Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5817_Библиотеки_им_академика_М_И_Перельмана-1.pdf
Скачиваний:
0
Добавлен:
09.09.2026
Размер:
18 Мб
Скачать
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
1
Margin Shadow
Margin shadows are often attributed to refraction. As shown in Figure impinging on the lateral margins of round, fluid-filled spaces are diverted into the surrounding tissue. This phenomenon also explains why some marginal struc­tures such as fetal skulls are not fully displayed.
1.22, mainly parallel ultrasound rays
Side Lobe
The actual sound field does not correspond to our simplified imaginary model of a thin, focused ultra­sonic beam. Rather, in the near field it forms so-called side lobes. Should a side lobe meet a strong reflection, this will return echoes in the direction of the side lobe. Since, however, a received echo is always interpreted as one returning vertically to the transducer, it is mis-
1.23). In modern ultrasound systems
interpreted ( complex runtime calculations and suppression of echoes entering laterally can reduce some of the side lobe artifacts.
Fig.
Slice Thickness Artifact
A slice thickness artifact comes about because of the finite penetration of an ultrasound beam, i. e., the “thickness” of the sound field. In an ideal case the sound field would be infinitely thin, but in reality it ex­tends over one to several millimeters (depending on
Figure
the frequency). pinging on a cyst, in one instance centrally, in the other at its margin. The depth of penetration of a sound field can be reduced by focusing through an acoustic lens.
1.24 shows a sound beam im-
Repetition Artifact
This artifact, also known as reverberation artifact, oc­curs when ultrasound waves impinge on a strong re­flector with a large surface in the near field. Some of the sound waves arriving there are reflected, return as echoes to the probe, where they are received. However, it is possible for some of the echoes to be reflected back from the probe into the tissues. In this way some of the sound waves can oscillate back and forth between the
14
Fig. 1.22 Margin shadow. Mode of generation (left) and simplified diagram of the ultrasound display (right).
Fig. 1.24 Slice thickness artifact. Fig. 1.25 Repetition artifact. Mode of generation (left) and
Fig. 1.23 Side lobe artifact. Mode of generation (left) and simplified diagram of the faulty ultrasound display (right).
simplified diagram of the ultrasound display (right).

Doppler Sonography

probe and the reflector. Hence the same structure will be displayed several times at several equidistant
Fig.
depths, with diminishing brightness (
verberation phenomenon is most often seen when in­terfaces close to the transducer present a large differ­ence in impedance (soft tissue/air).
1.25). The re-
Doppler Sonography
Fundamentals of Doppler Sonography
The Doppler effect is named after Christian Johann Doppler, who first described it in 1842. A change in dis­tance between an observer and a sound source changes the observed frequency. If, for instance, we stand on the sidewalk and, for example, an ambulance passes sounding its siren, we note that the tone of the
Fig.
siren changes abruptly ( changes is the pitch, not the intensity of the note.
While the vehicle is approaching the intensity in­creases, but diminishes again once the vehicle has passed. By contrast, the pitch remains constant as the
vehicle approaches. This remains true as the vehicle moves away, but the pitch is now lower. If the vehicle moves at a speed of 60 mph (100 km/h), the difference
1.26). Specifically, what
Geometrical Distortion
When an ultrasonic image is generated from received echoes, the assumption is that the ultrasonic wave has a constant velocity and spreads in a straight line. In fact, however, ultrasound moves at different velocities in different tissues. The sound ray is also deflected on its path through the tissues by refraction and diffrac­tion. All these deviations are not accounted for in the construction of the image. Hence the display may in­corporate small geometric errors.
is two whole notes. Of course, in reality the siren emits a constant pitch. The frequency heard by our ear, however, depends on whether the vehicle is moving toward or away from us. The same is true if we are in a vehicle moving past a stationary ambulance (sender stationary, receiver in motion) or if both vehicles are moving (sender and receiver move relative to each other). When receiver and sound source move toward each other, the frequency picked up by the receiver is higher than that sent by the sound source. Correspond­ingly, the frequency picked up by the receiver is lower than that emitted by the sound source when receiver and emitter are moving apart. The difference between the emitted and received frequency is called the Dop­pler frequency, or Doppler shift.
