of the delivery system also affects the ease
with which it is placed. Some devices generally provide a simple
maneuver to deploy the graft, whereas others have several complicated steps.
Endograft Features
The ideal endograft should be flexible enough to maneuver
through tortuous and angulated vessels but also rigid enough to
prevent kinking. It should have a low profile (having a small external diameter) that would allow it to be placed through as small an
arteriotomy as possible. Two general classifications of endografts
exist: unibody and modular. A unibody device is a single-piece
graft—including the main body and both limbs. Although this
decreases the risk of endoleaks at the graft-graft interface, the unibody design often requires a larger delivery system, and sizing can
be more difficult. The modular system includes endografts that are
composed of two to three pieces. Generally there is a main body
that may have one attached limb and one or two docking limbs.
These devices can be introduced through smaller delivery systems
and offer a greater degree of flexibility with regard to placement.
With multiple sites of graft-graft interface, however, there is an
increased risk of endoleak, as explained later.
Graft material is variable and can range from thin-walled polytetrafluorethylene (PTFE) to polyester. The graft material is typically
supported by a metal framework that is commonly stainless steel,
its modified version Elgiloy, or nitinol. The graft support can be
placed inside the graft material (endoskeleton) or outside the graft
(exoskeleton). Grafts can be fully supported, having stent material
throughout, or only partially supported, with aspects of the device
composed only of graft material and no metal. The graft skeleton
provides several key elements to endograft make-up. First, it assists
in graft fixation and in obtaining a seal. These stents provide some
degree of radial force that helps provide a seal, as well as providing
a point of fixation. Some devices have hooks or barbs in the proximal aspect of the skeleton that help anchor the graft onto the aortic wall and prevent migration. In addition, some devices employ
a metal framework that extends above the fabric and is used to
engage the aorta in the pararenal or suprarenal location. The second function of the skeleton is to provide columnar strength, which
may prevent graft migration. The skeleton can also prevent kinking and occlusion of limbs as they traverse the aortoiliac anatomy.
Lack of stents, however, may allow a graft to adapt more readily
to morphological changes without dislocation of attachment sites.
The interplay of the stent and fabric materials can lead to eventual
erosion of the fabric.
495
CH
40
EndovAsCulAR THERAPy foR AbdominAl AoRTiC AnEuRysms

496
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
TABLE 40-1 Description of a Variety of Commercially Available Endografts
COMPANY
Cook
CH
(Bloomington,
40
Ind.)
Gore
(Flagstaff, Ariz.)
Endologix
(Irvine, Calif.)
Medtronic
(Minneapolis,
Minn.)
Medtronic
(Minneapolis,
Minn.)
ePTFE, expanded polytetrafluoroethylene; FDA, U.S. Food and Drug Administration; OD, outer diameter.
INITIAL FDA
DEVICE
Zenith May 2003 Self-expanding Dacron Stainless steel Modular 21F-26F Suprarenal
Excluder November
Powerlink October 2004 Self-expanding High-density
AneuRx September
Talent June 2008 Self-expanding Dacron Nitinol Modular 22F-24F Suprarenal
APPROVAL
DAT E
2002
1999
DEPLOYMENT
TYPE
Self-expanding ePTFE Nitinol Modular 20F-23F Infrarenal
Self-expanding Dacron Nitinol Modular 21F Infrarenal
GRAFT
MATERIAL
ePTFE
STENT
MATERIAL
Cobalt chromium
alloy
BIFURCATED
DEVICE
DESIGN
Unibody 21F Anatomical (with
MAIN BODY
SHEATH SIZE
OD
FIXATION
either suprarenal or
infrarenal proximal
orientation)
Specific Grafts
Various endografts are currently commercially available or in clinical trials in the United States. A brief description of the currently
commercially available endograft systems (in the United States) is
outlined in
Table 40-1 and depicted in Figure 40-4.
Graft Placement and Postoperative
Management
Once the patient is deemed an endograft candidate, the best
graft has been chosen, and the device properly sized, the patient
can undergo implantation. The majority of endografts are placed
through the femoral arteries that have been operatively exposed.
The majority of surgeons prefer the use of the transverse incision
as it associated with a lower rate of wound complications (12.7% in
transverse incision vs. 47.5% in vertical incisions).
access for EVAR is growing in popularity, and its use will become
even more widespread with the further development of lowprofile devices. Suture-mediated closure devices facilitate this
process, and using a “preclose” technique has been described to
allow closure of sheaths as large as 24 F.27 Use of this procedure has
been associated with 70% to 100% technical success, and immediate failures mandate surgical exploration of the femoral artery.
Prospective analysis has demonstrated that use of a percutaneous approach may shorten operating times and reduce the rate
of wound-related complications, without a significant increase in
overall procedural cost.
27–29
The aorta is then cannulated with a
guidewire and catheter. Small boluses of contrast agent are delivered to further define the anatomy and localize the renal arteries.
With an angulated aorta, it is important to remember that the best
view of the renal arteries and visualization of the fixation zone may
not be in a direct anterior-posterior plane but at a more cranialcaudal angle. The device is then generally advanced over a stiff
guidewire and correctly positioned to allow the most extensive
coverage within the aortic neck without intruding on the orifice of
the renal arteries. Each device has its own unique instruction for
actual deployment. Once the main body and ipsilateral limb have
been placed, the contralateral limb has to be placed. The sequence
of events for this varies depending on graft design—whether unibody or modular.
Recovery following EVAR is generally rapid and uncomplicated, and most patients are discharged home on the first or second postoperative day. Return to activities of daily living has been
shown to be quicker following endovascular repair than open
surgery. In addition, most patients report less postoperative pain.
Aortic remodeling following EVAR, however, is a slow process that
26
Percutaneous
continues for several years. Anatomical changes in the native vessel, particularly at the proximal neck, can cause conformational
changes in the implanted device that mandate close follow-up. In
addition, late failures have been identified that have required reintervention.
30
Given these facts, routine surveillance following EVAR
is universally recommended, although there are no standard regimens, and the requirements of a standard intensive regimen are
debated. Nordon et al. performed a meta-analysis evaluating secondary intervention rates based on contemporary graft implants.
31
Their findings demonstrated that surveillance imaging alone initiated the secondary intervention in 1.4% to 9% of cases. Over 90%
of EVAR cases, however, received no benefit from surveillance
scans. Based on these findings, the group recommended that surveillance should be directed toward those patients identified as
having a high risk for postoperative complications. Identification of
this group, however, is not obvious but may be necessary in patients
with complicated aortic neck anatomy or in patients in whom the
device was used outside of the indications for use (IFU).
Contrast-enhanced CT is the most widely used modality for
follow-up after EVAR. It is widely available, has rapid data acquisition, reproducibility, and is uniform across institutions. The major
concerns associated with this modality are use of a contrast agent
and the potential associated nephrotoxicity, radiation exposure,
and cost. It is considered the gold standard for assessing aortic
diameter, with nearly 100% sensitivity and specificity. Sensitivity
and specificity rates for endoleak detection with CT are better
than those with conventional angiography: 92% and 90% for CT vs.
63% and 77% for angiography, respectively.
32–34
Triphasic CT (noncontrasted phase, arterial phase, and delayed phase) results in the
greatest amount of information but at the cost of increased radiation exposure. Unenhanced CT imaging is useful for differentiation of endoleaks from calcifications from the metallic portion of
a stent graft, and can help detect small perigraft leaks better than
arterial-phase images. Use of arterial phase alone has a lower diagnostic value than combined arterial and delayed-phase scanning.
35
Repeated CT scanning subjects the patient to potential carcinogenic risks associated with ionizing radiation exposure. The
estimated lifetime attributable risk of death from cancer following an abdominal CT scan in a patient older than 50 years of age
is 0.02%.
31,36
Although this effect in itself is small, the cumulative
effects over time with repeat imaging can be significant. Repetitive
use of iodinated contrast can have a cumulative deleterious effect
on renal function, especially in the elderly and those patients with
preexisting renal impairment.
37
Given this, as well as the expense,
the use of alternate modes of surveillance has been evaluated.
Magnetic resonance imaging (MRI) and MRA provide much of
the same imaging information that can be acquired by CT scanning.

