ECHO 2013 / Echocardiography in Advanced Heart Failure Mechanical Support
.pdf
Echo detected changes in right heart hemodynamics are associated with symptomatic status after LVAD implant
Lam et al, JASE 2009;22:1055-1062
•PVR, RVSP, TR reductions within 1 month
•No change in RV size, RV FAC, TAPSE
•Pts with larger reductions in PVR had better clinical status (6MWT, MLHFQ)
•Pts with significant deteriorations in RV function had poorer clinical status
PVR (wood units) = (TR velocity / RVOT TVI) X 10 + 0.16
Wood units X 80 = dyn.s/cm
Normal PVR = 1.25-2.5 WU; 100-200 dyn.s/cm
VTI = 12cm |
VTI = 14cm |
TR max vel |
TR max vel |
= 3.2 m/s |
= 2.1 m/s |
PVR = 2.8 |
PVR = 1.7 |
Wood units |
Wood units |
Abbas et al, JACC 2003:41;1021-7; Lam et al, JASE 2009;22:1055-1062
Potential complications
•Aortic dissection
•Insufficiency of native AV / Ao clot
•Conduit obstruction/kinking
•Worsened RV failure
•Inlet cannula thrombosis
•Overpumping
•Pump failure
•Controller/battery failure
complications
•Device malfunction – loss of circulatory support
–Pump/controller
–Cannulae
–Valves
•obstruction / degeneration - regurgitation
•Infection
•Thrombus formation / embolization
•Surgical bleeding / Leak around cannulae
•other
LVAD complications
Guiding device wean
•Reduce pump rates and observe
–Filling
–Ejection
•LV volumes and function
•Forward flow in LVOT
–Valve function
•AV opening
•Degree of MR / TR
ECMO wean: monitor SV or TVI as support decreased
|
|
|
|
|
|
|
|
|
Ecmo |
TVI = |
|
||||
|
flow = |
12 cm |
|
||||
|
3L/m |
|
|
|
|
1.5 L/m |
TVI = 13 cm |
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
0.5 |
|
L/m |
TVI=11 cm |
Guiding placement of percutaneous assist devices
•Guiding transeptal puncture (tandem heart)
–In LA away from MV, LA appendage, pulm veins
•Proper positioning of LV cannula (Impella)
–In LV cavity but away from walls
•Assessing filling
–LV/RV volumes
Impella
•Assess AV pre-device placement
•Guide cannula placement
• Monitor cannula position and function
Impella + TTE