In diagnostic medical procedures the Doppler shift is
utilized to measure the velocity of blood flow in blood
Basic Concepts
Fig. 1. 26 Doppler effect. The
frequency heard by the ear depends on whether a noise is moving
toward or away from the observer.
15
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
16
1
vessels, as first described by Satumora in 1959. The emitted sonic pulse is reflected from the moving blood corpuscles. The frequency of the echo received by the transducer differs from the ultrasound pulse it origi­nally emitted. In this case the sender and receiver are fixed and the frequency shift is due to the movement of a reflector (erythrocyte). The frequency therefore shifts twice: once when the ultrasonic pulse strikes the reflector, and again when the echo with its already shifted frequency returns. The frequency shift also de­pends on the angle between the axis of the blood ves-
1.27). This angle is
F
sel and the ultrasound beam (
ig.
known as the Doppler angle.
Mathematically the Doppler formula can be written:
f=2
f
v cos α
c
where: f = the Doppler shift frequency
f = transmitting frequency of the trans-
ducer c = velocity of sound v = blood velocity α = angle between the axis of the vessel and
the sound beam
If this formula is solved for v, the velocity of the ery­throcytes can be calculated:
v=∆f
c
2 f cos α
The Doppler frequency lies in the range of audible frequencies.
Table 1.3 shows how the measured Doppler
frequency depends on the Doppler angle. The calcula­tions are based on the assumption that the transmit­ting frequency is 6 MHz and the blood flow velocity 1
−1
, using the above Doppler formula.
ms
1.3 shows that at a 90° angle there is no Dop-
Table
pler shift, since the distance between erythrocyte and
α
V
Fig. 1.27 Doppler angle. The angle between vessel and ultra­sound beam (v = blood flow velocity; α = angle between axis of vessel and incident beam).
Table 1.3 Relationship between Doppler frequency and Dop­pler angle
α cos α∆f
1 7.79 0 30° 0.866 6.75 13 45° 0.707 5.51 29 60° 0.5 3.9 50 90° 0 0 100
Table 1.4 Correction factors and correction errors for different Doppler angles
Angle α Correction factor
1/cos α
30° 1.15 3% 45° 1.41 6% 60° 2.00 9% 70° 2.92 14 % 75° 3.86 21 % 80° 5.76 30 %
MHz
Deviation (%)
Correction error
transducer remains constant, i. e., sender and receiver move neither away from nor toward each other. Thus, what is measured is always the velocity vector parallel to the sound beam. The precise blood flow velocity in vessels running at an angle to the transducer can be calculated by determining the Doppler angle. This cal­culation is often termed angle correction. The Doppler angle can be measured by a measuring axis, which in B-mode is often aligned with the axis of the vessel. The ultrasound system then measures the angle between the axis of the vessel and the incident sound beam. The precise Doppler frequency is calculated, taking into ac­count an angle correction factor of 1/cos α. Table
1.4
contains a list of the various correction factors and cor­rection errors that may result from lack of precision in positioning the measuring axis.
As can be seen in
Table 1.4, as the angle increases the
correction error also increases. Hence during the ex­amination the Doppler angle must be kept as acute as possible by adjusting the tilt of the transducer.
The frequency of the altered echo received is then
compared to the original frequency of the transmitted ultrasound pulse to determine the Doppler frequency. In the ultrasound system the apparatus used for this is a demodulator. The demodulator delivers electrical signals that can be transmitted, for example, directly to a loudspeaker. In order also to determine the direction of blood flow, phase relationships are compared. Blood flow toward the transducer is defined as positive, away from the transducer as negative.