497
Multiple format MR images (T1, T2, gadolinium-enhanced) can be
viewed in 2D and be reformatted into 3D volumes, allowing dimensional measurements, assessment of luminal patency, device positioning, and detecting the presence of an endoleak. Limitations
of MRA/MRI include potential magnet-induced
heating or motion, which may prevent imaging in the immediate
postimplant time period. In addition, postimplant artifacts, in particular with ferromagnetic metallic stents, will limit morphological
assessments. Furthermore, there is a risk of nephrogenic systemic
fibrosis associated with gadolinium contrast use in patients with
A
renal insufficiency.
sure to ionizing radiation and low nephrotoxicity of MR contrast
38
Benefits of MRI are related to the lack of expo-
medium. Disadvantages of MRI are its lack of wide availability,
longer procedure time than CT, patient claustrophobia, and contraindications for patients with cardiac pacemakers.
Color duplex ultrasonography (US) is a convenient, noninvasive,
inexpensive portable means of postimplant surveillance. Its reliability as a useful surveillance tool, however, is still debated. Grayscale
US is accurate for measurement of aortic aneurysm diameter.
Endoleak detection by US requires color duplex. The reported
specificity rates of color duplex US for endoleak detection are high
(89%-97%), but the sensitivity and diagnostic power of color duplex
US for endoleak detection compared with CT is still debated.
B
use of duplex US to detect an endoleak has a sensitivity of approximately 77%, with a specificity approaching 94%.
US contrast agents increases sensitivity to 98%, with no significant
change in specificity.
40
Contrast agents are useful in identifying slow
leaks that are not readily discernible on CT.
direction of the endoleak is also an advantage of US that is not easily discernable with CT.
43
Presence of a “to-and-fro” flow pattern is
associated with spontaneous closure of the endoleak, whereas a
monophasic or biphasic waveform is consistent with endoleak
persistence.
43
Limitations of color duplex US include its operator
dependence, variation based on patient physical size, and the need
for optimal patient preparation. The substitution of duplex US imag-
C
ing for CT, however, may result in long-term cost savings.
in vivo metallic
40
The addition of
41,42
Detection of flow
44
39
The
CH
40
EndovAsCulAR THERAPy foR AbdominAl AoRTiC AnEuRysms
D
E
FIGURE 404 Several endograft systems illustrating different features.
A, Zenith endograft (Cook Inc., Bloomington, Ind.) represents a three-piece modular
system with a main body and separate bilateral limbs. This graft design uses a
bare suprarenal stent and internal stents at the sealing zones and is otherwise
supported by a stainless steel Z-stent exoskeleton. B, Powerlink graft (Endologix,
Irvine, Calif.) constructed of expanded polytetrafluoroethylene (ePTFE) and a cobalt
alloy skeleton. C, AneuRx stent graft system (Medtronic AVE, Santa Rosa, Calif.)
composed of nitinol exoskeleton. D, Excluder endograft (WL Gore and Associates,
Flagstaff, Ariz.), which represents a two-piece modular system. Graft is constructed
from ePTFE and is fully supported by a nitinol exoskeleton. E, Talent endograft
system (Medtronic AVE, Santa Rosa, Calif.) represents a two-piece modular system
composed of a suprarenal bare stent and then a nitinol endoskeleton.
Problems with Endografting
and Management
Various problems can arise in the planning and placement of
abdominal aortic endografts. Once the grafts are in place, several
complications can arise over time that may require intervention to
prevent subsequent expansion and possible rupture of the previously excluded aneurysm. In the following section, several of the
more common problems that occur following endograft placement are outlined.
Iliac Artery Disease
When iliac artery disease is present, whether it be aneurysmal
disease, atherosclerotic disease, or severe tortuosity, the use of an
iliac conduit can provide a safe route to deliver the endograft.
In cases of iliac artery lumen narrowing resulting from atherosclerotic disease or increased vessel tortuosity, advancement
of the device, despite the presence of resistance, can result in
rupture of the iliac artery. Iliac artery rupture has been reported
in 1% to 2% of cases.
46,47
To circumvent prohibitive iliac artery
anatomy, an iliac conduit can be used. An iliac conduit involves
suturing a prosthetic graft (generally 8-10 mm in diameter) to the
mid–CIA even if it is aneurysmal. This can be done in an endto-end or end-to-side fashion, although the latter often provides
a greater lumen for passage of the device. The device is placed
through the prosthetic graft, and the iliac limb of the endovascular graft traverses the CIA and anastomosis and seals within the
conduit. The distal end of the graft is tunneled along the natural course of the iliac artery and anastomosed to the femoral
artery. The distal end of the CIA is oversewn to allow retrograde
flow through the EIA to supply the ipsilateral hypogastric artery.
Alternatively, the hypogastric artery can be anastomosed directly
to the conduit.
45

498
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
Iliac artery ectasia or aneurysms can present a problem in
obtaining a distal seal. Enlarged CIAs are present in up to 30% of
patients presenting for endovascular aneurysm repair.
available endografts do not allow for ectatic or aneurysmal common iliac arteries, so the distal seal may have to be obtained within
the external iliac artery, which is often of normal caliber. If the distal
CH
seal occurs in the external iliac artery, the hypogastric artery is gen-
40
erally sacrificed using coil embolization. Presence of a hypogastric artery aneurysm would necessitate the same approach. Rarely,
bilateral hypogastric artery embolization is required. Hypogastric
artery embolization can occur before aneurysm repair or concurrently. If bilateral embolization is planned, it is generally performed
in a staged fashion, although its occlusion is not always planned.
Hypogastric artery embolization is not without risk, and side
effects can occur in up to 50% of patients.
the predominant complaint after hypogastric artery occlusion. This
occurs in 12% to 50% of patients, but in most it generally resolves
after several months.
onset erectile dysfunction (ED).
48–53
Some 5% to 25% of men complain of new-
51,52
ischemia requiring resection are of theoretical concern, but they
have not been described in any of the larger series. Patients requiring embolization in the more distal branches of the hypogastric
artery (as might be done with the presence of an IIA aneurysm)
and those in whom coil placement was not adequately controlled
are at higher risk of developing pelvic symptoms.
gastric artery embolization has not been associated with increased
symptoms when compared with unilateral occlusion.
embolization of the IIA is not necessary if it is not aneurysmal. In
the face of CIA aneurysms, Wyers et al.54 have shown that if there
is a 5-mm neck of iliac artery proximal to the hypogastric artery in
addition to a 15-mm neck in the external iliac artery, coil embolization of the hypogastric artery is not necessary to obtain a distal
seal. This may be possible in up to two thirds of patients requiring
coverage of the hypogastric artery.
49
Buttock claudication is
Buttock ischemia and bowel
53
Bilateral hypo-
48–51
49,50,52
Many
FIGURE 405 Angiogram demonstrating a type 1 endoleak. Contrast
can be seen leaking around proximal part of graft and filling aneurysm sac.
Patient subsequently had a giant Palmaz stent placed in the aortic neck, and this
ameliorated the endoleak.
Coil
Endoleaks
An endoleak is the persistence of blood flow outside the endograft, but in the aneurysm sac. Endoleaks are classified according
to their etiology, and currently five types have been described
(
Table 40-2). A type 1 endoleak (Fig. 40-5) arises from inadequate
sealing at either the proximal aortic (allowing antegrade flow)
or distal iliac (allowing retrograde flow) attachment sites. Type
2 endoleaks (
Fig. 40-6) arise from patent branch vessels off of
the aortic sac that allow for retrograde flow into the aneurysm.
Such branches may include a patent lumbar or inferior mesenteric artery (IMA). Type 3 endoleaks develop from defects in the
fabric of the graft or at the junction zone between modular components. Type 4 endoleaks develop secondary to diffuse “leaking” of blood between the interstices of the fabric or where the
graft is sutured to a stent. Type 5 endoleaks describe a scenario in
which the aneurysm sac remains pressurized and the aneurysm
enlarges, but no demonstrable flow of blood into the sac can be
visualized on current imaging modalities. These may be due to
imaging that is not sophisticated enough to discern these leaks
TABLE 40-2 Endoleak Classification
ENDOLEAK
Type 1 Inadequate seal at aortic or
Type 2 Patent branches off aneurysm
Type 3 Fabric defects or component
Type 4 Leak at fabric interstices Antegrade
Type 5 Endotension No clear leak
CAUSE
iliac attachment sites
sac
junctions
BLOOD FLOW INTO
SAC
Antegrade or
retrograde
Retrograde
Antegrade
47,48
FIGURE 406 Computed tomography (CT) scan representative of a
type 2 endoleak. There is contrast within aneurysm sac but outside limbs of
endograft. This aneurysm had continued expansion until patient underwent
embolization of inferior mesenteric artery (IMA).
or due to intermittent episodes of leakage.
to the aneurysm sac causing it to continue to expand in this situation, has been termed endotension.
