234 Stents, covered stents, stent–grafts
it until aer the stent is deployed. Perform a road
map with the patient having taken a half breath
in (not a deep sea diver breath that can change all
the positions and the landmarks). Advance the
stent into the lesion, have the patient take the exact
same half breath in, and hold it as he/she had done
during the angiographic run. During the breath
hold, the road map should match up and the stent
is deployed. Aer the stent is in place across the
lesion, but before it is deployed, if there is any doubt
about correct placement, pu contrast through the
sheath to reconrm the position or do another
road map. Aer the stent is in the lesion, additional
contrast runs may be performed, but the ow of
contrast around the stent and through the lesion
may be limited. e trailing end of the stent will
be visualized, and this is oen the most important
part of the stent, especially in aortic branch origin
lesions (e.g., innominate, subclavian, visceral, and
renal arteries). Sometimes, this process is repeated
a couple of times until the operator is satised with
the position for stent placement. Have a contin-
gency plan for what to do if the stent is placed too
far one way or the other with respect to the lesion.
Guidewire control must be maintained across
the stent until the reconstruction is complete. e
operator must be aware of and manage guidewire
bias. In the case of branch artery origin lesions, tra-
versal of the wire from the aorta into the branch
is accompanied by wire bias, which will direct the
wire toward one side of the lumen of the artery.
Anything that is passed over the wire, such as a
subsequent balloon or the next stent, will tend to
catch on the initial stent because of this wire bias.
If additional stents are required, the sheath is
advanced into the appropriate position. is can
sometimes be done over the wire alone and some-
times it is necessary to advance the wire over the
dilator. If the latter maneuver is selected, the opera-
tor must be careful to avoid damaging or dislodging
anything with the dilator tip. If numerous overlap-
ping stents are required, the distal stent is placed
rst and built proximally to create a “telescope”
eect (Figure19.4). If dilation to a larger diameter
is required, the deployment balloon is exchanged.
Aballoon-expandable stent does not taper well but
it can be dilated to a slightly larger size on one end if
necessary to are the end of the stent and attempt to
match vessel size and taper. is is especially true of
the newer balloon-expandable stents that are con-
structed of lighter and less rigid metal than the ear-
lier generation of stainless steel stents.
A completion arteriogram is obtained by plac-
ing the tip of the sheath at the distal end of the stent
and injecting contrast so that it reuxes through
the area of stent placement. Another option is to
place a 4-Fr or 5-Fr straight catheter through the
sheath, over the guidewire, and position the tip
of the catheter at a location proximal to the stent.
A 4-Fr or 5-Fr ush catheter may also be placed
in the same manner and positioned upstream from
the stent site.
A B C
Fig. 19.4 Telescope effect using balloon-expandable stents. (A) A long iliac artery stenosis requires
more than one stent for coverage. (B) The distal stent is placed first. (C) The second stent overlaps
the first and fits inside to create a telescope effect. This technique avoids a prominent metal edge
protruding into the flow stream. The proximal end of the proximal or second stent may also be flared
slightly with secondary balloon dilation.
https://t.me/med1917

Placement technique for self-expanding stents 235
PLACEMENT TECHNIQUE FOR
SELF-EXPANDING STENTS
Standard self-expanding nitinol stents must be
oversized by 1–2 mm so that they exert continu-
ous outward radial force at the site of deployment.
Placing a stent that is too small in diameter must be
avoided since these stents cannot be dilated beyond
their nominal diameter in the same manner that
a balloon-expandable stent can be overexpanded.
If there is doubt about the appropriate diameter
stent to use, consider balloon angioplasty rst with
an evaluation of the inated balloon prole. If still
in doubt, use a stent that is a bit bigger in diam-
eter. is is the opposite of balloon-expandable
stents. Placement of a self-expanding stent is per-
formed by withdrawing the covering catheter layer
that encloses the stent on a prepackaged delivery
catheter. e prepackaged stent delivery catheter is
placed through a 5-Fr or 6-Fr sheath in most cases
for peripheral arterial usage, as recommended
by the manufacturer. e largest self-expanding
stents used for aortic or venous applications may
require an 8-Fr or 10-Fr sheath. Stent length choice
varies signicantly from one stent type to another.
