Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3608_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
83 Мб
Скачать
162
The actual arteriovenous connection may be too small to be seen in postcatheterization AVF.
Evaluation of dialysis fistulae use specific criteria for the Doppler spectra obtained from arterial inflow and venous outflow. systolic velocity is recorded throughout the native system and the graft. The arterial limb should demonstrate high velocities and con-
CH
tinuous forward flow with a low-resistance waveform. The venous
12
limb is expected to have slightly lower velocities. The normal PSV at the anastomosis is 300 cm/sec. The normal outflow vein has a PSV greater than 180 cm/sec and appears distended. Peak systolic velocity less than 150 cm/sec indicates a fistula in jeopardy of fail­ure. The fistula may result in arterial steal from the distal circula­tion. If this is suspected, direction of distal flow should be evaluated before and after compression of the AVF.
Venous Duplex Ultrasound
45,46
Peak
FIGURE 12-30 Transverse gray-scale imaging of internal jugular vein
(IJV) and common carotid artery (CCA) without compression (left) and with gentle compression (right) obliterating vein lumen.
47
Normal veins have thin walls and an echo-free lumen. The vein lumen can be obliterated (compressed) with a small amount of extrinsic pressure (
Fig. 12-29). The walls do not co-apt, however, when the
lumen contains thrombus, even when enough pressure is applied to distort the shape of an adjacent artery. Vein compressibility is best tested in an image plane transverse to the vein axis. Veins are characterized by anatomical location as deep or superficial, and as proximal or distal. The major veins of the thigh and arm are larger in diameter than the corresponding arteries. Extremity veins have valves that permit only cephalad flow, and these increase in num­ber from proximal to distal. Valve sinuses are widened areas of the lumen that accommodate the valve cusps.
Doppler evaluation of flow in normal veins has four important characteristics: (1) respirophasic variation, (2) augmentation with distal compression, (3) unidirectional flow toward the heart, (4) and abrogation of flow in the lower extremities by the Valsalva maneu­ver. Complete analysis of venous spectral waveforms requires comparison of the waveforms from both right and left limbs. Presence of a flattened, unvarying waveform (loss of respirophasic variation in flow) on one side compared with the other suggests the presence of more proximal obstruction of venous return proximal to the site of the Doppler interrogation.
Neck and Upper-Extremity Venous Duplex Ultrasound
Neck and upper-extremity duplex evaluation includes assessments of the internal jugular, subclavian, axillary, brachial, cephalic, and basilic veins. cava cannot be evaluated with duplex ultrasound because of their location within the bony thorax. Examination begins with evaluation of the internal jugular (Fig. 12-30) and subclavian veins. The subclavian vein can be imaged from a supraclavicular or
48
The innominate veins and the superior vena
subclavicular approach. The arm is extended in a comfortable posi­tion for the evaluation of the axillary vein, paired brachial veins, basilic vein (medial), and cephalic vein (lateral). Examination includes color and spectral Doppler evaluation of flow in all these veins. Loss of respirophasic variation in the waveform in the sub­clavian or axillary veins suggests the presence of more proximal venous obstruction (due to thrombosis or extrinsic compression) (
Fig. 12-31). The subclavian vein cannot be compressed where
Loss of compressibility is the pathognomonic feature of venous thrombosis. As the thrombus progresses from acute to chronic, there is increased echogenicity of the thrombus and decreased diameter of the vein. Over time, collateral veins may develop and recanalization may occur in the thrombosed vessel. In the upper arm, both superficial and deep venous systems have a significant role in venous drainage. The majority of upper-extremity DVTs are secondary to indwelling venous catheters, pacemaker leads, or hypercoagulability. Primary upper-extremity DVT is a rare disorder that is idiopathic, attributed to effort thrombosis (Paget-Schrötter's syndrome), or related to thoracic outlet obstruction.
49
An unusual etiology of noncompressible veins is intravascular tumor. This is suspected when the echogenic material within the lumen appears to extend through the vessel wall and may contain arterial flow signals.
Lower-Extremity Venous Duplex Ultrasound
The venous ultrasound examination to evaluate the presence or absence of leg DVT begins at the inguinal ligament with identi­fication of the common femoral vein and extends to the calf. The proximal deep veins evaluated include the common femo­ral, femoral (previously known as superficial femoral), and pop­liteal veins. The deep calf veins include posterior tibial, peroneal, gastrocnemius (sural), and soleal veins. Special attention is given
50
FIGURE 12-29 Left, Transverse gray-scale imaging of superficial femoral artery (SFA) (A) and vein (V). Right, With gentle compression, artery is unchanged and vein is obliterated.
RT AX V
FIGURE 12-31 Spectral Doppler evaluation of axillary vein (AX) demonstrating loss of phasic variation in flow with respiration. This finding
suggests more proximal venous obstruction. Increase in flow on right results from compression of forearm (augmenting venous return).
163
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
Gray scale
A
Spectral Doppler
LT SFV
Color Doppler
*
CH 12
VASCULAR LABORATORY TESTING
SFV
LT SFV
B
SFA
SFA
SFV
SFV
CD
FIGURE 12-32 Ultrasound evaluation of femoral vein thrombosis. A, Gray-scale imaging with echogenic material (asterisk) seen within lumen of superficial femoral vein (SFV). B, Doppler in SFV shows a flow void within lumen. C, Spectral Doppler demonstrates diminished flow with respirophasic variation. D, Superficial femoral artery (SFA) and SFV without (left) and with (right) compression; vein is only partially obliterated due to thrombus.
to the saphenofemoral and saphenopopliteal junctions because thrombus in the superficial veins of these regions deserve more aggressive treatment than thrombus limited to other parts of the superficial venous system. Examination includes color and spec­tral Doppler evaluation of flow in all these veins Loss of respirophasic variation in the waveform of the common femoral vein suggests presence of obstruction proximal to the site of Doppler interrogation that is preventing venous return. Augmentation of flow with calf compression is not prevented by proximal venous obstruction. Proximal obstruction may be caused by extrinsic compression or venous thrombosis.
Bright-mode transverse images are used to determine compress­ibility along the entire course of the veins examined. Normally the vein walls fully coapt with gentle pressure. Lack of compressibility, which occurs because of a thrombosis in the vein, is the most reli­able finding for determining venous thrombosis. With acute throm­bosis, there is low echogenicity of the intraluminal thrombus, and the vein is dilated. As the thrombus ages, it becomes more echo­genic and less central within the lumen. Vein diameter decreases as the thrombus retracts. Recanalization occurs, and flow can be detected by pulsed or color Doppler. The thrombus often appears eccentric and adjacent to the vein wall. When image quality is poor because of significant soft-tissue edema, it is not possible to exclude the presence of small nonocclusive thrombi. Sensitivity for detection of common femoral vein thrombosis is 91%, and for both the femoral and popliteal veins is 94%.
