Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5817_Библиотеки_им_академика_М_И_Перельмана-1.pdf
Скачиваний:
0
Добавлен:
09.09.2026
Размер:
18 Мб
Скачать
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
1
the velocity of sound within the material. Table 1.1 provides an overview of the densities of various tissues and the sound velocities within these tissues.
The distance between two maximal compressions is the wavelength λ. The number of oscillations of a molecule in unit time is the frequency f, which is measured in hertz (Hz). Their depth of penetration and resolution capacity are listed in single oscillation per second, i.e., 1 Hz = 1/s.
Table 1.2. One hertz is a
1.2
igure
F provides an overview of the frequency range of sound waves and their applications. Ultrasound waves in the frequency range of 2−30 MHz are used for diagnostic medical imaging.
The velocity of a wave is determined by the product
of its wavelength and frequency:
c=λ × f
In medical diagnostic imaging the velocity of sound is assumed to be independent of tissue, namely a con­stant quantity of 1540 m/s.
The displacement of an individual molecule from its
resting point in diagnostic imaging amounts to about
−6
2×10 waves is typically 0.6 × 10
mm (0.000002mm). The pressure of sound
5
Pa. At this pressure each
oscillating molecule accelerates appreciably to the
5
order of 10
times gravitational acceleration.
Table 1.1 Densities, sound velocities, and attenuation of various substances
Substance Density
Fat 0.97 1470 0.5 Bone marrow 0.97 1700 Muscle 1.04 1568 2 Liver 1.055 1540 0.7 Brain 1.02 1530 1 Bone (compact) 1.7 3600 4−10 Water (20°) 0.9982 1492 0.002 Air (sea level) 0.0013 331
Table 1.2 Values for diagnostic ultrasound
Transmit­ting frequency (MHz)
2 0.78 25 3 0.8
3.5 0.44 14 1.7 0.5 5 0.31 10 1.2 0.35
7.5 0.21 6.7 0.8 0.25 10 0.16 5 0.6 0.2 15 0.1 3.3 0.4 0.15
Wave length (mm)
(g/cm
Depth of penetra­tion (cm)
Sound
2
)
velocity (m/s)
Lateral resolution (mm)
Attenuation (db/ MHz cm)
Axial res­olution (mm)
Fig. 1.2 Frequency ranges of sound waves and their applica­tions.
12
10 Hz
9
10 Hz
20 kHz
16 Hz
0 Hz
Ultrasound
Sound
Infrasound
Upper frequency limit for mechanical waves, as the wavelengths now become smaller than the distance between molecules
Medical diagnosis 2–30 MHz
Materials testing 50 kHz – 1MHz
Ultrasonic cleaners 25 kHz – 50 kHz
Sonar equipment, naval positioning 16 Hz – 16 kHz
Seismological waves

The Generation of Ultrasound

Ultrasonic waves can be generated in a variety of ways, the mechanical method being the simplest. Sounding a
4
tuning fork of less than 2 mm generates ultrasonic waves with a frequency up to 200 kHz. For engineering use such as cutting, drilling, or milling of very small
parts (e.g., to divide semiconductors) ultrasound is generated by magnetostriction. In this method a ferro­magnetic substance is placed in a magnetic field with alternating polarity, generating frequencies of up to about 50 kHz at considerable sonic intensities.
In 1880 the Curies discovered the piezoelectric ef-
fect, which is used for the generation of ultrasonic
waves in medical diagnostic imaging. If pressure is ex­erted on an ionic crystal and an elastic deformation in a defined direction is then imposed on the crystal, the internal charge shifts. Electric potentials result on its surface, negative on one side, positive on the other. The
voltage increases as the pressure increases. If, con-
versely,an electric potential is applied to the surface of a piezoelectric crystal, the latter will become elon-
gated or shortened according to the direction of the
voltage. If an alternating potential is applied, the crys­tal begins to oscillate. Quartz and tourmaline are espe­cially active piezoelectric substances. The piezoelectric materials used to build current ultrasonic probes are made of scintered ceramic such as barium titanate.
igures
F image of such a piezoelectric ceramic and a schematic
view of the multilayered arrangement used in its struc­ture.
1.3 and 1.4 show an electronmiscroscopic