Self-expanding stents are manufactured in mul-
tiple lengths, from 20 to 250 mm in fully expanded
form. Self-expanding nitinol stents, which are
commonly used in the noncoronary arteries, usu-
ally do not shorten much on placement, although
there are dierent designs. It makes sense to get
used to a couple of products that you will routinely
use in your practice and really understand the
extent to which the trailing end of the stent tends
to foreshorten. Self-expanding stents are generally
simple to place and they adapt to tortuosity and
manage longer lesions well.
Woven nitinol stents (Supera
®
and Wallstent
®
)
experience signicant changes in length with
deployment. e current application for the
Supera
®
is in the femoropopliteal vessel segment,
whereas the Wallstent
®
may be used in multiple
arterial and venous vascular beds. e Supera
®
stent is deployed by pushing it out of its delivery
catheter. e artery must be well prepared for the
stent to reach its nominal diameter. e Supera
®
stent is highly crush resistant but exerts no chronic
outward force. It is very eective at its nominal
diameter, but is quite ineective if it is deployed in
a crimped manner. e ultimate length of the stent
when deployed depends on the degree to which
the operator packs the stent into the lesion dur-
ing deployment. e Wallstent
®
has quite a bit of
chronic outward force and the nal length of the
stent depends on the nal overall resting diameter.
e constrained length of the Wallstent
®
(in the
package) is longer than the deployed length and,
aer placement, is constrained by the artery. If the
Wallstent
®
is opened on a tabletop and is completely
unconstrained, it is even shorter. For example,
a 10×42 Wallstent
®
(10-mm diameter × 42-mm
length) placed in a 9-mm iliac artery is 50–52 mm
long aer deployment. e Wallstent
®
in practice
is not intended to be completely expanded, since
it is oversized for the artery into which it is placed.
e self-expanding stent catheter is removed
from its package, ushed, wiped with heparin–
saline solution and advanced over the guidewire
(Figure 19.5). Because the delivery catheter and
the stent are somewhat exible, it can usually be
passed over the aortic bifurcation or into branches
without diculty. Standard self-expanding niti-
nol stents are marked by radiopaque markers on
their proximal and distal ends, which are observed
using uoroscopy. Woven nitinol stents typically
do not have markers. To deploy a standard nitinol
stent or a Wallstent
®
, the metal pushing rod is held
steady and the valve body is withdrawn, which
removes the covering membrane and permits the
stent to expand on its own. e view on uoros-
copy is magnied and continuous because it is
easy to move the stent with minimal force and it
makes sense to carefully adjust the stent location
asit begins to open. As the metal pushing rod is
held stationary and the valve body is withdrawn,
the proximal end of the stent begins to expand.
Asthe nitinol warms up, it becomes more rm.
e position of the stent is continually assessed
as it is being deployed. A road map can be used
to assist in placement. Self-expanding stents are
more likely to be used for lesions in conduit arter-
ies (e.g.,iliac artery, SFA), which may be longer
and have some degree of tortuosity (e.g., carotid
artery, iliac artery). Self-expanding stents have
the disadvantage that only the leading end of the
stent, usually the end toward the tip of the cath-
eter, can be placed with a high degree of accuracy.
e back end, trailing end, or end of the stent
that is closer to the hub end of the catheter will
land with much less placement accuracy than
the leading end of the stent. Somewhere in the
course of self-expanding stent deployment, but
https://t.me/med1917

236 Stents, covered stents, stent–grafts
A
B
C
D
Fig. 19.5 Placement technique for self-expanding stents. (A) The guidewire is placed across the
lesion. (B) The stent delivery catheter is placed over the guidewire so that the proximal marker on the
stent delivery catheter is proximal to the lesion. The metal push rod is held stationary while the valve
body is slowly withdrawn. (C) The position of the stent is continuously monitored during deployment.