Duplex ultrasound is accurate for diagnosing deep calf vein thrombosis in symptomatic patients, so long as the calf veins
51,52
(Fig. 12-32) .
FIGURE 12-33 Transverse gray-scale imaging of peroneal (deep) veins of calf. Vein appears dilated in noncompression image on left. Vein lumen is not
obliterated by gentle compression, indicating presence of calf vein thrombosis. A, artery; V, vein.
can be seen clearly (
Fig. 12-33). When compared with angiog-
raphy, sensitivity of compression ultrasound for deep calf vein thrombosis is 94% and specificity is 100%. Small calf veins can­not be visualized well in all patients. However, specificity and positive predictive value are high even when individuals with poor calf vein images are included in the evaluation. Thus, the diagnosis of DVT is made when calf veins are seen and cannot be compressed.
Lower-extremity venous ultrasound testing is often ordered when
a patient is undergoing evaluation for pulmonary embolism (PE).
164
( ) ( )
This test will provide information about presence or absence of venous thrombosis, but will not determine whether or not some­one has had a PE.
Duplex Ultrasound Evaluation of Venous
CH
Insufficiency
12
Use of duplex ultrasound has been extended to detect reflux or obstruction and determine the anatomical extent of venous dis­ease in patients with chronic venous insufficiency. ment has been facilitated by color Doppler imaging to provide instant determination of the direction of blood flow. A 4- to 7-MHz linear array transducer is used. The saphenofemoral junction is examined first with the patient standing, and then in the supine reverse-Trendelenburg position. Compressibility is determined in transverse views of the veins. A longitudinal view of the sapheno­femoral junction is then obtained.
One of two maneuvers can be used to elicit reflux. The first is the Valsalva maneuver. Intraabdominal pressure increases as the patient bears down, and venous outflow from the legs decreases. Venous return from the legs increases with release of the maneuver.54 The second is thigh cuff inflation and deflation. Venous return is stopped with inflation of a cuff, typically to a level approximating arterial diastolic pressure. There is a transient increase in venous return that accompanies cuff deflation. Color flow is evaluated before and after one of these two maneuvers to elicit reflux. Baseline antegrade flow is displayed by blue color Doppler. Red color after the maneuver indicates retrograde flow. Reflux is present if red color persists for more than 0.5 seconds after either maneuver. Spectral Doppler can also be used to evalu­ate reflux. The Doppler cursor is placed midstream with an angle of 60 degrees with respect to the wall. Reverse flow over 0.5 sec­onds in duration is consistent with reflux ( remainder of the examination is performed with the patient stand­ing with the weight on the leg not being examined.
The extent of reflux can be determined by repeating this assessment throughout the deep and superficial veins of the leg. For evaluation of the small saphenous and popliteal veins, the patient sits on the edge of the examination table with his/ her foot resting on a stool. The probe is placed over the popliteal fossa. The gastrocnemius veins can be seen between the popli­teal (which is deep) and small saphenous (which is superficial) vein. Compressibility and reflux following a Valsalva maneuver are determined in these veins. The posterior tibial and peroneal veins are assessed for reflux using the posteromedial and anterolateral views.
53
This develop-
Fig. 12-34). Ideally, the
Perforating veins are vessels connecting superficial and deep veins. Incompetent perforating veins are identified by sliding the transducer up and down dilated superficial varicose veins.
55
Color Doppler is then used while distal compression of the superficial vein is performed. The presence of different colors during com­pression and release indicates that the direction of venous flow changes with compression and relief. This finding is diagnostic of reflux in the perforator veins.
Plethysmographic Evaluation of Venous Reflux
Duplex ultrasound identifies reflux in individual veins, and plethysmographic methods evaluate the volume of venous reflux in the limb.56 Air or strain gauge plethysmography is a sim­ple screening test that has the potential to provide a complete analysis of venous hemodynamics (also see Chapter 55) . The air chamber is filled with air to 6 mmHg and connected to a pressure transducer and recorder. Changes in the volume of the leg as a result of emptying or filling veins produce changes in the pressure of the air chamber. Recordings are made with the patient supine, and the leg elevated at a 45-degree angle. The patient then stands with the leg flexed slightly and bearing weight on the nonstudy leg. Venous filling time and venous volume are determined. The time until the volume plateaus after the raised limb is dropped is the venous filling time. Venous volumes of 80 to 150 mL are normal. The venous filling index (VFI) correlates best with the clinical severity of reflux.
VFI .90 venous volume .9 venous filling time÷=
Values less than 2 mL/sec indicate the absence of significant reflux, whereas values over 10 mL/sec indicate high risks of edema, skin changes, and ulceration.
57
Post-exercise plethysmography can be used to evaluate the eject­ing capacity of the calf muscle pump. Venous volume is measured at rest, and again post-exercise. Rest volume minus post-exercise volume equals ejected volume. Ejection fraction is the ejected vol­ume/rest volume × 100. Calf ejection fractions below 40% indicate patients most likely to benefit from deep vein reconstruction.
Vascular Laboratory Accreditation
Laboratory accreditation is obtained through organizations such as the Intersocietal Commission for the Accreditation of Vascular Laboratories (www.icavl.org) and the American College of Radiology (www.acr.org). The accreditation process reviews the educational credentials of the interpreting physicians and sonog­raphers, as well as laboratory procedures. It provides excellent stan­dards for setting up examination protocols and quality assurance programs.
RT LSV Back of prox calf
FIGURE 12-34 Spectral Doppler of venous insufficiency. Lesser saphenous vein is imaged at saphenopopliteal junction. Spectral Doppler demonstrates prolonged retrograde flow at this site.
REFERENCES
1. Boyajian RA, Otis SM: Integration and added value of the modern noninvasive vascular
laboratory in vascular diseases management, J Neuroimaging 12:148, 2002.
2. Kupinski AM: Segmental pressure measurement and plethysmography, J Vasc Technol
26:32–38, 2002.
3. Rooke TW, Hirsch AT, Misra S et al: 2011 ACCF/AHA Focused Update of the Guideline for the
Management of patients with peripheral artery disease (Updating the 2005 Guideline): a report of the American College of Cardiology Foundation/American Heart Association Task Force on practice guidelines. Circulation 124:2020–2045, 2011.
4. Aboyans V, Criqui M, Abraham P, et al: The measurement and interpretation of the ankle
brachial index, Circulation In press.