There is no predominant blood flow inside a blood
vessel. Rather, areas of different velocities form. Flow rates close to the vessel wall are slower than flow rates in the vessel’s axis, because of friction between the er­ythrocytes and the wall. Thus, a laminated flow profile is formed. These profiles change over time, because of flow pulsatility, vessel wall elasticity, bends in the ves-
sels, vascular branching, and vascular stenoses. Signals derived from a blood vessel therefore consist not of a single Doppler frequency, but of a mixture of frequen­cies. This frequency mixture can be reduced to its basic oscillations by the use of the mathematical process of fast Fourier transform (FFT). The principle of FFT is il-
Figure
lustrated by means of a simple example in
The velocity distribution of the erythrocytes is repre-
sented by a tracing of their spectra on a time axis (Fig. 1.29). Corresponding to their magnitude positive frequency shifts are represented above the time axis, negative below. The rate of occurrence of individual frequencies is coded by the brightness of each point in the image.
1.28.
Continuous Wave Doppler Systems
Doppler Sonography
F F
T
Continuous wave (CW) Doppler systems transmit and receive sound waves continuously. This requires at least two elements: one for continuing transmission
Fig.
and one for constant receiving ( systems may be conceptualized as directional and nondirectional systems. Nondirectional systems can­not distinguish positive and negative flow directions. Directional systems, on the other hand, can detect flow direction by determining the phase shift. Since ultra­sonic pulses are being transmitted and received con­tinuously, it is not possible to measure the time of flight between a transmitted pulse and a received echo.
Thus, the depth from which the signals are received
cannot be determined for any scan. It follows that all
vessels on which the ultrasound beam impinges are scanned, and vessels running together in different planes are combined in a single display. Thus, it is not possible to distinguish the individual vessels. The depth of penetration is determined by the frequency. CW Dopplers are especially suited for the examination of vessels close to the surface.
1.30). CW Doppler
Pulsed Wave Doppler systems
The action of pulsed wave (PW) Doppler systems is similar to B-mode transducers (Fig. sound pulse is transmitted by an element and the echoes are received by the same element. Since the
velocity of the sound beam is known, the signals can be localized precisely by determining the time of flight.
Thus, the ultrasound system first transmits a brief ul­trasound pulse. The system is then switched off, i. e., all echoes entering subsequently are ignored. After a pre­determined time the system is switched to receive, and the echoes are processed. The duration of the receiving time defines a measured space. As a rule this measured space corresponds to the diameter of the vessel. It is called the sample volume, or Doppler window. When
1.31). A brief ultra-
Fig. 1.28 The mathematical procedure of FFT breaks down the frequency derived from the blood vessel into its component oscillations.
Fig. 1.29 The different velocities of the erythrocytes are dis­played as a tracing of spectra along a time axis.
Transmitting Receiving
V
Fig. 1.30 CW Doppler systems. At least two elements are re­quired to allow continuous transmission and reception (v = blood flow velocity).
the receiving period is complete, the system again sends an ultrasonic pulse. The number of ultrasound pulses transmitted per second is called the pulse re­petition frequency (PRF). The maximal transmitted PRF
Basic Concepts
17
1
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
Alias Phenomenon in Pulsed Doppler
Since reception in a pulsed Doppler is not continuous as in a CW Doppler, the pulsed Doppler does not have a
Sample volume
V
Fig. 1.31 PW Doppler systems. Display of the sample volume or Doppler window, corresponding as a rule to the diameter of the vessel.
T2
T1
Fig. 1.32 Oscillations with a frequency of 3 Hz being scanned at points T1 and T2, i. e., with a frequency of 2 Hz.
1s
constant Doppler signal. Rather, the Doppler signal is composed of a series of samples. The temporal resolu­tion is determined by the pulse interval, which is the inverse of the PRF. The actual frequency shift is there­fore unknown, and the curve between individual samples is interpolated. The maximal frequency that can be measured correctly is smaller than half the PRF. Higher frequencies contain too few samples to be dis­played accurately. The PRF must therefore be at least twice the Doppler frequency being measured. Figure
1.32 shows an oscillation with a frequency f. If we as- sume that the segment of the curve shown corre­sponds to one second, the oscillation has a frequency of 3 Hz. Suppose the signal is scanned at points T1 and T2. The oscillation is scanned twice in one second, i. e., the scanning frequency is 2 Hz. The signal constructed from the two sample measures has a frequency of 1 Hz (Fig. 1.33). To reconstruct the true frequency we would have to measure with a scan frequency of at
t
least 6 Hz in this example. The maximal frequency that can be displayed without error is called the Nyquist limit. If the Nyquist limit is exceeded, the result is an aliasing (Latin: alias = otherwise, at another time) phe­nomenon. This effect can be observed on a rotating wagon wheel. When the wagon begins to move, the direction of rotation of the wheel is seen correctly. As the wagon accelerates, the wheel’s motion is seen cor­rectly up to a certain maximal speed. As the speed con­tinues to increase, the motion is suddenly seen as backward motion. As shown in phenomenon can be seen in spectrum analysis when the Nyquist limit is exceeded. The maximal Doppler frequency is reflected in the opposite direction.