49
The pressure applied
50,51
Controversy with regard to
this concept exists, in particular with the ability of the thrombus
to transmit pressure to the aneurysm wall. It is argued that these
merely represent a type 1, 2, or 3 endoleak in which the defect
is large enough to allow blood to flow into the sac and transmit
pressure to the sac, but the exit site is not present or too small to
be detected.
Type 1 and type 3 endoleaks are associated with significant
risks of aneurysm enlargement and possible rupture, and these
should be treated.
55,56
This may be accomplished with placement
of an extension cuff limb over the site of the leak. If the leak is a
type 1 and the graft is juxtaposed to the inferior border of the renal
arteries, a large balloon-expandable stent can be placed in the
proximal aspect of the endograft. This provides increased radial
force, causing better juxtaposition of the graft and aortic wall, thus
ameliorating the leakage. If this is unsuccessful, open repair and
graft explantation are generally indicated. Fabric tears are easily managed if the site of the leak is localized. In these situations,
the tear can be covered with a cuff or extension. When it is more

diffuse, the entire endograft can be relined with a second endograft, or the device can be removed and the aneurysm repaired in
an open fashion.
Type 2 endoleaks are rarely associated with aneurysm rupture.
At least 10% to 15% of patients are identified with a type 2 endoleak
during follow-up.
58–61
Warfarin treatment is not associated with an
57
increased incidence of early or delayed postoperative endoleak,
but type 2 endoleaks are less likely to undergo spontaneous resolution in these patients.
62
Type 2 endoleaks are generally observed
unless they are associated with an increase in aneurysm size or
aortic pulsatility on physical examination. In these situations, arteriography is the next step to identify the source of the endoleak.
Superior mesenteric artery injection reveals retrograde IMA flow
as the source, whereas selective hypogastric artery injection demonstrates a lumbar artery filling the aneurysm. Super-selective arterial canalization can then be performed with embolization of the
feeding vessels. Another approach is through direct aneurysm sac
puncture.
63
With direct sac puncture, one can measure sac pressure
and inject the sac directly with contrast agent to precisely identify
the leak. The systolic sac pressure is related to the size of the leak,
and the pulse amplitude is related to the resistance of the outflow
vessels and sac compliance.
vessels can be directly accessed and embolized.
64
After localization of the leak, feeding
65
In addition, the
sac can be filled with substances such as coils, glue, or gel foam to
further prevent flow. Differences in outcomes between these two different approaches has not been realized.
66
Measurement of intrasac pressures may help determine whether
an endoleak is present at the time of the original surgery or if an
endoleak has been adequately treated if it has been approached
through direct sac puncture. In an ex vivo model of endoleaks,
Parodi et al.67 evaluated pressure changes in the aortic sac with
various types and sizes of endoleaks. In this model, sac pressures
were significantly higher than systemic pressures in the presence
of all endoleaks. This obviously places the aneurysm at significant
risk for rupture. Presence of patent side branches significantly
reduced the pressure within the sac, particularly the mean pressure and diastolic pressure. Clinically, persistent side branches augment the development of type 2 endoleaks and influence early
sac behavior.
68,69
Gawenda et al.70 evaluated the use of sac pressure
monitoring and found it helpful in the detection and treatment of
endoleaks. They noted, however, that intrasac pressure measurements did not correlate with AAA size change over ensuing followup. This may be an effective modality for monitoring aneurysms
after endograft exclusion once less invasive methods of pressure
measurement are developed.
Structural Failure
Material failure represents one of the most concerning problems
for potential failure of endograft placement. This is a difficult event
to identify because patients are often asymptomatic and may not
present with any acute changes in their endograft evaluation. Three
modes of structural failure have been described in aortic endografting and involve fabric erosion, suture disruption, and metal fracture.
Development of endoleaks secondary to graft erosion has been
documented with some first-generation endograft devices
(
Fig. 40-7). It has been speculated that the areas of graft erosion
are secondary to friction between the stent material and the fabric, which can be confounded by pulsation of the aorta. Predicting
the incidence of fabric fatigue is difficult, and although this does
occur in grafts placed by conventional open aneurysm repair, it
occurs much more rapidly and more commonly in the endograft
systems.
74,75
In several device designs, the graft fabric is attached to
the metal skeleton through the use of sutures. Disruption of these
sutures is believed to explain graft failure in some instances.
mechanism for suture failure is believed to be the same as for fabric
erosion—namely, motion of the stents with aortic pulsations causes
friction and wear of the sutures, with subsequent suture fracture.
The most common structural problem identified in aortic endo-
graft systems has been metallic stent fractures.
75
Stent and hook
76–78
71
72,73
The
FIGURE 407 Angiogram revealing a type 3 endoleak that developed
at site of a tear in graft fabric. This was a “homemade” aorto-uniiliac graft that
had been in place for approximately 5 years. Patient presented with new-onset
abdominal pain and had a computed tomography (CT) scan that revealed an
aneurysm sac that had significantly expanded in diameter. Tear was sealed by
placement of a new endograft.
fractures in the phase 1 trial of the Endovascular Technologie's graft
resulted in suspension of the program and redesign of the metallic attachment system.
79
In a review of 686 patients who underwent
endovascular aneurysm repair, Jacobs et al. identified 60 patients
who had material failure.75 Forty-three (72%) of these failures were
due to metallic stent fractures and occurred in various different
endografts with different stent composition. The cause of metal
failure has been attributed to stress fatigue and metal corrosion,
particularly in nitinol stents.
generation endografts and may reflect improved nitinol process-
81–83
ing.
Tortuosity of the arterial system can also stress the stent
80
Corrosion has not been seen in next-
graft system and lead to metal fracture. This has been reported in
the longitudinal bar of the Talent and Gore stent graft devices.
Limb Thrombosis
Endograft limb thrombosis after endovascular repair of infrarenal AAA is a recognized complication occurring in up to 11%
of patients.
place patients at increased risk for limb thrombosis. One reported
risk factor is the lack of device support. Although Carroccio et al.85
reported on the results of 351 bifurcated grafts with no significant
association between use of unsupported devices and graft thrombosis, others have suggested there is a significant relationship. Baum
et al.91 specifically evaluated the rates of graft limb kinking and
thrombosis between supported and unsupported abdominal aortic stent grafts. In total, 12% of the limbs in their series required an
intervention for kinking. In the supported limbs, 5% required subsequent placement of arterial stents; 2% required these for evidence of
kinking at the time of the initial operation, and 3% required stenting
in the postoperative period after the patients presented with limb
thrombosis. In the unsupported grafts, there was an intervention rate
of 44%. About half of these had an additional stent placed at the time
of the initial procedure, and the remainder had a subsequent stent
placed in the postoperative follow-up period, owing to limb thrombosis or severe stenosis.
84–90
Various underlying factors have been purported to
499
CH
40
EndovAsCulAR THERAPy foR AbdominAl AoRTiC AnEuRysms
75

500
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
Another factor increasing the risk of limb thrombosis is
oversizing of the iliac limb. Oversizing causes the graft material to have a significant amount of infolding, reducing the intraluminal dia meter.87 Along these lines, significant intraluminal
vessel narrowing from underlying atherosclerotic disease or tortuosity can result in flow abnormalities and eventually cause
CH
graft limb thrombosis.
40
has also been described as a risk factor for developing limb
thrombosis.
85
It was believed that transition into the EIA caused
91
Extension of the graft limb into the EIA
both a significant reduction in arterial diameter and a kink in
the graft due to acute angulation of the limb as it passed through
the pelvis. Damage to the distal iliac or femoral artery, such as dissection during graft placement, can subsequently cause outflow
obstruction and graft limb thrombosis.
84
Management of patients presenting with limb thrombosis
depends on the severity of the patient's symptoms. In the series
by Carroccio et al.,85 nearly a third of the patients presenting with
symptoms had such mild symptoms that no intervention was
required. Most patients, however, underwent a femoral-femoral
bypass to restore flow to the affected extremity. Few patients are
successfully treated with thrombolysis or graft thrombectomy followed by endovascular repair of the underlying problem. In most
series, patients with limb problems generally present early, within
the first 6 months following endograft repair.
84,91–95
In fact, Sampram
et al.93 reported that no limb occlusions presented after 30 months
of follow-up.
Migration
Distal stent graft migration after abdominal aortic endografting has
been reported to occur in 9% to 45% of patients.
certainly has been identified as a risk for development of a type
1 endoleak and delayed aneurysm rupture or late conversion to
open repair. The pathophysiology behind aortic endograft migration is complex, and various factors contribute to its occurrence.
