78
B.M. Howard and D.B. Christie III
III. Sounds again become crisper and louder. IV. Sounds are muffl ed, less distinct, and softer. V. Sounds disappear completely (diastolic pressure).
The systolic blood pressure corresponds to the pressure at which the sounds fi rst appear at phase I. The diastolic blood pressure corresponds to the pressure at which the sounds dis­appear at phase V [ limit the intensity of audible sound and can lead to falsely depressed systolic blood pressure recordings.
7 , 8 ]. Low-fl ow states or hypotension can

Considerations: Oscillometric/Automated Method

Oscillometric measurements have been reported to be con­sistently inaccurate when compared to invasive or direct blood pressure recording modalities. The disparity may be in part to mal-fi tting cuffs or individual manufacturer device variation [ 9 , 10 ]. Calcifi c arterial structures also impede the device’s capture of arterial oscillation amplitude resulting in falsely depressed blood pressure values. Automated oscillo­metric devices are quite accurate when capturing blood pressure values in normotensive patients but seem to overes­timate values in hypotensive states (i.e., systolic blood pres­sure less than 80 mmHg) and are consistently lower than direct blood pressure measurements achieved by arterial catheterization [ 9 , 10 ].

Gastric pH Monitoring

Tonometry is the measurement of the partial pressure of a gas in a fl uid or intravascular medium. Gastric tonometry measures luminal CO2 and mucosal CO2 partial pressures (pCO2) as they become equalized across a semipermeable membrane [ 8 ]. The general tenet of gastric pH monitoring projects that a decreased mucosal pH implies either an increase in anaerobic metabolic efforts or a low systemic fl ow state leading to tissue hypoperfusion and, subsequently, an ischemic insult.
Two general methods have been described to perform gastric pH monitoring, conventional gastric tonometry and air-semicontinuous gastric tonometry. Conventional gastric tonometry is the more commonly described technique and involves a transorally inserted catheter or tube with a saline­fi lled balloon often composed of silicon or latex polymers, as these agents are permeable to CO2. The saline PCO2 and the gastric mucosal PCO2 come to equilibrium. The PCO2 within the saline-fi lled balloon is now considered physiolog­ically representative of the gastric mucosal CO2 and is mea­sured using a blood gas analyzer. Utilizing the Henderson-Hasselbalch equation, and assuming that the par-
tial pressure of the stomach mucosa and atrial bicarbonate are equal, the intramucosal pH (pHi) can be calculated and perfusion trends established [ 11 ].
The air-semicontinuous gastric tonometry uses an air­fi lled CO2 permeable balloon and rapidly equilibrates with luminal and mucosal PCO2. This is directly measured in a semicontinuous manner, by an infrared sensor.
Considerations
The luminal and mucosal PCO2 generally equilibrate in approximately 30 min, although equilibration times up to 90 min have been reported [ 12 ]. A pHi of <7.32 represents mucosal acidosis and can be trended over time. Studies have found that H2 blockers have not been shown to infl uence lumi­nal PCO2, luminal pH, and pHi in the shock state [ 13 , 14 ].
Two major assumptions enter into the calculation of pHi
8 , 11 ]. It is assumed that the PCO2 measured through gas-
[ tric tonometric techniques approximates that of intramucosal PCO2. Secondly, it is assumed that intramucosal and arterial HCO3 measurements are the same. Animal studies exist that support the notion that no differences exist between luminal and mucosal CO2 measurements; however, arterial HCO3 measurements are likely an overestimation of the mucosal HCO3 in the ischemic or shock state [
Clinical investigators have shown that guiding resuscita­tions with gastric tonometry compared to traditional resusci­tation techniques do not yield improved outcomes in either overall survival rates or development rates of multi-organ system failure [ 15 , 16 ]. At one point, gastric tonometry seemed poised to evolve into a monitoring technique for the prevention of mesenteric ischemia insults during prolonged vasopressor use in the setting of refractory shock [ 8 ]; how- ever, consistent clinical application of the technique is lacking.
11 ].