1.34, the aliasing
Figure
18
T2
T1
Fig. 1.33 Signal with a frequency of 1 Hz, reconstructed from the two measurements T1 and T2 in Fig. 1.32.
depends on the depth of penetration: The deeper the sample volume lies in the tissue, the longer is the wait for the returning echo, i. e., the greater is the interval between successive pulses. Thus, the maximal PRF diminishes with deeper sample volumes.
1s
t
Baseline Shift
The PRF cannot be raised indefinitely if the Doppler signals are to provide accurate localization. A limited remedy is offered by the baseline shift. The Nyquist limit is divided equally between a positive and a nega­tive component. By shifting the reference axis (base line) for the direction of flow upward or downward, the range of the display can be extended in one direc­tion. Such an extension can reach twice the Nyquist limit. However, during this process the opposite flow cannot be displayed. shifting of the baseline during the recording of a Dop­pler curve.
Figure
1.35 shows a stepwise
Wall Filter
A wall filter is designed to eliminate low frequency noise artifacts and movement of vessel walls from the frequency spectrum. It is conceived as a high pass filter (HPF), i.e., high frequencies can pass the filter, low frequencies are filtered out and rejected. The setting of the wall filter is variable and can be changed by the user. The wall filter determines the minimal frequency range that can be displayed.
Color-Coded Doppler Sonography
As described in the section on PW Doppler, a pulsed Doppler system can derive Doppler signals with a de­fined sample volume from a precisely defined location. In order to obtain a flow velocity analysis from several locations simultaneously, many sample volumes are arranged along adjacent ultrasound lines of sight
ig.
F
1.36). This results in an analysis of flow velocity
( over a defined area, called a region of interest (ROI). In this situation a simultaneous spectrum analysis of each sample volume by FFT would require too much time and therefore cannot be applied. Moreover, the Doppler spectra obtained could not all be displayed simultaneously. Scanning with 20 lines of sight each
with 50 sample volumes would after all encompass 1000 samples! In color-coded Doppler sonography readings are therefore obtained not by FFT but, among other procedures, mostly by means of an autocorrela­tion procedure. In this procedure the Doppler signals for all the sample volumes in the time of one scan are collected and compared (correlated) with those of the following scan time. While 128 scan points are re­quired to analyze a signal by FFT, just four suffice for an autocorrelation procedure. The information about the phase difference so obtained is a direct measure of the mean velocity distribution in a sample volume. The re­sults of the scan are displayed in color. In B-mode flow toward the transducer is displayed in red, away from the transducer in blue. The velocity of flow is indicated by degrees of brightness in the colored pixels. A bright pixel indicates high velocity, a dark pixel slow blood flow. Color-coded Doppler sonography follows the same rules as PW Doppler. It is dependent on the Dop­pler angle and shows aliasing. Aliasing is clearly shown in the color display by color reversal.
Doppler Sonography
Fig. 1.34 Aliasing in a spectral analysis. When the Nyquist limit has been exceeded the maximal Doppler frequency is recorded in the opposite direction.
Basic Concepts
Fig. 1.35 Stepwise shift of the baseline during a Doppler re­cording.
Fig. 1.36 ROI.. A number of sample volumes are constructed along beams with their axes lying side by side to obtain a flow analysis measured at several locations simultaneously.