A number of forces are at play within the aortic endograft, but
blood flow acts as the main displacing force. As the tube of the aortic graft curves, there is a change in the velocity of the blood, resulting in an increased displacement force. For many endografts, the
forces providing protection against migration are friction forces of
the graft against the aortic wall and the columnar strength of the
graft. The friction forces depend on the apposition of the graft fabric and the aortic wall and obviously can be affected by aortic wall
composition (thrombus, calcifications), size of the aorta, radial
force of the stent, and the nature of the graft fabric. It has been suggested that the presence of barbs or hooks in the proximal portion
of the stent graft may provide additional protection.
The infrarenal aortic neck length and its maximum diameter,
shape, and angulation have all been implicated as causes of stent
graft migration.
96,102,103
All of these work to decrease the friction
between the stent graft and aortic wall. Albertini et al.
the development of proximal perigraft endograft leak and device
migration following endovascular aneurysm repair. Fifteen patients
had graft migration, and 31 of 184 repairs developed a proximal
endoleak. Neck angulation was the only factor found to be significant in the development of device migration, whereas neck
angulation and neck diameter were the two factors important in
developing a proximal perigraft endoleak.
102
were unable to identify any specific anatomical correlate and
device migration. They did observe that any device that migrated
distally by more than 1 cm subsequently required an intervention.
Other hypotheses as to the cause of device migration have
focused on morphological changes in the aneurysm and aortic
neck after endovascular AAA repair. Specifically, aortic neck dilation, longitudinal sac shrinkage, and graft shortening have been
described.
96,97,105,106
One of the more widely accepted hypotheses is
aortic neck dilation following aortic endografting. After endovascular aneurysm repair, the aneurysm neck has been documented
to dilate significantly, mostly in the first 2 years after graft place-
107
ment.
In a review by Cao et al.,99 17 (15%) of 148 patients had
61,96–99
101
Lee et al.,
Migration
102
evaluated
104
however,
an episode of device migration. The only two independent risk
factors for device migration were neck dilation postoperatively
and an AAA diameter of greater than 55 mm. Others have argued
that neck dilation is not a significant event, provided adequate
graft oversizing was performed at initial endograft placement.
The amount the aortic neck dilated did not exceed the size of the
original aortic endograft placed. Larger aneurysms have also been
noted to have increased risks of developing type 1 endoleak, graft
migration, and the subsequent need for open surgical conversion
compared with larger aneurysms.
108
Outcomes
Results of Aortic Endografting
Endovascular AAA repair generally has a low mortality rate (1%-3%)
compared to open repair, and subsequent rates of aneurysm rupture after endovascular repair are reduced to 1% per year.
Endograft placement is not free of adverse events, however, and
there is frequent need for secondary interventions. Naslund et al.
reported technical complications in 26% of 34 endografts placed.
Fairman et al. evaluated the occurrence of critical events during
deployment of their initial 75 endografts, and patients were divided
into three groups corresponding to the time period in which the
graft was placed.
technical difficulties that occurred during the course of operation
that threatened the success of the procedure. Difficulty in obtaining access occurred in nearly one quarter of all patients. Although
it would be expected that the latter 25 patients should not have
experienced as great a difficulty in obtaining access, these patients
had increased complexity of their aortoiliac anatomy compared
with endograft patients earlier in their experience. This group had
a greater frequency of iliac artery balloon angioplasty, as well as
the use of iliac artery conduits. Deployment difficulties existed and
100
were composed mostly of graft foreshortening, necessitating the
placement of additional distal covered extensions. Other deployment issues encountered included suprarenal graft displacement,
infrarenal graft displacement, and device-related issues such as
iliac limb kinking or twisting. Malplacement of the graft did not
correlate with anatomical complexity.
The need for subsequent secondary procedures has been evaluated by several large series of patients who had an abdominal
aortic endograft placed.
results of 1023 patients with a follow-up of 12 months or longer.
Overall, 186 (18%) patients required a secondary intervention. The
majority of these interventions (76%) involved a transfemoral procedure, whereas the remaining patients required transabdominal
(12%) or extra-anatomical (11%) surgery. The rates of freedom from
intervention at 1, 3, and 4 years were 89%, 67%, and 62%, respectively.
The transfemoral procedures performed most frequently were aortic or iliac limb extension for graft migration or endoleak. Late
death was more frequent in those patients requiring a secondary
intervention resulting in a 3-year cumulative survival of 85%, which
is lower than the 90% rate (P <0.05) observed in those that did not
require reintervention. In addition, death was more frequently associated with those requiring a transabdominal procedure.
The Montefiore Medical Center and the Cleveland Clinic
Foundation have published their single-institution results on the
durability of aortic endografting. Montefiore reported on 239 endografts placed over 9 years, with a technical success rate of 88.7%.88
The 5-year survival rate in this group was only 37%. Secondary interventions were required in 10% of the patients, with more than half
of the secondary procedures being performed for presence of an
endoleak. Sampram et al.93 reported the results from the Cleveland
Clinic Foundation on 703 patients undergoing endovascular aortic
endografting, with follow-up averaging 1 year. Survival in this group
was 90% at 1 year and declined to 70% at 3 years. Overall, 128 secondary interventions were performed in 105 patients (15%). Freedom
from intervention mirrored that of the Eurostar registry, with freedom from intervention rates of 88%, 76%, and 65% at 1, 2, and
94
Critical events were defined as unanticipated
88,93,111,112
The Eurostar registry reported the
104
56,61,93,109
110
111

3 years, respectively. Mortality related to the secondary procedure
was 8% but rose to 18% in those requiring a transabdominal procedure. Univariate analysis revealed that secondary procedures were
more common in patients with larger major and minor sac axes, in
patients who received a large aortic stent because of a proximal
endoleak present at initial aneurysm repair, and in patients who
received treatment later in the course of the review. This last finding is felt to be secondary to the increased complexity of cases
approached in an endovascular fashion.
The Cleveland Clinic Foundation review included the use of six
different devices, which included two Zenith grafts—one that was
part of the multicenter national trial and one group that was part of
a sponsor-investigator investigational device exemption trial.
25
The
overall freedom from risk of rupture was 98.7% at 2 years. Results of
this review reveal that there are significant differences in outcomes
between groups with different endovascular devices, in particular
with regard to limb occlusion and rate of endoleak. Limb occlusion occurred most frequently with the Ancure device, at a rate
of 11% at 2 years. Endoleak of any kind was most common with
the Excluder device, at a rate of 64% at 1 year. Modular separations
were the most frequent with the Zenith graft at 3.5%. Aneurysm sac
shrinkage correlated inversely with the frequency of endo leaks,
and aneurysm sac shrinkage was most common in the Zenith
and Talent groups but least common in the Excluder group. There
were no differences with regard to rate of secondary procedures,
conversion to open repair, or migration. Sternbergh et al.
113
have
reported similar findings. Outcomes were compared between the
Zenith device and the AneuRx device, and it was determined that
the Zenith graft was associated with fewer endoleaks and a higher
rate and amount of aneurysm sac shrinkage. Bertges et al.
114
also
reported similar findings in their evaluation. Regression of AAA
size after endograft placement was more significant after placement of the Talent and Ancure endografts than with the AneuRx
or Excluder devices. During the first 2 years of follow-up, the initial
size of the AAA, presence of an endoleak, and type of graft used
were significant predictors of sac shrinkage. After 2 years, however,
only graft type was significant. Ouriel et al.
108
have additionally concluded that the outcome after endovascular AAA repair depends
on the initial size of the aneurysm.
Comparison with Open Surgery
Endovascular AAA repair has been shown to be associated with
lower postoperative morbidity, shorter length of hospital stay, and
quicker return to normal function.
surgery utilizing randomized prospective trials, however, has only
recently been available for evaluation. There have been three randomized prospective trials evaluating the use of EVAR compared
with open surgery. The EVAR-1 trial enrolled 1082 patients with AAA
who were healthy enough to be suitable candidates for surgery.
They were randomized to either EVAR or open repair. Results from
this trial demonstrated that the 30-day mortality rate was lower
after EVAR (1.7%) than open surgery (4.7%, P <0.0001). At 4-years
follow-up, the aneurysm-related mortality rate in the EVAR group
was half that in the open group (P = 0.04), but there was no difference in all-cause mortality (26% for the EVAR group and 29% for
the open group). The Dutch Randomized Endovascular Aneurysm
Management (DREAM) trial was a prospective randomized trial
that enrolled 351 patients. As in EVAR-1, the 30-day mortality rate
was lower in patients who underwent endovascular repair than in
those who underwent open surgical reconstruction, but the 2-year
outcomes were similar between the two groups. In addition, the
results of the Open versus Endovascular Repair (OVER) Veterans
Affairs Cooperative Study Group results have been reported.
this trial, 881 patients suitable for open or EVAR were randomized
to one of the two surgical techniques. As with the other two trials,
30-day mortality was lower in the EVAR arm (0.5%) than the open
surgery arm (3.0%, P = 0.004). This difference, however, resolved by
2-year follow-up time points (7.0% vs. 9.8%, respectively). Patients
undergoing EVAR had shorter hospital stays, shorter operative
115
Direct comparison to open
115
116
In
durations, and required fewer blood transfusions, but they had
increased exposure to fluoroscopy and contrast. Given its promising initial results, it is not surprising EVAR has become increasingly
popular with both patients and providers over the past decade.