Sublingual Capnography

Similar to gastric PCO2 monitoring, sublingual PCO2 (PslCO2) is measured against the arterial PCO2 in order to defi ne a PslCO2 gap. Interestingly, investigators have dem­onstrated that the PslCO2 correlates closely with those obtained in the stomach during low-fl ow states [ 1719 ]. Sublingual capnography does not directly refl ect the perfu­sion quality to the splanchnic vasculature, however; the sub­lingual mucosal vascular bed shares similar characteristics with the splanchnic circulation, including a rich vascular supply and parasympathetic neural input.
A variety of commercially available devices have been manufactured and proved reliable in clinical application. One of the more widely used devices employs a specifi c CO2 sensing optode consisting of a CO2-permeable silicone bag with a buffering solution and a fl uorescent dye solvent. The
8 Hemodynamic Monitoring in Surgical Critical Care
79
silicone bag is attached to a light source that captures the changes in wavelengths of the fl uorescent solvent as it is acted upon by the pH of the buffering solution. This is con­verted to the measurement of PslCO2. A key to accurate measurement capture is that the patient’s mouth must be closed around the monitor [ 8 , 20 ].
Considerations
Initial research of sublingual capnography was predomi­nately with animal models, but consistent correlations between rising PslCO2 and systemic markers such as arterial lactate were identifi ed [ 17 ]. Comparative animal studies where gastric tonometry and sublingual capnography were performed noted parallel increases in both measured gastric PCO2 and PslCO2 gap [ 18 ]. Later identifi ed was that PslCO2 values greater than 70 mmHg were predictive of hospital sur­vival and that PslCO2 corrected more rapidly than lactate. A small study found that PslCO2 gap was a better predictor of survival than lactic acid. A recent study has also found cor­relation between hemorrhage in penetrating trauma and rise in PslCO3, coinciding with increases in pre-resuscitation base defi cit and lactate levels [ 2123 ].
infl uenced by volume status but by fl uctuations of venous tone, presence of tricuspid regurgitation, congestive heart fail­ure, right-sided diastolic dysfunction, pulmonary embolism, pulmonary hypertension, acute respiratory distress syndrome, chronic obstructive pulmonary disease, and mechanical venti­lation with positive end-expiratory pressure.
A zero reference point exists for central venous pressures in the thorax located at a point on the external thorax where the fourth intercostal space intersects the midaxillary line, the phlebostatic axis. This anatomical landmark corresponds to the position of the right and left atrium when the patient is in the supine position; therefore, in the lateral position, it is not a valid reference point and will provide an inaccurate CVP reading [
More contemporary research has challenged the utility of CVP as reliable hemodynamic monitoring method fi nding no association between CVP and circulating blood volume and identifying that CVP does not predict fl uid responsive­ness across a wide spectrum of clinical conditions [ 25 ]. Because of these infl uences, CVP must be interpreted with concern for variations in measurement with general recom­mendations encouraging the capture of trends in measure­ment rather than the absolute value.
1 , 24 ].

Central Venous Pressure Monitoring

Central venous pressure (CVP), as a measure of pressure in the superior vena cava, allows for the extrapolation of right­sided ventricular fi lling pressure and has been used to esti­mate intravascular volume and determine fl uid requirements in critically ill or injured patients for decades. CVP became rather ubiquitous as a hemodynamic monitoring modality due to its ease of insertion and seemingly straightforward measurement parameters.
While CVP provides the ability to estimate right-sided cardiac fi lling pressures, it is not a direct assessment of car­diac performance or output, nor a refl ection of end-organ perfusion adequacy. CVP can be measured from three sites of central venous access: the internal jugular vein, subcla­vian vein, and the femoral vein. However, the femoral vein is deemed less optimal as proper catheter positioning calls for the catheter tip to be situated at the atriocaval junction, a position unachievable with most commercially available catheters through femoral vein access. Furthermore, it should be noted that increases in intra-abdominal pressure or pres­sure of the accessed extremity from injury, edema, or infl am­mation can erroneously elevate values.
Considerations
Normal CVP in a critical care patient is estimated to range from 0 to 9 mmHg with a standard waveform being composed of three waves (a, c, and v) and two descents (x and y) (Fig.
8 ]. As a measure of intravascular volume, CVP is not only
[
8.1 )