A magnified view may help perfectly place the proximal end of the stent. As the stent is about to
open, it can be withdrawn slightly to reposition it, but not advanced. As the valve body is withdrawn
along the length of the pushing rod, the covering sheath or membrane that constrains the stent is
removed and the stent spontaneously expands. (D) After stent deployment, the delivery catheter is
removed. Poststent placement balloon angioplasty isperformed.
https://t.me/med1917

Placement technique for covered stents 237
before there is full engagement of the stent with
the artery wall, the stent can be moved and its
position readjusted. When the leading end of the
stent starts to open it “blossoms”. Once that blos-
som becomes big enough to touch the wall of the
artery, the stent cannot be advanced, but usually
can still be withdrawn. Most self-expanding niti-
nol stents can deploy about 15% to 20% of their
length, or 1 cm (whichever is less), and it still be
possible to withdraw the whole stent to readjust
its position. Ifthere is too much adherence of the
stent to the wall already in play, when the catheter
is withdrawn, the stent just hugs and hangs on to
the wall and the stent deploys. If the stent is moved
aer there is a lot of contact between the artery
wall and the stent, there is a risk that the end of the
stent will be mangled.
ere is a reconstrainable Wallstent
®
that can
be covered up again by its covering membrane,
provided the release process has stopped short
of a certain point that is marked on the delivery
catheter. e stent can then be moved to a dier-
ent place and deployed. e Wallstent
®
is a closed
cell structure. is permits a relatively smooth
outer surface without any “V”-shaped stent joints
extruding beyond the prole of the open or par-
tially open stent. is facilitates recovery of the
outer surface of the stent if desired. is recaptur-
ability will probably not be possible with open cell
nitinol self-expanding stents in the near future.
e design allows the stent to spring all the way
open when one end is deployed and one end is
still constrained in the delivery catheter. In other
words, the open cell stents contour well and they
are able to go from a very small diameter to a large
diameter over a very short longitudinal distance.
is property is needed for the management of tor-
tuous or tapered arteries, but it also creates rough
edges where stent joints are under stress and pro-
trude into the lumen.
Most of the self-expanding stents deploy in a
similar manner to each other: holding a pushing
rod in a stationary position while withdrawing
the covering membrane and allowing the stent to
expand. Some have ratcheted handles or turnwheel
knobs to activate for better control, possibly accu-
racy, but it is the same underlying mechanism.
Take the slack out of the guidewire and cath-
eter. Be aware that any withdrawal of the catheter
for placement accuracy will also tend to pull the
sheath out. Built-up energy in the stent delivery
catheter, when released, tends to make the stent
pop forward a bit (toward the tip end) during
deployment. Aer stent deployment, the delivery
catheter is removed and balloon angioplasty of the
length of the stent is performed, especially in sec-
tions where there is residual crimping of the stent
by the lesion. It is sometimes dicult to assess
whether the stent is fully expanded. Keep the
balloon within the stent if injury is to be avoided
to neighboring artery segments. When balloon
angioplasty is to be used within the Wallstent
®
, the
central part of the stent is dilated rst. e ends of
the stent are dilated last because the tines of the
stent may rupture the balloon.
e guidewire is maintained across the stented
segment until aer satisfactory completion stud-
ies are performed. Completion arteriography is
performed in the same manner as with balloon-
expandable stents. If guidewire control of the stent
is relinquished prematurely, advance a J-tip or very
oppy-tip guidewire back through the stent using
a small eld of view for magnication. is is done
to avoid passing the guidewire through the struts
of the stent. is is easy to do, especially prior to
postdeployment balloon angioplasty. If you have
to cross a fresh stent and you are not sure if you
are under its tines, pass a straight or simple curve
catheter over the guidewire. If the guidewire goes
under one of the stent struts, the catheter will hang
up instead of passing freely.