Amplitude-Coded Flow Display
Beside color-coding described above, an amplitude­coded flow display has become available in the last few
years. This principle was first described in the litera­ture in 1994 and is available for modern ultrasound
systems under various names, depending on the man­ufacturer. Amplitude-coded procedures are variously known as (among others) ultrasound angiography, power Doppler, power Doppler angio, color angio, angio color, color Doppler energy, or color perfusion
19
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
20
1
imaging. To distinguish between echoes of ultrasound signals from tissue and blood a HPF is inserted into the system in color-coded Doppler sonography (cf. Wall
p.
Filter, nals from tissue and erythrocytes are additionally coded by their intensity (amplitude). As a rule signals from tissue display an intensity 1000 times higher than those from erythrocytes, enabling significantly better separation between signals from tissue and ery­throcytes. While in color-coded Doppler sonography erythrocyte echoes are processed by frequency analy­sis, in amplitude-coded flow displays color attribution depends directly on the intensity of the echoes, the way intensities are displayed by a corresponding gray-
19). In an amplitude-coded flow display, sig-

Safety Aspects

The possible biological effects of medical ultrasound diagnosis have been discussed since their beginnings in the 1960s. Two dif ferent phenomena associated with the passage of ultrasonic rays through the body are known to influence biological systems. These are thermal effects, since part of the energy insonated into the body is transformed into heat and mechanical ef­fects due to changes in pressure on the molecules.
Thermal Effects
The rise in temperature in the tissues during insona­tion can be attributed to absorption of energy, absorp­tion being the transformation of ultrasound energy into heat. The property of absorption differs greatly in different tissues. The ability to absorb energy is poor in fluids (amniotic fluid, blood, urine), but very high in bones. The bones of an adult can absorb 60−80 % of the ultrasound energy impinging on them. Ultrasound frequency is the most important physical parameter influencing absorption, for the higher the frequency the faster ultrasound energy will be absorbed. A rise in temperature of 2.5 °C or more can cause severe damage to biological tissues, while temperature rises of 1 °C are not considered to be significant. However, experimen­tal studies (AIUM Bioeffects Committee 1988) suggest that in the diagnostic range no damaging thermal ef­fects are likely to occur.
Mechanical Effects
Biomechanical effects almost always refer to phenom­ena related to the generation and possible implosion of microscopic bubbles in tissues. The interaction be­tween ultrasonic rays and these bubbles is known as cavitation. An ultrasonic oscillation gives rise to nega-
scale value in B-mode. Since amplitude-coding deter­mines only intensity and not frequency shifts of the signals received from the erythrocytes, the procedure of amplitude-coded flow display does not produce color reversals due to changes in flow direction or ali­asing. Changes in blood flow velocity or direction can­not be distinguished in an amplitude-coded display. The advantages of amplitude-coded procedures in­clude in particular the ability to display flow velocities that are distinctly lower than those that can be dis­played by conventional Doppler methods. Amplitude signals are also less prone to noise and are almost inde­pendent of the probe angle.
tive as well as positive pressure phases. The positive phase is known as the compression phase, while the negative phase is called the cavitation phase, or rare­faction. If the negative pressure is sufficiently large, it is possible to generate microscopic bubbles or to en­large bubbles that are already present. The appearance and effect of cavitation depends not only on the selected ultrasound frequency and intensity, but also on the focus of the sound field.
For instance, cavitation appears in a stable form in vibrating, gaseous bodies that are stable in themselves, but have been set to oscillate by the ultrasound field. Once oscillations have reached a certain level, a fluid medium incorporating gas microbubbles is set in mo­tion. This is called microstreaming. It has been demon­strated that microstreaming exerts an extremely strong pressure that can cause cell membranes to burst.
Another aspect of cavitation is the so-called tran­sient cavitation in which existing microbubbles or cells undergoing cavitation inflate under the influence of negative pressure and then implode abruptly. These processes happen in microseconds. The implosion can then cause a sudden surge in temperature or a sudden rise in pressure in an area as small as a square micro­meter. In the end this may result in cell and tissue de­struction. However, research has shown that transient cavitation is purely a threshold phenomenon that only occurs in the presence of extreme pressure and ultra­sound frequency when cavitated cells are already pre­sent. If the pressure is below the cavitation threshold, it will never by itself lead to cavitation even during ex­tremely lengthy insonations. To this day there has been no known case in which the diagnostic insonation of a human body has led to even transient cavitation.
