Добавил:
Upload Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
DLYa_PEREVODA_PS_Oktyabrskaya_en.docx
Скачиваний:
0
Добавлен:
01.07.2025
Размер:
542.78 Кб
Скачать
      1. Technical data test set to test hf devices

Description

Information:

Note

Required Customer

Guaranteed Contractor

1.

Type

2.

Manufacturer

3.

Standards

IEC 61326-1

IEC 61000-6-2/4

IEC 61010-1

4.

Techincal description:

5.

LF - generator

5.1

Frequency range, Hz

20 – 24000

5.2

Output Type

The symmetrical and coaxial

5.3

Maximum output power, W, not less

10

6.

DC voltage source (LF - generator)

6.1

Highest voltage (full scale), V

15

6.2

The load resistance in ohms, at least

50

7.

HF- generator

7.1

Frequency range, kHz

24 – 2500

7.2

Type of output

The symmetrical and coaxial

7.3

Range issued stresses ,V

0 – 20

7.4

Maximum output power, W, not less

5

7.5

Internal resistance (Rgen), Ohm,

0; 75; 150

7.6

Damping asymmetry symmetrical output dB minimum

25

7.7

Output return loss in dB, not less than

45

7.8

The maximum voltage applied to the generator output from an external source via the BS-60 (from SPTA), no more than

100

8.

LF - Multimeter

8.1

Frequency range, Hz

45 – 24000

8.2

Input Type

The symmetrical and coaxial

8.3

The limits of measurement of current, A

0,1; 1,0

8.4

Measurement range of voltage, V

0,3; 3; 30; 300

8.5

Input impedance voltmeter, ohm, not less

100

8.6

Input resistance ammeter, Ohm, not more

2,5

9.

LF - frequency (LF - Multimeter)

9.1

Frequency range, Hz

45 – 24000

10.

HF - Multimeter

10.1

Frequency range, Hz

24 – 2500

10.2

Input Type

The symmetrical and coaxial

10.3

The limits of current measurement, A

0,1; 1.0

10.4

Measurement range of voltage, V

0,1; 1; 10; 100

10.4

The input impedance voltmeter, ohm

100

10.5

Input resistance ammeter, Ohm, not more

1,5

10.6

Damping asymmetry input voltage dB minimum

30

Damping asymmetry current input dB minimum

45

11.

(HF in - multimeter)

11.1

Frequency range, Hz

24 – 2500

12.

Selective voltmeter

12.1

Limits of measurements,

0,01; 0,1; 1; 10; 100

12.2

Measurement range voltage level dBc

-38; -18; 2; 22; 42

12.3

Frequency range, kHz

0,3 – 2500

12.4

The dynamic range when measuring selective levels dB

60

12.5

Noise level (full scale for all ranges), dB, no more

- 60

12.6

The bandwidth of the selective reception, Hz

25; 400; 1740; 3100

12.7

Selectivity (at offset from the edge of the strip at the reception 1,5 Δ), dB, not less

60

12.8

Frequency resolution setting Hz

0,1

12.9

Input impedance, Ohm

20

12.10

Damping asymmetry input (2.5 MHz) dB minimum

30

12.11

Type inputs

The symmetrical and coaxial

13.

Milli stop watch

13.1

Measuring range ,sec

0,001 – 99

13.2

Limits of relative error of measurement of time,%

±0,1

13.3

Resolution changes registration status contacts (time) msec ,not less

0,4

14.

Contact outputs

14.1

Amount

2

14.2

The resistance in the closed state, ohm, max

0,2

14.3

Resistance in the open position, MOhm, no less

50

15.

Digital inputs

15.1

Amount

16

15.2

Rhode signals

"Dry" contact; Contact for instance. Upost ≥ 300V

15.3

Inputs with TTL levels

yes

15.4

Inputs with CMOS levels

yes

16.

Generel technical information

16.1

Power devices - single phase

220 В, 50 Гц

16.2

Power consumption, W, no more

100

16.3

The insulation resistance between the housing and electrically insulated live parts of the device, Mohm, no less

20

Соседние файлы в предмете [НЕСОРТИРОВАННОЕ]