PLACEMENT TECHNIQUE FOR
COVERED STENTS
Choosing the appropriate stent diameter is impor-
tant no matter what kind of stent is planned, but
with self-expanding covered stents the diameter
choice is even more important. is is because
signicant oversizing of the stent results in place-
ment of a larger amount of material, which may
overwhelm the available lumen. In addition, when
treating longer segments with self-expanding
stents, there may be some overlap between dier-
ent stents. In the overlapped segment the amount
of material is even larger. When there is a change
in diameter or tapering of the artery, typically
the smallest diameter covered stents are placed
rst and the larger diameter self-expanding cov-
ered stents are placed subsequently. is permits
the development of a friction seal. When treating
occlusive disease, consider minimal overlap of the
https://t.me/med1917

238 Stents, covered stents, stent–grafts
covered stents in order to prevent build-up of mate-
rial. When treating aneurysmal disease, such as in
the popliteal artery, more overlap will be required
to prevent the stents from becoming dislodged or
disjointed. When selecting a balloon-expandable
covered stent, the stent is sized to the artery. In that
setting the operator has the luxury of a slight over-
expansion of the balloon if additional diameter is
required.
ere is typically a larger sheath required for
covered stents than for bare metal stents. In order
to make more room for delivery through the
sheath, many covered stents are delivered over
smaller diameter guidewires, such as 0.018-inch
diameter wires. e deliverability is partly
diminished with covered stents due to the larger
amount of material that usually ts very tightly
within the sheath. Aer the constrained covered
stent has passed through the sheath, the outside
surface of the covered stent is typically irregular,
and the device itself on the catheter is usually
slightly bulkier than a bare metal stent. All these
aspects aect the ability to deliver the covered
stent to its desired location. If the covered stent is
being delivered into a tortuous area, such as when
being used as a renal or visceral snorkel, it will
likely not pass bareback. In this case, the sheath
is placed where the covered stent is desired, the
stent is advanced into location, and then the
sheath is withdrawn.
Once the device is in the appropriate location,
the placement depends to some extent on the type
of covered stent. For balloon-expandable cov-
ered stents, the delivery is very similar to other
balloon-expandable stents. e device is placed
exactly where desired and the balloon is inated.
ere is typically minimal foreshortening. With
self-expanding covered stents, there may be some
adjustment of position during deployment of the
leading edge of the stent, and there is usually mini-
mal foreshortening.
Aer placement of self-expanding covered
stents, the operator performs aggressive balloon
angioplasty of the stent to make sure it is fully
expanded. e outward radial force is typically low
and the potential for the lumen to be compromised
is increased. With balloon-expandable covered
stands, additional balloon dilation is usually not
required, but sometimes, depending on the loca-
tion, the operator decides to dilate the end of the
stent to are it slightly.
Strategy: What to consider when
selecting a stent
1. Can you visualize the stent in this position?
2. What are the length and diameter require-
ments of the diseased segment?
3. What is the location of the lesion and what
type of lesion is it?
4. Are there delivery restrictions posed by dis-
eased access arteries, sheath size, vessel tortu-
osity, working room, or distance to the lesion?
5. Can the number of required stents be
minimized?
6. Is the stent in your inventory?
WHICH STENT FOR WHICH
LESION?
Table 19.3 oers a practical comparison of the
working properties of dierent stent categories. In
general, orice lesions and those that are heavily
calcied are best treated with balloon-expandable
stents. Lesions located in exible or tortuous arter-
ies, or that are more than several centimeters in
length, are usually treated with self-expanding
stents. Either balloon-expanding or self- expanding
stents may be used in the aorta (Table19.4). Short
focal lesions may be treated with balloon-expand-
able stents. Placement is precise and the single stent
is less expensive. Self-expanding stents are well
suited to longer aortic lesions (longer than 2–3cm).
Placement of a single longer self-expanding stent
is usually simpler, faster, and less expensive than
placing multiple balloon-expandable stents. Just be
careful to ensure that the trailing end of the stent
clears the aortic bifurcation and opens fully within
the aorta. e other advantage of self-expanding
stents in the aorta is when there is a heavily calci-
ed, shelf-like, or iceberg-like lesion, the stent can
be laid in and the lumen gradually enlarged with
successively enlarging angioplasty balloons. is is
done to avoid rupture.
Lesions in the aortic bifurcation are treated
by “raising” the aortic ow divider with balloon-
expandable stents. Precise, “kissing,” and side-
to-side placement is possible with self-expanding
stents, but is more challenging and may not be
adequate to manage heavily calcied plaque at the
aortic bifurcation. Aortic bifurcation plaque usu-
ally extends into the common iliac artery orices.