Pulmonary Catheter Monitoring

Pulmonary artery catheters (PACs) have historically been the defi nitive clinical tool for assessing hemodynamic perfor­mance and responsiveness to therapeutic interventions. PACs were introduced into clinical practice in the 1970s predomi­nately as a monitoring method for patients with acute cardiac insults and transitioned to use as a monitoring instrument for complex resuscitation scenarios, shock states, severe cardio­vascular dysfunction, respiratory failure, and intraoperative fl uid management.
The PAC allows for the capture of hemodynamic data that assists in defi ning the cardiovascular status of a systemic insult through the generation of physiologic profi le.
The measurement for Q, or cardiac output, and the calcu­lation of systemic vascular resistance (SVR) was touted to enable one to distinguish hypotensive states such hypovole­mic shock, cardiogenic shock, or a form of distributive shock (i.e., sepsis). Assessment of pulmonary artery occlusion pressure (PAOP) potentially provides for the differentiation between cardiogenic and noncardiogenic pulmonary edema processes. PAOP may also be used as a guide for fl uid resus­citation efforts, estimating left ventricular preload, while trending Q theoretically predicts the need for, or appropriate delivery of, inotropic support. In clinical scenarios of sys­temic infl ammatory response syndrome or sepsis with end­organ dysfunction, calculations of mixed venous oxygen saturation and oxygen delivery variables may assist in
80
A-line
ECG
B.M. Howard and D.B. Christie III
P
CVP
Atrial
contraction
Systole
aa
c
V
x
Isovolemic ventricular
contraction
Ventricular
contraction
y
Passive
Diastole
atrial filling
Passive
ventricular
filling
Fig. 8.1 A-line, ECG, and CVP tracings with cardiac function
identifying the presence of tissue ischemia and further direct resuscitation strategies [ 8 ].
Access sites for PAC insertion are similar to those of cen­tral venous catheters, but proper positioning of the PAC into the pulmonary artery can be challenging and is infl uenced by many factors. Volume-depleted states, aberrant vascular anatomy, and the diffi culties traversing normal vascular and cardiac contours can create diffi culty achieving appropriate
catheter placement, which is key to obtaining accurate physi­ologic readings. The most direct access site is the right inter­nal jugular vein, where the PAC only requires one gentle curve through the right ventricle to reach the right pulmo­nary artery. Left and right subclavian vein access requires the catheter to cross the innominate vein to reach the superior vena cava. Access through the left internal jugular vein also requires the catheter to cross the innominate vein and
8 Hemodynamic Monitoring in Surgical Critical Care
81
demands a greater travel distance. Femoral vein sites are possible alternatives but are often very diffi cult to pass into the right atrium making this access approach the least desir­able and often requiring fl uoroscopic control for placement.
After gaining central venous access, a stepwise approach is used to ensure proper catheter positioning. The catheter is fl ushed, zeroed, and advanced with the balloon defl ated to a distance of 12–15 cm. The PAC balloon is infl ated, the PAC is advanced, and the pressure waveforms measured at the tip of the PAC are analyzed. The initial waveform is that of a CVP and as the catheter is advanced will convert to a right atrial pressure waveform reading. Within the right ventricle, a high spiking pressure waveform of 20–30 mmHg is encountered, representing the systolic right ventricular pres­sure. Further advancement of the PAC into the pulmonary artery generates a characteristic step-up in diastolic pressure and a clear dicrotic notch within the waveform. When the infl ated balloon reaches a pulmonary arteriolar occlusion position, the waveform fl attens. At this point, the balloon should be defl ated to confi rm the reappearance of the pulmo­nary artery waveform. With the PAC tip now in an appropri­ate position, an occlusion waveform will be obtained with balloon infl ation of 1.5 mL of air [
8 ].
Considerations
If the occlusion waveform is obtained with less than 1.5 mL, the catheter tip is likely too distal and located in an inappro­priate branch. With the balloon defl ated to prevent pulmo­nary artery rupture, the catheter should be withdrawn several centimeters, followed by reinfl ation of the balloon and re­advancement back to the occlusion position. After register­ing an occlusion pressure, it is safest to pull the catheter back a few centimeters (with the balloon down), each time between readings to minimize the natural distal migration of the catheter [ 8 ]. PAOP is most precisely measured when there is minimal infl uence from external transmural pres­sures such as increased intrathoracic pressure. Therefore, accuracy is best achieved when the lungs return to their most relaxed volume state at end-expiration [ 8 ].
In review, the lungs are divided into three physiologic zones, based upon the relationship between the pulmonary vasculature and alveolar pressures. In lung zones (i.e., West zones) 1 and 2, the magnitude of alveolar pressure affects the pressure recordings of the PAC. The ideal intravascular pres­sure always exceeds the alveolar pressure, best achieved in zone 3 of the lung. In non-zone 3 circumstances, alveolar pressures may exceed intravascular pressure; therefore, the degree of alveolar pressure and PEEP may affect the mea­surement of PAOP [ 8 ].
The clinical application of the hemodynamic data gener­ated from the PAC placement requires careful consideration of how the information relates to the given clinical circum­stance. PAC hemodynamic profi les have been patterned to
suggest several broad clinical scenarios of hemodynamic aberration. However, in current medical practice, contro­versy surrounds PAC use as the devices’ safety and reliabil­ity have come into question. The popularity of the PAC has progressively declined over the past decade, as data has accumulated indicating that PAC-directed care does not improve outcome and may in fact worsen them [ 26 , 27 ]. PAC has also been associated with an increased incidence of pul­monary emboli in randomized controlled trials [ 26 ]. The profi cient and safe utilization of the PAC is directly depen­dent on the clinicians’ understanding of the measurements provided, the methods through which the data was derived, and how to deliver the proper therapeutic adjustments based on the interpretable data. It is these variables of PAC use that have come under scrutiny [ 28 , 29 ].