5. Carter SA, Tate RB: The value of toe pulse waves in determination of risks for limb
amputation and death in patients with peripheral arterial disease and skin ulcers or gangrene, J Vasc Surg 33:708–714, 2001.
6. Stein R, Hrilajac I, Halperin J, et al: Limitation of resting ankle brachial index in symptomatic
patients with peripheral artery disease, Vasc Med 11:29–33, 2006.
7. Anderson CA: Noninvasive assessment of lower extremity hemodynamics in individuals
with diabetes mellitus, J Vasc Surg 128:77S, 2010.
8. Feldman MK, Katyal S, Blackwood MS: US artifacts, Radiographics 4:1179–1189, 2009.
9. Grenier N, Basseau F, Rey MC, et al: Interpretation of Doppler signals, Eur Radiol 8:1295–1307,
2001.
10. Logason K, Barlin T, Jonsson M, et al: The importance of Doppler angle of insonation on
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
differentiation between 50-69% and 70-99% carotid artery stenosis, Eur J Endovasc Surg 21:311, 2001.
11. Wood MM, Romine LE, Richman KM, et al: Spectral Doppler signature waveforms in ultrasonography: a review of normal and abnormal waveforms, Ultrasound Q 26:83–99, 2010.
12. Pellerito JS: Current approach to peripheral arterial sonography, Radiol Clin North Am 39:553–567, 2001.
13. Needham TN: Review of pressure and flow in the arterial system, 26:27, 2002.
14. Phillips GW: Review of vascular ultrasound, World J Surg 24:232–248, 2000.
15. Rubin JM: Power Doppler, Eur J Radiol 9:318S–322S, 1999.
16. Bokor D: Diagnostic efficacy of SonoVue, Am J Cardiol 86:19G–24G, 2000.
17. Brott TG, Halperin JL, Abbara S, et al: 2011 ASA/ACCF/AHA/AANN/AANS /ACR/ASNR/CNS/ SAIP/SCAI/SIR/SNIS/SVM/SVS guideline on the management of patients with extracranial carotid and vertebral artery disease: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines, and the American Stroke Association, American Association of Neuroscience Nurses, American Association of Neurological Surgeons, American College of Radiology, American Society of Neuroradiology, Congress of Neurological Surgeons, Society of Atherosclerosis Imaging and Prevention, Society for Cardiovascular Angiography and Interventions, Society of Interventional Radiology, Society of NeuroInterventional Surgery, Society for Vascular Medicine, and Society for Vascular Surgery, Stroke 42:e420–e463, 2011.
18. Gerhard-Herman M, Gardin JM, Jaff M, et al: Guidelines for noninvasive vascular laboratory testing. American Society of Echocardiography; Society for Vascular Medicine and Biology, Vasc Med 11:183–200, 2006.
19. Grant EG, Benson CB, Moneta GL, et al: Carotid artery stenosis: gray-scale and Doppler US diagnosis—Society of Radiologists in Ultrasound Consensus Conference, Radiology 229:340, 2003.
20. Ochoa VM, Yeghezarians Y: Subclavian artery stenosis: a review for the vascular medicine practitioner, Vasc Med 16:29–34, 2011.
21. Staub D, Schinkel AF, Coll B, et al: Contrast-enhanced ultrasound imaging of the vasa vasorum: from early atherosclerosis to the identification of unstable plaques, J Am Coll Cardiol Cardiovasc Imaging 3:761–771, 2010.
22. Polak JF, Pencina MJ, Pencina KM, et al: Carotid-wall intima-media thickness and cardiovascular events, N Engl J Med 365:213–221, 2011.
23. Hirsch AT, Haskal ZJ, Hertzer NR, et al: ACC/AHA 2005 Practice Guidelines for the management of patients with peripheral arterial disease (lower extremity, renal, mesenteric, and abdominal aortic): a collaborative report from the American Association for Vascular Surgery/Society for Vascular Surgery, Society for Cardiovascular Angiography and Interventions, Society for Vascular Medicine and Biology, Society of Interventional Radiology, and the ACC/AHA Task Force on Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Patients With Peripheral Arterial Disease): endorsed by the American Association of Cardiovascular and Pulmonary Rehabilitation; National Heart, Lung, and Blood Institute; Society for Vascular Nursing; TransAtlantic Inter­Society Consensus; and Vascular Disease Foundation, Circulation 113:e463–e654, 2006.
24. Beales L, Wolstenholme S, Evans JA, et al: Reproducibility of ultrasound measurement of the abdominal aorta, Br J Surg 98:1517–1525, 2011.
25. Hong H, Yang Y, Liu B, et al: Imaging of the abdominal aor tic aneurysm: the present and the future, Curr Vasc Pharmacol 8:808–819, 2010.
26. Cao P, De Rango P, Verzini F, et al: Endoleak after endovascular aortic repair: classification, diagnosis and management, J Cardiovasc Surg (Torino) 51:53–69, 2010.
27. Hedayati N, Del Pizzo DJ, Harris SE, et al: Predictors of diagnostic success with renal artery duplex sonography, Ann Vasc Surg 25:515–519, 2011.
28. Fleming SH, Davis RP, Cravens TE, et al: Accuracy of duplex sonography scans after renal artery stenting, J Vasc Surg 52:953–957, 2010.
29. Ciccone MM, Cortese F, Fiorella S, et al: The clinical role of contrast enhanced ultrasound in the evaluation of renal artery stenosis and diagnosis compared to traditional echo color Doppler flow imaging, Int Angiol 30:135–139, 2011.
30. Radermacher J, Chavan A, Bleck J, et al: Use of Doppler ultrasonography to predict the outcome of therapy for renal-artery stenosis, N Engl J Med 344:410, 2001.
31. Gao J, Rubin JM, Xiang DY, et al: Doppler parameters in renal transplant dysfunction, J Ultrasound Med 30:169–175, 2011.
32. Collins R, Cranny G, Burch J, et al: A systematic review of duplex ultrasound, magnetic resonance angiography and computed tomography angiography for the diagnosis and assessment of symptomatic, lower limb peripheral arterial disease, Health Technol Assess 11:1–184, 2007.
33. Begelman S, Jaff M: Noninvasive diagnostic strategies for peripheral arterial disease, Cleve Clin J Med 73:22–29, 2006.
34. Roth SM, Bandyk DF: Duplex imaging of lower extremity bypasses, angioplasties and stents, Semin Vasc Surg 12:275–284, 1999.