Physical Effects

Basic Concepts
Fig. 1.3 Electronmiscroscopic photograph of a piezoelectric ceramic.
Fig. 1.4 Diagram of the multilayer structure of a piezoelectric ceramic.
Physical Effects
Reflection and Refraction
The propagation of sound in tissue follows the laws of optical waves. At the interface between two tissues of different densities there is a sudden change in im­pedance. Impedance is defined as the resistance Z to sonic waves, calculated by the product of sound veloc­ity × density. At such an acoustic interface, part of the incident sound wave is reflected (reflection), while another part is refracted and continues into the tissue
1.5). When the incidence of the
(transmission) ( sound wave on an interface is vertical, the reflection
gradient (R) is calculated by the formula:
Fig.
30 µm crystal layer
+ –
+ –
3 µm electrode
+ –
Reflection
Refraction
α
α
S
Fig. 1.5 Reflection and transmission of a sound wave at an acoustic interface (S = transducer; Z = impedance).
Tissue1
Z1
Tissue 2
Z2
Acoustic interface
5
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
1
The reflection gradient at the transition from liver tissue (Z can be calculated to be 0.0008. This means that less than a one hundred thousandth of the propagated sound energy is reflected at this boundary interface. At the transition from fatty tissue (Z (Z
2
lated. Over 99 % of propagated sound energy is re­flected here. Thus, all the ultrasound is reflected and behind this interface no more ultrasonic energy is available to penetrate further into the tissue. For this reason segments of gas-filled intestine and lungs can­not be reached by ultrasound, and the layer of air be­tween an ultrasonic transducer and skin must be elim­inated by contact gel.
= 1.66×105) to renal tissue (Z2= 1.63 × 105)
1
= 1.42 × 105) to air
= 43) a reflection gradient of 0.9987 can be calcu-
1
Scattering
As a rule the surfaces separating tissues are not smooth but rough or diffuse. Sound waves are not reflected directionally at diffuse interfaces but are scattered in
F
ig.
the form of a spherical wave ( smaller than the wavelength cause mainly scatter. Tissue structures exceeding the wavelength cause re­flection. Scatter echoes generate the typical textured pattern of parenchymatous organs.
1.6). Tissue structures
Interference
When two or more sound waves arrive at the same time, two waves in different phases may meet (i. e., the compression phase of one wave may coincide with the rarefaction phase of the other), and they may weaken each other. Similarly, waves in the same phase may
coincide and reinforce each other. This phenomenon is called interference, and the spatial distribution of areas of weakening and reinforcement are called inter­ference patterns. Interference patterns significantly determine the appearance of an ultrasound image.
Diffraction
When the straight propagation of a sound wave is impeded, for example, by the edge of an object, the waveis bent (diffracted) around such an edge. Thus the sound waves reach the space behind the object, which
1.7).
ig.
would normally lie in its shadow (
F
Absorption
The energy of a sound wave diminishes in the direction of itspropagation. The internal friction of the oscillating molecules is transformed into heat, i.e., the sound wave is absorbed. The values of attenuation listed in may be roughly averagedto 1 dB/(MHz cm). Absorption is dependent on frequency.On the one hand, high ultra­sound frequencies are useful because their short wavelengths provide precise localized resolution; on the other hand, it is also necessary to examine organs lying deep below the surface, and for this examination lower ultrasound frequencies with their longer wavelengths are more suited because they attenuate less. At a frequency of 10 MHz the attenuation is 10dB/ cm; at a frequencyof 3 MHz the attenuation is 3 dB/cm. Assuming an initial output of 100 dB, the depth of pene­tration at a frequency of 10 MHz can be calculated to be 5 cm (10 cm total path) and at a 3 MHz frequency the penetration would be 17 cm (33 cm total path).
Table
1.1
Diffuse interface
S
S
Tissue 1
Z1
Tissue 2
Z2
6
Fig. 1.6 Spherical scattering of a sound wave at a diffuse inter­face (S = transducer; Z = impedance).
Fig. 1.7 Diffraction of a sound wave at a diffracting edge (S = transducer).