Tissue Oxygenation

The ultimate goal of hemodynamic monitoring is to ensure adequate tissue perfusion and subsequent oxygen delivery to end organs in order to optimize performance in the physio­logic stressed state; however, there are no monitoring modal­ities that can directly measure tissue oxygenation. When cellular metabolic demands for oxygen outpace its supply or delivery, a condition termed dysoxia results with its clinical manifestation being that of systemic shock [ 30 ]. Dysoxic states result from cardiogenic and hypovolemic shock states as the oxygen supply is insuffi cient or in contrast, with septic shock where mitochondrial oxygen utilization appears impaired.
Tissue Oxygenation: VO2
The rate of oxygen uptake from the systemic capillaries (VO2) is an indirect measure of oxygen availability in the tissues [ 1 ]. As oxygen is not stored in tissues, the VO2 can also be interpreted as a measure of tissue O2 consumption. Monitoring of VO2 has been suggested as a means to assist in determining the tissue dysoxia that results from hypoper­fusion with inadequate oxygen supply to the tissue level. VO2 = Q × (CaO2 − CvO2).
Considerations
VO2 can be calculated or measured. The calculated VO2 is captured from a physiologic profi le from an indwelling pul­monary artery catheter. The directly measured VO2 requires specifi c equipment and staff trained with the modality use and is not pervasively available in most critical care units, making this method less practical. An abnormally low VO2 (less than
2
100 ml/min/m dysoxia due to impaired tissue oxygenation [
) can be the result of hypometabolism or tissue
1 ].
Calculated VO2 cannot be interpreted as a measure of
whole-body VO2 as it does not include the VO2 of the
82
B.M. Howard and D.B. Christie III
lungs. During pulmonary infl ammatory conditions such as acute respiratory distress syndrome or pneumonia, the VO2 of the lungs has been estimated 20 % of the whole-body VO2, allowing for signifi cant underestimation of whole­body VO2 [
The derivation of VO2 utilizes four different variables (cardiac output, concentration of hemoglobin in blood, and the percent oxyhemoglobin saturation in arterial and mixed venous blood) and each has its own inherent variability [ 1 ]. The variability of the calculated VO2 is 18 % that is equiva­lent to the summed variability of its components [ 1 , 3234 ]. The calculated VO2 must change by at least 18 or 20 % for the change to be considered signifi cant [ 1 ]. Correction of VO2 defi cits will therefore require discerning clinical data relative to anemic states, inadequate cardiac output, physi­ologic hypoxemia, and augmenting these components appropriately.
VO2 may not appropriately refl ect aerobic metabolism in the septic clinical scenario as tissue oxygenation is not impaired in the septic state [ 1 ]. The accelerated activity of neutrophils and macrophages is accompanied by a marked increase in cellular oxygen consumption coined the respira­tory burst. The exaggerated oxygen consumption in this pro­cess is used to generate toxic oxygen intermediates (superoxide radical and hydrogen peroxide) that are elabo­rated as part of the infl ammatory process contributing to the septic state [ 35 ]. Therefore, VO2 will be artifi cially elevated, through the contribution of the respiratory burst, and inac­curately refl ect metabolic VO2 or the rate of aerobic metabolism.
1 , 31 ].
Tissue Oxygenation: Mixed Venous O2 Saturation (SvO2)