35. Shrikhande GV, Graham AR, Aparjita R, et al: Determining criteria for predicting stenosis with ultrasound duplex after endovascular intervention in infrainguinal lesions, Ann Vasc Surg 25:454–460, 2011.
36. Slim H, Tiwari A, Ritter JC, et al: Outcome of infrainguinal bypass grafts using vein conduit less than 3 mm in diameter in critical limb ischemia, J Vasc Surg 53:421–425, 2011.
37. Tinder CN, Bandyk DF: Detection of imminent graft occlusion: what is the optimal surveillance program? Semin Vasc Surg 22:252–260, 2009.
38. Tinder CN, Chavapun JP, Bandyk DF, et al: Efficacy of duplex ultrasound surveillance after infrainguinal vein bypass may be enhanced by identification of characteristics predicting stenosis development, J Vasc Surg 43:613–618, 2008.
39. Kapoor BS, Haddad HL, Saddekni S, et al: Diagnosis and treatment of pseudoaneurysms, Curr Probl Diagn Radiol 38:170–188, 2009.
40. Kang SS, Labropoulos N, Mansour MA, et al: Expanded indications for ultrasound-guided thrombin injection of pseudoaneurysms, J Vasc Surg 31:289, 2000.
41. Mohler ER III, Mitchell ME, Carpenter JP, et al: Therapeutic thrombin injection of pseudoaneurysms: a multicenter experience, Vasc Med 6:241, 2001.
42. La Perna L, Olin JW, Goines D, et al: Ultrasound-guided thrombin injection for the treatment of postcatheterization pseudo-aneurysms, Circulation 102:2391, 2000.
43. Seay T, Soares G, Dawson D: Postcatheterization arteriovenous fistula: CT, ultrasound and arteriographic findings, Emerg Radiol 9:269–299, 2002.
44. Rose SC: Noninvasive vascular laboratory for evaluation of peripheral arterial occlusive disease. Part III—clinical applications: nonatherosclerotic lower extremity arterial conditions and upper extremity arterial disease, J Vasc Interv Radiol 12:11, 2001.
45. Polkinghorne KR, McMahon LP, Becker GJ: Pharmacokinetic studies of dalteparin (Fragmin), enoxaparin (Clexane), and danaparoid sodium (Orgaran) in stable chronic hemodialysis patients, Am J Kidney Dis 40:990, 2002.
46. Gonzalez SB, Busquets JC, Figueriras RG, et al: Imaging arteriovenous fistulas, AJR Am J Roentgenol 31:1425–1433, 2009.
47. Tan M, van Rooden CJ, Westerbeek RE, et al: Diagnostic management of clinically suspected acute deep vein thrombosis, Br J Haematol 146:347–360, 2009.
48. Kucher N: Clinical practice: deep vein thrombosis of the upper extremities, N Engl J Med 3:861–869, 2011.
49. Joffe HV, Goldhaber SZ: Upper-extremity deep vein thrombosis, Circulation 106:1874, 2002.
50. Bounameaux H, Perrier A, Righini M: Diagnosis of venous thromboembolism: an update, Vasc Med 15:399–406, 2010.
51. Vogel P, Laing FC, Jeffrey RB Jr, et al: Deep venous thrombosis of the lower extremity: US evaluation, Radiology 163:747, 1987.
52. Elliot CG, Lovelace TD, Brown LM, et al: Diagnosis: imaging techniques, Clin Chest Med 4:641–657, 2010.
53. K alodiki E, Nicolaides AN: Out of a recent CVI consensus: some features of a basic statement, Int Angiol 21:2, 2002.
54. Min RJ, Khilnani NM, Golia P: Duplex ultrasound evaluation of lower extremity venous insufficiency, J Vasc Interv Radiol 14:1233–1241, 2000.
55. Stuart WP, Adam DJ, Allan PL, et al: The relationship between the number, competence, and diameter of medial calf perforating veins and the clinical status in healthy subjects and patients with lower-limb venous disease, J Vasc Surg 32:138–143, 2000.
56. Miyazaki K, Nishibe T, Kudo F, et al: Hemodynamic changes in stripping operation or saphenofemoral ligation of the greater saphenous vein for primary varicose veins, Ann Vasc Surg 18:465–469, 2004.
57. Nicolaides AN: Investigation of chronic venous insufficiency: a consensus statement. Cardiovascular Disease Educational and Research Trust; European Society of Vascular Surgery; The International Angiology Scientific Activity Congress Organization; International Union of Angiology; Union Internationale de Phlebologie at the Abbaye des Vaux de Cernay, Circulation 102:E126–E163, 2000.
165
CH 12
VASCULAR LABORATORY TESTING
CHAPTER
13 Magnetic Resonance Imaging
Cihan Duran, Piotr S. Sobieszczyk, Frank J. Rybicki
Magnetic resonance angiography (MRA) is widely accepted for the majority of vascular imaging applications and is the modality of choice for many of them, particularly peripheral and renal arterio-
1–3
graphy. high-quality spatial resolution images with relatively short acquisi­tion times. Detailed comparison with other imaging technologies is beyond the scope of this chapter, but general advantages and dis­advantages appear in applications such as aortography, where multiple modalities (e.g., computed tomography [CT] and MRA) are routinely diagnostic and capable of providing images for planning most interventions.
Gadolinium-based contrast-enhanced (CE) MRA yields
Table 13-1. These become important for many
Basic Principles
frequency of the wobble is proportional to the strength of B radiofrequency (RF) pulse is applied at the resonance frequency of the wobble, protons can absorb energy and jump to a higher energy state. This RF pulse deflects the protons, creating a new net magnetization vector distinct from the major axis of the applied magnetic field. The net magnetization vector tips from the longitu­dinal to the transverse plane (transverse magnetization). The pro­tons are “flipped” by the RF pulse, and the net magnetization vector is defined by a “flip angle.” The stronger the RF pulse applied, the greater the angle of deflection for the magnetization. Common flip angles for spin echo are 90° and 180°. For gradient echo (GRE) MRI, flip angles typically range between 10° and 70°. After the RF pulse tips the spinning protons out of alignment with the main magnetic field, new protons begin to align with the main magnetic field at a rate determined by the T1 relaxation time.