Generating the Image

Generating the Image
Pulse-Echo Procedure
Almost all ultrasound procedures used in medical di­agnosis are based on the so-called pulse-echo pro­cedure. A brief electrical impulse is applied to the pie­zoelectric crystal of the transducer, and this impulse is transformed into an ultrasonic pulse by the piezoelec­tric crystal. The duration of the pulse is about 1 µs. The transducer is then switched to act as a receiver. The sound wave penetrates the tissue and is reflected from the internal boundary interface. The part of the sonic pulse returning to the piezoelectric element is called an echo; it elicits an electric impulse in the element.
The time differences between the emission of the ul­trasonic pulse and the reception of each echo are then measured. The product of the ultrasound velocity c and the time difference t gives the distance z covered by the ultrasonic pulse. Dividing this number into 2 gives the exact position of the reflecting structure with re­spect to the transducer:
Z = ct/2
For a time difference of, for example, 0.13ms calcula­tion shows a distance of 10cm between transducer and reflector. Depending on the construction of the trans­ducer, 3000−5000 ultrasonic pulses a second may be emitted. At the same time echoes are captured, calcu­lated, and transformed into images.
Time Gain Compensation
Because some of their sound is absorbed, energy (amplitude) from echoes traveling over a longer period of time from deeper tissues is lower than that of echoes with shorter travel times. Thus, interfaces having equal reflection gradients will emit signals of different amplitudes according to their depth. To equalize the representation of signals, those received at the ele­ment after a longer travel time are proportionally amplified. Such amplification is known as time gain compensation (TGC), or depth gain compensation (DGC), and can be controlled by the user of the ultra­sound apparatus.
Basic Concepts
A-Mode
The simplest and also the oldest imaging procedure used in diagnosis is the A-mode procedure. The ampli­tudes of the electric signals generated at the trans­ducer are displayed on a cathode ray oscilloscope, pro­portional to the distances of the boundary surfaces in the tissue being examined (Fig. 1.8). Because it repre­sents amplitude, this procedure is known as amplitude mode, or A-mode. A-mode procedures provide only one-dimensional information. Today they are still used in ophthalmology (to determine the thickness of the cornea) and otorhinolaryngology (for noninvasive ex­amination of the nasal sinuses).
Fig. 1.8 A-mode scan. The ampli-
tude of the electric signals generated at the transducer are dis­played on a cathode ray oscillo­scope. S = transducer RV = right ventricle LV = left ventricle
AO = aorta LA = left atrium
LV
S
RV
AO
LA
7
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
1
B-Mode
In contrast to A-mode, in this procedure amplitudes are represented not as deflections (peaks) but as bright spots on a monitor. The brightness steps of these spots are proportional to the amplitudes of the electric sig­nals and hence those of the echoes. The stronger the signal, the brighter is the point on the image. This pro­cedure is known as brightness mode, or B-mode. In current ultrasonic systems about 256 degrees of brightness can be represented in shades of gray (gray­scale display), though the human eye can only distin­guish 80 shades of gray. The individual brightness spots are arranged in a straight line, and the sound beam is now shifted before sending a new pulse. If the new lines acquired by this procedure are displayed corresponding to their localization, a two-dimensional
Fig.
cross-sectional image is obtained ( needed to display a line at a depth of, for example,
15cm (cf. Pulse-Echo Procedure). For the display x of a
width of 5 cm, an interval between lines of x of 1 mm, and the known ultrasound velocity c of 1540ms scan time T of 10 ms can be obtained by the formula:
T = (2zx)/(Cx)
1.9). 0.2 ms are
−1
, the
M-Mode
In contrast to B-mode display, in M-mode (motion mode), the sound beam is not shifted, but is kept in a fixed position over the organ to be examined. The in­dividual lines are again displayed side by side on a time axis. By this means it becomes possible to visualize movements as they occur in the imaged tissue
ig.
F
1.10). At a depth of 15 cm of the displayed tissue
( the scan time is 0.2 ms, i. e., an image can be generated about 5000 times a second. Thus, even very rapid movements, for example, of the heart valves can be displayed. Since the scales of the image and of the time axis are known, movements, velocities, and accelera­tion can be accurately measured. Hence M-mode is of great value in echocardiography.
From this may be calculated a repetition frequency for the image of 100 Hz, i. e., 100 separate images per sec­ond. It is therefore practicable to generate an image in real time.
S
RV
AO
LV
LA
S
8
Fig. 1.9 B-mode scan. Generation of a two-dimensional cross-sectional image. S = transducer, RV = right ventricle, LV = left ven­tricle, AO = aorta, LA = left atrium
Fig. 1.10 M-mode scan. The course of movements in the insonated field is displayed as it develops over time. S = transducer RV = right ventricle LV = left ventricle
AO = aorta LA = left atrium