Mixed venous oxygen saturation of hemoglobin (SvO2) can be utilized to assess the balance of satisfactory systemic oxy­gen delivery and effective oxygen uptake. Derived through the established relationship between the rate of tissue oxy­gen uptake (VO2) and variations in systemic oxygen delivery (DO2), extrapolating SvO2 can imply the effi ciency of sys­temic oxygen delivery [ 1 ]. VO2 = DO2 × (SaO2 − SvO2). The SaO2 and SvO2 are the oxygen saturation of hemoglobin in arterial and mixed venous blood, respectively. The difference in the two variables equates to the degree of oxygen extrac­tion from hemoglobin in capillary blood.
Considerations
Normal values for SaO2 and SvO2 are 95 % and 70 %, respectively, implying a satisfactory O2 extraction of 25 % or, alternatively, that 25 % of the hemoglobin molecules have desaturated as they traverse the capillaries and microvascula­ture of end-organ targets. A decrease in SvO2 below 70 % indicates that systemic O2 delivery is impaired (or utilization is increased) while a SvO2 below 50 % indicates a global
state of tissue hypoperfusion or impending dysoxia. The pos­sible sources of impaired O2 delivery are identifi ed by deter­minants of DO2 DO Q Hgb SaO2134 2=´ ´ ´ in DO2 can be the result of a low cardiac output (Q), anemia (Hgb), or hypoxemia (SaO2) [ 1 ]. Increased O2 utilization is related to hypermetabolism.
The measurement of SvO2 historically required a pulmo­nary artery catheter as blood from the pulmonary artery is considered to be admixed with venous blood from all tissue beds. A PAC equipped with an infrared sensor at the distal catheter tip was later developed that provided the enhanced feature of performing continuous measurements of the SvO2 in the pulmonary artery blood. Continuous monitoring of SvO2 with this more specialized pulmonary artery cathe­ter has revealed a spontaneous variation that averages 5 % but can be as high as 20 % [ 1 , 36 ]. In general, a greater than 5 % change in SvO2 that persists longer than 10 min is con­sidered a signifi cant change [ 1 , 37 ].
Central venous catheters were subsequently developed that allowed for SvO2 for continuous monitoring but have fallen out of favor as the utility of SvO2 as a reliable resus­citation measure has come into question. More recently, transcutaneous devices have been engineered to measure radial artery and ipsilateral digital artery pressures and extrapolate SvO2 through computerized computation meth­ods providing a noninvasive option for capturing this hemo­dynamic data.
()
.
. A decrease
Tissue Oxygenation: Central Venous O2 Saturation (ScvO2)
Central venous O2 saturation (ScvO2) has become an accept­able alternative to SvO2 for monitoring tissue oxygen deliv­ery and utilization as the modality does not require a PAC and can be drawn from central venous catheter. The two methods are reported to correlate well with an estimated variation up to 5 % if multiple measurements are aggregated [ 38 ]. However, if comparing single measurements, ScvO2 can differ from SvO2 by as much as 10 % implying that establishing a trend in measurements of ScvO2 is prudent prior to delivery diagnostic and therapeutic decisions [ 1 ].
The ScvO2 became an acceptable substitute measure of mixed venous O2 saturation because it precludes the expense and technical challenges associated with the PACs. Previously established guidelines for the manage­ment of patients with severe sepsis and septic shock have included achieving an ScvO2 of greater than 70 % as a therapeutic end point [ 39 ]. One should note that the foun- dational basis of this value stemmed from the early goal­directed therapy trial whose conclusions have been directly challenged by three multicenter and randomized trials demonstrating no specifi c outcome advantage to proto­colized as opposed to non- protocolized care for those with severe sepsis or septic shock.
8 Hemodynamic Monitoring in Surgical Critical Care
83