One of the most controversial aspects of AAA repair, however, is when to perform EVAR and when to perform conventional open surgery. Open surgical repair of AAA has long been
considered the gold standard, and there is evidence this option
provides good long-term durability.
117,118
Endovascular AAA
repair however, given its young age, does not have similar timetested outcomes data. Recently, longer-term outcomes from both
EVAR-1 and DREAM have been reported.
30,119
For EVAR-1,30 the
median follow-up was 6 years (range, 5-10 years), and at followup, the overall aneurysm-related mortality was 1.0 deaths per
100 person-years in the EVAR group and 1.2 deaths per 100 person-years in the open repair group (P = 0.73). All-cause mortality
was 7.2 deaths per 100 person-years (EVAR) and 7.1 deaths per
100 person-years (open surgery). Graft-related complication rates
were higher in the EVAR group (12.6 per 100 person-years) compared to the open surgical arm (2.5 per 100 person-years; P <
0.001), and significantly more patients in the EVAR group required
reintervention (5.1 per 100 person-years vs. 1.7 per 100 personyears; P <0.001). In fact, new graft-related complications and
reinterventions were reported for as long as 8 years following EVAR.
For DREAM,
119
at a median follow-up of 6.4 years (5.1-8.2 years),
cumulative survival rates were 69.9% for open repair and 68.9% for
EVAR. The cumulative rates of freedom from secondary interventions were 81.9% for the open repair group and 70.4% for EVAR (P
= 0.03). Based on these data, it is clear that EVAR is not without its
drawbacks. These factors may change as the technology improves
and we gain a better understanding of the long-term implications
of placing an endovascular graft in the aorta.
Initial applications of EVAR were geared toward patients considered high risk for conventional surgery, but this concept has
come under some scrutiny after the results of the EVAR-2 trial.
In this trial, the outcomes of 404 patients with large AAAs (≥
in diameter) who were considered to be physically ineligible
for open repair were evaluated. Of this cohort, 197 patients were
assigned to undergo endovascular repair, while 207 were assigned
to have no intervention. The 30-day operative mortality rate for the
EVAR group was 7.3%, and the overall rate of aneurysm rupture in
the observation group was 12.4 per 100 person years. Aneurysmrelated mortality was lower in the endovascular repair group, but
this did not provide an advantage when evaluating all-cause mortality, and during follow-up, EVAR required a considerable increase
in expense. These results called into question the appropriateness
of using EVAR in high-risk patients. The results, however, have been
refuted by others demonstrating lower rates of perioperative mortality and better long-term survivals in these high-risk patients.
These improved outcomes are likely due to an aggressive multidisciplinary approach to managing these patients’ comorbidities.
Other Considerations
EVAR for Small Abdominal Aortic Aneurysms
Randomized prospective trials have demonstrated that there is
no benefit to open repair of AAA for aneurysms that are less than
5.5 cm in diameter.
in these two trials raised the question of whether a procedure
with lower operative mortality might provide benefit compared
with observation in patients with smaller AAA. The Positive Impact
of Endovascular Options for Treating Aneurysms Early (PIVOTAL)
trial sought to evaluate whether endovascular repair of small AAA
(4-5 cm) might provide a survival advantage compared with surveillance.
124
either EVAR (n = 366) or ultrasound surveillance (n = 362). Of the
patients randomized to EVAR, 89% underwent repair, and of those
assigned to surveillance, 31% subsequently underwent repair during the course of the study (mean follow-up 20 ±
122,123
Operative mortality rates of 2.7% and 5.8%
In this trial, 728 patients were randomly assigned to
12 months, range
120
5.5 cm
121
501
CH
40
EndovAsCulAR THERAPy foR AbdominAl AoRTiC AnEuRysms

502
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
0-41 months). The unadjusted hazard ratio (95% confidence interval [CI]) for mortality after EVAR was 1.01 (0.49-2.07; P = 0.98), with
15 deaths (4.1%) occurring in each group. No survival advantage
was demonstrated with EVAR.
Endovascular Treatment of Ruptured Abdominal
CH
Aortic Aneurysms
40
With the widespread application of EVAR for repair of AAA, its use
in the repair of ruptured AAA has similarly expanded. Initially there
were several limitations to the application of this technology to
treat the devastating problem of ruptured AAA: (1) unavailability
of preoperative CT in patients with ruptured AAA, (2) unavailability of a dedicated operating room and ancillary staff equipped to
perform emergent EVAR at all times, (3) unavailability of off-theshelf stent grafts; and (4) lack of data from multicenter randomized trials.
125
Many large-volume centers, however, have developed
protocols that allow treatment teams to overcome these hurdles
and provide emergent care for ruptured AAA with endovascular
devices.
125–128
Gerassimidis et al.
126
reported on the treatment of 69
patients with a ruptured AAA. Of these, 42 patients (63%) were suitable for EVAR. The in-hospital and 30-day mortality rates were 36%
and 41%, respectively. Veith et al.
128
reported the worldwide experience with treating ruptured AAA with endovascular grafts. Data
were collected from 49 centers at which 1037 patients were treated
with EVAR for ruptured AAA. Overall 30-day mortality was 21%,
which was significantly lower than the 30-day mortality rate for the
763 patients undergoing open repair (36%, P < 0.001) for ruptured
AAA at these same institutions during the same time period. Given
these experiences, there is a trend toward even more centers instituting a program of EVAR for ruptured AAA. Certainly, longer-term
follow-up and larger series will be required to assess whether EVAR
has ultimately affected outcomes from ruptured AAA.
Fenestrated and Branched Aortic Endografts
The most common reason for excluding patients from EVAR is lack
of a suitable proximal implantation site between the renal arteries and the aneurysm. Although commercially available devices
provide a mechanism for supplementing fixation within the suprarenal aorta without detrimentally affecting renal function,
a practice has not been advocated to treat juxtarenal aneurysms.
Despite evidence of short-term success with treatment of short
necks with devices intended to treat infrarenal aneurysms,
risk of later failure remains high.
stent graft technology was developed. The devices used currently
are hybrids of original abdominal devices. The primary goal of
treating an aneurysm with a fenestrated graft is to move the sealing and fixation region of the repair into healthy aorta with a parallel neck and without wall defects. A fenestration (hole in the
graft) allows the stent graft to occupy this more proximal location
while providing for transgraft flow to the renal arteries (or other
significant branches) (
Fig. 40-8). It is designed to incorporate the
minimum number of visceral vessels required to achieve fixation
131
To overcome this, fenestrated
129
such
130
the
and seal within healthy aorta. The fenestrations are constructed to
match the ostial diameter of the visceral vessels and maximize the
sealing zone. Several large series of fenestrated endograft deployments have been reported, demonstrating the midterm safety and
efficacy of fenestrated stent grafting.
is reported by O'Neill et al.
132
132–134
One of the largest series
They outline a series of 119 patients
with mean follow-up of 19 months. There was only one perioperative death, and survival at 12, 24, and 36 months was 92%, 83%, and
79%, respectively. The 30-day endoleak rate was 10%, and all endoleaks were type II in nature. Regression of the aneurysm sac was
noted in 79% of the patients by 12 months. Complications related to
the renal arteries was noted in 10 of the 231 stented renal arteries,
and only one patient who did not have significant renal dysfunction preoperatively went on to require dialysis. Incorporation of the
renal arteries raises questions about the effect of fenestrated stent
graft repair on long-term renal function.
Application of fenestrated technology has advanced to allow for
the treatment of thoracoabdominal aortic aneurysms (TAAA) in
which the aneurysm involves the renal and visceral vessels. When
treating TAAA with an endograft, however, use of a simple fenestration is inadequate. Unlike fenestrated grafts where a hole in the graft
suffices, in more complex aneurysms such as TAAA, the branch
arteries arise from the aneurysm. In this scenario, blood flow has
to be carried from the endograft, across the aneurysm, and to the
target vessel, without extravasation into the aneurysm (
There are two modes by which this can be assured. The first is the
fenestrated branched stent graft
(see
Fig. 40-8). In this style, the addition of a covered bridging stents
converts a fenestrated stent graft into a form of branched stent
137
graft.