The Sound Field

S
t
RV
AO
The Sound Field
Resolution
Resolution describes the smallest interval between two structures that allows them to be displayed on the monitor as two distinct objects. Resolution is measured in millimeters. Axial resolution in the direc­tion of the sound beam must be distinguished from lateral resolution, which occurs in a plane at right an-
gles to the axis of the sound beam (Table olution is determined by the pulse length of the ultra­sound beam. It usually measures one or more
wavelengths. Higher resolution can be attained by using higher ultrasound frequencies with their shorter
wavelengths. However, as demonstrated when dis­cussing absorption (p. 6), the penetration of ultra­sound diminishes with increasing frequency. Hence it is not possible to avoid the use of lower frequencies to display deeper tissue structures. Lateral resolution is
1.2). Axial res-
LV
LA
determined by the width of the ultrasound beam, i.e., it is proportional to the diameter of the sound beam. F
e 1.11 is a schematic representation of a simple ul-
igur trasonic transducer. The sound field is composed of a narrow focused near field and a divergent far field. Pre­cise scanning is only possible in the focused near field. To improve the resolving power of the ultrasonic beam it must be focused.
Focusing
An ultrasonic beam can be focused in various ways. The simplest is the use of an acoustic lens ( since essentially the physical laws of wave optics are valid for the spread of sound (cf. Reflection and Calcu-
p.
lation,
5). Hence the ultrasonic beam is maximally focused at a fixed point. This point is known as the focal point. The element may be given a concave
Fig. 1.12),
Basic Concepts
Fig. 1.11 Diagram of the sound
field of a simple ultrasonic trans-
ducer.
Near field Far field
Acoustic lens
Focal point
9
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
Fig. 1.13 Generation of a con­cave wave front by time delay of individual crystal elements in an array.
Array of elements
10
1
shape. This is known as internal focusing and is used in the single element systems of mechanical sector scanners.
The ultrasound beam is focused electronically in order to reach as variable a depth of focus as possible. Current ultrasonic transducers are built in the form of arrays, consisting of a greater number of individual ele­ments arranged next to each other. The total number of individual elements amounts to between 60 and 256, according to the type of transducer. Several of these in­dividual crystal elements are grouped together to send and receive ultrasound rays. A concave wave front can be generated by regulating the timing of individual
Fig.
elements in the group ( converges to a focal point. A sound beam generated in this way is maximally focused at this point and pro-
1.13). Such a wave front