Self-expanding stents do not have as much hoop
https://t.me/med1917

Which stent for which lesion? 239
Table 19.3 Practical stent comparison: working qualities of balloon-expandable, self-expanding, and woven nitinol stents
Working
quality
Balloon-
expandable
Self-expanding
(nitinol)
Woven nitinol
(Wallstent
®
)
Woven nitinol
(Supera
®
) Advice
Method of
expansion
Balloon-
expandable
Self-expanding Self-expanding Self-expanding Requirement for balloon expansion to
deploy makes balloon-expandable
stent placement a little more
complicated, but self-expanding
stents always require postexpansion
balloon dilation
Maximum
length
>5 cm 25 cm >9.4 cm Approximately 15
cm
Balloon-expandable stent length
limited; lesions more than 3 or 4 cm
need more than one stent; adds
complexity, time, and cost to the
case. Self-expanding stents provide
longer length
Length changes
during
placement
Shortens by 5% to
25%, depending
on final diameter
Less than 8%, does
not depend on
final diameter
Shortens by more
than 30% of
constrained length
Significant; depends
on how tightly the
stent is packed
Final resting Wallstent
®
length can be
difficult to predict; nitinol stents
have minimal shortening
Hoop strength High Low to medium Low High Balloon-expandable stent is a better
choice for orifice lesions and heavily
calcified arteries
Flexibility None High Moderate High Self-expanding stents are better
choices for tortuous or tapered
arteries
Contourability None High Moderate Poor; does not
contour to
diameter changes
(Continued)
https://t.me/med1917

240 Stents, covered stents, stent–grafts
Table 19.3 (Continued) Practical stent comparison: working qualities of balloon-expandable, self-expanding, and woven nitinol stents
Working
quality
Balloon-
expandable
Self-expanding
(nitinol)
Woven nitinol
(Wallstent
®
)
Woven nitinol
(Supera
®
) Advice
Maximum
diameter
>24 mm 14 mm 24 mm 7 mm Self-expanding stents a better choice
for vessels >12 mm in diameter;
balloon-expandable stents can be
pushed to larger sizes, but with
severe foreshortening
Precision of
placement
Precise Precise on one
end
Precise only on one
end
Only somewhat
precise and only
on one end
Proximal end (first end to be
deployed) of self-expanding stent
can be placed precisely and moved
a bit prior to full deployment;
location of second/distal end of
stent is not precise
Delivery sheath 6–9 Fr 6–8 Fr 6–9 Fr 6-Fr or 7-Fr Balloon-expandable stent requires a
long sheath to be placed through
the lesion, which adds risk of
instrumenting lesion and creates
problems for remote delivery.
Self-expanding stents are on
delivery catheters
Biohazard Sharp edges;
cannot be
clamped; can
tear balloon
No sharp edges;
can be clamped
in emergency
Loose wire ends are
sharp; can be
clamped in
emergency
Will not expand to
nominal diameter
if lumen is not
adequate
In an emergency, self-expanding
stents can be clamped with a
shodded clamp
https://t.me/med1917

How to select the best stent for the job 241
strength, which is desirable for the aortic bifurca-
tion and other orice lesions.
Routine common and external iliac artery
lesions may be treated with either self-expanding
or balloon-expandable stents. Focal lesions are
treated with balloon-expandable stents and self-
expanding stents are used for longer lesions and
those located in tortuous arteries or arteries with
signicant taper. Distal iliac artery lesions that
are close to the groin should be treated with self-
expanding stents to maintain exibility.
Self-expanding stents are better for stenting in
exible arteries, such as the supercial femoral,
popliteal, and distal subclavian arteries. Lesions
in an aortic branch orice, such as the proximal
innominate, common carotid, subclavian, vis-
ceral, or renal artery, are best treated with balloon-
expandable stents. e combination of placement
accuracy and hoop strength is better in this setting.
HOW TO SELECT THE BEST STENT
FOR THE JOB
Although there are multiple considerations when
selecting a stent for a given case, there is substantial
overlap in the capabil ities of the various stents . Most
specialists develop a short list of a couple of favor-
ites in each stent category, balloon- expandable and
self-expanding, in each vascularbed. Onetrend of
note is that self-expanding stents are being used
more widely than before. ey are becoming more
versatile, better engineered, provide more choices,
and there is better data on results. What follows is
a discussion of some of the issues that drive stent
preferences in clinical practice.