states. The absorbed RF energy is retransmitted at the resonance frequency and can be detected with RF antennas or “coils” placed around the patient. These signals are compiled, and after mathe­matical processes become the MR images. Proton excitation with an externally applied RF field is repeated at short intervals to obtain signals. This MR parameter is referred to as repetition time (TR). For conventional MRI, TR is typically 0.5 to 2 seconds, whereas for MRA, TR ranges from 30 to less than 5 milliseconds. When the spins are tipped to the transverse plane, they all precess in phase. The speed of wobbling depends on the strength of the magnetic field each proton experiences. Some protons spin faster while others spin slower, and they quickly get out of phase relative to one another. Throughout the dephasing process, the MR signal decays. This loss of phase is termed T2 relaxation time or transverse relaxation. T2, like T1, is unique among tissues and is used for image contrast. In addition to the intrinsic T2 of tissue, inhomogeneity of B time that reflects the sum of these random defects with tissue T2 is
) to create a net magnetization vector. On a quantum level,
0
. However, a slight excess of spins aligns with the field,
0
Spinning protons wobble or “precess” about the axis of B
Energy is given off as the spins move from high to low energy
results in rapid loss of transverse magnetization. The relaxation
0
. The
0
. If a
0
called T2*. To obtain an MRI signal, these spins must be brought back in phase and produce a signal or echo. The time at which it happens is referred to as echo time (TE). In spin echo imaging tech­nique, the echo is obtained by using a refocusing 180° RF pulse, after which the spins begin to dephase. Another 180° RF pulse can be applied to generate a second echo and so on. Signal loss at longer echo times reflects tissue T2. In GRE imaging, the echo is obtained by gradient reversal rather than RF pulse. Because this includes effects from tissue homogeneity, TE-dependent signal loss reflects T2*. Recently, GRE sequences (balanced GRE steady-state free precession [SSFP]) have been developed that are insensitive to magnet field inhomogeneities and reflective of actual tissue T2.
Longitudinal and transverse magnetizations occur simultane­ously but are two different processes that reflect properties of vari­ous tissues in the body. Since T1 measures signal recovery, tissues with short T1 are bright, whereas tissues with long T1 are dark. Fat has a very short T1. In contrast, T2 is a measure of signal loss. Therefore, tissues with short T2 are dark, and those with long T2 are bright. Simple fluids, such as cerebrospinal fluid and urine, have long T2. To differentiate between the tissues based on these relaxation times, MR images can be designed to be T1-weighted, T2-weighted, or proton-density weighted. Exogenous contrast such as gadolinium-based agents are routinely used to alter tis­sue conspicuity. Spatial encoding of signals obtained from tissues is required for imaging. Additional external time-varying mag­netic fields are applied to spatially encode the MR signal. Spatially dependent gradients are used to locate the MR signal in space. In two-dimensional (2D) MRI, these are slice-selection, frequency­encoding, and phase-encoding gradients. In three-dimensional (3D) MRI, the slice-selection gradient is replaced by a second phase-encoding gradient.
Magnetic resonance echoes are digitized and stored in “k-space” composed of either two axes (for 2D imaging) or three axes (for 3D imaging). K-space represents frequency data and is related to image space by Fourier transformation. An important feature of k-space is that tissue contrast is determined by the center of k-space (cen­tral phase encoding lines), whereas the periphery of the k-space encodes the image detail. The order in which k-space lines are collected can be varied, strongly influencing tissue contrast. For example, in CE-MRA, the central contrast-defining portion of k-space may be acquired early in the scan (centric acquisition) during peak intraarterial contrast concentration to maximize arterial contrast. In addition to simple line-by-line k-space acquisition schemes, more complex schemes have been described. In spiral imaging, data acquisition begins at the center of k-space and spirals to the periph­ery. Slice-selective gradients applied along the z-axis will form axial images. Those along the y-axis will yield coronal images, and the x-axis gradients will provide sagittal images. An oblique slice can be selected by a combination of two or more gradients.
Magnetic Resonance Angiography Techniques
Magnetic resonance imaging relies on selective imaging of moving blood where signals from blood vessels are maximized, whereas signals from the stationary tissues are suppressed. Algorithms then enable reformatted images similar to those found in conventional x-ray angiography (Table 13-2).
Magnetic resonance angiography methods can depict blood as either black or white. For those “black-blood” methods that use standard spin echo (SE) sequences (Fig. 13-1A), the excitation RF pulse is applied at 90° and followed by a refocusing pulse at 180°. If the imaging slice cuts across a vessel, then depending on the flow
166
TABLE 13-1 Advantages and Disadvantages of Vascular Imaging Methods
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
METHOD ADVANTAGES DISADVANTAGES
MR No ionizing radiation Expensive hardware
CE High signal from gadolinium-based agents
Non-contrast Eliminates toxicity concern Longer acquisition times compared to CE protocol
CT Rapid, high signal, high-quality image acquisition
DSA Intervention can be performed at time of diagnosis
Sonography
CE, contrast-enhanced; CNR, contrast-to-noise ratio; CT, computed tomography; DSA, digital subtraction angiography; MRA, magnetic resonance angiography; MRI, magnetic resonance imaging; SNR, signal-to-noise ratio.
Less nephrotoxic than comparable iodine-based imaging (CT)
Less technical expertise required compared to MR
Highest spatial resolution
No ionizing radiation Less expensive Flow information readily obtained Portable
Acquisition requires technical expertise
Nephrogenic systemic fibrosis can occur rarely in patients with
severe renal insufficiency
Increased risk of artifacts that are largely mitigated with CE-MRA
Ionizing radiation Nephrotoxicity of iodine
Invasive Nephrotoxicity of iodine Projectional data may be inferior to volumetric acquisitions (CT, MR)
that can be viewed in any plane
Better for superficial imaging Limited by artifacts from bone, air, and sonographic interfaces Lower CNR, SNR compared to CT and MRI Operator-dependent
TABLE 13-2 Types of Magnetic Resonance Angiography Sequences
NAME OF
SEQUENCE
TOF Bright vessels produced by inflow of
PC Bright vessels produced by applica-
Dynamic CE Bright vessels produced by rapid infusion
Postcontrast Bright vessels produced by equilibrium
Black blood
2D, two-dimensional; 3D, three-dimensional; CE, contrast-enhanced; PC, phase-contrast; RF, radiofrequency; TOF, time-of-flight.