Sealing between the covered stent and the fenestration is
135
or reinforced fenestrated graft
Fig. 40-9).
136
tenuous because there is very limited overlap of material. A nitinol
ring is added to the fenestration to reinforce the site of interaction
between the covered stent and the fenestration. These are termed
reinforced fenestrations. The second mode of branched graft design
is the cuffed branched stent graft
136
graft
(see Fig. 40-8). The cuff or branch creates an overlap zone
138
or directional branched stent
between the stent graft and the branch artery. It provides a segment
of overlap that can be used to provide better sealing and fixation
than the thin joint between a reinforced fenestration and mating
visceral stent graft. A longer overlap affords one the ability to use
a self-expanding stent graft rather than a balloon-expandable stent
graft. This may provide a means to better accommodate tortuosity
and diameter discrepancies and may limit type 1 endoleaks and
component separation from this region.
Investigators tend to pool results of fenestrated branch grafts
and cuffed branched grafts, with few series containing significant
numbers of patients.
has been reported by the Cleveland Clinic. Greenberg et al.
139–145
The largest single-center experience
143
performed a retrospective analysis on patients who underwent
elective open surgical repair (N = 372) or endovascular repair
(N = 352) of descending thoracic or thoracoabdominal aortic aneurysms. The group of patients treated with endovascular repair was
older and had more comorbid conditions than those undergoing
open repair. Open repair, however, was more frequently applied
A BCD
FIGURE 408 A, Example of endograft with fenestration (arrow) within endograft body to accommodate a renal artery. B, Reinforced fenestration (arrow) that
allows for placement of (C) covered stent graft, creating a form of a branched endograft for treatment of a thoracoabdominal aortic aneurysm (TAAA). D, True
directional branch (arrow) used to allow for continued flow to a visceral vessel when treating thoracoabdominal aortic aneurysm.

FIGURE 409 Example of thoracoabdominal aortic aneurysm (TAAA)
that has been treated with a branched endograft that incorporates all
of the visceral vessels.
to patients with type II or type III aneurysms and those that were
associated with a chronic dissection. Despite the differences in
patient age and comorbid conditions, mortality rates at 30 days
(5.7% vs. 8.3%) and 12 months (15.6% vs. 15.9%) were not different
between endovascular repair and surgical repair, nor was there a
difference in the development of spinal cord ischemia (4% vs. 8%,
respectively; P = 0.08). Bakoyiannis et al. performed a meta-analysis
of all English language literature published between 2000 and
2009 on endovascular procedures using fenestrated and/or
branched technology.
145
The results of this analysis demonstrated
that complex endografting can be performed with a technical success rate of 94%, with a 30-day mortality of 7%. Typically these procedures were performed in patients who were deemed high risk
for conventional surgery. The 1-year mortality was 1.3%. The application of fenestrated and branched technology is very much in
its infancy. As the technology progresses, we will be able to better
discern who will best benefit from these procedures, and ultimately replace open surgery with this less invasive option.
Summary
Abdominal aortic endografting provides a less invasive method
of treating AAA. It provides a beneficial way of treating aneurysms in patients who are at high risk for conventional open surgical repair, and results of randomized trials suggest, at least in
the short-term, it provides clinical equipoise with conventional
surgery. The durability of this procedure is still under evaluation.
It is likely, however, that the application of endovascular technology will replace current open surgical options, and ultimately the
entire aortic tree will be treated with endovascular options.
REFERENCES
1. Criado FJ, Barnatan MF, Lingelbach JM, et al: Abdominal aortic aneur ysm: overview of
stent-graft devices, J Am Coll Surg 194(1 Suppl):S88, 2002.
2. Power D: Palliative treatment of aneurysms by wiring with Colt's apparatus, Br J Surg
9:27, 1927.
3. Blakemore A, King B: Electrothermic coagulation of aortic aneurysms, JAMA 111:1821,
1938.
4. Matas R: Ligation of the abdominal aorta: report of the ultimate result. 1 year, 5 months
and 9 days after the ligation of the abdominal aorta for aneurysm of the bifurcation, Ann
Surg 81:457, 1925.
5. Rea C: Surgical treatment of aneurysm of the abdominal aorta, Minn Med 31:153, 1948.
6. Dubost C, Allary M, Oeconomos N: Resection of an aneur ysm of the abdominal aorta:
resection of the continuity by a preserved arterial graft, with result after five months, Arch
Surg 64:405, 1952.
7. Voorhees A, Jaretski A, Blakemore A: Use of tubes constructed from Vinyon “N” cloth in
bridging arterial defects: a preliminary report, Ann Surg 135:322, 1952.
8. Edwards W, Tapp J: Chemically treated nylon tubes as arterial grafts, Surgery 38:61, 1955.
9. Creech O: Endo-aneurysmorrhaphy and treatment of aortic aneurysm, Ann Surg 164:935,
1966.
10. Parodi J, Palmaz J, Barone H: Transfemoral intraluminal graft implantation for abdominal
aortic aneurysms, Ann Vasc Surg 5:491, 1991.
11. Ernst C: Abdominal aortic aneurysm, N Engl J Med 1993:1167, 1993.
12. Szilagyi D, Smith R, DeRusso F, et al: Contributions of abdominal aortic aneurysmectomy
to prolongation of life, Ann Surg 164:678, 1966.
13. United Kingdom Small Aneurysm Trial Participants: Long-term outcomes of immediate
repair compared with surveillance of small abdominal aortic aneurysms, N Engl J Med
346:1445, 2002.
14. Hertzer NR, Mascha EJ, K arafa MT, et al: Open infrarenal abdominal aortic aneur ysm repair:
the Cleveland Clinic experience from 1989 to 1998, J Vasc Surg 35:1145, 2002.
15. Cronenwett J, Johnston K: The United Kingdom Small Aneurysm Trial: implications for
surgical treatment of abdominal aortic aneurysms, J Vasc Surg 29:191, 1999.
16. Baxendale B, Baker D, Hutchinson A, et al: Haemodynamic and metabolic response to
endovascular repair of infra-renal aortic aneurysms, Br J Anaesth 77:581, 1996.
17. Laheij RJ, Van Marrewijk CJ: Endovascular stenting of abdominal aortic aneurysm in
patients unfit for elective open surgery. Eurostar group. EUROpean collaborators registry
on Stent-graft Techniques for abdominal aortic Aneurysm Repair, Lancet 356:832, 2000.
18. Beebe H, Kritpracha B, Serres S, et al: Endograft planning without preoperative
arteriography: a clinical feasibility study, J Endovasc Ther 7:8, 2000.
19. Eide K, Odegard A, Myhre H, et al: DynaCT dur ing EVAR - a comparison with multidetector
CT, Eur J Endovasc Surg 37:23, 2009.
20. Nordon I, Hinchcliffe R, Malkawi A, et al: Validation of DynaCT in the morphological
assessment of abdominal aortic aneurysm for endovascular repair, J Endovasc Ther 17:183,
2010.
21. Beebe H, Kritpracha B: Imaging of abdominal aortic aneurysm: current status, Ann Vasc
Surg 17:111, 2003.
22. Beebe H, Jackson T, Pigott J: Aortic aneurysm morphology for planning endovascular
aortic grafts: limitations of conventional imaging methods, J Endovasc Surg 2:139, 1995.
23. Truijers M, Resch T, Van Den Berg JC, et al: Endovascular aneurysm repair: state-of-art
imaging techniques for preoperative planning and surveillance, J Cardiovasc Surg (Torino)
50:423, 2009.
24. Resch T, Ivancev K, Brunkwall J, et al: Distal migration of stent-grafts after endovascular
repair of abdominal aortic aneurysms, J Vasc Interv Radiol 10:257–264, 1999.
25. Ouriel K: Endovascular repair of abdominal aortic aneurysms: the Cleveland Clinic
experience with five different devices, Semin Vasc Surg 16(2):88–94, 2003.
26. Swinnen J, Chao A, Tiwari A, et al: Vertical or transverse incisions for access to the femoral
artery: a randomized control study, Ann Vasc Surg 24:336–341, 2010.
27. Lee W, Brown M, Nelson P, et al: Total percutaneous access for endovascular aortic
aneurysm repair (“preclose” technique), J Vasc Surg 45:1095–1101, 2007.
28. McDonnell C, Forlee M, Dowdall J, et al: Percutaneous endovascular abdominal aortic
aneurysm repair leads to a reduction in wound complications, Ir J Med Sci 177:49–52,
2008.