Most practices have at least some restriction
on the variety of stents stocked and the number
of dierent stent diameters and lengths avail-
able. Specialists with a limited inventory will ask
the question sooner, but everyone must face the
availability issue at some point. Work with what
is available or plan ahead well enough so that spe-
cic items are anticipated and ordered. e devel-
opment of various sizes, lengths, designs, proles,
and qualities of stents has been signicant over the
past few years.
e single most important thing to do when
selecting a stent is to visualize the stent in the
intended location. Will it expand to oppose the
wall of the artery? Can it handle the tortuosity
and/or diameter changes? Does it have the hoop
strength to stand up to the amount of calcication
present in the lesion? Will one end of the stent be
oating free in a segment of poststenotic dilation?
Will it cover the segment of artery most in need
of treatment? Can the stent be passed safely to the
target site? Will it cover or impinge on an impor-
tant branch and, if so, is that acceptable? Can the
Table 19.4 Some examples of which stent to use
Angioplasty site Lesion type Stent to use
Reason
Aorta Focal lesion Balloon-expandable Length match
Bulky lesion Self-expanding Gradual expansion, avoid rupture
Long (>2–3 cm) Self-expanding Fewer stents required
Aortic bifurcation Iliac origin Balloon-expandable Hoop strength
Iliac Focal Balloon-expandable Length match; simple
Tortuous Self-expanding Flexible
Long (especially
external iliac)
Self-expanding Fewer stents required; handles
tortuosity well
Superior femoral/
popliteal
Long Self-expanding Fewer stents required
Across joint Self-expanding Flexible
Focal, short Self-expanding Better patency
Heavily calcified Woven nitinol Supera
®
hoop strength
Renal Orifice Balloon-expandable Hoop strength, rigid
Body of artery Either
Subclavian Tortuous Self-expanding Flexible
Orifice Balloon-expandable Hoop strength
https://t.me/med1917

242 Stents, covered stents, stent–graf ts
lesion be treated with a single stent or will multiple
implants be required?
e length and diameter requirements of the
lesion must be taken together with the type and
location of the lesion in order to come up with a
stent choice. Lesion type and location were dis-
cussed in an earlier section. It is not always appar-
ent how long a length of artery should be stented if
there is mild or even moderate disease juxtaposed
to the lesion. e operator must make an arbi-
trary decision in many cases. e diameter may
be sized the same way as for balloon angioplasty,
but the consequences of a bad guess are greater. If a
balloon-expandable stent is undersized, it may not
be securely adherent to the artery. If it is oversized
too much, the artery may be injured. One method
of dealing with this dilemma is to undersize the
balloon-expandable stent just slightly. Deploy the
stent so that it is held in place by the newly dilated
lesion. en dilate the stent again to the desired
size using a larger balloon if necessary. e diam-
eter choices for self-expanding stents provide more
leeway since the stents are oversized from 1 to 3
mm. Nevertheless, if too small a stent is used, it
may become free-oating. Required stent lengths
can be challenging. All external marking methods
have some degree of parallax.
Finally, delivery restrictions may be posed by
diseased access arteries, the sheath size of the
selected stent, vessel tortuosity, adequacy of work-
ing room, a long distance to the lesion, or branch
points between the access point and the lesion. e
risk of puncture site thrombosis increases when a
large sheath is passed through a diseased common
femoral artery. Sheath size may inuence the choice
of stent, since relatively larger diameter sel f-expand-
ing stents may be placed through 6-Fr sheaths.
Vessel tortuosity or a branch point on the way to
the lesion can be overcome with a guiding access
sheath of the appropriate length. Occasionally, a
more exible stent is required to make these turns
(e.g., over a very narrowly angled aortic bifurca-
tion) and a self-expanding stent is used instead of a
balloon-expandable stent. e distance to the lesion
is an important variable. Balloon-expandable stents
must be mounted on a balloon with an adequate
sha length, and self-expanding stents must be
delivered on a delivery catheter of adequate length.