blood with full magnetization into a slice or volume where magnetization has been reduced by RF saturation
tion of flow encoding radiofrequency pulses that produce a phase image where intensity is proportional to velocity
of gadolinium contrast, with timing of scan to arterial and/or venous transit; mask subtraction may be used
enhancement of gadolinium contrast with fat saturation
Dark vessels produced by use of an
inversion prepulse to null the signal of blood, based on its T1 recovery time
DESCRIPTION
2D OR 3D GATING USEFUL
Both Occasionally No Intracranial, carotid, pedal
Both Occasionally Yes None in current practice
3D No No Carotid, chest, abdomen,
Both No No Veins, aorta
2D
Necessary
FLOW QUANTIFICATION
POSSIBLE
No
APPLICATIONS
extremity runoff
Plaque imaging
167
CH 13
MAgnETiC REsonAnCE iMAging
velocity and time interval between the pulses, the blood volume originally excited by the first pulse may not “see” the second pulse. This results in a black-appearing signal void in the vessel lumen. The use of thin sections or long echo times can further emphasize this flow void. This technique allows detailed examination of arte­rial wall morphology. Fast spin echo (FSE) sequences, in which a long train of echoes is obtained by use of repeated 180° pulses, pro­duce images more rapidly. The double inversion recovery (DIR) FSE offers a new approach to enhance black-blood sequences. The technique uses two consecutive inversion pulses: the first nulls or blackens the blood everywhere in the coil, and the second restores magnetization in the slice being imaged. Between these pulses and image production, blood within the slice is replaced by nulled blood from outside. This produces more reliable black blood than conventional approaches, making this sequence ideal for exam­ining wall thickness, dissection flaps, and the presence of mural thrombus or inflammation.
4
This provides a clear advantage over
traditional x-ray angiography.
“Bright blood” MRA techniques use GRE sequences and are gen­erally divided into those measuring signal amplitude (time-of-flight [TOF]) and those based on phase effects (phase-contrast [PC]). In each GRE sequence, a single RF pulse is applied in short time inter­vals, eliminating signal loss due to flow void. The stationary protons occupying a given tissue slice do not have sufficient time to relax to their equilibrium state.
TOF-MRA techniques depend on the inflow of unsaturated protons in blood from outside the field of view (FOV) into the stationary tissue within a section already saturated by its expo­sure to repeated RF pulses. These “saturated” protons are unable to contribute signal to the image. The signals in the stationary tis­sues of GRE images used in MRA are therefore typically low. The “unsaturated” protons in blood flowing into the imaging plane have not experienced the RF pulses and yield maximum signal. The unsaturated blood appears bright compared to background tissue (
Fig. 13-1B). The time required for blood to flow through an
image slice and its effect on the resulting signal is known as TOF.
168
CH
13
A
B
FIGURE 131 “Black-blood” and “bright-blood” imaging techniques. A, Cross-sectional T1-weighted image of ascending and descending aorta; lumen appears black. B, Time-of-flight (TOF) image of carotid-vertebral system. C, Maximum-intensity projection (MIP) reconstruction of arterial phase of contrast-enhanced magnetic resonance angiography (CE-MRA) images in same patient shows normal carotid-vertebral arteries at higher spatial resolution than corresponding TOF acquisition.
Saturation of signals can occur in vessels with slow-moving blood as a result of repeated RF excitation in the acquisition plane. This can create artifacts in vessels with stenotic lesions or reduced blood flow.
Time-of-flight techniques can be obtained in 2D or 3D. The 2D TOF utilizes multiple sequentially acquired, overlapping thin slices to form an image. The patient is instructed to hold his/her breath to minimize motion artifact. However, spatial misregistration may occur if patients cannot hold their breath at the same level each time. Thus, only one or two slices are typically acquired per breath­hold. 2D TOF has good sensitivity for identifying vessels with slow flow because blood must move only 3 to 5 mm to refresh a slice. 3D TOF consists of GRE acquisition of a volume into which blood is flowing. The advantage of this technique is higher signal-to-noise ratio (SNR) and improved resolution. The thick volumes of tissue imaged require rapid flow to fully refresh signals within the arter­ies. The technique is flow dependent and superior in vessels with rapid steady flow without respiratory motion. Additional saturation pulses can be applied to eliminate signal from veins. Segmented GRE sequence with cardiac triggering can be used to eliminate arterial pulsation artifacts.
A successful TOF image requires the section to be thin enough to allow for sufficient inflow between RF pulse repetitions, but thick enough to ensure adequate SNR and anatomical coverage. Section thickness of 3 to 4 mm is used for large vessels, and 1 to 2 mm for smaller vessels. Spatial presaturation pulses are applied above or below the imaged slice or volume to eliminate unwanted signal from arteries or veins, depending on which part of the vas­cular tree is being imaged. Optimal TR for TOF is 20 to 50 millisec­onds. Short TR keeps background tissues saturated, but it must be
C
long enough to allow for satisfactory inflow of unsaturated blood between successive repetitions. The best flip angle is usually 30° to 60°. With phasic flow in the extremities, systolic flow signal may be increased (because of greater transverse magnetization created), and distal flow may be decreased, creating view-to-view intensity changes and phase artifacts from pulsatile variations. This pulsa­tion artifact is greatest at higher flip angles. Cardiac gating can be used to minimize these artifacts at the expense of increased imag­ing time.
Whereas TOF uses differences in signal amplitude to differ­entiate between stationary and flowing spins, the PC technique observes the phase shifts of signals. Moving spins experience dif­ferent phase shifts in the presence of the applied magnetic fields used in MRA. Strength and orientation of the applied magnetic field are varied to encode different phase shifts for flowing pro­tons relative to stationary protons. The faster the spins are moving, the greater their phase shift, and protons of flowing blood may be discriminated from stationary protons. The phase shifts result in a contrast between moving and stationary tissues and form the basis for PC imaging. Pairs of images are acquired that have different sensitivities to flow and are then subtracted to cancel background signal, leaving only the signal from flowing blood. Phase shift is proportional to velocity, allowing flow quantification with this modality. Phase-contrast acquisitions may be acquired in two or three dimensions; although used rarely in angiography today, Phase-contrast offers a reliable way to quantify amount and direction of flow. It requires long imaging times: two data sets in each direction are acquired by using flow-encoding gradients of opposite polarity, and up to three measurements in the orthogonal planes are needed to image flow in all directions.
Visualization of the arterial system with PC and TOF is adequate5
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
but has limitations. Acquisition times can be long and prevent imaging within the time span of a single breath-hold. This increases the chance of movement artifacts. Some of the limitations are caused by flow-related artifacts such as in-plane saturation and phase dispersion. Flow-based imaging also has limits in areas of slow flow, such as aneurysms. Overgrading of stenotic lesions is most commonly a manifestation of signal loss in the areas of complex flow. Undergrading is a matter of inadequate spatial reso­lution. Complex turbulent flow patterns in areas of stenoses can create signal loss and mimic a critical lesion. This is due to “intra­voxel dephasing.” An accelerated flow across a stenosis consists of a wide distribution of velocities and thus a large distribution of pro­ton phases. In the smallest volume element, a “voxel” of the image, this distribution of phases can result in cancellation rather than coherent addition of signals, accounting for the presence of signal voids at the site of stenosis. A short TE minimizes flow-phase dis­persion artifacts. Phase dispersion is further decreased when voxel size is minimized using thin sections. Small voxels and short TE are most easily obtained with 3D TOF methods. The biggest drawback of the thick volumes used with 3D techniques is that slow or recir­culating flow can become saturated. The MOTSA (multiple overlap­ping thin-slab acquisitions) technique of sequential 3D TOF gives better flow enhancement than single-slab 3D TOF techniques and less dephasing than 2D techniques. However, the need for substan­tial overlap of adjacent slabs increases acquisition time.