Direct Perfusion Measurement

Invasive monitoring of arterial pressure was fi rst performed in 1733 with the cannulation of a horse’s left crural artery by Reverend Stephen Hales [ 8 , 40 ]. In 1865, the fi rst human inva- sive blood pressure measurement was taken by Dr. Faivre, a French physician, who cannulated an amputee’s superfi cial femoral artery [ 8 ]. The routine application of invasive arterial blood pressure measurement did not become commonplace until the mid-1960s, associated with the expansion of cardio­thoracic and vascular surgery as well as the Angiocath can­nulation technique, innovated by Barr in 1961 [ 41 ].
Intra-arterial cannulation with continuous blood pressure transduction is considered the gold standard for arterial blood pressure monitoring. While there are many indications for invasive arterial monitoring, the most common is the need for continuous blood pressure recordings in a patient at risk for cardiovascular instability [ 8 , 42 ].

Considerations

Mean arterial pressure is the true driving pressure for periph­eral blood fl ow and does not alter as the pressure waveform migrates distally in a vascular structure, nor is it subject to distortions generated by recording systems, thus making mean arterial pressure superior to systolic pressure for arte­rial pressure monitoring [ 1 ]. In contemporary medical prac- tices, mean arterial pressure is generally measured, although estimated values can be achieved. Modern electronic moni­toring devices can measure mean arterial pressure by inte­grating the area under the pressure waveform and dividing this by the duration of the cardiac cycle (Fig. 8.1 ) [ 1 ]. Intuitively, the electronic measurement is more accurate, reliable, and preferred to the estimated mean pressure. Estimated mean arterial pressure is derived as the diastolic pressure plus one-third of the pulse pressure; a formula is based on the assumption that diastole represents two-thirds of the cardiac cycle, which corresponds to a heart rate of 60 beats/min. Therefore, heart rates faster than 60 beats/min, which are common in critically ill patients, lead to inaccura­cies in the estimated mean arterial pressure [ 1 ].
Essentially any peripheral artery can be accessed as a can­nulation site. Most commonly used sites include the radial, femoral, dorsalis pedis, brachial, and axillary arteries. The appropriate site selection for arterial cannulation is infl u­enced by many factors. Typically the radial artery of the non­dominant hand is utilized [ 8 ]. When this is unsuitable, alternate sites are chosen based upon the following four cri-
1 ] the artery should be large enough to accurately
teria: [ refl ect systemic blood pressure [ free of cellulitis or nearby infected or devitalized tissue [ 3 ], there should be suffi cient collateral fl ow to prevent distal
2 ], the chosen site should be
ischemia, and [ tomic aberrations [
4 ] the artery should be proximal to any ana-
8 ].
Invasive arterial blood pressure monitoring carries two categories of complications, infectious or mechanical. The most common mechanical complication is arterial thrombosis, which may be partial or complete, and has been noted to occur in 25 % of patients [ 43 ]. However, despite this seemingly high thrombosis risk, investigators identifi ed no signifi cant disability or ischemic damage occurred in most catheter-associated radial artery throm­bosis and noted most recanalized within 2–4 weeks after catheter removal [ 8 ].
Many clinicians recommend performing a modifi ed Allen’s test prior to radial artery catheter site selection in order to assess for adequacy of perfusion and collateral fl ow to the distal extremity. The modifi ed Allen’s test consists of compression of both the radial and ulnar arteries, clinching the patient’s fi st until it blanches, then relaxing the fi st and releasing pressure from the ulnar artery. If the patient’s skin becomes hyperemic, the collateral fl ow is deemed satisfac­tory for radial artery cannulation [ 8 ]. While historically rec- ommended, the Allen’s test has not demonstrated a reduction in digit ischemia, nor has an abnormal Allen’s test proven to guarantee distal ischemic changes. However, if a grossly abnormal Allen’s test is obtained, arterial cannulation should be sought elsewhere [ 44 ].
Brachial artery cannulation has conventionally been avoided due to the anatomic consideration of limited col­laterals at this level of the upper extremity, theoretically predisposing to increased risks for thrombosis and extrem­ity loss. While an anatomically sensible consideration, research has found no increased Doppler evidence of thrombosis or vascular compromise in a study of brachial artery cannulations, leading clinicians to consider brachial artery cannulations as an acceptable risk in the proper clini­cal scenario [ 45 ]. However, even with this documented sup- port, it is generally recommended to access the axillary artery preferentially to the brachial artery due to its gener­ous collateral circulation.
An additional mechanical complication encountered with arterial cannulation is pseudoaneurysm formation. Improper technique is believed to contribute to pseudoan­eurysm development if the posterior wall of the vessel is punctured or when multiple passes at the artery are made, enhancing the risk. Prolong catheter placement and can­nulation site infection also predispose to pseudoaneurysm formation [ 8 ].
Arterial cannulation infection is the second signifi cant category of complications with the insertion site serving as the most common access for catheter-associated bacteremia. Compared to central venous catheters, research has demon­strated that arterial catheter infections do not increase over time of indwelling, and disciplined aseptic technique on
84
B.M. Howard and D.B. Christie III
insertion, access, and dressing care minimizes the infectious risk [
46 ]. One should also be cognizant of impedance mis-
match between the patient’s artery and the catheter which may lead to over- or underestimation of systolic and diastolic values. Moreover, erroneous placement with partial occlu­sion may similarly lead to inaccurate pressure measurement. Therefore, one should always assess the waveform quality, to look for whip artifact or a dampened waveform before using the numeric values to trigger intervention.