Both balloon-expandable and self-expanding stents
are generally available on two dierent lengths of
catheter, either 80cm or 120 cm. If an 80-cm sha
is opened when a longer sha is required, it will
have to be discarded. ese challenges can usu-
ally be solved by using the balloon sha length to
estimate distance if the lesion was dilated prior to
stent placement. If not, a standard arteriographic
catheter with a bright tip may be placed over the
guidewire and advanced close to the lesion and
used to estimate the length from the access to the
lesion. Ifthe distance from the access to the lesion
is farther, there are some devices up to 150 cm in
length, but care must be taken to ensure that these
are available. If not, or if the longest devices are
inadequate, the operator must make a plan for an
alternative access closer to the treatment site.
TRICKS OF THE TRADE
Raising the flow divider with kissing
stents
A bifurcation can be reconstructed by modifying
the ow divider with kissing stents. Two stents
are placed simultaneously with the leading edge
of each stent abutting the other at a point proxi-
mal to the location of the native ow divider. e
need for kissing stents arises most commonly at
the aortic bifurcation (see Chapter 23 for a detailed
discussion of this technique). e atherosclerotic
disease that occurs at the aortic bifurcation tends
to be extensive and circumferential. With kissing
stents, the aortic bifurcation can be reconstructed.
ere is also a technique for reconstruction of
the aortic bifurcation in which a covered stent is
placed in the infrarenal aorta and covered balloon-
expandable stents are placed at the aortic bifurca-
tion in the fashion of kissing stents. is provides
a completely covered reconstruction of the aortic
bifurcation. Aer any of these techniques for rais-
ing the ow divider are performed, it is usually not
possible to deliver endovascular devices up-and-
over the aortic bifurcation.
Tapering a stent
e various self-expanding stents tend to taper
naturally with diminishing distal arterial diameter.
Standard open cell nitinol stents should be dilated
with the appropriately sized balloon. All of the self-
expanding nitinol stents exhibit variable degrees
of foreshortening. Because of this, balloon dilation
should be performed rst at the location along the
https://t.me/med1917

Tricks of the trade 243
stent where some shortening would be least desir-
able (Figure 19.6). Once the nal diameter of the
stent is xed in this location, dilation of other seg-
ments of the stent with the appropriately sized bal-
loon is performed. e Wallstent
®
tends to shorten
quite a bit with expansion. e very end is dilated
last since it may lead to rupture of the balloon.
Woven nitinol stents (Supera
®
) cannot taper and
cannot be manipulated much with balloon dilation.
e shorter, more rigid balloon-expanding
stents can also be tapered, but to a much lesser
degree (1–2 mm maximum). One end of the stent
is selectively dilated using only the shoulder of
the balloon (Figure 19.7). e operator may also
consider aring the proximal end of a balloon-
expandable stent at the origin of a renal or visceral
artery. In this case, only a very short segment of
balloon shoulder is placed in the stent.
Moving a self-expanding stent
To varying degrees, self-expanding stents can be
moved a bit to adjust deployment position. e
delivery catheter can be withdrawn or pulled back
but not advanced forward aer beginning partial
deployment. e entire deployment catheter appa-
ratus must be withdrawn in a retracted position to
move the stent (Figure 19.8). Most standard niti-
nol stents tend to move a little during initiation of
deployment. Moving the stent can be helpful in
achieving very precise placement of its proximal
end. It is not possible to move a stent aer it has
been deployed. Under magnied uoroscopy, the
stent delivery catheter tip is carefully observed.
When the proximal end of the stent begins to
expand and change shape, this is the opportunity
to adjust its position. If too much of the stent is
A
C D E
B
Fig. 19.6 Tapering a self-expanding stent. (A) The guidewire is passed through an iliac artery stenosis.
(B) The self-expanding stent is deployed across the lesion. (C) The self-expanding stent tends to taper
naturally with decreasing distal arterial diameter. (D) The distal end of the stent is ballooned to the
diameter appropriate for the external iliac artery. (E) Angioplasty is performed in the larger proximal
end of the stent, which lies in the common iliac artery, using a slightly larger balloon.