Scanning Procedures

Principle of Operation
The principle of operation of modern scanners is shown in Figure dividual elements in an array are grouped together to generate an ultrasound beam. Such a group generates an ultrasound pulse and also receives the returning echo signals. When an element on the left side is switched on, while an element on the right side is switched off, the new group of elements generates another ultrasound ray that has been shifted by the width of one element. The linear density can also be in­creased by varying the breadth of the group: First a group of elements generates an ultrasound pulse, then one element, for example, on the left side, is switched off, while on the right none is switched on. The axis of the next ultrasound beam is now shifted relative to the previous group by the width of half an element. If, now, an element on the right side is switched off, while none is activated on the left, the number of lines of sight has been increased by a factor of 2. On the other hand, the time required to scan a complete image in­creases, i. e., the image repetition frequency is halved.
1.14. As described above, several in-
vides high lateral resolution. The focal length can be shifted by varying the width of a group and the timing of the elements. In this way the user can set the focus and with it the point of maximal resolution to the area of diagnostic interest. In modern ultrasound systems it is possible to set several such transmission foci simul­taneously. It should be noted that this requires that each sound beam must be repeated for each focal length, and this reduces the repetition frequency of the image. The focusing of the received beam has a special feature: Since the echoes from deeper regions at the transducer arrive later than those from closer regions it is possible in practice to let the receiving focus shift to the deep area. This procedure is known as dynamic receiving focusing. The groups used for focusing vary in width between 8 and 128 elements.
Array of elements
Direction of scan
Fig. 1.14 Mode of operation of a modern ultrasound scanner.
Scanning Procedures
Various scanning devices are described in what fol-
lows.
Linear Array Scanner
In a linear array scanner, also known as a parallel scan­ner, the crystal elements are arranged in a straight line
ig.
F
1.15). The individual ultrasound beams run in par-
( allel lines, generating a rectangular cross-sectional image. The resolution is more or less equally good over the whole depth being displayed. The number of ele­ments in a linear array scanner varies between 60 and 196, with a width of each element of 1−4λ. The frequency range of linear array transducers lies be­tween 5 MHz and 13 MHz. An acoustic lens is generally used to focus a linear array scanner in a plane trans-
verse to the direction of the sound beam.
Curved or Convex Array Scanner
Curved or convex array scanners are effectively a special type of linear array scanner. The mode of opera­tion corresponds to that of a linear array scanner. They differ in the curved arrangement of their elements,
1.16).
which generates a pie-shaped sound field ( Since the density of the lines of sight diminishes at a distance from the transducer, lateral resolution is re­duced with increasing depth. Typically a convex array scanner has more than 96 elements. Its radius varies between 25 mm and 80 mm and the frequency ranges between 3 MHz and 7 MHz. Most often the sound field extends over an angle of 60−90°.
Fig.
Crystal elements
Basic Concepts
Fig. 1.15 Linear array or parallel scanner. The crystal elements are arranged side by side in a straight line.
Crystal elements
Sector Scanner
This differs from the curved array scanner by having a smaller radius (25 mm). This results in a smaller ap­plication surface, a narrow near field, and an angle of departure of 90°. These properties are utilized when imaging through a small sound window such as the in­tercostal spaces in echocardiography or in transvaginal sonography.
Phased Array Scanner
The arrangement of elements in a phased array scan­ner is the same as that in a linear array scanner. However, instead of a group of elements, all the ele­ments take part in the generation of a scanned line of sight.
It is possible to generate a wavefront running at an
angle to the surface of the transducer surface by
Fig. 1.16 Curved or convex array scanner. The crystal arrange­ment is curved, generating a pie-shaped sound field.
using a time-delayed firing of the elements
F
1.17).
ig.
(
A pie-shaped sound field is generated by changing
the settings. Phase array transducers have a small contact area of 12−20 mm using between 64 and 128 elements.
The sector angle is between 80° and 90°. The
frequency range is between 2 MHz and 7 MHz.
These transducers are very costly because of their complex electronics and are used chiefly in cardiology and transcranial sonography.
11
1 Physical and Technical Fundamentals of Ultrasound and Doppler Ultrasound
Mechanical Sector Scanners
1
T2 T1
W1
W2
Fig. 1.17 Phased array scanner. By using time delay the generated sound waves are made to run at an angle to the transducer surface (T = point in time; W = sound wave front).
Single element
In contrast to the electronic phased array scanner, a mechanical sector scanner requires little investment in its regulation and signal processing. The resulting favorable price−performance ratio has allowed me­chanical transducers to compete with electronic trans­ducers in low-cost systems. Mechanical sector scan­ners are divided into rotary transducers and wobbler transducers.
Rotary Principle
Figure 1.18 illustrates schematically the operation of a rotor transducer. Three to five single elements are usu­ally arranged on a rotor at equal angular distances. A motor in the handle turns the rotor. The element rotat­ing past the sound window is activated and covers a sector-shaped sound field. The next element then ro­tates past the window and generates a second image.
12
Single element
Fig. 1.18 Diagram of the operation of a rotary transducer. Three to five individual elements mounted on a rotor at equal intervals generate a pie-shaped sound field as they rotate past the sound window.
Fig. 1.19 Wobbler transducer. A single element defines a sec­tor between 60° and 100° by angulating to and fro.