29. Smith S, Timaran C, Valentine R, et al: Percutaneous access for endovascular abdominal
aortic aneurysm repair: can selection criteria be expanded? Ann Vasc Surg 23:621–626,
2009.
30. The United Kingdom EVAR Trial Investigators: Endovascular versus open repair of
abdominal aortic aneurysm, N Engl J Med 362:1863–1871, 2010.
31. Nordon I, Kar thikesalingam R, Hinchliffe R, et al: Secondary interventions following
endovascular aneurysm repair (EVAR) and the enduring value of graft surveillance, Eur J
Endovasc Surg 39:547–554, 2010.
32. Armerding M, Rubin G, Beaulieu C, et al: Aortic aneurysmal disease: assessment of
stent-graft treatment - CT versus conventional angiography, Radiology 215:138–146,
2000.
33. Rozenblit A, Marin M, Veith F, et al: Endovascular repair of abdominal aortic aneurysm:
value of postoperative follow-up with helical CT, Am J Roentgenol 165:1473–1479,
1995.
34. Gorich J, Rlinger N, Sokirnaski R, et al: Leakages after endovascular repair of aortic
aneurysms: classification based on findings at CT, angiography, and radiography,
Radiology 213:767–772, 1999.
35. Buth J, Harris P, Marrewijk C: Causes and outcomes of open conversion and aneurysm
rupture after endovascular abdominal aortic aneurysm repair: can type II endoleaks be
dangerous? J Am Coll Surg 194:S98–S102, 2002.
36. Brenner D, Hall E: Computed tomography - an increasing source of radiation exposure,
N Engl J Med 357:2277–2284, 2007.
37. Solomon R, DuMouchel W: Contrast media and nephropathy: findings from systematic
analysis and food and drug administration reports of adverse effects, Invest Radiol 41:
651–660, 2006.
38. Broome D, Giguis M, Baron P, et al: Gadodiamide-associated nephrogenic systemic
fibrosis: why radiologists should be concerned, Am J Roentgenol 189:586–592, 2007.
39. Sun Z: Diagnostic value of color duplex ultrasonography in the follow-up of endovascular
repair of abdominal aortic aneurysm, J Vasc Interv Radiol 17:759–764, 2006.
40. Mirza T, Karthikesalingam A, Jackson D, et al: Duplex ultrasound and contrast-enhanced
ultrasound versus computed tomography for the detection of endoleak after EVAR:
systematic review and bivariate meta-analysis, Eur J Vasc Endovasc Surg 39:418–428, 2010.
503
CH
40
EndovAsCulAR THERAPy foR AbdominAl AoRTiC AnEuRysms

504
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
41. Napoli V, Bargellini I, Sardella S, et al: Abdominal aortic aneurysm: contrast-enhanced US
for missed endoleaks after endoluminal repair, Radiology 233:217–225, 2004.
42. Henao E, Hodge M, Felkai D, et al: Contrast-enhanced duplex surveillance after
endovascular abdominal aortic aneurysm repair: improved efficacy using a continuous
infusion technique, J Vasc Surg 43:259–264, 2006.
43. Parent F, Meier G, Godziachvili V, et al: The incidence and natural history of type I and II
endoleak: a 5-year follow-up assessment with 1 color duplex ultrasound scan, J Vasc Surg
CH
40
35:474–481, 2002.
44. Beeman B, Doctor L, Doerr K, et al: Duplex ultrasound imaging alone is sufficient for
midterm endovascular aneurysm repair surveillance: a cost analysis study and prospective
comparison with computed tomography scan, J Vasc Surg 50:1019–1024, 2009.
45. Yao O, Faries P, Morrissey N, et al: Ancillary techniques to facilitate endovascular repair of
aortic aneurysms, J Vasc Surg 34:69, 2001.
46. Zarins CK: The US AneuRx Clinical Trial: 6-year clinical update 2002, J Vasc Surg 37:
904–908, 2003.
47. May J, White G, Waugh R, et al: Improved survival after endoluminal repair with secondgeneration prostheses compared with open repair in the treatment of abdominal aortic
aneurysms: a 5-year concurrent comparison using life table method, J Vasc Surg 33:
S21–S26, 2001.
48. Lee WA, O'Dorisio J, Wolf YG, et al: Outcome after unilateral hypogastric artery occlusion
during endovascular aneurysm repair, J Vasc Surg 33:921–926, 2001.
49. Wolpert LM, Dittrich KP, Hallisey MJ, et al: Hypogastric artery embolization in endovascular
abdominal aortic aneurysm repair, J Vasc Surg 33:1193–1198, 2001.
50. Criado FJ, Wilson EP, Velazquez OC, et al: Safety of coil embolization of the internal iliac
artery in endovascular grafting of abdominal aortic aneurysms, J Vasc Surg 32:684–688,
2000.
51. Schoder M, Zaunbauer L, Holzenbein T, et al: Internal iliac artery embolization before
endovascular repair of abdominal aortic aneurysms: frequency, efficacy, and clinical
results, AJR Am J Roentgenol 177:599–605, 2001.
52. Mehta M, Veith F, Ohki T, et al: Unilateral and bilateral hypogastric artery interruption
during aortoiliac aneurysm repair in 154 patients: a relatively innocuous procedure, J Vasc
Surg 33:S27–S32, 2001.
53. Kritpracha B, Pigott JP, Price CI, et al: Distal internal iliac artery embolization: a procedure
to avoid, J Vasc Surg 37:943–948, 2003.
54. Wyers MC, Schermerhorn ML, Fillinger MF, et al: Internal iliac occlusion without coil
embolization during endovascular abdominal aortic aneurysm repair, J Vasc Surg
36:1138–1145, 2002.
55. Zarins CK, White RA, Hodgson KJ, et al: Endoleak as a predictor of outcome after
endovascular aneurysm repair: AneuRx multicenter clinical trial, J Vasc Surg 32:90–107,
2000.
56. Holzenbein T, Kretschmer G, Thurnher S, et al: Midterm durability of abdominal aor tic
aneurysm endograft repair: a word of caution, J Vasc Surg 33:S46–S54, 2001.
57. Buth J, Harris PL, Van MC, et al: The significance and management of different types of
endoleaks, Semin Vasc Surg 16:95–102, 2003.
58. Chuter TA, Faruqi RM, Sawhney R, et al: Endoleak after endovascular repair of abdominal
aortic aneurysm, J Vasc Surg 34:98–105, 2001.
59. Buth J, Laheji R: Early complications and endoleaks after endovascular abdominal aortic
aneurysm repair: report of a multicenter study, J Vasc Surg 31:134–146, 2000.
60. Dattilo JB, Brewster DC, Fan CM, et al: Clinical failures of endovascular abdominal aortic
aneurysm repair: incidence, causes, and management, J Vasc Surg 35:1137–1144,
2002.
61. Zarins C: The US AneuRx clinical trial: 6-year clinical update 2002, J Vasc Surg 37:904–908,
2003.
62. Fairman R, Carpenter J, Baum R, et al: Potential impact of therapeutic warfarin treatment
on type II endoleaks and sac shrinkage rates on midterm follow-up examination, J Vasc
Surg 35:679, 2002.
63. Baum RA, Carpenter JP, Cope C, et al: Aneurysm sac pressure measurements after
endovascular repair of abdominal aortic aneurysms, J Vasc Surg 33:32–41, 2001.
64. Mart y B, Sanchez LA, Ohki T, et al: Endoleak after endovascular graft repair of experimental
aortic aneurysms: does coil embolization with angiographic “seal” lower intraaneurysmal
pressure? J Vasc Surg 27:454–461, 1998.
65. Baum R, Cope C, Fairman R, et al: Translumbar embolization of type 2 endoleaks after
endovascular repair of abdominal aortic aneurysms, J Vasc Interv Radiol 12:111–116, 2001.
66. Stavropoulos S, Park J, Fairman R, et al: Type 2 endoleak embolization comparison:
translumbar embolization versus modified transarterial embolization, J Vasc Interv Radiol
20:1299–1302, 2009.
67. Parodi J, Berguer R, Ferreira L, et al: Intra-aneurysmal pressure after incomplete
endovascular exclusion, J Vasc Surg 33:909, 2001.
68. Back MR, Bowser AN, Johnson BL, et al: Patency of infrarenal aortic side branches
determines early aneurysm sac behavior after endovascular repair, Ann Vasc Surg 17:27–
34, 2003.
69. Fan CM, Rafferty EA, Geller SC, et al: Endovascular stent-graft in abdominal aortic
aneurysms: the relationship between patent vessels that arise from the aneurysmal sac
and early endoleak, Radiology 218:176–182, 2001.