Contrast-Enhanced Magnetic Resonance Angiography
The introduction of CE-MRA has revolutionized MRA. nique overcomes many of the limitations of traditional bright blood modalities: respiratory motion artifacts, poor SNR, and flow and saturation-related artifacts (Fig. 13-1C). Gadolinium increases the signal intensity of blood on contrast-enhanced 3D T1-weighted (spoiled) GRE images. Blood contrast is not flow dependent. It is determined by the concentration of contrast agent within the arterial system while imaging data are being col­lected. Reliable images can be acquired irrespective of whether flow is laminar, turbulent, or stagnant. This technique acquires large-volume data sets in coronal or sagittal orientation within a single breath-hold during the first pass of the contrast material. The contrast agent, gadolinium, is a heavy metal but becomes inert when bound to a chelator. Intravenous (IV) administration of gadolinium-diethylenetriamine pentaacetic acid (DTPA) results in a marked reduction of the T1 or longitudinal relaxation time of blood, therefore reducing the effects of spin saturation. Signal reduction is also problematic in 3D TOF sequences. Moreover, the very short TE reduces spin dephasing and allows accurate evaluation of vascular stenoses.
Multiple refinements have resulted in a technique that is much faster than TOF-MRA. The development of high-performance gradient systems with ultra-short repetition TR and TE has short­ened acquisition time in CE-MRA to allow imaging within a single breath-hold and minimize motion artifacts. Administration of agents shortening T1 allows selective visualization of contrast- containing structures and better visualization of circuitous collaterals. Digital subtraction, spoiling, and fat saturation techniques suppress back­ground signal and enhance signal from the contrast agent in the vessels. The subtracted data sets can be postprocessed to provide 3D projectional images. CE-MRA still provides a luminogram, and conventional or FSE images are needed for a complete study so that true lumen diameter and presence of thrombus can be established.
Optimal images are generated when gadolinium concentration is highest in the vessel of interest. To make blood bright compared to background tissues, the gadolinium bolus must be adminis­tered in a way that ensures the majority of the contrast to be pres­ent in the arterial tree. This requires exact timing of the arrival of the gadolinium bolus. Acquisition prior to contrast arrival creates
6,7
This tech-
a “ringing” artifact, whereas late acquisition creates venous and tissue enhancement, contaminating the arterial signal. This is espe­cially problematic in MRA of the extremities, where the images are obtained in multiple segments. Contrast transit time can be affected by low cardiac output, valvular regurgitation, large abdom­inal aneurysms, and flow-limiting stenoses. Proper timing can be achieved by empirical estimation of transit time or a test bolus in the anatomical field of interest. Alternatively, with automated trig­gering, a pulse sequence can be designed to sense the arrival of contrast and automatically trigger image acquisition. Magnetic resonance fluoroscopy allows the user to visualize arrival of the contrast bolus directly on the image and manually trigger the start of the scan. Areas that require higher spatial resolution, such as the lower extremities, also need larger doses of contrast for longer acquisition times.
Imaging during the arterial phase of gadolinium infusion takes advantage of higher arterial SNR and eliminates overlapping venous enhancement. This is a brief moment in time, but sev­eral methods allow slower MR image acquisition to capture that moment. The phase reordering (mapping k-space) technique acquires central k-space data (i.e., the low spatial frequency data) when contrast concentration is high in arteries but lower in veins. This allows a relatively long MR acquisition to achieve the image contrast associated with the shorter arterial phase of the contrast bolus. It is critical to time the contrast bolus to achieve maximum arterial gadolinium concentration during acquisition of central k-space data.
CE-MRA is limited by venous and soft-tissue enhancement. Contrast media not only passes into venous structures, depen­dent on the arteriovenous transit time of the tissue, but also rap­idly leaks out of the vascular compartment, creating significant tissue enhancement. New “blood pool” agents, which are currently undergoing clinical trials, are retained within blood vessels and selectively enhance the blood pool on T1-weighted MR images. These use either gadolinium compounds that bind to albumin, or are large enough to stay within the vascular space or ultra-small iron particle. Another agent, gadobenate, has a higher T1 relaxation time because of its capacity for weak and transient interaction with serum albumin. This may enhance vascular signal intensity and thus increase diagnostic efficacy at doses comparable to those used for current gadolinium agents. It is approved for imaging use in Europe but is under clinical investigation in the United States. It provides a higher and longer-lasting vascular signal enhancement in the abdominal aorta compared with gadolinium, which does not interact with proteins.
Gadolinium-based contrast has a very favorable safety pro­file. However, gadolinium is nephrotoxic. For patients with under­lying chronic renal insufficiency, gadolinium chelates can cause acute renal failure. Nephrogenic systemic fibrosis (NSF) is linked to gadolinium-based contrast agents9 and largely involves the skin, though it may also affect the muscle, joints, or internal organs such as the lungs, liver, and heart in patients with renal failure. Nephrogenic systemic fibrosis occurs in patients with severe renal disease who are exposed to high doses of gadolinium agents, or in patients who receive multiple standard doses of contrast agents in a short period of time. The reported prevalence of NSF among patients with glomerular filtration rate (GFR) less than 30 mL/min is 3% to 5%.10 Therefore, the MR protocol should aim to minimize contrast volume, especially in patients with moderate to severe renal failure.
Metal objects, such as surgical clips, lead to susceptibility artifacts in MRA. The increasing use of stents has important implications for MRA. Cavagna et al. evaluated CE-MRA of seven stent types in the aortic, iliac, and popliteal positions.12 Few of the commonly used stents permitted visualization of the lumen. Susceptibility artifact results in significant signal loss that can preclude proper visualization of the stent lumen even with gadolinium-enhanced MRA. Some nitinol, tantalum, or polytetra­fluoroethylene (PTFE)-based devices, on the other hand, cause less artifact on CE-MRA.