Direct Visualization Methods

Transthoracic Echocardiography

Transthoracic echocardiography (TTE) has been widely accepted as an excellent, noninvasive, direct visualization modality for the assessment of cardiac anatomy and func­tion. Point of care TTE has been utilized successfully in ICUs for many years, but more recently, intensivists have strongly advocated for its incorporation as a standard of care for hemodynamic monitoring purposes [ 47 ]. However, point of care TTE, or intensivist-performed TTE, should be distin­guished from a formal, cardiology-driven TTE in several respects. Formal cardiology-driven TTE creates a detailed anatomic report with cardiac functional descriptors relative to the given moment of the exam. Point of care TTE in the ICU is a more limited exam anatomically but allows for the assessment of cardiac fi lling volumes and chamber perfor­mances over a continuum of time [ 48 ]. In the critical care setting, TTE can be quickly applied to assess for cardiac fi ll­ing volumes, chamber functionality, global wall motion abnormalities, and cardiac responsiveness to therapeutic efforts. The exam is quick, noninvasive, and repeatable on demand.
Image Acquisition
Point of care TTE can be performed on any ultrasonographic device with cardiac penetrance. Most modern critical care units have access to advanced ultrasonographic equipment often with exchangeable probes and various examination modes, such as breast, vascular, or cardiac that adjusts pen­etration needs based on exam site and type.
TTE transducers have a square surface, as opposed to the elongated, rectangular-shaped transducer used for vascular imaging. An inverse relationship exists between sound wave frequency and depth of penetration, with better resolution being achieved with high frequency. Higher-frequency, lower-depth probes are used for vascular imaging with ultra­sound. By contrast, an echocardiography transducer emits a variety of frequencies to provide optimal resolution at a range of depths. It also uses a high frame rate to capture car­diac motion and display it smoothly [ 49 ].
Examination Views
The views obtainable are essentially those of a standard TTE and include the parasternal long axis (PLA), parasternal short axis (PSA), apical four chamber, and subxiphoid win­dows (Fig. 8.2 ):
• Subxiphoid view: The subxiphoid view allows assess­ment of the IVC, its diameter, and diameter variation with respiration.
• Parasternal long-axis view: The PLA is obtained with prone positioning of the patient between the second and sixth intercostal spaces with the transducer notch oriented toward the right shoulder. Through this view the mitral valve, aortic valve, and left ventricular function can be assessed.
• Parasternal short-axis view: The PSA view is captured with prone positioning of the patient between the second and sixth intercostal spaces with transducer probe now rotated 90°. Through this window the classic donut view of the left ventricle is seen along with the right ventricular chamber. Tilting of the probe allows imaging of the left ventricle from base to apex revealing the papillary muscu­lature, mitral, and aortic valve.
• Apical four-chamber view: The apical four-chamber view is visualized with prone positioning of the patient with the probe placed over the left chest wall at the level of the cardiac apex. The transducer notch is turned to the bed. This view allows for visualization of the left and right ventricular functions.
• Subxiphoid view: The subxiphoid view is obtained by placing the probe under the xiphoid process with the patient in supine positioning. The view is the same as that achieved with a routine assessment with sonography for trauma exams and allows for visualization of the IVC and for presence or absence of pericardial effusions. Rotating the probe counterclockwise opens the view so that the IVC is in its long axis, allowing assessment of its diame­ter and collapsibility.
Image Interpretation
The information obtained is to be interpreted as a physio­logic assessment relative to a given clinical scenario, allow­ing for the analysis of cardiac function, preload, afterload, and cardiac anatomy:
• Cardiac function can be assessed by evaluating ejection fraction, stroke volume, cardiac output, cardiac index, diastolic function, and right-sided heart function.
• Preload can be analyzed through the calculation of ven­tricular cavity size, IVC diameter, and diameter change.
• Afterload is estimated with interpretation of systemic vas­cular resistance and systolic pulmonary artery pressure.
8 Hemodynamic Monitoring in Surgical Critical Care
85
Fig. 8.2 Transthoracic echocardiography probe positions and corresponding views
• Cardiac anatomy such as severe valvular dysfunction resulting in hemodynamic compromise, pericardial fl uid, or global wall motion abnormalities can be identifi ed.
systole, to be entered whereupon physiologic data such as CO, CI, EF, SVR, etc., may be generated. IVC diameter and variation can be measured manually and analyzed. Alternatively, the skilled echocardiographer can visu-
Software-driven computations provided by manufac-
turers of advanced ultrasound devices are imbedded in their individual operating systems which allow for exam fi ndings such as ventricular chamber sized in diastole and
ally assess for chamber size, variation, and responsive­ness to therapy in order to estimate hemodynamic status. The technique of image capture and data point entry for algorithmic analysis for TTE is beyond the scope of this
86
B.M. Howard and D.B. Christie III
chapter but is readily available in a variety of well-estab­lished teaching modules or manufacture device- specifi c literature.
Considerations
The TTE examination can be limited by diffi culty in obtain­ing clear images in the presence of subcutaneous emphysema, anasarca or signifi cant edematous states, large patient body habitus, or presence of abdominal wall or thoracic bandages.
A variety of physiologic data can be gleaned from a thor-
ough TEE exam; however, guidelines for provider applica­tion have yet to be standardized. IVC size and index of change are currently being applied to hemodynamic moni­toring scenarios and studied as a marker for hemodynamic performance. Many researchers and clinicians alike have endorsed the analysis of IVC size and variability of change with therapeutic interventions as a guide for determining fl uid needs and responsiveness to therapy [
5052 ]. Others
utilize a composite of all views relative to patient perfor­mance and base clinical decisions of measurable data such as computation-derived CO, EF, SVR, as well as other related variables.
Whether computation-driven measurements or visual estimation is utilized is a provider or operator preference. Research is ongoing to identify the validity of both methods as well as domains of intersection. Specifi c pattern recogni­tion for paradigms of cardiac evaluation relative to shock management is beyond the scope of this chapter but is avail­able through a variety of references and reviews [ 49 ].