However, biological effects are difficult if not im­possible to demonstrate, since the negative effects of these phenomena may in some circumstances affect
only individual or a few cells, while cavitation may occur in any location in the tissue. Currently available ultrasound systems are designed in such a way that their settings allow the operator to avoid such me­chanical effects as cavitation at peak pressures by lim­iting average pulse intensity.
ously. The sample volume is therefore distinctly smaller than in B-mode, but the PRF at ca. 1 kHz is also clearly lower. As far as damage to the patient is con­cerned, M-mode scanning may also be considered safe.
CW Doppler
Important Definitions
Acoustic Output
Acoustic output is defined as the ultrasound waves that are insonated by the transducer into the body of the patient under examination. The intensity of ultra­sonic waves is usually measured in fluids with known absorption coefficients.
Acoustic Power
Acoustic power is defined as the total acoustic output (in mW) insonated by the transducer into the tissue.
Intensity
Intensity is defined as the energy acting on a specific area. Intensity is measured in mW/cm over time (I
).
ta
2
and is averaged
Intensity Special Peak Time Average
Intensity special peak time average (ISPTA) measures the maximum of the intensity averaged over time. Its unit is mW/cm
2
.
Risks of Individual Ultrasound Procedures
B-Mode
As in M-mode, the insonated tissue volume in a CW Doppler procedure is relatively small. Energy is trans­mitted not as ultrasound pulses, but continuously. The insonated energy therefore attains a power of ca. 100 mW. Thus, thermal effects can no longer be ig­nored in all cases. However, despite the higher energy output, the mechanical effects remain far below those of B-mode or M-mode procedures. In CW Doppler the transmitted energy must be adjusted to the depth of the examination. It is important to keep examination times as brief as possible.
PW Doppler
The insonated tissue volume is also very small in this procedure. Moreover, PW Doppler involves very high PRFs. The pulse time is often double that of B-mode or M-mode. Therefore, considerably elevated ultrasonic energy values may be attained. The implied danger of temperature effects can no longer be ignored. Since the intensities of the stimulating energy are similar to those of B-mode and M-mode, mechanical effects may be ignored.
Color-Coded Doppler Sonography
The energies used in color-coded Doppler sonography lie between those used in B-mode and those used in PW Dopplers. Mechanical effects can be ignored. The insonated ultrasound pulses are distributed over a relatively large volume. Thermal effects are higher than in B-mode, but lower than in PW Doppler.
Basic Concepts
The insonated ultrasound energy is very small in B­mode (10 mW/cm possible very brief (5µs) single pulses are trans­mitted at a PRF under 5 kHz. The insonated energy is distributed over a large sample volume and generates immeasurably small temperature changes in the tis­sues. As far as damage to the patient is concerned, B­mode scanning may be considered absolutely safe.
2
). To attain as great a resolution as
M-Mode
In this case an elevation of temperature might be con­sidered in view of the higher energy used. As opposed to B-mode, a single ultrasound ray is emitted continu-
Summary
To date there have been no indications of any injuries due to the use of ultrasound procedures anywhere in the world with the currently used sound pressure in­tensities of less than 100 mW/cm
Nevertheless, each user is advised to adhere to the ALARA principle (as low as reasonably achievable), i.e., to use as low an intensity as possible for no longer than necessary. Consequently, Doppler procedures using higher intensities should be avoided during the first trimester of pregnancy.
2
.
21
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound

Important Instrument Settings

1
A great many ultrasound instruments of various types with a variety of different controls are available on the market. The most important parameters that can be controlled will be described in the following.