11
8
169
CH 13
MAgnETiC REsonAnCE iMAging
170
Postprocessing Techniques
Magnetic resonance data can be viewed as source images or be displayed in projections with any orientation. Image postprocess­ing allows reformation in any desired plane to improve conspi­cuity of overlapping vessels (
CH
13
ple, can overlap in coronal projections, whereas the origins of the right subclavian artery and right CCA can overlap in some oblique views. The renal ostia are usually best seen in either coronal or slightly oblique view. The celiac axis and superior mesenteric artery (SMA) are best depicted on sagittal projections. One advan­tage of MR versus digital subtraction angiography (DSA) is that the latter may require multiple injections to assess the origins of these vessels. The details of image interpretation are beyond the scope of this review, but source image data, multiplanar reconstruction (MPR), maximum-intensity projection (MIP), and volume render­ing (VR) are used (
Fig. 13-2). Source images are the initial recon-
structions and should be used for problem solving and to confirm
Table 13-3). The origins of the left
TABLE 13-3 Types of Postprocessing Techniques
TECHNIQUE DESCRIPTION
MPR Production of cross-sectional images in planes
MIP projection Production of full- or partial-volume images
Volume rendering
MIP, maximum intensity projection; MPR, multiplanar reconstruction; MRI, magnetic resonance imaging.
different from acquisition plane
along any desired axis from a stack of image slices
Manipulation of MRI slices to produce full
volumetric images; structures segmented for viewing by application of intensity thresholds and removal of unwanted structures
findings. Interpretation often begins with a vascular survey using MPR and MIP data sets. Multiplanar reconstructions are very use­ful in volumetric acquisitions because the desired imaging plane
B
A
C
FIGURE 132 Postprocessing techniques; contrast-enhanced magnetic resonance angiography (CE-MRA) of abdominal aorta and branches. A, Coronal thin-section image of abdominal aorta. Summation of these images, projected with maximum intensity, is used for (B) coronal maximum-intensity projection (MIP) that includes normal renal and mesenteric arteries. C, Axial multiplanar reconstruction (MPR) image at level of left renal ostium does not include entire extent of both renal arteries, giving false impression of a proximal right renal artery occlusion (arrow). D, Three-dimensional volume-rendered (3D-VR) image shows entire course of abdominal aorta and its branches.
D
can be prescribed to enhance vascular separations. Subtracted
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
MIPs are routinely created from CE-MRA. The noncontrast (mask) images are subtracted from the enhanced images, and resulting high SNR data sets undergo projection of maximum intensity. By performing the projections of all angles around the z
-axis of the patient, the data sets can be viewed in cine. These projections are referred to as rotating MIPs.
Clinical Applications
Extracranial Carotid and Vertebral Arteries
Atherosclerosis, dissection, and inflammatory diseases affect the extracranial carotid and vertebral arteries. Time-of-flight meth­ods (2D and 3D) have been largely replaced by CE-MRA because CE-MRA allows imaging of the entire course of these vessels (Fig. 13-3). A problem with TOF imaging is turbulent flow at or near the carotid bifurcation where most of the lesions occur; this may lead to overestimation of lesion severity. Gadolinium-enhanced MRA has greater SNR than noncontrast imaging, is less sensitive to intravoxel dephasing from turbulence, and does not have signal loss from saturation effects. Phase-contrast sequences can supple­ment anatomical data for flow direction or quantification.
Transcranial flow has a rapid arteriovenous transit time, and thus venous contamination can limit image quality; therefore, time­resolved imaging may be required. Assessing the test characteristics of carotid MRA is challenging because technologies evolve, and the patients and methods are both heterogeneous. In a 41-study meta-analysis, carotid CE-MRA had high sensitivity (94%) and high specificity (93%) for diagnosis of severe (70%-99 %) carotid artery stenosis.13 The sensitivity and specificity of CE-MRA for ostial steno­sis also is very high.
The intimal flap of CCA dissection can be visualized with either CT or MRA. For a patient with poor or uncertain hemodynamic sta­bility, CT is the preferred modality because it is more rapid and has better patient monitoring capabilities. However, MR is preferred for stable patients, since it does not impart ionizing radiation to the thyroid gland for either initial or follow-up studies.
For vertebral artery dissection, MRA can detect the level of ste­nosis or obstruction and distinguish residual flow from intramural hematoma.
14
FIGURE 134 Maximum-intensity projection (MIP) image shows normal thoracic aorta and origin and course of supraaortic vessels.
Thoracic Aorta and Its Branches
Both CT and MRI provide comprehensive imaging of the aorta. Single breath-hold 3D CE-MRA imaging of the thoracic aorta (Fig. 13-4) is typically performed with electrocardiographic (ECG) gating to eliminate pulsation artifact.
AORTIC DISSECTION
A properly performed modern MRA is 100% sensitive for dissection and intramural hematoma. and extent of the intimal tear, and the relationship of the tear to the branch vessels are readily tracked in multiple planes. Cine images of the proximal aorta can identify aortic regurgitation complicating type A dissection (
Fig. 13-5). Delayed phase images allow identifica-
tion of intramural hematoma, ulceration, and complications includ­ing rupture. On T1-weighted SE sequences, intramural hematoma is seen as a concentric thickening of the wall, with increased intramu­ral signal intensity.
16
Inflammatory changes are seen as arterial wall thickening and enhancement. When used with clinical parameters and other testing (e.g., blood pressure changes), MRA is useful to tri­age patients into either medical or surgical management options.
15
The true and false lumens, location
17
171
CH 13
MAgnETiC REsonAnCE iMAging
THORACIC AORTIC ANEURYSM
Aneurysm imaging should consider slow blood flow through the lesion. Magnetic resonance angiography can demonstrate the location and size of an aneurysm, presence of a mural throm­bus, and the relationship of the aneurysm to the branch vessels. Because MIP images from 3D CE-MRA are designed to highlight the lumen of the aorta, the source (including noncontrast) images should be evaluated to determine the extent of mural thrombus and the actual aneurysm size for accurate measurements. Time­resolved imaging shows delayed enhancement from slower flow.18 Patients with a clinically suspected dissection and contraindica­tion to gadolinium (e.g., severe allergy, acute renal failure) can undergo DIR imaging or often noncontrast techniques to evaluate the thoracic aorta.
19
ARCH VESSEL DISEASE
Occlusive disease of the great vessels is usually due to atheroscle-
FIGURE 133 Carotid artery disease. Left internal carotid artery (ICA) dissection resulted in a thrombotic occlusion of proximal vessel. Internal carotid artery reconstitutes more distally.
rosis. Vasculitis, fibromuscular dysplasia (FMD), and radiation arte­riopathy also can cause branch vessel stenoses. CE-MRA is an established tool for rapid and accurate definition of brachiocephalic