Transesophageal Echocardiography

Transesophageal echocardiography (TEE) is utilized in the management of the critically ill over the last decade due to its accuracy and reproducibility as an assessment tool for car­diac anatomy and performance. However, formal multiplane TEE requires a full-scale TEE module and monitor system, utilizes relatively bulky probes, requires probe reinsertion for each exam, and potentially limits the clinicians’ ability to accurately follow changes in cardiopulmonary performance over time. Since the US Food and Drug Administration’s approval for the clinical use of a newly developed, miniatur­ized TEE probe that may remain indwelling in the esophagus for up to 72 h, TEE as an application in the critical setting has become more pervasive. Contemporarily known as hemodynamic TEE or hTEE, this exam modality has the unparalleled advantage of allowing for anatomical and func­tional assessments of the heart and great vessels in real time for ventilated, critically ill, or injured patients. The ability to continuously monitor and assess for effects of interventions facilitates intensivists by guiding complex resuscitations in
order to better tailor therapeutic maneuvers to patient needs and tolerance [ 53 ].
Image Acquisition and Interpretation
The hTEE probe is approximately 17 French in size, inserted transorally, and advanced into the stomach. The probe is slowly retrieved where fl exion, antefl exion, and gentle rota­tional torque of the probe facilitate image capture.
The views obtainable with hTEE include (Fig. 8.3 ):
• Transgastric view: a short-axis view allowing visualiza­tion of both cardiac ventricles, permitting measurement of LV size, calculation of LV fractional area contraction, and examination of septal size, shape, and wall motion abnormalities.
• Four-chamber view: a long-axis view providing visualiza­tion of the four cardiac chambers and measurement opportunity of LV and RV end-diastolic and end-systolic size and calculation of LV ejection fraction.
• Superior vena cava view: a long-axis view, at the level of the SVC/right atrial junction where examination of the variation in the SVC diameter during the respiratory cycle can be assessed.
Software-driven computations provided by the device
manufacturer are embedded in the operating system which allow for exam fi ndings such as ventricular chamber sized in diastole and systole, to be entered whereupon physiologic data such as CO, CI, EF, SVR, etc., can be generated. SVC diameter and variation can be manually measured and ana­lyzed. Alternatively, the skilled echocardiographer can visu­ally assess for chamber size, variation, and responsiveness to therapy in order to estimate hemodynamic status and resus­citation needs. The technique of image capture and data point entry for algorithmic analysis for hTEE is beyond the scope of this chapter but is readily available in a variety of established teaching modules and device manufacturer literature.
Considerations
The simple yet robust TEE images and cardiac performance patterns achieved with hTEE have previously been validated in ventilated patients with septic shock against a more com­prehensive and quantitative assessment [ 53 , 54 ]. The hTEE examination has been demonstrated to provide relevant infor­mation for the hemodynamic monitoring of ventilated ICU patients with cardiorespiratory compromise and has proven to be therapeutically impactful in several studies and continues to be applied to various, specifi c clinical scenarios for more accurate hemodynamic monitoring efforts [
5356 ].
Through the continuous evaluation of the superior vena
cava and the right and LV chambers and their response to therapeutic maneuvers, hTEE can identify the precise cause
8 Hemodynamic Monitoring in Surgical Critical Care
87
Fig. 8.3 Hemodynamic transesophageal echocardiography probe position and corresponding image (With permission of ImaCor, Inc.)