Ultrasound Artifacts

Wobbler Principle
In a wobbler transducer a single element angulates back and forth, covering a sector of between 60° and
F
ig.
100° ( considerably less energy is required to regulate it. In addition, the sector angle can be adjusted, in contrast to a rotary transducer. Because both types of trans­ducer use only one element at a time, their focus is al-
ways fixed.
1.19). Since only a single element is used,
Annular Phased Array Transducer
The annular phased array transducer works on the wobbler principle, and combines a mechanical and
electronic scanner. It consists of an array of ring ele-
Ultrasound Artifacts
Ultrasonic imaging is plagued with far more artifacts (imaging errors) than other diagnostic imaging pro­cedures such as computed tomography (CT) or mag­netic resonance imaging (MRI) scans, because the as­sumed values of parameters, such as sound velocity, straight line sound propagation, attenuation, often de-
viate from actual values. Artifacts may also be due to inadequate instrument settings. However, experience has shown that some of the important artifacts de­scribed below may be diagnostically useful and can provide additional information about the properties of the examined tissue.
ments arranged concentrically inside each other in­stead of single elements. Each individual ring can be regulated separately. This allows variable focusing in two dimensions (cf. Focusing, p.
9).
Disadvantages of Mechanical Scanners
Regardless of their principle, mechanical sector scan­ners are subject to wear and require maintenance. Moreover, rapid switching between scan modes (B­mode, M-mode, Doppler) is impossible because of in­ertia. In general, real-time display of B-mode/M-mode or B-mode/Doppler cannot be performed.
wave leaving, for example, a cyst retains nearly as much energy as it had when entering it. The area be­hind the cyst therefore appears brighter than the sur­rounding tissue ( in differential diagnosis.
Fig. 1.21). Such an artifact can be used
Basic Concepts
Distal Acoustic Shadowing
One of the most commonly found artifacts is distal acoustic shadowing. When meeting strong reflectors, i.e., structures with a resistance to sound waves that
greatly deviate from that of the surrounding tissues (e.g., air) or structures that strongly attenuate sound energy such as bone or calculi, the greater part of the ultrasound energy is reflected or absorbed (cf. Reflec­tion and Calculation, p. tor obviously less energy is available than in the sur­rounding tissue. This phenomenon is called an acoustic shadow (
Fig. 1.20).
5). Behind such a strong reflec-
Dorsal Sound Amplification
With optimal TGC (cf. p. 7) the phenomenon of dorsal sound amplification may be observed behind areas offering weak attenuation. In hollow fluid-filled spaces ultrasound is subject to less reflection and absorption than in the surrounding tissues. Thus, an ultrasonic
Fig. 1.20 Distal acoustic shadowing. Mode of generation (left) and simplified diagram of the ultrasound display (right).
13
Fig. 1.21 Dorsal sound amplification. Mode of generation (left) and simplified diagram of the ultrasound display (right).