70. Gawenda M, Heckenkamp J, Zaehringer M, et al: Intra-aneurysm sac pressure--the holy
grail of endoluminal grafting of AAA, Eur J Vasc Endovasc Surg 24:139–145, 2002.
71. Jacobs T, Teodorescu V, Morrissey N, et al: The endovascular repair of abdominal aortic
aneurysm: an update analysis of structural failure modes of endovascular stent grafts,
Semin Vasc Surg 16:103–112, 2003.
72. Stelter W, Umscheid T, Ziegler P: Three-year experience with modular stent-graft devices
for endovascular AAA treatment, J Endovasc Surg 4:362–369, 1997.
73. Beebe HG: Lessons learned from aortic aneur ysm stent graft failure; observations from
several perspectives, Semin Vasc Surg 16:129–138, 2003.
74. Riepe G, Loos J, Imig H, et al: Long-term in vivo alterations of polyester vascular grafts in
humans, Endovasc Surg 13:540, 1997.
75. Jacobs TS, Won J, Gravereaux EC, et al: Mechanical failure of prosthetic human implants: a
10-year experience with aortic stent graft devices, J Vasc Surg 37:16–26, 2003.
76. Alimi YS, Chakfe N, Rivoal E, et al: Rupture of an abdominal aortic aneurysm after
endovascular graft placement and aneurysm size reduction, J Vasc Surg 28:178–183, 1998.
77. Riepe G, Heilberger P, Umscheid T, et al: Frame dislocation of body middle rings in
endovascular stent tube grafts, Eur J Vasc Endovasc Surg 17:28–34, 1999.
78. Krajcer Z, Howell M, Dougherty K: Unusual case of AneuRx stent-graft failure two years
after AAA exclusion, J Endovasc Ther 8:465–471, 2001.
79. Moore W, Rutherford R: Transfemoral endovascular repair of abdominal aortic aneurysm:
results of the North American EVT phase I trial, J Vasc Surg 34:353, 1996.
80. Heintz C, Riepe G, Birken L, et al: Corroded nitinol wires in explanted aortic endografts: an
important mechanism of failure? J Endovasc Ther 8:248–253, 2001.
81. Trepanier C, Tabrizian M, Yahia L, et al: Effect of modification of oxide layer on NiTi stent
corrosion resistance, J Biomed Mater Res B Appl Biomater 43:433, 1998.
82. Duerig T, Pelton A, Stockel D: An overview of nitinol medical applications, Mater Sci Eng
A273–275:149, 1999.
83. Starosvetsky E, Gotman I: Corrosion behavior of titanium nitride coated Ni-Ti shape
memory surgical alloy, Biomaterials 22:1853, 2001.
84. Fairman R, Baum R, Carpenter J, et al: Limb intervention in patients undergoing treatment
with an unsupported bifurcated aortic endograft system: a review of the phase II EVT trial,
J Vasc Surg 36:118–126, 2002.
85. Carroccio A, Faries P, Morrissey N, et al: Predicting iliac limb occlusion after bifurcated
aortic stent grafting: anatomic and device-related causes, J Vasc Surg 36:679–684, 2002.
86. Baum R, Shetty S, Carpenter J, et al: Limb kinking in supported and unsupported
abdominal aortic stent-grafts, J Vasc Interv Radiol 11:1165–1171, 2000.
87. Amesur NB, Zajko AB, Orons PD, et al: Endovascular treatment of iliac limb stenoses or
occlusions in 31 patients treated with the Ancure endograft, J Vasc Interv Radiol 11:421–
428, 2000.
88. Ohki T, Veith FJ, Shaw P, et al: Increasing incidence of midterm and long-term complications
after endovascular graft repair of abdominal aortic aneurysms: a note of caution based on
a 9-year experience, Ann Surg 234:323–334, 2001.
89. Carpenter JP, Neschis DG, Fairman RM, et al: Failure of endovascular abdominal aortic
aneurysm graft limbs, J Vasc Surg 33:296–302, 2001.
90. Cao P, De RP, Verzini F, et al: Endoleak after endovascular aortic repair: classification,
diagnosis and management following endovascular thoracic and abdominal aortic repair,
J Cardiovasc Surg (Torino) 51:53–69, 2010.
91. Baum RA, Shetty SK, Carpenter JP, et al: Limb kink ing in supported and unsupported
abdominal aortic stent-grafts, J Vasc Interv Radiol 11:1165–1171, 2000.
92. Carroccio A, Faries PL, Morrissey NJ, et al: Predicting iliac limb occlusions after bifurcated
aortic stent grafting: anatomic and device-related causes, J Vasc Surg 36:679–684, 2002.
93. Sampram ES, Karafa MT, Mascha EJ, et al: Nature, frequency, and predictors of secondary
procedures after endovascular repair of abdominal aortic aneurysm, J Vasc Surg 37:930–
937, 2003.
94. Fairman RM, Velazquez O, Baum R, et al: Endovascular repair of aortic aneurysms: critical
events and adjunctive procedures, J Vasc Surg 33:1226–1232, 2001.
95. Conners M III, Sternbergh W III, Carter G, et al: Secondary procedures after endovascular
aortic aneurysm repair, J Vasc Surg 36:992, 2002.
96. Resch T, Ivancev K, Brunkwall J, et al: Distal migration of stent-grafts after endovascular
repair of abdominal aortic aneurysms, J Vasc Interv Radiol 10:257–264, 1997.
97. Harris P, Vallavhaneni S, Desgranges P, et al: Incidence and risk factors of late rupture,
conversion, and death after endovascular repair of infrarenal aortic aneurysms: the
EUROSTAR experience, J Vasc Surg 32:739–749, 2000.
98. Albertini N, Kalliafas S, Travis S, et al: Anatomical risk factors for proximal perigraft endoleak
and graft migration following endovascular repair of abdominal aortic aneurysms, Eur J
Endovasc Surg 19:308–312, 2000.
99. Cao P, Verzini F, Zannetti S, et al: Device migration after endoluminal abdominal aortic
aneurysm repair: analysis of 113 cases with a minimum follow-up period of 2 years, J Vasc
Surg 35:229–235, 2002.
100. Lawrence-Brown M, Semmens J, Hartley D, et al: How is durability related to patient
selection and graft design with endoluminal grafting for abdominal aortic aneurysm? In
Greenhalgh R, editor: Durability of vascular and endovascular surgery, London, 1999, WB
Saunders, pp 375–385.
101. Resch T, Malina M, Lindblad B, et al: The impact of stent-graft development on outcome of
AAA repair–a 7-year experience, Eur J Vasc Endovasc Surg 22:57–61, 2001.
102. Albertini J, Kalliafas S, Travis S, et al: Anatomical risk factors for proximal perigraft endoleak
and graft migration following endovascular repair of abdominal aortic aneurysms, Eur J
Vasc Endovasc Surg 19:308–312, 2000.
103. Malina M, Lindblad B, Ivancev K, et al: Endovascular AAA exclusion: will stents with hooks
and barbs prevent stent-graft migration? J Endovasc Surg 5:310–317, 1998.
104. Lee J, Lee J, Aziz I, et al: Stent-graft migration following endovascular repair of aneurysms
with large proximal necks: anatomical risk factors and long-term sequelae, J Endovasc
Ther 9:652–664, 2002.
105. White G, May J, Waugh R, et al: Shortening of endografts during deployment in
endovascular AAA repair, J Endovasc Ther 6:4, 1999.
106. Prinssen M, Wever J, Mali W, et al: Concerns for the durability of the proximal abdominal
aortic aneurysm endograft fixation from a 2-year and 3-year longitudinal tomography
angiography study, J Vasc Surg 33:S64–S69, 2001.
107. Badran MF, Gould DA, Raza I, et al: Aneurysm neck diameter after endovascular repair of
abdominal aortic aneurysms, J Vasc Interv Radiol 13(9 Pt 1):887–892, 2002.
108. Ouriel K, Srivastava SD, Sarac TP, et al: Disparate outcome after endovascular treatment of
small versus large abdominal aortic aneurysm, J Vasc Surg 37:1206–1212, 2003.
109. Becker G, Kovacs M, Mathison M, et al: Risk stratification and outcomes of transluminal
endografting for abdominal aortic aneurysm: 7-year experience and long-term follow-up,
J Vasc Interv Radiol 12:1033–1046, 2003.
110. Naslund TC, Edwards WH Jr, Neuzil DF, et al: Technical complications of endovascular
abdominal aortic aneurysm repair, J Vasc Surg 26:502–509, 1997.