Selecting the Most Suitable Transducer
As shown in the section on “Resolution” (cf. p. 9), the axial as well as the lateral resolution in addition to the depth of penetration are dependent on the frequency of the sound wave. The higher the frequency of the ul­trasound waves, the higher the resolution, but the lower the depth of penetration. The optimal transducer frequencies for abdominal and obstetric examinations lie between 3 MHz and 5 MHz.
B-Mode Settings
Depth of Penetration
The depth displayed by an ultrasound system can be varied within the maximal depth of penetration set by the frequency of the transducer. This depth should be set to correspond to the areas of diagnostic interest, i.e., the depth of the display should correspond to the depth of the anatomical structure to be examined.
Gain
For optimal diagnostic accuracy, the tissue must ap­pear uniformly bright over the whole depth of the image. If the gain is too weak, the result is a nonhomo­geneous display pattern with inadequate echoes. Sub­tle tissue differences cannot be distinguished. If the gain is too high, the individual structures are insonated excessively, so that they cannot be distinguished from each other. Electronic noise will also be displayed when the gain is too high and will be seen, for example, as spontaneous echoes in fluid-filled spaces.
The intensity of echoes diminishes with increasing penetration. Therefore, the echoes from the deepest penetration must be intensified relative to those from
p.
the near field by using the TGC (cf.
7).
Focusing
Setting the Doppler Parameters
Sample Volume
In PW systems it is of paramount importance to adapt the size of the sample volume to the diameter of the vessel to be examined, for if the sample is too small not all parts of the flow can be encompassed.
PRF and Baseline Shift
The sample volume must also be adapted to the type of flow to be measured in the vascular segment to be ex­amined. If the flow is displayed in only one direction, it is possible to increase the sample volume by shifting the baseline to the Nyquist limit (cf. Baseline Shift, p. 19). Next, PRF must be adapted to flow velocity. Set­ting the PRF too low results in the phenomenon of ali­asing described on page 18. If, on the other hand, the PRF is set too high, the tracing of spectra displayed on the monitor is too small. The result is that markers for measurements cannot be set accurately, leading to er­rors in the determination of the Doppler indices. PRF and baseline should be set so that the tracings of spec­tra displayed on the monitor are as large as possible without aliasing.
Scaling the Time Axis
The scale of the time axis can also be set. If it is too long, the tracings of the spectra are too compressed, again resulting in inaccuracy in applying markers for measurements. If, on the other hand, the time axis is spread too wide, it becomes difficult to evaluate the temporal course of several cardiac cycles.
Wall Filter
The wall filter defines the minimal frequency range that can be displayed. If the wall filter is set too high, low-frequency end-diastolic signals can be cut off and are not displayed. If end-diastolic signals are not dis­played in the tracings of the spectra, the setting of the wall filter must be shifted toward zero in order to ex­clude any pathology artificially created by an error in the setting of the system.
22
As described in the section on focusing (cf. p. 9), the highest resolution is attained at the focal point. The focal zone is displayed as a mark beside the image on the monitor. The position of the focus can be adapted to the plane of diagnostic interest using electronic transducers.
Orientation of the Tracings of Spectra
The basic setting of ultrasound systems generally dis­plays positive flows (toward the transducer) above the line and negative flows (away from the transducer) below the line. As a rule, however, it is desirable to dis-
Important Instrument Settings
play and measure the deflection of spectra above the baseline, regardless of the direction of flow. To accom­plish this, the spectrum often needs to be inverted.
Color-Coded Doppler
Size of the Color Window
In color-coded Doppler procedures it is often neces­sary to adapt the size of the color window (ROI) to the size of the region to be examined. The greater the selected ROI, the smaller is the refresh rate of the image. If the refresh rate is too small, a dynamic flow display becomes impossible.
PRF
Color-coded Doppler also requires that the PRF be adapted to the type of flow in the examined area. Too low a PRF here, too, leads to aliasing, which in this case is seen as color reversal.
Color Gain
If color gain is exaggerated, noise artifacts are dis­played as a mosaic of color evenly filling the whole color window. For optimal color gain, it is recom­mended that one raise the gain above the noise limit, and then reduce it gradually until noise artifacts are no longer displayed.
Basic Concepts
23